Confocal Application Notes Vol. 5 July 2010

Similar documents
Leica TCS SP8 Quick Start Guide

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Leica TCS SP8 Quick Start Guide

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

LSM 800 Confocal Microscope Standard Operation Protocol

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

Leica SP8 TCS Users Manual

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Cell Biology and Bioimaging Core

Swept-Field User Guide

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

LSM 780 Confocal Microscope Standard Operation Protocol

TRAINING MANUAL. Olympus FV1000

Supplemental Method Information Zeiss LSM710

Things to check before start-up.

Title: Leica SP5 Confocal User Manual

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Stitching MetroPro Application

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Zeiss LSM 780 Protocol

Zeiss 880 Training Notes Zen 2.3

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Quick Start Guide. Leica SP5 X

OPERATING INSTRUCTIONS

ZEN 2012 SP5 black edition Hotfix 12

Nikon SIM-E & A1-R System

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

Zeiss 780 Training Notes

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

LSM 710 Confocal Microscope Standard Operation Protocol

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Practical work no. 3: Confocal Live Cell Microscopy

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Leica SP8 TCS Users Manual

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Grid Assembly. User guide. A plugin developed for microscopy non-overlapping images stitching, for the public-domain image analysis package ImageJ

Working With Drawing Views-I

Zeiss LSM 880 Protocol

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Guide to Confocal 5. Starting session

The Zeiss AiryScan System, Confocal Four.

Leica Sp5 II Confocal User Guide

Nikon A1Rsi Confocal Start-Up Sequence

Microscopy from Carl Zeiss

Instructions for the Leica SP5 II laser scanning confocal microscope

Title: Nikon A1R Confocal User Manual

Fire CR Calibration Guide

Further Mathematics Support Programme

Zeiss AxioImager.Z2 Brightfield Protocol

Instructions for the Experiment

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Operating Instructions for Zeiss LSM 510

The image regions are all in focus, producing no contrast resulting from changes in signal level.

Generations Automatic Stand-Alone Lace By Bernie Griffith Generations Software

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

Leica SPEII confocal microscope. Short Manual

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

Tiling. 1. Overlapping tiles with fixed number of tiles. Tutorial

Renishaw InVia Raman microscope

LSM 510 Meta Training Notes

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

How to combine images in Photoshop

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

LSM 510 Training Notes

Leica DMi8A Quick Guide

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Getting Started. with Easy Blue Print

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

NCSS Statistical Software

Microscope Confocal Sp2 Upright.

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Confocal Application Letter No. 13. Sequential Scan for Leica TCS NT/SP systems

2 How to operate the microscope/obtain an image

b. Turn the power switch and key to on position for blue laser.

Operating Rausch ScanCam within POSM.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

User manual for Olympus SD-OSR spinning disk confocal microscope

ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur

Focus Strategies with ZEN2. Sven Terclavers ZEISS Embedded 3D Imaging Specialist Harvard University,

Lab 3: Low-Speed Delta Wing

Image-Pro Plus 7.0 Product Note

Leica SP8 Resonant Confocal. Quick-Start Guide

4 Use the adjustable Focus meter tool to take the subjectivity out of focusing the image, to get the best possible image

Software Manual for the Economy LBIC Demo

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

Information & Instructions

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

Transcription:

Tile Scan Prepared by Myriam Gastard, PhD Application and Technical Support Group, Leica Microsystems, Inc. In this issue of our Confocal Application Notes, proper set up of the Tile function enables you to scan multiple partial images from a specimen and stitch them together to form a complete image. By extension, this feature makes possible to scan a much larger field of view than a high magnification objective alone would be able to give you, and still keeping the high resolution of the objective. In order to have all the images aligned in x, y and possibly z (when z-stack are acquired simultaneously), and to avoid any gap and/or misalignment of the resulting tile, you must precisely adjust the tile scan. One important tool supporting this feature is the Tile Scan Calibration. First requisite is to have the stage initialized at startup of LASAF. Place a sample onto the stage and setup your system in order to be able to image the sample (laser, PMT, gain etc ). To assure a precise calibration, an oil objective with a magnification of at least 40x, and high numerical aperture, is recommended. If for any reason an oil objective cannot be used, then a 20x dry is appropriate. You should zoom in order to have enough details to adjust accurately the different structures of interest. 1- In acquisition mode select the Tile Scan function. Acquire Acquisition Select the Tile Scan function 2- When Tile Scan function is selected, a drop down window named Tile Scan appears: Tile Scan window interface 1

3- Click on the Calibration function. 4- A new interface window Tile Scan Calibration appears. 2

5- Activate the Live imaging and click on the preview image in order to activate the marked line. This will be used as reference in order to adjust the rotation. 6- While in Live mode zoom in to find a structure in your sample that you can use as a horizontal reference point. Reference structure 3

7- Using the Smartmove and xy precise move the structure along the line and pay attention to its relative position to the marked line. Reference structure moved on the other side of the field of view. If the sample needs to have some rotation adjustment (like shown in this example), the structure will not be parallel when moved from side to side of the field of view. 4

8- Adjust the rotation by moving the rotation cursor or by adjusting the degrees numbers in the window while in Live mode. Check the reference point by moving the structure back and forth along the marked line and be sure that it is parallel to the marked line. 5

9- Once the new rotation position is found, stop the live scanning and click on the Run Test tab, then click on the Run Test button. If a region of the resulting tile scan is showing some tiling discrepancies (like in this image noted in yellow), then go to Set parameters. After each change, the Run Test needs to be activated and a full tiling is taken. Adjust then the parameters in order to compensate accordingly. 6

Once the Tiling parameters have been correctly modified, you should be able to get a very good tiling. 10- Once the Set parameters is chosen, the Stage Configuration window appears: 1 2b 2a 7

In this window, 2 different areas may need adjustments: 1- Tile Scan: set overlap sizes to: X % and Y% The percentage of overlap needed to have the best tiling can be adjusted by entering positive values corresponding to the X and Y axes overlap. Try to change one parameter at the time in order to understand which way it will go. 2a- Tile scan: set merge settings to: Factory and Custom Here are the settings needed to assemble the tiling. It is essential to change only one parameter at the time. Factory: It is by default on Factory settings. Custom: If the factory settings are not compensating completely (or the tiled image seems to be scrambled), then these settings must to be adjusted. You can flip the X, Y or even swap XY axes in order to reorder the tiles. 2b- Tile scan: set merge settings to: Smooth, Multi frame Search, Auto Stitching The stitching process can be started either automatically right after the tile scan acquisition, or any time post acquisition in Process Mosaic Merge tab. To enable the stitching right after acquisition, check the Merge box in Tile Scan interface. In this case, the Auto-Stitching/Smooth parameters are set in Configuration Stage tab. Smooth: It enable the stitching of the images together with a function which will even the sharp grid lines (after stitching), in order to make it disappear from human eyes. The Smooth function can be turned on or off during Auto-Stitching and it will not affect the image alignment. Multi frame Search: This function applies the smooth and/or Auto Stitching functions to every sections in x, y and z allowing a better stitching at every level of the merged sections. The calculations will make this step slower than if smooth and/or Auto Stitching functions are selected alone. Auto Stitching: When this box is checked prior the Tile Scan acquisition, it allows having a full tiled image without having to request it in Process Mosaic Merge and hence acting as a short cut for this post-processing method. 8

11- Tricks and Tips: To enable the stitching right after acquisition, check the Merge box in Tile Scan interface. If the Merged Tile Scan is poor, try to open a little bit the pinhole to 1.5 or 2 Airy Units. When using a low magnification objective (5x or 10x) the corners of each tile scan can appear less illuminated, which can be very noticeable once the Tile Scans are merged. To resolve this, please use the zoom in function (~ 1.7 zoom in) to obtain a better field uniformity. 9