In essence this means, that a certain proportion of a signal in one channel is actually derived from another dye spilling over into the channel.

Similar documents
Multifluorescence The Crosstalk Problem and Its Solution

Practical work no. 3: Confocal Live Cell Microscopy

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Instruction Manual. Mark Deimund, Zuyi (Jacky) Huang, Juergen Hahn

TN378: Openlab Module - FRET. Topic. Discussion

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1

Zeiss Axio Imager.A1 manual

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Adobe Studio on Adobe Photoshop CS2 Enhance scientific and medical images. 2 Hide the original layer.

ZEISS LSM510META confocal manual

Nasmyth Ultraview Vox User Protocol

personal DELTAVISION (pdv)

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Zeiss 780 Training Notes

Quick Guide. NucleoCounter NC-3000

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Leica_Dye_Finder :53 Uhr Seite 6 Dye Finder LAS AF

Microscopy from Carl Zeiss

AxioVision User's Guide. Release 4.1

Zeiss 880 Training Notes Zen 2.3

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Spatial intensity distribution analysis Matlab user guide

GIMP WEB 2.0 ICONS. Web 2.0 Icons: Paperclip Completed Project

Motic Live Imaging Module. Windows OS User Manual

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

The following units are required for an ApoTome imaging workstation:

Zeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005

ID Photo Processor. Batch photo processing. User Guide

Using the Nikon TE2000 Inverted Microscope

Creating a Contrast Mask. Text and images Copyright (C) 2002 Eric R. Jeschke and may not be used without permission of the author.

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Zeiss Axioplan 2 imaging microscope and Axiovision software

Zeiss Deconvolution Microscope: A Quick Guide

OPERATING INSTRUCTIONS

Calculate Ratiometric FRET-Images with the FRET-Image-Script

Quick Start Guide. Leica SP5 X

IncuCyte ZOOM Scratch Wound Processing Overview

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras

Leica TCS SP8 Quick Start Guide

Spectral Imaging with the Opterra Multipoint Scanning Confocal

Color and More. Color basics

FLIM Protocol. 1. Install IRF sample on the microscope using the stage insert. IRF sample options include urea crystals or BBO crystal.

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Takeoff Guide Tutorial to get started with the AxioVision Imaging System (Based on Release 4.8 June 2009)

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Version 6. User Manual OBJECT

ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

LSM 510 META in Chang Gung University

Optika ISview. Image acquisition and processing software. Instruction Manual

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Leica TCS SP8 Quick Start Guide

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

IncuCyte ZOOM Fluorescent Processing Overview

Operating Instructions for Zeiss LSM 510

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

Image Analysis for Fluorescence

Simplified Instructions: Zeiss Brightfield Microscope S1000

Exercise 4-1 Image Exploration

A guide to SalsaJ. This guide gives step-by-step instructions on how to use SalsaJ to carry out basic data analysis on astronomical data files.

Zeiss AxioObserver with ApoTome

TEMScripts Auto Image Measurement (Particle) Manual. TEMScripts LLC. Last updated: 12/5/2016

Image Pro Ultra. Tel:

TRAINING MANUAL. Olympus FV1000

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

The operation manual of spotlight 300 IR microscope

Fundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy. Integrated Microscopy Course

CELL-IQ IMAGEN USER MANUAL

Color Correction and Enhancement

Manual. ZEN 2011 (blue edition)

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

inform ADVANCED IMAGE ANALYSIS SOFTWARE inform User Manual

SlideBook 5. FRAP Imaging Module

Photoshop Project 1: Create Vector Art

LSM 510 Meta Training Notes

TEMScripts Auto Image Measurement (Pattern) Manual. TEMScripts LLC. Last updated: 12/5/2016

Proudly serving laboratories worldwide since 1979 SPECIFICATIONS

Leica Sp5 II Confocal User Guide

ADOBE 9A Adobe Photoshop CS3 ACE.

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

LSM 510 Training Notes

Microscopy Live Animal Imaging

Quick Operation Guide

1. Editorial. N 9 June Content

Getting Started in Eagle Professional Schematic Software. Tyler Borysiak Team 9 Manager

Thermo ImageQuest Version 1.0.1

ImageJ: Introduction to Image Analysis 3 May 2012 Jacqui Ross

Ph 3455 The Photoelectric Effect

LSM 780 Confocal Microscope Standard Operation Protocol

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Pattern Matching. Open an Image

Using Curves and Histograms

Motorized Axio Observer Start-up instructions

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

GXCapture 8.1 Instruction Manual

Digital Imaging - Photoshop

Sensitive imaging of spectrally overlapping fluorochromes using the LSM 510 META

Transcription:

NNKQ táçéñáéäç=jìäíáåü~ååéä=råãáñáåö= _~ÅâÖêçìåÇ= Widefield Multichannel Unmixing is a new function for the removal of crosstalk between fluorescent dyes in multichannel images with up to eight fluorescence channels using a traditional epifluorescence microscope. Crosstalk is a phenomenon which occurs, whenever dyes are excited by excitation light of more than one filter combination. This problem occurs for example, when dyes with broadly overlapping excitation spectra are used concurrently in one sample. Good examples are the spectral variants of fluorescent proteins, e.g. BFP, CFP, GFP and YFP. With traditional reflector filter sets it has been often difficult to achieve 100% signal separation for such dyes. In essence this means, that a certain proportion of a signal in one channel is actually derived from another dye spilling over into the channel. A parameter based separation is now available with the Widefield Multichannel Unmixing module. The procedure is separated into three steps: Measurement of the amount of crosstalk in 2 alternative procedures: With appropriate reference samples by automatic component extraction (ACE) from the sample to be unmixed Creation of the unmixing matrix file Unmixing of the image to be corrected by means of using the unmixing matrix file generated in step 2 11-62 B 48-0038 e 04.2005

NOTES: Non-fluorescent channels such as DIC or Phase contrast are automatically detected and disregarded by all unmixing functions. Such channels are automatically copied into the unmixed result images to facilitate merged views with fluorescent channels. The module Widefield Multichannel Unmixing is designed to work with images created using the AxioVision Multichannel Fluorescence module. For the creation of correct reference images the use of the AxioVision Multichannel module is required. The images used in these examples you will find on the AxioVision installation CD in the folder "Images". qüé=mêáååáéäé=çñ=ìåãáñáåö= The following example tries to demonstrate the problem of cross talk as well as show the potential of unmixing such images. In this case, determination of cross talk is done using reference samples and not ACE. Reference sample A (CFP-Reference1.zvi): HeLa cells containing a virus proteins coupled to cyan fluorescent protein (CFP). This protein accumulates especially in the nucleolar regions of the cell nucleus. Reference sample B (GFP-Reference1.zvi): HeLa cells containing a virus proteins coupled to green fluorescent protein (GFP). This protein accumulates in the nucleolar regions of the cell nucleus too. Sample C: (RevCFP-H2-GFP_3.zvi): HeLa cells containing two proteins with different fluorescence: coupled to CFP is the same viral protein as in reference sample A (nucleoli); coupled to GFP is a histone protein, which stains the chromosomes and thus the entire nucleus a lesser degree of staining of the nucleoli. AxioVision User's Guide, Release 4.4 11-63

Sample Channel 1(CFP) Channel 2 (GFP) A. Reference sample CFP only: A small degree of crosstalk is visible in channel 2 coming from channel 1. Image histogram: x-axis: pixel gray values. y-axis: relative number of pixels in logarithmic scale; identical settings of the display curve. B. Reference sample GFP only: A moderate degree of cross talk from channel 2 is visible in channel 1. Image histogram: x-axis: pixel gray values. y-a xis: relative number of pixels in logarithmic scale; identical settings of the display curve. 11-64 B 48-0038 e 04.2005

Sample Channel 1(CFP) Channel 2 (GFP) C. 2-channel fluorescence sample (CFP and GFP): Before Unmixing: How much signal from channel 2 is visible in channel 1 (cross talk) and vice versa? Image histogram: x-axis: pixel gray values. y-axis: relative number of pixels in logarithmic scale; identical settings of the display curve. D. 2-channel fluorescence sample (CFP and GFP): After Unmixing: A net percentage of pixel intensity values from channel 2 are added to the signal in channel 1. Image histogram: x-axis: pixel gray values. y-axis: relative number of pixels in logarithmic scale; identical settings of the display curve. AxioVision User's Guide, Release 4.4 11-65

E. 2-channel fluorescence sample (CFP and GFP): Comparison "before" and "after" unmixing: shows a markedly improved signal separation (shown at identical display settings. CFP&GFP-sample original pseudocolor mode CFP&GFP-sample unmixed pseudocolor mode 11-66 B 48-0038 e 04.2005

qüé=táçéñáéäç=jìäíáåü~ååéä=råãáñáåö=ñìååíáçå= The Widefield Multichannel Unmixing function is the central function for unmixing samples: U sing the function dialog, you can select whether the unmixing should take place via automatic component extraction (left-hand image) or using a reference matrix (right-hand image). If you choose unmixing using a reference matrix, you will need a reference matrix file containing the corresponding information for the unmixing. AxioVision User's Guide, Release 4.4 11-67

If such a file does not yet exist, you can start a wizard via the button, which will lead you through the process of generating a reference matrix file. tçêâáåö=ïáíü=íüé=ä~ëáå=ñìååíáçåë= The basic functions in the Widefield Multichannel Unmixing Basic f unctions menu are the functions for unmixing that will already be familiar to you from the previous version of AxioVision. These functions are still available, for compatibility reasons and for use in Commander scripts. The following two examples describe how you can use these functions for unmixing. råãáñáåö=ïáíü=êéñéêéååé=ë~ãéäéë= This example shows you how to measure cross talk using reference samples and use this information to unmix a 2 channel fluorescence image. The same images are used as in example 1. Open the first reference image "CFP-Reference1.zvi". It is a 3 channel image. The first channel contains transmitted light information (DIC). The CFP signal is visible strongest in channel 2 (Zeiss filter set #47). Change into b/w mode ( ) and channel. 11-68 B 48-0038 e 04.2005

Open the properties dialog (menu View Properties) and adjust the display line in a way to enhance the gray value visibility (e.g. Best Fit and Gamma value ~ 0.5). Now it is easier to distinguish true image background from signal derived from the fluorescent dye. In the workarea select Widefield Multichannel Unmixing and then Reference measurement. The image "CFP- Reference1.zvi" is selected automatically as Input image because it the active image in the foreground. AxioVision User's Guide, Release 4.4 11-69

If it is not selected, switch the Create new function to Off gallery. and select the image from the pop up In the edit field on the button. click Select a suitable region in the image for background correction (ideally without any signal coming from the dye). The pixel coordinates (x/y) as well as the gray value of the selected pixel are shown. Click at an suitable position once in the image. The x coordinate is shown in the input field. Click the input field button in the. Search for a suitable region in the image for measurement. Ideally search for a region with an intensity as high as possible. 11-70 B 48-0038 e 04.2005

Click the button to carry out measurement. The regions taken for measurement are shown in the image. Save the image, because it s used later for generation of the unmixing matrix. Open the second reference image GFP-Reference1.zvi. The third channel contains the GFP fluorescence (fluorescence filter set # 44). Repeat the steps 2-7 also for this image. In the workarea select the function Create Unmixing Matrix from the Widefield Multichannel Unmixing function group. 10. Select the input images accordingly: CFP-Reference1 as ReferenceImage1, GFP- Reference1.zvi as ReferenceImage2. Please take care to enter channels and reference images in the same order. If you use images with more than two channels, the function is extended accordingly. AxioVision User's Guide, Release 4.4 11-71

11. To define a name for the unmixing matrix file, click the button in the input field. In the following dialog you can enter a name for the matrix file. Click th e button to accept the filename. Click the button to generate the unmixing matrix file. The file is saved like other AxioVision files in the user folder. In the workarea select the function Unmix Multichannel Image from the Widefield Multichannel Unmixing function group. Open the image to unmix and select it as Input image. In this example, the image "RevCFP- H2-GFP_3.zvi" is used. 11-72 B 48-0038 e 04.2005

Switch to the black and white view ( ) and activate channel 2 ( ). Right click in the image and select Properties. Set the display characteristic line in such a way, that the gray values are displayed extremely amplified. So you can easily detect, where the background is not caused by the sample. Switch to channel 3 and repeat the setting of the display characteristic line like in previous step. Switch back to the color view and switch off channel 1 Now you can separate the two fluorescence colors clearly from the background signal.. AxioVision User's Guide, Release 4.4 11-73

Click the input field button in the. In the image search for a suitable region for the background measurement (preferably without a signal caused by the fluorescence dye). The pixel coordinates (x/ y) as well as the gray value of the selected pixel are shown. Click at an suitable position once in the image. The x coordinate is shown in the input field. Enter a name for the Output image. Click the button in the input field to select the unmixing matrix file. Click the button to load the file. The dialog is closed. 11-74 B 48-0038 e 04.2005

Click the button to start the unmix ing. The result image shows a clear separation of the fluorescence channels 2 (CFP) and 3 (GFP). Channel 1 without fluorescence (DIC) is copied to the output images unchanged. To optimize the display, you should set the display characteristic line for channel 2 and 3 accordingly: Save the output image. AxioVision User's Guide, Release 4.4 11-75

råãáñáåö= çñ=~=ãìäíáåü~ååéä=áã~öé=ìëáåö=^`b= This example shows you how to measure cross talk directly an image for unmixing with Automatic Component Extraction (ACE) and how to use it for unmixing of samples. The same images are used as in example 1. 1. Open the image to unmix. In this example the image "RevCFP-H2-GFP_3.zvi" is used. Switch to the black and white view ( ) and activate channel 2 ( ). Right click in the image and select Properties. Set the display characteristic line in such a way, that the gray values are displayed extremely amplified. So you can easily detect, where the background is not caused by the sample. 11-76 B 48-0038 e 04.2005

Switch to channel 3 and repeat the setting of the display characteristic line like in previous step. Switch back to the color view and switch off channel 1 Now you can separate the two fluorescence colors clearly from the background signal.. In the workarea select the function Automatic Component Extraction from the Widefield Multichannel Unmixing function group. The reference image "RevCFP- H2-GFP_3.zvi" is selected automatically as Input image because it the active image in the foreground. AxioVision User's Guide, Release 4.4 11-77

If it is not selected, switch the Create new function to Off gallery. and select the image from the pop up Click the input field button in the. In the image search for a suitable region for the background measurement (preferably without a signal caused by the fluorescence dye). The pixel coordinates (x/y) as well as the gray value of the selected pixel are shown. Click at an suitable position once in the image. The x coordinate is shown in the input field. The predefined threshold in the input field Threshold can be used unchang ed. For further information about this parameter, please read in the online help. To define a name for the unmixing matrix file, click the button in the input field. In the following dialog you can enter a name for the matrix file. Click the button to accept the filename. 11-78 B 48-0038 e 04.2005

Click the but ton to generate the unmixing matrix file. The file is saved like other AxioVision files in the user folder. The measurement regions defined by ACE are drawn to the image using the channel color. Switch to the black and white view ( ) and activate the accordant channel. Channel 2: Channel 3: In the workarea select the function Unmix Multichannel Image from the Widefield Multichannel Unmixing function group. Open the image to unmix and select it as Input image. In this example, the image "RevCFP- H2-GFP_3.zvi" is used. AxioVision User's Guide, Release 4.4 11-79

Now repeat the background measurement: Click the button in the input field and then click in a background region (see also step 7). Enter a name for the Output image. Click the button in the input field to select the unmixing matrix file. Click the button to load the file. The dialog is closed. Click the button to start the unmixing. The result image shows a clear separation of the fluorescence channels 2 (CFP) and 3 (GFP). Channel 1 without fluorescence (DIC) is copied to the output images unchanged. 11-80 B 48-0038 e 04.2005

To optimize the display, you should set the display characteristic line for channel 2 and 3 accordingly: Save the output image. AxioVision User's Guide, Release 4.4 11-81