Laser interferometric measuring system for positioning in nanometrology

Similar documents
Contribution of the Refractive Index Fluctuations to the Length Noise in Displacement Interferometry

Precision displacement interferometry with stabilization of wavelength on air

Stability of a Fiber-Fed Heterodyne Interferometer

High frequency stability semiconductor laser sources at 760 nm wavelength

Absolute distance interferometer in LaserTracer geometry

1.6 Beam Wander vs. Image Jitter

Wavelength Control and Locking with Sub-MHz Precision

Improving a commercially available heterodyne laser interferometer to sub-nm uncertainty

ULTRASONIC TRANSDUCER PEAK-TO-PEAK OPTICAL MEASUREMENT

A Multiwavelength Interferometer for Geodetic Lengths

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing D2.2. Ger Folkersma (Demcon)

EVALUATION OF A FOCUSED LASER SPOT DIAMETER FOR AN OPTICAL ANGLE SENSOR. Yuki Shimizu, Taiji Maruyama, Shota Nakagawa, Yuan-Liu Chen, Wei Gao

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

DISPLACEMENT INTERFEROMETRY WIN PASSIVE FABRY-PEROT CAVITY

High stability multiplexed fibre interferometer and its application on absolute displacement measurement and on-line surface metrology

Investigation of an optical sensor for small angle detection

SUPPLEMENTARY INFORMATION DOI: /NPHOTON

Agilent 5527A/B-2 Achieving Maximum Accuracy and Repeatability

Keysight Technologies Using a Wide-band Tunable Laser for Optical Filter Measurements

Length section: New calibration and research services

Supplementary Figures

Laser Telemetric System (Metrology)

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components

UNMATCHED OUTPUT POWER AND TUNING RANGE

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

3 General Principles of Operation of the S7500 Laser

Swept Wavelength Testing:

Comparison of resolution specifications for micro- and nanometer measurement techniques

Optical generation of frequency stable mm-wave radiation using diode laser pumped Nd:YAG lasers

Agilent 10717A Wavelength Tracker

(51) Int Cl.: G01B 9/02 ( ) G01B 11/24 ( ) G01N 21/47 ( )

A novel tunable diode laser using volume holographic gratings

Optics and Laser Heads for Laser-Interferometer Positioning Systems Product Overview

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

R. J. Jones Optical Sciences OPTI 511L Fall 2017

Keysight Technologies Optics and Laser Heads for Laser-Interferometer Positioning Systems

7 WAVEMETER PROJECT #6 MODEL OEK-100. Measure the Wavelength of An Unknown laser Using 633nm and 543 nm HeNe lasers

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Spectrally resolved frequency comb interferometry for long distance measurement

Jones matrix analysis of high-precision displacement measuring interferometers

6 THICKNESS MEASUREMENT OF TRANSPARENT MEDIA

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

An Optical Characteristic Testing System for the Infrared Fiber in a Transmission Bandwidth 9-11μm

IST IP NOBEL "Next generation Optical network for Broadband European Leadership"

M-041 M-044 Tip/Tilt Stage

Fiber Optic Device Manufacturing

Difrotec Product & Services. Ultra high accuracy interferometry & custom optical solutions

Simple System for Active Frequency Stabilization of a Diode Laser in an External Cavity

Uncertainty in measurements of micro-patterned thin film thickness using Nanometrological AFM - Reliability of parameters for base straight line -

A transportable optical frequency comb based on a mode-locked fibre laser

SENSOR+TEST Conference SENSOR 2009 Proceedings II

OPTICS IN MOTION. Introduction: Competing Technologies: 1 of 6 3/18/2012 6:27 PM.

la. Smith and C.P. Burger Department of Mechanical Engineering Texas A&M University College Station Tx

Fiber Optic Sensing Applications Based on Optical Propagation Mode Time Delay Measurement

Doppler-Free Spetroscopy of Rubidium

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

The Henryk Niewodniczański INSTITUTE OF NUCLEAR PHYSICS Polish Academy of Sciences ul. Radzikowskiego 152, Kraków, Poland.

Physics 476LW. Advanced Physics Laboratory - Microwave Optics

Mode analysis of Oxide-Confined VCSELs using near-far field approaches

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

Results from the Stanford 10 m Sagnac interferometer

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs

LOS 1 LASER OPTICS SET

High-power semiconductor lasers for applications requiring GHz linewidth source

OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY

Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI)

LOPUT Laser: A novel concept to realize single longitudinal mode laser

THE TUNABLE LASER LIGHT SOURCE C-WAVE. HÜBNER Photonics Coherence Matters.

A Fast Phase meter for Interferometric Applications with an Accuracy in the Picometer Regime

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

Application Note #548 AcuityXR Technology Significantly Enhances Lateral Resolution of White-Light Optical Profilers

Lecture 21. Wind Lidar (3) Direct Detection Doppler Lidar

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

It s Our Business to be EXACT

University of Huddersfield Repository

LISA and SMART2 Optical Work in Europe

2 CYCLICAL SHEARING INTERFEROMETER

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

Lecture 27. Wind Lidar (6) Edge Filter-Based Direct Detection Doppler Lidar

Absolute distance measurement with an unraveled femtosecond frequency comb Steven van den Berg

REAL TIME THICKNESS MEASUREMENT OF A MOVING WIRE

DIODE LASER SPECTROSCOPY (160309)

Installation and Characterization of the Advanced LIGO 200 Watt PSL

Investigation of some critical aspects of on-line surface. measurement by a wavelength-division-multiplexing. technique

Development of a Low Cost 3x3 Coupler. Mach-Zehnder Interferometric Optical Fibre Vibration. Sensor

Fabrication of large grating by monitoring the latent fringe pattern

DESIGN OF COMPACT PULSED 4 MIRROR LASER WIRE SYSTEM FOR QUICK MEASUREMENT OF ELECTRON BEAM PROFILE

z t h l g 2009 John Wiley & Sons, Inc. Published 2009 by John Wiley & Sons, Inc.

A Narrow-Band Tunable Diode Laser System with Grating Feedback

An Auto-Locked Diode Laser System for Precision Metrology

OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY

Plane Mirror Interferometer Configurations. Functional description. Interferometeranordnung Plane Mirror Interferometer

Kit for building your own THz Time-Domain Spectrometer

Ph 77 ADVANCED PHYSICS LABORATORY ATOMIC AND OPTICAL PHYSICS

Optical phase-locked loop for coherent transmission over 500 km using heterodyne detection with fiber lasers

Evaluation of Scientific Solutions Liquid Crystal Fabry-Perot Etalon

Nano Beam Position Monitor

Fast Widely-Tunable CW Single Frequency 2-micron Laser

FREQUENCY COMPARISON AT 633 NM WAVELENGTH: DETERMINATION OF DIAGONAL ELEMENTS OF MATRIX MEASUREMENTS BY USING A MASTER-SLAVE He-Ne LASER SYSTEM

Transcription:

Laser interferometric measuring system for positioning in nanometrology JOSEF LAZAR, ONDŘEJ ČÍP, ARTIN ČÍŽEK, JAN HRABINA, AND OJÍR ŠERÝ Department of Coherence Optics Institute of Scientific Instruments, Academy of Sciences of the Czech Republic Královopolská 147, 612 64 Brno CZECH RAPUBLIC PETR KLAPETEK Department of Nanometrology Czech etrology Institute Okružní 31, 638 00 Brno CZECH RAPUBLIC joe@isibrno.cz http://www.isibrno.cz Abstract: - In this contribution we present a development of a system for dimensional nanometrology based on scanning probe microscopy techniques (primarily atomic force microscopy, AF) for detection of sample profile combined with interferometer controlled positioning. The key goal for introduction of interferometer measurement is not only improvement of resolution but the direct traceability to the primary etalon of length. Interferometry compared to a host of other optical length measuring techniques [1,2,3...] represents the most precise measuring technique available. The system is being developed to operate at and in cooperation with the Czech metrology institute for calibration purposes and nanometrology. Keywords- Nanometrology, Interferometry, Traceability, Local probe microscopy, Nanopositioning. 1. Introduction Dimensional metrology dealing with objects in the micro- and nanoworld relies predominantly on AF (Atomic Force icroscope) and related local probe microscopy techniques where the object topology, dimensions and other properties are examined by scanning the sample. Positioning of the probe in AF microscopes through piezoelectric (PZT) transducers offers sub-nm resolution but small range over several tens of micrometers. ore, PZT transducers suffer errors of repeatability, non-linearity and hysteresis. etrological AF can be calibrated using etalon gratings or samples of height staircase type in the scale of hundreds or thousands of nanometers. The problem of traceability [4] is a complex one where a system independent on an etalon samples should be linked directly to the primary etalon of length. This means employment of laser interferometry techniques for measurement of the probe position and taking care for all other sources of error starting with the uncertainty of the laser optical frequency [5]. Systems designed to follow this demand represent a setup mostly consisting of an AF head, positioning stage and displacement measuring arrangement where a multiaxis laser interferometers dominate [6,7,8] but other approaches based on optical methods may represent suitable solution [9]. easurement of the sample position in three orthogonal axes is sufficient when the guides of the stage can ensure negligible angle errors. Full control of the stage position should engage evaluation of tilting of the stage where non-contact optical methods a preferred [10]. A more complex interferometric measuring system needs also a complex approach to all sources of errors caused by angle deviations from orthogonality of the measuring beams, angle errors of reflecting surfaces, etc. [11]. Interferometric measuring techniques in dimensional metrology are well established and represent a link between the fundamental etalon of length and mechanical measuring systems. Significant effort has been invested into improvement of their performance in the nanoscale through linearization of the fringe interpolation [12]. Transforming the local probe microscope from an imaging tool into a measuring system for metrological purposes means replacement of the often small scale positioning of the probe with an external stage moving ISSN: 1109-2734 660 Issue 10, Volume 9, October 2010

the sample and interferometric monitoring or even control. Laser interferometry seems to be a solution not only due to its direct traceability to the fundamental etalon of length but also the incremental interferometer in its fringe counting mode gives an excellent dynamic range limited only by the fluctuations of the refractive index of air and offers nanometer or even subnanometer resolution over large range. We concentrated onto a small range flexture three-axis nanopositioning stage equipped with closedloop motion control with capacitive sensors embedded in a frame with six-axes interferometric system supplied from a stabilized single-frequency frequency doubled 532 nm Nd:YAG laser. 2. Stage Design In the design presented here we concentrated on a commercial positioning stage with 200 x 200 x 10 μm travel and its enclosure into a frame containing interferometric displacement monitoring system. Full control of the stage and evaluation of all its positioning errors needs six-axis measurement. In our arrangement we equipped the stage with a top plate a sample holder and a set of flat mirrors. It overlaps the stage and makes possible the measurement of the vertical displacement around the stage by three interferometers pointing upwards. Together with the mount of the local probe microscope the side view is in Figure 1. Thus the vertical position in the z-axis together with pitch and roll angles can be evaluated. F B C H D Figure 1: Side view of the stage with vertical interferometers, B: baseplate, I: microscope, S: sample, F: fiber light delivery, C: collimator, H: sample holder, T: stage, I: interferometer, D: homodyne detection unit, : mirror Horizontal measurement of x- and y- axes is ensured by three interferometers which allow also evaluation of the yaw. Interferometers are with flat-mirror reflector and I I T S a fixed corner-cube reflector in both reference arm and measuring arm. Double-pass arrangement enhances resolution in simple fringe-counting regime to λ/4 (Figure 2). The resolution of the interferometric detection and data processing system here is 10 bit with 1 LSB being the 1/1024 of one cycle of the interferometric signal. Together with the double beam pass it results in λ/4096 which means for the 532 nm wavelength resolution 130 pm B I D S S D I H S S Figure 2: Top view of the interferometric configuration with measurement in the x- and y-axes, S: beam steering mirror, others see Figure 1. Compensation for the fluctuations of the refractive index in interferometric systems is traditionally ensured through the evaluation of the Edlen formula and under laboratory conditions results in relative uncertainty between 10-6 and 10-7. Here when due to small dead length (0.1 mm) and travel range the maximum length of the measuring arm is 0.3 mm. The influence of the refractive index of air may prove significant only at the 0.3 nm level. The small range of positioning ranging within 200 μm in the horizontal plane and only 10 μm in the vertical axis enhances the importance of the linearity of the scale. Linearity of the fringe division is further improved by software linearization algorithm embedded directly into the signal processing of the interferometer signal [13,14]. With the shorter wavelength of the green laser (compared with the traditional 633 nm red He-Ne laser of commercial interferometers) another small resolution improvement was achieved. D I C F ISSN: 1109-2734 661 Issue 10, Volume 9, October 2010

tip axis y a x - z PZTs y b x z c z a sample board z b Figure 3: Arrangement of the additional three PZT transducers compensating for the angle errors. x, y a, y b, z a, z b, z c : orientation of interferometers. The chance to evaluate the angle deviations offers a good opportunity to compensate for them. Angle deviations of the reflective mirrors in a plane-mirror interferometric design results in angle-induced errors due to difference between the returning beam path and the axis of motion. We introduced a set of three piezoeletric transducers (PZT) each with two directions of motion: vertical and sheering horizontal (Figure 3). Their arrangement around the centre of the table with the horizontal axes of motion being oriented tangentially gives a chance to control all the angle errors. In the first pilot design we looked for the level of angle deviations and decided for PZT transducers with 3 μm travel in the vertical axis and 5 μm in the horizontal. The horizontal motion of all three transducers oriented tangentially rotates the stage with respect to the vertiocal axis. 3. Laser Source Traditional dimensional interferometry has been a domain of stabilized He-Ne lasers operating at the 633 nm wavelength. Their single- or two-frequency operation, narrow linewidth, modest output power and relative simplicity were among their advantages. Stabilization of optical frequency was based here on various detection techniques of the Doppler-broadened spectral profile of the active line in Ne. With the fluctuations of the refractive index of air influencing the interferometric measurements in the 10-6, in the best case in the 10-7 order even with precise compensation through the Edlen formula the stability achieved proved to be sufficient. Output power of a single-frequency He-Ne laser in the range of 1 mw is enough to supply an interferometric system with a single-axis measurement. Splitting of the laser beam into several discrete interferometers usually needs to rise gain in the detection chain and thus to worsen the noise performance of the system. In our case we intended to assemble a displacement measurement arrangement which would monitor the nano-positioning table in all six axes of freedom. Feeding six discrete interferometers would be a difficult task for a low-power He-Ne laser especially when it seems suitable to apply a fiber optic light delivery system. A free-space to fiber coupler and a splitter causes considerable losses. In the first design we arranged the laser and light delivery scheme with two fibre optic branches and a stabilization scheme with iodine cell according to fig. 4. D L I 2 C Figure 4: First setup of laser stabilization and twobranche fibre optic light delivery, L: laser, : mirrors, D: photodetector, F: polarization maintaining fiber, C: collimator, S: non-polarizing beamsplitter, I 2 : iodine cell. S F ISSN: 1109-2734 662 Issue 10, Volume 9, October 2010

Figure 5: echanical design of the interferometer frame for embedding of the nanopositioning stage. The stage itself is omitted here, the transparent upper board serves as a mirror holder and sample holder. We decided for a more powerful and proven frequency doubled Nd:YAG laser operating at the 532 nm wavelength, a laser with high-efficiency intracavity doubling designed for primary laser optical frequency etalons. Interferometric detection systems used in our designs are based on DC, homodyne detection principle and need a single-frequency laser operation. This made our choice easier while generation of two closely spaced stable optical frequencies would not be easy with a solidstate laser. Our Nd:YAG laser features low amplitude noise and output power on the 100 mw level. The short-term intrinsic optical frequency stability of the free running laser satisfies well the requirements of interferometry in the presence of air where the fluctuations of index of refraction are the main limiting factor. This is ensured by thermal stabilization of the active crystal and the laser cavity. The laser is stabilized through a linear spectroscopy in molecular iodine by locking to a Doppler broadened absorption of a transition which is in coincidence with the laser tuning range. This follows the recommendations of CIP [5] and ensures the traceability to fundamental etalon. Stabilization to the Doppler broadened spectral profile is able to deliver stabilities in the range of 10-8 which is here at least one order better than the influence of variations of the refractive index of air. 4. Light Delivery The first configuration presented here with twobranche fiber optic delivery is in Fig. 5. Interferometers are arranged in two levels with steering of the laser beams to discrete interferometric units via semireflecting mirrors within each level. On the right of the stage mount there are collimators for each level. Simplicity of the fiber optic assembly with cheap components (fiber optic splitter) needed a complex set of precisely adjustable mirrors. This proved to be a great obstacle to elimination of primary angle errors, cosine and Abbe errors in all axis of measurement. Independent and precision angle steering of each of the interferometric unit proved to be neccessary. The the first configuration the laser radiation was delivered to the interferometers via a single-mode polarization maintaining fiber and split into the measuring axes by a 50/50% non-polarizing splitter incorporated into the fiber-optic assembly. The two arms of the fiber-optic delivery supply three interferometers each. First are the three interferometers measuring in the z-axis are arranged in one plane and the other are the x and double y measuring interferometers in other plane. The key problem of a complex fiber optic assembly for independent light delivery proved to be the polarization marinating fiber. A set of tests was performed to find a suitable fiber and a splitter with ISSN: 1109-2734 663 Issue 10, Volume 9, October 2010

Figure 6: Laser source with stabilization of frequency through linear spectroscopy in iodine and light delivery arrangement. LASER: Frequency doubled Nd:YAG laser, C: fiber collimators, FS: fiber polarization maintaining splitter, D: photodetector, IC: iodine cell. acceptably low sensitivity to stress and thermal effects. We applied a polarization maintaining splitter with one input and 8 outputs where the light delivery to the iodine cell was via a fiber as well. The final 8 th output was reserved for a 7 th interferometer which is intended to monitor the vertical positioning of the AF tip. Configuration of the full independent fiber optic light delivery system is in Fig. 6. The arrangement of the modified frame with interferometers supplied from a discrete fiber each is in Fig. 7. The three vertical interferometers are hidden below the sample table and a board on the left represents the fiber splitter. Each output collimator is housed in a positioner with four degrees of freedom and fine adjusting screws. Figure 7: New mechanical design of the interferometer frame with independent fiber light delivery and 1/8 fiber optic polarization maintaining splitter. ISSN: 1109-2734 664 Issue 10, Volume 9, October 2010

5. Evaluation and Testing The positioning performance of the free-running stage (without feedback control of angle via additional PZTs) was tested through evaluation of the angle errors. The calculated angle deviations of the stage during fullrange travel are presented in Figure 8. The recordings show strong non-linearities and the maximum angle deviation approx. in proportion to the travel range. All positioning and recordings were performed in stable laboratory conditions and the wavelength of the laser source was stabilized to Doppler broadened absorption transition R(56) 32-0 and monitored during the whole process by a wavemeter with a 0.5 GHz resolution. A scanning regime of the stage allows linear motion at selected speed over the whole range which allows fast scannig of a larger sample without delays needed for mechanical settlement in the stepping mode. 7 1 6 5 0.8 0.6 rotation [ rad] 4 3 2 1 pitch, resp. roll [ rad] 0.4 0.2 0 0-0.2-1 -0.4-2 0 50 100 50 0 position x, resp. y [ m] -0.6 0 5 10 5 0 position z [ m] Fig. 8. Recordings of angle deviations. First diagram represents rotation of the stage in the horizontal plane (yaw) during diagonal motion in both x and y axes, the second pitch and roll during vertical motion along z axis. In the horizontal plane the angle deviation (yaw) was measured during diagonal motion incrementing both x and y position simultaneously with the vertical z position fixed to mid-travel in forward and reverse direction. Pitch and roll angle deviations are in a separate diagram related to the vertical position. The x and y value was again set to mid-range. The stage was tested again by recording of the difference between the value measured through capacitive sensors and interferometers. In Fig. 9 the deviations are expressed in a greyscale over the whole horizontal plane. The first image represents deviations of the capacitive sensor value from theoretical linear position of the stage that is related to time, the second represents similarly the deviations of the interferometer Figure 9: Two-dimensional expression of the differences between the value measured through capacitive sensors and interferometers. ISSN: 1109-2734 665 Issue 10, Volume 9, October 2010

value. The third shows their differences. The differences recording express significant periodicity (modulation) in the horizontal (x) axis which is also clearly visible in the recording of the capacitive sensor output. This periodicity is visible also in Fig. 9, left, bottom curve which indicates it is a systematic error of the capacitive sensor independent on the dynamic effects of the stage positioning. This shows how important improvement in resolution and precision the interferometric displacement monitoring can be. 6. Operation with AF microscope The operation of the positioning and measuring system was tested with an AF (Atomic Force icroscope) head mounted on a tripod upon the frame. A simple AF scanning head with optical feedback based on laser diode, set of mirrors and quadrant detector was assembled. For fast scanning a small range z-axis piezo was added to the cantilever holder. The signal from fourquadrant detector was processed digitally by single chip microcontroller in order to obtain feedback data used by both table and z-axis piezo in the head. The Z-axis piezo in the head was recalibrated using the scanning table and interferometer before each measurement as a part of sample feedback adjustment. being previously calibrated through the interferometric measured vertical positioning of the stage. 7. Etalon Calibration The primary intention of the whole system operation is at the department of nanometrology of the Czech etrology institute in Brno. The effort that is presented here was at this stage finalized with evaluation of the overall uncertainty by testing on a calibrated etalon of a grating which could be independently verified through diffraction techniques. The overall values are in this stage of development limited primarily by mechanics of the assembly but represent an overview of the performance. A two-dimensional grating with 2 μm steps was selected. The evaluation of overall uncertainty included verification of the laser wavelength by a calibrated wavemeter and comparison of the grating dimensions derived from measurement via the metrological AF and through laser diffraction technique. The image of a fraction of the grating is in Figure 11 together with a typical profile in the x-axis. Evaluation of the grating spacing was done by comparing of the position derived from built-in capacitive sensors and position measured through the interferometric frame. Figure 10: Left: image of a fragment of an etalon of two-dimensional grating; right: an example of a recorded profile achieved by vertical calibrated positioning of the AF tip. The whole assembly with AF microscope was tested by imaging and measuring of an etalon with twodimensional grating. The image of a fragment of the grating is in figure 10 together with a scanned profile over a small range. Positioning on the horizontal plane is performed with the nanopositioning table while in the vertical axis the PZT of the AF microscope is used Laser based diffraction technique allowed measurement of the overall value of the groove spacing with small uncertainty compared to statistical evaluation of average value from AF measurement. Results are summarized in table 1. ISSN: 1109-2734 666 Issue 10, Volume 9, October 2010

Figure 11: Image of the two-dimensional etalon grating used for calibration and uncertainty eavluation together with a cross-section of the profile x- direction y- direction capacitive sensors interferometer diffraction 3995 ± 6 nm 4001 ± 6 nm 3996.7 ± 1.8 nm 4002 ± 10 nm 4001 ± 10 nm 3994.2 ± 1.1 nm Table 1: Spacing of the measured grating with estimated statistical uncertainty. Statistical uncertainty associated with the AF measurement either through capacitive sensors or interferometers includes angle errors caused by nonlinear motion of the stage. Together with uncertainty of the coincidence of the measuring tip and measuring axes of the interferometers this introduces additional errors. Further improvement towards closed-loop operation derived from the interferometers will help significantly. The evaluation of uncertainty was performed with the first version of the system where the correction of the angle errors was not yet operational. It seems that it will represent further improvement. 8. Conclusion Interferometric system presented here represents a nanometrology tool still under development. Interferometric monitoring and further full feedback servo control of the positioning derived from the interferometer value represents significant improvement for calibration of grating-type etalons through local probe microscopy where the resulting image can be referenced to the interferometer measured position. First experiments showed that the six-axis interferometric monitoring with a wide base for independent angle evaluation can give information about angle errors with a resolution on the level of few tens of nanoradians. This seems to be an excellent value and a full control of the straightness of motion will be an answer to problems with alignment of the measuring axis and center of the AF tip. With a positioning of the stage without angle shifts the full traceability can be ensured and it is possible to suppress uncertainty associated with Abbe errors resulting from different motion path of the probe and the interferometer value due to angle tilt. Overall angle deviations over the whole range of motion did not exceed 10 μrad. Introduction of small- ISSN: 1109-2734 667 Issue 10, Volume 9, October 2010

range PZT transducers for real-time control (Fig. 8) and closed-loop operation is able to eliminate all the angle deviation errors with additional positioning on several μm level. Independent adjustment of all collimators feeding the laser light into interferometric units brought a significant improvement of performance where any cosine errors could be also eliminated. An overview of the whole setup is given in the photo in Fig. 12 where the stage and interferometers are visible with removed sample table. References: [1] Wang, T.-H., Lu,.-C., Hsu, C.-C., Lu, Y.Y., Image ruler and its applications in distance and area measurement, WSEAS Transactions on Systems, 2007, Vol. 6, No. 5, pp. 901-907. [2] èndez, J., Parsiani, H., Sanchez, E., Remote sensing of atmospheric particles using LIDAR, Calipso satellite, & AERONET: Algorithm development, WSEAS Transactions on Systems, 2009, Vol. 8, No. 1, pp. 86-95. Figure 12: Photo of the first arrangement of the interferometric positioning assembly for AF based nanometrology. 9. Acknowledgements The authors wish to express thanks for support to the grant projects from inistry of Education, Youth and Sports of CR, projects No.: LC06007, the AS CR, project No.: KAN311610701, inistry of Industry and Commerce, projects No: FR-TI1/241 and GA CR, project: GA102/09/1276. [3] Kumarajah, K., enon, P.S., Ismail,., Yeop, B.Y., Shaari, S., QW design parameter variation in a 1.5?m wavelength InP-based LW-VCSEL, WSEAS Transactions on Electronics, 2008, Vol. 5, No. 11, pp. 437-446. [4] Korpelainen V and Lassila A (2006) Calibration of a commercial AF: traceability for a coordinate system, eas. Sci. Technol., 18: 395-403. [5] Quinn T J (1992) ise en pratique of the definition of the metre, etrologia 1993/94, 30: 523-541. [6] Jäger G, Gruenwald R, anske E, Hausotte T and Fuessl R (2004) A nanopositioning and nanomeasuring ISSN: 1109-2734 668 Issue 10, Volume 9, October 2010

machine: operation measured results Nanotechnology and Precision Engineering 2: 81-4. [7] Weckenmann A, Hoffmann J and Schuler A (2008) Development of a tunnelling current sensor for a long-range nano-positioning device, eas. Sci. Technol. 19: 064002, 7pp. [8] Dai G, Pohlenz F, in Xu, Koenders L, Danzebrink H U and Wilkening G (2006) Accurate and traceable measurement of nano- and microstructures, eas. Sci. Technol., 17: 545 552. [9] Otsuka J, Ichikawa S, asuda T, et al. (2005) Development of a small ultraprecision positioning device with 5 nm resolution, eas. Sci. Technol., 16: 2186-2192. [10] Kim J W, Kang C S, Kim J A, et al. (2007) A compact system for simultaneous measurement of linear and angular displacements of nano-stages Optics Express, 15: 15759-15766. [11] Koops K R, van Veghel G A, Kotte G J W L and oolman C (2007) Calibration strategies for scanning probe metrology, eas. Sci. Technol., 18: 390 394. [12] Hou W and Wilkening G (1992) Investigation and compensation of the nonlinearity of heterodyne interferometers, Prec. Eng, 14: 91-98. [13] Číp O and Petrů F (2000) A scale-linearization method for precise laser interferometry. eas. Sci. Technol., 11: 133-141. [14] Petrů F and Číp O (1999) Problems regarding linearity of data of a laser interferometer with a single frequency laser. Precision Engineering, 23: 39-50. ISSN: 1109-2734 669 Issue 10, Volume 9, October 2010