AC-DC TRANSFER DIFFERENCE STANDARDS AND CALIBRATIONS AT NRC CANADA

Similar documents
IMPROVEMENTS IN THE NIST CALIBRATION SERVICE FOR THERMAL TRANSFER STANDARDS

NRC Publications Archive Archives des publications du CNRC

Publishable JRP Summary for Project T4 J03 JOSY. Next generation of quantum voltage systems for wide range applications

Josephson Voltage Sources

Ac-dc transfer standard shunts for frequencies up to 1 MHz. U. Pogliano, C.G. Bosco, M. Lanzillotti, D Serazio

Comparison of the NIST and NRC Josephson Voltage Standards (SIM.EM.BIPM-K10.b)

Establishing a High-precision AC/DC Transfer Standard using ET2001 ADS System

The Metrology Behind Wideband/RF Improvements to the Fluke Calibration 5790B AC Measurement Standard

AC-DC TCC with Built-in Tee Connector for Accurate Calibrations

New generation of cage-type current shunts at CMI

The Effect of Changing the Applied Sequence Using the TVC on the Accuracy of the AC Signal Calibration

50 OHM MULTIJUNCTION THERMAL CONVERTERS FOR AC VOLTAGE MEASUREMENTS UP TO 100 MHZ

THE National Institute of Standards and Technology

Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals

DC Voltage Linearity Measurements and DVM Calibration with Conventional and Programmable Josephson Voltage Standards

Chapter 6. The Josephson Voltage Standard

RF-POWER STANDARD FROM AC-DC THERMAL CONVERTER. L. Brunetti, L. Oberto, M. Sellone

SIM Regional Comparison of AC-DC VOLTAGE TRANSFER DIFFERENCE. SIM.EM-K6a, SIM.EM-K9, SIM.EM-K11 and SIM.EM-Supplementary 120 V / 53 Hz FINAL REPORT

Comparison of the Josephson Voltage Standards of the INTI and the BIPM

New generation of cage type current shunts developed using model analysis

A New Method for the Calibration of the mv Ranges of an AC Measurement Standard

6 - Stage Marx Generator

Wideband resistive voltage divider for a standard wattmeter

Power Quality Measurements the Importance of Traceable Calibration

IEEE/CSC & ESAS SUPERCONDUCTIVITY NEWS FORUM

Mitigation of Electromagnetic Interference Generated by GPIB Control-Network in AC-DC Transfer Measurement System

CCEM-K9 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS

Final Report Key Comparison COOMET.EM.BIPM-K10.b. Comparison of the 10 V Josephson Voltage Standards COOMET 542/RU/11. A.S. Katkov 1, P.A.

Traceability and Modulated-Signal Measurements

AFTER nearly 19 years since the invention of the

MASSACHUSETTS INSTITUTE OF TECHNOLOGY HAYSTACK OBSERVATORY

Validation Of The New Automatic System For AC Voltage Comparisons. Umberto Pogliano, Gian Carlo Bosco and Marco Lanzillotti

Development of thin-film multi-junction thermal converters

- Datasheet - Features: Version 1.1. Cryogenic Low Pass Filter Unit Type KA-Fil 2a

Comparison of the Josephson Voltage Standards of the NIMT and the BIPM

ACCREDITED LABORATORY

Design and Construction of a150kv/300a/1µs Blumlein Pulser

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CCEM Meeting, March 2017

Microwave measurements for planar circuits and components: State of the art and future directions. Dr. Uwe Arz PTB

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z

Comparison of the Josephson Voltage Standards of the LNE and the BIPM

at cryogenic temperatures

Performance of the Single Junction Thermal Voltage Converter at 1 MHz via Equivalent Circuit Simulation

Price List. Effective date 29 Sep of 13. Electrical Metrology Department. National Institute of Metrology (Thailand)

THE ELECTROMETRIC AC-DC TRANSFER STANDARD AS PRIMARY STANDARD AT IEN FOR AC VOLTAGES FROM 300 V TO 1000 V

An improvement for dual channel sampling wattmeter

Calibration of 100 MΩ Hamon resistor using current-sensing Wheatstone bridge. Ivan Leniček 1, Roman Malarić 2, Alan Šala 3

EUROMET 557 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Model 745 Series. Berkeley Nucleonics Test, Measurement and Nuclear Instrumentation since Model 845-HP Datasheet BNC

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Comparison of the Josephson Voltage Standards of the CENAM and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)

Application of Digital Sampling Method for Voltage Transformer Test Set Calibrations. Hüseyin Çaycı

Calibration Techniques for Precision Power Measurement in Semiconductor Proces Applications

Report on the Activities in Electricity and Magnetism within National Institute of Metrology (NIM), China CCEM 2017

ACCREDITED LABORATORY. LIBERTY LABS, INC. Kimballton, IA for technical competence in the field of Calibration

Verifying the Wideband Input of an AC Measurement Standard

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

Technical Reference for ET2001 Thermal Voltage Converters

CALIBRATION PROCESS QUANTITY REDUCTION OF THE THERMAL VOLTAGE CONVERTER STANDARD USING A THREE-STAGE BUILD-UP AND BUILD-DOWN METHOD

A COMPACT, AGILE, LOW-PHASE-NOISE FREQUENCY SOURCE WITH AM, FM AND PULSE MODULATION CAPABILITIES

5790A Automated AC Measurement Standard

Josephson Circuits I. JJ RCSJ Model as Circuit Element

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

P H Y S I C A L P R O P E R T Y M E A S U R E M E N T S Y S T E M. Quantum Design

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

AC Bias Characterization of Low Noise Bolometers for SAFARI Using an Open-Loop Frequency Domain SQUID-based Multiplexer Operating Between 1 and 5 MHz

EVOLUTION OF THE CRYOGENIC EDDY CURRENT MICROPROBE

Gain Slope issues in Microwave modules?

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Over-voltage Trigger Device for Marx Generators

IEC Electrical fast transient / Burst immunity test

AC Voltage Standards With Quantum Traceability

Improving CDM Measurements With Frequency Domain Specifications

CALIBRATION CMC 2, 4 ( )

Calibration Techniques for the Home Lab

Comparison of the Josephson Voltage Standards of the DMDM and the BIPM

Compact Series: S5065 & S5085 Vector Network Analyzers KEY FEATURES

Manufacture and Performance of a Z-interconnect HDI Circuit Card Abstract Introduction

Assoc. Prof. Dr. Mohammed Helmy Abd El-Raouf

Coaxial-type water load for measuring high voltage, high current and short pulse of a compact Marx system for a high power microwave source

Microwave Dielectrometer. Microwave Dielectrometer. Non- Destructive

Ultrastable Low-Noise Current Amplifiers With Extended Range and Improved Accuracy

Discover the. Blue Box. Difference. Electrical and Temperature Metrology Products Guide

A Guide for establishing primary AC-DC transfer standard using ET2001 ADS system

CCAUV/ CCAUV. Activities in KRISS. Wan-Cho Cho

Sub-micron SNIS Josephson junctions for metrological application

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PYLON ELECTRONICS INC. 147 Colonnade Road Ottawa, Ontario Canada K2E 7L9 Jim Mullins Phone:

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK

A Synchronized Two-Phase Sinewave Generator for AC Metrology System Compensations

PLANAR 814/1. Vector Network Analyzer

COAXIAL CURRENT SHUNTS FROM 1 ma TO 100 A

Calibration of High-Voltage Test Equipment

Voltage Sensors URV5-Z

A 30 GHz highly integrated LTCC antenna element for digital beam forming arrays

Transcription:

AC-DC TRANSFER DIFFERENCE STANDARDS AND CALIBRATIONS AT NRC CANADA Piotr S. Filipski, Michael Boecker Institute for National Measurement Standards, National Research Council Canada 1200 Montreal Rd. Bldg. M-36, Ottawa, Ontario, Canada K1A 0R6 Phone: 613 993 2313 Fax: 613 952 1394 E-mail: Peter.Filipski@nrc-cnrc.gc.ca Abstract: The paper presents the National Research Council (NRC) activities to establish and maintain ac-dc transfer difference standards of voltage and current at the lowest achievable uncertainty levels. It describes the AC Josephson Voltage Standard, a quantum-effect based standard of ac voltage based on pulse-driven Josephson junction arrays, recently introduced at NRC. 1. INTRODUCTION Over the last several years, a significant effort has been undertaken at the National Research Council (NRC) to establish ac-dc transfer difference standards of voltage and current at the lowest achievable uncertainty levels. The primary standards have been re-evaluated. The working standards based on thermal voltage converters (TVC), and voltage range extenders, were redesigned. The NRC ac-dc comparators were fully automated. The current range extenders, ac-dc shunts, were custom built, expanding calibration ranges in current and frequency, and significantly reducing uncertainties. These efforts have been successfully validated by NRC participation in all CCEM Key Comparisons of the ac-dc transfer difference: CCEM-K6.a ac-dc transfer difference at the lowest uncertainty level (less than 1 µv/v), CCEM-K6.c at high frequency (3 V, 1 MHz to 100 MHz), CCEM-K9 at high voltage, (200 V to 1 kv, 1 khz to 100 khz), CCEM- K11 at low voltage (10 mv to 100 mv, 1 khz to 1 MHz), and CCEM-K12 current (10 ma and 5 A, 10 Hz to 100 khz). Recently, in a close cooperation with the National Institute of Standards and Technology (NIST), USA, NRC has established a prototype ac Josephson Voltage Standard (ACJVS) based on a pulse-driven Josephson junction array. Our objective is to integrate this standard into the NRC system as the primary standard of ac-dc transfer difference. 2. PRIMARY STANDARD OF AC/DC TRANSFER DIFFERENCE The NRC primary standard of ac-dc transfer difference is a group standard, a set of four 20-mA multijunction thermal voltage converters (MJTC). They have been manufactured by a Canadian company, Guildline Instruments, closely following a design developed by Wilkins at the National Physical Laboratory, UK, [1]. The ac-dc transfer difference of these converters has been evaluated theoretically. The four members of the group have been mutually compared at different voltage and current levels. Over the years they were compared to other MJTCs (manufactured by PTB, Germany), and tested as voltage and current converters using two different Fast-Reversed DC Sources (both manufactured by AIST, Japan). As a result of these in-depth evaluations, a value of 0.0 µv/v, with an extended uncertainty of 0.4 µv/v, was assigned to the average ac-dc transfer difference of this group of MJTCs. The frequency range in which the MJTCs can be considered practically errorless extends from 20 Hz to 5 khz. The converters serve as primary standards of voltage ac-dc transfer difference at approximately 1 V (when used as voltage converters) and current ac-dc transfer difference at (10 to 20) ma (when used as current converters). 3. CALORIMETRIC THERMAL VOLTAGE CONVERTER The frequency range of the NRC working standards is extended by comparing them to a Calorimetric Thermal Voltage Converter (CTVC). This converter, originally designed at NRC, has a calculable and flat frequency characteristic between 1 Hz and 100 MHz. Its thermoelectric ac-dc transfer difference is less than 5 µv/v. A simplified schematic of the converter is shown in Fig. 1. The CTVC has been built around an internal tee. A calculable wideband thermal voltage converter was assembled at the end of one arm of the tee. The second arm of the tee, terminated by a type-n female connector, has been designed to have the same electrical parameters as the arm of a type-n female tee. During calibrations, the converters compared to the CTVC are attached 1

dedicated thermal voltage converter. For calibration of lower voltages, down to 2 mv, we use commercial micropotentiometers and a commercial thermal transfer standard, the Fluke model 792A. An original technique, minimizing the uncertainty of comparison, was developed at NRC to transfer the value of the ac-dc transfer difference from the volt level, the operating voltage range of the thermal voltage converters, to the millivolt level. Fig. 1. Simplified schematic of a CTVC. to this arm. The heater of the reference TVC is a miniature microwave rod resistor, mounted at the end of a short coaxial line. The stainless steel coaxial line, copper plated at the conducting surfaces, thermally insulates the heater from the body of the converter. The temperature of the resistor is sensed by a thermopile: 100 thermocouple junctions manufactured by partially copper plating a spiral of constantan wire. The mechanical design of the converter is very simple, and its frequency characteristic can be calculated. However, the thermoelectric ac-dc transfer difference of the CTVC is not negligible and is level dependent. The converter is therefore calibrated at low frequency, 1 khz, using a reference MJTC. This low frequency ac-dc transfer difference is then added to the calculable frequency dependent component of the ac-dc transfer difference. The CTVC is small, easily transportable and mechanically stable. It is mostly used as a standard voltage converter with a calculable frequency characteristic, for voltages between 1 V to 2 V, and also as a standard current converter. We have used it as a traveling standard in international comparisons with the National Metrology Institutes of the USA, the Netherlands, Germany, [4], Mexico and Japan. 4. VOLTAGE RANGE EXTENSION AND AC/DC TRANSFER COMPARATOR The voltage range of the working standards has been extended upwards to 1 kv by a set of custom range resistors, frequency compensated for a maximum flatness in the operating range of 10 Hz to 1 MHz. At the highest voltage level, 0.5 kv to 1 kv, we use a commercial high-voltage resistor with a The NRC ac-dc transfer voltage and current comparators have also been assembled in house, [5]. The commercial instruments such as generators, nanovoltmeters and amplifiers, are controlled via the IEEE-488 bus. However, the custom-built hardware and the IEEE-488 bus controller communicate with the computer through the serial bus - USB. The controlling software has been written in Visual Basic. We are in a process of rewriting it in LabView. The comparison between two thermal voltage converters or two current shunts is fully computerized. While TVCs still remain the most accurate standards of ac-dc transfer difference, the less accurate but much more user-friendly electronic instrumentation, based on digital sampling converters or integrated circuit rms-dc converters, is used in industry. An electronically-aided thermal voltage standard, such as Fluke model 792A, spans the gap between these two applications, and is a good compromise when the ultimate accuracy is not required. It is very stable, and reliable, and was very successfully used as a traveling standard in international comparisons, such as piloted by CENAM, Mexico, in the SIM comparison of ac-dc voltage transfer difference, [6]. The NRC has automated the calibration of the Fluke 792A, by designing computer controlled stepping motor range changing. The automated calibration of a Fluke 792A at 312 points is time and cost efficient. Only a test at the highest range requires a short physical intervention of the operator, for the addition of a 1 kv range resistor. 5. AC/DC SHUNTS The NRC ac-dc current transfer difference standards consist of thermal converters and wideband ac shunts. All shunts have been built in house, either from discrete resistors or from a resistive foil. Depending on the current range, three different configurations were used. For low currents, (10 to 50) ma, shunts were built using surface mount resistors arranged in a star-like 2

The middle cylinder, the resistor of the shunt, is manufactured from a thin manganin foil. The two remaining cylinders are made of a copper foil. The outside cylinder creates the return path for the input current. The inner-most cylinder serves as the highvoltage output potential lead. Fig. 2 shows NRC 100 A coaxial shunt during calibration of a commercial shunt. Details of construction of the NRC standard shunts are fully described in [7] and [8]. Fig. 2 Calibration of a 100 A commercial shunt (right) by comparing it to NRC100 A coaxial shunt configuration on a single-sided printed circuit board. attached to an output connector (type-n female). This assembly, together with an input connector, is mounted in a small component box. Shunts for medium range currents, (0.5 to 10) A, were assembled from a number of bulk metal foil resistors. The structure of each shunt consists of three printed circuit board plates, two output plates and one input plate, connected by a number of ribs. The multiple resistors of each shunt, 112 for the largest shunt of 10 A, are soldered in parallel between the output plates, around their outside perimeter. The input plate is removed from the output plates by the length of the ribs. Each linking rib supplies to the corresponding resistors only a fraction of the total current. An example of a commercial shunt of a similar construction is shown in Fig. 2. If a large number of parallel resistors is used, a high power shunt can be assembled from precise, low power components. By spatially separating the large input current connection from the output voltage leads, and supplying particular resistors separately through the dedicated ribs, one can considerably limit the electromagnetic coupling between the input and the output, otherwise a source of a significant shunt error. The four high current shunts, (20, 40, 80 and 100) A, are of a cylindrical coaxial design. They have been manufactured in cooperation with BEV, Austria. Each shunt consists of three coaxial foil cylinders mounted between four copper plates (two input and two output plates), on a glass-fiber epoxy cylinder. The presented shunt set is used in the frequency range of 10 Hz to 100 khz. The ac-dc transfer differences of the shunts in the set are lower than 100 µω/ω at 100 khz. At this frequency, the customer shunt calibration uncertainty varies between 10 µa/a for a 10 ma shunt to 50 µa/a at 100 A for a 100 A shunt. 6. AC JOSEPHSON VOLTAGE STANDARD The primary standards of dc voltage and resistance are based on quantum effects, the Josephson effect and the quantum Hall effect. Only the primary ac voltage standard relies on an artifact, a multijunction thermal voltage converter, the parameters of which depend on the fabrication, not on fundamental constants. Lead by the NIST, a significant effort, has been undertaken by the international community to extend application of the Josephson effect based circuits to the realization of a quantum primary standard of the ac voltage. Two methods of ac voltage synthesis are being pursued: a step-wise synthesis of an ac voltage using programmable Josephson junction arrays, and a pulse-modulation synthesis, using pulse-driven Josephson arrays. NRC has established, in cooperation with NIST, a prototype ac Josephson Voltage Standard (ACJVS) based on a pulse-driven Josephson junction array. The operation of a pulse-driven ACJVS is based on a pulse-quantizing property of the Josephson junction. A Josephson junction excited by a short pulse of an appropriate amplitude and duration will respond with a perfectly quantized voltage pulse. The time-integrated area of such a voltage pulse is given by the ratio of two fundamental constants h/2e, the Planck constant h and the charge of an electron e. The polarity of the quantum-accurate Josephson pulse is changed by changing the polarity of the exciting pulse. The exciting pulse does not have to be very accurately known or formed, as long as it is good enough to stimulate the junction response. By modulating the density of 3

Fig. 3. In response to a train of exciting pulses, top, a Josephson array generates a corresponding train of quantum-accurate pulses, bottom. these exciting but imperfect pulses, one can synthesize a quantum-accurate ac voltage, Fig. 3. To increase the generated voltage, a large number of junctions connected in series is excited simultaneously. The NIST pulse-driven ACJVS has been described in [9], a detailed list of literature is given in a review paper [10]. Fig. 4 shows the schematic of the pulse-driven ACJVS, a photograph of the NRC system is shown in Fig. 5. At the heart of the system is a superconducting chip containing two independent arrays of 5120 Josephson junctions. The chip is mounted in a cryoprobe and placed in a liquid helium Dewar. Each array can generate up to 110 mv of low frequency rms ac voltage; 100 mv in ordinary applications. Two arrays can be connected in series to generate 200 mv. The exciting pulses are formed by combining the output of a 10 GHz Pulse Code generator with a 15 GHz sinewave generator in a 6 db stripline directional coupler. The proper timing between these two signals is achieved by shifting the 15 GHz sinewave phase. The array is galvanically insulated from the directional coupler by a DC block (inner and outer capacitors). These capacitors attenuate the low frequency component of the exciting pulses, shown in Fig. 3, top as a sinewave. The exciting pulses are Fig. 4. Schematic of the AC Josephson Voltage Standard based on an impulse-driven array. reconstructed to their proper form by reinserting the low frequency component from a separate arbitrary waveform generator, followed by a current amplifier. The arbitrary waveform generator has to be synchronized with the Pulse Code Generator and its phase and amplitude adjusted to accurately replace the attenuated signal. The output of an array is connected to a room temperature cryoprobe head through a twisted pair copper wire. The output voltage can be then applied to a tested instrument, such as a thermal transfer standard Fluke 792A. In the NRC system, usually a small 6 MHz Low Pass RC Filter (LPF) and a coaxer (equalizer) is inserted between the ACJVS output and the input of the tested instrument. One of the factors limiting the accuracy of the ACJVS is an error due to the considerable transmission line length, over 1.5 m, between the quantum-accurate voltage reference plane at the array output and the input of the instrument under test. This frequency-dependent probe-leads error is relatively large and requires correction. It can be determined either theoretically or experimentally, [10]. The 6 MHz LPF has been adjusted to partially compensate the probe leads error. It additionally filters out residual high-frequency components of the array output voltage. The coaxer suppresses common mode error voltage. 4

Fig. 5. Photograph of AC Josephson Voltage Standard at NRC. The thermal voltage standard was compared to the ACJVS at (2, 10, 20, 100, 200) mv in the frequency band from 2.5 khz to 100 khz (and up to 1 MHz with reduced accuracy). At 100 mv and 2.5 khz, the ACJVS and TVC results agree better than 1 µv/v, [11]. We have also conducted with NIST the first international comparison of Quantum AC Voltage Standards, [12]. Results of this comparison, parameters of the NRC ACJVS and some detailed measurement results will be discussed at the Conference. 7. SUMMARY NRC standards of ac-dc voltage transfer difference are based on multijunction thermal converters and calorimetric thermal voltage converters of NRC design. The current standards are based on custom set of wideband ac shunts. A quantum-based AC Josephson Voltage Standard is being integrated into the NRC system to serve as a primary standard of ac-dc transfer difference. ACKNOWLEDGEMENTS The ACJVS has been developed by Sam Benz, Charlie Burroughs, Paul Dresselhaus, et al., at NIST and NRC benefits from their significant support and involvement. REFERENCES [1] F.J. Wilkins, Theoretical analysis of the ac-dc transfer difference of NPL multijunction thermal converter over the frequency range dc to 100 khz, IEEE Trans. Instrum. Meas., vol. IM-21, pp. 334-340, 1972. [2] R.F. Clark, P.S. Filipski, D.C. Paulusse, Improvements in the NRC AC-DC Transfer Capabilities, IEEE Trans. Instrum. Meas., vol. 46, April 1997, pp. 365-368. [3] P.S. Filipski, R.F. Clark, D.C. Paulusse, Calorimetric Thermal Voltage Converter as a Wideband Calculable Standard of AC-DC Difference, IEEE Trans. Instrum. Meas., vol. 48, April 1999, pp. 387-390. [4] P.S. Filipski, C.J. van Mullem, D. Janik, J.R. Kinard, T. Lipe, B. Waltrip, Comparison of High Frequency AC-DC Voltage Transfer Standards at NRC, VSL, PTB and NIST, IEEE Trans. Instrum. Meas., vol. 50, April 2001, pp. 349-352. [5] P.S. Filipski, R. Rinfret, An Automated AC-DC Transfer Calibration System, IEEE Trans. Instrum. Meas., vol. 49, April 2000, pp. 279-284. [6] S. Campos, P.S. Filipski, D. Izquierdo, E. Afonso, R.P. Landim, L. Di Lillo, T. Lipe, "Final report: SIM regional comparison of ac dc voltage transfer difference (SIM.EM.K6a, SIM.EM-K9 and SIM.EM- K11)", Metrologia, 46 (Technical Supplement), 01004, 2009. doi:10.1088/0026-1394/46/1a/01004 [7] P.S. Filipski, M. Boecker, AC-DC Current Shunts and System for Extended Current and Frequency Ranges, IEEE Trans. Instrum. Meas., vol. 55, August 2006, pp. 1222-1227. [8] P.S. Filipski, M. Boecker, M. Garcocz "20-A to 100-A AC-DC Coaxial Current Shunts for 100 khz Frequency Range," IEEE Trans. Instrum. Meas., vol. 57, August 2008, pp. 1637-1641. [9] S.P. Benz and C.A. Hamilton, A Pulse-Driven Programmable Josephson Voltage Standard, Appl. Phys. Lett., vol. 68, no. 22, pp. 3171-3173, May 1996. [10] B. Jeanneret, S.P. Benz, Application of the Josephson effect in electrical metrology, Eur. Phys. J. Special Topics vol.172, No. 1, pp. 181 206, June 2009. [11] P.S. Filipski, M. Boecker, S.P. Benz, C.J. Burroughs, Establishing an AC Josephson Voltage Standard at NRC, 2010 Conference on Precision Electromagnetic Measurements digest : CPEM 2010 (Daejeon, Korea, June 13-18, 2010), p. 20-21, [cdrom] 2010. ISBN 9781424467945, submitted to IEEE Trans. Instrum. Meas. [12] P.S. Filipski, J.R. Kinard, T.E. Lipe, Y. Tang, An International Comparison of Quantum AC Voltage Standards between NIST and NRC, Proceedings of the NCSL International Workshop and Symposium, Providence, RI, USA, July 25-29, 2010, paper 1185, 6 pp., [cd rom] 2010. ISBN 1584640634 5