3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

Similar documents
TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

1 Introduction. 2 The basic principles of NMR

High Field MRI: Technology, Applications, Safety, and Limitations

SIEMENS MAGNETOM Skyra syngo MR D13

[Maestro Class] MAGNETOM Harmony The New Degree of Perfection

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

HETERONUCLEAR IMAGING. Topics to be Discussed:

2014 M.S. Cohen all rights reserved

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

Reach out for the power in mobile x-ray imaging MULTIMOBIL 10. Answers for life.

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Field Simulation Software to Improve Magnetic Resonance Imaging

Background (~EE369B)

Image Quality/Artifacts Frequency (MHz)

MRI Metal Artifact Reduction

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

Liver imaging beyond expectations with Ingenia

Cook Medical Peripheral Intervention MRI Information

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T)

MR Advance Techniques. Flow Phenomena. Class II

2SK1056, 2SK1057, 2SK1058

Encoding of inductively measured k-space trajectories in MR raw data

Human Subjects in fmri Research

Retrospective Transmit Beamformation. Whitepaper. ACUSON SC2000 Volume Imaging Ultrasound System. Answers for life.

2SB648, 2SB648A. Silicon PNP Epitaxial. Low frequency high voltage amplifier complementary pair with 2SD668/A

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

HA17080 Series. J-FET Input Operational Amplifiers. Description. Features

Voltage Feedback Op Amp (VF-OpAmp)

2D, 3D CT Intervention, and CT Fluoroscopy

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging

Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging

MRI RF-Coils. Innovation with Integrity. Highest sensitivity for your preclinical MRI and MRS applications. Preclinical Imaging

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

ACRIN 6686 / RTOG 0825

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

PF0030 Series. MOS FET Power Amplifier. ADE (Z) 1st. Edition July Features. Ordering Information. Pin Arrangement

Downloaded from by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved

Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners

Power Amplifiers. Power with Precision

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy

2015 Spin echoes and projection imaging

Pulse Sequence Design and Image Procedures

2SC4927. Silicon NPN Triple Diffused. Application. Features. Outline. TV/character display horizontal deflection output. High breakdown voltage V CES

MR Basics: Module 8 Image Quality

Potential Risks of MRI in Device Patients

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve

BOLD fmri: signal source, data acquisition, and interpretation

MR Basics: Module 6 Pulse Sequences

Magnetic Resonance Research Facility (MRRF) Resources

See the big picture in urology. UROSKOP Omnia.

Application Guide & Release Notes

from signals to sources asa-lab turnkey solution for ERP research

The development of the RF-pulse for the low level SAR used by the MRI.

RF PULSE DESIGN FOR PARALLEL TRANSMISSION IN ULTRA HIGH FIELD MAGNETIC RESONANCE IMAGING. Hai Zheng. B.S., Xi an JiaoTong University, 2005

Diffusion and Functional MRI of the Spinal Cord Methods and Clinical Applications

LG1627BXC Clocked Laser Driver

Iterative Reconstruction in Image Space. Answers for life.

The compact all-rounder for surgical imaging. SIREMOBIL Compact L. Answers for life.

MRI with the VNS Therapy System October 2017

1.5T HIGH FIELD SMALL ANIMAL MRI

Pulse Sequence Design Made Easier

H 2 O and fat imaging

Annual Ceiling Amount: $

Supplementary Material

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)

THIS DOCUMENT IS FOR MAINTENANCE PURPOSES ONLY AND IS NOT RECOMMENDED FOR NEW DESIGNS

MRI Systems and Coil Technology

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

Example for a Gain Tracking Algorithm Required in a GSM Terminal Based on the Lucent Technologies Sceptre Platform

A Breakthrough in Sputtering Target Inspections: Ultra-High Speed Phased Array Scanning with Volume Focusing

Data. microcat +SPECT

2SK1949(L), 2SK1949(S)

RAD 229: MRI Signals and Sequences

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging

HA17358/A Series. Dual Operational Amplifier. ADE (Z) 1st Edition July Description. Features. Features only for A series

Magnetic Resonance Imaging Principles, Methods, and Techniques

2SK2554. Silicon N-Channel MOS FET. November Application. Features. Outline. High speed power switching

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging

Supplementary Figure 1

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel.

COMPACT POWER TWIN RELAY. 1 POLE x 2 30A (Dual relay) (FOR AUTOMOTIVE APPLICATIONS)

Ysio Max. The most direct way to the image. Answers for life.

Designing Better Industrial Robots with Adams Multibody Simulation Software

Magnetic Resonance Imaging

LM837 Low Noise Quad Operational Amplifier

The promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners

Compact yet Sophisticated

LG1628AXA SONET/SDH Gbits/s Transimpedance Amplifier

Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle

Besides long acquisition times and restricted spaces, acoustic noise

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

WU-Minn HCP 1200 Subjects Release: Reference Manual

2 Hardware for Magnetic Resonance Imaging

Transcription:

3T Unlimited ipat on MAGNETOM Allegra The Importance of ipat at 3T s medical

ipat on MAGNETOM Allegra The Importance of ipat at 3T The rise of 3T MR imaging Ultra High Field MR (3T) has flourished during recent years. The advantages of 3T field strength have led to the successful introduction of MAGNETOM Allegra. MAGNETOM Allegra is the first and only machine providing the best of MR for brain applications: 3T field strength, extremely compact magnet, easy to site design and ultra-fast gradients (slew rate: 400 mt/m/ms). Indeed, 3T field strength offers many advantages such as twice the signal-to noise ratio (SNR) of 1.5T field strength or increased sensitivity to T2* effects for better functional MRI studies. These benefits have boosted the interest for 3T MR systems, especially in brain applications. However, some challenges still remain. Susceptibility effects and consequently image distortions are more pronounced compared to 1.5T. The Specific Absorption Rate (SAR or the energy deposited in the body by the radio frequency transmission) increases by factor of 4 when going from 1.5T to 3T. Siemens overcomes these challenges using integrated Parallel Acquisition Techniques (ipat). ipat describes imaging strategies that use multiple element coils to receive MR signals in parallel and therefore speed up acquisition time or increase spatial resolution. MAGNETOM Allegra, similarly to other members of the MAGNETOM family, supports ipat. How does ipat serve 3T applications? 1.Overcome susceptibility effects and blurring ipat in single-shot techniques reduces blurring and minimizes susceptibility artifacts which are typically more pronounced at 3T than 1.5T. While already compensating for these effects using the ultra fast gradient performance of the MAGNETOM Allegra, ipat further overcomes susceptibility-induced geometric distortions. This is particularly useful in echo planar imaging-based techniques. 2. Keep SAR low SAR increases by a factor of 4, when going from 1.5T to 3T. This means that SAR limits set for patient safety are more rapidly reached. ipat reduces the number of RF excitation pulses and therefore reduces SAR. Note that ipat is only one component of the unique Siemens 3TCare initiative, which consists of developing new methods to reduce SAR. 3. Utilize the extra 3T SNR to minimize acquisition time or maximize resolution Because SNR at 3T is twice compared to 1.5T, the signal loss associated with ipat proportional to the square root of the acceleration factor is less critical at 3T. This means you can use higher ipat factors to speed up your acquisitions or increase image resolution. For example, ipat factor of 2 allows acquiring 27 64x64 slices per second. Alternatively, ipat factor of 2 allows matrix sizes of 96x96 in a time that conventional EPI techniques need for the reduced matrix sizes of 64x64. The increased resolution allows more specific interpretation and diagnosis due to better image quality.

MAGNETOM Allegra RF configuration MAGNETOM Allegra comes with 1 or 4 (option with syngo MR 2003A) fast 1 MHz bandwidth fully equivalent ipatcompatible channels. MAGNETOM Allegra ipat coil A 4-channel transmit/receive ipat-compatible head coil is available with syngo MR 2003A on MAGNETOM Allegra. This coil enables running ipat sequences with acceleration factors of 2. MAGNETOM Allegra to take advantage of ipat benefits at 3T ipat and 3T provides best image quality for brain imaging in regions affected by susceptibility effects. Secondly ipat is an alternative to reduce deposited energy for SAR intensive sequences. It offers also an alternative to trade-off the extra 3T SNR for acquisition speed. MAGNETOM Allegra with its very fast 400 T/m/s gradients and ipat, is the fastest MR scanner on the market.

[Figure 1] Single-shot EPI without and with ipat, in coronal orientation. Constant TE = 30 ms for optimal BOLD sensitivity at 3T. 4-channels head array, ipat-compatible transmit/receive ipat can be used to increase spatial resolution, i.e. the specificy of fmri experiments without an increase of the temporal resolution. Image quality improvements in regions prone to susceptibility artifacts are evident. Conventional EPI (matrix size 64x64), 22 slices/sec EPI with ipat 2 EPI (matrix size 64x64), 27 slices/sec EPI with ipat 2 EPI (matrix size 128x128), 16 slices/sec

[Figure 2] 2D TSE without and with ipat. TR/TE 5480/107 ms, 220x220 mm, 254x512 matrix, SL 3 mm 4-channels head array, ipat-compatible transmit/receive Conventional high resolution TSE (total scan time 1:44 min) High resolution TSE with ipat 2 (total scan time 1:00 min) [Figure 3] High resolution 3D ToF with ipat 2 TR/TE 42/6.2 ms, 704 matrix, TA 6:12 min 4-channels head array, ipat-compatible transmit/receive [Figure 4] High resolution single shot diffusion EPI with ipat 2 b-value of 1000 mm 2 /s, TE 77 ms, 256x384 matrix 4-channels head array, ipat-compatible transmit/receive

[UHF Class] MAGNETOM Allegra 3T Dedicated The ultimate brain machine The fastest MR system in the most compact design for brain imaging and spectroscopy. Fastest gradients in the industry: 400 T/m/s The information in this document contains general descriptions of the technical options available, which do not always have to be present in individual cases. The required features should therefore be specified in each individual case at the time of closing the contract. Siemens reserves the right to modify the design and specifications contained herein without prior notice. Please contact your local Siemens Sales representative for the most current information. Original images always lose a certain amount of detail when reproduced. All devices listed herein may not be licensed according to Canadian Medical Devices Regulations. The most compact 3T MR system available Cost-efficient siting: your 3T system can be installed in a 1.5T suite. High patient-friendliness with a total length of only 1.25 m The fastest MR system in the industry Outstanding image quality and speed, thanks to the Siemens gradient and gradient amplifier technology combined with integrated Parallel Acquisition Techniques (ipat). Please contact in USA: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Tel.: 610-448-4500 Tel.: 1 888-826-9702 Fax: 610-448-2254 Japan: Siemens-Asahi Medical Technologies Ltd. Takanawa Park Tower 14 F 20-14, Higashi-Gotanda 3-chome Shinagawa-ku Tokyo 141-8644 (+81) 354238489 Asia: Siemens Medical Solutions Asia Pacific Headquarters c/o Siemens Pte Ltd. Block 28 Ayer Rajah Crescent No. 06-06 Singapore 139959 (+65) 6341 0990 Or contact your local Siemens sales representative Siemens AG Medical Solutions Magnetic Resonance Henkestr. 127, D-91052 Erlangen Germany Telephone: ++49 9131 84-0 www.siemensmedical.com Order-No. A91100-M2220-B85-1-7600 Printed in Germany CC 64085 WS 08033.