Johnson Noise and the Boltzmann Constant

Similar documents
Thermal Johnson Noise Generated by a Resistor

Goals of the Lab: Photodetectors and Noise (Part 2) Department of Physics. Slide 1. PHYSICS6770 Laboratory 4

1 Lock-in Amplifier Introduction

USE OF BASIC ELECTRONIC MEASURING INSTRUMENTS Part II, & ANALYSIS OF MEASUREMENT ERROR 1

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

INTRODUCTION TO COMMUNICATION SYSTEMS LABORATORY IV. Binary Pulse Amplitude Modulation and Pulse Code Modulation

ECE 2274 Lab 2. Your calculator will have a setting that will automatically generate the correct format.

ECE 2274 Lab 2 (Network Theorems)

Group: Names: Resistor Band Colors Measured Value ( ) R 1 : 1k R 2 : 1k R 3 : 2k R 4 : 1M R 5 : 1M

EE 3302 LAB 1 EQIUPMENT ORIENTATION

B. Equipment. Advanced Lab

DC Circuits, Ohm's Law and Multimeters Physics 246

PHY 122 Shot Noise. Complete Shot Noise Pre- Lab before starting this experiment

EE 210: CIRCUITS AND DEVICES

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope.

Laboratory 2. Lab 2. Instrument Familiarization and Basic Electrical Relations. Required Components: 2 1k resistors 2 1M resistors 1 2k resistor

Wave Measurement & Ohm s Law

Lab E2: B-field of a Solenoid. In the case that the B-field is uniform and perpendicular to the area, (1) reduces to

Laboratory 2 (drawn from lab text by Alciatore)

Group: Names: (1) In this step you will examine the effects of AC coupling of an oscilloscope.

Pre-Lab. Introduction

Lab 1: Basic Lab Equipment and Measurements

Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm

ADC and DAC converters. Laboratory Instruction

Ph 3455 The Franck-Hertz Experiment

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons

PHY 123/253 Shot Noise

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS

ECE 53A: Fundamentals of Electrical Engineering I

Experiment 3 Topic: Dynamic System Response Week A Procedure

Lab #2: Electrical Measurements II AC Circuits and Capacitors, Inductors, Oscillators and Filters

CHAPTER 6. Motor Driver

Experiment 1: Instrument Familiarization (8/28/06)

10: AMPLIFIERS. Circuit Connections in the Laboratory. Op-Amp. I. Introduction

Operational Amplifier

Laboratory 6. Lab 6. Operational Amplifier Circuits. Required Components: op amp 2 1k resistor 4 10k resistors 1 100k resistor 1 0.

BME 3512 Bioelectronics Laboratory Five - Operational Amplifiers

Experiment 1: Instrument Familiarization

Physics 323. Experiment # 1 - Oscilloscope and Breadboard

Lab 6: Instrumentation Amplifier

EE EXPERIMENT 2 ANALOG AND DIGITAL MULTIMETERS INTRODUCTION. Figure 1: Internal resistance of a non-ideal ammeter.

Lab 2: Linear and Nonlinear Circuit Elements and Networks

Experiment 2: Transients and Oscillations in RLC Circuits

ECE 4670 Spring 2014 Lab 1 Linear System Characteristics

ADC, FFT and Noise. p. 1. ADC, FFT, and Noise

Model 7000 Low Noise Differential Preamplifier

Using Circuits, Signals and Instruments

total j = BA, [1] = j [2] total

AME140 Lab #2 INTRODUCTION TO ELECTRONIC TEST EQUIPMENT AND BASIC ELECTRONICS MEASUREMENTS

Laboratory 4: Amplification, Impedance, and Frequency Response

Sonoma State University Department of Engineering Science Spring 2017

PHYSICS 330 LAB Operational Amplifier Frequency Response

Exercise 4 - THE OSCILLOSCOPE

2 Oscilloscope Familiarization

When you have completed this exercise, you will be able to relate the gain and bandwidth of an op amp

BME/ISE 3512 Bioelectronics. Laboratory Five - Operational Amplifiers

Integrators, differentiators, and simple filters

SENSOR AND MEASUREMENT EXPERIMENTS

PHYS 1402 General Physics II Experiment 5: Ohm s Law

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION

Tektronix digital oscilloscope, BK Precision Function Generator, coaxial cables, breadboard, the crystal earpiece from your AM radio kit.

Introduction to the Laboratory

Lab #2: Electrical Measurements II AC Circuits and Capacitors, Inductors, Oscillators and Filters

Laboratory 3 (drawn from lab text by Alciatore)

PHYS Contemporary Physics Laboratory Laboratory Exercise: LAB 01 Resistivity, Root-mean-square Voltage, Potentiometer (updated 1/25/2017)

Experiment 2. Ohm s Law. Become familiar with the use of a digital voltmeter and a digital ammeter to measure DC voltage and current.

LAB 1: Familiarity with Laboratory Equipment (_/10)

ECE 2274 Lab 1 (Intro)

Physics 310 Lab 2 Circuit Transients and Oscilloscopes

Oscilloscope and Function Generators

AP034-OM-E Rev D ISSUED: January 2000 ²

BLACKBODY RADIATION PHYSICS 359E

Experiment 3 Topic: Dynamic System Response Week A Procedure

ECE 440L. Experiment 1: Signals and Noise (1 week)

Model 4402B. Ultra-Pure Sinewave Oscillator 1Hz to 110kHz Typical Distortion of % Serial No. Operating Manual

INTRODUCTION TO ENGINEERING AND LABORATORY EXPERIENCE Spring, 2015

ECE 6416 Low-Noise Electronics Orientation Experiment

EET 223 RF COMMUNICATIONS LABORATORY EXPERIMENTS

RLC Frequency Response

I = I 0 cos 2 θ (1.1)

DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139

Check out from stockroom:! Two 10x scope probes

DC MHZ PXI Differential Instrumentation Amplifier

2 AC and RMS. To pass this lab you must solve tasks 1-2. Tasks 3 and 4 are included in the grading of the course.

Notes on Experiment #1

Demonstrating Electromagnetic Noise in an Undergraduate Measurement and Instrumentation Course

University of Jordan School of Engineering Electrical Engineering Department. EE 204 Electrical Engineering Lab

PHYSICS 107 LAB #9: AMPLIFIERS

PHY203: General Physics III Lab page 1 of 5 PCC-Cascade. Lab: AC Circuits

Study of Analog Phase-Locked Loop (APLL)

ENG 100 Lab #2 Passive First-Order Filter Circuits

LLS - Introduction to Equipment

Introduction to Electronic Equipment

Experiment 3. Ohm s Law. Become familiar with the use of a digital voltmeter and a digital ammeter to measure DC voltage and current.

ECE 480: SENIOR DESIGN LABORATORY

Op-Amp Simulation Part II

Exp. 1 USE OF BASIC ELECTRONIC MEASURING INSTRUMENTS, PART I

Professional Article. What is noise?

Transcription:

Johnson Noise and the Boltzmann Constant 1 Introduction The purpose of this laboratory is to study Johnson Noise and to measure the Boltzmann constant k. You will also get use a low-noise pre-amplifier, a nice bandwidth limiter, and a very good AC voltmeter to directly observe the Johnson noise on a series of metal-film resistors. This should give you an appreciation for the effect of Johnson noise in circuits with high-gain. References: Melissinos and Napolitano, Section 3.6 (p. 122 forward). Elementary Statistical Physics, C. Kittel, Chapter 29. J. B. Johnson, Phys. Rev. 32, 97 (1928). H. Nyquist, Phys. Rev. 32, 110 (1928). Boltzmann s constant k relates microscopic degrees of freedom to temperature. It is probably most familiar as the constant in the ideal gas law, which relates the pressure P, volume V, number of particles N, and temperature T : pv = NkT Because the electrons (the charge carriers) in a metal can be thought of as a gas it should come as no surprise that their random motion can bring in Boltzmann s constant. Consider an ordinary metal resistor as a conductor made out of electrons. When we apply no electromotive force to the resistor, there is no potential difference across it on the average. But instantaneously, the random thermal motion of the charge carriers produces some instantaneous current, and so voltage. Although this voltage averages to zero, its square does not there is a finite RMS voltage across the conductor that depends on temperature. That is, there is electrical noise associated with a resistor that cannot be removed except by cooling. This effect was first studied experimentally by Johnson, and analyzed theoretically by Nyquist, in 1928. The power spectrum of this noise is given by the Nyquist formula dp = 4kT df (1) where dp is the amount of power in a frequency interval of width df, k is Boltzmann s constant, and T is the temperature. The factor of 4 in this formula is very interesting. I will not present a derivation of the Nyquist formula here, but the reference to Kittel s book above is excellent and readable. The derivation by Melissinos and Napolitano is sloppy and unconvincing, but provides a beautiful physical picture. The derivation by Nyquist is just physics salad good luck with that. 1

An interesting and important feature of the Nyquist formula is that it is flat. That is, the power spectral density dp/df is independent of the frequency. For this reason, Johnson noise is often called white noise. It is clear that this must be a classical approximation. The total power cannot be infinite, so at some high frequency this formula cannot be correct. Equation (1) assumes classical electromagnetism. When the photon energy hf becomes comparable to kt Eq. (1) breaks down, and dp/df approaches 0. For room temperature, this frequency f = kt/h is well into the THz range, so we will not be concerned with those effects for the low frequencies used in this lab. If we consider a resistor as discussed above, the power is voltage squared divided by resistance. Then the root-mean-square (rms) voltage across the resistor is d V 2 = 4kT Rdf Of course, to measure this we have to integrate df over some bandwidth which we observe. To make life easier, we will amplify this voltage before measuring it. If we have an amplifier with frequency dependent gain g(f), then the voltage we measure will be: V 2 measured = 4kT R g 2 (f)df (2) Procedure The process of this lab is straightforward. First, we will measure the integral in Eq. (2). That is, let B g 2 (f)df (3) Then our first step is to measure B. This is a long and tedious process. Then, we will measure V 2 for a series of different resistors. By plotting V 2 versus R, you can fit a straight line whose slope is 4kT B. Since you have already measured B, checking the temperature in the lab will give you Boltzmann s constant k. Before continuing, make sure you understand the chain of reasoning above. 2

2 Measurements Figure 1 shows a block diagram of the basic apparatus. You will also use a function generator and scope, which are not shown. Low noise high gain preamplifier band pass filter rms ac voltmeter output differential inputs Figure 1: Simplified block diagram of the apparatus. 2.1 Low-noise preamp The Signal Recovery model 5113 is a nice low-noise preamplifier. In order to measure the very small Johnson noise on resistors, we need high gain. In order to avoid other noise sources such as ground currents in the lab, we will operate the preamp in differential input mode. Set the input coupling for DC A-B. Set the gain to 1000. Set the Low and high frequency roll-off to flat. Connections to the input should be done carefully. Please note that there are two aluminum Bud boxes on the table, labeled A and B. On each of these boxes, one side has two female BNC connectors approximately the same distance apart and height off the table as the preamp inputs. Connect the Bud-box to the preamp using two short male-male BNC adapters. This is shown in Fig. 2. Figure 2: Connection of Bud box B to preamplifier. 3

2.2 Bandwidth filter In order to obtain a precise value for the integral B in Eq. (3). we need to have the gain function g(f) go to zero outside some well defined range. For this purpose, we will use a Krohn-Hite model 3362 4-pole filter. The model 3362 is actually a box with two independent filters. We will set the first filter to high-pass mode (blocking low frequencies) and the second filter to low-pass mode (blocking high frequencies). In this way, only frequencies between the low- and high-pass cut-offs are passed. Setting the filter is a bit tricky. Before you turn the filter on, check that the cables are connected as follows: Connect the output of the preamp to the CH 1+ filter input. Connect the CH1 output to the CH 2+ filter input. Connect the CH2 output to the Fluke 8846A multimeter. Now set up Channel 1: Look at the Channel display, and press the small button with a triangle under it until it reads 1. Press the Mode button until the Cutoff Frequency display reads h.p. (high pass). Press the Type button until the same display reads bu. (Butterworth). Press the Freq button once. Type 10 into the numeric display. Press the kilo button, and the then Freq button. This sets Channel 1 to a high-pass filter with cut-off of 10 khz. Perform an analogous procedure to set Channel 2 for a low-pass filter with a cutoff of 30 khz. You should now have a band-pass filter that passes signals between 10 and 30 khz. At this point, ask your professor to come check your work. 2.3 Digital multimeter We will make measurements with a Fluke 8846A digital multimeter. Connect the output of the filter to the voltage-input (not the sense input) of the fluke with a BNC-banana adapter (also known as a Pomona connector ). This device makes it easy to measure noisy signals and get a useful measure of the uncertainty in their value. This is done with the yellow Analyze button. Pressing Analyze and then Stats (blue button labeled F2) starts a series of measurements. The average and standard deviation of these measurements is continuously updated. 4

2.4 Calibrating the gain-squared-bandwidth integral First we need to measure the integral B from Eq. (3). To do this, we will put a known small signal into the preamplifier, and measure the output of the filter with the digital multimeter. We will do this for many different frequencies. The ratio of the filter output to the preamp input is the gain, g(f). By plotting the square of this function, we can perform a numerical integral to get B. For the above procedure to be accurate, we must choose a frequency range that is wide enough that the filter output has dropped to zero, and we must choose enough points to get an accurate numerical integral. Before you start, consider that the preamplifier has a gain of 1000, and cannot sustain an output voltage much greater than a volt. This means that we will need to use a voltage divider to reduce the input from something easy to measure. First, open up Bud-box A. You will see that the single input drops across two resistors, and that the two outputs are connected across the second resistor. This means that if you connect a sine wave to the input, the voltage across the two outputs will just be this input voltage multiplied by R 2 /(R 1 + R 2 ). You need to measure these two resistors now. Use the fluke, and the BNCalligator clip lead on the table. Make sure nothing is connected to the Bud-box before making your measurements. (Do you see why?) After you have measured the resistors, you will see that this box is designed to divide the input by approximately a factor of 1000. Make sure your measurements support this. Next, close up the box, and connect it to the preamplifier as shown in Fig. 2. Turn on the Wavetek Model 182A function generator and the Tektronix TDS 2002B oscilloscope. Connect the Low output of the function generator to the scope, and arrange to have a sine wave of approximately 1 V amplitude (2 V peak-to-peak) with zero DC offset. Set the frequency to something around the middle of the passband of the filter. Next, connect the output of the function generator to the input of Bud-box A. Using a BNC tee adapter, continue this signal to the multimeter. Set the multimeter to DC V (DC volts) and make sure your DC offset on the Wavetek is really nearly zero. Then set the multimeter to AC V and make sure the rms is nearly 0.7 V (which would be 1 V amplitude). Next, connect the filter output to the multimeter, and again set it for AC Volts. You should get a reading that is similar to what you just measured for the rms output of the Wavetek. It is important to get this right. You are dividing the Wavetek s output by something that is approximately 1000 with the bud-box, and then multiplying it by about 1000 with the preamp. Since your frequency should be within the pass-band of the filter, you should get about the same voltage out. 5

Next, set the Wavetek frequency to something well below the pass-band of the filter. Now you should get a voltage out of the filter that is very small. If you are far enough below the pass-band of the filter, this signal will just be the noise in the preamp and filter. Vary the frequency until you see the output of the filter start to rise. This is the frequency in which the filter output just starts to rise above the noise. Now, increase the frequency of the Wavetek to something several times the upper cut-off frequency of the filter, and do the same thing to find some upper frequency where the filter output drops to background noise. This gives you frequency bounds for your measurements. Now, you are ready to measure g(f). Set the Wavetek frequency to something low, outside the pass-band of the filter. Connect the BNC cable from the tee at the Bud-box to the Fluke multimeter, and measure the frequency and amplitude. Note: This measurement should be very precise. There is no need to use the Analyze feature. Disconnect that cable from the multimeter, and connect the output of the filter to the multimeter. Measure the output of the filter. Note: This measurement should be somewhat noisy. Use the Analyze Stats feature described above to average about 25-30 readings, and record the average value and standard deviation as uncertainty. Increase the frequency somewhat, and repeat. Continue the above procedure, working from a frequency so low that only noise is passed through the filter, slowly through the passband of the filter until you have again only noise. This should allow you to make a plot of the gain of the preamp/filter combination as a function of frequency. When the laboratory staff did this experiment, we got the following: Figure 3: Sample data for gainversus-frequency. In this case, 63 different frequency points were used starting at 0.1 khz and ending at 150 khz. The passband of the filter was set at 10 khz to 30 khz. You should carefully consider what frequencies to use in constructing the plot above, since this procedure can take a long time. For example, looking at the plot above it seems too many points were taken on the high frequency side and not enough on the low. Also consider spacing your points roughly logarithmically. 6

Here is an example of the kind of data you should get: f (khz) V in (V) V out (mv) V out (mv) 0.516 0.67948 6.790 0.200 1.011 0.67988 6.730 0.110 2.044 0.68122 6.610 0.150 3.021 0.68299 8.717 0.079 4.085 0.68508 19.908 0.078 5.031 0.68715 43.644 0.023 6.043 0.68897 89.982 0.016 6.967 0.69041 156.391 0.020 8.111 0.69193 272.437 0.042 9.049 0.69295 382.836 0.030 10.065 0.69386 492.876 0.040 11.000 0.69456 567.994 0.053 12.125 0.69525 623.219 0.031 13.077 0.69573 648.414 0.032 14.189 0.69619 663.630 0.038 15.020 0.69647 669.259 0.032 16.076 0.69678 672.417 0.036 17.125 0.69704 672.783 0.036 This is just a sample! You need to cover the entire frequency range from a point low enough that the output is just noise to a point high enough that again the output is just noise. But look at this example: you see that we have started at a low enough frequency that the output output is constant and very low, and then as we step the frequency up the voltage rises until it reaches a plateau that is approximately equal to the input voltage. With these data, and the values of the resistors in Bud-box A you can calculate the gain at each point. From that you can numerically integrate the gain squared (Don t forget the squared ) to get B. 2.5 Measuring Johnson noise Now that you have data for determining the integral B from Eq. (3), you will measure the Johnson noise on a series of resistors. This will allow you to plot V 2 versus R. According to Eq. (2) the slope will be 4kT B. First, we need to measure the resistance of the resistors you will use. In a cup you will find 11 high quality metal-film resistors. First, use the Fluke multimeter to measure the resistance of each one. Try to get ridiculously precise values. Also, be careful to handle the resistors as little as possible, and try to hold them by the wires. Remember the slope you are going to measure depends on the temperature in the lab. You don t want to heat up the resistors with your hands. 7

Next connect Bud-box B to the preamplifier. Gently open the box. Use two hands be careful not to stress the connections to the preamp. Inside you will see two alligator clips. Gently clamp one of the resistors with the alligator clips. Make sure the connection is good, and make sure the leads from the resistors do not touch the box. This is shown in Fig. 4. Figure 4: Connection of resistors in Bud -box B. For each resistor, measure the output of the filter with the Fluke multimeter. Remember you are looking for a small value, on the order of a milli-volt. Make sure you use the Analyze feature of the multimeter. We recommend letting it do about 100 averages. Don t forget to record the standard deviation this is your uncertainty. Here is an example of the kind of data you should get: R (kω) V out (mv) V out (mv) 0.99890 694.6 3.7 2.00383 896.2 4.5 3.00156 1061.7 5.0 4.00195 1202.8 5.4 4.98222 1330.4 6.5 6.03090 1453.8 6.5 6.98997 1559.7 8.2 8.02847 1659.7 8.6 9.05905 1758.0 7.9 9.98767 1841.6 9.3 Again, this is just an example. But please note that the uncertainty (standard deviation) gently rises with the voltage. Remember: this is the uncertainty in V out, you must calculate the uncertainty in V 2 from this. 2.6 Measuring temperature The last thing you need to measure is the temperature in the lab. The thermostat on the East wall will suffice. 8

3 Analysis Next, you need to analyze your data. We will determine B, then the slope of V 2 versus R. Finally, we will obtain a value for Boltzmann s constant. 3.1 Gain-squared-bandwidth integral You have measured the voltage out of the filter V out and the voltage into Bud-box A as a function of the frequency. You need to get the integral B. Enter those data into columns in a spreadsheet. Using the measured values of the voltage-divider resistors in this box, convert this to gain in another column. Pay attention to milli-volts and Volts so you don t blow any factor of 1000. In the pass-band of the filter, the gain should be on the order of 1000. Next, convert gain to gain-squared. Finally, perform a numerical integral of gain-squared versus bandwidth. You may use the trapezoid rule, or something more sophisticated To make sure your value is sensible, the answer you get for B should be approximately equal to 1 million (gain 1000, squared) times the difference between the high-pass cutoff and low-pass cutoff frequencies of the filter. So, for 10 khz and 30 khz cutoff frequencies, you should get B 2 10 10 Hz. Hopefully, you can get quite a few significant figures. 3.2 Slope of the Johnson noise You have measured the amplified Johnson noise as a function of resistance. You need to find the slope of V 2 versus R, and its uncertainty. The uncertainty is tricky! You must use a weighted least-squares fit, because your uncertainty will vary a lot from the low resistance to high. Why is this? You may think your uncertainty in the voltage was about the same for each different resistor. But wait! What you measured was V rms, or rather ( V 2 ) 1/2. Thus, you have to calculate your uncertainty in V 2. Do you know how to do this?? (Hint: the error bars will increase with increasing voltage.) Enter your V rms versus R data into columns in a spreadsheet. (11 rows one for each resistance.) Create two new columns for V 2, and its uncertainty. 9

Using the formulae for weighted least squares, find the slope of V 2 versus R and the uncertainty in this slope. You will need to create several new columns in your spreadsheet for the sums involved in the least squares fit. Note that you cannot simply ask Excel to generate a trend line! It will not weight the points correctly. You might want to plot the fitted line and its residuals to show how nice a job you did. Figure 5 shows an example. Note how the error bars increase with resistance. This is because the uncertainty in voltage was approximately constant, so the uncertainty in voltage-squared increases with voltage. Figure 5: Fit of Johnson-noise data. Note the error bars. 4 Boltzmann s Constant Finally, you are in a position to extract Boltzmann s constant. Eq. (2) says the slope of the line just measured is equal to 4kT B. From your slope, your measured value of B and from the temperature in the lab, derive a value for Boltzmann s constant and its uncertainty. Make sure you include some uncertainty in the temperature in your estimate. Half a degree Fahrenheit might be a good estimate. For the sample data shown in this write-up, the value obtained is: k = 1.389(9) 10 23 J/K This is to be compared with the known value of 1.38066 10 23 J/K. For the sample data, we are within about half a standard deviation, with no systematic effect. Can you do better? 10

5 Discussion The gain-squared-bandwidth integral B is sort of proportional to the bandwidth. We used 10 khz to 30 khz, which is 20 khz of bandwidth. Why not run the bandwidth way up more and get more signal? Why not set the upper cut-off of the filter to 100 khz, or even 1 MHz? Along the same lines, why not increase the gain? Instead of 1000, why not use 10000? Since the Johnson-noise voltage (squared) is proportional to resistance (we did get a nice straight line) why not use bigger resistors? We only used 1 kω to 10 kω. Why not use 100 kω or even a MΩ? Can you think of any way to improve the measurement? 11