FLATS: SBC INTERNAL LAMP P-FLAT

Similar documents
SBC Internal Lamp P-flat Monitoring

HRC AND WFC FLAT FIELDS: DISPERSORS, ANOMALIES, AND PHOTOMETRIC STABILITY

FLAT FIELD DETERMINATIONS USING AN ISOLATED POINT SOURCE

FLAT FIELDS FOR FILTER WHEEL OFFSET POSITIONS

Wavelength Calibration Accuracy of the First-Order CCD Modes Using the E1 Aperture

WFC3 SMOV Proposal 11422/ 11529: UVIS SOFA and Lamp Checks

Pre-Launch NUV MAMA Flats

The Flat Fielding and Achievable Signal-to-Noise of the MAMA Detectors 1

UVIS 2.0: Chip-Dependent Flats

CCD reductions techniques

WFC3/IR Cycle 19 Bad Pixel Table Update

WFC3 TV3 Testing: IR Channel Nonlinearity Correction

WFC3 Thermal Vacuum Testing: UVIS Science Performance Monitor

SBC Dark and Cumulative Images

Improvements to the STIS First Order Spectroscopic Point Source Flux Calibration

WFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields

WFC Zeropoints at -80C

COS: NUV and FUV Detector Flat Field Status

WFC3 TV2 Testing: UVIS Filtered Throughput

Determination of the STIS CCD Gain

Baseline Tests for the Advanced Camera for Surveys Astronomer s Proposal Tool Exposure Time Calculator

STIS CCD Saturation Effects

COS Near-UV Flat Fields and High S/N Determination from SMOV Data

Flux Calibration of the ACS CCD Cameras III. Sensitivity Changes over Time

WFC3/IR Channel Behavior: Dark Current, Bad Pixels, and Count Non-Linearity

WFC3 SMOV Program 11433: IR Internal Flat Field Observations

Making a Panoramic Digital Image of the Entire Northern Sky

Temperature Dependent Dark Reference Files: Linear Dark and Amplifier Glow Components

WFC3 UVIS Ground P-flats

WFC3/IR Bad Pixel Table: Update Using Cycle 17 Data

Cross-Talk in the ACS WFC Detectors. II: Using GAIN=2 to Minimize the Effect

WFC3 Post-Flash Calibration

Use of the Shutter Blade Side A for UVIS Short Exposures

New Bad Pixel Mask Reference Files for the Post-NCS Era

SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA

Interpixel Capacitance in the IR Channel: Measurements Made On Orbit

A Test of non-standard Gain Settings for the NICMOS Detectors

Railroad Valley Playa for use in vicarious calibration of large footprint sensors

Photometric Aperture Corrections for the ACS/SBC

Astronomy 341 Fall 2012 Observational Astronomy Haverford College. CCD Terminology

Flux Calibration Monitoring: WFC3/IR G102 and G141 Grisms

New Bad Pixel Mask Reference Files for the Post-NCS Era

Master sky images for the WFC3 G102 and G141 grisms

No Evidence Found for WFC3/UVIS QE Overshoot

WFC3 SMOV Program 11427: UVIS Channel Shutter Shading

WFC3/UVIS Sky Backgrounds

WFC3 TV2 Testing: UVIS Shutter Stability and Accuracy

2017 Update on the WFC3/UVIS Stability and Contamination Monitor

Temperature Reductions to Mitigate the WF4 Anomaly

Assessing ACS/WFC Sky Backgrounds

Camera Calibration Certificate No: DMC III 27542

WFC3/UVIS Updated 2017 Chip- Dependent Inverse Sensitivity Values

STIS Cycle 7 Calibration Close-out Report

Cerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf)

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

The predicted performance of the ACS coronagraph

Anomalies and Artifacts of the WFC3 UVIS and IR Detectors: An Overview

This release contains deep Y-band images of the UDS field and the extracted source catalogue.

F/48 Slit Spectroscopy

ON-ORBIT FLAT FIELDS AND ABSOLUTE CALIBRATION OF STIS

Module 10 : Receiver Noise and Bit Error Ratio

Camera Calibration Certificate No: DMC IIe

Observing*Checklist:*A3ernoon*

WFC3/UVIS TV3 Post-flash Results

Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming)

Southern African Large Telescope. RSS CCD Geometry

Overview of the WFC3 Cycle 17 Detector Monitoring Campaign

STIS CCD Anneals. 1. Introduction. Instrument Science Report STIS Revision A

Resampling in hyperspectral cameras as an alternative to correcting keystone in hardware, with focus on benefits for optical design and data quality

ACS/WFC: Differential CTE corrections for Photometry and Astrometry from non-drizzled images

Camera Calibration Certificate No: DMC II

Camera Calibration Certificate No: DMC II

Camera Calibration Certificate No: DMC II

Camera Calibration Certificate No: DMC II

STARDUST NAVCAM DATA

Camera Calibration Certificate No: DMC II Aero Photo Europe Investigation

Camera Calibration Certificate No: DMC II

On spatial resolution

Calibrating VISTA Data

Cycle 24 HST+COS Target Acquisition Monitor Summary

WFPC2 Status and Plans

Selecting the NIR detectors for Euclid

Advanced Camera for Surveys Instrument Handbook for Cycle 11

STIS Cycle 10 Calibration Close-out Report

M67 Cluster Photometry

OmegaCAM calibrations for KiDS

Update to the WFPC2 Instrument Handbook for Cycle 9

PACS SED and large range scan AOT release note PACS SED and large range scan AOT release note

LANDSAT 8 Level 1 Product Performance

Advanced Camera for Surveys Instrument Handbook for Cycle 12

WISE Photometry (WPHOT)

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

The 0.84 m Telescope OAN/SPM - BC, Mexico

Advanced Camera for Surveys Instrument Handbook for Cycle 13

High Contrast Imaging using WFC3/IR

PACS photometry on extended sources

Post-Flash Calibration Darks for the Advanced Camera for Surveys Wide Field Channel (ACS/WFC)

ENGINEERING CHANGE ORDER ECO No. COS-057 Center for Astrophysics & Space Astronomy Date 13 February 2001 University of Colorado, Boulder Sheet 1 of 6

Properties of a Detector

NIRSpec Technical Note NTN / ESA-JWST-TN Authors: G. Giardino, S. Birkmann, M. Sirianni Date of Issue: 9 Nov Version: 1.

Transcription:

Instrument Science Report ACS 2005-04 FLATS: SBC INTERNAL LAMP P-FLAT R. C. Bohlin & J. Mack May 2005 ABSTRACT The internal deuterium lamp was used to illuminate the SBC detector through the F125LP filter for 20.5 hours and produced about 12,000 counts/pixel. This illumination does not simulate the OTA optics and, therefore, is not suitable for the production of a low frequency L-flat. However, the pixel-to-pixel P-flat is an improvement over the laboratory SBC flat currently used in the ACS pipeline processing for the six SBC imaging filters. 1. Introduction The full low-frequency and pixel-to-pixel high-frequency LP-flat field correction for the SBC filters on ACS are difficult to construct. There was no attempt in the lab to simulate flat field OTA illumination, and the bright earth cannot be observed. Sufficiently dense star fields with appropriate magnitudes are hard to find for deriving the on-orbit L-flat, and the only significant contribution for a sky flat is from the geocoronal Lyman-alpha line that has only ~1 count/min per pixel in F115LP. The internal deuterium lamp does produce a smoothly varying illumination pattern that can be used to define a P-flat. Extensively dithered observation of ~50 stars in NGC6681 are currently the best available data for defining L-flats for the SBC imaging modes. This ISR addresses the creation of a new P-flat for the SBC. 2. Data 1

The internal lamp exposure times ranged from 780 to 2500s and were obtained in programs 9562, 9657, and 10047 over the time period of 2002-06-11 to 2004-09-09. The total exposure time is 20.5 hours, which resulted in ~12,000 total counts/pixel near the center of the SBC MAMA detector. Figure 1 shows the degradation of the lamp output vs. exposure time. Based on pre-launch testing, the declining lamp output is probably caused by polymerization of contaminants on the lamp window. The 37 individual exposures through the F125LP filter are co-added and fit following the procedures summarized in Bohlin et al. (1999) and Bohlin, Hartig, and Meurer (BHM 1999). The IDL routine flatall along with its main subroutines flats and fitflat are used to produce the P-flat by fitting the overall vignetting pattern with low order splines. This internal vignetting of the lamp illumination is about 10% from image center to the edges, while the P-flat is equal to the original co-added image divided by the fitted pattern. The data quality flags in the third extension of the flat field file are all zero, except for values of 512 around the edges and in rows 599:604 with dead MAMA diodes. Figure 1 - Relative count rate vs. total time of illumination of the SBC detector by the internal deuterium flat field lamp. 2

3. Results 3

Figure 2 - Ratio of the on-orbit SBC P-flat to the laboratory P-flat. The name of the lab flat, psbcex99marsm77super, is the sum 77 images with external illumination obtained in March of 1999 and is also know as jref$m5t12223j_pfl in the ACS pipeline processing system. The grey scale calibration is indicated on the reference bar at the top right. The line of text "P-flat rms(%)= " indicates onesigma values in the 101x101 pixel standard region (dashed box) for the total pixelto-pixel scatter, the Poisson counting statistic, and the intrinsic rms variation of the ratio. The horizontal grey strip is the region of bad anodes, where the flats are set to unity. The vertical feature near the center is caused by the slightly different illumination in the region of the repeller wire. There are slight differences in the steep drop of the flat field at the extreme right hand edge where the vertical white and black regions appear. The ratio of the new to the pre-flight P-flat is illustrated in Figure 2, while these differences are quantified more precisely in Table 1 with the program ratstat.pro. This ratio of new/old is divided into 8x8 128x128 pixel blocks, and Table 1 contains the statistical analysis for the 16384 pixels in each block. Columns in Table 1 correspond to successive blocks in x on the image, while each set of three rows corresponds to blocks in y where the top of the Table corresponds to the top of the image. The three numbers for each block in Table 1 are from top to bottom: the expected rms scatter due to counting statistics, the actual rms scatter, and the residual scatter attributed to actual systematic differences between the two flats. The right hand column should be ignored, because the larger scatter there is caused by slight differences in the steep drop in the flats, as seen at the right side of Figure 2. Ignoring all 30 of the boxes around the edges, the actual differences average 1.3% and range from 1.0 to 1.6%. Thus, the in-flight flat differs from the lab flat by more than the typical ~0.9% Poisson uncertainty in the new flat. The old flat has a better statistical uncertainty of ~0.4%; but the 0.9% Poisson uncertainty of the new flat will begin to contribute significantly to the final uncertainty only for observations with more than ~10,000 counts/pixel. Evidence for stability of the on-orbit SBC flat field is that the ratio of the first half to the second half of the on-orbit exposures are identical to 0.3% rms per pixel after correcting for the Poisson statistics. The first half of the total counts covers the interval 2002 June to 2003 March, while the second half includes 2003 April to 2004 September. 4

Table 1. RESIDUAL % RMS DEVIATIONS FOR NEW/OLD 1.04 1.03 1.03 1.02 1.02 1.03 1.04 1.08 1.95 1.98 2.13 2.06 2.05 2.00 1.93 2.67 1.65 1.70 1.87 1.79 1.78 1.71 1.63 2.45 1.02 1.00 0.99 1.00 1.00 1.01 1.02 1.07 1.85 1.85 1.77 1.72 1.76 1.74 1.73 2.81 1.54 1.56 1.47 1.40 1.45 1.42 1.40 2.60 1.02 0.99 0.99 0.99 0.99 1.00 1.01 1.06 1.75 1.74 1.67 1.66 1.66 1.70 1.67 2.62 1.42 1.43 1.34 1.34 1.32 1.37 1.33 2.40 1.02 0.99 0.99 0.99 0.99 1.00 1.01 1.05 1.69 1.72 1.72 1.68 1.63 1.69 1.63 2.42 1.35 1.41 1.41 1.36 1.30 1.36 1.29 2.18 1.03 0.99 0.99 0.99 0.99 1.00 1.01 1.06 1.63 1.58 1.59 1.62 1.62 1.65 1.65 2.29 1.26 1.23 1.25 1.29 1.28 1.31 1.30 2.03 1.03 1.00 0.99 0.99 1.00 1.00 1.01 1.06 1.51 1.49 1.53 1.61 1.66 1.66 1.69 2.30 1.10 1.11 1.16 1.26 1.33 1.33 1.36 2.04 1.04 1.01 1.00 1.00 1.01 1.01 1.02 1.07 1.35 1.40 1.50 1.62 1.73 1.85 1.90 2.51 0.87 0.97 1.12 1.27 1.41 1.55 1.60 2.27 1.06 1.03 1.02 1.02 1.02 1.02 1.03 1.08 1.36 1.36 1.47 1.66 1.81 1.95 2.11 2.68 0.85 0.90 1.07 1.31 1.50 1.66 1.84 2.46 Order: Poisson, actual, residual sigmas Central 36 bins: Average residual & rms= 1.34 0.12 Because BHM found that the SBC P-flat for the imaging modes is independent of wavelength to <0.5% and averaged the P-flats from five filters to make the original pipeline super flat, the new F125LP P-flat should be used for all six SBC filters for future ACS pipeline data processing. Errors in the existing pre-flight SBC prism P-flat should be comparable to the typical 1.3% differences between the pre-flight P-flat and the new internal lamp P-flat. 5

4. Future Observations Even though the P-flat is independent of wavelength, the L-flat may not be. The deuterium lamp illumination of the detector-filter combination should be the same for all six filters. Thus, the internal flat field lamp can be used to find differences in the L- flat with wavelength. In cycle 14, short exposures of 2500s in each of the filters require only about 4hrs of burn time and would be sufficient to define the delta L-flats relative to the existing F125LP data set. These delta L-flats will help us understand the anomalous ratio of F165LP/F140LP Jupiter images from proposal 10140 reported by C. Proffitt. In addition, the ratios of L-flats for the six filters that will be derived from the NGC6681 observations can be checked; and perhaps, a more robust L-flat can be computed for any filter that may have a poor NGC6681 data set. The last set of these six 2500s internal observations should be for F125LP to verify that there is no change from the data discussed here. The prism P-flat differs from the imaging filters because the illumination has a different angle of incidence (BHM). Because the P-flat for the prisms has probably changed by an amount that is comparable to the change observed from ground to space for the imaging filters, an additional set of internal lamp exposures are required to define a P- flat appropriate for on-orbit use for the prism modes. About 10,000 counts/pixel are required in cycle 14 and will be split equally between the two prism modes to verify that their P-flats are still identical. The count rates measured in the lab for the internal lamp are in Table 2 and correspond within a few percent to the on-orbit rates for F115LP and F125LP. A new prism P-flat can be obtained with about 23 hours of exposure time split equally between PR110L and PR130L, respectively. Following the proposed total 27 hours burn time in cycle 14, the internal deuterium lamp should still have ~68% of its brightness after about four years in space. Table 2. Pre-launch Count Rates per Pixel for the SBC Internal Deuterium Lamp F115LP 0.185 F125LP 0.17 F140LP 0.16 F150LP 0.15 F165LP 0.03 F122M 0.004 PR110L 0.17 PR130L 0.17 6

REFERENCES Bohlin, R., Hartig, G., Lindler, D., Meurer, G., & Cox, C. 1999, Instrument Science Report, ACS 99-01, (Baltimore:STScI) Bohlin, R., Hartig, G., & Meurer, G. 1999, Instrument Science Report, ACS 99-02, (Baltimore:STScI), BHM 7