A SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING. R.L. Baer and G.S. Kino. Edward L. Ginzton Laboratory Stanford University Stanford, CA 94305

Similar documents
Ginzton Laboratory, W. W. Hansen Laboratories of Physics Stanford University, Stanford, CA 94305

DETERMINATION OF FLAW DEPTH PROFILES FROM ACOUSTIC IMAGES. D.K. Peterson, S.D. Bennett and G.S. Kino

SURFACE ACOUSTIC WAVE STUDIES OF SURFACE CRACKS IN CERAMICS. A. Fahr, S. Johar, and M.K. Murthy

EFFECT OF SURFACE COATINGS ON GENERATION OF LASER BASED ULTRASOUND

ENHANCEMENT OF SYNTHETIC APERTURE FOCUSING TECHNIQUE (SAFT) BY ADVANCED SIGNAL PROCESSING

Standard Guide for Evaluating Characteristics of Ultrasonic Search Units 1

A SAM BEARING BALL INSPECTION SYSTEM. C-H. Chou, P. Parent, and B. T. Khuri-Yakub

DESIGN OF GLOBAL SAW RFID TAG DEVICES C. S. Hartmann, P. Brown, and J. Bellamy RF SAW, Inc., 900 Alpha Drive Ste 400, Richardson, TX, U.S.A.

Measurement of phase velocity dispersion curves and group velocities in a plate using leaky Lamb waves

NUMERICAL MODELING OF AIR-COUPLED ULTRASOUND WITH EFIT. D. E. Chimenti Center of Nondestructive Evaluation Iowa State University Ames, Iowa, USA

FATIGUE CRACK DETECTION IN METALLIC MEMBERS USING SPECTRAL

ULTRASONIC GUIDED WAVE ANNULAR ARRAY TRANSDUCERS FOR STRUCTURAL HEALTH MONITORING

Imaging using Ultrasound - I

A MEMS Transducer for Ultrasonic Flaw Detection

Liquid sensor probe using reflecting SH-SAW delay line

ACCURACY IMPROVEMENT ON NON-INVASIVE ULTRASONIC-DOPPLER FLOW MEASUREMENT BY UTILZING SHEAR WAVES IN METAL PIPE

FATIGUE CRACK CHARACTERIZATION IN CONDUCTING SHEETS BY NON

DETECTION OF CORROSION IN BOTTOM PLATES OF GAS AND OIL TANKS USING GUIDED ULTRASONIC WAVES AND ELECTROMAGNETIC ULTRASONIC (EMAT) TRANSDUCERS

REAL-TIME B-SCAN ULTRASONIC IMAGING USING A DIGITAL PHASED. Robert Dunki-Jacobs and Lewis Thomas General Electric Company Schenectady, New York, 12301

ACOUSTIC MICRO IMAGING ANALYSIS METHODS FOR 3D PACKAGES

ULTRASONIC SIGNAL CHARACTERIZATIONS OF FLAT-BOTTOM HOLES IN

CRACK DETECTION AND DEFECT CLASSIFICATION USING THE LLT - TECHNIQUE. Wolfgang Gebhardt and Friedhelm Walte

Chapter 4 Results. 4.1 Pattern recognition algorithm performance

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

Implementation of Orthogonal Frequency Coded SAW Devices Using Apodized Reflectors

Detectability of kissing bonds using the non-linear high frequency transmission technique

Application of Ultrasonic Guided Waves for Characterization of Defects in Pipeline of Nuclear Power Plants. Younho Cho

Effect of coupling conditions on ultrasonic echo parameters

SPARSE ARRAY TOMOGRAPHY SYSTEM FOR CORROSION EXTENT MONITORING H. Bian, H. Gao, J. Rose Pennsylvania State University, University Park, PA, USA

ULTRASONIC DETECTION OF CRACKS BELOW BOLTS IN AIRCRAFT SKINS

Micromachined ultrasonic transducers for air-coupled

Introduction To NDT. BY: Omid HEIDARY

Ultrasonic Transmission Characteristics of Continuous Casting Slab for Medium Carbon Steel

Equipment for Attenuation and velocity of ultrasound in solid state materials (transmission), experimental set-up

Miniature, High Efficiency Transducers For Use IN Ultrasonic Flow Meters

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

USE OF GUIDED WAVES FOR DETECTION OF INTERIOR FLAWS IN LAYERED

ISO INTERNATIONAL STANDARD. Non-destructive testing Acoustic emission inspection Secondary calibration of acoustic emission sensors

Attenuation and velocity of ultrasound in solid state materials (transmission)

MEASUREMENT OF SURFACE ACOUSTIC WAVE USING AIR COUPLED TRANSDUCER AND LASER DOPPLER VIBROMETER

Crack Detection with Wireless Inductively-Coupled Transducers

Factors Affecting Ultrasonic Waves Interacting with Fatigue Cracks

GUIDED WAVES FOR DAMAGE MONITORING IN PLATES FOR NOTCH DEFECTS

12/26/2017. Alberto Ardon M.D.

G. A. Alers and D. T. MacLauchlan

Module 2 WAVE PROPAGATION (Lectures 7 to 9)

IMAGING OF DEFECTS IN CONCRETE COMPONENTS WITH NON-CONTACT ULTRASONIC TESTING W. Hillger, DLR and Ing. Büro Dr. Hillger, Braunschweig, Germany

THE ANALYSIS OF ADHESIVE BONDS USING ELECfROMAGNETIC

Structural Integrity Monitoring using Guided Ultrasonic Waves

Ultrasonic Imaging of Tight Crack Surfaces by Backscattered Transverse Wave with a Focused Transducer

Experimental and theoretical investigation of edge waves propagation and scattering in a thick plate with surface-breaking crack-like defect

A. Fahr* Ontario Research Foundation, Mississauga, Ontario, Canada. ItV.R. Sturrock Defence Research Establishment, Pacific, Victoria B.C.

Characterization of High Q Spherical Resonators

DYNAMIC ANALYSIS OF CMUTs IN DIFFERENT REGIMES OF OPERATION

ASSESSMENT OF WALL-THINNING IN CARBON STEEL PIPE BY USING LASER-GENERATED GUIDED WAVE

Lamb Wave Ultrasonic Stylus

Further Developments in Ultrasonic Phased Array Inspection of Aging Aircraft

A SELF-COMPENSATING TECHNIQUE FüR THE CHARACTERIZA TION OF A

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล

Acknowledgment. Process of Atmospheric Radiation. Atmospheric Transmittance. Microwaves used by Radar GMAT Principles of Remote Sensing

New Multi-Technology In-Line Inspection Tool For The Quantitative Wall Thickness Measurement Of Gas Pipelines

NONDESTRUCTIVE EVALUATION OF CLOSED CRACKS USING AN ULTRASONIC TRANSIT TIMING METHOD J. Takatsubo 1, H. Tsuda 1, B. Wang 1

LASER GENERATION AND DETECTION OF SURFACE ACOUSTIC WAVES

Design and Analysis of Resonant Leaky-mode Broadband Reflectors

BEHAVIORAL STUDY OF RAYLEIGH, LAMB, AND CRITICALLY REFRACTED LONGITUDINAL WAVES FOR CRACK DETECTION IN PAINTED AND RUSTED STEEL STRUCTURAL COMPONENTS

ULTRASONIC FIELD RECONSTRUCTION FROM OPTICAL INTERFEROMETRIC

Use of Lamb Waves High Modes in Weld Testing

Lesson 02: Sound Wave Production. This lesson contains 24 slides plus 11 multiple-choice questions.

ACOUSTIC EMISSION MEASUREMENTS ON SHELL STRUCTURES WITH DIRECTLY ATTACHED PIEZO-CERAMIC

Passive Polymer. Figure 1 (a) and (b). Diagram of a 1-3 composite (left) and a 2-2 composite (right).

NARROWBAND ULTRASONIC SPECTROSCOPY FOR NDE OF LAYERED STRUCTURES T. Stepinski and M. Jonsson 1 Uppsala University, Uppsala, Sweden

Performance of UT Creeping Waves in Crack Sizing

PHYS2090 OPTICAL PHYSICS Laboratory Microwaves

1112. Dimensional evaluation of metal discontinuities by geometrical parameters of their patterns on imaging flaw detector monitor

ULTRASONIC IMAGING of COPPER MATERIAL USING HARMONIC COMPONENTS

Theory and Applications of Frequency Domain Laser Ultrasonics

DETECTION OF LEAKY-RAYLEIGH WA YES AT AIR-SOLID INTERFACES BY

CHARACTERIZATION OF PIEZOELECTRICS USING LINE-FOCUS TRANSDUCER

First Observation of Stimulated Coherent Transition Radiation

Standard Practice for Ultrasonic Examination of Turbine and Generator Steel Rotor Forgings 1

A Breakthrough in Sputtering Target Inspections: Ultra-High Speed Phased Array Scanning with Volume Focusing

ISO INTERNATIONAL STANDARD. Non-destructive testing Ultrasonic thickness measurement

Multi-Element Synthetic Transmit Aperture Method in Medical Ultrasound Imaging Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki and Marcin Lewandowski

EVOLUTION OF THE CRYOGENIC EDDY CURRENT MICROPROBE

STUDY ON SAW ATTENUATION OF PMMA USING LASER ULTRASONIC

B.R. Tittmann, R.S. Linebarger and R.C. Addison, Jr.

Investigation of Woven Fiber Reinforced Laminated Composites Using a Through Transmission Ultrasonic Technique

2. Pulsed Acoustic Microscopy and Picosecond Ultrasonics

1. Introduction. 2. Concept. reflector. transduce r. node. Kraftmessung an verschiedenen Fluiden in akustischen Feldern

Development of Automatic Neural Network Classifier of Defects Detected by Ultrasonic Means

ECHO-CANCELLATION IN A SINGLE-TRANSDUCER ULTRASONIC IMAGING SYSTEM

Advanced Ultrasonic Imaging for Automotive Spot Weld Quality Testing

Measurement of the loss due to grooved bottom structure intended for use as a backing in Capacitive Micromachined Ultrasonic Transducers

(A) 2f (B) 2 f (C) f ( D) 2 (E) 2

G. Hughes Department of Mechanical Engineering University College London Torrington Place London, WClE 7JE, United Kingdom

Air Coupled Ultrasonic Inspection of Steel Rubber Interface

High contrast air-coupled acoustic imaging with zero group velocity Lamb modes

Time Reversal FEM Modelling in Thin Aluminium Plates for Defects Detection

DAMAGE DETECTION IN PLATE STRUCTURES USING SPARSE ULTRASONIC TRANSDUCER ARRAYS AND ACOUSTIC WAVEFIELD IMAGING

From concert halls to noise barriers : attenuation from interference gratings

Transcription:

A SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING R.L. Baer and G.S. Kino Edward L. Ginzton Laboratory Stanford University Stanford, CA 94305 INTRODUCTION In this paper we describe a contacting shear wave array which we have used to characterize flaws in aluminum samples. By using shear waves instead of longitudinal waves. this array has doubled the resolution of our real-time synthetic aperture imaging system. The images we have obtained are significantly better than those obtained with longitudinal or surface wave arrays. The array is built on a buffer rod of the same material as the test specimen. Consequently there is no refraction, and aberrations are not introduced at the interface. Coupling loss, which is typically large with shear wave transducers. is negligible in this array. Off-normal wave incidence at the buffer rod-specimen interface is utilized to couple shear waves from the array to the specimen through longitudinal waves in the fluid coup1ant. The geometry of the array causes reflections from the end of the buffer rod and the back of the specimen to be eliminated from the image. This simplifies the task of isolating the reflections of flaws from the larger reflections of surfaces. THEORY AND DESIGN We have developed a contacting shear wave arrayl for use with our real-time digital synthetic aperture imaging system. 2 The images obtained with this array have been used to characterize flaws in aluminum samples. The location, width, and in some cases depth, of cracks have been clearly indicated. In all cases. the image of the crack has been free from obscuring front-wall and back-wall echoes. Currently, we are developing techniques to estimate crack size from echo data obtained with this array. 1669

1670 R. L. BAER AND G. S. KINO o Q ~ SIGNAL MEMORY SAMPLE TRANSDUCER BUFFER ROD \ J Fig. 1. CONTACTING ARRAY Relevant features of the digital synthetic aperture imaging system. The relevant features of our synthetic aperture imaging system are presented in Fig. 1. The system pulses each of the 32 array elements individually and then digitizes the received signal. These digitized signals are stored in an electronic memory and are used to synthesize an image based on the acoustic echoes. This type of imaging system places special requirements on the array. The digitization rate of the AID converter limits the operating frequency. In order to achieve at least three samples per cycle, the center frequency of the array is limited to 3 MHz. Only one element is operated at a time; hence, the elements must be highly efficient. The impulse response must be narrow, as this determines the depth resolution of the image. In addition, the angular spectrum must be as broad as possible, allowing each element to illuminate the entire target field. These criteria imply that cross-coupling in the array must be minimized, as cross-coupling adversely affects the impulse response and angular spectrum of an array. In order to satisfy these requirements, our array is designed to be placed in direct contact with the target specimen. By doing so, we avoid the high losses and aberrations associated with immersion techniques. Direct contact permits the use of shear waves. The reduced wavelength of shear waves provides our array with twice the resolution of a comparable longitudinal wave array. This is achieved without having to raise the operating frequency, which has been shown to be limited. Since we are using direct contact, the array is constructed on a buffer rod. The acoustic impedance of the buffer rod heavily loads the array elements, providing high efficiency and a good impulse response. It also eliminates the need for a backing. The buff~r rod is composed of aluminum and the array elements of PZT-5H. This ceramic has a high electromechanical

SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING 1671 coupling constant for shear wave operation. The 'elements are separated by sawing to inhibit cross-coupling. The dimensions of the array for 3 MHz operation appear in Fig. 2. The maximum resolution obtainable at this frequency is about one wavelength, or 1 mm using shear waves in an aluminum test specimen. In order to provide a broad angular response, the individual elements are a half wavelength wide (0.5 mm). This width was also chosen so as to achieve an acceptable electrical input impedance of about 300 ohms. The spacing between elements is one wavelength. In their longest dimension, the array elements are 12 mm. This allows the array to focus through more than 5 cm of aluminum without significant beam diffraction in the transverse direction. Aberrations are caused by refraction at the buffer rod-specimen interface. Refraction occurs when the velocities of the buffer rod and specimen are dissimilar. Under these circumstances, the transit time between an array element and a point in the specimen is a function of relative position as well as spacing. The reconstruction algorithm used by the imaging system assumes that the transit time is only a function of spacing. Aberrations may be eliminated by choosing the buffer rod velocity to be the same as that of the specimen. For this reason we have chosen to use a buffer rod composed of the same material as the specimen, to provide aberration-free images. 12tl" DO D~ / I;----~ 0.4 mm L I Imm PZT - 5H ~ALUMINUM 0.5mm ARRAY SUBSTRATE ELEMENTS Fig. 2. Array dimensions.

1672 R. L. BAER AND G. S. KINO This solution to the problem of aberrations cannot be adopted in arrays designed for immersion, which suffer from aberrations at the fluid-specimen interface. In order to achieve high efficiency. losses at the buffer rodspecimen interface must be considered. Contact between the two is accomplished by using a thin layer of fluid couplant. Figure 3 plots the transmission through such a buffer rod-couplant-specimen interface as a function of incidence angle. In this case, an aluminum buffer rod and specimen are assumed, as well as.04 A of water couplant. The wave incidence is shear vertical, as used in our array design. The transmission is zero for normal incidence. This occurs because the shear wave has no component of particle motion normal to the surface. Even when viscous fluid couplants are used, the transmission is very low. Beyond 28, the longitudinal wave exceeds its critical angle and does not propagate in either the buffer rod or specimen. This eliminates a potential source of spurious echoes which would clutter the image. Finally, the transmission is near its maximum at 45 incidence. This is the value we have chosen for our array. For comparison, transmission through the same interface, with longitudinal wave incidence instead of shear, is plotted in Fig. 4. In this case, the transmission has its maximum at normal incidence. For all values of incidence angle other than 0. shear waves are excited. These waves would create additional echoes, and interfere with the imaging process. Hence, the comparable longitudinal wave array design is optimized at normal incidence, rather than 45 incidence. z o ~ 60 i Z <f~ ~!z 40 a: w w ~5 U On. n.~ 20 w > i= <f -I W a: LONGITUDINAL...----.. O~~~-=~~L- ~ L-~~ L-~~~ o 20 INCIDENCE ANGLE (degl Fig. 3. Transmission vs. incidence angle at the buffer rodspecimen interface: shear vertical incidence.

SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING 1673 ----- -- LONGITUDINAL --- ----- --- ~" o L ~~~~S~H:E:A:RJ:======~::::::~~'~~ o 20 40 INCIDENCE ANGLE (deg) Fig. 4. Transmission vs. incidence angle at the buffer rodspecimen interface: longitudinal incidence. The efficiency of the two optimized designs is compared in Fig. 5. Transmission loss through the interface is plotted as a function of couplant layer thickness. As the figure shows, the 45 SV incidence design is considerably more efficient than the normal incidence longitudinal design over a wide range of couplant thicknesses. At the couplant thickness we generally operate with 0.04 A; the shear wave design is 5 db more efficient in round-trip travel. In addition, it is 17 db more efficient than an immersion design, which would suffer an 11 db one-way loss at the fluid-specimen interface. The surprising conclusion, which applies to single transducers as well as arrays, is that the use of shear waves and off-normal incidence, with direct contact, is the most efficient technique for exciting an acoustic wave in a solid sample. The choice of 45 incidence provides several advantages other than efficiency, which are depicted in Fig. 6. Longitudinal waves are not excited, as they are beyond their critical angle. Shear waves which reflect off the interface between the buffer rod and specimen reflect away from the array at an angle of 90 and do not return to complicate the image. In the same fashion, waves reflecting off the back wall of the specimen continue to propagate in the specimen and do not return to the array. Such echoes typically overwhelm and obscure the reflections from flaws in arrays which utilize normal incidence. Forty-five degrees incidence makes the array a very sensitive detector of flaws which run normal to the back wall of the sample. These flaws combine with the back wall to form corner reflectors of large effective aperture to the incident acoustic wave. Echoes from these reflectors are much greater in magnitude than those emanating from crack tips (which have a smaller effective aperture). With normal incidence, only the crack tips are visible. Consequently an array utilizing normal incidence is much less sensitive to these sorts of cracks.

1674 R. L. BAER AND G. S. KINO 10 NORMAL INC IDENCE LONGITUDINAL WAVE o..j Z o ::ii z ~ Cl: I- 45 VERTICAL INCIDENCE SHEAR WAVE 0.1 FLUID LAYER THICKNESS (LONGITUDINAL WAVELENGTHS) Fig. 5. Transmission loss vs. layer thickness at the buffer rodspecimen interface. ARRAY OF PIEZOELECTR IC CERAMICS ALUMINUM ""--WEOGE Fig. 6. Array coupled to specimen.

SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING 1675 1000 MEASURED THEORY ".... FREQ.(MHz) 5.0-1000 Fig. 7. Electrical input impedance of a single array element. EXPERIMENTAL RESULTS The electrical input impedance of a single array element is compared to theory in Fig. 7. There is reasonable agreement around the 3 MHz resonant frequency. At 0.7 MHz there is a significant disparity due to acoustic cross-coupling. Cross-coupling manifests itself most strongly in the impulse response, which appears in Fig. 8. The theoretical impulse response is approximately four half-cycles long. The measured response parallels the theoretical response for the first four half-cycles, but extends for another five half-cycles beyond that and possesses a long tail which is 20 db below the amplitude of the main pulse. As a consequence of this extended impulse. response, our depth resolution is only half of what we predicted. In addition, the long tail.of the response obscures small features, such as crack tip echoes, which we wish to observe. As the elements are separated by sawing, and there is no acoustic backing, the only path available for cross-coupling is the buffer rod. Surface skimming SH waves propagate along the surface of the buffer rod, coupling the elements. The existence of this wave is allowed by the placement of the elements themselves, which serve as periodic mass 10ads. 3 These waves are strongly excited by the narrow array elements and decay slowly. We view the elimination of this cross-coupling mechanism as being the most important improvement which could be made on the array. We have obtained excellent images of EDM slots and cracks on the back side of aluminum samples using this array. These images were very easy to obtain. A few drops of fluid couplant were applied to the buffer rod, and the sample was placed on top. No pressure

1676 R. L. BAER AND G. S. KINO THEORY EXPERIMENT Fig. 8. Theoretical and measured impulse responses. was required, due to the high transmission at the interface. The images appeared as soon as the target entered the field of view of the array. Figure 9 is an image of an EDM slot 5 mm long and 2 mm deep. In this image, the vertical axis represents lateral position in the sample, and the horizontal axis represents depth. The depth increases to the right, starting from the right-hand side of the series of bars which appears in the image. Note the absence of back wall and interface reflections. Figure 10 is an image of a crack which was created with a three-point bending jig. The irregular shape of the crack is visible in the image. The intensity of the image varies with the crack depth, being most intense at the center. In some cases, we have been able to estimate crack depth based on the time difference between the crack tip echo and the crack base echo. Unfortunately, the tip echo is much weaker, often rendering it invisible. CONCLUSIONS Several important conclusions can be drawn from the design and operation of this array. The most significant, which applies to single transducers as well as arrays, is that shear waves may be

SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING 1677 Fig. 9. Image of an EDM slot. Fig. 10. Image of a crack.

1678 R. L. BAER AND G. S. KINO excited in solid samples with greater efficiency than is possible with longitudinal waves. This is accomplished by using off-normal wave incidence, and direct contact. A second conclusion which applies specifically to arrays is that aberrations may be eliminated by using direct contact, and choosing a buffer rod composed of the same material and specimen. The third conclusion is that 45 incidence increases sensitivity to cracks which run normal to the back surface of the array, and at the same time eliminates spurious echoes from the buffer rod-specimen interface, and the back wall of the specimen. Finally, we believe that we can double the depth resolution of this array by using techniques that we have developed to eliminate crosscoupling. This will increase the utility of this tool in quantitative NDE. ACKNOWLEDGEMENT This work was supported by the Department of Energy under Contract No. DE-AT03-81-ER10865. REFERENCES 1. R.L. Baer, A.R. Selfridge, B.T. Khuri-Yakub, G.S. Kino and J. Souquet, "Contacting transducers and transducer arrays for NDE," Proc. of the Ultrasonics Symposium, 1981. 2. G.S. Kino, D. Corl, S.D. Bennett and K. Peterson, "Real time synthetic aperture imaging system," 1980 Ultrasonics Symposium Proceedings. 3. B.A. Auld, J.J. Gagnepain and M. Tan, "Horizontal shear surface waves on corrugated surfaces," Elect. Lett. 12, 650-651, 1976.