X-RAY COMPUTED TOMOGRAPHY

Similar documents
Principles of CT scan

Reconstruction Filtering in Industrial gamma-ray CT Application

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation

Alignment of the camera

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

TEP Principles of Digital X-ray Imaging

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

An Activity in Computed Tomography

Digital Radiology with Photon Counting Detectors

ME 6406 MACHINE VISION. Georgia Institute of Technology

Operating Procedures for MICROCT1 Nikon XTH 225 ST

Applications of the Cracow X-Ray Microprobe in Tomography

NOT FOR DISTRIBUTION JINST_128P_1010 v2

Real Time Linear Array Imaging. Brian Caccamise

Digital Radiography : Flat Panel

An Activity in Computed Tomography

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Sensors and Sensing Cameras and Camera Calibration

Single Photon Interference Katelynn Sharma and Garrett West University of Rochester, Institute of Optics, 275 Hutchison Rd. Rochester, NY 14627

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Fig.2: Scanner VistaScan for image plates

Charged Coupled Device (CCD) S.Vidhya

Synchrotron X-ray tomographic microscopy Theory vs. practice

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y

High Energy Digital Radiography & 3D-CT for Industrial Systems

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor

Control of Noise and Background in Scientific CMOS Technology

RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING

Figure 1 HDR image fusion example

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

Towards accurate measurements with synchrotron tomography Problems and pitfalls. Robert C. Atwood. Nghia T. Vo, Michael Drakopoulos, Thomas Connolley

Distortion Correction in LODOX StatScan X-Ray Images

X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P )

Advanced Camera and Image Sensor Technology. Steve Kinney Imaging Professional Camera Link Chairman

Geometric Functions. The color channel toolbar buttons are disabled.

MY ASTROPHOTOGRAPHY WORKFLOW Scott J. Davis June 21, 2012

TEP Optimization of the CT scan quality. Related topics Number of projections, stop angle, exposure time, image statistics, binning

HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY

Very short introduction to light microscopy and digital imaging

System and method for subtracting dark noise from an image using an estimated dark noise scale factor

Camera Image Processing Pipeline

Philip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany.

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14

a simple optical imager

NEUROIMAGING DATA ANALYSIS SOFTWARE

Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

AST Lab exercise: CCD

COMPUTED TOMOGRAPHY 1

Time Delay Integration (TDI), The Answer to Demands for Increasing Frame Rate/Sensitivity? Craige Palmer Assistant Sales Manager

The design and testing of a small scale solar flux measurement system for central receiver plant

Hardware for High Energy Applications 30 October 2009

Collimation Assessment Using GAFCHROMIC XR-M2

EMVA1288 compliant Interpolation Algorithm

A Short History of Using Cameras for Weld Monitoring

Patents of eye tracking system- a survey

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Time of Flight Measurement System using Time to Digital Converter (TDC7200)

Backgrounds in DMTPC. Thomas Caldwell. Massachusetts Institute of Technology DMTPC Collaboration

Imaging Photometer and Colorimeter

Projection Data and Scanograms in X-ray Computed Tomography

The 0.84 m Telescope OAN/SPM - BC, Mexico

Moving from biomedical to industrial applications: OCT Enables Hi-Res ND Depth Analysis

An Electronic Eye to Improve Efficiency of Cut Tile Measuring Function

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

Data Sheet SMX-160 Series USB2.0 Cameras

Pixel Response Effects on CCD Camera Gain Calibration

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Camera Calibration Certificate No: DMC III 27542

Highly Miniaturised Radiation Monitor (HMRM) Status Report. Yulia Bogdanova, Nicola Guerrini, Ben Marsh, Simon Woodward, Rain Irshad

Cameras CS / ECE 181B

Image Processing for Comets

Exercise questions for Machine vision

ScanArray Overview. Principle of Operation. Instrument Components

ULS24 Frequently Asked Questions

USB Line Camera 8M. Coptonix GmbH

FTA SI-640 High Speed Camera Installation and Use

v tome x m microfocus CT

Photometry. Variable Star Photometry

PLazeR. a planar laser rangefinder. Robert Ying (ry2242) Derek Xingzhou He (xh2187) Peiqian Li (pl2521) Minh Trang Nguyen (mnn2108)

The Noise about Noise

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

A machine vision system for scanner-based laser welding of polymers

CCD reductions techniques

Pixel CCD RASNIK. Kevan S Hashemi and James R Bensinger Brandeis University May 1997

Digital Industrial Radiography

6.098/6.882 Computational Photography 1. Problem Set 1. Assigned: Feb 9, 2006 Due: Feb 23, 2006

X-rays in medical diagnostics

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

QC Testing for Computed Tomography (CT) Scanner

A 120dB dynamic range image sensor with single readout using in pixel HDR

PD233: Design of Biomedical Devices and Systems

Camera Overview. Digital Microscope Cameras for Material Science: Clear Images, Precise Analysis. Digital Cameras for Microscopy

Stitching MetroPro Application

Putting It All Together: Computer Architecture and the Digital Camera

Transcription:

X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner structure of various objects. It is based on penetrating radiation (such as X-Rays) which passes through an object and is attenuated by the material of the object. If the object is rotated and the passed radiation is detected at many angles, a cross-section of the object can be reconstructed using computers. Computed tomography was invented in 1972 by Hounsfield and because of its non-destructive nature, it has found its widest application in medicine for diagnosing various diseases. In this project, MATLAB was used as the key tool in building a device for X- Ray computed tomography. The system consists of an X-Ray radiation source, a rotation stage used for rotating the object under investigation, the radiation detector, and a computer. MATLAB controls the rotation stage by means of ActiveX drivers which are included in the device. After each step, an image is acquired by an external program that is called from MATLAB (using dos command). When the object has been rotated by 180 or 360 degrees, the cross-section of the body can be computed. This is done by MATLAB routines fan2para and iradon (both are part of the Image Processing Toolbox which comes as a part of MATLAB). Before calculating the resulting images, filtering of the individual images from the noise and various errors is essential. The system has been successfully built and provided satisfactory results. Further work is expected to enhance the quality and resolution of the images. 1 Block scheme and function overview The system built was an instrument for performing X-Ray computed tomography. It consists of four main parts: X-ray source (tube) rotation stage X-ray detector computer First, images of the object under investigation must be acquired. This is done at different angles to cover the full cycle, i.e. 360 degrees. The object is rotated by the rotation stage about 1 degree per a step. The next step is the mathematical calculations. The goal is to get a cross-section of the object at a given height. After enhancing the quality of the acquired images (offset and openbeam corrections), sinograms are calculated. A sinogram is a new image which consists of all 360 previously acquired images at a particular height. Let s say we want to obtain a cross-section at a height corresponding to 100th pixel of the images. We take the 100th line of each image and put it gradually after each other in the resulting sinogram. For each slice (cross-section) of the object, one sinogram is necessary. Having computed the sinograms, we can arrive at the the desired cross-sections of the object using MATLAB function iradon. 2 Used devices Let us briefly describe devices used in our setup.

Figure 1: Block scheme of the system 2.1 Shad-o-Box X-ray camera Shad-o-Box X-ray camera was used to detect the X-Ray radiation. It is based on a CMOS photo diode sensor which has 10 lines per millimeter resolution in a 2000 by 2048 pixel array. The sensor is produced for two energy ranges (10-50 kv and 10-160 kv). We used the extended version for up to 160 kv. The digital output is 12 bits. The area of the sensor covers a square with a side of 10 cm. 2.2 PXD1000 frame grabber The X-Ray camera was connected to a digital frame grabber which is a device that acquires the images from the sensor and provides the data to a computer through the computer s PCI bus. The PXD1000 frame grabber was used. The frame grabber was provided with libraries that enabled us to develop custom applications in C programming language. 2.3 Rotation stage For rotating a sample, a rotation stage by Thorlabs was used. It is a worm driven device which is controlled by T-Cube DC Servo Controller (TDC001) by the same company. The controller is connected to a computer by USB interface. It has its own software called APT User where the motion can be easily controlled. It also provides ActiveX components which provide programmable interfaces that can be incorporated into client applications (e.g. in MATLAB). 2.4 X-Ray tube Since there was not a separate X-Ray tube available which we could use only for the purposes of our device, we used the X-Ray tube from tomographic scanner BT-400 which is installed in our department. Even though the maximum voltage of the Isovolt 420/5 tube is 420 kv, we could only use 160 kv maximum because the X-Ray camera range was limited by this energy.

3 Images acquisition Figure 2: Setup of the whole system 3.1 Sample rotation The principle of the computed tomography is to take projections of the sample at many different angles. To achieve this, either the sample or the source and the detector are rotated. In our setup, the sample was rotated using the rotation stage CR1/M-Z6 together with the servo controller TDC001 by Thorlabs company. The controller was connected to a computer using the USB interface. We could not use the supplied software for controlling the motion because we needed to acquire the images after each step. Therefore, we used the ActiveX components to control the motion of the rotation stage directly from MATLAB. 3.2 Image capturing The Shad-o-Box X-Ray camera was supplied with a Windows software called ShadoCam. It provides a simple user interface to capture, display, and save images from the camera. However, the program needs an interaction with a user. There is neither a command line mode nor a batch mode. For our purposes it was necessary to be able to acquire a large number of images without any user interaction. Thus, ShadoCam software was not suitable for this task. It was necessary to code a simple program because there was not any other utility supplied to do this. To control the frame grabber, we used the libraries supplied with the frame grabber. 3.3 Correct exposure In order to get well exposed images which are not saturated or all black, the correct exposure has to be determined so that the image captures all details of the sample. The exposure consists of three adjustable parameters: integration time (set on the flat panel) voltage (set on the X-ray tube) current (set on the X-ray tube)

Both voltage and current are set on the X-Ray tube control panel and influence the characteristics of the X-Ray radiation emitted. The number of emitted photons is influenced by the current and it corresponds to the brightness of the image (together with the integration time). The voltage controls X-Ray penetration as it determines the energy range of the radiation. It can be related to the contrast of the image. 3.4 Image corrections The Shad-o-Box X-Ray detector which we used is based on CMOS technology. As opposed to CCD, CMOS active pixel sensors are less linear in their response. To decrease the influence of the non-linearity effects on the resulting picture, it is necessary to employ pixel corrections methods. The standard gain correction consists of two steps a dark offset calibration and gain correction (also referred to as flat-field or open-beam correction). The first one compensates for the readout noise, ADC offset, and dark current. The dark offset image is taken before of after tomography measurements with no radiation applied and with the same integration time as the one which will be used during the measurements. The dark offset image is simply subtracted from each acquired image. Figure 3: Comparison of a corrected image to the original The flat-field images are obtained at some arbitrary input signal level (usually around 50% of the saturation signal but for best results, it should be similar to the signal levels expected in the final images). Each final image is divided by the flat-field image and multiplied by the average value of the flat-field image. Since every final image is processed using the dark offset and flat-field image, every imperfection or error in those images will show in all the images (resulting in ring artefacts, for example). Thus, it is very advisable to take many calibration images (10 to 20) and then calculate their average to reduce noise and statistical errors. 3.5 Spot filtering There are some pixels in the detector which are defective their output level is either saturated, dark or very different from all the surrounding pixels. Such pixels cause ring artefacts in the final reconstruction. We have implemented a spot filter to filter out these pixels. Since it consumes a lot of computation time, we used it to filter only the flat-field images which leads to good results in removing the ring artefacts. The spot filter is a modified median filter. Whereas the median filter acts on all pixels and replaces them with the median value of pixels in some surroundings (typically 3 3), we implemented a thresholded median filter.

Figure 4: Ring artefacts: reconstruction without and with spot-filtered flat-field correction 4 Image reconstruction Figure 5: Images: original image and spot-filtered image After images over the whole cycle (360 degrees) have been acquired, the image reconstruction can be carried out. The reconstruction is based on the Radon transform and filtered back-projection. It uses MATLAB script iradon which is a part of basic MATLAB distribution. 4.1 Sinograms A set of projection data for a given slice through an object is called a sinogram. It is visualized as an image where the number of lines corresponds to the number of projections taken. Each line then represents a projection at a given angle. The number of columns is equal to the width of the detector. The reconstruction of a slice is performed from the data recorded in a sinogram. For each desired cross-section of the object, the sinogram must be built. It is done by taking the appropriate line from each acquired image (corresponding to all angles covering the full cycle) and putting it gradually after each other. 4.2 Fan beam geometry Before iradon can be applied, the conversion of the measured data from the fan beam geometry to the parallel beam geometry. The reason is the input requirements of the iradon function. It calculates the reconstructions from the parallel beam projections whereas we acquire the data using the fan beam because the source of the radiation is a small spot of the X-ray tube. To convert the sinogram which we acquired using the fan beam, script fan2para was used. It is a part of the Image Processing Toolbox of MATLAB.

Figure 6: Reconstruction difference between not converted data (left) and fan beam data transformed to the parallel beam data (right) 4.3 Center of rotation calculation The next step is to determine the center of rotation position in the images. It is essential that the center of rotation of the sample is perfectly aligned with the center of the detector (image). Best effort is made to align the physical geometry of the system so that the rotation stage axis is in the middle of the detector. However, it is hard to do with a pixel accuracy. It is better to use an algorithm which calculates the center of rotation from the measured data. Figure 7: The influence of the center of rotation position. On the left: center of rotation wrongly placed by 12 pixels. On the right: Correct position of the center of rotation. 5 Filtered back-projection After all three previous preparatory steps have been performed, the sinogram is passed to the iradon function which calculates the final reconstruction. The algorithm is as follows: 1. Projections are filtered. This is an essential step as the reconstructions without the filtering would appear blurred. There is a wide range of filters to choose from: Ram-Lak, Shepp-Logan, Hamming, Hann, and none. They differ in speed and efficiency. The ideal filter is a ramp filter which is not physically viable so its cropped version (Ram-Lak) is used. As the Ram-Lak filter is sensitive to noise in projections, we used the Hamming filter which multiplies the Ram-Lak filter by a Hamming window. 2. Backprojection. The filtered projections are backprojected to obtain the final reconstruction. It is done in a loop over the angles of rotation. Interpolation is necessary during the backprojection; we used linear interpolation because of the speed. 6 Tomography measurements procedure In this section, we are going to describe the overall procedure to get a final image (slice, crosssection) of the object.

1. Put the object on the rotation stage. 2. Initialize the rotation stage control. 3. Set the correct exposition: Turn on the X-ray radiation, set the integration time, and acquire an image of the object. Load the test image in MATLAB and display its histogram. Adjust the X-ray voltage and/or current and the integration time to get the correct exposure. 4. Acquire the dark offset images for calibration: While X-rays are turned off, acquire at least 10 dark offset images. Remember that the integration time should be the same as for all the other measurements (as determined in the above step). 5. Start the image acquisition: Turn on the X-rays and start the acquisition. This step will take the most time depending on the integration time (typically around 30 minutes). The system will rotate the sample about 360 degrees in 360 steps and after each step, it will take an X-ray image which is saved on the hard drive in raw format. 6. Acquire the flat-field image: Turn off the X-rays, remove the sample from the rotation stage, turn on the X-rays, and acquire the flat-field image. The input signal level should be similar to the one used in previous measurement but the detector must not be in saturation. Again, it is important to take more images because of the noise and statistical errors. 7. Average the correction images: Calculate the average of the dark offset and flat-field images to improve the quality. 8. Normalize the images: Perform pixel corrections on the acquired images. 9. Make the sinogram. 10. Reconstruct the cross-section (slice). 7 Images appendix Figure 8: An aluminium profile on the rotation stage.

Figure 9: Sinogram of an aluminium profile. Figure 10: Final reconstruction of an aluminium profile. 8 Conclusion Figure 11: Final reconstruction of an aluminium cylinder. During the course of the work, we managed to build a system which is now able to visualize the inner structure of bodies under investigation. We started from scratch having just the components of the resulting instrument an X-ray detector and a source, a rotation stage, and a computer. The components were physically put together and software to control each of them was written. We used C language for controlling the detector and MATLAB for all other work. MATLAB was also the environment from which the system was controlled. We explored various effects on image quality of the reconstruction images. We found a couple of different artefacts which were present in the reconstructions and found a way how to limit them. Among others, let us mention the ring artefact and the influence of a correctly determined center of rotation. Bc. Jan Kratochvíla jan.kratochvila@gmail.com Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Department of Physical Electronics Bøehová 7, 115 19 Praha 1