Better Imaging with a Schmidt-Czerny-Turner Spectrograph

Similar documents
Improved Spectra with a Schmidt-Czerny-Turner Spectrograph

Improving the Collection Efficiency of Raman Scattering

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

SpectraPro 2150 Monochromators and Spectrographs

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember

MS260i 1/4 M IMAGING SPECTROGRAPHS

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Oriel MS260i TM 1/4 m Imaging Spectrograph

DESIGN NOTE: DIFFRACTION EFFECTS

The designs for a high resolution Czerny-Turner spectrometer are presented. The results of optical

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club

AST Lab exercise: aberrations

Specifications. Offers the best spatial resolution for multi-stripe spectroscopy. Provides the user the choice of either high accuracy slit mechanism

Waves & Oscillations

Lecture 4: Geometrical Optics 2. Optical Systems. Images and Pupils. Rays. Wavefronts. Aberrations. Outline

A TUTORIAL By J.M. Lerner and A. Thevenon TABLE OF CONTENTS. Section 1:DIFFRACTION GRATINGS RULED & HOLOGRAPHIC

Applications of Optics

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

Advances in Hyperspectral Imaging Technologies for Multi-channel Fiber Sensing

Historical. McPherson 15 Mount

The Optics of Spectroscopy A Tutorial. By J.M. Lerner and A. Thevenon

Observational Astronomy

Optical design of a high resolution vision lens

ECEN 4606, UNDERGRADUATE OPTICS LAB

Optical basics for machine vision systems. Lars Fermum Chief instructor STEMMER IMAGING GmbH

Tutorial Zemax Introduction 1

Application Note (A11)

OPTICAL SYSTEMS OBJECTIVES

UltraGraph Optics Design

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

Basic spectrometer types

UV/Optical/IR Astronomy Part 2: Spectroscopy

Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

ECEN 4606, UNDERGRADUATE OPTICS LAB

Chapters 1 & 2. Definitions and applications Conceptual basis of photogrammetric processing

DIMENSIONAL MEASUREMENT OF MICRO LENS ARRAY WITH 3D PROFILOMETRY

Εισαγωγική στην Οπτική Απεικόνιση

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

DESIGN AND CHARACTERIZATION OF A HYPERSPECTRAL CAMERA FOR LOW LIGHT IMAGING WITH EXAMPLE RESULTS FROM FIELD AND LABORATORY APPLICATIONS

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers

EE-527: MicroFabrication

Comparison of low-cost hyperspectral sensors

!!! DELIVERABLE!D60.2!

Notes from Lens Lecture with Graham Reed

Aberrations and adaptive optics for biomedical microscopes

Chemistry Instrumental Analysis Lecture 7. Chem 4631

SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals

instruments Solar Physics course lecture 3 May 4, 2010 Frans Snik BBL 415 (710)

R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad.

Long-Range Adaptive Passive Imaging Through Turbulence

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

Reflection! Reflection and Virtual Image!

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Astro 500 A500/L-8! 1!

REAL TIME THICKNESS MEASUREMENT OF A MOVING WIRE

Maya2000 Pro Spectrometer

Image Slicer for the Subaru Telescope High Dispersion Spectrograph

Collimation Tester Instructions

The Optics of Mirrors

Investigations towards an optical transmission line for longitudinal phase space measurements at PITZ

Image Formation. Light from distant things. Geometrical optics. Pinhole camera. Chapter 36

High throughput spectrometer for fast localized Doppler measurements

Big League Cryogenics and Vacuum The LHC at CERN

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

Eric B. Burgh University of Wisconsin. 1. Scope

arxiv: v1 [astro-ph.im] 26 Mar 2012

plasmonic nanoblock pair

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

Applying of refractive beam shapers of circular symmetry to generate non-circular shapes of homogenized laser beams

Potential benefits of freeform optics for the ELT instruments. J. Kosmalski

Physics 308 Laboratory Experiment F: Grating Spectrometer

Astigmatism Particle Tracking Velocimetry for Macroscopic Flows

Spatially Resolved Backscatter Ceilometer

Introducing Celestron s EdgeHD Optical System

OCT Spectrometer Design Understanding roll-off to achieve the clearest images

Performance Factors. Technical Assistance. Fundamental Optics

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Guide to SPEX Optical Spectrometer

Opto Engineering S.r.l.

SPECTRAL SCANNER. Recycling

Using Optics to Optimize Your Machine Vision Application

Sharpness, Resolution and Interpolation

BEAM HALO OBSERVATION BY CORONAGRAPH

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Be aware that there is no universal notation for the various quantities.

Microscopy: Fundamental Principles and Practical Approaches

Image Formation and Capture. Acknowledgment: some figures by B. Curless, E. Hecht, W.J. Smith, B.K.P. Horn, and A. Theuwissen

Applied Optics. , Physics Department (Room #36-401) , ,

Supplementary Materials

ME 297 L4-2 Optical design flow Analysis

Imaging Optics Fundamentals

Use of Computer Generated Holograms for Testing Aspheric Optics

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

Astronomical Cameras

Chapter 17: Wave Optics. What is Light? The Models of Light 1/11/13

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

THE WASATCH ADVANTAGE

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

Transcription:

Better Imaging with a Schmidt-Czerny-Turner Spectrograph Abstract For years, images have been measured using Czerny-Turner (CT) design dispersive spectrographs. Optical aberrations inherent in the CT design give images that are blurred and have poor spatial resolution. The aberrations grow larger towards the edges of the focal plane, causing some researchers to abandon these regions of their sensors. A new dispersive spectrograph using a Schmidt-Czerny-Turner (SCT) design greatly reduces optical aberrations, giving sharp images with excellent spatial resolution across the entire focal plane and enabling researchers to use the entire CCD sensor to take images. The Czerny-Turner Spectrograph For decades, scientists have measured images and spectra using Czerny-Turner (CT) spectrographs. The imaging applications of this spectrograph include looking at multiple optical fibers, combustion processes, plasmas, and living cells. A schematic diagram of a CT spectrograph is seen in Figure 1. Figure 1. A schematic diagram of the traditional Czerny-Turner spectrograph (not to scale). In a CT spectrograph, light passes through an entrance slit, reflects off a collimating mirror, is dispersed by a diffraction grating, and is then brought to a focus by a focusing mirror at the focal plane. 1

Figure 2. A white light source imaged on the focal plane of a CT spectrograph through a bundle of 14 optical fibers (100 micron optical fibers, 1200 g/mm grating, first order, and backilluminated CCD). The optical aberration called field astigmatism occurs when lenses and mirrors are used to focus light off axis, which is the case for the CT spectrograph. Field astigmatism causes vertical and horizontal distortion of images as seen in Figure 2. In the center of Figure 2, note how the optical channels are thin, straight, and spatially well resolved. However, towards the left and right edges of the focal plane, the channels are broadened, distorted, and overlap because of field astigmatism, giving what is sometimes called the bow tie effect. At the very edges of the sensor, the spatial resolution is poor so there is significant crosstalk between the channels (i.e., photons from different optical fibers strike the same pixels on the CCD). The optical aberration called coma occurs when mirrors are used to image a source off axis. The distortion causes a tail to appear on images, similar to that on a comet. The net effect of the optical aberrations present in a CT spectrograph is seen in Figure 3. Note in Figure 3 that the optical fibers are well imaged in the center of the focal plane, but that along the top, bottom, left, and right edges of the sensor the images are distorted due to optical aberrations. This is why some researchers avoid using the edges of their CCD sensors. The net effect of these optical aberrations is a degradation of image crispness and a loss of spatial resolution. This makes it more difficult, for example, to follow a particle during a combustion process or visualize a fluorescing organelle inside a cell. 2

Figure 3. An image of a stack of 14 optical fibers stepped across a focal plane as measured by a Czerny-Turner spectrograph. Note the smearing of the image at the top, bottom, left, and right of the sensor caused by optical aberrations. (100 micron diameter core fibers, 435 nm source, 1200 g/mm grating, backilluminated CCD, zero order.) The Schmidt-Czerny-Turner Spectrograph The new Schmidt-Czerny-Turner (SCT) spectrograph from Princeton Instruments, the IsoPlane SCT 320, has a unique design that eliminates field astigmatism and reduces coma. The IsoPlane is shown below in Figure 4. Figure 4. The Princeton Instruments IsoPlane SCT 320 Schmidt-Czerny-Turner spectrograph shown with a PIXIS 400 CCD camera. 3

The IsoPlane has a 320 mm focal length and a three position on axis grating turret. Camera focus is optimized using a micrometer attached to the focusing mirror. By calculating the magnitudes of the optical aberrations present in the Czerny-Turner design a spectrograph can be built to compensate for them, which lead to the development of the IsoPlane. The IsoPlane has zero astigmatism at all wavelengths across the entire focal plane and has greatly reduced levels of coma, giving images that are sharp and well focused as seen in Figure 5. Figure 5. Top: Same image as Figure 2. Bottom: A white light source imaged on the focal plane of an SCT spectrograph through a bundle of 28 optical fibers (200 micron optical fibers, 1200 g/mm grating, first order, and back-illuminated CCD). Improved Imaging with the IsoPlane The IsoPlane has zero field astigmatism at all wavelengths and at all points across the focal plane. The resulting improved image quality is seen in Figure 5. The top of Figure 5 is the same image shown in Figure 2. Again, optical aberrations destroy spatial resolution causing the optical channels to overlap and create significant crosstalk, particularly at the focal plane edges. The bottom of Figure 5 shows 28 optical channels measured by the IsoPlane. The image is crisp and clear at all points on the focal plane. There is little crosstalk between the channels thanks to the reduction in optical aberrations. Figure 5 shows that an IsoPlane can easily image 28 channels while a CT spectrograph struggles with 14. 4

Another example of the imaging power of the IsoPlane is seen in Figure 6. Figure 6. Top: Same as Figure 3, a stack of optical fibers imaged with a CT spectrograph. Bottom: An image of the same optical fiber bundle seen in Figure 3, as measured with the IsoPlane. The top of Figure 6 is the same image shown in Figure 3. Again note how blurred the image is, particularly at the focal plane edges. The bottom of Figure 6 is an image of the same optical fiber bundle, obtained with an IsoPlane. Because of the greatly reduced levels of optical aberrations present in the IsoPlane, this image is sharp at the top, bottom, left, and right edges of the focal plane. 5

A final example of the improved imaging available with an IsoPlane is seen in Figure 7, which shows images of six round optical fibers taken from the upper left hand corner of a CCD. Figure 7. Images of six optical fibers taken from the upper left hand corner of a CCD. Left: Traditional Czerny-Turner. Right: IsoPlane. The optical aberrations inherent in the CT design cause the light from the optical fibers to become smeared out, giving an elliptical image at the focal plane. The image obtained with the IsoPlane is sharper and rounder. Conclusions Optical problems inherent in Czerny-Turner spectrographs include field astigmatism and coma. These aberrations distort images measured with CT spectrographs, particularly at the edges of the focal plane. The Schmidt-Czerny-Turner, or IsoPlane, spectrograph from Princeton Instruments has a unique optical design that completely eliminates field astigmatism at all wavelengths and at all points across the focal plane, and reduces coma to negligible levels. This means the IsoPlane gives sharp and spatially well resolved images across the entire CCD sensor. 6