Cytometry tutorial: The impact of adjusting PMT voltages on spillover and compensation

Similar documents
Flow cytometer instrument set up

PAVING THE ROAD TO UTOPIA VIA INSTRUMENT STANDARDIZATION

Setup Procedure for Beckman Coulter CytoFLEX Flow Cytometer

BD FC Beads 7-Color Kit

capabilities today. Flexibility for tomorrow.

BD FC Beads 7-Color Kit

BD LSRFortessa X-20. Special Order Product. Designed for limited space and boundless potential

BD OneFlow Setup Beads

COMPENSATION: AN INTRODUCTION

Technical Bulletin. Guide for Using BD Cytometric Bead Array (CBA) Flex Sets with the BD Accuri C6 Flow Cytometer. Introduction

SP6800 Spectral Analyzer

quantiflash Calibration Light Source for Cytometry

Perform instrument startup and verification check following the manufacturer s recommendations.

Perform instrument startup and verification check following the manufacturer s recommendations.

Quick Guide. NucleoCounter NC-3000

Practical work no. 3: Confocal Live Cell Microscopy

Leica SPEII confocal microscope. Short Manual

Scanning and Image Processing -by Steve Clough

Průtokový cytometr CytoFLEX

ab Firefly Cytometer Setup Particles

BD LSRFortessa X-20. Special Order Product. Designed for limited space and boundless potential

GenePix Application Note

GenePix Application Note

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Pixel shift in fluorescence microscopy

Leica TCS SP8 Quick Start Guide

LSM 780 Confocal Microscope Standard Operation Protocol

ScanArray Overview. Principle of Operation. Instrument Components

Instructions for Howto Scan µarrays

a Beckman Coulter Life Sciences: White Paper

The Zeiss AiryScan System, Confocal Four.

flowbeads: Bead Normalisation in Flow Cytometry

Comparison of the Analysis Capabilities of Beckman Coulter MoFlo XDP and Becton Dickinson FACSAria I and II

Image Analysis for Fluorescence

Confocal Microscopy. Kristin Jensen

(Quantitative Imaging for) Colocalisation Analysis

Quality control of microarrays

SEPTEMBER VOL. 38, NO. 9 ELECTRONIC DEFENSE SIMULTANEOUS SIGNAL ERRORS IN WIDEBAND IFM RECEIVERS WIDE, WIDER, WIDEST SYNTHETIC APERTURE ANTENNAS

contents TABLE OF The SECOM platform Applications - sections Applications - whole cells Features Integrated workflow Automated overlay

Cross-Talk in the ACS WFC Detectors. II: Using GAIN=2 to Minimize the Effect

Last updated: May 2014 Y.DeGraaf

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Leica_Dye_Finder :53 Uhr Seite 6 Dye Finder LAS AF

Leica Sp5 II Confocal User Guide

LSM 710 Confocal Microscope Standard Operation Protocol

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

BD Trucount Controls IVD

Manual Rosetta Calibration v

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

In our previous lecture, we understood the vital parameters to be taken into consideration before data acquisition and scanning.

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons

Cell Biology and Bioimaging Core

Spatial intensity distribution analysis Matlab user guide

Image Capture TOTALLAB

Nikon Instruments Europe

Product Information. Introduction

Supplemental Method Information Zeiss LSM710

Guide to Confocal 5. Starting session

Perkin Elmer Envision 2100 Multilabel Reader

Trademarks Cytek, the Cytek logo, and all other trademarks are property of Cytek Biosciences Cytek

Multi-channel imaging cytometry with a single detector

TCSPC at Wavelengths from 900 nm to 1700 nm

Confocal Application Letter No. 13. Sequential Scan for Leica TCS NT/SP systems

Agilent Cary 610/620 FTIR microscopes and imaging systems RESOLUTION FOR EVERY APPLICATION

Zeiss 780 Training Notes

Zeiss 880 Training Notes Zen 2.3

Quick Start Guide. Leica SP5 X

Dr. Bob on Colocalization or MSL Experiments In Learning Colocalization Using Image J

Development of a spectrometry system Using lock-in amplification technique

WFC3 TV2 Testing: UVIS Shutter Stability and Accuracy

Leica TCS SP8 Quick Start Guide

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Automatic Locating the Centromere on Human Chromosome Pictures

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1

Confocal, hyperspectral, spinning disk

Title: Leica SP5 Confocal User Manual

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Reikan FoCal Fully Automatic Test Report

A Simple and Rapid Method for Determining the

Balanced Constant Current Excitation for RTD Sensor Measurements

Overview Full Featured Silicon Photomultiplier Module for OEM and Research Applications The is a solid state alternative to the Photomultiplier Tube (

Creating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II

Sensitivity Measurement and Compensation in Spectral Imaging

CytoFLEX Flow Cytometer Platform. Join the Resolution REVOLUTION

Illumination Correction tutorial

Bi Imaging. Multicolor Imaging: The Important Question of Co-Localization. Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

Operating Instructions for Zeiss LSM 510

Pupil Lumens and their impact on the choice of lighting

Microsoft Excel: Data Analysis & Graphing. College of Engineering Engineering Education Innovation Center

(Refer Slide Time: 00:10)

Multifluorescence The Crosstalk Problem and Its Solution

STEM Spectrum Imaging Tutorial

Robust, Highly-visible and Facile Bioconjugation Colloidal Crystal Beads for Bioassay

Enabling Legendary Discovery. LEGENDplex. LEGENDplex. Mul -Analyte Flow Assay Kit

IncuCyte ZOOM Fluorescent Processing Overview

Advanced Live Cell Imaging

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Instruction Manual. Mark Deimund, Zuyi (Jacky) Huang, Juergen Hahn

Transcription:

Cytometry tutorial: The impact of adjusting PMT voltages on spillover and compensation Thomas Myles Ashhurst 1,2,3, Adrian Lloyd Smith 3,4 1Viral Immunopathology Laboratory, Discipline of Pathology, Sydney Medical School, The University of Sydney 2Marie Bashir Institute for Infectious Disease and Biosecurity (MBI), The University of Sydney 3Ramaciotti Facility for Human Systems Biology (RFHSB), The Centenary Institute and The University of Sydney) 4Sydney Cytometry Facility, The Centenary Institute and The University of Sydney) Contact: thomas.ashhurst@sydney.edu.au, a.smith@centenary.org.au

Voltage adjustment to reduce spillover? Misconception: I need to adjust my voltage to lower the compensation required between two overlapping fluorophores. A B C R780: 800 V COMP: 64.5% R780: 700 V COMP: 24.6% Case study My APC comp bead (A) spills into the APC/Cy7 detector (R780) (B) The compensation value is 64.5% (B) My supervisor thinks this is too high and wants me to reduce the compensation value by lowering the voltage of R780 If I reduce the voltage from 800V to 700V or 600V I can reduce the compensation value from 64.5% (high) (B) to 24.6 (medium) (C) or 7.78% (low) (D) D R780: 600 V COMP: 7.78%

Refresher: what is compensation? Compensation is correcting fluorescence spillover Target detector (YG582) 40,362 PE + comp bead Compensation value (extracting PE from the PECF594 (YG610) detector): 6,824 / 40,362 x 100 = 16.9% This compensation value is dependent on voltages Comp(YG610) = Measure(YG610) - %Measure(YG582) 6,824 Non-target detector (YG610) Why is this spreading? Correct compensation: Median Fluorescence Intensity (MFI) of PE+ and PE- beads in the PECF594 (YG610) detector are equal Over-compensated 0% 2% 5% 10% 16.9% 17.5%

What is spreading error? Spreading error is what causes problems for resolving true positive populations due to spreading of the negative population. Diagrammatic example: PerCP/Cy5.5 (with marker A ) spreads into the BV711 and BV650 detectors, leading to a loss of resolution in both. Figure 1: BV711 is bright enough, and marker B is expressed highly enough that A+B+ cells can be resolved from A+B- cells. Figure 2: However, BV650 is not bright enough to resolve A+B+ cells from A+B- cells. Figure 1: resolvable Figure 2: not resolvable A+B- cells A+B+ cells A+B- cells A+B+ cells PerCP/Cy5.5 anti-a A-B+ cells PerCP/Cy5.5 anti-a A-B+ cells BV711 anti-b BV650 anti-b The process of correct compensation reveals (not creates) measurement error, known as spreading error (SE)

What is spreading error? For more information on spreading error, see the following publication Reference: Nguyen, R., Perfetto, S., Mahnke, Y. D., Chattopadhyay, P., Roederer, M. 2013. Quantifying spillover spreading for comparing instrument performance and aiding in multicolor panel design. Cytometry A, 83(3): 306-315

What is spreading error? The process of correct compensation reveals (not creates) measurement error, known as spreading error (SE) Uncompensated Compensated Low SE Medium SE High SE PE+ beads BV711+ beads BV711+ beads COMP: ~8% COMP: ~71.43% COMP: ~57.14% SE: 1.93 SE: 6.68 SE: 11.60 Example: Higher compensation values tend to correlate with higher spreading error values (i.e. more spillover, more the measurement error will spread) However, higher compensation values cannot predict higher spreading error values Example: BV711 into BV785 detector Comp = 71.43% SE = 6.68 BV711 into AF700 detector Comp = 57.14% SE = 11.60 Despite the higher compensation value, the SE is lower

Detector voltages What happens when we adjust the voltages for a given detector CS&T beads (neg, dim, and bright) run at various voltages FITC detector Electronic noise rsd: 53.7 A B CS&T beads Negative bead Dim bead Bright bead Separation ratio Dim bead:negative bead Separation ratio Bright bead:negative bead Each detector uses a photomultiplier (PMT) that amplifies any signal picked up in that detector. The voltage applied dictates the level of amplification At a specific voltage (~600 volts for the FITC detector in this example), fluorescence signal is a) detectible above background, and b) is consistently and proportionately separated Voltages above this value do not increase the separation of signal (A) (including dim-neg, dim-bright, bright-neg) Voltages below this value compromise the detection of low-end (dim) signal (B) This is true of any signal (target fluorophore or spillover fluorophore) appearing in the detector Voltage scan approach to determine minimum voltage: Perfetto, S. P., Ambrozak, D., Nguyen, R., Chattopadhyay, P. K., & Roederer, M. (2012). Quality assurance for polychromatic flow cytometry using a suite of calibration beads, 7(12), 2067 2079. doi:10.1038/nprot.2012.126i

APC/Cy7 and APC in the R780 detector Let s take a look at the performance of the APC/Cy7 detector (R780) 550 volts A B COMP beads were recorded, where the R780 voltage was increased in 50V increments from 350 V up to 800 V At a specific voltage (550 volts in this case), fluorescence signal is a) detectible above background, and b) is consistently separated for any voltages above this This is clearly seen with the APC/Cy7 comp bead in the R780 detector However, this is also true of contaminating signals, such as the uncompensated APC comp bead in the R780 detector The minimum voltage required for optimal separation of signal from noise is between 500-550 volts Voltages above the minimum for separation (550 volts) do not increase the separation of signal (A) Voltages below the minimum for separation (550 volts) compromise the detection of low-end (dim) signal (B) The separation of bright from dim signal occurs at lower voltages (~350 volts) This is true of any signal (target fluorophore or spillover fluorophore) appearing in the R780 detector In this case lowering the voltage below 500-550 volts would result in the loss of dim APC/Cy7 signal 550 volts A B Voltage scan approach to determine minimum voltage: Perfetto, S. P., Ambrozak, D., Nguyen, R., Chattopadhyay, P. K., & Roederer, M. (2012). Quality assurance for polychromatic flow cytometry using a suite of calibration beads, 7(12), 2067 2079. doi:10.1038/nprot.2012.126i

APC/Cy7 and APC in the R780 detector Looking at the uncompensated plots individually: above the 550 volt cut off the separation of the target fluorophore (APC/ Cy7) from the spillover fluorophore (APC) is fixed, even though the compensation value increases. This is also true for the separation of APC/Cy7 from the background. Voltage: 350 V Comp: 0.148% 400 V 0.38% 450 V 0.915% 500 V 1.99% 550 V 4.06% 600 V 7.78% 650 V 13.9% 700 V 24.6% 750 V 40.2% 800 V 64.5% APC R780: APC/Cy7 Consistent with the principles of our previous slide The minimum voltage required for optimal separation of signal from background is between 500-550 volts (APC/Cy7+ signal and APC+ signal above background) The separation of bright from dim signal occurs earlier (separation of APC+ from APC/Cy7+ signal is consistent above above 350V)

APC/Cy7 and APC in the R780 detector Looking at the compensated plots individually: above the 550 volt cut off the separation of the target fluorophore (APC/ Cy7) from the spread of the compensated spillover fluorophore (APC) is fixed, even though the compensation value increases. Voltage: 350 V Comp: 0.148% 400 V 0.38% 450 V 0.915% 500 V 1.99% 550 V 4.06% 600 V 7.78% 650 V 13.9% 700 V 24.6% 750 V 40.2% 800 V 64.5% APC R780: APC/Cy7 The minimum voltage required for optimal separation of signal from background is between 500-550 volts The separation of compensated APC+ (now in the negative/low region scale of the R780 detector) from APC/Cy7+ signal is consistent above above 550 V

APC/Cy7 and APC in the R780 detector Overlay of the compensated data: 350-800 volts in R780 d 500 400 800 750 700 650 600 550 450 350 V APC R780: APC/Cy7 To replay

So what does compensation >100% look like? Despite compensation values of ~140%, the spreading error, and resolution of target (UVLD) from nontarget (BUV395) signal is reasonable.

So how do I set my voltages? On-the-fly method for setting voltages (not a full optimisation procedure) Start with the CS&T assigned voltages (this will set an approximate minimum for separation of dim signal) Run through your full set of compensation beads/cells reduce voltages for any detector where the signal is off-scale/ in the non-linear range (typically reduce if the peak is above 10^5). Even if reducing the voltage compromises the resolution of dim signal, it is more important to have linear signals (signal on-scale), as it is absolutely required for correct compensation Run through the compensation controls again and check the spillover values. Voltages should not be modified to reduce these values, as this will not impact resolution Check each of your sample types (especially samples from any stimulation or inflammatory situation) to ensure signals are all on scale/in the linear range Record compensation controls, apply compensation, and record samples Full optimisation procedure: see the following resources Sydney Cytometry Facility optimisation and tracking (email thomas.ashhurst@sydney.edu.au or a.smith@centenary.org.au) BD Application settings using unstained cells (https://www.bdbiosciences.com/documents/ BD_FACSDiva_Stndrd_App_Setup_TechBulletin.pdf) VRC bead suite optimisation method (http://www.nature.com/nprot/journal/v7/n12/abs/nprot.2012.126.html)

Summary and application Summary: changing voltages does not impact resolution Users are often concerned by high compensation values, especially values above 100%, and want to reduce the voltage of the detector receiving spillover to reduce the compensation value However, decreasing/increase voltages will only modify the compensation value, and will not change the impact of spillover/spreading on the resolution of positive signal PMT voltages: properly optimised Each detector will have a minimum voltage where signal is consistently and proportionally detected above background Decreasing the voltage for a given PMT below this minimum, then resolution of dim populations will be lost. Increase the voltage for a given PMT above this gives no advantage to population resolution. Often the cause for abnormally high compensation values (values >100% in particular) are a result of improperly optimised PMT voltages, for example, such as unnecessarily high default voltages being set by the CS&T program in some detectors. Use a proper optimisation protocol (or at least an on-the-fly method) to optimise PMT voltages What do I do if I have spillover/spreading error causing a loss of resolution in my data Firstly, optimise your instrument If spillover/spreading error is causing a loss of resolution in your data on an optimised system, the solution is to modify your panel, not to modify the voltages