Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses

Similar documents
*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes

Aberrations Before and After Implantation of an Aspheric IOL

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations

Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature

Dr. Magda Rau Eye Clinic Cham, Germany

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO

ROTATIONAL STABILITY MAKES THE DIFFERENCE

University of Groningen. Young eyes for elderly people van Gaalen, Kim

Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens

Causes of refractive error post premium IOL s 3/17/2015. Instruction course: Refining the Refractive Error After Premium IOL s.

Objective Analysis of Toric Intraocular Lenses. Affiliation (all authors): Aston University, School of Life and Health Sciences, Ophthalmic

Contrast Sensitivity after Refractive Lens Exchange with A Multifocal Diffractive Aspheric Intraocular Lens

Unique Aberration-Free IOL: A Vision that Patients

Long-term quality of vision is what every patient expects

IOL Types. Hazem Elbedewy. M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university

Raise your expectations. Deliver theirs.

Improving Lifestyle Vision. with Small Aperture Optics

Roadmap to presbyopic success

COMPARISON OF THE MEDICONTUR 860FAB

Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up

Visual Simulation: application to monofocal intraocular lens analysis

Postoperative Wavefront Analysis and Contrast Sensitivity of a Multifocal Apodized Diffractive IOL (ReSTOR) and Three Monofocal IOLs

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric.

Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation

Clinical Evaluation 3-month Follow-up Report

AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

Evolution of Diffractive Multifocal Intraocular Lenses

Comparison of Visual Acuity, Contrast Sensitivity and Spherical Aberration after Implantation of Aspheric and Spheric Intraocular Lenses

The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1

Prospective sual evaluation of apodized diffractive intraocular lenses

The Blueprint for Improved Image Quality

Comparison of Contrast Sensitivity, Higher- Order Aberrations and Subjective Visual Function after Different Aspheric Intraocular Lenses Implantation

University of Groningen. Young eyes for elderly people van Gaalen, Kim

NEW. AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

7 DO IT. A SAFER WAY TO TRIFOCALITY * : ELEVATED PHASE SHIFT (EPS) ** 7 DIFFRACTIVE RINGS FOR OPTIMAL LIGHT DISTRIBUTION AND LESS DISTURBANCE

Maximum Light Transmission. Pupil-independent Light Distribution. 3.75D Near Addition Improved Intermediate Vision

Comparison between clinical results of two diffractive multifocal lenses with the same platform but different additions

Clinical Update for Presbyopic Lens Options

Centre Hospitalier Universitaire et Psychiatrique de Mons-Borinage. B-Flex Multifocal. Dr Emmanuel Van Acker Belgium

Update on Aspheric IOL Technology

Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)?

Treatment of Presbyopia during Crystalline Lens Surgery A Review

Pichit Naripthaphan MD*, Pongsak Pachimkul MD*, Somporn Chantra MD*

Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects

SEE BEYOND WITH FULLRANGE OPTICS. Developed by Hanita Lenses

Effect of an aspheric intraocular lens on the ocular wave-front adjusted for pupil size and capsulorhexis size

Objective and subjective outcomes in comparing three different aspheric intraocular lens implants with their spherical counterparts

Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN

Refractive Power / Corneal Analyzer. OPD-Scan III

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12

Advanced Technology IOLs

Product Portfolio. Sulcoflex Pseudophakic Supplementary IOLs

Principles and clinical applications of ray-tracing aberrometry (Part II)

MODERN CATARACT SURGERY AND LENS REplacement

Evaluate Your Patient s Total Visual System With the OPD-Scan III

This is the author s version of a work that was submitted/accepted for publication in the following source:

Effects of Pupil Center Shift on Ocular Aberrations

THE XTRAFOCUS IS AN ELEGANT SOLUTION TO COMPLEX CASES.

Visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses with a D3.0 D addition

Maximising Treatment Outcomes with Premium IOL Technology. Saturday 13 September 2014 XXXII Congress of the ESCRS London, UK.

Product Portfolio. Sulcoflex Pseudophakic Supplementary IOLs. Your skill. Our vision.

Wave Front Topography. ReSeeVit Evolution Topography Module for Modi Topographer

Surgical data reveals that Q-Factor is important for good surgical outcome

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

Akreos AO Intraocular Lens

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses

American National Standard for Ophthalmics. Extended Depth of Focus Intraocular Lenses

The Aberration-Free IOL:

Special Publication: Ophthalmochirurgie Supplement 2/2009 (Original printed issue available in the German language)

Clinical Evaluation 3-month Follow-up Report

Optical Connection, Inc. and Ophthonix, Inc.

Role of Asphericity in Choice of IOLs for Cataract Surgery

4/2/2015. Bonnie An Henderson MD Clinical Professor of Ophthalmology Tufts University School of Medicine Ophthalmic Consultants of Boston

Accommodating IOL s History and Clinical Management

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12

PreciSAL Preloaded Lens System

LEAVE A LEGACY OF VISUAL FREEDOM. TECNIS PRESBYOPIA-CORRECTING IOLs

THE ASPHERIC PRELOADED INJECTION

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

TRANSLATIONAL SCIENCE. Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Visual Performance after AcrySof ReSTOR Aspheric Intraocular Lens Implantation

NEW THE WORLD S FIRST AND ONLY SINUSOIDAL TRIFOCAL IOL

Technicians & Nurses Program

LEAVE A LEGACY OF VISUAL FREEDOM. TECNIS PRESBYOPIA-CORRECTING IOLs

Customized intraocular lenses

Comparative Study Between the High Order Aberrations Before and After Cataract Surgery Using Two Different Types of IOLS

PATIENT SELECTION THE RIGHT PATIENT UNDERPROMISE AND OVERDELIVER THE PERFECT SPECTACLE FREE TREATMENT. Desires Less Dependence on glasses

HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design

Intraocular lens Difficulties

Design and qualification of a diffractive trifocal optical profile for intraocular lenses

VISUAL ACUITY AND IMAGE QUALITY IN FIVE DIFFRACTIVE. Short title: Visual acuity and image quality in five diffractive intraocular lenses

In Vitro Strehl Ratios with Spherical, Aberration-Free, and customized spherical aberration-correcting METHODS

CLINICAL SCIENCES. Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Were Implanted

Sutureless, Glueless, Scleral Fixation of Single-Piece and Toric Intraocular Lens: A Novel Technique

Clinical Study Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular Lenses

The Impact of New Generation Aspherical Soft Contact Lenses on Quality of Vision: A Comparison with Spherical Contact Lenses and Spectacle Correction

Transcription:

ARTICLE Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses Francisco Pastor-Pascual, PhD 1 ; Aitor Lanzagorta-Aresti, MD 1 ; Pablo Alcocer-Yuste, MD 1 ; Cristina Puchades-Lladró 1 ; José Luis Menezo-Rozalén, PhD 1 INTRODUCTION: Study aimed to compare Visual Acuity, ocular Spherical Aberration (SA) and Contrast Sensitivity (CS) in eyes implanted with the AcrySof toric IOL and AcrySof Natural IQ. METHODS: Fifty-eight eyes of 46 patients participated in this prospective randomized study. After phacoemulsification the eyes received the AcrySof toric or AcrySof IQ IOLs depending on their preoperative corneal astigmatism. Three month after surgery, all patients underwent complete ophthalmological examination including corneal topography, best corrected visual acuity, wavefront analysis for 6 mm pupil diameter and CS measurements with the CSV-1000HGT test under photopic (85 cd/m 2 ) and mesopic conditions (5 cd/m 2 ). RESULTS: BCVA in toric group was 0.054±0.043 logmar and 0.030±0.037 logmar in aspheric group (all patients achieved a BCVA of 20/25 or better). The mean postoperative SA was higher in the AcrySof toric group than the AcrySof IQ group (0.540±0.887 vs. 0.216±0.047 µm) and this difference was statistically significant (p<0.001). CS results showed no statistically significant differences between the values post-implantation at any lighting condition and spatial frequency (p>0.01). The AcrySof toric lens provided excellent rotational stability in the capsular bag with an average lens rotation of less than 3.3±2.5 degrees. CONCLUSIONS: The AcrySof toric IOL and aspheric AcrySof IQ yield similar and good visual outcomes. However, the AcrySof IQ IOL reduced significantly the SA for large pupils in relation to toric IOL. Surgeons should consider that patients may benefit from aspheric designs. Future studies, should consider direct comparison between spherical and aspheric toric IOLs to analyze properly the benefit in visual performance. KEYWORDS: Visual acuity, Contrast Sensitivity and Spherical Aberration with AcrySof Natural IQ and Toric IOLs. J Emmetropia 2010; 1: 68-73 Submitted: 5/23/2010 Revised: 6/7/2010 Accepted: 6/13/2010 1 Fundación Oftalmológica del Mediterráneo. Valencia. Spain. The authors have no proprietary interest in any of the materials mentioned in this article. The authors have no any financial support. Corresponding author: Francisco Pastor-Pascual. Fundación Oftalmológica del Mediterráneo. Bifurcación General Avilés-Pío Baroja s/n. 46015 Valencia. Spain. Telephone: +34 96 232 81 00. Fax: +34 96 232 81 02. E-mail: francisco_pastor@yahoo.es INTRODUCTION Approximately 15 20% of patients with cataracts have more than 1.5 diopters (D) of pre-existing astigmatism 1. Reducing this preoperative astigmatism can improve visual outcomes after cataract surgery. The astigmatism can be reduced with the phacoemulsification incisions, with corneal or limbal relaxing incisions and with excimer laser keratectomy. Another alternative for astigmatism correction in the cataract surgery is to implant a toric intraocular lens (IOL). Several reports have shown good outcomes in terms of safety, predictability, and efficacy after implantation of toric 68 2010 SECOIR Sociedad Española de Cirugía Ocular Implanto-Refractiva ISSN: 2171-4703

COMPARISON OF CONTRAST SENSITIVITY BETWEEN TORIC AND ASPHERIC IOLs 69 IOLs 2-8. Rotation of the toric IOL after implantation is the main problem associated with this modality 9-11. The majority of this occurs within the first three months after implantation 12. Even a small rotational deviation of the toric IOL from its intended axis can result in large reduction of the astigmatic correction 13. For example, a deviation of 10 degrees minimizes potential correction by 35%. The optical quality of the eye is determined by the combination of corneal and lens aberrations. Combination of both surfaces is important to achieve the best optical quality of the whole eye 14. There are aspheric designs to produce negative spherical aberration (SA) in order to compensate the positive SA of the cornea. However low-order optical aberrations should be corrected efficiently also to provide the best optical quality, and thus visual quality, for cataract patients. The purpose of this study is to assess and compare the visual performance in patients implanted after cataract surgery with two IOLs, a spherical toric IOL, the AcrySof toric SN60Tx and an aspheric IOL, the AcrySof IQ SN60WF. PATIENTS AND METHODS Study Design Fifty-eight eyes of 46 patients who underwent cataract extraction by uncomplicated phacoemulsification and implantation of the AcrySof toric SN60Tx or AcrySof IQ SN60WF, depending on their preoperative corneal astigmatism, were included. If it was larger than 1.50 D, the IOL implanted was the AcrySof toric IOL. If it was smaller than 1.50 D the IOL was the AcrySof IQ IOL. The study was performed at the Fundación Oftalmológica del Mediterráneo, Valencia, Spain, between February 2008 and July 2008. Inclusion criteria included cataract, age between 45 and 85 years, and preoperative regular corneal astigmatism smaller than 5.00 D. Exclusion criteria included preoperative astigmatism larger than 5.00 D, history of glaucoma or retinal detachment, corneal disease, previous corneal or intraocular surgery, macular degeneration or retinopathy, and history of ocular inflammation. The AcrySof Natural IQ SN60WF (Alcon Laboratories Inc, Ft Worth, TX) is an aspheric IOL with negative SA (-0.20 µm with a 6.0 mm pupil) designed with prolate posterior surface to reduce the total amount of ocular SA when implanted after cataract or clear lens extraction. The AcrySof toric IOL SN60Tx (Alcon Laboratories Inc, Ft Worth, TX) is an open-loop, modified single piece, acrylic polymer IOL with L-shaped haptics. The IOL s toricity, located on the posterior surface, extends across the entire surface of the spherical optic. The IOL is available in 3 models: SN60T3, SN60T4, and SN60T5. The U.S. Food and Drug Administration (FDA) approved the AcrySof toric IOL in September 2005 for 3 cylindrical powers: 1.50 D, 2.25 D, and 3.00 D at the IOL plane (Available at: http://www.fda.gov/cdrh/pdf/p930014s015.html. Accessed January 9, 2008). Before cataract surgery, patients had a complete ophthalmologic examination including manifest refractions, keratometry, topography (Pentacam system), slit lamp examination, applanation tonometry, and ocular fundus examination. Axial length and keratometry were measured with IOL Master (Carl Zeiss Meditec, Jena, Germany). The SRK/T formula was used for IOL power calculation. The targeted refraction was emmetropia. In the case of the Acrysof toric IOL was implanted the IOL model and axis placement was using a program available from the IOL manufacturer (www.acrysoftoriccalculator.com). Preoperative keratometry and biometry, data, incision location, and the surgeon s estimated surgically induced corneal astigmatism were used to determine the appropriate AcrySof toric IOL model, spherical equivalent lens power and axis of placement in the eye. Surgical Technique Two experienced surgeons (A.L.,F.P.) performed all surgeries using peribulbar anesthesia. With the patient seated at the slitlamp and with a coaxial thin slit turned to the 0- to 180-degree axis, the corneal limbus was marked at the 0-degree and 180-degree positions with a sterile marker after vertical alignment of the patient s head. Next, with the patient lying on the surgical table, the steep corneal meridian was identified and marked using a Marquez gauge with the aid of the preplaced reference points. Phacoemulsification was performed through a 2.75 mm temporal corneal incision. In all cases, the target capsulorhexis diameter was 5.50 mm to ensure overlap of the IOL border. After phacoemulsification, a foldable AcrySof toric IOL was inserted in the capsular bag using the Monarch II injector (Alcon Laboratories Inc, Ft Worth, TX). The IOL was rotated to align the cylinder axis with the marked steep corneal meridian. There were no complications in any of the cases. The tenets of the Declaration of Helsinki were followed in this research. Informed consent was obtained from all patients after the nature and possible consequences of the study were explained. Institutional Review Board approval was obtained. Visual Performance Measurements Best spectacle-corrected visual acuity (BCVA) measurement was performed by means of the logarithm of the minimum angle of resolution (logmar) acuity

70 COMPARISON OF CONTRAST SENSITIVITY BETWEEN TORIC AND ASPHERIC IOLs charts under photopic conditions (85 cd/m 2 ). Ocular wavefront aberrations were measured with the LADARWave Hartmann-Shack aberrometer (Alcon Laboratories Inc, Ft Worth, TX) for 6 mm pupil. Monocular photopic and mesopic contrast sensitivity (CS) was measured with best distance correction using CSV-1000HGT (VectorVision, Dayton, Ohio) in both groups at 3 month after implantation. Absolute values of log 10 CS were obtained for each combination of patient, spatial frequency, and luminance; means and standard deviations were calculated. The CS measurements were taken under 2 illumination conditions: photopic at 85 cd/m 2 (luminance recommended in manufacturer s guidelines) and mesopic at 5 cd/m 2. IOL rotation was measured at the slitlamp in 1-degree steps using an eyepiece for angle measurement through pupils dilated with tropicamide. A thin coaxial slit was projected in front of the eye and rotated until the thin slit projection overlapped the axis marks of the IOL. Data Analysis All examinations were performed at 3 month after implantation by one ophthalmic technician who was unaware of the objective of the study. To explore any correlation between the BCVA, CS at 2 lighting conditions and SA measured between the two IOLs t-test was carried out. Differences were considered to be statistically significant when the p-value was <0.01 (i.e., at the 1% level). RESULTS The mean ages of the AcrySof toric and AcrySof IQ groups were 65.1±7.78 and 63.7±4.78 years, respectively. Mean IOL power was 18.97±4.30D in the AcrySof toric group and 21.61±2.28 D in the AcrySof IQ group. Patient demographics are shown in Table 1. No statistically significant differences were found between both groups in relation to age, IOL power, and pre- and post- sphere (p>0.01). We found statistically significant differences between both groups for pre- and post-cylinder (p=4.48x10-13 and p=1.77x10-3, respectively). After the surgery and IOL implantation, the pupils of all patients were round, without iris trauma, and showed a good responsiveness to light. Visual Acuity In the toric group the BCVA was 0.054±0.043 logmar and in the aspheric group was 0.030±0.037 logmar (in both groups, all the patients achieved a BCVA of 20/25 or better). No statistically significant differences were found between both groups for BCVA (p=0.015). Contrast Sensitivity The mean values of log 10 CS after implantation with the AcrySof toric and the Acrysof IQ IOLs were plotted as a series of CS Functions (CSF) in Figure 1. Left and right images shows distance CSFs at the different luminance levels (85 and 5 cd/m 2 ). To explore the statistical significance of differences between lenses, t-tests were performed on the comparable data of the two Figure 1. Contrast sensitivity functions after implantation of the AcrySof toric IOL SN60Tx and AcrySof Natural IQ SN60WF under photopic (85 cd/m 2 ) and mesopic conditions (5 cd/m 2 ). Table 1: Demographic Characteristics of Participants AcrySof Toric IOL (SN60Tx) AcrySof IQ IOL (SN60WF) P value Number of eyes 31 27 Age (years) 65.1±7.78 63.7±4.78 0.08 Gender (Male/Female) 20/11 13/14 IOL Power (D) 18.97±4.30 21.61±2.28 0.09 Preoperative Sphere (D) -1.70±4.03-0.10±2.12 0.03 Preoperative Cylinder (D) -2.61±0.99-0.73±-0.44 4.48x10-13 Postoperative Sphere (D) 0.34±0.73 0.33±0.48 0.38 Postoperative Cylinder (D) -0.82±0.51-0.44±0.41 1.77x10-3 IOL = intraocular lens; D = diopters; means ± standard deviation.

COMPARISON OF CONTRAST SENSITIVITY BETWEEN TORIC AND ASPHERIC IOLs 71 groups (absolute log 10 CS values) at each spatial frequency and illumination level. The results showed no statistically significant differences between the values post-implantation at any lighting condition and spatial frequency (p>0.01). Spherical Aberration The mean SA for each group before and after surgery is summarized in Table 2. The mean postoperative SA for a 6mm pupil was higher in the AcrySof toric group than the AcrySof IQ (0.540±0.887 versus 0.216±0.047 µm). This difference was statistically significant (p<0.001; Table 2). Rotational Stability The AcrySof toric lens provided good rotational stability in the capsular bag with an average lens rotation of less than 3.3±2.5 degrees, verifying the placement with photographies. DISCUSSION Previous clinical trials evaluated clinical, optical and functional outcomes after spherical and aspheric IOLs 21-31, however this is the first study that compares toric IOLs with aspheric IOLs. Toric IOLs are a predictable and stable alternative for correcting astigmatism in patients with cataracts and preexisting corneal astigmatism 2-20. Success in achieving favourable postoperative visual function depends on several factors. First, patients with regular bowtie astigmatism may be benefit with toric IOL implantation. The second factor is to obtain accurate corneal astigmatism measurements to determine the actual amount of cylinder requiring correction and the spherical IOL power. The AcrySof toric IOL calculator program has been proved efficient for determining the toric IOL to be implanted and calculation of the implantation axis 3,16,18-20. The stability of the toric IOL is highly important to reduce the pre-existing cylindrical power. In our study the toric lens provided rotational stability in the capsular bag with an average lens rotation of less than 3.3±2.5 degrees from the lens initial placement at 3 months after surgery. This study did not evaluate the incidence of later postoperative rotation because previous trials found good long-term rotational stability of toric IOLs 7,15-17. Our results revealed that both IOLs provided good BCVA (about 20/20). Mendicute et al. 3 using the AcrySof toric reported similar values and percentages of binocular BCVA in 30 eyes at 3 month post-surgery (mean BCVA of 0.02±0.05 logmar; 100% of the patients with 20/25 or better). Zuberbuhler et al. 16 on a sample of 44 eyes implanted with the same IOL at 3 months found a mean BCVA of 0.01±0.11 logmar (about 20/20); and recently, Chang 19 reported that the 94% of a sample of 50 patients achieved a BCVA of 20/40 or better. Cadarso et al.(29) at 6 months and Mester et al.(30) at 6 weeks in 30 and 52 eyes implanted with the AcrySof IQ aspheric IOL, respectively, reported that 100% of the patients achieved BCVA of 20/25 or better (logmar 0.1 and -0.03±0.15, respectively). Different authors 21-24,26-30 compared the results of BCVA in patients implanted with the AcrySof IQ aspheric IOL with those implanted with the AcrySof Natural spherical IOL and none found differences statistically significant between both IOLs. Probably the visual performance metrics used were not accurate enough to detect subtle visual changes resulting from spherical aberration reduction by asphericity. In relation to the CS our results revealed photopic CSFs in both groups within the normal range (see Figure 1). Comparing both lenses we found similar photopic CS for the toric and aspheric IOLs at 3 month post-surgery without statistically significant differences at any spatial frequency analysed (p>0.01). From seven studies reporting photopic CS (85 cd/m 2 ) with the AcrySof IQ IOL, 3 didn t find statistically significant differences at any spatial frequency compared with the SN60AT IOL 21,24,25 and 4 revealed differences 23,26,27,30 at 3 cpd 30, 6 cpd 23,30, 12 cpd 23, and 18 cpd 23,26,27. Six studies evaluated the AcrySof IQ compared with SN60AT under mesopic conditions 21,23,26-28. All found significant differences specially at 1.5 cpd 23,26,27,30, 3 cpd 21,23,26-28, 6 cpd 23,29, 12 cpd 23,26-28, and 18 cpd 21,23,26-29. Our CS results with AcrySof IQ were similar to the others but the differences in conditions and measure systems between studies may play a role in the differences found. No comparison with previous studies about the CS of the Table 2: Ocular spherical aberration values Z(4,0) with AcrySof toric (SN60Tx) and AcrySof IQ (SN60WF) IOLs for 6 mm of pupil AcrySof Toric IOL (SN60Tx) AcrySof IQ IOL (SN60WF) P value Preoperative Spherical aberration Z(4,0) 0.243 ± 0.225 0.301 ± 0.114 0.041 Postoperative Spherical aberration Z(4,0) 0.540 ± 0.887 0.216 ± 0.047 3.021 x 10-5 IOL = intraocular lens; means (µm) ± standard deviation.

72 COMPARISON OF CONTRAST SENSITIVITY BETWEEN TORIC AND ASPHERIC IOLs AcrySof toric IOL is possible because this is the first study that evaluates it. The ocular SA for a 6 mm pupil found was 0.216±0.047 µm for the AcrySof IQ group and 0.540±0.887 µm for the AcrySof Toric group (p=3.021 x 10-5 ). Rocha et al. 21 showed a mean value of SA of 0.24 ±0.04 µm for a 5 mm pupil at 3 months with the AcrySof IQ. Cadarso et al. 29 found an ocular SA of 0.114±0.147 µm in 20 eyes for a 6 mm pupil at 6 months implanted with the AcrySof IQ. No comparison with previous studies about the SA of the AcrySof toric IOL is possible because this is the first study that evaluates it. We have found that the SA was significantly lower in those eyes implanted with the AcrySof Natural IQ IOL. Marcos et al. 31 found that the tolerance to defocus was significantly higher with spherical IOLs than with aspherical IOLs; it was necessary to add 1.5 D with the spherical IOL and 1.1 D with the aspherical IOL to make the 20/20 line illegible on simulation. Recently, Rocha et al. 22 concluded that the reduction in SA after aspherical IOL implantation may degrade distance-corrected near visual acuity and intermediate visual acuity. They point out that residual SA can improve depth of focus and that the tolerance to defocus seems to be higher in eyes with a spherical IOL than in eyes with an aspherical IOL. However, it must be kept in mind that this refers to a monofocal IOL that was not designed to provide a large depth of focus, as are pseudoaccommodating IOLs. From a theoretical point of view, an aspheric IOL design improves the optical quality and therefore, the visual performance in a patient. However, in our study, the SA reduction achieved in those eyes implanted with the aspheric IOL does not correlate with an improvement in the mesopic CSF. We have to note that we didn t control the real pupil size of our patients and then differences among eyes may overlap the role of the aspheric design for large pupils. SA was measured for a 6 mm pupil but we cannot be sure that the real pupil size of those patients would be of 6 mm under mesopic conditions. Alfonso et al. 32 measured the pupil size in 650 eyes (61.2±9.3 years old) under the same lighting conditions (5 cd/m 2 ) reporting a mean value of 5.4±0.7 mm. 0.6 mm of difference from this mean value and the pupil size used to compute the SA in our study may be responsible to not report differences in the mesopic CSF between both lenses. In conclusion, both AcrySof spherical toric IOL and AcrySof aspheric IOL provided good visual acuity results. Lower SA was found in those eyes implanted with the aspheric IOL in relation to the spherical model, although, no differences were achieved in visual acuity, and photopic and mesopic CS between lenses. Surgeons should consider that patients may benefit from aspheric designs (i.e. reduction of the central thickness of the lens). Custom selection of the asphericity of the lens for coupling with the corneal SA may reduce highly the ocular SA and then expect a visual performance improvement in those patients implanted with aspheric IOLs. Future studies, should consider direct comparison between spherical and aspheric toric IOLs to analyze properly the benefit in visual performance. SYNOPSIS Acrysof IQ IOL reduces the spherical aberration more than Acrysof Toric IOL but no differences were achieved in Visual Acuity, photopic and mesopic contrast sensitivity between lenses. REFERENCES 1. Hoffer KJ. Biometry of 7500 cataractous eyes. Am J Ophthalmol 1980; 90: 360-368; correction 890. 2. López-Gil N, Montes-Micó R. New intraocular lens for achromatizing the human eye. J Cataract Refract Surg 2007; 33: 1296-1302. 3. Mendicute J, Irigoyen C, Aramberri J, Ondarra A, Montés- Micó R. Foldable toric intraocular lens for astigmatism correction in cataract patients. J Cataract Refract Surg. 2008; 34: 601-7. 4. Grabow HB. Early results with foldable toric IOL implantation. Eur J Implant Refract Surg 1994; 6: 177-178. 5. Werblin TP. Do three-piece PMMA IOLs rotate after implantation in the capsular bag? J Refract Surg 1995; 11: 468-471. 6. Sun X-Y, Vicary D, Montgomery P, Griffiths M. Toric intraocular lenses for correcting astigmatism in 130 eyes. Ophthalmology 2000; 107: 1776-1781. 7. Ruhswurm I, Scholz U, Zehetmayer M, et al. Astigmatism correction with a foldable toric intraocular lens in cataract patients. J Cataract Refract Surg 2000; 26: 1022-1027. 8. Leyland M, Zinicola E, Bloom P, Lee N. Prospective evaluation of a plate haptic toric intraocular lens. Eye 2001; 15: 202-205. 9. Gills JP, Van der Karr M, Cherchio M. Combined toric intraocular lens implantation and relaxing incisions to reduce high preexisting astigmatism. J Cataract Refract Surg 2002; 28: 1585-1588. 10. Viestenz A, Seitz B, Langenbucher A. Evaluating the eye s rotational stability during standard photography; effect of determining the axial orientation of toric intraocular lenses. J Cataract Refract Surg 2005; 31: 557-561. 11. Weinand F, Jung A, Stein A, et al. Rotational stability of a singlepiece hydrophobic acrylic intraocular lens: new method for highprecision rotation control. J Cataract Refract Surg 2007; 33: 800-803. 12. Strenn K, Menapace R, Vass C: Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring. J Cataract Refract Surg 1997; 23: 1543-1547. 13. Sanders DR, Grabow HB, Shepherd J: The toric IOL. In Sutureless Cataract Surgery; An Evolution Toward Minimally Invasive Technique Edited by: Gills JP, Martin RG, Sanders DR. Thorofare, NJ, Slack; 1992; 183-197. 14. He JC, Gwiazda J, Thorn F, Held R. Wave-front aberrations in the anterior corneal surface and the whole eye. J Opt Soc Am A Opt Image Sci Vis. 2003; 20: 1155-1163. 15. Sun X-Y, Vicary D, Montgomery P, Griffiths M. Toric intraocular lenses for correcting astigmatism in 130 eyes.

COMPARISON OF CONTRAST SENSITIVITY BETWEEN TORIC AND ASPHERIC IOLs 73 Ophthalmology 2000; 107:1776 1781; discussion by RM Kershner, 1781-1782. 16. Zuberbuhler B, Signer T, Gale R, Haefliger E. Rotational stability of the AcrySof SA60TT toric intraocular lenses: A cohort study. BMCOphthalmology 2008; 8: 8. 17. Patel CK, Ormonde S, Rosen PH, Bron AJ. Postoperative intraocular lens rotation; a randomized comparison of plate and loop haptic implants. Ophthalmology 1999; 106: 2190-2196; discussion by DJ Apple, 219. 18. Bauer N, De Vries N, Webers C, Hendrikse F, Nuijts R, Astigmatism management in cataract surgery with the AcrySof toric intraocular lens. J Cataract Refract Surg 2008; 34: 1483-1488. 19. Chang DF. Comparative rotational stability of single-piece open-loop acrylic and plate-haptic silicone toric intraocular lenses. J Cataract Refract Surg 2008; 34: 1842-1847. 20. Hill W. Expected effects of surgically induced astigmatism on AcrySof toric intraocular lens results. J Cataract Refract Surg 2008; 34: 364-367. 21. Rocha KM, Soriano ES, Chalita MR, Yamada AC, Bottós K, Bottós J, Morimoto L, Nose W. Wavefront analysis and contrast sensitivity of aspheric and spherical intraocular lenses: a randomized prospective study. Am J Ophthalmol 2006; 142: 750-756. 22. Rocha KM, Soriano ES, Chamon W, Chalita MR, Nose W. Spherical aberration and depth of focus in eyes implanted with aspheric and spherical intraocular lenses; a prospective randomized study. Ophthalmology 2007; 114: 2050-2054. 23. Caporossi A, Martone G, Casprini F, Rapisarda L. Prospective randomized study of clinical performance of 3 aspheric and 2 spherical intraocular lenses in 250 eyes. J Refract Surg 2007; 23: 639-648. 24. Sandoval HP, Fernández de Castro LE, Vroman DT, Solomon KD. Comparison of visual outcomes, photopic contrast sensitivity, wavefront analysis, and patient satisfaction following cataract extraction and IOL implantation: aspheric vs spherical acrylic lenses. In press, Eye 2008. 25. Awwad ST, Lehmann JD, McCulley JP, Bowman RW. A comparison of higher order aberrations in eyes implanted with Acry-Sof IQ SN60WF and AcrySof SN60AT intraocular lenses. Eur J Ophthalmol 2007; 17: 320-326. 26. Pandita D, Raj SM, Vasavada VA, Vasavada VA, Kazi NS, Vasavada AR. Contrast sensitivity and glare disability after implantation of AcrySof IQ Natural aspherical intraocular lens: prospective randomized masked clinical trial. J Cataract Refract Surg 2007; 33: 603-610 27. Tzelikis PF, Akaishi L, Trindade FC, Boteon JE. Ocular aberrations and contrast sensitivity after cataract surgery with AcrySof IQ intraocular lens implantation; clinical comparative study. J Cataract Refract Surg 2007; 33: 1918-1924. 28. Awwad ST, Warmerdam D, Bowman RW, Dwarakanathan S, Cavanagh HD, McCulley JP. Contrast sensitivity and higher order aberrations in eyes implanted with Acrysof IQ SN60WF and AcrySof SN60AT intraocular lenses. J Refract Surg 2008; 24: 619-625. 29. Cadarso L, Iglesias A, Ollero A, Pita B, Montés-Micó R. Postoperative optical aberrations in eyes implanted with AcrySof spherical and aspheric intraocular lenses. J Refract Surg. 2008; 24: 811-816. 30. Mester U, Kaymak H. Comparison of the AcrySof IQ aspheric blue light filter and the Acrysof SA60AT intraocular lenses. J Refract Surg. 2008; 24: 817-825. 31. Marcos S, Barbero S, Jiménez-Alfaro I. Optical quality and depth-of-field of eyes implanted with spherical and aspheric intraocular lenses. J Refract Surg 2005; 21: 223-235. 32. Alfonso JF, Fernández-Vega L, Baamonde B, Montés-Micó R. Prospective visual evaluation of apodized diffractive intraocular lenses. J Cataract Refract Surg. 2007; 33: 1235-1243. First author: Francisco Pastor-Pascual, PhD Fundación Oftalmológica del Mediterráneo. Valencia, Spain