Measurement of the Modulation Transfer Function (MTF) of a camera lens. Laboratoire d Enseignement Expérimental (LEnsE)

Similar documents
PROCEEDINGS OF SPIE. Measurement of the modulation transfer function (MTF) of a camera lens

Optical design of a high resolution vision lens

Exam 4. Name: Class: Date: Multiple Choice Identify the choice that best completes the statement or answers the question.

ECEN 4606, UNDERGRADUATE OPTICS LAB

Physics 3340 Spring Fourier Optics

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Fast MTF measurement of CMOS imagers using ISO slantededge methodology

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

WaveMaster IOL. Fast and Accurate Intraocular Lens Tester

Be aware that there is no universal notation for the various quantities.

Optical basics for machine vision systems. Lars Fermum Chief instructor STEMMER IMAGING GmbH

Optical Performance of Nikon F-Mount Lenses. Landon Carter May 11, Measurement and Instrumentation

LEOK-3 Optics Experiment kit

Design Description Document

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

WaveMaster IOL. Fast and accurate intraocular lens tester

Supplementary information Phase-sensitive plasmonic biosensor using a portable and large field-of-view interferometric microarray imager

Aberrations of a lens

Testing Aspheric Lenses: New Approaches

Imaging Optics Fundamentals

Very short introduction to light microscopy and digital imaging

MASSACHUSETTS INSTITUTE OF TECHNOLOGY. 2.71/2.710 Optics Spring 14 Practice Problems Posted May 11, 2014

An Indian Journal FULL PAPER. Trade Science Inc. Parameters design of optical system in transmitive star simulator ABSTRACT KEYWORDS

LENSES. a. To study the nature of image formed by spherical lenses. b. To study the defects of spherical lenses.

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

OptiSpheric IOL. Integrated Optical Testing of Intraocular Lenses

Some of the important topics needed to be addressed in a successful lens design project (R.R. Shannon: The Art and Science of Optical Design)

1 Laboratory 7: Fourier Optics

This experiment is under development and thus we appreciate any and all comments as we design an interesting and achievable set of goals.

Sensitive measurement of partial coherence using a pinhole array

SUPPLEMENTARY INFORMATION

EE-527: MicroFabrication

BEAM HALO OBSERVATION BY CORONAGRAPH

ECEN 4606, UNDERGRADUATE OPTICS LAB

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Bias errors in PIV: the pixel locking effect revisited.

Improving the Collection Efficiency of Raman Scattering

Chapter 25. Optical Instruments

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Confocal, hyperspectral, spinning disk

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

Microscope anatomy, image formation and resolution

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Mechanical Engineering Department. 2.71/2.710 Final Exam. May 21, Duration: 3 hours (9 am-12 noon)

Image Formation and Capture. Acknowledgment: some figures by B. Curless, E. Hecht, W.J. Smith, B.K.P. Horn, and A. Theuwissen

Determination of Focal Length of A Converging Lens and Mirror

ADVANCED OPTICS LAB -ECEN Basic Skills Lab

EUV Plasma Source with IR Power Recycling

Laser Speckle Reducer LSR-3000 Series

Phys 102 Lecture 21 Optical instruments

Laboratory experiment aberrations

Introduction to Optics Work in Y1Lab

plasmonic nanoblock pair

Open Access Structural Parameters Optimum Design of the New Type of Optical Aiming

Overview: Integration of Optical Systems Survey on current optical system design Case demo of optical system design

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

Properties of optical instruments. Projection optical systems

Week IV: FIRST EXPERIMENTS WITH THE ADVANCED OPTICS SET

Optics: An Introduction

Chapter Ray and Wave Optics

Test procedures Page: 1 of 5

Diffraction. Interference with more than 2 beams. Diffraction gratings. Diffraction by an aperture. Diffraction of a laser beam

Defocusing Effect Studies in MTF of CCD Cameras Based on PSF Measuring Technique

The spectral colours of nanometers

Εισαγωγική στην Οπτική Απεικόνιση

Collimation Tester Instructions

Lens Design I. Lecture 5: Advanced handling I Herbert Gross. Summer term

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

Lecture Notes 10 Image Sensor Optics. Imaging optics. Pixel optics. Microlens

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

P202/219 Laboratory IUPUI Physics Department THIN LENSES

Chapter 18 Optical Elements

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope

Lens Design I Seminar 5

OPTICAL SYSTEMS OBJECTIVES

Image Formation. Light from distant things. Geometrical optics. Pinhole camera. Chapter 36

Telecentric lenses.

ECEG105/ECEU646 Optics for Engineers Course Notes Part 4: Apertures, Aberrations Prof. Charles A. DiMarzio Northeastern University Fall 2008

Study on Imaging Quality of Water Ball Lens

Resolving Power of a Diffraction Grating

OPAC 202 Optical Design and Instrumentation. Topic 3 Review Of Geometrical and Wave Optics. Department of

Phy Ph s y 102 Lecture Lectur 21 Optical instruments 1

Fiber Optic Communications

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature:

OCT Spectrometer Design Understanding roll-off to achieve the clearest images

Converging and Diverging Surfaces. Lenses. Converging Surface

Optical Design with Zemax for PhD - Basics

Mirrors and Lenses. Images can be formed by reflection from mirrors. Images can be formed by refraction through lenses.

Chapter 34 The Wave Nature of Light; Interference. Copyright 2009 Pearson Education, Inc.

LENSES. INEL 6088 Computer Vision

Optical System Design

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

LOS 1 LASER OPTICS SET

Point Spread Function Estimation Tool, Alpha Version. A Plugin for ImageJ

Transcription:

Measurement of the Modulation Transfer Function (MTF) of a camera lens Aline Vernier, Baptiste Perrin, Thierry Avignon, Jean Augereau, Lionel Jacubowiez Institut d Optique Graduate School Laboratoire d Enseignement Expérimental (LEnsE) 2 Avenue Augustin Fresnel, 91127 Palaiseau (France) Abstract An accurate characterization of optical elements is one of the pivotal skills to be learnt in the context of undergraduate optics lab work or that of continuing education. We will show an experimental apparatus designed to measure the Modulation Transfer Function (MTF) of a camera lens. The educational scope of this apparatus ranges from the study of the aberrations of the lens and its suitability for use with a given sensor, to the design of the test bench itself. A first, purely visual observation allows the trainees to single-out the aberrations exhibited by an objective. The presence of the spherical and chromatic aberrations on the axis of the objective is thus identified, and a first estimate of the cut-off frequency of this objective can be given. There, the importance of the geometrical extent of the source and the impact of the optical elements of the test bench on the measurement can also be assessed. The second part of the experiment makes use of a CMOS camera to measure the Linear Spread Function (LSF) of the camera lens and the MTF is then retrieved by Fourier transform. The experimental setup consists of a fine slit as a source, a collimator, the studied objective exhibiting conspicuous aberrations and a microscope objective to re-image the LSF either for direct observation with an eye-piece or for analysis with the camera. The latter is performed with a homedeveloped Matlab software. Keywords: Line Spread Function, characterization of optical elements, optical test bench for aberration measurements, Modulation Transfer Function, cut-off frequency and sampling frequency.

1. Scope of the experiment This apparatus allows our students to measure the modulation transfer function of a camera lens with f = 25mm and N min = 0.95 which is used with a CMOS chip with 5472 3648 pixels and a pixel size of 2.4 µm. Such a delicate measurement is also a good opportunity to demonstrate several key aspects of optical instrumentation, namely, the implementation of a linear source for a given optical system, the impact of the finite size of the source on the measurement, and the choice of the optical elements to image the response of the studied lens on the camera. The linear source is a practical solution to measure the MTF with light sources which would not be bright enough for use with a point source. The MTF thus measured is in 1D, orthogonally to the direction of the line. This can be proven as follows, given a source SS(xx, yy) = δδ(xx). CC and a lens of diameter aa: The response of the objective can be expressed as the square of the Fourier transform of the product of the source with the aperture of the lens RR 2 kk xx, kk yy, with kk xx, kk yy the spatial frequencies associated with resp. (xx, yy). RR kk xx, kk yy = aa/2 aa/2 ee ii kk xx xx+kk yy yy δδ (xx). CC dddddddd Recalling dd HH(xx) dddd = δδ(xx), and integrating by parts yields: RR 2 kk xx, kk yy sin2 aa kk yy aa kk yy 2 This corresponds to the Linear Spread Function (LSF). The Fourier Transform of the LSF then gives the 1D MTF in the yy-direction. 2. Experimental apparatus Collimator f = 500mm, Ø = 100mm Source Microscope objective Camera lens CMOS camera Figure 1: Experimental apparatus picture

Microscope objective 40 + tube + camera Ground plate Adjustable slit LED array Collimator f = 500mm Ø = 100mm Camera lens f = 25mm Nmin = 0.95 The setup shown on Figure 2 goes as follows: Figure 2: Experimental apparatus schematic Linear source: bright, white or coloured LED light is first diffused by a ground glass diffuser to homogenize the illumination of a thin adjustable slit which acts as a linear source Shaping: the source is collimated with a collimator with a focal length ff = 500 mm and a diameter = 100 mm Camera lens Measurement apparatus: a microscope objective x40 re-images the focal spot onto a CMOS chip with 1.3 10 6 square, 5.3 µm pixels. 3. Preliminary considerations Prior to diving into the experiment, the students must work out the best theoretical specifications of the studied camera lens: the MTF cut-off frequency ff cc and the size of the point spread function ΦΦ AAAAAAAA. For λ = 630 nm, N = 0.95 and γγ the magnification of the optical system: ff cc = 1 λλ NN = 1671 mm 1 ΦΦ AAAAAAAA 2.44 λλ NN 1.5 µm Because the camera has a pixel size of 2.4 µm, the system (lens + camera) is thus theoretically limited by the pixel size (ff cc,cchiiii = 417 mm 1 ). From ΦΦ AAAAAAAA they can deduce the maximum size of the point source dd, given the optical system. dd γγ ΦΦ AAAAAAAA Given that the focal length of the collimator is 500 mm and that of the objective is 25 mm, the magnification γγ = 20. Therefore, dd 30 µm in order to study the camera lens at full aperture. The students are then asked to work out the sampling spatial frequency of their measurement apparatus to verify that the combination of the x40 microscope objective and the CMOS chip with 1.3 10 6 square, 5.3 µm pixels is appropriate to measure the MTF of the camera

lens. A quick calculation yields a sampling frequency of about 7500 mm 1, which is clearly sufficient. Another important point concerns the choice of the collimator and the microscope objective. It goes without saying that the two must be devoid of aberrations. Moreover, the collimator aperture must be significantly larger than that of the camera lens, and the object numerical aperture of the microscope objective must obviously be greater than the image numerical aperture of the lens not to limit the measurement. 4. Observations and measurements Before making use of the camera and the analysis software, it is informative to make visual observations of the linear spread function. This allows the students to get an intuitive grasp of the aberrations, estimate the size of the LSF and for which aperture size the objective starts to be diffraction-limited. A more in-depth quantitative analysis is then carried out using the Matlab software provided, the interface of which is shown on Figure 3. Image integrated vertically Ideal MTF and computed MTF Acquisition properties and MTF computation parameters Image on camera Figure 3: Matlab software interface After a careful calibration of the magnification of the imaging of the focal spot onto the camera and tuning the exposure time to an appropriate value, the students can view the LSF (e.g. Figure 4) and the measured MTF (e.g. Figure 5) in real time. This allows them to identify the maximum aperture for which the camera lens is diffraction-limited and to quantify the effects of the

different aberrations, for example the longitudinal displacement of the best focus for different aperture sizes or for different wavelengths, on- and off-axis. Figure 4: On-Axis Linear Spread Function (integrated vertically from camera image) λλ = 530 nm, NN = 8 Measured MTF Ideal MTF Figure 5: Modulation Transfer Function, λλ = 530 nm, NN = 8

5. Conclusion We have shown an undergraduate lab setup designed to perform the measurement of the modulation transfer function in 1D of a camera lens. The use of a slit as a linear source allows to work with sources that would not be bright enough for use as a point source. Moreover, the adjustability of the slit width is educationally interesting as the students can directly see the impact of the size of the source on the LSF and subsequently on the measured MTF which they can visualize in real time thanks to a simple GUI. The aberrations of the camera lens can then be studied quantitatively in terms of the MTF. They can then identify that the objective is diffraction-limited from NN = 8.