MaxxFlow HTC Flow measurement for dry bulk solids

Similar documents
MaxxFlow HTC. Product Information. Flow Measurement for Dry Bulk Solids

MaxxFlow HTC Flow Measurement for Dry Bulk Solids

SolidFlow 2.0. Solid mass flow measurement. Product Information. SWR engineering Messtechnik GmbH enveatm a trademark of Environnement S.

Manufactured by M-Sens 2 Online-Moisture Meter for Solids. Moisture % Product Information.

ProGap Microwave barrier

Operating Instructions. MaxxFlow HTC Measurement of Bulk Solids for High Flow Rates

WATERFLUX 3000 Quick Start

WATERFLUX 3000 Quick Start

Technical Specification FLUXUS ADM 8127B

J Series induq VMZ MAGNETIC INDUCTIVE FLOW SENSORS. DB_Magnetic_Inductive_Flow_Sensors 05/2016

OPTIFLUX 5000 Technical Datasheet

Ambient Conditions. Operating Conditions. Pressure

Electromagnetic flowmeters and switches DWM 1000/2000

SITRANS F flowmeters SITRANS F M. Transmitter Transmag 2 with sensor 911/E 4/77. Overview

Level sensor Magnetostrictive, high-resolution measuring principle For sanitary applications, model FLM-H

OPTIFLUX 5000 Technical Datasheet

Level Sensors Reed Contact Chain

Flow - Rotor. Characteristics. Applications. Mechanical-magnetical, Infrared. Quantity measurement for filling processes

Type PLC Fitting. Valve for Continuous control. Diaphragm valve. Output

[Pick the date] Moisture Sensor

WATERFLUX 3070 Technical Datasheet

TIDALFLUX 2300 F Technical Datasheet

OPTIMASS 1000 Technical Datasheet

Level Measurement Techniques In A Modern Quarry

Digital conductivity transmitter

Rosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna

Weighing Instruments

Bulk Flow Meter LB 442 CONTACTLESS MEASUREMENT

Digital flow transmitter for continuous flow measurement

Electromagnetic Flow Transmitter

Technical Specification FLUXUS ADM 8x27

Digital inductive conductivity transmitter

Technical Information Nivector FTC968, FTC968Z

DOG-4. Oscillation Flowmeter. for gases. Measuring ranges: m 3 /h air. : PN 40; t max. p max : 120 C

prosonic T FMU230, FMU231

ph or ORP Transmitter

Magnetic Inductive Flow Sensors VMI Series

Magnetic-inductive flow meter. PITe / UMF2. Technical Datasheet. Maintenance-free Nearly no pressure drop Robust design Easy installation and start-up

Refer to manual. Failure to read message could result in personal injury or serious damage to the equipment or both.

6GHz Intelligent Radar Level Meter Product Manual. Model:KRD520 SERIES

Wind Transmitter First Class

SAT.. Electromotoric actuators for valves ACVATIX. with a 5.5 mm stroke and 300 N positioning force

UFM 800 W, C Ultrasonic flowmeter for water and wastewater

Magnetic Inductive Flow Sensors induq. Series VMI

Ultrasonic level transmitter, non-contact

Drinking water monitoring

Weighing Instruments. Applications. Characteristics. DMS strain gauges (load cells), DMS amplifier

Target disk flowmeters Series DP Metallic flowmeter for liquids and gases

ChuteMaster Microwave-blocked chute detection and point level instrument

Digital flowmeter for continuous flow measurement

Digital flowmeter for continuous flow measurement

INSTRUCTION MANUAL VIBRATION SWITCH CVS 100 LC

SN+ MID MASTER + APPLICATION WATER METERS

INLINE flowmeter for continuous flow measurement

SITRANS F flowmeters. SITRANS F System information MAGFLO electromagnetic flowmeters 4/9

OPTIFLUX 1000 Technical Datasheet

INLINE flowmeter for continuous flow measurement

Flow Measurement SITRANS F M

SCHMIDT Flow Sensor SS SS FB

SITRANS F flowmeters. SITRANS F M System information MAGFLO electromagnetic flowmeters. 4/18 Siemens FI

DIGITAL LENGTH GAUGES

Insertion Flow sensor for continuous flow measurement

INLINE flowmeter for continuous flow measurement

Digital flow ELEMENT transmitter for continuous flow measurement

Magnetic-Inductive flow Meter fmi

Digital batch controller

Target disk flowmeter for liquids and gases

Coriolis Mass Flow Measuring System promass 40 E

KMS305W. ManuFlo Page 1 MANU ELECTRONICS PTY LTD Flow Measurement Products. Electromagnetic Wafer Flowmeters with S300 display

KFS. Flap-type flow meter. Design and range of application

Rotamass TI Coriolis Mass flow meter

SITRANS F flowmeters SITRANS F US SONOCAL /184 Siemens FI

T40FH. Torque flange. Special features. Data sheet

Installation and Operating Instructions Differential pressure transmitter for potentially explosive areas. MG-1000-EEx USAGE STRUCTURE AND DESIGN

SALES SERVICE +358 (0)

T40B. Torque Flange. Special features. Data sheet. Overall concept

FH CMF Coriolis Mass Flow Meters

KFS-EM. Overall length. Al. PTFE coated Alu, painted Al. coated Methacrylate / acrylic nitrite

Wind Direction Transmitter - compact - GMR, analogue output xxx

Temperature Sensor for Build-In System G3/8" FLEXadapt

Electronic timer CT-ERS ON-delayed with 2 c/o contacts Data sheet

MICROWAVE MOISTURE SENSOR WITH WIRELESS DATA TRANSMISSION FL-WAPP

Liquid Level Switches

Size : Connection End : Min Temperature : Max Temperature : Materials : Cast iron body

DPS 100 DPS 100. Gauge- / Differential- / Absolute Pressure Transmitter

Magnetic Inductive Flow Sensor induq

TEK-F LEX 4100A. Guided Wave Radar Level Transmitter. LEVEL. Technology Solutions

Capacitive levels controllers

Model Sono-Trak Transit Time

Temperature Sensor for Build-In System G3/8" FLEXadapt

Measuring and Controlling Unit Multronic OC

Coriolis Massflowmeter

Technical specification FLUXUS F401 H2O

T10FS. Data Sheet. Torque Flange. Special features. Installation example T10FS. B en

6.00 INFRARED ANALYZER

RHE11. Hazardous Area Multifunction Coriolis Flow Transmitter. Features. Applications. Benefits

Serie NW - Acoustic Wave. Option list. Applications / Overview

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

Pharma Temperature Sensor G3/8"

NUBIS. R t o WATER MEASUREMENT

Transcription:

Product Information EN MaxxFlow HTC Flow measurement for dry bulk solids SWR engineering Messtechnik GmbH envea TM a trademark of Environnement S.A Group

MaxxFlow HTC MaxxFlow HTC MaxxFlow HTC Use The MaxxFlow HTC is specially developed for the flow measurement of dry bulk solids without any built-in in the flow. Due to its independence from angle of the pipework and its low installation height, the MaxxFlow HTC is especially suitable in situation in which it was previously only possible to use complicated or expensive mechanical solutions, e. g. impact plates or measuring chutes. The installation of the MaxxFlow HTC takes place independently of the direction of the line (vertical/inclined), but always after mechanical conveyor elements, e. g. rotary feeder, screw conveyors, air slides or chain conveyors. 120 t/h 80 t/h Installation after Rotary conveyor Installation after Screw conveyor Function After the conveyor, the material to be measured falls or slides through an inlet path and runs through the sensor. During the throughput, the MaxxFlow HTC records the material type and speed. Since the material falls from a constant height after emerging from the conveyor element, the speed of the product stream is accelerated, but is constant at the installation position of the sensor. Due to this constant speed, the speed measurement does not need to be activated in every case, but can be calculated as a constant depending on the height of the fall. The mass flow is determined as follows: h = constant v = 2 x g x h v = Speed g = Gravity acceleration h = Height of fall Q (kg/s) = K (kg/m 3 ) x v () x A (m 2 ) Ceramic inner pipe Through the input coupling of a high-frequency, electromagnetic alternating field, a homogenous measuring field is generated in the measuring tube. The measuring tube (interior tube of the sensor) consists of wear-resistant Al 2 O 3 ceramics. Dry bulk solids inside the measuring field reduce the amplitude of the alternating field. This leads to a measuring signal that is in proportion to the concentration of the dry bulk solids in the sensor (kg/m 3 ) If the material speed varies, for example due to a change in initial speed, then this can also be measured. This takes place through a runtime measurement with the help of two additional electrodes behind the interior ceramics tube. 2

Calibration The recording of the speed is in every case independent of the type of the dry bulk solids that are to be measured, since they can either be calculated or based on a runtime measurement. Thus the speed measurement requires no calibration. This makes a new and simple type of calibration possible: A material sample (approx. 10 liter = sensor volume) can be poured into the sensor via a filling nozzle in the inlet path above the sensor. There is a knife-gate below the sensor that is closed for the calibration. If the sensor is completely full, then the measured density value must correspond to the bulk density of the material. This bulk density (set point) can simply be determined as a gram per liter weight and entered in the evaluation unit (full calibration). Even with large product flows, the measurement is completely calibrated with a material sample of only about 10 liter. Material drop-tests using calibration points during a running process are thus a thing of the past. The handling of several tons of material as a reference therefore is no longer necessary. System A complete measuring station consists of these components: Measuring feeder for installation in the conveyor line Transmitter MFE 100 Sensor + Transmitter The measuring feeder is available in sizes DN 100 mm, DN 150 mm and DN 200 mm. On request the components inlet path (between conveyor element and sensor) and gate valve (for calibration) can also be configured and supplied. The transmitter is connected to the sensor using 4-wired, shielded cable. The maximum distance between measuring feeder and transmitter is 300 m. DN 100mm DN 150mm DN 200mm 3

Configuration 70,0 MaxxFlow DN 100 throughput at max. 50 % fill level in m 3 /h For the configuration of the sensors, the knowledge of the maximum volume flow is fundamentally important in order to configure the measuring point in such a way that the dry bulk solids can flow through the sensor unhindered and the product flow is not influenced. The diagrams in figure 1 to 3 show the maximum conveyable volume depending on material speed for the three sensor sizes DN 100, DN 150 mm and DN 200 mm. The sensor cross-section in both cases is filled to 50 %. m³/h Fig. 1 60,0 50,0 40,0 30,0 20,0 10,0 0,00 0 0,5 1 1,5 2 2,5 3 3,5 4 4,5 5 Example: If the max. mass flow amounts to 80 t/h, and if the material has a bulk density of 0.8 t/m 3, then the maximum volume flow amounts to 100 m 3 /h. When using a MaxxFlow HTC DN 150 mm, a speed of approx. 3 would be necessary. When using a MaxxFlow HTC DN 200 mm, a speed of approx. 1.7 will be required. Figure 4 shows that for these speeds to be reached, an inlet path of 0.5 respectively 0.15 m will be necessary between the conveyor and the sensor. m³/h 300.0 250.0 200.0 150.0 100.0 50.0 0.0 Fig. 2 MaxxFlow DN 150 throughput at max. 50 % fill level 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 Turndown and measuring range 300.0 250.0 MaxxFlow DN 200 throughput at max. 50 % fill level Which sensor type is being used depends on the available standing installation height and the turndown or the desired measuring range. The sensor itself has an installation height of 220 mm (DN 100 mm), 290 mm (DN 150 mm) or 310 mm (DN 200 mm). To equip the measuring point from the above example with a DN 150 mm sensor, an installation height of 500 mm + 290 mm = 790 mm would be necessary. The turndown, i.e. the ratio between the end and the beginning of the measuring range, amounts to 1:5 for all MaxxFlow HTC sensors. With regard to the example provided, it is possible to measure with a MaxxFlow HTC DN 150 mm between 20 m 3 /h and 100 m 3 /h. m³/h m 200.0 150.0 100.0 50.0 0.0 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 Fig. 3 Inlet path 1.2 1.0 0.8 0.6 0.4 0.2 0 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 Fig. 4 4

A MaxxFlow HTC DN 200 mm can also be used for this measuring task. Care must be taken that the angle for the contractions of the inlet path at a height of 0.15 m does not become too flat. The angle α (see figure 5) must be at least 60 in order to guarantee a flawless product flow. If the inlet path (height of fall) must be enlarged to maintain the demand after an angle > 60, then it must be noted that the lower limit of the measuring range moves upward. If for example an inlet path of 0.3 m would become necessary, then the speed would be approx. 2.5 and thus the maximum volume for a MaxxFlow HTC DN 200 mm 150 m 3 /h. With a turndown of 1:5, the lower measuring range limit would then be 30 m 3 /h. Figure 5 α 60 α p q Inlet path Advantages installation free of obstacles in the cross-section, therefore no dead spots independent of the arrangement of the line (vertical or inclined) simple refitting dust proof zero-contact measuring procedure (no mechanics) maximum material temperature 120 C pressure-resistant up to 10 bar on request abrasion-resistant ceramic inner pipe Dimensions DN 150 / 200 DN 100 F B 8 x Ø18 Grounding M8 in both flanges Ø100 C DN PD 8 x E Ø220 Ø180 A Flange EN 1092-1/ 01A/DNxxx/PN10/ DN 100 150 200 A 220 290 310 B 250 269 294 C 220 285 340 PD 180 240 295 E 18 22 22 F 31 36 23,5 5

Technical data Sensor Housing Steel St52, powder-coated (optional stainless steel 1.4571) NW 100/150/200 mm, Flange according EN 1092-1/PN10 Transmitter Power supply 110 / 240 V AC, 50 Hz, 24 V DC Power consumption 20 W / 24 VA Protection category IP 65 to EN 60 529/10.91 Inner pipe Ceramic Al 2 O 3 Operating temperature -10 +45 C [14 113 F] Protection category IP 65 Operating temperature Max. working pressure Working frequency Transmitting power Sensor pipe: -20 +120 C Sensor electronic: 0 + 60 C 1 bar, optional 10 bar 88 khz Max. 2 mw Dimensions 258 x 237 x 174 (W x H x D) Weight Approx. 2.5 kg Cable glands 3 x M16 (4.5-10 mm Ø) Terminal clamp wire size 0.2-2.5 mm 2 [AWG 24-14] Current output signal Alarm output Error output 2 x 4 20 ma (0 20 ma), load < 500 Ω Relay with toggle switch - max. 250 V AC, 1 A Weight Depending on model Data backup Flash memory Accuracy ± 3 % (based on end of measuring range and calibrated material) Impulse output RS 485 interface Open collector - max. 30 V, 20 ma ModBus SWR engineering Messtechnik GmbH Gutedelstraße 31 79418 Schliengen (Germany) Fon +49 7635 827248-0 Fax +49 7635 827248-48 www.swr-engineering.com envea TM a trademark of Environnement S.A Group (All rights reserved.) 6 EN 16/01/2018