Directions for running IC samples on the Shimadzu TOC analyzer. (5/18/2015 DER)

Similar documents
30 Plex Human Luminex (Invitrogen Kit, Single Plate)

just below the screen. Data collection will begin, and a graph will show your data being plotted in real time.

SOP-P094. BioMek 2000 Compound Protocol Cyan/Hypercyt + Analysis

Measuring Nitrate in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

SOLiD EZ Bead Emulsifier

Corrie Moreau Page 1 7/23/10

Fire CR Calibration Guide

Introduction to Spectrometry

Installation Instructions: Epson 1430 CFS

Metzger Lab Protocol Book Written by Pierre Coutu 08/06/2003. Ionoptix Procedure

NEULOG CALCIUM ION LOGGER SENSOR GUIDE

The CO2 Sensor Calibration Kit

Creating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II

Page 1. Center for Environmental Health Sciences, MIT NIEHS Community Outreach and Education Program Core Center Award P30-ES 02109

When to Perform the Test CCD 4-Color and/or Seq Fill Capillary Run Modules

Ocean Optics R-2000 Raman Spectrometer Setup and Operating Instructions Arlen Viste and Deanna Donohoue April 2000 Update 2003, DEW

Sodium Thiosulphate and hydrochloric acid

Robotic Manipulation Lab 1: Getting Acquainted with the Denso Robot Arms Fall 2010

Alon s SCN ChIP Protocol

MicroLab 500-series Getting Started

INSTRUCTION MANUAL DISSOLVED OXYGEN-METER MODEL CDO-01

What is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS

STUDENT LABORATORY WORKSHEET EXPERIMENT B: NANOSCALE THIN FILMS

Kitchen and Bath Design Tutorial

Saponification and the Making of Soap - An Example of Basic Catalyzed Hydrolysis of Esters

CELL-DYN SYSTEM

CHM 152 Lab 1: Plotting with Excel updated: May 2011

Outline of Standard Protocols for Particulate Organic Carbon (POC) Analyses

illumina TruSeq RNA Sample Prep. (LT) Protocol 1

Olympus xcellence Software - basic user guide

ISONIC PA AUT Spiral Scan Inspection of Tubular Parts Operating Manual and Inspection Procedure Rev 1.00 Sonotron NDT

5 Using KODAK EKTACOLOR RA Chemicals with Tray, Drum, and Rotary-Tube Processors

Volumetric (transfer) Has been calibrated to deliver only one volume. Do not blow out last drop (there are no etched rings).

USING ABSOLUTE BLACK INKS FOR MAKING SCREEN POSITIVES ON THE EPSON STYLUS PRO 4900

Setup and Walk Through Guide Orion for Clubs Orion at Home

Familiarization with the Servo Robot System

LIGHT-SCENE ENGINE MANAGER GUIDE

ab Firefly Cytometer Setup Particles

Refillable Cartridge Ink Fill and Use Instructions for Epson 7800/9800 and 7880/9880 Printer Models

Kitchen and Bath Design Tutorial

TUSCANY 4PK CUSHION ROCKERS ASSEMBLY INSTRUCTIONS ITEM# SC-J-250-1ER/4

Table of Contents. Chapter 1: Software Installation...1. Chapter 2: Running the Software II. Daily Practical Operation...10

NAME SECTION PERFORMANCE TASK # 3. Part I. Qualitative Relationships

STUDENT LABORATORY WORKSHEET EXPERIMENT A: DRUG DELIVERY

Finer Points of ICP-OES Setup and Operation

SLICEWare. Operator Manual for software. Notes and comments

Experiment 2B Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART B. MOLECULAR FLUORESCENCE SPECTROSCOPY

What is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS

Illumina TruSeq Stranded mrna (LT) Protocol 1

Procedure & Checklist - 1 kb Template Preparation and Sequencing

LABORATORY SAFETY SERIES: Safe Handling Of Glassware

Recording EPR spectra using the Loop Gap Resonator (LGR)

Installation Instructions: Epson R200 CFS

Autodesk Inventor Drawing Manager Tips & Tricks

Authors: Doug Bail, Angela Ziebell, Maria Triantis Figure 1 ph Electrode.

QwikCheck Beads Precision and Linearity Kit Instructions QwikCheck GOLD Analyzer

The Basic RAW Image Processing Workflow In PSE 2018

KODAK T-MAX 100 Direct Positive Film Developing Outfit

Experiment 2A. Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART A. MOLECULAR ABSORPTION SPECTROPHOTOMETRY

Supplementary User Manual for BSWA Impedance Tube Measurement Systems

Procedure & Checklist - 2 kb Template Preparation and Sequencing

Rev F 8/30/ O Neil Rd, Macedon NY Telephone: (315) Fax: (315)

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

Kitchen and Bath Design Tutorial

COM-300 ph / EC / TDS / TEMP METER

PURDUE UNIVERSITY PCB BLACK HOLE APPLICATION FOR THRU-HOLE PREPARATION FOR ELECTROPLATING RF LAB. Professor in charge: William Chappell MSEE 289

ILFORDINFO HOW TO PROCESS YOUR FIRST BLACK AND WHITE FILM

What is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS

Preparing Smaller-Than-Package-Size Amounts of KODAK Processing Chemicals

mi-mag mrna Isolation Kit

PRINTING SETUP with OMEGA ENLARGER

Xcalibur. LCquan. Tutorial. Quantitative Analysis of a Three-Drugs Data Set Software Version 2.8

Using the Hitachi 3400-N VP-SEM

Basic P-XRD instructions for Operating the Instrument

Inventor Modeling Procedure By: Eric Small January 18, 2011

Olympus Time-lapse Microscope Basic operation

How to join your Online Tutoring Appointment. Step 1: Go to Log in using your username and password.

Procedure & Checklist - 20 kb Template Preparation Using BluePippin Size-Selection System

Procedure & Checklist bp Amplicon Library Preparation and Sequencing

TITLE: GUIDELINES FOR SAMPLE COLLECTION U + B

1. Setup Output mode. 2. Using a Fixed tile size

Procedure & Checklist - 10 kb Template Preparation and Sequencing (with Low-Input DNA)

MODEL SETUP FOR RENOVATION PROJECTS INSTRUCTIONS AND TUTORIALS

Hall Effect User Manual v1.0

PROFILING WITH WASATCH SOFTRIP

Interior Orientation Introduction Project Open File

Procedure and Checklist - 20 kb Template Preparation Using BluePippin Size-Selection System

ChE 436 Lab Project 1 Armfield Level Control

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Procedure & Checklist - 10 kb Template Preparation and Sequencing

The Discussion of this exercise covers the following points: On-off control On-off controller with a dead band. Conductivity control

HP 564 and 920 InkJet Cartridges Refill Instructions (Professional Version)

TECHNICAL INFORMATION

1. Setup Output mode. 2. Using a Fixed tile size

We want to thank and acknowledge the authors for sharing this protocol and their contributions to the field.

EDUCATION GIS CONFERENCE Geoprocessing with ArcGIS Pro. Rudy Prosser GISP CTT+ Instructor, Esri

RayBio anti-mouse IgG Magnetic Beads

10.2. Scanning Document Camera Scoring. Page 1 of 5. How do I score answer sheets using a document camera? STEP 1

ACCU-GOLD QUICK START MANUAL

History of Control Charts

Transcription:

Directions for running IC samples on the Shimadzu TOC analyzer. (5/18/2015 DER) Turn on the TOC analyzer by pressing the on switch located in the lower left corner of the panel on the front of the instrument. It will take at least 30 minutes for the instrument to warm up to temperature and stabilize. Open the TOC-L sample table editor by double clicking the shortcut on the desktop. User name and password are both blank. Open a new sample table by clicking the New button near the upper left corner of the window. The system should be TOC-TN for internal sparging and toc-tn2 for external sparging. Press OK. Press the connect button near the center of the tool bar. A status window should open with a message like open port 100 % then initializing. Set up calibration curves In the sample table editor, select the cal curve tab and pres the New button. The System should be TOC-TN. Leave the user name blank. Add comments as desired and press next. Select normal and use dilution from standard solution. Press Next. Select IC for the analysis.. The calculation method should be Linear Regression, the zero shift should be off and Multiple Injections should be on. Enter a file name; the name should include the full date, a person or project name, the concentration of the top standard, the analysis type. The preferred format for all file names is: 2013 May 30 project analysis Press Next. The standard parameters for DIC are: Units: mg/l Injections: 2/3 No. of washes: 2 SD max: 0.1 CV max: 2% Press Next In the calibration points table, click on add. The first calibration point is usually 0 ppm so the Standard Solution Conc. and the Cal. Point Conc. should both be 0. The no. of Injections should be 2/3, SD Max =0.1 and the CV Max=2% for all calibration points. Press OK to enter the point into the standard curve.

Press Add for more calibration points. Set the value of the Standard Solution Conc. to the value of the highest standard and the Cal. Point Conc. should be set to the desired value of that point. Press OK to enter the point into the standard curve. Cal curves should have 5 points ranging from 10 % to 100% of the highest standard plus a zero. Curves should be set up starting with the lowest concentration and ending with the highest. Note the Injection volume after entering the highest standard. Normally, 100ul is used for DIC measurements. Press Next after all calibration points have been entered The Use Default Setting box should be checked and the correlation Coeff. Check box should not be checked. Press Finish to complete the standard curve. Create the second calibration curve. The standard practice is to create a calibration curve with a top standard of 10 mg/l as the first calibration curve and a second curve with a top standard of 50 mg/l as the second. If a sample exceeds the 10 mg/l concentration of the first curve, the instrument will automatically jump to the second curve. This minimizes the number of sample needing to be rerun. After completing the high and low calibration curves, set up a method Set up a method Click on the method tab in the lower left corner of the software window, then press New button. The System should be TOC-TN. Leave the user name blank. Add comments as desired and press Next. Select IC for the analysis. Enter the default sample name and ID you want (leaving it blank is fine). Manual dilution is 1 and number of determinations is 1. Enter a file name, please include the date and project or person name. The preferred format for all file names is: 2013 may 30 project analysis Press Next. Select the first desired calibration curve which should be your low concentration standard curve as calibration curve 1. Then select the next highest calibration curve as curve 2. Then press Next. The standard parameters for IC are: Units: mg/l Injections: 2/3 No. of washes: 2 SD max: 0.1 CV max: 2%

Multiple Injections should be selected. Press Next. Use Default Settings should be selected. Press Next. On the Pharmaceutical Water Testing page, None should be selected. Press Finish to save the method. Set up a sample table If there is not a new sample table open already, open a new sample table by clicking the New button near the upper left corner of the window. The system should be TOC-TN. Press OK. Press save to save the sample table. File name should follow the format: yyyy mon dd project. To add samples to the table, highlight the line in the table and right click. Choose insert multiple samples. Use the browse button to find and select the method to use to create the sample parameters, then hit next. Type in or select the number of samples and the starting vial number. Then hit finish. A list of the samples and vial numbers will appear along with a diagram of the sampler. Verify that the samples and vial numbers are correct. Make sure the vial type in the lower right corner is 40 ml. Press OK to close the diagram window. To add a standard curve to the table, highlight the line in the table and right click. Select insert standard curve. Highlight the desire standard curve and press Open. Check that the vial numbers for the curve are correct. The vial number is in the last column on the right of the sample table. To edit the vial numbers, click the sampler diagram icon in the upper right of the sample table window. A list of the sample/standards and vial number will appear along with a diagram of the sampler. Edit the vial numbers as needed then press OK. Repeat the above step for the second(high) standard curve. Save the sample table when it is complete. The general setup for a sample run is: 3 samples of DI water at the beginning of the run to prime the instrument. The low concentration calibration curve. The high concentration calibration curve. A DI water and a high and low check standard to verify the calibrations. 10 samples. A DI water and a check standard.

Repeat the 10 samples followed by a DI water and a check standard to the end of the run. Alternate between the high and low check standards. At the end of the samples, run a DI and both the high and low check standards. Other notes on sample table setup. If vial number 0 is enter for a sample or standard, the instrument will sample from the tube on the left side of the instrument next to the lines going to the autosampler. This tube may be placed in a flask or bottle. It is commonly used for the calibration standard as it can be sampled many times. The auto sampler is random access, meaning it can sample the numbered vials in any order and can sample a vial multiple times. This is useful for check standards and DI waters as the same vial may be sampled up to three times. A vial filled to the shoulder may be sampled up to 3 times. Standard Preparation The excel file: standard concentration calcultaionsv2 on the desktop of the instrument computer contains a list of commonly used calibration and check standard concentrations as well as a tool for calculating other dilutions. Typically, two standards are prepared, a calibration standard and a check standard. The calibration standard contains the highest concentration used in the calibration and the instrument is programmed to internally dilute it to create the other points on the calibration curve (see calibration curve section above). The check standard is usually prepared with about half the concentration of the calibration standard and is analyzed periodically though the run to verify instrument performance. Remove the stock solutions (IC) from the refrigerator and allow them to warm to near room temperature. A warm water bath may be used speed up this step. There are separate stock solutions for the calibration and check standards. Take the appropriate size volumetric flasks from the acid washed glassware cabinets above the instrument. The flasks are store full of DI water and should be emptied before use. Select the appropriate size volumetric pipets from the acid washed pipet drawer below the instrument. Pipet the desired amount of the C stock, into the volumetric flasks for the calibration and check standards. Use different pipets for each component! Add DI water to the volumetric flask to bring the total volume to the line. Cap the flasks, invert and shake gently several times to ensure complete mixing. The used pipets and flasks must be acid washed before the next use. Return the stock solutions to the fridge.

Starting a run. Pour the samples and standards into acid washed vials (located in the cupboard under the instrument) and place them into the autosampler according to the sample table. Check the instrument waste carboy on the floor in front of the instrument. Empty it down the sink if it is more than half full. Be sure to replace both waste tubes after emptying. Empty, rinse and refill the sampler wash bottle (left side of the sampler) with DI water. Make sure the tube to the sample goes to the bottom of the bottle. Check the level of the 20% H 3 PO 4 bottle between the instrument and the sampler. Add more H 3 PO 4 if it is less than half full. Wear safety glasses when handling H 3 PO 4! Make sure the end of the tube going to the instrument is at the bottom of the bottle. Empty, rinse and refill the water for dilution bottle (between the instrument and the sampler) with DI water. If you are running samples in a matrix other than water, move the tube going into the instrument to a container of sample matrix. Make sure the end of the tube goes to the bottom of the bottle. Press the Start button on the upper tool bar. Select Shut down Instrument. Press OK. Cleanup Users are expected to promptly clean up volumetric pipets, volumetric flasks, sample vials, caps, and any other materials used in running the instrument. Pipets, flasks and vials should be rinsed 3 times with DI water. Then place them in the 1.5 M HCl bath in room A254. Take care that there are no air bubbles in any of the items. Always wear a lab coat, gloves and safety glasses when working with or around the acid baths! Allow the items to soak in the acid bath overnight. Remove the items from the acid bath and rinse individually with DI 6 times. Pipets should then be placed in the105 C oven in A258. When they are completely dry, put them back in the drawer under the instrument. Volumetric flasks should be filled with DI and placed in the cabinets above the instrument. Vials should be placed upside down in the plastic racks and air dried before putting them in the cupboard below the instrument. The caps should be rinsed 6 times with DI water and aired dried before going back into the cupboard under the instrument.

Exporting results Open the sample table containing the results. Highlight the rows containing the desired results. Go into file, ascii export ; there will be 2 choices: normal and detailed. Normal will give one line for each sample with the average result for C and N for that sample. Detailed will give one line for each injection. Normal is the best one for obtaining the data in a form that can be used for subsequent calculations. Detailed is principally used for maintaining a record of the sample run that includes the details like how many injections were performed for each sample/analyte combination, peak areas and standard deviations.