OP735. Benchtop Optical Power Meter Instruction Manual

Similar documents
OP710. Multichannel Optical Power Meter Instruction Manual. (Also supports the OP710-ANX)

Using the Integrating Sphere in the Multiple Application Platform (MAP)

Agilent 8703B Lightwave Component Analyzer Technical Specifications. 50 MHz to GHz modulation bandwidth

Agilent 86030A 50 GHz Lightwave Component Analyzer Product Overview

Advanced Test Equipment Rentals ATEC (2832)

TIA-952 Optical/Electrical Converter. Operating Instructions

TIA-525 Optical/Electrical Converter Operating Instructions

PSW-002. Fiber Optic Polarization Switch. User Guide

Agilent N4373B Lightwave Component Analyzer Testing advanced 40Gb/s components with highest relative and absolute accuracy

TIA-3000 Optical / Electrical Converter Operating Instructions

Model 7000 Low Noise Differential Preamplifier

Model MV106J/MV116J. ±10nVdc to ±11Vdc Precision DC Voltage Standard Source. Operating Manual

GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier

TIA-4000 Optical/Electrical Converter. Operating Instructions

Agilent 81980/ 81940A, Agilent 81989/ 81949A, Agilent 81944A Compact Tunable Laser Sources

TIA-500 Optical/Electrical Converter Operating Instructions

OVA-50 Optical Variable Attenuator User s Manual

USER MANUAL OLS-100 Optical Light Source

Agilent 81600B All-band Tunable Laser Source Technical Specifications December 2002

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

Pro. SimpliFiber. Getting Started Guide. Optical Power Meter and Fiber Test Kits

UNICOM 2500 UNICOM Installation Operation Programming. Universal Communication Converter

MODEL W Power Amplifier

Advanced Test Equipment Rentals ATEC (2832)

80-MHz Balanced Photoreceivers Model 18X7

TIA-1200 Optical / Electrical Converter Operating Instructions

80-MHz Balanced Photoreceivers Model 18X7

TIA-3000 Optical/Electrical Converter. Operating Instructions

IQS-2100/FLS-2100 OPTICAL LIGHT SOURCE. Telecom Test and Measurement

OM7AT & OM7AR1 (Stand alone) ROM7AT & ROM7AR1 (Rack mount)

TDS-535 Tuned Dipole Set Operation Manual

TIA-527 Balanced Optical/Electrical Converter. Operating Instructions

NOW WITH BROADER TUNING RANGE

R&S ZVT Vector Network Analyzer Specifications

TIA-1200 Optical/Electrical Converter. Operating Instructions

Copyright Teletronics International, Inc. Patent Pending

We bring quality to light. MAS 40 Mini-Array Spectrometer

Spectrum Analyzers 2680 Series Features & benefits

GM8036 Laser Sweep Optical Spectrum Analyzer. Programming Guide

DC to 3.5-GHz Amplified Photoreceivers Models 1591 & 1592

Agilent 83440B/C/D High-Speed Lightwave Converters

MADE FOR THE TRADE! XL fibertools Series

Copyright Teletronics International, Inc. Patent Pending

SPM-50 RF Spectrum Power Meter PC Software User Manual

DC to 12-GHz Amplified Photoreceivers Models 1544-B, 1554-B, & 1580-B

PIECAL Model 541 Frequency Calibrator with Totalizer Operating Instructions

Agilent N4375B 20 GHz and 26.5 GHz Single-Mode Lightwave Component Analyzer Data Sheet

E/O and O/E Measurements with the 37300C Series VNA

Power Measurment. Hand-held Design Brings Peak Performance to the Field. Introduction. Features

LI-1500 Light Sensor Logger

LIGHTWAVE, OPTICAL TEST EQUIPMENT

PIECAL 211 Automated Universal RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

R&S NRP-Zxx Power Sensors Specifications

Agilent 8614xB Optical Spectrum Analyzer Family Technical Specifications

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc.

Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016

PIECAL 311 Automated Universal RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

Agilent N4376B 20 GHz and 26.5 GHz Multimode Lightwave Component Analyzer. Data Sheet

Artisan Scientific is You~ Source for: Quality New and Certified-Used/Pre:-awned ECJuiflment

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

A-CUBE-Series High Sensitivity APD Detector Modules

Contents. CALIBRATION PROCEDURE NI PXIe-5698

R&S ZNC Vector Network Analyzer Specifications

R&S NRP Power Meter Family Specifications

PIECAL 211 Automated RTD Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

Amplified Photodetectors

SAS-551 Passive Monopole Antenna Operation Manual

Keysight N4373E 43.5/50/67 GHz Single-Mode Fiber Lightwave Component Analyzer for 100G/400G/1T Electro-Optical Test

ME7220A. Radar Test System (RTS) Target Simulation & Signal Analysis for Automotive Radar Exceptional Performance at an Affordable Price.

RHS SERIES RH Set Point Controller/Probe

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

Model 3210C. 100 Ampere AC Current Standard. Operating Manual

TIA-2000 Optical/Electrical Converter. Operating Instructions

PXI Modules 3066 PXI Multi-Way Active RF Combiner Data Sheet

Agilent 86146B Optical Spectrum Analyzer Technical Specifications

MA24104A. Inline High Power Sensor. True-RMS, 600 MHz to 4 GHz

Handheld Infrared Thermometers

N432A Thermistor Power Meter DATA SHEET

DIRECTIONAL FIBER OPTIC POWER MONITORS (TAPS/PHOTODIODES)

Instruction Manual Model M Switch, DPDT, Manual Select

It s Our Business to be EXACT

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

INGAAS FAST PIN (RF) AMPLIFIED PHOTODETECTORS

TEGAM, INC. SINGLE/DUAL CHANNEL HIGH VOLTAGE AMPLIFIER MODEL 2340/2350. Instruction Manual PN# CD Publication Date: June 2006 REV.

KRAMER ELECTRONICS LTD. USER MANUAL MODEL: 912 Power Amplifier. P/N: Rev 2

SAS-543 Biconical Antenna Operation Manual

Photon Count. for Brainies.

R&S RSC Step Attenuator Specifications

ASX SERIES PACIFIC OPERATION MANUAL AC POWER SOURCE POWER SOURCE

Headend Optics Platform (CH3000)

Agilent 71400C Lightwave Signal Analyzer Product Overview. Calibrated measurements of high-speed modulation, RIN, and laser linewidth

SPECIAL SPECIFICATION 1118 Data Fiber Optic Transceiver

S3602A/B Vector Network Analyzer Datasheet

HP 8509B Lightwave Polarization Analyzer. Product Overview. Optical polarization measurements of signal and components nm to 1600 nm

instructions for Models QC-10 and QC-20 Sound Calibrators MODELS QC-10/QC-20 SOUND CALIBRATORS GENERAL DESCRIPTION

VITESSE SEMICONDUCTOR CORPORATION. Bandwidth (MHz) VSC

PowerPXIe Series. Analog Power Meter ADVANCE SPEC SHEET

SAS-562B Active Loop Antenna Operation Manual

Variable Optical Delay Line

1995 Metric CSJ SPECIAL SPECIFICATION ITEM Multimode Video Fiber Optic Transmission Equipment

Transcription:

Benchtop Optical Power Meter Instruction Manual www.optotest.com 1.805.987.1700

Contacting OptoTest Corporation 1.805.987.1700 (7:30 a.m. to 5 p.m. PST) www.optotest.com engineering@optotest.com OptoTest Corp. 4750 Calle Quetzal Camarillo, CA 93012 USA Notice of Proprietary Rights The design concepts and engineering details embodied in this manual, which are the property of OptoTest Corporation, are to be maintained in strict confidence. No element or detail of this manual is to be spuriously used or disclosed without the express written permission of OptoTest Corporation. All rights are reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without prior written permission from OptoTest Corporation. COPYRIGHT 2016 by OptoTest Corp ALL RIGHTS RESERVED PRINTED IN THE UNITED STATES OF AMERICA MnOP735-RevC 1 of 12

Table of Contents Overview 3 Initial Preparation 4 Definition of Specifications 5 Front Panel Operation 9 Display Operation 10 Smart Remote Head Option 11 USB Control of the OP735 11 Warranty Information 12 2 of 12

Overview The OP735 can be configured up to 4 channels with a mix of InGaAs, Silicon, and High Power Detectors. This unit is ideal as a compact stand-alone power meter or used in conjunction with a stabilized light source to measure insertion loss. Designed for a high speed data acquisition on fiber optic components, the OP735 provides absolute or relative measurements with high resolution and linearity. The OP735 is available starting at one channel up to four channels and can be configured for a variety of detector and connector interfaces. Available detector options: IN1 IN3 IN5 IN10 SI3 R S 1mm InGaAs detector with Universal Adapter 3mm InGaAs detector with Universal Adapter 5mm InGaAs detector with Universal Adapter 10mm InGaAs detector with Universal Adapter 3mm Silicon detector with Universal Adapter Electrical port for Remote Head Detector Digital port for Integrating Sphere 3 of 12

Initial Preparation Unpacking and Inspection The unit was carefully inspected, mechanically, electrically and optically before shipment. When received, the shipping carton should contain the items listed in Standard Contents. Account for and inspect each item. In the event of a damaged instrument, write or call OptoTest Corp, California. Note: Be aware that accessories such as detector adapters, remote head detectors, and high performance reference cables will be located inside a small box labeled Accessories Inside. If this box is not included with the original shipment, contact OptoTest of their nearest distributor. Please retain the shipping container in case re-shipment is required for any reason. Damaged In Shipment All instruments are shipped F.O.B. Camarillo when ordered from OptoTest. If you receive a damaged instrument you should: 1. Report the damage to your shipper immediately. 2. Inform OptoTest Corporation. 3. Save all shipping cartons. Failure to follow this procedure may affect your claim for compensation. Standard Contents 1. Model OP735 Multichannel Optical Power Meter 2. Power Cord (U.S. Shipments only) 3. USB A-B cable 4. Certificate of Calibration and if requested the Metrology Report 5. Instruction Manual(s) 6. CD with applicable software and documentation (if ordered) 7. Rack mount kit (optional) 4 of 12

Definition of Specifications Dynamic Range The dynamic range spans from the maximal power level the instrument can measure without major saturation to the detector to the minimal power level where the thermal noise of the detector becomes greater than the current produced by the incident light. For accurate power measurements it is NOT recommended to measure power levels at either end of the dynamic range. (see Linearity) Linearity Photodetectors are, by nature, very linear over a wide range of optical input powers, but the power meter electronics can affect the overall system linearity. The power meter linearity is characterized and specified to know the measurement accuracy and linearity over the full dynamic range. For accurate insertion loss measurements only power levels that fall within the range with the best linearity (+/-0.05dB) should be measured. Calibration Wavelength The calibration wavelengths are the nominal wavelengths of the instrument is calibration points. The exact wavelength of each particular calibration is stated in the certificate of calibration. Calibration Traceability The detector s absolute calibration data is directly traceable to N.I.S.T. at the specified calibration wavelength and the specified power level, typically -10dBm. 5 of 12

Definition of Specifications Spectral Responsivity Depending on the detector type, InGaAs (Indium Gallium Arsenide) or Silicon the spectral responsivity, the efficiency of the detector to convert optical power into electrical current changes with wavelength. Responsivity of InGaAs Detectors Responsivity of Silicon Detectors Note that other detector types are available such as IN5 (5mm InGaAs) IN10 (10mm InGaAs) as well as WSR (wide spectral range) and might exhibit a different spectral responsivity. Absolute Accuracy The absolute accuracy specification includes the total measurement uncertainties involved in the calibration process including the transfer of the absolute power standard from N.I.S.T. Optical Power Meter, Channel Performance For multichannel instruments, the power meter circuit converts and digitizes the optical power level with the given sampling interval. Changes in light levels such as modulation will be averaged within that sampling interval. Instrument, Warm up Time Optical power meters, in general, do not need any warm-up time unless the instrument has to acclimate to a changing environment. In order to calibrate the instrument or to perform stable measurements, the instrument should be acclimated for 15 minutes for each 5ºC of temperature differential. For example if the instrument was stored at 18ºC and brought into an environment of 28ºC the instrument should be allowed to warm up for 30 minutes. 6 of 12

Definition of Specifications Recommended Recalibration Period This is the recommended time period for re-calibration in order to maintain accuracy specifications. The recommendation is made based upon statistics on detector aging; however it is up to the metrology policies and procedures within each company to define the calibration cycles on optical power meters. Optical Power Meter, Fiber Compatibility The amount of aerial coverage of the detector, or the portion of the light emitted from the fiber being measured, depends on the mechanical features of the optical interface, the active area of the detector and the numerical aperture (NA) of the fiber. A fiber with a large NA, for example 100/140 multimode fiber, might not under fill a small area detector hence the absolute power reading will be less than actual. Return Loss Range The lower end of the return loss (low return loss = high reflection) defines the level where the instrument is saturated by large reflections. The higher end of the return loss (high return loss = very weak reflections) is given by capability of the instrument to amplify and resolve reflection out of the noise floor. Return Loss Accuracy The Return Loss Accuracy is measured using an optical variable attenuator connected to a >98% reflector. The insertion loss of the attenuator is initially quantified against a reference optical power meter. The actual attenuation is then used to calculate the generated reflection, where the resulting reflection = 2x (variable attenuation + insertion loss of attenuator) + reflector coefficient. Accuracy of return loss measurements can also be affected by the reference cable and any excessive losses at the front panel interface. 7 of 12

Definition of Specifications Reference Cable The reference cable is the cable with which the DUTs will be measured against. Typically reference cables are required to be of a defined quality with a specified connector/endface polish. Instrument, Environmental Operating Temperature: This is the temperature range in which the instrument will conform to the specifications after the specified warm up time. Storage Temperature: This is the temperature range at which the instrument can be stored with the power off without any damage or any loss of specification to the instrument. It is required that the instrument be brought back to within the operating temperature range before it is turned on. Humidity: The relative non-condensing humidity levels allowed in the operating temperature range. 8 of 12

Front Panel Operation Channel Indicator Wavelength Indicator Ch1-12.45 1310nm 22.4 C dbm db dbm Wavelength Relative Measurement Absolute Measurement Channel Selection Power Reading Internal (ambient) Temperature Wavelength The wavelength button toggles through the available calibration wavelength. Typically for power meters with InGaAs this is 850nm, 980nm, 1300nm, 1310nm, 1480nm, 1550nm, and 1625nm. Relative Measurement The db button switches the power meter into relative measurement mode. At the same time it stores the current absolute power reading as the reference. The reference power is displayed above the relative power reading (see illustration of Display). If the instrument is already in relative measurement mode pressing the db button stores the current power level as the new reference. For each wavelength and for each channel a reference reading can be stored. Absolute Measurement The dbm button switches the power meter into absolute measurement mode. Channel Selection By pressing the right button the instrument display switches to the next channel, it will stop at the last channel. Similarly the left button switches the instrument to the previous channel. Each channel retains the calibration wavelength, absolute or relative measurement mode and the corresponding reference power levels. 9 of 12

Display Operation Channel Display Wavelength Reference Power Ch 01 1310nm 22.4 C -12.45 dbm Ch 01 Ref: -12.45dBm 1310nm 22.4 C -0.316 db Absolute Power Internal (ambient) Temperature Relative Power Channel Display Shows the current selected channel. Wavelength Displays the currently selected calibration wavelength. Absolute Power The absolute power is displayed in dbm. Relative Power The relative power is displayed in db. It is the difference between the reference power and the measured absolute power. Internal Temperature The internal, ambient temperature is displayed in either ºF (Fahrenheit) or ºC (Celsius), that selection is performed with a USB command. USB When communicating with the instrument via USB, the unit will display the OptoTest logo. 10 of 12

Smart Remote Head Option The OP735 when manufactured has the option of having a digital OPM port. This port interfaces with a smart remote head and/or a smart integrating sphere. When no remote head is connected the unit will function normally as described in this manual. For the remote head to be properly loaded by the system, the remote head should be connected while the unit is powered down. Connect the remote head to the mainframe and power up the system. Linearity and absolute calibration are established by the remote head, not the mainframe. USB Control of the OP735 The OP735 can be controlled via the USB bus. Upon request, OptoTest can supply the appropriate DLLs along with sample programs to facilitate the software creation process. For these DLLs please contact sales@optotest.com. 11 of 12

Warranty Information OptoTest Corp. warrants this product to be free from defects in material and workmanship for a period of one year from date of shipment. During the warranty period we will, at our option, either repair or replace any product that proves to be defective. To exercise this warranty contact OptoTest Corp. headquarters. You will be given prompt assistance and return instructions. Repairs will be made and the instrument returned, transportation prepaid. Repaired products are warranted for the balance of the original warranty period, or at least 90 days. NOTE: Do not send instruments for any reason without contacting OptoTest headquarters first. 12 of 12

For Application Notes, more detailed Testing Instructions, and the most up-to-date OptoTest News go to www.optobuzz.com

www.optotest.com 1. 805.987.170 0