Better. Hints. Microwave Counter. for making. Measurements

Similar documents
Permittivity Measurements of PC Board and Substrate Materials using the HP 4291B and HP 16453A

Hints. for making. Better. Spectrum Analyzer. Measurements. Application Note

HP 8491A/B, 8492A, 8493A/B/C, 11581A, 11582A and 11583A/C Coaxial Attenuators. Product Overview

HP 8509B Lightwave Polarization Analyzer. Product Overview. Optical polarization measurements of signal and components nm to 1600 nm

HP 8921A Cell Site Test Set. Product Note AMPS Base Station Testing

E-series E9300 Average Power Sensor Specifications

HP 8560 E-Series Spectrum Analyzers Technical Specifications

6.5-Digit Multimeter, High- Accuracy, C-Size HP E1410A Technical Specifications

Wide Range DC Current Biased Inductance Measurement. Application Note with HP 4284A Precision LCR Meter / HP 42841A Bias Current Source

product note Using Power Leveling to Control Test Port Output Power Product Note 8510XF XF Network Analyzer

Network, Spectrum, and Impedance Evaluation of Electronic Circuits and Components

product overview HP 8970 Series Noise Figure Measurement Products


Fundamentals of Microwave Frequency Counters. Application Note Electronic Counters Series

HP 8647A and 8648A/B/C/D Signal Generators Technical Specifications

ProBER 2 2 Mb/s Handheld Test Set. The handheld that takes 2 Mb/s testing beyond convention

9 Hints for Making Better Measurements Using RF Signal Generators. Application Note 1390

Introduction to BER testing of WDM systems

Agilent AN 1275 Automatic Frequency Settling Time Measurement Speeds Time-to-Market for RF Designs

Agilent 8703A Lightwave Component Analyzer Technical Specifications

Optimizing Electronic Component and Material Impedance Measurements

CDPD Conceptual Overview. Product Note

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Phase Noise Measurements

Agilent AN Applying Error Correction to Network Analyzer Measurements

HP 4286A RF LCR Meter. Technical Specifications. Specifications. Test Signal

Bird Model 7022 Statistical Power Sensor Applications and Benefits

profile New Generation Precision Impedance Analyzer for functionality and efficiency in engineering 40 Hz to 110 MHz 4294A

Meet the new E4438C ESG vector signal generator...

Agilent PSA Series Spectrum Analyzers Noise Figure Measurements Personality

Agilent 8657A/8657B Signal Generators

100 Hz to 22. HP 8566B Spectrum Analyzer. Discontinued Product Support Information Only. Outstanding Precision and Capability

Agilent 8902A Measuring Receiver

Agilent 71400C Lightwave Signal Analyzer Product Overview. Calibrated measurements of high-speed modulation, RIN, and laser linewidth

Phase Noise Measurement Personality for the Agilent ESA-E Series Spectrum Analyzers

Keysight Technologies Network Analyzer Measurements: Filter and Amplifier Examples. Application Note

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for GSM and EDGE Measurements

Base Station Installation and Maintenance

GSM/EDGE Application Firmware R&S FS-K5 for R&S FSP and R&S FSU

Agilent 81600B All-band Tunable Laser Source Technical Specifications December 2002

Agilent Highly Accurate Amplifier ACLR and ACPR Testing with the Agilent N5182A MXG Vector Signal Generator. Application Note

Advanced Test Equipment Rentals ATEC (2832)

8 Hints for Better Spectrum Analysis. Application Note

Agilent 83711B and 83712B Synthesized CW Generators

Agilent PSA Series Spectrum Analyzers Noise Figure Measurements Personality

Spectrum Analyzers R3465/3463 R3465/3463. Modulation Spectrum Analyzers. Measurement For Digital Mobile Communications

Voltage Sensors URV5-Z

Agilent EPM-P Series Single- and Dual-Channel Power Meters Agilent E9320 Family of Peak and Average Power Sensors

8 Hints for Better Spectrum Analysis. Application Note

POWER-MEASUREMENT needs can vary greatly among different

EXHIBIT 7: MEASUREMENT PROCEDURES Pursuant 47 CFR 2.947

Obtaining Flat Test Port Power with the Agilent 8360 s User Flatness Correction Feature. Product Note

INSTRUCTION SHEET WIDEBAND POWER SENSOR MODEL Copyright 2008 by Bird Electronic Corporation Instruction Book P/N Rev.

GSM Transmitter Modulation Quality Measurement Option

Power Measurement Basics

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Spectrum Analysis

Handheld 3.3GHz Spectrum Analyzer

Appnote - Realtime Spectrum Analyzer vs Spectrum Analyzer

Agilent EPM Series Power Meters

Data Sheet. Peak, CW & Average. Power Sensors. Taking performance to a new peak

Advanced Test Equipment Rentals ATEC (2832)

Vector Signal Analyzer FSE-B7 for Spectrum Analyzers FSE

Keysight Technologies Optimizing RF and Microwave Spectrum Analyzer Dynamic Range. Application Note

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

FCC PART 15C TEST REPORT FOR CERTIFICATION On Behalf of. DEI Sales Inc. dba Definitive Technology. Model Number: STUDIO SLIM SUBWOOFER

HP 4155B/4156B Semiconductor Parameter Analyzer Product Note-3 Prober Connection Guide. HP 4155B/4156B Semiconductor Parameter Analyzer

Agilent PNA Microwave Network Analyzers

HP 8901B Modulation Analyzer. HP 11722A Sensor Module. 150 khz MHz. 100 khz MHz. Technical Specifications. Four Instruments In One

Agilent 8902A Measuring Receiver Product Note

LB480A Pulse Profiling USB PowerSensor+ Data Sheet

Agilent 81980/ 81940A, Agilent 81989/ 81949A, Agilent 81944A Compact Tunable Laser Sources

Agilent E9300 Power Sensors E-Series Technical Overview

Agilent N9342C Handheld Spectrum Analyzer (HSA)

Phase Matrix, Inc. 545B 548B. Phase Matrix, Inc. EIP 545B and 548B CW Frequency Counters. Instruments You Can Count On

A Guide to Calibrating Your Spectrum Analyzer

Direct-Conversion I-Q Modulator Simulation by Andy Howard, Applications Engineer Agilent EEsof EDA

of Switzerland Analog High-Speed Products

Agilent N9342C Handheld Spectrum Analyzer (HSA)

RF Fundamentals Part 2 Spectral Analysis

LB680A Pulse Profiling USB PowerSensor+ Data Sheet

Keysight Technologies N1918A Power Analysis Manager and U2000 Series USB Power Sensors. Demo Guide

Agilent 8360B Series Synthesized Swept Signal Generators 8360L Series Synthesized Swept CW Generators

FCC PART 15C TEST REPORT FOR CERTIFICATION On Behalf of. ION Audio, LLC. Portable Karaoke PA speaker with vocal effects

Test & Calibration Benefits from a New Precision RF/Microwave Calibrator

Agilent E8267C/E8257C/E8247C PSG

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Spectrum Analysis

Keysight Technologies 8 Hints for Making Better Measurements Using RF Signal Generators. Application Note

HP 8901B Modulation Analyzer. HP 11722A Sensor Module. 150 khz MHz. 100 khz MHz. Technical Specifications. Four Instruments In One

Unprecedented wealth of signals for virtually any requirement

Agilent E8267C PSG Vector Signal Generator

Understanding the Precision Antenna, Cable, and Power Measurements on the 3550 Radio Test System

High-speed programmable attenuator MAT800

Agilent 8703B Lightwave Component Analyzer Technical Specifications. 50 MHz to GHz modulation bandwidth

Stephen Plumb National Instruments

Agilent E8247/E8257C PSG CW and Analog Signal Generators

khz to 2.7 GHz Spectrum Analyzer

World s Most Trusted Family of Handheld RF and Microwave Analyzers

Multi-function Site Passive Intermodulation Analyzer.

Spectrum Analyzers. 2394A 1 khz to 13.2 GHz Spectrum Analyzer.

Advanced Test Equipment Rentals ATEC (2832)

Impedance 50 (75 connectors via adapters)

Transcription:

This literature was published years prior to the establishment of Agilent Technologies as a company independent from Hewlett-Packard and describes products or services now available through Agilent. It may also refer to products/services no longer supported by Agilent. We regret any inconvenience caused by obsolete information. For the latest information on Agilent s test and measurement products go to: www.agilent.com/find/products Or in the U.S., call Agilent Technologies at 1-800-452-4844 (8am 8pm EST) 4 Hints for making Better Microwave Counter Measurements

4 Hints for making Better Microwave Counter Measurements With the advancement of technology, there is an increasing tendency for measurement tools to overlap each other s capabilities. Spectrum analyzers can now measure frequency (a function once reserved for counters), microwave counters can now measure power (a function once reserved for power meters), universal counters can now measure microwave frequencies (a function once reserved for microwave counters), and so on. Furthermore, this same advancement in technology in general has created unique measurement problems. This note will deal with some of these measurement issues for microwave counters. There are several types of counters that can make microwave frequency measurements: Universal counters (timer/ counters) and RF counters (frequency counters) can now extend their frequency measurement capability into the low to mid microwave range (usually to 5 GHz, and in some cases to 12.4 GHz). Continuous Wave (CW) microwave counters typically measure frequency to at least 20 GHz, and to as high as 50 GHz. Most of these counters can make power measurements as well, using the same input for both frequency and power measurements. CW microwave counter/power meters are a combination of a microwave frequency counter and a true power meter in one package. Frequency and power measurements are made using separate inputs. Power is measured with an external power sensor, yielding a more accurate measurement. Pulse and CW microwave counters measure pulsed microwave signals (typically found in military applications) as well as CW signals. Since CW microwave counters are far and away the most popular of these types of microwave counters, this note will deal primarily with them. 2

When to use a Microwave Counter versus a Spectrum Analyzer A spectrum analyzer is a powerful instrument that is considered by most to be a basic microwave tool. However, using a spectrum analyzer to make microwave frequency measurements may not be the optimum solution. Purchasing a microwave counter is definitely the more economical decision. Typically, a 20 GHz spectrum analyzer is at least three times the price of a 20 GHz microwave counter: a high premium to pay if the requirement is only for measuring the frequency of microwave signals. HINT 1 A microwave counter can offer several other advantages as well. If high resolution frequency measurements and/or rapid measurement times are required, a microwave counter will always yield a 1 Hz resolution in 1 second. Not all spectrum analyzers will. Of course a counter is easier to use, especially for novices. The size and weight of a counter are generally less than those of a spectrum analyzer, making counters a truly fieldportable instrument when an internal battery is added. The HP 53150 Series Microwave Counters are good examples of all of these points. 3

HINT 2 Measuring Digitally Modulated Signals Measuring the carrier or center frequency of a modulated signal with a microwave counter has always been a difficult task. If frequency modulation (FM) was used, the accuracy of the measurement was related to both the modulating signal s symmetry and the linearity of the modulator, neither of which were typically well known prior to making the measurement. With amplitude modulation (AM), the signal always needed to be within the counter s sensitivity, making pulsemodulated or single-sideband signals almost impossible to measure without special techniques, often involving delay lines or envelope detection. These difficulties resulted in a classic rule-of-thumb: either the carrier s modulation must be turned off (thereby interrupting service) or the signal must be measured prior to modulation. This rule-of-thumb still has merit with digitally modulated signals. Modern digital modulation is a combination of AM, phase modulation, and often pulse modulation. The latter is employed in Time Division Multiple Access (TDMA) plans where each user is assigned not only a specific channel frequency, but also a specific time slot during which transmission is allowed. The most typical implementations of this TDMA technique are the NADC (US), PHS/ PDC (Japan) and GSM (worldwide) standards which are applied to cellular/pcs and wireless local loop (WLL) implementations for both voice and data. In addition, the signal spectrum is often deliberately spread using either frequency hopping (FHSS) or a psuedorandom sequence (CDMA). The former is used in several wireless local area network (WLAN) implementations and the latter is the heart of the various CDMA networks including WLAN, cellular/pcs, and low earth orbit satellite (Iridium and Globalstar). Digitally-modulated signals represent special problems for measuring the carrier frequency with any degree of accuracy. Figure 1 (a, b, and c) shows the typical spectrums of three of these signals: PHS, NADC and GSM. Each one of these standards uses different data rates, modulation schemes and spectrum shaping filters that result in different degrees of uncertainty when measuring the signal. Since a microwave counter displays an average frequency during its measurement (gate) time, longer gate times will result in better accuracy if the transmitted data is truly random. If it is not (due to either the data itself or an imperfect modulator), the spectrum is skewed to one side or another resulting in an unpredictable reading bias. This bias would, at first glance, seem like an insurmountable measurement problem. However, it can be used to actually improve the measurement accuracy under certain circumstances. 4

REF.0 dbm hp 10 db/ ATTEN 10 db CENTER 900.000 MHz CENTER 900.000 MHz SPAN 2.000 MHz RES BW 100 Hz VBW 100 Hz SWP 500 sec Figure 1a. PHS typical spectrum To understand why, it is useful to consider that the modulator typically used in digital modulation schemes is a quadrature modulator. This type of modulator is essentially a single sideband, suppressed carrier modulator. When presented with an all-zero data stream, the spectrum basically becomes a single-tone (sideband) with a suppressed carrier. The offset of this sideband from the suppressed carrier is related to the data rate and the modulation scheme employed, but it is always fixed in any given digital standard. This is graphically illustrated in Figure 2 where a GSM spectrum is shown with all zeros as data. The single sideband is precisely offset one quarter of the bit rate (67.708 khz or 270.833 kb/s divided by 4) above the suppressed carrier. Thus, the exact carrier frequency can be calculated easily by measuring the sideband itself, assuming the data rate is known and the data can be set to all zeros. REF.0 dbm hp 10 db/ CENTER 900.0000 MHz ATTEN 10 db CENTER 900.0000 MHz SPAN 200.0 khz RES BW 30 Hz VBW 30 Hz SWP 500 sec REF.0 dbm hp 10 db/ CENTER 900.000 MHz ATTEN 10 db CENTER 900.000 MHz SPAN 2.000 MHz RES BW 100 Hz VBW 100 Hz SWP 500 sec Figure 1b. NADC Figure 1c. GSM 5

A modern microwave counter, like the HP 53150 Series Microwave Counters with offset capability, can perform this calculation automatically. For transmission standards which use other modulation techniques, the offset is similarly fixed and predictable as summarized in Table 1. REF.6 dbm hp 10 db/ MARKER 67 khz 1.20 db ATTEN 10 db MKR 67 khz 1.20 db Another factor to consider when employing this all zero technique is whether or not the carrier signal is also being pulse modulated or burst (TDMA standards only). The direct measurement of such a burst signal must be made with a counter capable of pulse measurements. Alternatively, the frequency can be measured before the pulse modulator by a CW microwave counter. Spread spectrum systems cannot be measured using this technique as the spreading method (CDMA or FHSS) inherently interferes with the measurement. For these systems a microwave counter will only measure the approximate center value of the assigned channel. CENTER 900.000 MHz SPAN 1.000 MHz RES BW 10 khz VBW 30 khz SWP 530 msec Table 1 Figure 2. GSM spectrum with an all zeros pattern Standard Modulation Type Data Rate All Zero Sideband GSM GMSK 270.833 kb/s Data Rate/4 above (+67.708 khz) NADC π/4 DQPSK 48.608 kb/s Data Rate/16 above (+3.038 khz) PHS π/4 DQPSK 384 kb/s Data Rate/16 above (+24 khz) PDC π/4 DQPSK 42 kb/s Data Rate/16 above (+2.625 khz) DECT GFSK 1152 kb/s Data Rate/4 below ( 288 khz) TETRA π/4 DQPSK 36 kb/s Data Rate/16 above (+2.250 khz) 6

Making Sense of Power Measurements with a Microwave Counter Most CW microwave counters have the ability to measure the input signal s microwave power as well as its frequency. Typically, a Schottky diode is used as the power sensing device. This diode is part of the sampler downconversion assembly, which converts the microwave signal to an intermediate frequency that the counter actually counts. The effective match (SWR) is limited by the entire complex sampler structure. As a result, the typical match of a microwave counter is inferior to that of a power meter sensor, which has only a single diode or thermocouple device in its microwave structure. In addition, as the power measurement and downconversion of a microwave counter are done virtually at the front panel input connector, a cable must be used to bring the signal to the instrument s input. This cable adds its loss and match characteristics to the measurement, a factor typically not part of power measurements using a microwave power meter. It is clear that the match and, therefore, the accuracy of a power measurement using a microwave counter cannot be as accurate as that of a microwave power meter. Nevertheless, the HP 53150 Series Microwave Counters still make very useful power measurements. With short, high-quality cables, typical overall measurement accuracy can be better than ± 2.5 db up to about 20 GHz. The key to this accuracy is the careful factory calibration of the counter s power measurement with respect to both frequency and level. For example, the counter diode s frequency response is measured every 50 MHz from 50 MHz to the counter s maximum frequency: approximately 400 points for a 20 GHz counter and almost 1,000 points for a 46 GHz counter. Since the counter measures the frequency of the incoming signal, it is a simple matter of the internal microprocessor making the adjustment for the power measurement diode s frequency response. This is the reason the counter must make a frequency measurement before it can make a power measurement. Without the frequency measurement, the correction for frequency response of the power measuring diode would not be known and the resulting measurement uncertainty could be quite large. Since a microwave counter operates over a typical input power range of +10 to 30 dbm, its power-sensing diode operates over a similar range. Unfortunately, over this range the diode goes from square law to linear characteristics with a transition zone in between. Rather than trying to correct this mathematically via a complex polynomial, a powercalibration look-up table is stored in the counter s memory. This table characterizes the actual HINT 3 power sensing diode in the sampler in 1 db steps over the 40 db dynamic range at 4 to 6 frequencies (depending on the counter s frequency range). This ensures proper compensation for deviations from square law. The result is that the accuracy of the HP 53150 Series power measurements is quite good at the counter s input connector (better than ±1.5 db to 20 GHz over its 55 C operating range). When combined with a short, highquality, microwave cable, measurement accuracies of better than ± 2.5 db can be expected to about 20 GHz. 7

HINT 4 Improving the Accuracy of Counter Power Measurements When better accuracies are required, a microwave counter may still be able to do the job. Several measurement techniques can be used to improve accuracy of the counter s power measurement. Improving the match of, or compensating for, the insertion loss of the cable associated with getting the desired signal to the counter can improve measurement uncertainties. These techniques typically improve the power measurement uncertainty at the test point to below ±1.0 db over a wide frequency range, and even better at low frequencies (below 3 GHz) or over narrow bandwidths (1 GHz or so). The first of these techniques improves the match of the measurement by minimizing the interaction between the counter s input and the microwave cable connected to it. This involves putting a quality, fixed microwave attenuator (pad) between the input cable and the counter. A value as small as 3 db is useful although 4 to 6 db is preferred. For example, since the HP 53150 Series input match is typically 10 db (2:1 SWR) to 10 GHz, a 3 db pad improves it to roughly 16 db (1.4:1 SWR) while a 6 db pad improves it to 22 db (1.17:1 SWR). This assumes the match of the pad is much better than 25 db. Since real-life, highquality pads have a match of around 20 to 25 db, more than about 6 db of attenuation will not improve the effective match of the counter s input. Note that a 16 to 20 db figure is the typical match of power meter sensors in this frequency range. A similar pad at the test point will isolate the cable from the source to be measured. This preferred measurement technique is illustrated in Figure 3. The down side to this technique is that the signal level is reduced by the total attenuation of the test set-up: the insertion loss of the cable up to 1.5 db for a quality 1 meter cable at 10 GHz) plus the 2 pads (6-12 db). Thus, the effective lower end of the measurement is reduced by roughly 10 db, changing a 30 dbm measurement limit to 20 dbm, for example. If this amount of insertion loss is a problem, using a single pad at the worst SWR point (typically the counter) is an effective compromise. 8

Device Under Test 4 db Pad 4 db Pad Quality Microwave Cable Microwave Counter Figure 3. Recommended power measurement set-up. An alternative technique is to reduce uncertainty by calibrating the cable s insertion loss variation with frequency over the range of frequencies that are expected to be measured. This technique involves making an insertion loss measurement with a reference power meter at each frequency point of interest. This insertion loss can then be subtracted from the actual microwave counter s reading to improve measurement uncertainty. With the HP 53150 Series unique Power Correction mode, the storage of the insertion loss data and subsequent subtraction can be done automatically. This mode stores up to 10 insertion loss/frequency points, then linearly interpolates between these points should the measurement fall between them. Since nine different front-panel setups can be stored, characteristics for up to nine separate cables can be stored. Alternatively, a single cable with up to 90 calibration points can be stored. Of course, the best possible measurement accuracy will be achieved by combining the above two techniques when possible. 9

Hewlett-Packard CW Microwave Counter Selection Guide Three models: HP 53150A 20 GHz Counter HP 53151A 26.5 GHz Counter HP 53152A 46 GHz Counter Simultaneous frequency and power measurements with analog peaking indicator. Power measurement accuracy to ±2 db and better. Power Correction mode for cable loss compensation. Truly field portable: small, lightweight, rugged, optional internal battery, optional soft carrying case. HP s full-featured CW microwave counters for field, factory, or lab ATE ready: HP-IB and RS-232 interfaces standard, SCPI language programming, optional rack mounting kit. Laboratory features: frequency and power offsets, measurement averaging, relative (delta) frequency and power measurements, resolution and sample rate control, instrument set-up save and recall, optional oven timebase. For further information, visit our web site at Access HP: http:// www.tmo.hp.com/ or call your local HP office and ask for the HP 53150 Series Product Overview. 10

11

For more information visit our web site at Access HP: http://www.tmo.hp.com/ For more information about Hewlett-Packard test and measurement products, applications, services, and for a current sales office listing, visit our web site, http://www.hp.com/go/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative. United States: Hewlett-Packard Company Test and Measurement Call Center P.O. Box 4026 Englewood, CO 80155-4026 1 800 452 4844 Canada: Hewlett-Packard Canada Ltd. 5150 Spectrum Way Mississauga, Ontario L4W 5G1 (905) 206-4725 Europe: Hewlett-Packard European Marketing Centre P.O. Box 999 1180 AZ Amstelveen The Netherlands (31 20) 547 9900 Japan: Hewlett-Packard Japan Ltd. Measurement Assistance Center 9-1, Takakura-Cho, Hachioji-Shi, Tokyo 192-8510, Japan Tel: (81-426) 56-7832 Fax: (81-426) 56-7840 Latin America: Hewlett-Packard Latin American Region Headquarters 5200 Blue Lagoon Drive 9th Floor Miami, Florida 33126 U.S.A. Tel: (305) 267 4245 (305) 267-4240 Fax: (305) 267-4288 Australia/New Zealand: Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130 Australia Tel: 1 800 629 485 (Australia) 0800 738 378 (New Zealand) Fax: (61 3) 9210 5489 Asia Pacific: Hewlett-Packard Asia Pacific Ltd. 17-21/F Shell Tower, Times Square, 1 Matheson Street, Causeway Bay, Hong Kong Tel: (852) 2599 7777 Fax: (852) 2506 9285 Data Subject to Change Printed in U.S.A. September 1998 12 Hewlett-Packard Company Copyright 1998 5967-6195E