Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Similar documents
Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

WITec Alpha 300R Quick Operation Summary October 2018

Renishaw InVia Raman microscope

RENISHAW INVIA RAMAN SPECTROMETER

Confocal Raman Microscopy (WITec Alpha 300R)

Instructions for the Experiment

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

SUPPLEMENTAL MATERIAL

Leica TCS SP8 Quick Start Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

LumaSpec 800S User Manual

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

UltraGraph Optics Design

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Renishaw InVia Quick Operation Summary October 2018

Supplementary Materials

DISCO DICING SAW SOP. April 2014 INTRODUCTION

CONFOCAL MICROSCOPE CM-1

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

IBIL setup operation manual for SynerJY software version

Zeiss AxioImager.Z2 Brightfield Protocol

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Fast Laser Raman Microscope RAMAN

Leica TCS SP8 Quick Start Guide

Practical work no. 3: Confocal Live Cell Microscopy

Swept-Field User Guide

Quick Start Guide. Leica SP5 X

ENSC 470/894 Lab 3 Version 6.0 (Nov. 19, 2015)

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.

STEM Spectrum Imaging Tutorial

Zeiss LSM 880 Protocol

Improving the Collection Efficiency of Raman Scattering

Exp. No. 13 Measuring the runtime of light in the fiber

Fast Laser Raman Microscope RAMAN

LSM 710 Confocal Microscope Standard Operation Protocol

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Guide to Confocal 5. Starting session

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Standard Operating Procedure

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

Figure 1. Oil-immersion objectives available for use with the Lionheart FX.

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Sabeeh Irfan Ahmad, Physlab, 23 July 2016

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

LSM 510 Meta Training Notes

FEI Tecnai G 2 F20 Operating Procedures

University of Wisconsin Chemistry 524 Spectroscopic Components *

Be aware that there is no universal notation for the various quantities.

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Things to check before start-up.

Characterization Microscope Nikon LV150

Basic Users Manual for Tecnai-F20 TEM

Title: Leica SP5 Confocal User Manual

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Tools Needed 3/32 Allen Wrench which is located in your accessory kit Masking Tape

Leica SP8 TCS Users Manual

2. Refraction and Reflection

LSM 510 Training Notes

Exercise 8: Interference and diffraction

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

MS260i 1/4 M IMAGING SPECTROGRAPHS

Nasmyth Ultraview Vox User Protocol

Oriel MS260i TM 1/4 m Imaging Spectrograph

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

3B SCIENTIFIC PHYSICS

Motorized Axio Observer Start-up instructions

b. Turn the power switch and key to on position for blue laser.

Standard Operating Procedure

LEICA TCS SP5 AOBS TANDEM USER MANUAL

RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Microscopy from Carl Zeiss

3B SCIENTIFIC PHYSICS

Zeiss 880 Training Notes Zen 2.3

Operating Instructions for Zeiss LSM 510

User manual for Olympus SD-OSR spinning disk confocal microscope

Zeiss LSM 780 Protocol

Instructions for Tecnai a brief start up manual

Reference and User Manual May, 2015 revision - 3

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Spotlight 150 and 200 FT-IR Microscopy Systems

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter.

Systematic Workflow via Intuitive GUI. Easy operation accomplishes your goals faster than ever.

University Libraries ScanPro 3000 Microfilm Scanner

Supplementary Figure 1

TRAINING MANUAL. Olympus FV1000

Speed of light E Introduction

Leica Sp5 II Confocal User Guide

LSM 780 Confocal Microscope Standard Operation Protocol

2 How to operate the microscope/obtain an image

Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope. Advanced Lab Course University of Basel

FEI Falcon Direct Electron Detector. Best Practice Document

Raman images constructed from. Raman Imaging: Defining the Spatial Resolution of the Technology

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Transcription:

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Please contact Dr. Amanda Henkes for training requests and assistance: 979-862-5959, amandahenkes@tamu.edu Hardware LN 2 FTIR FTIR camera 1 optics box spectrometer InGaAs detector LN 2 objective turret 2 3 4 5 objective lens Raman camera CCD detector Laser Diode Micro OFF Scanner Camera Laser CCD joystick microscope stage Superhead InGaAs STEMS: 1. UP: measurement; DOWN: video 2. IN: Raman acquisition/video; MIDDLE: FTIR acquisition/video; OUT: look through eyepieces 3. Always IN (OUT is for superhead) 4. Aways OUT 5. IN: InGaAs detector; OUT: CCD detector Start-up procedure 1. Fill detector with liquid nitrogen (CCD or InGaAs) 2. Turn 633 laser key and flip on laser switch on silver control box. Ensure that shutter is on micro for microscope; detector on CCD; diode off; camera on; scanner is nonfunctional on our system LabSpec software 1. Open LabSpec CCD software on computer 2. On top toolbar, choose Acquisition Detector. A window will open after synching the computer with the detector displaying the detector temperature. Ensure CCD is selected and check that the temperature is -130 C or lower. If you haven t already added liquid nitrogen, top off the CCD detector dewar. Within 1 hr of filling the dewar, it should cool to optimal operating temperature. **Never change any of the numerical values in this window** 1

Setting acquisition parameters **Exercise the motor for each parameter on the bottom toolbar at the beginning of your session!** 1. Set acquisition parameters on bottom toolbar -Laser: choose 633 nm or 785 nm. Choosing the appropriate laser ensures proper scale on X axis -Filter: optical density filter; controls the intensity of the laser light that hits your sample; higher number = more light blocked out (i.e., D 4 has a very feint beam intensity whereas D 0.3 is very strong) -Hole: confocal pinhole; controls the thickness of the plane of focus; larger pinhole size = thicker focal plane -Spectrometer: spectral position of the selected Laser 633 nm 785 nm Filter Power for 200 µm pinhole 10x 50x 50x SLM --- 8.61 mw 7.87 mw 8.00 mw D 0.3 4.39 mw 4.02 mw 4.01 mw D 0.6 1.91 mw 1.75 mw 1.75 mw D 1 660 µw 613 µw 613 µw D 2 94.7 µw 86.6 µw 78.4 µw D3 12.3 µw 11.8 µw 10.9 µw D4 1.45 µw 1.43 µw 0.96 µw --- 22.4 mw 12.0 mw 4.16 mw D 0.3 16.1 mw 11.3 mw 1.63 mw D 0.6 7.94 mw 6.23 mw 840 µw D 1 2.51 mw 1.87 mw 388 µw D 2 265 µw 201 µw 39.0 µw D 3 50.9 µw 42.4 µw 5.25 µw D 4 3.36 µw 3.32 µw 0.42 µw grating. The grating will move along the spectral range in several steps. This number indicates the center of the grating s position. It also refers to the spectral window for RTD acquisition (see table below for spectral windows for each grating/laser). You can type in a number or press the left arrow to go the zero position or the right arrow to go to the diode position. -Grating : 300 or 1800 lines per mm. The higher number indicates better spectral resolution. -Objective : x10, x50, x100, x50slm or x50leica; ensure appropriate objective is selected to correlate with objective in sampling position on the objective turret (only important if doing mapping) -Tag : name given to the spectrum -RTD exposure time : duration of RTD scan that will continually refresh and only scan within the spectral window selected in Spectrometer -Extended scan exposure time : exposure time during an extended scan for each spectral window within the range chosen in Extended Range ( ). longer exposure time = higher signal intensity -Accumulations : number of times the extended scan will be repeated. Accumulations are averaged together, thus creating better contrast between the signal and noise. more accumulations = less noise -XYZ Coords: X, Y and Z position of the stage and objectives. 2

initialize 2. Initialize grating to ensure that it is properly zeroed 3. Choose the Extended Range icon from the top toolbar and set the scan range. Mounting and focusing your sample ***Joystick: twist to raise/lower objectives (clockwise=down, counterclockwise=up); move sideways or up/down to translate stage in XY plane. ALWAYS MOVE OBJECTIVES SLOWLY!*** 1. Raise joystick to highest position. 2. Pull STEM 2 out slightly to block laser from shining on your hand 3. Insert sample mounted on glass microscope slide (or other similar solid support) onto stage. Secure with clips. 4. Push STEM 2 completely in to open laser pathway to sample. 5. Move 10x objective (yellow) into the sample position. Always focus with 10x first before moving to higher magnifications. The objectives have been mounted and adjusted so that they are nearly par-focal (meaning they will all be in focus at the same Z-position). Because the 50x and 100x have such short working distances, it is safer for our microscope (and easier for users) to first focus using 10x, then switch to 50x, slightly adjust Z-position (down), then switch to 100x, and slightly adjust Z-position again (down). ***NEVER remove an objective lens for the turret; if you want to use your OWN objective, please consult me first so I can calibrate it and best advise you where to mount it!*** Diffraction-limited spatial resolution on video Objective Lens 633 nm 785 nm Working laser laser Distance Coverslip Olympus MPlan 10x, N.A. 0.25 1544 nm 1915 nm 10.6 mm no Olympus MPlan 50x, N.A. 0.75 514 nm 638 nm 0.38 mm no Olympus MPlan 100x, N.A. 0.90 429 nm 532 nm 0.21 mm no Olympus SLMPlan 50x, N.A. 0.45 858 nm 1064 nm 15.0 mm no **Objective lens spatial resolution is dependent on numerical aperture (N.A.) and laser wavelength: Resolution = (wavelength)(0.61)/(n.a.)** Beam cross-sectional diameter in microns (mapping spatial resolution) Pinhole (µm) Objective 10x 50x 100x 50x SLM 50 8 2.5 1 2 75 15 5 1.4 4 100 30 7.5 1.8 7 125 44 10 2.2 9 150 55 14.5 2.5 11.5 175 65 17 3 14.5 200 75 20 3.7 17.5 250 100 25 4.2 21 **Beam cross-section diameter correlates to the sampling area for each spectrum** 3

Focusing with visible light (for visible image acquisition) 1. On the top toolbar, go to Video and ensure that Micro Camera is checked 2. Turn on visible light source. 3. Flip off laser switch (controls shutter) on silver control box. 4. Lower STEM 1 to enable video visualization. 5. Open the video screen in the LabSpec software (camera icon on top toolbar ). You will see a reflected light visible image of your sample. 6. Lower the objective until the image appears in focus. When the objective is just above the focal point, you will see an image of the fiber optic cable that supplies the visible light source. Continuing to lower the objectives just slightly past this point will bring your sample into focus. 7. Press the stop sign on the LabSpec software on the top toolbar (far right) to stop video acquisition. 8. Press the disk icon to save the image. fiber optic image sample surface Focusing with visible light (through the microscope eyepieces) 1. Turn on the visible light source. 2. Pull STEM 2 all the way out to enable eyepiece visualization of the sample. 3. Looking through the eyepieces, lower the objective until you see the sample in focus. When the objective is just above the focal point, you will see an image of the fiber optic cable that supplies the visible light source. Continuing to lower the objectives just slightly past this point will bring your sample into focus. Focusing with the laser (for Raman acquisition): 1. On the top toolbar, go to Video and ensure that Micro Camera is checked 2. Lower STEM 1 to enable video visualization. 3. Open the video in the LabSpec software (camera icon laser focused on sample surface on top toolbar ). You should see a red-tinted noisy background for the 633 nm laser, and black/white noisy background for the 785 nm laser. 4. Lower the objective (twist the joystick clockwise) until you see the white, circular laser spot condense into the center of the video screen. Try to shrink the white laser spot until it s as small as possible by slowly lowering the objective. A completely condensed laser spot means that the laser is in focus on the SURFACE of your sample. **Over-focusing (objective lower) will be focused underneath the surface of your sample, whereas under-focusing (objective higher) will be focused above your sample.** 5. Press the stop sign on the top toolbar (far right) to stop video acquisition and close the video window. **Not closing the video window often causes it to freeze the next time you start the video.** 4

Acquiring spectra 1. Extended scan is for scanning the full selected range (as input in Extended Range ), usually with longer exposure times and several accumulations. This spectrum can allow you to see low intensity peaks. This is the typical type of acquisition used for measuring spectra for publication. 2. RTD scan is for quick analysis. The scan will be continually refreshed and will only be able to scan within the selected grating s spectral window (see table above). This is mainly used to see if you are focused on a particular spot on your spectrum. (Some users like to adjust the Z-position with the joystick to maximize the intensity of a peak of interest.) You will only see very intense peaks in the spectrum. Analyzing Spectra 1. After collecting a spectrum, you can analyze the position of the peaks by using the peak search tool. You can adjust the intensity and width of peaks to be labeled by dragging the Level and Size slider bars. Pressing Search labels all peaks within click & drag chosen height/width range. Pressing Clear removes all previous peak labels. You can choose the Gaussian or Lorentzian shaped peaks in the Function field. (LabSpec uses these characteristic shapes to find and fit the peaks of your spectrum.) 5

Saving Spectra After you have labeled/analyzed your spectrum, you can save it in several ways: 1. Select spectrum you wish to save on right toolbar by clicking its colored circle 2. Choosing File Save will create either a.txt or.ngs file of each spectrum displayed on the screen. (If you don t wish to save a spectrum, it s easiest to delete it before you press save by selecting it and choosing the trashcan icon.) A save window will open for the first spectrum and it will have the name you gave it in the tag field. After choosing a save location and adjusting the name if necessary, clicking Save will save that spectrum and open a new save window for the next spectrum. This will occur for each spectrum as the program goes down the list of all the spectra you have displayed. **A.ngs file can only be opened in LabSpec; a.txt file can be opened in both LabSpec and in any spreadsheet program.** 3. You can write the spectrum to a PDF file by choosing File Print. It will save the screen shot of all active spectra. (If you have the 1 button activated on the right toolbar, only one spectrum is visible on the screen. In this case, only the active spectrum is saved as a PDF. If you have the + button activated on the right toolbar, all spectra will be seen on the screen and saved in one PDF image.) This saving option removes peak labels. 4. You can also save a screen shot of the spectra by pressing Ctrl+C when the mouse is in the spectrum window. You can then paste this screen shot into a WordPad or Paint file. This option retains the peak labels. 5. After you save your data, you can either email the files to yourself (the computer has internet access) or you can retrieve them with a USB drive. Shutdown procedure 6

1. Rotate to 10x objective (yellow) 2. Raise objective 3. Remove sample 4. Save spectra 5. Close LabSpec 6. Turn off laser (key switch for 633 nm, both buttons on Pilot box for 785 nm) 7. Flip laser switch off Switching lasers 1. Make sure all lasers are turned off **Do not go behind Raman table when lasers are on; there is risk of exposure to laser radiation** 2. In the optics box behind the microscope, carefully slide back door and remove magnetic lid. 3. For the 633 nm laser, ensure the mirror is in the UP position; for the 785 nm laser, ensure that it is in the DOWN position 4. Make sure the correct edge filter is in place. Carefully and SLOWLY remove unwanted filter and replace with the desired filter. Put unused filter in the EXACT SAME SPOT the other filter was stored. (This location is out of the beam path.) **Make sure that you ONLY touch the filter and the mirror inside the optics box, lest other mirrors be bumped out of alignment** 5. Turn on appropriate laser: -633 nm: key on silver control box; ensure laser switch is on -785 nm: on the pilot box; press Line and then Enable (cougar button); ensure the laser switch is on 6. Choose the appropriate laser on the bottom toolbar in LabSpec. Analyzing Liquid Samples 1. Fully raise objectives 2. Remove objective turret using allen wrench 3. Switch objective turret to the one with the cuvet holder; always put unused turret in plastic storage box 4. Ensure cuvet holder is in beam path by rotating turret until it is vertical 5. Fill cuvet with sample 6. Insert cuvet into holder 7. Make sure STEM 1 is in the up position for Raman acquisition (no focusing necessary) 8. Make sure STEM 2 is fully in so the laser hits your sample 9. Adjust acquisition parameters as above 10. Acquire spectra as above **There is no focusing the laser with the cuvet holder** 7

Troubleshooting 1. The peaks seem to be in the wrong position. -Check that you selected the appropriate laser on the bottom toolbar of LabSpec. -Check that the units of the y-scale are in wavenumbers (cm -1 ) and not nanometers. On the top toolbar of LabSpec, choose Options Units 1/cm. -Check the location of the silicon peak by mounting the standard Si sample (small rectangle of Si glued onto a glass slide). When using either laser and the 1800 grating, it should appear at 520.5 cm -1 ± 1 cm -1. If it is not at this location, 1) re-initialize the grating; if that doesn t solve the problem 2) put in the D3 or D4 filter and move the Spectrometer position to zero (press the left arrow). When the spectrometer arrives at the zero position < 0 >, observe the RTD spectrum (set RTD exposure to 1 sec). If the laser s peak is at exactly zero (full width at half max, FWHM; not apex), then the Si peak should be in the proper position. If it is not, please tell me so that I can help you recalibrate the grating. 1. I don t see any signal. -Check that you have the STEM 1 in the sampling position (up) -Check that STEM 2 is pushed all the way in -Check that the laser has been properly focused on the sample -Increase the exposure time 2. My spectrum is very noisy. -Increase the number of accumulations. Adding more accumulations will average out the noise, thus increasing your ability to distinguish your signal from the noise. -Check that the laser is properly focused initialize 3. I see a very intense background in my spectrum, particularly at low wavenumbers. -Your sample may fluoresce. Try changing to the 785 nm laser to see if this helps. -You are using the 633 nm laser with the 300 grating at low wavenumbers. The filter for this laser using this grating does not perform well at low wavenumbers. Try switching to the 1800 grating or the 785 nm laser. 8