Solving Connection Challenges in On-Wafer Power Semiconductor Device Test. Application Note Series. Introduction

Similar documents
What Is An SMU? SEP 2016

Ensuring that Power Supply Performance Meets Your Requirements. Application Note Series

Techniques for Proper and Efficient Characterization, Validation, and Reliability Testing of Power Semiconductor Devices. applications guide

Application Note Se ries

Application Note Series

The Challenge of Integrating Three Critical Semiconductor Measurement Types into a Single Instrument Chassis

200 V channels 1-3 Common LO channel Maximum DCV Both 3030 V 202 V 40 V 42 V Maximum DCI 1. DC 122 ma A 4.5 A AC + DC 100 µa 100 µa

Model 4210-MMPC-W. Multi-measurement Prober Cable Kit. Overview

Switching Between C-V and I-V Measurements Using the 4200A-CVIV Multi-Switch and 4200A-SCS Parameter Analyzer APPLICATION NOTE

Application Note Series

Solutions for Production Testing of Connectors. Application Note Series. Introduction. Test Description. Number 2208

High Voltage Component Production Testing with Two Model 2410 SourceMeter Units. Application Note Series. Introduction. Test System Configuration

Figure 1 Figure 3 Figure 2

Next Generation Curve Tracing & Measurement Tips for Power Device. Kim Jeong Tae RF/uW Application Engineer Keysight Technologies

Model 4210-MMPC-L. Multi-measurement Prober Cable Kit. Overview. Quick start guide topics. Related documents

Application Note Se ries. Low Current Measurements. Basic Current Measurements. Shunt vs. Feedback Ammeters. Number 100.

of High Power Semiconductor Device Testing

2520 Pulsed Laser Diode Test System

Keysight Technologies Direct Power MOSFET Capacitance Measurement at 3000 V

Ultra-Fast I-V Module for the Model 4200-SCS

Applications Overview

Welcome! Device Characterization with the Keithley Model 4200-SCS Characterization System.

Welcome! Device Characterization with the Keithley Model 4200-SCS Characterization System.

Varistor Verification with the 2400 Digital SourceMeter Instrument. Application Note Se ries. Introduction. Test System Configuration

Using the isppac-powr1208 MOSFET Driver Outputs

Application Overview: Simplified I/V Characterization of DC-DC Converters

Application Note Series. Solutions for Production Testing of Connectors

Controlling Inrush current for load switches in battery power applications

IRF130, IRF131, IRF132, IRF133

TLP/VF-TLP/HMM Test System TLP-3010C/3011C Advanced TLP/HMM/HBM Solutions

SOURCE-MEASURE UNITS INCREASE PRODUCTIVITY AND ACCURACY IN AUTOMATED TESTING. Lee Stauffer. Keithley Instruments, Inc.

Physical Test Setup for Impulse Noise Testing

Super Junction MOSFET

2302 Battery Simulator 2306, 2306-PJ Battery/Charger Simulators

Using the 4200A-CVIV Multi-Switch to Make High Voltage and High Current C-V Measurements APPLICATION NOTE

CHAPTER 2 A SERIES PARALLEL RESONANT CONVERTER WITH OPEN LOOP CONTROL

Super Junction MOSFET

Test Structure Design for Parallel Testing

UNISONIC TECHNOLOGIES CO., LTD

Features. Symbol JEDEC TO-204AA GATE (PIN 1)

New Wide Band Gap High-Power Semiconductor Measurement Techniques Accelerate your emerging material device development

Model 4210-MMPC-C. Multi-Measurement Prober Cable Kit Quick Start Guide. Overview. Prober cable kit contents

Battery Simulator Battery/Charger Simulators

TECHNICAL REPORT: CVEL

Quad SPST JFET Analog Switch SW06

Keithley Instruments, Inc.

CHAPTER 7 HARDWARE IMPLEMENTATION

Fast IC Power Transistor with Thermal Protection

Achieving 3000 V test at the wafer level

Power Semiconductor Devices

Op Amp Booster Designs

Using a Linear Transistor Model for RF Amplifier Design

Electronic Circuits EE359A

UNIT 3: FIELD EFFECT TRANSISTORS

MOSFET = 0V, I D = 10V, 29A) = 500V, V GS = 0V) = 0V, T C = 400V, V GS = ±30V, V DS = 0V) = 2.5mA)

Experiment (1) Principles of Switching

High Voltage Power Operational Amplifiers EQUIVALENT SCHEMATIC R1 R2 C1 R3 Q6 4 CC1 5 CC2 Q8 Q12 3 I Q Q16. +V s

MOSFET UNIT V DSS. Volts I D W/ C T J. Amps E AR = 0V, I D = 10V, I D = 88A) = 200V, V GS = 0V) = 160V, V GS = 0V, T C = ±30V, V DS = 0V) = 5mA)

Welcome! Device Characterization with the Keithley Model 4200-SCS Characterization System.

APPLICATION NOTE. Wide Range of Resistance Measurement Solutions from μω to PΩ

FREDFET FAST RECOVERY BODY DIODE UNIT V DSS. Volts I D W/ C T J. Amps E AR = 0V, I D = 10V, I D = 88A) = 200V, V GS = 0V) = 160V, V GS = 0V, T C


Discover. Blue Box. the. Difference. High Resistance Metrology Products Guide

MOSFET = 0V, I D = 16.5A) = 10V, I D = 200V, V GS = 0V) = 0V, T C = 160V, V GS = 0V) = ±30V, V DS. = 2.5mA)

RF/Microwave Switching Systems Subtleties: Achieving the Performance You Need

Keysight Technologies Accurate Capacitance Characterization at the Wafer Level

DatasheetArchive.com. Request For Quotation

HRLD150N10K / HRLU150N10K 100V N-Channel Trench MOSFET

CONSONANCE. 4A, Standalone Li-ion Battery Charger CN3761. General Descriptions: Features: Pin Assignment: Applications:

UNDERSTANDING HORIZONTAL OUTPUT STAGES OF COMPUTER MONITORS

P D Storage Temperature Range T stg 65 to +150 C Operating Junction Temperature T J 200 C

PA92. High Voltage Power Operational Amplifiers PA92

Enhancement Mode N-Channel Power MOSFET

TPS1120, TPS1120Y DUAL P-CHANNEL ENHANCEMENT-MODE MOSFETS

Application Note 5011

Customized probe card for on wafer testing of AlGaN/GaN power transistors

Application Note 5012

Methods to Achieve Higher Currents from I-V Measurement Equipment. Application Note Series. Pulse sweeps. Number 3047

COMPONENT VENDORS have come

Enhancement Mode N-Channel Power MOSFET

High Voltage Off-Line Linear Regulator by Jimes Lei, Applications Engineering Manager

Precision IC Comparator Runs from a5v Logic Supply

Production Testing of High-Intensity, Visible LEDs. By Doug Rathburn, Keithley Instruments, Inc.

Agilent 4070 Series Accurate Capacitance Characterization at the Wafer Level

Difference between BJTs and FETs. Junction Field Effect Transistors (JFET)

Enhancement Mode N-Channel Power MOSFET

Ultra-Fast NBTI/PBTI Package for the Model 4200-SCS

Enhancement Mode N-Channel Power MOSFET

APT5010B2FLL APT5010LFLL 500V 46A 0.100

12N60 12N65 Power MOSFET

APT1003RBLL APT1003RSLL

Enhancement Mode N-Channel Power MOSFET

GS61008T Top-side cooled 100 V E-mode GaN transistor Preliminary Datasheet

HCA80R250T 800V N-Channel Super Junction MOSFET

TPH3207WS TPH3207WS. GaN Power Low-loss Switch PRODUCT SUMMARY (TYPICAL) Absolute Maximum Ratings (T C =25 C unless otherwise stated)

8N Amps, 600/650 Volts N-CHANNEL POWER MOSFET 8N60 MOSFET N 600V 7.5A 1,2 OHM. Power MOSFET. DESCRIPTION FEATURES

results at the output, disrupting safe, precise measurements.

UNISONIC TECHNOLOGIES CO., LTD

Selecting Telecommunication Test Equipment To Maximize Throughput and Accuracy. By Robert Green Keithley Instruments, Inc.

APT30M30B2FLL APT30M30LFLL

Transcription:

Number 3276 pplication Note Series Solving Connection Challenges in On-Wafer Power Semiconductor Device Test Introduction Measuring DC and capacitance parameters for high power semiconductor devices requires sufficient expertise to optimize the accuracy of various measurements. Even for those with this level of expertise, managing set-up changes between ON-state, OFF-state and capacitancevoltage (C-V) measurements can be time consuming and prone to errors; this is especially true in the on-wafer environment. This application note provides an overview of how the Keithley Model 8020 High Power Interface Panel greatly simplifies the user test experience by: Integrating key measurement-enabling accessories for ON-state, OFF-state and C-V measurements Enabling all C-V measurement parameters, at both the componentlevel and circuit-level, with minimal connection changes at the device under test (DUT) Building in flexibility with a variety of connector options Expanding easily to support future measurement requirements and new device terminals can involve tens to hundreds of amps. Therefore, creating these curves requires a high current instrument, such as the Keithley Model 2651 High Power System SourceMeter instrument; when two Model 2651s are configured in parallel, they can generate up to 100 in pulsed current. This instrument is coupled with a lower power instrument, such as Keithley s Model 2636B, to drive the control terminal (e.g., gate) of a threeterminal device. To produce optimal high current measurements, managing the resistance and inductance in the cabling is imperative. This can be achieved using low inductance coaxial Kelvin connections and minimizing the inductive loop area at the DUTs. (Probe station geometries often dictate this inductive loop area.) The gate instrument also requires Kelvin connections so that high current pulses at the FET s source terminal don t affect the gate voltage and change the operating point of the device. High power devices often exhibit high gain and may be prone to oscillation during testing. dding series resistors helps eliminate such oscillation. In typical test systems, the resistor is inserted between the lower power instrument and the gate terminal. This resistor must be housed in a shielded enclosure so that electrostatic interference is not coupled in when characterizing other parameters, such as gate leakage. Otherwise, the series resistor would have to be removed for those other tests. Series resistance values can vary between devices and it may be necessary to try several values in order to select the appropriate value. It can be a design challenge to house and connect to these resistors properly when performing on-wafer measurements. Figure 1 illustrates the typical setup for high current ON-state measurements. OFF-State Measurements OFF-state characterization includes measurement of the breakdown voltage and leakage current of the device. These parameters are typically tested at the Configuring Common Semiconductor Device Measurements using Traditional ccessories and Standalone Instrumentation review of the requirements for the DC and capacitance measurements to be made on the DUT is an important first step. ON-State Measurements Characteristic curves for a power transistor, FET, or diode include a plot of its typical output characteristics. Output characteristics for some power devices Figure 1. Typical ON-state measurement setup.

device s maximum rated voltage, which can be hundreds to thousands of volts. high voltage instrument, such as Keithley s Model 2657 High Power System SourceMeter instrument, which has 3kV capability, is ideal for measuring the OFF-state parameters. The Model 2636B instrument can also be used to bias the device in a certain state or to measure current at other device terminals. Because device breakdown can subject this lower power instrument to high voltage from the Model 2657, the addition of a protection module is critical to protecting the Model 2636B from damage. The breakdown profile of some devices is very sharp, with device impedance changing by several decades in microseconds. Such transitions can exceed the speed of the instrument in transitioning to current limit control, as well as lead to high currents in the test system due to rapid discharge of cable and system capacitance. Inserting a resistor in series with the Model 2657 enforces an absolute current limit that can protect devices and probes from high transient current. It s important to ensure the resistor is rated for high voltage and is in a grounded enclosure to protect the operator from exposure to high voltage. This will require creating a custom enclosure to mate with the high voltage instrument and the connector at the probe station, as well as extra cables to connect the resistor box to both the instrument and the positioner. Measurements of leakage current range from picoamps to microamps for most power semiconductor devices. t these levels, system leakage can affect the accuracy of the measurement, requiring the use of low noise triaxial cabling and connections. Both the Model 2657 and the Model 2636B have triaxial connectors. Triaxial cabling allows for a driven guard, which shunts extraneous system leakage current from the measurement node and reduces settling time by nulling the effect of cable capacitance. Figure 2 illustrates the typical setup for a high voltage breakdown voltage measurement. Figure 2. Typical OFF-state measurement setup. CVU HCUR LCUR Bias Tee Figure 3. Typical high voltage C-V measurement setup. Capacitance-Voltage Measurements power MOSFET device engineer or circuit designer may have an interest in various capacitance parameters, including input capacitance (C ISS ), output capacitance (C OSS ), reverse transfer Bias Tee capacitance (C RSS ), gate-source capacitance (C GS ) and drain-source capacitance ( ). Similar capacitance parameters exist for BJTs and IGs. Capacitance changes as a function of voltage and tests are performed up to the rated voltage of the 3kV 3kV

C Signal Coax-to-Triax dapter, version 1 High Voltage Source must connect here Inner C Return Outer C ISS Figure 4(a). Coax-to-triax adapter that connects coax outer shield to triax outer shield. Using this adapter for low current DC I-V measurements would result in degraded measurements because DC guard cannot be carried out to the DUT. C Signal C Return device. The rated voltage of power devices likely exceeds the built-in DC voltage capability of most capacitance meters, so bias tees are required to achieve these measurements. Because of the high voltage involved, it s essential to protect any lower power instruments used in the setup, requiring the use of protection modules to accomplish this task. Figure 3 illustrates a typical high voltage capacitance measurement. Coaxial cabling and connections are common with capacitance measurements. To minimize series inductance in the measurement, the shields of the coaxial cables are tied together as close as possible to the DUT. If the test fixture or probe station uses triaxial connections, it is important to consider how to adapt coaxial C-V connections to triax. Figure 4 provides the schematic for two available coax-to-triax adapters. In both cases, the adapters must be removed when transitioning to low-level DC current measurements in order to achieve optimal performance. In Figure 4a, the adapter prevents carrying the DC guard to the DUT. Therefore, any measurements made would include error due to system Coax-to-Triax dapter, version 2 Inner Outer Figure 4(b). Coax-to-triax adapter that connects coax outer shield to triax inner shield. Using this adapter requires making connection changes between DC l-v measurements and high voltage C-V measurements. leakage. In Figure 4b, the adapter would result in problems when DC guard is routed to the DUT if the shields were still shorted together at the DUT as necessary for C-V measurements. Therefore, it would be necessary to change connections when moving from C-V to DC to avoid error or instrument damage when moving from C-V to DC I-V measurements. The different capacitance parameters of the device require different connections for C source and C measurement terminals of the capacitance meter. These connections must be moved around the device while maintaining bias at the device s drain or collector. lthough capacitance measurements are inherently two-terminal measurements, for three-terminal devices, it s important to consider the capacitance between either of the two terminals to the third terminal. If this third terminal is not handled from an C perspective, capacitance to this terminal will be included as an error in the measurement. For example, a C ISS measurement on a power MOSFET requires the bias tee at the drain to enable high voltage bias. However, C ISS is a combination of the gate-to-source and Need C EXT >> in order to eliminate from C ISS measurement Figure 5. Need for external capacitor when measuring C ISS in order to remove error due to. C EXT gate-to-drain capacitance. Therefore, it s necessary to eliminate the drain-source capacitance ( ) from the measurement. To eliminate, it s essential to short this capacitance from an C perspective while still allowing for hundreds to thousands of DC volts across the same terminals. This requires a high voltage rated capacitor that is of higher capacitance (or lower impedance) than. Note Figure 5 for an illustration of the measurement problem. The presence of high voltages also requires that this capacitor be housed in a safe enclosure. Finally, it s necessary to switch out the capacitor when measuring other capacitance parameters on this device. Setting up this measurement is no easy task. Moving from Measurement to Measurement review of Figures 1 3 illustrates the setup changes required when transitioning from ON-state to OFF-state to capacitance measurements. Table 1 summarizes the various accessories required throughout the testing of a particular device. It is easy to see how the opportunities for user error can increase with the number of devices to be tested. Configuring Common Semiconductor Measurements using the Model 8020 High Power Interface Panel The Model 8020 High Power Interface Panel integrates several of the

Interconnects and ccessories Cables Series Resistor Bias Tee Bias Tee Overvoltage Protection Module Measurement Configuration ON-State (Output OFF-State Characteristics, RdsOn) (Breakdown voltage) Low resistance, low High voltage, inductance Kelvin triaxial cabling connections Yes for measurement stability measurement-enabling accessories shown in Table 1, thereby eliminating cable and setup transitions. The Model 8020 supports one high voltage instrument (up to 3kV), up to three lower power instruments and up to two high current instruments to achieve 100 pulsed capability. That allows the Model 8020 to be used in testing a large variety of power devices, especially at the current and voltage levels being tested on wafers. Yes for absolute current limiting Capacitance- Voltage Coaxial RF Frequency (up to 1MHz) No No Yes No Yes Yes Table 1. Summarizes use of critical accessories in the various power semiconductor device measurements. No ll channels support full Kelvin (remote sense) connections, so that ON-state measurements can even be performed with instruments (up to 10 pulse capability). The Model 8020 includes built-in overvoltage protection for the lower power channels, eliminating the need for external protection modules and the complications involved in connecting them. Screw terminals located under the interface panel s top cover provide for inserting any two-terminal resistor in series with the high voltage or channels. Resistors are housed within a safe, shielded enclosure, which manages the safety and electrostatic interference issues associated with using series resistors for oscillation suppression and gate leakage measurements, as well as when using resistors to enforce absolute current limits for breakdown voltage measurements. high voltage rated resistor kit is provided with the Model 8020, which includes a wide range of commonly used values so that users can get up and running quickly. Transient current limit protection is also available on the high voltage and channels. This can be useful in breakdown voltage testing to protect sensitive probes from the high transient currents that may appear when the device has a sharp breakdown characteristic. This transient protection affects currents greater than 100m with transitions less than 100µs. This protection can also be easily bypassed in order to achieve the maximum pulse current rating on the channels. One of the key advantages the Model 8020 offers is evident when measuring capacitance. Rather than employing a single bias tee at the high voltage terminal of the device, the Model 8020 includes a bias tee at each of the high voltage and channels to allow a bias tee at up to #1 #1 HV C RSS = C GD C ISS C OSS #2 #2 #1 C GS C GS #3 #3 #2 C Guard Figure 6. Measurement configurations for common high voltage capacitance parameters: C ISS, C RSS, C OSS. Configurations possible using Model 8020 with 8020-CVU option. Overvoltage protection is not shown but is integrated in the Model 8020.

four device terminals. Having a bias tee at each device terminal improves isolation between the capacitance meter and the instrument. When transitioning between measurements of different capacitance parameters, the connections between the Model 8020 and the DUT remain static. The setup transition between the tests of several capacitance parameters occurs at the input of the bias tee, where the user can easily manipulate the C source and measurement terminals of the capacitance meter using connectors that are low voltage and fixed. Figure 6 depicts connection changes between circuit level capacitance parameters (C ISS, C OSS, C RSS ). Note that connections at the device do not change. llowing a bias tee at each device terminal also simplifies the capacitance measurement. s shown in Figure 5, a C ISS measurement requires a fairly large high voltage rated capacitor to short out from an C perspective. Using bias tees at every terminal allows for the use of the bias tee capacitance as an C short. This is accomplished by using a single coaxial cable to connect the C inputs of the bias tees, which are connected to the drain and source terminals. See Figure 7. Capacitance measurements are available on triaxial connectors with the Model 8020. The return of the capacitance meter is routed on the outside shield of the triax, and this eliminates the need for coax-to-triax adapters, as well as cabling changes between DC I-V measurements and C-V measurements. This means that the shields can remain shorted together for C-V measurement without negatively impacting DC I-V measurements. Typical bias tees induce significant leakage current. The bias tees used in the Model 8020 are of guarded triaxial design and have programmable multimode capability so that low current DC OFF-state measurements, ON-state DC measurements, and high voltage C-V measurements can be made through the bias tee without changing the connections between the Model 8020 and the DUT (except for connection of the Model 2651 to the FET s drain terminal when performing ON-state measurements). C ISS Figure 7. Having a bias tee at each terminal enables simpler capacitance measurement. The capacitance of bias tee #3 can be used to short out and thus get a true measurement of C ISS. Figure 8. 50 50 CVU #2 In addition to these testing advantages, the Model 8020 manages common reference for all the instruments that connect to it. It manages the system #1 #3 Model 8020 High Power Interface Panel Manages Series resistors High voltage bias tees Overvoltage protection Cabling differences between measurements Connector interface to probe stations and fixtures C 3 >> so C MESURED = C ISS interlock so that a single normally-open switch at the device access point can disable the high voltage output of all connected instruments. The Model

8020 also has daisychain capabilities, which allows for system expansion when it s necessary to add channels to the system. It continues to manage the common reference and system interlock across all connected Model 8020 mainframes. The Model 8020 provides the flexibility to mate with a variety of connectors. The lack of standardized commercial connectors for high voltage and high current measurements means probe station positioners and test fixtures have a variety of connector options. Because the Model 8020 has several connector card options for use with high voltage,, and common channels, it can be used with a variety of probe stations or with a custom fixture. Conclusion The Model 8020 High Power Interface Panel greatly reduces the frustration test engineers encounter when performing comprehensive DC I-V and C-V testing of power semiconductor devices. The role of the Model 8020 for on-wafer power device measurements is shown in Figure 8. By integrating the capabilities of several measurement-enabling accessories, the Model 8020 minimizes both connection changes and opportunities for user error in technology development applications. Specifications are subject to change without notice. ll Keithley trademarks and trade names are the property of Keithley Instruments. ll other trademarks and trade names are the property of their respective companies. Greater Measure of Confidence KEITHLEY INSTRUMENTS 28775 UROR RD. CLEVELND, OH 44139-1891 440-248-0400 Fax: 440-248-6168 1-888-KEITHLEY www.keithley.com BENELUX +31-40-267-5506 www.keithley.nl FRNCE +33-01-69-86-83-60 www.keithley.fr ITLY +39-049-762-3950 www.keithley.it MLYSI 60-4-643-9679 www.keithley.com SINGPORE 01-800-8255-2835 www.keithley.com.sg BRZIL 55-11-4058-0229 www.keithley.com GERMNY +49-89-84-93-07-40 www.keithley.de JPN 81-120-441-046 www.keithley.jp MEXICO 52-55-5424-7907 www.keithley.com TIWN 886-3-572-9077 www.keithley.com.tw CN 86-10-8447-5556 www.keithley.com.cn INDI 080-30792600 www.keithley.in KORE 82-2-6917-5000 www.keithley.co.kr RUSSI +7-495-664-7564 www.keithley.ru UNITED KINGDOM +44-1344-39-2450 www.keithley.co.uk For further information on how to purchase or to locate a sales partner please visit www.keithley.com/buy Copyright 2014 Keithley Instruments Printed in the U.S. No. 3276 9.05.14