Applying Virtual Oscilloscope to Signal Measurements in Scintillation Detectors

Similar documents
Measuring Voltage and Time Quantities of a Signal Through a Virtual Oscilloscope

GENERATION OF SIGNALS USING LABVIEW FOR MAGNETIC COILS WITH POWER AMPLIFIERS

A Method to Reveal 137 Cs Gamma Spectrum by a Multi-Pixel Photon Counter

CHAPTER 7 HARDWARE IMPLEMENTATION

Development of an Amplifier for Nuclear Spectrometers

Development of 4/16-Channel Data Acquisition System Using Lab VIEW

DEVELOPMENT OF A CHARGE-SENSITIVE PREAMPLIFIER USING COMMERCIALLY AVAILABLE COMPONENTES

INVESTIGATION AND DESIGN OF HIGH CURRENT SOURCES FOR B-H LOOP MEASUREMENTS

Radiation Detection Instrumentation

Precalculations Individual Portion Introductory Lab: Basic Operation of Common Laboratory Instruments

Lab 12 Laboratory 12 Data Acquisition Required Special Equipment: 12.1 Objectives 12.2 Introduction 12.3 A/D basics

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics

EKT 314/4 LABORATORIES SHEET

ni.com Sensor Measurement Fundamentals Series

Generation of Gaussian Pulses using FPGA for Simulating Nuclear Counting System

PX4 Frequently Asked Questions (FAQ)

Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source

Laboratory set-up for Real-Time study of Electric Drives with Integrated Interfaces for Test and Measurement

User Guide. SIB Channel APD Array Interface Board Hamamatsu S8550 series

A Virtual Instrument for Automobiles Fuel Consumption Investigation. Tsvetozar Georgiev

K 223 Angular Correlation

Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel

400 W up to 21,000 W 260 V up to 440 V 3 A up to 100 A. Electronic Load ZSAC

Electronic AC Load ZSAC Series. 400 W up to 21,000 W 260 V up to 440 V 3 A up to 100 A. Rev. 3.03

TF TF Analyzer 2000 Measurement System

Faculty of Information Engineering & Technology. The Communications Department. Course: Advanced Communication Lab [COMM 1005] Lab 6.

User Guide. SIB616 4 x 4 SiPM Sensor Interface Board SensL ArrayC P

Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission

Physics 472, Graduate Laboratory DAQ with Matlab. Overview of data acquisition (DAQ) with GPIB

Data acquisition and instrumentation. Data acquisition

Positron Emission Tomography

Electro Optical Components, Inc. SUNSTAR 传感与控制 TEL: FAX: Skylane Boulev

Labview Based Gain scheduled PID Controller for a Non Linear Level Process Station

Measurement & Control of energy systems. Teppo Myllys National Instruments

LabVIEW and MatLab. E80 Teaching Team. February 5, 2008

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Data Acquisition Boards and USB-DAQ

Development of front-end readout electronics for silicon strip. detectors

PC-based controller for Mechatronics System

MAROC: Multi-Anode ReadOut Chip for MaPMTs

Getting started with OPENCORE NMR spectrometer. --- Installation and connection ---

Fig. 1. NI Elvis System

Lab VIEW Programming for Vibration Analysis

Things to consider while selecting a Digitizer

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

Lab 2: Designing an Optical Theremin. EE 300W Section 5 Team #3: Penn Power United Gregory Hodgkiss, Nasser Aljadeed 10/23/15

Universally Applicable for Piezoelectric Measuring Technology. Technical Data. Charge Input. Instant value. Extreme values ors statistic measurements

Atomic and Nuclear Physics

Universally Applicable for Piezoelectric Measuring Technology. Technical Data. Charge Input. Extreme values ors statistic measurements

GDS MHz Handheld Digital Oscilloscope New Product Announcement

DEVELOPMENT OF HIGH STABLE MONITOR FOR MEASURERING ENVIRONMENTAL RADIATION

An Incremental Measurements and Data Acquisition Project

Charge Meter. Electronics & Software. Universally Applicable for Piezoelectric Measuring Technology. Type 5015A...

Final Report (Group 15-22)

Effective Teaching Learning Process for PID Controller Based on Experimental Setup with LabVIEW

Corona Current Data Acquisition Card Based on USB Bus in Extra High Voltage Environment

Specifications Test Conditions Vs = ± 15 V, Ta = 25 C Gain Gain Values 40, 60, 80, 100 db indicated by four LEDs Gain Accuracy ± 0.1 % (between settin

GPS Tracking System Using Car Charger

User Guide. SIB Channel MAPMT Interface Board Hamamatsu H7546 series

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z

Kistler LabAmp. Electronics & Software. Charge Amplifier and Data Acquisition Unit for Dynamic Measurements. Type 5165A...

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers

DATA ACQUISITION AND DATA PROCESSING SOFTWARE FOR MAGNETIC MEASUREMENTS FOR NICA MAGNETS

BACKPLANE ETHERNET CONSORTIUM

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS

nanodpp datasheet I. FEATURES

Signal Generation in LabVIEW

Design And Application Of A Control System For DC Motors Over Power Line

nanomca datasheet I. FEATURES

Optical Power Meter Basics

Mass Spectrometry and the Modern Digitizer

Metrological Characterization of Low Power Instrument Transformer Integrated in MV Recloser

GDS-2000A 300/200/100/70 MHz DSO New Product Announcement

Laboratory Preliminaries and Data Acquisition Using LabVIEW

e t Development of Low Cost γ - Ray Energy Spectrometer

WaveAce Integration with WaveStation

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1

A novel acquisition method of nuclear spectrum based on pulse area analysis *

A SMART METHOD FOR AUTOMATIC TEMPERATURE CONTROL

SonoLab Echo-I User Manual

Using Signal Express to Automate Analog Electronics Experiments

Uniformity and Crosstalk in MultiAnode Photomultiplier Tubes

EXPERIMENT 2 DIGITAL STORAGE OSCILLOSCOPE

Virtual Digital Control Experimental System

Exp 3 COLCULATE THE RESPONSE TIME FOR THE SILICON DETECTOR

BIOMEDICAL DIGITAL SIGNAL PROCESSING

A NOVEL FPGA-BASED DIGITAL APPROACH TO NEUTRON/ -RAY PULSE ACQUISITION AND DISCRIMINATION IN SCINTILLATORS

Getting started with Mobile Studio.

Pulse Shape Analysis for a New Pixel Readout Chip

LabVIEW Based Biomedical Signal Acquisition and Processing

LabVIEW Based Instrumentation and Experimental Methods Course

High Stability Voltage Source

What the LSA1000 Does and How

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination

PHYSICS EXPERIMENTS BASED ON DATA ACQUISITION THE LONGITUDINAL ACOUSTIC DOPPLER EFFECT

1 Purpose of This Lab Exercise:

PC-OSCILLOSCOPE PCS500. Analog and digital circuit sections. Description of the operation

CAEN Tools for Discovery

Gamma Spectrometer Initial Project Proposal

Transcription:

Radiation Science and Technology 2015; 1(1): 1-5 Published online July 16, 2015 (http://www.sciencepublishinggroup.com/j/rst) doi: 10.11648/j.rst.20150101.11 Applying to Signal Measurements in Scintillation Detectors Gozde Tektas *, Cuneyt Celiktas Ege University, Faculty of Science, Physics Department, Bornova, Izmir, Turkey Email address: gozdetektas90@hotmail.com (G. Tektas) To cite this article: Gozde Tektas, Cuneyt Celiktas. Applying to Signal Measurements in Scintillation Detectors. Radiation Science and Technology. Vol. 1, No. 1, 2015, pp. 1-5.doi: 10.11648/j.rst.20150101.11 Abstract: A virtual oscilloscope was designed by using the LabVIEW software. Signals supplied by a pulse generator and background radiation signals from a NaI(Tl) scintillation detector were displayed in a real and a virtual oscilloscope, respectively. Amplitude, maximum voltage, rise time and fall time values through the oscilloscopes for both type signals were measured. They were acquired in different time/div. values to test and compare their performances. Obtained results and the signal shapes from them were meticulously compared. It was observed that they were highly comparable to each other. Results indicate that the developed virtual oscilloscope would reliably be able to be used for data acquisition as well as a real oscilloscope. Keywords: NaI(Tl) Detectors, LabVIEW Software, and GPIB 1. Introduction LabVIEW software is one of the virtual instrumentation software platforms. LabVIEW expanded as Laboratory Virtual Instrumentation Engineering Workbench. It provides an easy-to-use application development environment designed specifically with the needs of engineers and scientists. It is a powerful graphical development environment for signal acquisition, measurement analysis and data presentation [1]. The new way in the design of computer-based measurement systems can be seen in the use of up-to-date measurement, control and testing systems based on reliable devices. Digital signal processing (DSP) is used to replace conventional analog systems [2]. The virtual instrumentation technique performed in LabVIEW is used for nuclear DSP system realization [3]. A virtual oscilloscope is composed of two parts in that hardware and software. The hardware includes a general computer and a data acquisition (DAQ) device [4]. The DAQ devices typically connect directly to the computer s internal bus through a plug-in slot. Some DAQ devices are external and connected to the computer via serial, GPIB (General Purpose Interface Bus) or Ethernet ports [5]. GPIB facilitates the communication between computer and instrument. It is simply the means by which computers and instrument transfer data. Its purpose is to provide computer control of test and measurement instrument [6]. Their software include their driver programs also [4]. An oscilloscope is a device for recording and visualizing the instantaneous value of alternating voltage [7]. It can be used in order that amplitude, maximum voltage, rise time and fall time values of a signal are determined. Amplitude is a difference between max. and min. voltage values of a signal. Maximum voltage is a value of positive peak voltage [8]. Rise time is the passed time to rise from 10 to 90% of the signal amplitude. Fall time is the passed time to fall from 90 to10% of the signal amplitude [9]. A NaI(Tl) scintillation detector is widely used to particle detection in nuclear physics. The scintillation can be converted into electrical pulses which can be analyzed and counted electronically to give information concerning the incident radiation. A preamplifier amplifies its input signal from a detector by filtering it from the electronic noise together with the impedence matching. A main amplifier amplifies the output of the preamplifier and shapes it to a convenient form for further processing [9]. A multifunctional virtual oscilloscope was designed by Gong and Zhou using LabVIEW software [4]. A virtual oscilloscope was designed with LabVIEW by D. ShengLi et al. Voltage and time parameters were determined from a signal generator to design the test signals by ShengLi et al. [10]. In this study, a generator signal was compared with a real

2 Gozde Tektas and Cuneyt CeliktasApplying to Signal Measurements in Scintillation Detectors. oscilloscope and a virtual oscilloscope developed with LabVIEW. The quantities containing voltage and time values of the signal and the signal shapes were compared with each other. These processes were repeated for the background signal from a NaI(Tl) scintillation detector. 2. Materials and Methods In this work, a GW Instek Model 2204 oscilloscope and a pulse generator (Ortec 419) as a signal source were used. The virtual oscilloscope was operated in a desktop PC by generating a virtual instrument (VI) through LabVIEW. The connection between the computer and the real oscilloscope was performed by GPIB interface. Pulse generator output was first sent to the real oscilloscope, then, the signal was transferred to the virtual oscilloscope through GPIB cable. The real oscilloscope s interface has a USB port, RS232C and GPIB port. Due to the fact that data transfer rate via GPIB is fast, GPIB cable was preferred. A block diagram of the used setup is given in Figure 1. Figure 1. A block diagram for the measurement of the generator signal. (DSO: Digital Storage Oscilloscope, VI: Virtual Instrument, GPIB: General Purpose Interface Bus). The generator signal was analyzed with different time/div. values of the real and During the measurements, volt/div. adjustment of both oscilloscopes was set to 500 V. Amplitude, maximum voltage, rise time and fall time values of the signal were determined. The measurement was repeated five times for each time/div. setting. Each quantity was determined between 250.0 and 2.5 time/div. intervals since the entire signal could not be seen on the scope. In the second part of the work, background signal from a Bicron NaI(Tl) scintillation detector (3 x 3 ) was displayed in both of the real and the virtual oscilloscopes. The spectrometer used for this process composed of a preamplifier (Ortec 113), an amplifier (Ortec 485) and the GW Instek 2204 oscilloscope together with the virtual oscilloscope installed in a PC via GPIB connection were used. A block diagram is shown in Figure 2. Figure 2. A block diagram for the measurement of NaI (Ti) scintillation detector background signal. (DT: Detector, HV: High Voltage, PA: Preamplifier, A: Main Amplifier, DSO: Digital Storage Oscilloscope, VI: Virtual Instrument, GPIB: General Purpose Interface Bus). In the block diagram, the detector output was sent to the preamplifier. Its output was transmitted to the amplifier, and its output signal was displayed in the real oscilloscope. The signal was transferred from the real oscilloscope to the virtual oscilloscope through GPIB cable. High voltage to the detector, preamplifier capacitance, amplifier coarse gain and fine gain adjustments were set to 1000 V, 100 pf, 64 and 10 settings, respectively, thus, the background signal had nearly same amplitude with the generator signal to compare and test them easily. Amplitude, maximum voltage, rise time and fall time values of the background signal were measured five times for each time/div. values from both oscilloscopes. Volt/div. adjustments of the oscilloscopes were kept unchanged. Since the signal shapes deteriorated in the time/div. under 2.5 and over 50.0, all measurements were performed between these time values. 3. Results and Discussion Amplitude, maximum voltage, rise time and fall time values for the generator signal and the background signal were determined from the real and the virtual oscilloscopes. To test the performance of the virtual oscilloscope signal measurements were performed at different time/div. adjustments. All measurements were performed after the oscilloscopes were stopped. For the generator signal, time/div. setting was set to 250.0, 500.0, 1.0 and 2.5, respectively. The comparison between real and virtual oscilloscopes for the generator signal is given in Tables 1-4. Background signal from the NaI(Tl) scintillation detector was analyzed for 2.5, 5.0, 10.0, 25.0 and 50.0 time/div. adjustments. Obtained results from the oscilloscopes are given with their relative errors in Tables 5-9. Besides, signal shape comparison for the generator signal for 250.0 time/div. adjustment is shown in Fig. 3. Background signal shapes from the scintillation detector through both oscilloscopes are compared in Fig. 4 for 2.5 time/div. setting. Table 1. Amplitude (V amp), maximum voltage (V max), rise time (T R) and fall time (T F) values of the generator signal for 250.0 time/div. in the real and 1 1.020±0.109 1.400±0.011 5.912±0.143 778.500±0.096 2 1.020±0.109 1.380±0.002 6.182±0.093 813.500±0.049 3 1.240±0.087 1.380±0.002 7.402±0.086 815.000±0.047 4 1.240±0.087 1.380±0.002 7.403±0.086 896.000±0.047 5 1.140±0.007 1.380±0.002 6.908±0.021 964.500±0.115 1 1.020±0.109 1.400±0.011 5.913±0.143 778.500±0.096 2 1.020±0.109 1.380±0.002 6.182±0.093 813.500±0.049 3 1.240±0.087 1.380±0.002 7.403±0.086 815.000±0.047 4 1.240±0.087 1.380±0.002 7.403±0.086 896.000±0.047 5 1.140±0.007 1.380±0.002 6.909±0.021 964.500±0.115

Radiation Science and Technology 2015; 1(1): 1-5 3 Table 2. Amplitude (V amp), maximum voltage (V max), rise time (T R) and fall time (T F) values of the generator signal for 500.0 time/div. in the real and 1 1.340±0.011 1.380±0.011 16.000±0.002 904.000±0.035 2 1.300±0.018 1.340±0.017 16.250±0.018 854.900±0.019 3 1.320±0.003 1.360±0.002 15.760±0.012 869.900±0.002 4 1.340±0.011 1.380±0.011 15.760±0.012 867.600±0.004 5 1.320±0.003 1.360±0.002 16.000±0.002 862.900±0.010 1 1.340±0.011 1.380±0.011 16.000±0.002 904.000±0.035 2 1.300±0.018 1.340±0.017 16.250±0.018 855.000±0.019 3 1.320±0.003 1.360±0.002 15.760±0.012 870.000±0.002 4 1.340±0.011 1.380±0.011 15.760±0.012 867.500±0.005 5 1.320±0.003 1.360±0.002 16.000±0.002 862.900±0.010 Table 3. Amplitude (V amp), maximum voltage (V max), rise time (T R) and fall time (T F) values of the generator signal for 1.0 time/div. in the real and 1 1.340±0.017 1.380±0.017 31.520±0.018 897.400±0.045 2 1.320±0.003 1.360±0.002 32.490±0.012 818.900±0.045 3 1.220±0.078 1.260±0.076 32.000±0.003 827.300±0.034 4 1.360±0.032 1.400±0.031 32.480±0.011 905.200±0.054 5 1.340±0.017 1.380±0.017 32.000±0.003 832.400±0.028 1 1.340±0.017 1.380±0.017 31.530±0.018 897.500±0.045 2 1.320±0.003 1.360±0.002 32.490±0.012 819.000±0.045 3 1.220±0.078 1.260±0.076 32.000±0.003 827.200±0.035 4 1.360±0.032 1.400±0.031 32.470±0.011 905.100±0.054 5 1.340±0.017 1.380±0.017 32.000±0.003 832.300±0.028 Table 4. Amplitude (V amp), maximum voltage (V max), rise time (T R) and fall time (T F) values of the generator signal for 2.5 time/div. in the real and 1 1.300±0.000 1.340±0.000 79.990±0.003 831.500±0.046 2 1.280±0.015 1.320±0.015 79.990±0.003 894.300±0.026 3 1.300±0.000 1.340±0.000 78.780±0.012 819.200±0.062 4 1.300±0.000 1.340±0.000 79.990±0.003 873.200±0.003 5 1.320±0.015 1.360±0.014 79.990±0.003 932.800±0.067 1 1.300±0.000 1.340±0.000 80.000±0.003 831.600±0.046 2 1.280±0.015 1.320±0.015 80.000±0.003 894.200±0.026 3 1.300±0.000 1.340±0.000 78.780±0.012 819.100±0.062 4 1.300±0.000 1.340±0.000 80.000±0.003 873.300±0.003 5 1.320±0.015 1.360±0.014 80.000±0.003 932.800±0.067 Table 5. Amplitude (V amp), maximum voltage (V max), rise time (T R) and fall time (T F) values of the background signal for 2.5 time/div. in the real and 1 1.120±0.025 1.220±0.049 1.215±0.002 3.590±0.343 2 1.140±0.042 1.180±0.016 1.272±0.042 5.240±0.079 3 1.180±0.074 1.220±0.049 1.315±0.073 5.938±0.187 4 1.000±0.092 1.040±0.115 1.200±0.015 5.350±0.098 5 1.020±0.070 1.140±0.017 1.088±0.119 3.992±0.207 1 1.120±0.025 1.220±0.049 1.215±0.002 3.590±0.343 2 1.140±0.042 1.180±0.016 1.272±0.042 5.240±0.079 3 1.180±0.074 1.220±0.049 1.315±0.073 5.938±0.187 4 1.000±0.092 1.040±0.115 1.200±0.015 5.350±0.098 5 1.020±0.070 1.140±0.017 1.088±0.119 3.992±0.207 Table 6. Amplitude (V amp), maximum voltage (V max), rise time (T R) and fall time (T F) values of the background signal for 5.0 time/div. in the real and 1 1.700±0.228 1.740±0.216 1.275±0.000 7.840±0.115 2 1.180±0.111 1.200±0.136 1.278±0.001 7.530±0.078 3 1.240±0.058 1.300±0.049 1.315±0.029 5.900±0.175 4 1.320±0.006 1.380±0.011 1.296±0.015 7.230±0.040 5 1.120±0.171 1.200±0.136 1.216±0.049 6.176±0.122 1 1.700±0.228 1.740±0.216 1.275±0.000 7.840±0.115 2 1.180±0.111 1.200±0.136 1.278±0.001 7.530±0.078 3 1.240±0.058 1.300±0.049 1.315±0.029 5.900±0.175 4 1.320±0.006 1.380±0.011 1.296±0.015 7.230±0.040 5 1.120±0.171 1.200±0.136 1.216±0.049 6.176±0.122 Table 7. Amplitude (Vamp), maximum voltage (Vmax), rise time (TR) and fall time (TF) values of the background signal for 10.0 time/div in the real and Vamp (V) Vmax (V) TR (s) TF (s) 1 1.120±0.250 1.200±0.223 1.173±0.034 6.853±0.225 2 1.700±0.176 1.720±0.146 1.151±0.054 6.983±0.240 3 1.440±0.027 1.520±0.034 1.279±0.051 4.442±0.194 4 1.300±0.076 1.380±0.063 1.273±0.046 3.824±0.387 5 1.440±0.027 1.520±0.034 1.191±0.018 4.421±0.199 1 1.120±0.250 1.200±0.223 1.173±0.034 6.853±0.225 2 1.700±0.176 1.720±0.146 1.151±0.054 6.983±0.240 3 1.440±0.027 1.520±0.034 1.280±0.051 4.444±0.193 4 1.300±0.076 1.380±0.063 1.274±0.047 3.825±0.387 5 1.440±0.027 1.520±0.034 1.192±0.018 4.423±0.199 Table 8. Amplitude (V amp), maximum voltage (V max), rise time (T R) and fall time (T F) values of the background signal for 25.0 time/div. in the real and 1 1.300±0.009 1.400±0.028 1.603±0.005 5.406±0.240 2 1.200±0.093 1.220±0.114 1.582±0.018 6.098±0.099 3 1.240±0.058 1.320±0.030 1.584±0.017 6.352±0.055 4 1.100±0.192 1.140±0.192 1.687±0.044 8.250±0.187 5 1.720±0.237 1.720±0.209 1.600±0.007 7.416±0.095 1 1.300±0.009 1.400±0.028 1.603±0.005 5.408±0.240 2 1.200±0.093 1.220±0.114 1.583±0.017 6.100±0.099 3 1.240±0.058 1.320±0.030 1.584±0.017 6.357±0.054 4 1.100±0.192 1.140±0.192 1.687±0.044 8.250±0.187 5 1.720±0.237 1.720±0.209 1.600±0.007 7.418±0.095 Table 9. Amplitude (V amp), maximum voltage (V max), rise time (T R) and fall time (T F) values of the background signal for 50.0 time/div. in the real and 1 1.700±0.174 1.700±0.157 1.600±0.334 13.960±0.020 2 1.440±0.025 1.460±0.019 1.719±0.241 15.110±0.057 3 1.160±0.210 1.200±0.193 1.657±0.288 15.910±0.104 4 1.080±0.300 1.160±0.234 2.462±0.132 12.750±0.117 5 1.640±0.143 1.640±0.126 3.236±0.340 13.510±0.054 1 1.700±0.174 1.700±0.157 1.600±0.334 13.950±0.020 2 1.440±0.025 1.460±0.019 1.719±0.241 15.110±0.057 3 1.160±0.210 1.200±0.193 1.657±0.288 15.910±0.104 4 1.080±0.300 1.160±0.234 2.462±0.132 12.740±0.117 5 1.640±0.143 1.640±0.126 3.236±0.340 13.500±0.054

4 Gozde Tektas and Cuneyt CeliktasApplying to Signal Measurements in Scintillation Detectors. (a) Figure 3. Generator signal shapes in (a) the virtual and (b) the real oscilloscope for 250.0 time/div. (b) (a) Figure 4. Background signal shapes in (a) the virtual and (b) the real oscilloscope for 2.5 time/div. (b) When compared all results from the oscilloscopes above, it can be easily realized that voltage values (V amp and V max ) were very close to each other whereas very few differences on rise time and fall time values have been occurred in some measurements. Even so, it can be deduced from the overall results that the measurement performance of the virtual oscilloscope was quite satisfactory. 4. Conclusions A signal generator and the background radiation signals from a NaI(Tl) scintillation detector were separately analyzed. Amplitude, maximum voltage, rise time and fall time values of both type signals were determined for different time/div. adjustments. Obtained values from the designed virtual oscilloscope were highly comparable to those of the real oscilloscope as can be seen from tables (1-9). Also, it was observed that generator and background signals in both oscilloscopes were similar in shape. Finally, our designed virtual oscilloscope (VI) showed highly good performance in comparison with the real one. It was deduced from the obtained results that the developed virtual oscilloscope would reliably be able to use for the data acquisition as well as a real oscilloscope. This helps us to analyze many types of detector signals in scintillation detectors.. Acknowledgment This work was supported Scientific Research Foundation of Ege University under project No. 14 FEN 052. Additionally, the Authors thank to Dr. Jiri Pechousek for his precious help. References [1] Sumathi, S., Surekha, P., 2007. LabVIEW based Advanced Instrumentation Systems. Springer Berlin Heidelberg New York. [2] Pechousek, J., 2011. Application of Virtual Instrumentation in Nuclear Physics Experiments in Practical Applications and Solutions Using LabVIEW, InTech. [3] Pechousek, J., Prochazka, R., Prochazka, V., Frydrych, J., 2011. Virtual instrumentation technique used in the nuclear digital signal processing system design: Energy and time measurement tests. Nuclear Instruments and Methods in Physics Research A. 637, 200-205.

Radiation Science and Technology 2015; 1(1): 1-5 5 [4] Gong, P., Zhou, W., 2012. Design and Implementation of Multifunctional Using USB Data- Acquisition Card. Procedia Engineering. 29, 3245-3249. [5] NI LabVIEW Measurements Manual. http://fy.chalmers.se/~f7xlh/lv_kurs/measur_man_2000.pdf [6] Travis, J., Kring, J., 2006. LabVIEW for Everyone: Graphical Programming Made Easy and Fun. Third Edition, Prentice Hall, U.S.A. [7] Shinde, S., Prabhu, S., 2013. Labview Based Digital CRO For Electronic Measurement Techniques. IJERA. 3 (1), 693-698. [8] GDS-2000 Series Digital Oscilloscope User Manual. https://fenix.tecnico.ulisboa.pt/downloadfile/3779578030076/ GDS-2000_User_Manual_20080416[1].pdf [9] Leo, R.W., 1987. Techniques for Nuclear and Particle Physics Experiments. Springer Verlag Berlin Heidelberg, Germany. [10] ShengLi, D., JunFeng, H., ShengHui, P., JianFeng, D., 2011. Design of Based LAbVIEW. Springer- Verlag Berlin Heidelberg. 34-41.