Ring-Down versus Comet Tail: Two artifacts uncovered

Similar documents
Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Computer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

Correction of the local intensity nonuniformity artifact in high field MRI

Improvement of CT image quality with iterative reconstruction idose4

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Evaluation of no-grid radiography using the digital scattered x-ray removal processing

Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

The impact of increasing SID on patient dose in pa abdomen imaging

The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)

Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing

Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis

Multi-slice computed tomography analysis of bullet trajectory in forensic investigation.

Lumbar disc height assessment: a comparative study between EOS and digital radiography

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

Practical guidelines for color calibration and quality assurance of medical displays

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging

Aim. Images for this section: Page 2 of 13

Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)

The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution

Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Development of new dosimeter for measuring dose distribution in CT

Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography

Artifacts. Artifacts. Causes. Imaging assumptions. Common terms used to describe US images. Common terms used to describe US images

Ultrasound & Artifacts

Voice recognition versus transcriptionist

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล

Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners

Indicator of Small Calcification Detection in Ultrasonography using Decorrelation of Forward Scattered Waves

Echo Artifacts: The Cause and Solution

Interpreting RF Radiation Patterns

Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study

Data Sheet of SAW Components

Data Sheet of SAW Components

Challenges to Autonomous and Continuous Fluid Level Sensing

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time

Interaction of Sound and. logarithms. Logarithms continued. Decibels (db) Decibels (db) continued. Interaction of Sound and Media continued

Datasheet of SAW Device

Datasheet of SAW Device

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

MULTI-FREQUENCY ULTRASOUND IMAGING: PHANTOM STUDY

ORDERING A PHOTOGRAPH FOR PERSONAL USE

Credits January 9, ASCRS Course, Paper, and Poster Categories/Topics and Subtopics February 4 March 4

Introduction to Medical Engineering (Medical Imaging) Ultrasound Imaging. Ho Kyung Kim Pusan National University

Datasheet of SAW Device

Lesson 06: Pulse-echo Imaging and Display Modes. This lesson contains 22 slides plus 15 multiple-choice questions.

Datasheet of SAW Device

Datasheet of SAW Device

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

Datasheet of SAW Device

United States Patent (19)

Datasheet of SAW Device

Datasheet of SAW Device

Datasheet of SAW Device

Ultrasound Imaging Ultr Michael Dadd 2007

Package Pin Out Configuration. Part Number Compliance Marking Reel Size (inch) Tape Width (mm) Quantity per Reel UMC4NQ-7 Automotive NP ,000

Datasheet of SAW Device

A Guide To Writing And Submitting Your Written Testimonial.

ADVANCED INFORMATION

Datasheet of SAW Device

Datasheet of SAW Device

Isolation Scanner. Advanced evaluation of wellbore integrity

Proceedings of Meetings on Acoustics

Black Science Fiction Script Movie Project 2012

Datasheet of SAW Device

Data Sheet of SAW Components

Photomultiplier Tube

Datasheet of SAW Device

TI Designs: Biometric Steering Wheel. Amy Ball TIDA-00292

ADC114YUQ. Mechanical Data. Features. Ordering Information (Notes 4 & 5) Marking Information NXX 1Y7 YM NXX YM

NONDESTRUCTIVE EVALUATION OF CLOSED CRACKS USING AN ULTRASONIC TRANSIT TIMING METHOD J. Takatsubo 1, H. Tsuda 1, B. Wang 1

Image Optimization: The Sonographer s Responsibility. Prepared by Cathy Daniels, EdD, RTR, RDMS, RDCS, RVT

Datasheet of SAW Device

Datasheet of SAW Device

ULTRASONIC GUIDED WAVE ANNULAR ARRAY TRANSDUCERS FOR STRUCTURAL HEALTH MONITORING

Process Leak Detection Diagnostic with Intelligent Differential Pressure Transmitter

Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures

Datasheet of SAW Device

Blood Flow Analysis Using Ultrasound Doppler Technique

Vivid S5. Cardiovascular ultrasound system. GE Healthcare. Davis Medical

Sonic crystal noise barrier using locally resonant scatterers

Medical Imaging (EL582/BE620/GA4426)

SOD123. Top View. M5X = Product Type Marking Code YM = Date Code Marking Y = Year (ex.: E = 2017) M = Month (ex: 9 = September)

Green. SOD123F (Standard) Top View. Part Number Qualification Case Packaging US1GWF-7 AEC-Q101 SOD123F (Standard) 3,000/Tape & Reel

Shear Wave elastography on the Toshiba Aplio 500

Datasheet of SAW Device

MEDICINE LICENSE TO PUBLISH

Datasheet of SAW Device

Top View BAT54T BAT54AT BAT54CT BAT54ST

Datasheet of SAW Device

Transcription:

Ring-Down versus Comet Tail: Two artifacts uncovered Poster No.: C-2288 Congress: ECR 2013 Type: Educational Exhibit Authors: S. Méndez Garrido 1, J. Arce Duran 2, A. Talegón Meléndez 1 ; 1 2 Sevilla/ES, Seville/ES Keywords: DOI: Calcifications / Calculi, Artifacts, Physics, Ultrasound, Ultrasound physics, Anatomy, Abdomen 10.1594/ecr2013/C-2288 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. www.myesr.org Page 1 of 17

Learning objectives - To recognize these artifacts in ultrasound. - To learn the physical basis that support in order to discriminate them. - To guide the possible underlying pathology thanks to the correct interpretation of these artifacts. Background From the standpoint of radiologic imaging, we can define the term artifact as an image we cannot correlate with real anatomical structures. In ultrasound, artifacts can produce structures that appear in the image but they are not present anatomically, although structures which are present anatomically may be completely absent. But that is not all, because artifacts can even show structures that are present but incorrect in location. In our daily practice it is very usual to find them, and we can say they are a double-edged sword because they can interfere in a correct interpretation of the image, even they can be very useful to help us to do a precise diagnosis. Therefore, there are many reasons to know, even not necessary with much depth, the physical bases that are underneath these phenomena. If we do not exercise often, we can easily forget, some of the less unknown artifacts. The fact is that if we understand, at least once, its production mechanism, it will be much less arduous its evocation in subsequent occasions. To understanding the differences between the two, in first place we have to keep in mind their common aspects, considering that both artifacts are caused by multiple echoes. We are going to describe below in a precise and concise way, the production mechanisms that underlie both artifacts. Imaging findings OR Procedure details COMET TAIL ARTIFACT Page 2 of 17

We can consider this phenomenon as a form of reverberation: We assume that sounds travels in a straight line, at constant speed. Therefore, we can determinate the origin of an echo that goes back if we quantify the time it takes to return. In the presence of two reflective interfaces, the echoes generated from the main beam may be repeatedly reflected, in repeated trips before go back to the transducer, where they may be detected. Each echo is received erroneously transcribed as a band located at a greater depth, so we may see multiple parallel lines of decreasing intensity and equidistant from each other (Fig. 2 on page 4). In this artifact, the two reflective interfaces, and therefore the generated echoes are closely spaced. In the image, sequential echoes can be so close together that the individual signals cannot be seen. In addition, delayed echoes have lower amplitude, secondary to attenuation process; this decrease in amplitude is shown as a width of echoes increasingly diminished. The result is an artifact caused by the principle of reverberation, but with a conical or triangular shape. This phenomenon occurs when the beam comes into contact with metal objects (Fig. 3 on page 5 ) such as metal clips of suture, needles, foreign bodies or calcification (granulomas, cholecystolithiasis (Fig. 4 on page 6), calcified myomas, nephrolithiasis ). A classic example is cholesterolosis of the gallbladder wall (Fig. 5 on page 7 ). RING-DOWN In the past, it was thought this artifact was a variant of the previous. This statement was based in the similar appearance both artifact exhibit. However, it is known that Ring-Down appears mostly due to gas (rather than metal, as in the case of comet tail). We need multiple gas bubbles to produce it. When the ultrasound beam reaches the gas bubbles, it is capable of exciting the liquid trapped between the bubbles, which causes the liquid resonates. These vibrations create a continuous sound wave which is transmitted back to the receiver. This phenomenon is shown as a line or Page 3 of 17

series of bands extending parallel after the image corresponding to gas (Fig. 6 on page 8 ). Some examples we can found in our daily practice are abundant gas in the colon structure ( Fig. 7 on page 9 ) or those produced after surgery, such as aerobilia after cholecystectomy or hepaticojejunostomy (Fig. 8 on page 10), air bubbles after injections (Fig. 9 on page 11) or in pathological processes for example in Fournier gangrene, where the display of gas adjacent to the scrotum is charasteristic of the disease (Fig. 10 on page 12). Considering air is the cause of this artifact, seen as "dirty shadows" in the image, it is possible to think this phenomenon is not useful, because you can hide deeper structures. Nothing is further from the truth. If you find air in the thickness of the gallbladder wall, or in the context of pyelonephritis, this artifact can help us to diagnose emphysematous cholecystitis (Fig. 12 on page 14) or pyelonephritis. Likewise, if we visualize the Ring-Down in the thickness of a fluid collection we probably are in the face of an abscess (Fig. 11 on page 13), or if we see that image in peritoneal cavity is probably we are in front of a perforated hollow viscera. Images for this section: Page 4 of 17

Fig. 2: Physical basis of comet tail artifact Page 5 of 17

Fig. 3: Endoprothesis. Page 6 of 17

Fig. 4: Microlithiasis Page 7 of 17

Fig. 5: Cholesterolosis polyps inside the gallbladder. Page 8 of 17

Fig. 6: Physical basis of Ring-Down artifact Page 9 of 17

Fig. 7: Gas in duodenal frame Page 10 of 17

Fig. 8: Aerobilia after hepaticojejunostomy. Page 11 of 17

Fig. 9: Gas after intravitreal injection. Page 12 of 17

Fig. 10: Fournier gangrene. Page 13 of 17

Fig. 11: Intrahepatic abscess with gas. Page 14 of 17

Fig. 12: Emphisematous cholecystitis Page 15 of 17

Conclusion Comet tail and Ring-Down artifact usually are used as synonyms, due to their similar appearance on ultrasound. But their physical basis are completely different and the mechanism that produce them: - Comet tail artifact is a form of reverberation; therefore it has the same production basis, i.e, repeated trips between the transducer and two reflective surfaces. - Ring-down artifact occurs when the ultrasound beam excites the liquid trapped between gas bubbles, producing the liquid to vibrate or resonate. - Metallic elements/calcifications cause comet tail artifact, being thus useful for diagnosing from foreign bodies, metal clips of suture, catheters to calculus or granulomas. - Gas is responsible for Ring-Down artifact. Its display can translate from gas than can be found in physiological situations (for example inside bowel), after surgical procedures (aerobilia after cholecystectomy or hepaticojejunostomy) or pathological processes (pneumatosis intestinalis, emphysematous cholecystitis ). It is necessary to emphasize that knowledge of clinical data and patient history are essential to guide in the ethiology. References - US Artifacts. Feldman M, Katyal S, Blackwood M. Radiographics 2009; 29: 1179-1189. - Talegón Meléndez A, Molina Chacón J. Imagen por Ecografía. En: JL del Cura, S Pedraza, A Gayete. Radiología Esencial. Madrid: Editorial Médica Panamericana; 2010. p. 35-47. - Física práctica. En: Middleton, Kurtz, Hertzberg. Ecografía. Madrid: Marbán; 2007. p. 3-27. - Duplication Images in vascular sonography. Rubin J, Gao J, Hetel K, Min R. J Ultrasound Med 2010; 29: 1385-1390. Page 16 of 17

- Is it real or is it an artifact. Keogh C, Coopberg P. Ultrasound Quarterly 2001; 17 (4): 201-210. - Phisics tutorial for residents: topics in US. Hangiandreou,N. Radiographics 2003: 1019-1033. - Gas at abdominal US: appearence, relevance, and analysis of artifacts. Wilson S, Burns P, Wilkinson L, Simpson D, Muradali D. Radiology 1999; 210: 113-123. - Ring-down artifacts on abdominal sonography to predict pulmonary abnormalities in the emergency department. Tsai CL, Wang HP, Lien WC, Chen CC, Lai TI, CHen WJ. Emerg Med 2005; 22 (10): 747-748. - The V-Shaped artifact ot the gallbladder wall. Lafortune M, Gariépy G, Dumont A, Breton G, Lapointe R. AJR AM J Roentgenol 1986 sep; 147 (3): 505-508. - Color comet tail artifact: clinical applications. Tchelepi H, Ralls PW. Am J Roentgenol 2009 Jan; 192 (1): 11-18. - The comet tail artifact. An ultrasound sign of alveolar-intestitial syndrome. Lichtenstein D, Mézière G, Biderman P, Gepner A, Barré O. Am J Respir Crit Care Med 1997 Nov; 156 (5): 1640-1646. - Sonography detection of pneumoperitoneum in patients with acute abdomen. Lee DH, Lim JH, Ro YT, Yoon Y. AJR Am J Roentgenol 1990 jan; 154 (1): 107-109. Personal Information Page 17 of 17