BNC/TC-2095 RACK-MOUNT ADAPTER

Similar documents
VXI-TB CHANNEL ISOTHERMAL TERMINAL BLOCK

TBX CHANNEL ISOTHERMAL TERMINAL BLOCKINSTALLATION GUIDE

TBX-68S Isothermal Terminal Block

The SCB-68 is a shielded board with 68 screw terminals for easy connection to National Instruments 68-pin products.

TBX-1329 AC/DC COUPLING TERMINAL BLOCK

Analog Signal Conditioning Accessories

SCXI HIGH-ACCURACY ISOTHERMAL TERMINAL BLOCK

SCXI 8-Channel Isolated Analog Input Modules

SCXI TERMINAL BLOCK

Model LIA100. Lock-in Amplifier

DIN Thermocouple Conditioner MODEL 5M14(V)

SCC-ACC01 Accelerometer Input Module

DBK45. 4-Channel SSH and Low-Pass Filter Card. Overview

GE Fanuc IC695ALG312. Rx3i PacSystem

Isolated, Linearized Thermocouple Input 7B47 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated, Thermocouple Input 7B37 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated, Linearized Thermocouple Input 5B47 FEATURES APPLICATIONS PRODUCT OVERVIEW

TBX-1329 AC/DC COUPLING TERMINAL BLOCK

Isolated Linearized 4-Wire RTD Input 5B35 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

VT1586A Rack Mount Terminal Panel Installation and User s Manual

This Errata Sheet contains corrections or changes made after the publication of this manual.

Isolated, Frequency Input 5B45 / 5B46 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Signal Conditioning Fundamentals for PC-Based Data Acquisition Systems

Improved Second Source to the EL2020 ADEL2020

PACSystems* RX3i IC695ALG600

BNC-2090 User Manual. Rack-Mount BNC Accessory. March 1996 Edition Part Number A-01

Isolated High Level Voltage Output 7B22 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

PACSystems* RX3i IC695ALG600-DD

DBK17. 4-Channel Simultaneous Sample and Hold Card. Overview

ADC-20/ADC-24 Terminal Board

PACSystems* RX3i IC695ALG600-EE Universal Analog Input Module

USB-TEMP and TC Series USB-Based Temperature Measurement Devices

Contents NUDRIVE ACCESSORY USER GUIDE

SCC-FV01 Frequency Input Module

SCXI HIGH-VOLTAGE ATTENUATOR TERMINAL BLOCK

ni.com Sensor Measurement Fundamentals Series

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9213E 16-Channel Thermocouple Input Module

AMP-13 OPERATOR S MANUAL

Quantum I/O Modules. At a Glance. The following section provides information on the Quantum Input/Output (I/O) modules. What s in this Chapter?

SCC-CO20 Isolated Current Output Module

12/4/ X3 Bridge Amplifier. Resistive bridge amplifier with integrated excitation and power conditioning. Logos Electromechanical

Model SR554 Transformer Preamplifier

Isolated, Linearized RTD Input 7B34 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

GE Fanuc IC695ALG600. Rx3i PacSystem

SCXI. SCXI-1120/D User Manual

User s Manual for Integrator Short Pulse ISP16 10JUN2016

F4-08RTD 8-Channel RTD Input

Isolation (ISO), Special Overvoltage functions protection (OP) Voltage measurement PAD-V8-P 8 Voltage, up to ±50 V BW: 6 Hz ISO: 350 V DC

2750 Integra Mainframes

Product Description. SIGnal Workbench. Programmable Signal Conditioning System

Model 7000 Low Noise Differential Preamplifier

INSTRUCTION SHEET. Eaton Logic Controller ELCM TC Temperature Modules. [Applicable Temperature modules] ELCM-TC04ANNN MTN1

Signal Conditioning Amplifier

Tel: (949) Fax: (949) IAA100 Product Manual

SCXI CURRENT INPUT,HIGH-VOLTAGE TERMINAL BLOCK

SEM104 SERIES. SEM104P Pt100 Temperature Transmitter. SEM104TC Thermocouple Temperature Transmitter INDEX SECTION CONTENTS PAGE NO.

Ethernet-Based Temperature, Voltage and Strain Measurement Modules

Modicon Momentum automation platform 0

IC Preamplifier Challenges Choppers on Drift

MINI 048 / JUN 2003 PAGE 1. Temperature I/O Module HE500OCS048 / HE500OCS078 HE500RCS078

DAQMeter 4350 User Manual

OP5000 Signal Conditioning & I/O Products for RT-LAB Engineering Simulators OP5511 High-current and high-voltage input conditioning module

6.6 Analog input module SM 331; AI 8 x 13 Bit; (6ES7331-1KF02-0AB0)

Film Scanner 24V Inverter Replacement

Universal Alarm Indicator

BNC-208X Series User Manual

SC-2040 User Manual. Eight-Channel Simultaneous Sample-and-Hold Accessory. September 1994 Edition. Part Number A-01

High Performance Current Transducer IT 200-S ULTRASTAB = A. ε L

CALYS 150. Advanced documenting multifunction calibrator thermometer

USB-TEMP. Multi-sensor Temperature Measurement. User's Guide. October Rev 14A Measurement Computing Corporation

Jaguar Motor Controller (Stellaris Brushed DC Motor Control Module with CAN)

OPERATING INSTRUCTIONS NI Channel, 24-Bit, Universal Analog Input Module

Thermocouple Conditioner and Setpoint Controller AD596*/AD597*

Programming restrictions when operating SM 331; AI 8 x RTD with PROFIBUS masters which only support DPV0.

User's Manual: Series 350T AC Current Input (External Sensor), DC-Powered Transmitters

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Exercise 2: Temperature Measurement

Seismometer Preamplifier Manual

SDS TAP Retrofit Install Bartlett 3 Key Installation of TAP Control for Models Originally Built with the Bartlett 3 Key Control

Brushed DC Motor Control. Module with CAN (MDL-BDC24)

SCXI Terminal Block

SCM5B48 ACCELEROMETER INPUT MODULE USER S MANUAL

Uni-Mux XQL Multi-Channel Data Acquisition Module

SDI SPECTRADYNAMICS, INC. HIGH PERFORMANCE DISTRIBUTION AMPLIFIER OPERATING MANUAL

MICROTC LINEAR THERMOCOUPLE AMPLIFIER OPERATOR S MANUAL

PRECISION CURRENT TRANSDUCERS. DC Current Transducers CT-100 CT-150. User s Manual. All Rights Reserved CAEN ELS d.o.o. Rev. 1.

VersaPoint I/O Module

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

BAPI-Stat 4 X-Combo, Room %RH and Temp. Sensor BA/BS4XC Installation & Operating Instructions

ALM473 DUAL MONO \ STEREO AUDIO LEVEL MASTER OPERATION MANUAL IB

Octal Sample-and-Hold with Multiplexed Input SMP18

AD596/AD597 SPECIFICATIONS +60 C and V S = 10 V, Type J (AD596), Type K (AD597) Thermocouple,

NI 4350/4351 User Manual

LF442 Dual Low Power JFET Input Operational Amplifier

Correlation of Voltage and Temperature Measurement

USER'S MANUAL DMR-6700

Servicing North America: For immediate technical or application assistance: Servicing Europe:

High Precision 10 V IC Reference AD581

AMP-12 OPERATOR S MANUAL

Transcription:

ISTALLATI GUIDE BC/TC-095 RACK-MUT ADAPTER Introduction This installation guide describes how to install the BC-095 and the TC-095 rack-mount adapters and use them with 3-channel SCXI modules and VXI-SC-0 submodules. To set up and get started using your adapter, you need to read through the Signal Connections section only. Read this rest of this guide only for special configuration instructions. The BC-095 adapter consists of a shielded box with 3 BC connectors to connect your field signals to the input of 3-channel SCXI modules or VXI-SC submodules. The 96-position connector on this adapter connects to the front connector of a 3-channel SCXI module or a VXI-SC submodule using an SH96-96 shielded cable or an R96-96 ribbon cable. This adapter is designed specifically for ease of use with field signals using BC cables. It features rack-mount construction for use with 9-in. racks and a shielded, metal enclosure to minimize noise. The TC-095 adapter consists of a shielded box with 3 miniconnectors to connect your thermocouples to the inputs of 3-channel SCXI modules or VXI-SC submodules. The 96-position connector on this adapter connects to the front connector of an 3-channel SCXI module or a VXI-SC submodule using an SH96-96 shielded cable or an R96-96 ribbon cable. This adapter is designed specifically for ease of use with thermocouples. It features rack-mount construction for use with 9-in. racks, a shielded metal enclosure to minimize noise, isothermal construction to minimize the temperature gradients across the thermocouple junctions, and a high-accuracy thermistor for cold-junction temperature sensing. Both the BC-095 and the TC-095 have pull-up resistors connected between CH+ and +5 V and ground-referencing resistors connected between CH and chassis ground via switches located on the rear of the adapter. These resistors help you detect open thermocouples by detecting module amplifier output saturation and to help you ground reference floating signals. SCXI is a trademark of ational Instruments Corporation. Product and company names are trademarks or trade names of their respective companies. 373A-0 Copyright 997 ational Instruments Corp. All rights reserved. September 997

What You eed To Get Started You need the following items to set up and use your BC-095 or TC-095: ne of the following: BC-095 3-Channel, Rack-Mount Adapter with BC Connectors TC-095 3-Channel, Rack-Mount Adapter with Thermocouple Miniconnectors BC/TC-095 Installation Guide ne package of four 0 MΩ resistor networks (included) /8 in. flathead screwdriver (included) An SCXI chassis ne of the following SCXI 3-channel modules or VXI-SC submodules: SCXI-00 SCXI-0/B/C VXI-SCXI-0/B/C ne of the following cable assemblies: SH96-96 cable assembly (recommended) R96-96 cable assembly o. Phillips-head screwdriver BC/TC-095 Rack-Mount Adapter ational Instruments Corporation

Signal Connections Connecting Field Signals This section describes how to connect field signals to your adapter and how to connect the adapter to your SCXI module and VXI-SC submodule. BC-095 The BC-095 has 3 BC connectors, one for each channel of a 3-channel SCXI module or VXI-SC submodule. Each BC connector is labeled with the corresponding channel number. The center pin of the BC connector is the CH+, and the outer shield or metal is the CH of each channel. The CH is isolated from the chassis ground so you can use it for differential signals. Connect the BC connector of your cable to the BC connector of the BC-095 by pushing the connectors together and turning the cable connector clockwise. ote Do not force the BC connector on the BC-095. If you have difficulty making your connection, remove the connector completely and try again. TC-095 The TC-095 has 3 universal thermocouple miniconnectors that you can connect your thermocouple to each of the channels on an SCXI 3-channel module and VXI-SC submodule. Each connector is labeled with the corresponding channel number. The positive and negative terminals are indicated on the miniconnector with raised lettering. Connect the thermocouple wire to the TC-095 by pushing the connectors together. Each miniconnector consists of two spades of different widths, which polarize the miniconnector. ote Do not force the miniconnector; if you have difficulty inserting the miniconnector, check to see if the polarity is correct. ational Instruments Corporation 3 BC/TC-095 Rack-Mount Adapter

Configuring pen-thermocouple/signal Detection and Signal Ground-Referencing Use Table to configure your BC/TC-095 for operating with open-thermocouple or signal detection and for ground-referencing your signals properly. Table. Configuring pen-thermocouple Detection and Signal Ground-Referencing Switch Settings Signal Type pen Thermocouple Comments CH 3 0 FF GD REF GD REF GD REF GD REF Floating signal or thermocouple Yes Use this setting on the BC-095 for detecting open signal leads if the measured signal is <5 V by pulling the positive input to 5 V and when your signal is floating. Use this setting on the TC-095 to detect open thermocouples and when your thermocouple is floating. CH FF GD REF 3 GD REF GD REF GD REF 0 Ground-referenced signal or thermocouple o Use this setting with either the BC-095 or TC-095 when your signal is ground-referenced and you want to disable open-thermocouple detection. BC/TC-095 Rack-Mount Adapter 4 ational Instruments Corporation

Table. Configuring pen-thermocouple Detection and Signal Ground-Referencing (Continued) Switch Settings Signal Type pen Thermocouple Comments CH 3 0 FF GD REF GD REF GD REF GD REF Floating thermocouple or signal o Use this setting with either the BC-095 or TC-095 when your signal or thermocouple is floating and you want to disable open-thermocouple detection. CH FF GD REF 3 GD REF GD REF GD REF 0 Ground-referenced thermocouple or signal Yes Use this setting on the TC-095 to detect open thermocouples for ground-referenced thermocouples. Use this setting on the BC-095 for detecting open signal leads if the measured signal is <5 V by pulling the positive input to 5 V and when your signal is ground-referenced. Warning Connecting an external ground-referenced signal with the 0 Ω resistor network in place can cause permanent damage to the resistor network and the traces on the BC/TC-095 printed circuit board. ational Instruments is T liable for any damage or injuries resulting from improper signal connections. Refer to the Configuring the Resistor etworks section for more information. Connecting the BC/TC-095 to Your SCXI Module and to Your VXI-SC Submodule You can mount the BC/TC-095 onto a 9-in. rack or place it on a workbench near the SCXI or VXI chassis. If you do not rack-mount the BC/TC-095, attach the four adhesive rubber feet included in your kit to the bottom of the adapter to help keep it stationary. ote To minimize the temperature gradient inside the TC-095 and thus maintain its isothermal nature for accurate cold-junction compensation, place the adapter away from extreme temperature differentials. ational Instruments Corporation 5 BC/TC-095 Rack-Mount Adapter

! Caution Refer to Figure as you perform the following steps to connect the BC/TC-095 to the SCXI module or to the VXI-SC submodule.. Connect one end of the 96-pin cable to the BC/TC-095 96-pin connector and screw down the backshell.. Connect the other end of the cable to the SCXI module or the VXI-SC submodule in the same manner as above. The cable connectors of the cables are polarized. Do not force the cable. If you encounter difficulty, check to see if the polarity is correct. 3 4 5 BC/TC-095 96-Pin Backshell 3 SCXI Chassis 4 VXI-SC Submodule 5 VXI Chassis Figure. BC/TC-095 Connected to SCXI Module or VXI Chassis BC/TC-095 Rack-Mount Adapter 6 ational Instruments Corporation

Your BC/TC-095 is ready to use. Read the Configuration section only if you need to learn more about ground-referencing or resistor network configuration with the SCXI-0 module or the VXI-SC-0 submodule or for further configuration of your adapter shield and cold-junction sensor. Configuration This section describes how to open your BC/TC-095 to access your internal jumpers and resistor networks for changing the factory default settings. Refer to Figure as you perform the following steps to access jumpers W and W and your resistor networks.. Using the o. Phillips-head screwdriver, remove the nine screws that secure the adapter cover.. Pull the cover up and off. 3. Set jumpers W and W as needed. For more information about setting your jumpers, refer to Tables and 3. 4. Reassemble in reverse order. ational Instruments Corporation 7 BC/TC-095 Rack-Mount Adapter

7 8 9 0 3 4 5 6 RP8 RP4 3 RP7 4 RP3 5 W 6 W 7 RP6 8 RP 9 RP5 0 RP Cover Screws (9) Figure. Assembly Figure (BC-095) BC/TC-095 Rack-Mount Adapter 8 ational Instruments Corporation

Shield Ground Configuration Jumper W, located inside the BC/TC-095, connects the metal case and shield to the analog ground of the SCXI module directly or through a 00Ω resistor. r, you can disconnect jumper W. The factory-default setting is through the 00 Ω resistor. Table. BC/TC-095 Shield Ground Jumper Configuration Jumper W Configuration Description When to Use a Setting 4 3 W 0 Ω 00 Ω PE 00 Ω (factory default) This setting connects the BC/TC-095 housing to SCXI or VXI chassis ground through a 00 Ω resistor, while keeping any potential ground loop current very small. Use this setting when you are using the SH96-96 cable, and one of the following is true: You have not rack-mounted the BC/TC-095. You have rack-mounted the BC/TC-095 but are unsure whether the rack and the BC/TC-095 are grounded. 4 3 0 Ω 00 Ω PE 0 Ω This setting connects the BC/TC-095 housing to SCXI or VXI chassis ground. Use this setting only if the factory-default setting is inadequate for noise rejection, when you are using the R96-96 cable, and one of the following is true: W You have not rack-mounted the BC/TC-095. You have rack-mounted the BC/TC-095 and are sure that the rack and the BC/TC-095 are not grounded. 4 3 W 0 Ω 00 Ω PE pen This setting disconnects the BC/TC-095 metal housing from SCXI or VXI chassis ground. Use this setting only if the factory-default setting was inadequate for noise rejection, when you are using either the SH96-96 cable or the R96-96 cable, you have rack-mounted the BC/TC-095, and you are sure that the rack and the BC/TC-095 are grounded. ational Instruments Corporation 9 BC/TC-095 Rack-Mount Adapter

Cold-Junction Sensor Configuration (TC-095 nly) The TC-095 has a thermistor temperature sensor for cold-junction compensation to enable you to use thermocouples with your SCXI module or VXI-SC submodule. You can connect the cold-junction sensor to an SCXI module or VXI-SC submodule in either of two ways: Multiplexed temperature sensor (MTEMP) mode Set jumper W to the MTEMP position. This is the factory default setting and the mode you need to use in most cases. Refer to Table to learn more about jumper W settings for the TC-095. Direct temperature sensor (DTEMP) mode Set the jumper W to the DTEMP position. This mode connects the temperature sensor to a separate DAQ channel through your SCXI module. Refer to your SCXI module documentation to configure your SCXI module for DTEMP mode. Table shows the terminal block jumper settings for the TC-095. Table 3. Jumper W Settings for the TC-095 Jumper W Position DTEMP W Description MTEMP mode factory default setting; preferred mode. MTEMP DTEMP W DTEMP mode connect to a separate DAQ channel on some SCXI modules. MTEMP ote n the SCXI-0 module and the VXI-SC-0 submodule, the MTEMP and DTEMP modes are equivalent. BC/TC-095 Rack-Mount Adapter 0 ational Instruments Corporation

Configuring the Resistor etworks If you are an SCXI-0 or a VXI-SC-0 user, you can use a package of 0 MΩ resistor networks included in the BC/TC-095 kit. These resistor networks can be installed as RP, RP, RP3, and RP4. See Figure 3 for the location of RP through RP8. If you use this configuration and you have the ground-referencing switch turned on, the thermocouples or signals with low impedance source (less than 00 Ω) can be either ground-referenced or floating without affecting your measurement. Figure 3 shows how the pull-up and ground-referencing resistors are connected to the CH± inputs of the 3-channel SCXI module or VXI-SC submodule. 5 V CH+ SCXI Module or VXI-SC Submodule CH- Pull-Up Switch CH+ Screw Terminals CH- R pullup (RP5, RP6, RP7, RP8) Ground Referencing Switch R ground referencing (RP, RP, RP3, RP4) (in sockets) Figure 3. Resistor Connections Table 3 shows which resistor networks to use for your SCXI module, signal type, and application. Channel Table 4. Channel Input Signals and Resistor etwork Pull-up Resistor etwork Ground-Referencing Resistor etwork 0 7 RP8 RP4 8 5 RP7 RP3 6 3 RP6 RP 4 3 RP5 RP ational Instruments Corporation BC/TC-095 Rack-Mount Adapter

Using Resistor etworks with SCXI Modules and VXI-SC Submodules This section describes how to use resistor networks with the SCXI-00 and SCXI-0 modules, and the VXI-SC-0 submodule. SCXI-0 Module and VXI-SC-0 Submodule You can replace the 0 Ω ground-referencing resistor networks (factory-default configuration) in the BC/TC-095 with the 0 MΩ resistor networks supplied in the kit. With the 0 MΩ resistor networks, it does not matter whether your signal is ground-referenced or floating. The channels with open thermocouples saturate at all sample rates. Use long-nose pliers to remove or replace the resistor networks in the sockets; be careful not to damage the network package. Make sure pin of each network is in the correct socket (see Figures 4 and 5). Each resistor network is labeled with descriptive numbers on the left front side, and pin is located directly beneath the darkened symbol within these numbers. The 0 ΜΩ resistor network is labeled 06 (0 * 0 6 Ω). Figures 4 and 5 show examples of these resistors. 0x--00 Mfr. code 0x--06 Mfr. code Pin Pin a. 0 Ω Resistor etwork b. 0 MΩ Resistor etwork Figure 4. Resistor etworks RP Pin Figure 5. Pin Location on BC/TC-095 for Ground-Referencing Resistor Socket SCXI-00 Module Your SCXI-00 module must be set at a gain of 00 or higher if you want the open thermocouple to saturate when the open-thermocouple detection is switched. If you do not want the measurements on any other channel to be disturbed, you must use an interchannel delay of 00 µs, which corresponds to a sample rate of 5 khz. BC/TC-095 Rack-Mount Adapter ational Instruments Corporation

Faster sampling rates will still result in accurate measurements; however, if a channel has an open thermocouple, this channel might not saturate and if it does, it affects the measurement accuracy of other channels on the same module. Errors Due to pen-thermocouple Detection Circuitry pen-thermocouple detection circuitry can cause two types of measurement errors. These errors are the results of common-mode voltage at the input of the SCXI module or VXI-SC submodule and current leakage into your signal leads. Common-Mode Voltage at the Input of the SCXI Module or VXI-SC Submodule With 0 MΩ pull-up and ground-referencing resistors, a common-mode voltage of.5 VDC develops if the thermocouple is floating. At a gain of 00, the common-mode rejection of the SCXI-0 module or VXI-SC submodule is sufficiently high that the resulting offset error is negligible. If your application demands extremely high accuracy, you can eliminate this offset error by calibrating your system with both the pull-up and ground-referencing resistor switches. You also can turn off the pull-up resistor switch, which eliminates the open-thermocouple detection feature, or use the 0 Ω ground-referencing resistor networks, which brings the common-mode voltage down to nearly 0 VDC. Current Leakage The open-thermocouple detection circuitry results in a small current leakage into the thermocouple. With the 0 MΩ ground-referencing and pull-up resistor networks, the current leakage results in a negligible error. With the 0 Ω ground-referencing resistor, the 0 MΩ pull-up resistor connected to 5 VDC causes a current leakage of approximately 0.5 µa (5 V/0 MΩ) to flow into the unbroken thermocouple. If the thermocouple is very long, a voltage drop develops in the thermocouple because of lead resistance. For example, if you have a 4 AWG J-type thermocouple that is 0 ft long, a voltage drop of approximately 8.78 µv (0.878 Ω/double ft 0 double ft 0.5 µa) can develop in the thermocouple, which corresponds to an error of 0.8 C. If your application demands very high accuracy, you can eliminate this error by turning off the appropriate pull-up resistor switch or by calibrating the system offset. See Figure 3 for the location of these switches. ational Instruments Corporation 3 BC/TC-095 Rack-Mount Adapter

Temperature Sensor utput and Accuracy The TC-095 temperature sensor voltage output varies from.9to 0.58 V over the 0 to 55 C temperature range. Table 5 shows the temperature sensor output accuracy. To select and read the temperature sensor, refer to your data acquisition software documentation for programming information. Alternatively, you can use the following formula to convert the cold-junction sensor voltage to cold-junction temperature: T( C) = T K 73.5 where T K is the temperature in kelvin a =.9536 x 0 3 b =.34359 x 0 4 c =.08703 x 0 7 Table 5. Temperature Sensor Voltage utput Accuracy Temperature Range Voltage utput Accuracy * 0 to > 5 C ±.0 C 5 to 35 C ±0.65 C < 35 to 55 C ±.0 C * Includes the combined effects of the temperature sensor accuracy and the temperature difference between the temperature sensor and any screw terminal. The terminal sensor accuracy includes tolerances in all component values, the effects caused by temperature and loading, and self-heating. T K = --------------------------------------------------------------- 3 a + b ( lnr T ) + c ( lnr T ) R T = resistance of the thermistor in Ω V TEMPUT R T = 5000 ---------------------------------------.5 V TEMPUT V TEMPUT = output voltage of the temperature sensor T F ( ) [ T C ( ) ]9 = ---------------------- + 3 5 BC/TC-095 Rack-Mount Adapter 4 ational Instruments Corporation

where T( F) and T( C) are the temperature readings in degrees Fahrenheit and Celsius, respectively. ote V TEMPUT varies from.9 V (at 0 C) to 0.58 V (at 55 C). For best resolution, use the maximum gain for this signal range on the analog input channel of your DAQ device. The SCXI-0 and the VXI-SC-0 have a Hz filter on the V TEMPUT signal. The SCXI-00 does not have a filter on the V TEMPUT signal. Therefore, use an average of a large number of samples to obtain the most accurate measurement. oisy environments require more samples for greater accuracy. Temperature Sensor Circuit Diagram The circuit diagram in Figure 6 shows details about the BC/TC-095 temperature sensor. +5 V 4.7 kω %.5 V LM 4040.5 V 0.% 0. µf o -t 5 kω 0.% 5 kω + 0 µf at 5 o C 6 V 0. µf W MTEMP DTEMP Figure 6. Temperature Sensor Circuit Diagram ational Instruments Corporation 5 BC/TC-095 Rack-Mount Adapter

Specifications Cold-junction sensor (TC-095 nly) Accuracy...0.65 from 5 to 35 C.0 from 0 to 5 and 35 to 55 C Repeatability...0.35 from 5 to 35 C utput...9 VDC (at 0 C) to 0.58 VDC (at 55 C) pen-thermocouple detection Pull-up resistor...0 MΩ Ground-referencing resistor...0 Ω default or 0 MΩ Maximum working voltage (signal + common mode)...each input should remain within ±0 V of chassis ground. Dimensions...47.6 by 4.4 cm (9 in. by.75 in.) Includes the combined effects of the temperature sensor accuracy and the temperature difference between the temperature sensor and any screw terminal. The temperature sensor accuracy includes tolerances in all component values, the effect caused by temperature and loading, and self-heating.