Overview of performance and improvements to fixed exit double crystal monochromators at Diamond. Andrew Dent, Physical Science Coordinator, DLS

Similar documents
Characterisation of a novel super-polished bimorph mirror

Nano Beam Position Monitor

Air Bearing Monochromator at APS 13-ID-E (GSECARS)

CESRTA Low Emittance Tuning Instrumentation: x-ray Beam Size Monitor

Experience with Insertion Device Photon Beam Position Monitors at the APS

x-ray Beam Size Monitor

Mechanical Design of an Ultra-High-Vacuum Compatible Compact Hard X-ray Monochromator with Artificial Channel-Cut Crystal Mechanism

Supplementary Information

P61 BEAMLINE. Calculations of beamline characteristics Version 2.3

Structure in out-of-focus beams of X-ray focusing mirrors: Causes and possible solutions. Fiona Rust Department of Physics, University of Bath

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL

Experience of synchrotron sources and optics modelling at Diamond Light Source

Status of the Electron Beam Transverse Diagnostics with Optical Diffraction Radiation at FLASH

Supplementary Figure 1

Nonintercepting Diagnostics for Transverse Beam Properties: from Rings to ERLs

LCLS-II-HE Instrumentation

Design of Photon Beamlines at the European XFEL

High-Power, Passively Q-switched Microlaser - Power Amplifier System

Photon Diagnostics. FLASH User Workshop 08.

LCLS project update. John Arthur. LCLS Photon Systems Manager

Sources & Beam Line Optics

VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION Suren Arutunian

SURVEY AND ALIGNMENT FOR THE SWISS LIGHT SOURCE

X-Ray Beam Size Monitor for CESRTA

Bioimaging of cells and tissues using accelerator-based sources

Hartmann wavefront sensing Beamline alignment

Operation of a Single Pass, Bunch-by-bunch x-ray Beam Size Monitor for the CESR Test Accelerator Research Program. October 3, 2012

Wir schaffen Wissen heute für morgen

Focusing X-ray beams below 50 nm using bent multilayers. O. Hignette Optics group. European Synchrotron Radiation Facility (FRANCE) Outline

Diamond X-ray Rocking Curve and Topograph Measurements at CHESS

CXI 1 micron Precision Instrument Stand

First test experiments with FMB- Oxford direct drive DCM at the Sirius beamline of Synchrotron SOLEIL

LCLS Injector Diagnostics. Henrik Loos. Diagnostics overview Transverse Beam Properties Longitudinal Beam Properties

LUSI Pulse Picker System

Adaptive Optics for LIGO

Manufacturing Metrology Team

High-Precision Positioning Mechanism Development at the Advanced Photon Source

BEAM SIZE MEASUREMENTS USING SYNCHROTRON RADIATION INTERFEROMETRY AT ALBA

Performance of the SASE3 monochromator equipped with a provisional short grating. Variable line spacing grating specifications

Fast Laser Raman Microscope RAMAN

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

Shenglan Xu. GM/CA CAT Argonne National Laboratory

Outline of the proposed JLAMP VUV/soft X-ray FEL and the challenges for the photon beamlines and optics

Introduction to High-Resolution Accelerator Alignment Using X-ray Optics

High Precision Positioning Mechanisms for a Hard X-ray Nanoprobe Instrument. Abstract

PETRA III Extension Project. Beamline P24 Chemical Crystallography. Technical Design Report

OPTICS IN MOTION. Introduction: Competing Technologies: 1 of 6 3/18/2012 6:27 PM.

Optics for next generation light sources

Orbit Stability Challenges for Storage Rings. Glenn Decker Advanced Photon Source Beam Diagnostics March 8, 2012

BL39XU Magnetic Materials

Commissioning of National Synchrotron Light Source-II (NSLS-II) Fast Orbit Feedback System

Small Angle Scattering Platform for Structural Biology

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Spatial resolution. Spatial resolution

Continuous-Wave (CW) Single-Frequency IR Laser. NPRO 125/126 Series

X-ray FEL Oscillator (XFEL-O) Gun Requirements and R&D Overview FLS2010: WG5: High Brightness Guns March 1, 2010

Scanning X-ray microscopy with a single photon counting 2D detector

X-Ray Transport, Diagnostic, & Commissioning Plans. LCLS Diagnostics and Commissioning Workshop

On-line spectrometer for FEL radiation at

RAPID: A Revolutionary Fast Low Noise Detector on Pionier

Feedback Requirements for SASE FELS. Henrik Loos, SLAC IPAC 2010, Kyoto, Japan

Wavelength Control and Locking with Sub-MHz Precision

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

SpectraPro 2150 Monochromators and Spectrographs

Measurements of Small Vertical Beamsize using a Coded Aperture at Diamond Light Source. C. Bloomer G. Rehm J.W. Flanagan

Development of scalable laser technology for EUVL applications

3 General layout of the XFEL Facility

Optimization of Beamline BL41XU for Measurement of Micro-Protein Crystal

Investigation of a Next Generation Piezo Bimorph Mirror

Towards accurate measurements with synchrotron tomography Problems and pitfalls. Robert C. Atwood. Nghia T. Vo, Michael Drakopoulos, Thomas Connolley

CSPADs: how to operate them, which performance to expect and what kind of features are available

DCS laser for Thomson scattering diagnostic applications

Workshop IGLEX Andromède & ThomX 23 June 2016, LAL Orsay. The X-line of ThomX.

FLASH II. FLASH II: a second undulator line and future test bed for FEL development.

CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

Infrared Single Shot Diagnostics for the Longitudinal. Profile of the Electron Bunches at FLASH. Disputation

Introduction to X-ray Detectors for Synchrotron Radiation Applications

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs

Optical Correlator for Image Motion Compensation in the Focal Plane of a Satellite Camera

Fast Laser Raman Microscope RAMAN

FLASH 2. FEL seminar. Charge: 0.5 nc. Juliane Rönsch-Schulenburg Overview of FLASH 2 Hamburg,

Talbot- Lau interferometry with a non- binary phase grating for non-destructive testing

Challenges of Optics for High Repetition Rate XFEL Source

Single Photon Interference Laboratory

UV/Optical/IR Astronomy Part 2: Spectroscopy

Fabrication, testing, and performance of a variable-focus x-ray compound lens

Vacuum Ultra Violet Monochromator

Vibration studies of a superconducting accelerating

The CMS Outer HCAL SiPM Upgrade.

Beam Infrared Detection with Resolution in Time

Newly Developed Experimental Facilities

MEASUREMENT OF BEAM LOSSES USING OPTICAL FIBRES AT THE AUSTRALIAN SYNCHROTRON

Thorough Small Angle X-ray Scattering analysis of the instability of liquid micro-jets in air

Usage and Setup From Actuators to Controls

Sub-ps (and sub-micrometer) developments at ELETTRA

Part 2: Second order systems: cantilever response

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

the need for an intensifier

Large area position-sensitive CVD diamond detectors for X-ray beam monitoring with extreme position resolution

SIL for improved sensitivity and spatial resolution

Transcription:

Overview of performance and improvements to fixed exit double crystal monochromators at Diamond Andrew Dent, Physical Science Coordinator, DLS

Overview Diffraction limit Geometric magnification Source movement Beamline and instrumentation movement

Positional stability Source Error ~ 2Δx Δx Positioning resolution of mono motion usually << source size Floor vibrations ~ 20 nm << source size

Angular Stability Source Sample Source Sample Intensity stability Beam size/stability Transverse coherence length Error ~ 2Δθ*d

Frequency response At the end (sample, slit, focussing optic...) Blurred image if collection time >> rate of vibration Lots of movement if collection time ~ rate of vibration Collection time < rate of vibration but beam position is somewhere within +/- 4σ vibration ) Two frequency requirements: Slow --- drifts Fast --- vibrations

Example of impact on DLS beamline I04 Example Compound Refractive lens to focus source Optics design value - 0.5 µm at sample Actual focus = 2.0 µm (averaged) Monochromators vibrations produce effective source size x4 larger than actual

Requirements (Past) Instability < %10 beam size 2nd Gen source, mono typically at 20m, Source size 100 µm (rms) For 1:1 focusing 10% movement corresponds ~ 250 nrad rms vertically For DLS, mono typically at 30m, Source size 123 x 6.8 µm (rms) For 1:1 focusing 10% movement corresponds ~ 11 nrad rms vertically

Requirements (now and future) Instability < 2% beam size, coupling reduced from 1% to 0.3% (8 pm V emittance) Mono typically at 30m, Source size 123x3.7 µm (rms) For 1:1 focusing 2% movement corresponds ~ 1 nrad rms vertically. Other sources already proven 1 pm V emittance Much more relaxed in the horizontal direction, but with new Multi-achromat sources and upgrades, FEL s; will become more important Strongly depends on beamline design and requirements

Monochromator performance Performance of cryo monochromators Source Vibration (rms) Type I18/I22 450 nrad* Vertical bounce-up MX (I04) 350 nrad Vertical bouncedown I11 ~500 nrad Vertical bounce-up C Thomas et al 2012 JINST 7 P01014 *The initial value was higher than this, but simple clamping of pipes and other minor changes helped a lot. A full redesign was required to reduce further, see A. Peach talk on crystal cage update

Monochromator performance Best of (some) of the rest. Source Australian Light Source Vibration (vertical) Estimate (rms) Comment 0.5 µm @ 20m <20 nrad Horizontal DCM- low power Performance water cooled monochromator Source Vibration Estimate (rms) B18 1µm rms @ 12m <40 nrad

Overview of the Measurements Motion measured with fast X-ray camera (400 Hz) Assumed no contribution from mirrors DCM motion calculated from geometrical optics Source motion measured using BPM s sampled at 10 KHz D DCM D M Ds source DCM Focussing mirror Motion amplitude: 1. Centroid motion recorded with camera images 2. FFT of the centroid 3. Cumulative integral of the FFT Fast Camera at sample position

Photon Beam Movement on I15 Source motion in unfocussed mode

Photon Beam Movement on I15 Source motion in focussed mode Modulation 71Hz Bloomer et al J Phys Conf Ser 425 (2013), 042010

Photon Beam Movement on I11 Bloomer et al J Phys Conf Ser 425 (2013), 042010

Long Term Stability on I11 and B16 Depending on crystal cage mass, can reject vibrations to ~ 50 Hz

Geometric Calculations source D DCM D M D s DCM Vert. Focussing mirror Fast Camera at sample position α: angular motion of the DCM inducing 1µm vertical displacement at sample; no focusing α = 1/ 2( DS DDCM ) General beamline layout above, but it is different for I10, I11, I16 For each case the layout as given on beamline web pages has been used to calculate a geometrical factor α Focusing: M = DS D D M M α = 1 2D DCM M

Geometric Values Beamlines D DCM (m) D M (m) D s (m) α (nrad / µm) I07 25 28 47 30 I04 28 33 40 84 I10 (PGM) - - - 286 J04 (Horz) 24 38 44 132 I15 25 29 47 23/32 I11 27-47 25 I16 DCM now channel-cut with additional focussing: example ID01 A. Diaz et al. JSR 17, (2010), 299 I24 33.6 39.2, 45.7 46.4 590 B18 22 25 37.5 45

Vertical beam stability for DLS beamlines Beamline (date) I07 (08/2010) I04 (07/2010) I04 (02/2013) I10 (09/2012) Vert. r.m.s Beam size (µm) 80 23 27 45 Resona nce (Hz) 24.8 40 43 43 39 34 25 21 27 73 1/f noise motion Amplitude DCM angular motion (nrad) Total motion : @200Hz (µm) (% σ V ) (µm) (% σ V ) 1.7 3.5 1.5 0.8 0.3 1.5 1.1 0.06 0.5 0.4 2.1 4.3 1.8 3.4 1.3 6.5 4.8 0.2 1.8 1.5 100 200 90 130 50 250 180 3.6 30.7 24.6 - - - DCM Total angular motion (nrad) 7.9 9.8 240 3.8 16.5 320 1.6 5.9 134 Hor. : 0.35 Vert. : 0.4 0.9 100 114

Vertical beam stability for DLS beamlines Beamline (date) J04 (02/2013) (hor.) I15 (03/2013) I11 (03/2012) Vert. r.m.s Beam size (µm) 24 37 (focusse d) 120 (slit) Resona nce (Hz) 22 24 71 1.6 22 24 104 65 43 24 15 Amplitude DCM angular motion (nrad) Total motion : @200Hz (µm) (% σ V ) (µm) (% σ V ) 0.04 0.03 0.16 0.2 0.2 0.1 1.0 2 2 2.4 2.6 0.16 0.12 0.66 0.54 0.54 0.27 2.7-5 3 19 6.7 6.7 3.3 33 50 50 60 65 DCM Total angular motion (nrad) 0.46 1.9 61 2.6 7.0 122 19-475

Vertical beam stability for DLS beamlines Beamline (date) Vert. r.m.s Beam size (µm) Resona nce (Hz) Amplitude DCM angular motion (nrad) Total motion : @200Hz (µm) (% σ V ) (µm) (% σ V ) DCM Total angular motion (nrad) I16 (04/2007) 60 68 48 31 19 4.7 1.4 0.7 0.9 7.8 2.3 1.1 1.5-8.7 14.5 - I24 (12/2012) 4.4 21 39 0.7 0.11 15 2.5 330 47 1.0 23 590 I24 (12/2012) piezo off 4.4 21 39 1/f noise 0.1 0.15 2.5 3.4 47 51 0.6 13 355 B18 (03/2010) 60 50 19 1/f noise 0.02 0.02 0.03 0.03 1 1 0.96 mostly 1/f 1.6 43

Monochromator improvements I18/I22 Improved crystal cage: Andy Peach I09, I23 DLS built DCM using DC motor, air bearing: Jon Kelly Source Vibration (rms) Type I18/I22 80 nrad Vertical bounce-up I09 78-49 nrad* Vertical bounce-up I23 49-27 nrad* Vertical bounce-up *Vibration data was measured in commissioning and under different conditions, hence variation

B16 Test Beamline Flexible & versatile to enable wide range of experiments Large energy range (4 kev 45 kev) Several operational modes: mono, white, micro-focused, Range of beam sizes : 1 micron to 100 mm Essentially a general purpose beamline Experimental Hutch Optics: DCM Toroidal Mirror DMM - CRLs

B16 Experiments Hutch High Spatial Resolution Detector <0.5µm ideal resolution 4008x2672 pixels, 44dia 2x to 40x objectives 16 bit dynamic range several scintillators Detectors PCO4000 & PCO.edge high resolution Pilatus 300k area detector Image Star 9000 :135mm dia CCD Merlin (medipix based) VORTEX spectroscopy, APD, X-ray eye, PIPS diode

Thermal Issues On I18 at long acquisition times (2-8s) get better data by reducing the heat load on the monochromator. Thermal load on I20 wiggler beamline not being well managed. Measurements carried out on B16: Simon Alcock talk Optimisation of Monochromator crystal configurations for increasing Synchrotron Powers (PhD program) Dr Peter Docker Diamond Light Source, Campus, Didcot, Oxon OX11 0DE Professor Mike Ward Birmingham City University B4 7XG

Conclusions Scope for improvement across Diamond beamlines New in-house designs and retrofits I09/I23/I20/I13/I22/I18 Need better interaction and collaboration with supplier If we can get x6 improvements, so can they. (precision metrology lab and B16): Simon Alcock Licensing our knowledge or defining preferred designs Mono tenders still specification based Leaving the supplier to come up with solutions?? Feedback control - A standard approach or hardware solution?

Acknowledgements Paul Quinn Cyrille Thomas and Chris Bloomer Simon Alcock Andy Peach Jon Kelly Peter Docker