Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Similar documents
TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Non-Descanned FLIM Detection in Multiphoton Microscopes

Things to check before start-up.

In Vivo Two-Photon Laser Scanning Microscopy with Concurrent Plasma- Mediated Ablation

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Supplementary Materials

How to align your laser for two-photon imaging

BASICS OF CONFOCAL IMAGING (PART I)

Leica TCS SP8 Quick Start Guide

Practical work no. 3: Confocal Live Cell Microscopy

You won t be able to measure the incident power precisely. The readout of the power would be lower than the real incident power.

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging

MULTIPHOTON MICROSCOPY. Matyas Molnar Dirk Pacholsky

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

The Zeiss AiryScan System, Confocal Four.

Basics of confocal imaging (part I)

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Be aware that there is no universal notation for the various quantities.

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Imaging Retreat - UMASS Customized real-time confocal and 2-photon imaging

Quick Start Guide. Leica SP5 X

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

3D light microscopy techniques

Zeiss 880 Training Notes Zen 2.3

Spectral phase shaping for high resolution CARS spectroscopy around 3000 cm 1

In Vivo Two-Photon Laser Scanning Microscopy with Concurrent Plasma-Mediated Ablation Principles and Hardware Realization - In Vivo Op...

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Zeiss 780 Training Notes

PAD Correlator Computer

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Kit for building your own THz Time-Domain Spectrometer

Light Microscopy. Upon completion of this lecture, the student should be able to:

ADVANCED OPTICS LAB -ECEN Basic Skills Lab

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

How-to guide. Working with a pre-assembled THz system

BUILDING A SINGLE MOLECULE DEVICE AND MEASURING THE DIFFUSION CONSTANT OF ZSYELLOW

Multiphoton Detection Unit (MDU)

Systems Biology. Optical Train, Köhler Illumination

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots

ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006

Leica TCS SP8 Quick Start Guide

Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope. Advanced Lab Course University of Basel

BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP LASER SCANNING CONFOCAL MICROSCOPY PRACTICAL CONSIDERATIONS

Technology Note ZEISS LSM 880 with Airyscan

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

plasmonic nanoblock pair

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Period 3 Solutions: Electromagnetic Waves Radiant Energy II

The DCS-120 Confocal Scanning FLIM System

Multi-channel imaging cytometry with a single detector

a) How big will that physical image of the cells be your camera sensor?

:... resolution is about 1.4 μm, assumed an excitation wavelength of 633 nm and a numerical aperture of 0.65 at 633 nm.

Cell Biology and Bioimaging Core

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

3D light microscopy techniques

Adaptive optics two-photon fluorescence microscopy

Rates of excitation, emission, ISC

Compact Dual Field-of-View Telescope for Small Satellite Payloads

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM

Last updated: May 2014 Y.DeGraaf

PCS-150 / PCI-200 High Speed Boxcar Modules

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

Design Description Document

Beam Shaping and Simultaneous Exposure by Diffractive Optical Element in Laser Plastic Welding

[4] (b) Fig. 6.1 shows a loudspeaker fixed near the end of a tube of length 0.6 m. tube m 0.4 m 0.6 m. Fig. 6.

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

University of Wisconsin Chemistry 524 Spectroscopic Components *

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Εισαγωγική στην Οπτική Απεικόνιση

Operating Instructions for Zeiss LSM 510

Laser Speckle Reducer LSR-3000 Series

NIH Public Access Author Manuscript Opt Lett. Author manuscript; available in PMC 2010 August 9.

Development of a High-speed Super-resolution Confocal Scanner

CONFOCAL MICROSCOPE CM-1

Heisenberg) relation applied to space and transverse wavevector

Leica SP8 TCS Users Manual

Laser stabilization and frequency modulation for trapped-ion experiments

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Leica SP8 Resonant Confocal. Quick-Start Guide

TCSPC at Wavelengths from 900 nm to 1700 nm

FPPO 1000 Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual

Optical Design of. Microscopes. George H. Seward. Tutorial Texts in Optical Engineering Volume TT88. SPIE PRESS Bellingham, Washington USA

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Microscopy from Carl Zeiss

LSM 710 Confocal Microscope Standard Operation Protocol

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

Instructions for the Experiment

Supplementary Information for. Surface Waves. Angelo Angelini, Elsie Barakat, Peter Munzert, Luca Boarino, Natascia De Leo,

Transcription:

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It has been widely used in neuroscience because it can image internal structures of the brain at fine resolution. Principle of fluorescence microscopy Molecules emits fluorescence are called fluorophores. These molecules have multiple electron orbits, which is approximately spaced by the energy of a photon of visible light. The electrons of fluorophores are usually in low-energy orbits, but they go into high-energy orbits by absorbing a photon. The probability of absorption depends on the wavelength, which is called excitation spectrum. After absorbing a photon, first the electron relaxes to other orbits with slightly lower energy, and then goes back to the lower-energy orbits by emitting light. The wavelength of emission is probabilistic, and follows emission spectrum. Importantly, emission spectrum has longer wavelength than excitation spectrum because of the relaxation before going back to the lower-energy orbits. Using dichrotics to dissociate excitation and emission wavelength, fluorescence microscopy can achieve higher contrast than the conventional microscopy using the same wavelength of light. In addition, the excitation and emission spectrum is specific to a fluorophore, which also leads to higher contrast. Two-photon microscopy Instead of using a light within the excitation spectrum, two-photon microscopy utilizes a light with twice as long wavelength. Two photons with half energy can be simultaneously absorbed by a fluorophore to elicit fluorescence. As the probability of this phenomena is proportional to square of the light intensity, the intensity of fluorescence emission drops biquadratically to the distance from the focal plane, which is faster than quadratic decrease of single-photon excitation. With this smaller focal volume, an image with finer resolution can be obtained. Furthermore, a light with longer wavelength is less scattered, which enables imaging deeper planes of the tissue. Methods System overview

We followed closely the design principles previously described (Tsai and Kleinfeld, 2009, overview shown in Figure 1). Briefly, a Ti:sapphire laser was pumped and mode-locked to emit a spectrum of light centered at 780nm. An optical path was designed that split the 500mW beam into two beams so that another group could construct a similar system in tandem. The beam was then wrapped around a vertical tower via a system of mirrors. A set of scanning mirrors directed the beam down through a series of lenses to expand the beam in order to completely fill the back aperture of the objective. Light was collected from a sample through the objective, reflected off a dichroic, passed through lenses, and directed into a photomultiplier tube (PMT). The PMT signal was converted to a voltage signal, and read out by a data acquisition board into a PC computer running custom imaging software. PMT collection lens sample dichroic tube lens scan lens scanning mirrors Laser Figure 1 System overview. Dashed lines indicate optical path. Solid lines indicate reflective surfaces (mirrors or dichroics). Laser and optical path We used a Spectra Physics Millenia V diode-dumped, CW visible laser to pump a spectra Physics Mode-locked Ti:sapphire laser. The difference in intensity profile between a mode-locked and non-mode locked laser is significant (see Troubleshooting section for more information on mode-locking and Figure 2).

Figure 2 Laser mode-locking. (Top) Laser not mode-locked. Intensity distribution is peaked. (Bottom) Mode-locked laser displaying significantly broadened intensity distribution. The power of the laser drops off as it traverses the optical path (see Troubleshooting section for more information on losing beam power across the scanning mirrors, and Figure 3). The path was sequentially aligned using a twomirror two-aperture method to center the beam on the succeeding point in the pathway. The proximal mirror was used to fit the beam through the proximal pinhole and the distal mirror was used to fit the beam through the distal pinhole. After bringing the beam to the top of a vertical tower it was bent at 90 degrees by the scanning mirror system. It then passed though a scan lens (focal length 50.8mm) and a tube lens (focal length 76.2mm), the purpose of which were to collimate the beam for an infinity-corrected 20X Ph1 Zeiss water-immersion objective. The constraints on path length and the placement of the objective and lenses is previously described (Tsai and Kleinfeld, 2009) but briefly, the scan lens is placed to optimally expand the beam, the tube lens collimates the beam coming out of the scan lens, and together they create an imaging condition between the scan mirror and the back aperture of the objective. Optimal expansion of the beam depends on the diameter of the back aperture and the characteristics of the beam, which is Gaussian in its intensity profile. Ideally the

full width at half maximum is taken as the diameter of the beam, which is then expanded to slightly overfill the back aperture. 28.9 scanning mirrors 78.5 59.7 68.5 79.2 76.5 74.5 72.3 89.9 99.3 Laser 100 Figure 3 Percentage of maximal intensity of laser beam at different points along the optical path (normalized to 100 at the laser outlet). Measured by a handheld Coherent FieldMate Laser Power Meter. Epi-fluorescence collection is used to direct light from the sample to the PMT. Collected light is collimated coming out of the objective, and reflected off of a dichroic that passes excitation wavelengths down to the sample and reflects emission wavelengths towards the detector apparatus. It passed through a collection lens (focal length 50.2mm), the purpose of which is to create an imaging condition between the back aperture of the objective and the active surface of the PMT, as well as to magnify the beam to fit the active surface of the PMT. In our system, given our lens setup, beam profile, and detector sizes, the path length from back aperture to collection lens was 101.6mm and from collection lens to detector was 6.5mm. Scanning system We used a set of open-loop X-Y scanners from Midwest Laser to complete X-Y scanning. Initially we assumed that these were silvered mirrors, but later learned that they are dielectric mirrors (see Troubleshooting section for more information). Each mirror had its own control board, which was powered by a custom power box Figure 4A), and received analog inputs from a National Instruments (NI) BNC-2090 attached to a PC. Output control from the PC was done through the custom program MPScope 2.0 (Figure 4B). Adjusting the Maximal Mirror Command Voltage changed the amplitude of the scan, and adjusting the frame rate or magnification changed the speed of the scan.

A B C D Figure 4 Custom electronics and software. (A) Power supply to drive the actuators in the scanning mirrors. (B) Screen shot of MPScope 2.0 showing at example of a real-time image. (C) Power supply for the PMT. (D) I to V converter and pre-amplifier that sits between the PMT and the data acquisition board. Detection electronics and software

We used a Hamamatsu GaAsP H7422 PMT for epi-fluorescence detection and powered it with a custom power supply (Figure 4C). A laser-blocking filter was mounted on the detector to reduce background from scattered photons. The PMT output was fed into a custom I-to-V converter box (see Troubleshooting section for more information and Figure 4D), where it was also amplified, and then into the same NI BNC-2090 and PC computer. The output was read out by MPScope and displayed in a window that allowed real-time manipulation of the image (Figure 4A). Sample preparation We prepared a sample for testing the microscope using thin lens paper, which has a fine microstructure clearly visible under 20X and 40X objectives. The paper was dipped in fluorescein, which should emit fluorescence with two-photon excitation in the 750-800nm range. The fluorescein paper was then placed under a 1.5 coverslip on a glass slide and sealed with nail polish. The paper must be dry after dipping in fluorescein so that the fluorescein molecules are forced to adhere to the paper and not remain dissolved in water. Results Two-photon fluorescence from fluorescein We succeeded in taking an image of a lens paper with fluorescein dye (Figure 5).

Figure 5 Image of a lens paper with fluorescein dye taken with two-photon excitation microscopy. Discussion/Troubleshooting Finally we could image the fibers of a paper with our two-photon microscope. At least at the center of the image, we could identify each fiber. However, the signal intensity and the contrast dropped rapidly at the peripheral of the image. We could not completely solve this problem, but this could be partly because of misalignment in the optical path. For example, if the back aperture of the objective is not exactly at the imaging plane of the scanning mirrors, the laser moves around and does not fill the back aperture properly when imaging peripheral pixels. The same could happen when the PMT is not at the image of the back aperture. Here are some other problems and solutions we came up with during setting up the system: 1. Mirrors/pinholes were not fixed properly.

Optical path alignment requires sequential alignment of mirrors. One day after finished the alignment, we noticed the first few mirrors and pinholes were loose. We had to tighten them and align the rest again. 2. Battery of laser pointer was not easy to remove. The laser pointers used for alignment require two AA batteries. The second one was not easy to replace. We taped the two batteries together so that they come out together. 3. Some parts required metric screws. We had to be careful not to use different screws. 4. Scanning mirror was broken. It worked properly at the beginning, but failed to work at high frequencies. We had to replace the scanning mirror. The broken one was labeled as broken. 5. I-V converter had too long time constant. First, the image taken had horizontal lines (Figure 6). This might be because the I-V converter had too long time constant and blurred the image along the axis of scanning. We replaced the I-V converter box. Figure 6 Example image taken with a improper I-V converter. 6. Dichroic mirror was not reflecting fluorescence. After we switched from single-photon setup with green-laser pointer to two-photon setup with Ti:Sapphire laser, we had to switch to a different dichroic mirror because the wavelength was different. 7. Scanning mirror was not reflecting laser. Because the scanning mirrors were not silvered mirrors but dielectic mirros, they did not reflect the infrared laser properly, and large portion of the laser power was lost through the mirrors. To compensate that, we took out the beam splitter and used all the laser power in this setup rather than sharing a laser with the other group. 8. Laser was not mode-locked properly. Sometimes the laser turned to be not mode-locked. We had to monitor the spectrum regularly to make sure it is mode-locked.

References In vivo two-photon laser scanning microscopy with concurrent plasma-mediated ablation: Principles and hardware realization. P. S. Tsai and D. Kleinfeld. In Methods for In Vivo Optical Imaging, Second Edition, R. Frostig, editor, 2009, CRC Press, 3:59-115. Appendix I Pictures of the completed system