IR Antibunching Measurements with id201 InGaAs Gated SPAD Detectors

Similar documents
TCSPC at Wavelengths from 900 nm to 1700 nm

PZ-FLIM-110. Piezo Scanning FLIM System. Based on bh s Megapixel FLIM Technology. Complete FLIM Microscopes FLIM Upgrades for Existing Microscopes

Non-Descanned FLIM Detection in Multiphoton Microscopes

DCS-120. Confocal Scanning FLIM Systems. Based on bh s Multidimensional Megapixel FLIM Technology

BDS-MM Family Picosecond Diode Lasers

Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN Detector

BDS-SM Family Picosecond Diode Lasers

DCS-120. Confocal Scanning FLIM Systems. Based on bh s Multidimensional Megapixel FLIM Technology

BDS-SM Family Picosecond Diode Lasers

Multiphoton FLIM with the Leica HyD RLD Detectors

Instrument response function. Left linear scale, right logarithmic scale. FWHM is 120 ps.

An 8-Channel Parallel Multispectral TCSPC FLIM System

TCSPC measurements with the InGaAs/InP Single- photon counter

Picosecond Light Sources

PCS-150 / PCI-200 High Speed Boxcar Modules

SHM-180 Eight Channel Sample & Hold Module

The DCS-120 Confocal Scanning FLIM System

Megapixel FLIM with bh TCSPC Modules

Becker & Hickl GmbH. Technology Leader in Photon Counting

NUV and Blue ps Diode Lasers

User Handbook. DPC Channel Photon Correlator

PML Channel Detector Head for Time-Correlated Single Photon Counting

Time-Correlated Single Photon Counting Systems

Time Correlated Single Photon Counting Systems

Multi-wavelength TCSPC lifetime imaging Wolfgang Becker a, Axel Bergmann a, Christoph Biskup b, Thomas Zimmer b, Nikolaj Klöcker c, Klaus Benndorf b

Boston Electronics Corporation 91 Boylston Street, Brookline MA USA (800) or (617) fax (617)

Lab4 Hanbury Brown and Twiss Setup. Photon Antibunching

High Performance Photon Counting. User Manual PML-16-C. 16 Channel Detector Head for Time-Correlated Single Photon Counting. Becker & Hickl GmbH

Boston Electronics Corporation 91 Boylston Street, Brookline MA USA (800) or (617) fax (617)

Time-Correlated Single Photon Counting

Supplementary Information:

Ultraviolet and Blue Picosecond Diode Laser Modules

OCF-401 Optical Constant Fraction Discriminator

Solea. Supercontinuum Laser. Applications

Supplemental Information

Photon Count. for Brainies.

3.003 Lab 3 Part A. Measurement of Speed of Light

InGaAs SPAD BIOMEDICAL APPLICATION INDUSTRIAL APPLICATION ASTRONOMY APPLICATION QUANTUM APPLICATION

Cavity QED with quantum dots in semiconductor microcavities

Supporting Online Material for

Redefining Measurement ID101 OEM Visible Photon Counter

Theoretical Approach. Why do we need ultra short technology?? INTRODUCTION:

A New Single-Photon Avalanche Diode in 90nm Standard CMOS Technology

Characterizing a single photon detector

InGaAs SPAD freerunning

TCSPC for FLIM and FRET in Microscopy

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Detectors for High-Speed Photon Counting

FPPO 1000 Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual

PoS(PhotoDet 2012)051

Becker & Hickl GmbH DCS-120. Confocal Scanning FLIM Systems. An Overview

PERFORMANCE OF PHOTODIGM S DBR SEMICONDUCTOR LASERS FOR PICOSECOND AND NANOSECOND PULSING APPLICATIONS

DeltaMyc. Fluorescence Lifetime Mapping Microscope. Affordable Fluorescence Lifetime Imaging Microscopy (FLIM)

Setting up High Gain Detector Electronics for TCSPC

The All New HarmoniXX Series. Wavelength Conversion for Ultrafast Lasers

High Power Supercontinuum Fiber Laser Series. Visible Power [W]

arxiv: v2 [quant-ph] 9 Jun 2009

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging

Nanowires for Quantum Optics

Direct Measurement of Optical Cross-talk in Silicon Photomultipliers Using Light Emission Microscopy

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

SUPPLEMENTARY INFORMATION DOI: /NPHOTON

MEGAFRAME: a fully integrated, timeresolved SPAD pixel array with microconcentrators

Direct Measurement of Optical Cross-talk in Silicon Photomultipliers Using Light Emission Microscopy

Simultaneous Phosphorescence and Fluorescence Lifetime Imaging by Multi-Dimensional TCSPC and Multi-Pulse Excitation

W32 ID Quantique: Entanglion Game and Quantum Optics Demo room DS 141

Nano-structured superconducting single-photon detector

Photon Counters SR430 5 ns multichannel scaler/averager

GFT1504 4/8/10 channel Delay Generator

Metrology for QKD an industrial quantum optical communication technology

MSA-200 MSA-300 MSA-1000 Ultrafast Photon Counters / Multiscalers

REU Student: Si (Athena) Pan Connecticut College Mentor: Dimitre Ouzounov Graduate Student Mentor: Heng Li Summer 2008

Measure the roll-off frequency of an acousto-optic modulator

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

DDG-210 Preliminary Manual Version A4

Single-Photon Counting Detectors for the Visible Range Between 300 and 1,000 nm

Spider Pulse Characterization

DIODE LASER SPECTROSCOPY (160309)

LABORATÓRIUMI GYAKORLAT SILLABUSZ SYLLABUS OF A PRACTICAL DEMONSTRATION. financed by the program

SPM Series Quick Start Experiment Guide Rev.1.0, May 2011

GFT Channel Digital Delay Generator

Heterodyne Interferometry with a Supercontinuum Local Oscillator. Pavel Gabor Vatican Observatory, 933 N Cherry Ave., Tucson AZ 85721, USA

APE Autocorrelator Product Family

Xiuliang Chen, E Wu, Guang Wu, and Heping Zeng*

3D light microscopy techniques

Instruction manual and data sheet ipca h

FLIM Protocol. 1. Install IRF sample on the microscope using the stage insert. IRF sample options include urea crystals or BBO crystal.

Fiber Laser Chirped Pulse Amplifier

Measuring the speed of light

Measuring Kinetics of Luminescence with TDS 744 oscilloscope

Dead-time optimized time-correlated photon counting instrument with synchronized, independent timing channels

Short-Pulse X-ray at the Advanced Photon Source Overview

Supplementary Figure S1: Schematic view of the confocal laser scanning STED microscope used for STED-RICS. For a detailed description of our

SiPMs for solar neutrino detector? J. Kaspar, 6/10/14

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by

Practical work no. 3: Confocal Live Cell Microscopy

Femtosecond to millisecond transient absorption spectroscopy: two lasers one experiment

Single Photon Sources: Nano-diamond Characterization and Ani-bunching Investigation

Supporting Material 1. Actual design in FPGA

PRELIMINARY. Specifications are at array temperature of -30 C and package ambient temperature of 23 C All values are typical

Transcription:

IR Antibunching Measurements with id201 GaAs Gated SPAD Detectors Abstract. Antibunching measurements with GaAs SPAD detectors are faced with the problems of high background count rate, afterpulsing, and the requirement to gate the SPADs. This application note describes how anti-bunching measurements can be performed by using pulsed excitation and gated detection. Possible problems of the principle are discussed and hints for the buildup of suitable experiments are given. General Principle Antibunching measurements based on GaAs avalanche photodiodes (SPADs) are used to verify single-photon emission from single quantum dots in the IR at 1300 nm and beyond. However, GaAs SPADs differ from detectors for the visible range in some important details: - Compared with detectors for the visible range, GaAs SPADs have extremely high dark-count rates. - GaAs SPADs have strong afterpulsing. When a photon was detected the diode is likely to produce a dark count in the next gate interval. This dark count produces additional afterpulses. If the total multiplication factor is larger than one instability occurs. - GaAs SPADs can therefore operated only in a gated mode. That means, the diode is turned on for a gate time of 2.5 to 100 ns. If a photon is detected within this time the diode delivers an put pulse. Even if no photon is detected in the gate interval the diode has to be turned off for at least 1µs until the next gate pulse is applied. - To further reduce the amount of afterpulsing the diode is usually turned off for a period of time longer than the gate period after the detection of a photon. This dead-time is adjusted to keep the afterpulsing intensity at an acceptable level for the particular experiment. The traditional Hanbury-Brown-Twiss setup for anti-bunching measurements is shown in Fig. 1, left. antibunching-id201.doc 1

Gate 1 MHz 20ns Beam Splitter start Start (CFD) Light Detector 2 Detector 1 Delay stop TCSPC 0 delay time Gate trigg. 20ns Delay 50 ns TCSPC Stop (SYNC Fig. 1: Left: Traditional Hanbury-Brown Twiss experiment for antibunching measurement. Right: With id201 gated GaAs SPADs. the experiment the sample is excited by a continuous laser. The light from the sample is split into two channels. Each part is detected by an individual single-photon detector, A and B. A TCSPC device measures the times between the detection events in channel A and channel B and builds up the photon distribution over these times. A single photon emitted by the sample can go only to detector A or to detector B. Therefore the distribution obtained from a single-photon emitter shows a dip at a time that corresponds to events that appear simultaneous in detector A and B. Obviously, the Hanbury-Brown-Twiss setup requires a continuously working detector. Therefore, it cannot be directly used in combination with gated GaAs SPADs. A possible remedy is shown in Fig. 1, right. Detector A is gated periodically on by its internal gate generator. The gate of detector B is synchronised with that of detector A via a connection from the gate clock put of detector A to the gate trigger of detector B. A TCSPC module receiving a start from detector A and a Stop from detector B would detect an anti-bunching curve over a correlation time interval equal to the gate time interval. Unfortunately, this setup works reasonably only for a correlation time interval shorter than 10 ns. For longer correlation times the high dark count rate of the SPAD results in a substantial background of random correlation events, and in a noticeable drop of the correlation curve with increasing correlation time. Moreover, there is crosstalk of the gate pulse into the detection probability, which, in turn, results in ringing in the correlation curve recorded. An example is shown in Fig. 2. Because of these problems, results obtained from the setup shown in Fig. 1, right, have not been published so far. Fig. 2: Correlation background form the setup shown in Fig. 1, right. Gate width 20 ns. 2 antibunching id201.doc

Gated Antibunching Measurements with Pulsed Excitation A solution to the problem of counting background is gated detection in combination with pulsed excitation. The setup is shown in Fig. 3. A picosecond or femtosecond pulsed laser must be used to excite the sample. This is usually a titanium-sapphire (Ti:Sa) laser. Due to the short laser pulse a single quantum dot can only be excited one time within the duration of the laser pulse. Consequently, it can also emit only a single photon for one laser pulse, no matter how high the laser pulse energy is. To allow the id201 to gate the photons excited by the laser the repetition rate has to be reduced to ab 1 MHz. A pulse picker is therefore needed in combination with the Ti:Sa laser. An electrical reference pulse is derived from the laser via a photodiode. The reference signal triggers the gates in both id201 SPADs. The gate time should be as short as possible to keep the background count rate low. However, it should be no shorter than 50% of the excited-state lifetime of the sample. The gate delays are adjusted so that the photons emitted by the sample are inside the gate, see timing diagram in Fig. 3, right. Please note that there is some delay in the optical system. The gate delays required may therefore be in the range of several ns or even 10 ns. The exact values depend on the length of the laser light path, the length of the reference cable, the length of the detection fibres, and the internal delays of the id201. Sample splitter Gate trigg. 5ns Start (CFD) Laser 1MHz Photodiode TCSPC Reference from laser Pulse Picker 80 MHz Reference from Laser (ps pulsed) Gate trigg. 5ns Delay 250 ns Stop (SYNC Gate Photons Laser Ti:Sa Timing Diagram Fig. 3: Gated antibunching measurement with ps or fs laser The put pulses of the id201 detectors are connected to the start and stop inputs of the TCSPC module [1]. The stop signal is delayed by ab 250 ns in order to shift the zero-correlation peak into the recordable time interval of the TCSPC module. 250 ns correspond to ab 50 m of 50-Ω cable. Any bh SPC (TCSPC) module or the bh DPC-230 photon correlator can be used. Please note that the start input of the SPC modules is marked CFD, the stop input SYNC. Please see [2] and [3] for details. The recommended SPC system parameters for gated anti-bunching measurement with a pulsed laser of 1 MHz repetition rate are shown in Fig. 4. The setup shown left is for alignment purpose. It uses the Oscilloscope Mode of the SPC or DPC module to display the result in intervals of 1 second. antibunching-id201.doc 3

The setup shown in Fig. 4, right, accumulates the photons until a count number of 65565 is reached in the highest channel or until the measurement is stopped by the operator. Fig. 4: Recommended SPC system parameters for gated anti-bunching measurements. Laser pulsed at 1 MHz. Left: Oscilloscope mode for alignment of experiment. Right: Single mode for accumulating the photons over a longer time. Typical results are shown in Fig. 5 and Fig. 6. The curve shown in Fig. 5 has been obtained from a sample that did not show any anti-bunching effect. The result of the measurement is a number of correlation peaks. The peak on the right is the correlation of the photons of detector A versus detector B for the same excitation pulse. Please note that the SPC module uses reversed start stop; consequently the time axis is reversed. Please see [2, 3] for details. The peak in the middle and the peak on the left is the correlation between the photons of different laser pulses. Fig. 5: Result for an uncorrelated signal from a sample with any antibunching effect For a sample that shows antibunching the peak on the right becomes smaller, the other peaks remain the same. A result is shown in Fig. 6. 4 antibunching id201.doc

Fig. 6: Result from a sample with anti-bunching. The peak at the right becomes smaller. Please note that the apparent peak amplitude may vary due to the distribution of the data points on the peaks. To evaluate the exact number of photons within the individual peaks, please enable the cursors and zoom into the peak of interest. Then click on the Trace Statistics button and check the number of photons in this peak, see Fig. 7. Fig. 7: Evaluating the number of photons in a correlation peak. Enable the cursors of the SPC main panel, zoom into the peak, and click on the Trace Statistics button It should be noted that antibunching results obtained from GaAs SPADs are by far not ideal. Even for a perfect single-photon emitter the zero-correlation peak (on the right in Fig. 6) is not zero because it contains a number of random correlation events form background pulses. Moreover, the size of the correlation peaks obtained from an uncorrelated signal is not necessarily constant. The sizes are influenced by afterpulsing and by the dead time of the SPAD. Please note that these effects depend on the count rate in the start and stop channel. Reference measurements should therefore be performed at similar start and stop rates as the antibunching measurement. Optical System should be noted that anti-bunching measurements require to excite and detect signals from single emitters, such as single molecules, nanoparticles or single quantum dots. The measurements can therefore only be done in microscope or a similar optical system. A confocal setup has to be used to confine the detection volume to a single particle. This requires perfect alignment of the optics. The antibunching-id201.doc 5

laser has to be focused exactly on the particle under investigation. The light emitted by the particle is detected through the same microscope lens. It is separated from the excitation light by a dichroic beamsplitter. After passing a filter the emission light is focused on the input of an optical fibre. The input of the fibre works as an optical pinhole. The fibre input musts exactly conjugate with the particle in the laser focus. Any misalignment in x, y, or z results in loss in signal intensity and in loss of the correlation signature. References 1. W. Becker, Advanced time-correlated single-photon counting techniques. Springer, Berlin, Heidelberg, New York, 2005 2. W. Becker, The bh TCSPC handbook. 3 rd edition, Becker & Hickl GmbH (2008), available on www.beckerhickl.com 3. Becker & Hickl GmbH, DPC-230 16 Channel Photon Correlator, available on www.becker-hickl.com 6 antibunching id201.doc