RAD 229: MRI Signals and Sequences

Similar documents
Background (~EE369B)

Pulse Sequences: Rapid Gradient Echo

H 2 O and fat imaging

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012

Pulse Sequence Design Made Easier

Pulse Sequence Design and Image Procedures

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 )

2015 Spin echoes and projection imaging

RF pulse design and the Small Tip Angle Approximation

System/Imaging Imperfections

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

EE469B: Assignment 1 Solutions

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

MAGNETIC RESONANCE IMAGING

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

k y 2k y,max k x 2k x,max

MRI: From Signal to Image

RF Pulse Toolkit: Application Specific Design

MR Basics: Module 6 Pulse Sequences

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

EE469B: Assignment 2 Solutions

EE469B: Assignment 4 Solutions

NMR Basics. Lecture 2

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Lecture 7: Basics of magnetic resonance imaging (MRI): one dimensional Fourier imaging

MRI SYSTEM COMPONENTS Module One

1 Introduction. 2 The basic principles of NMR

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93

RF Pulse Design. Multi-dimensional Excitation II. M229 Advanced Topics in MRI Kyung Sung, Ph.D Class Business

functional MRI: A primer

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21.

2014 M.S. Cohen all rights reserved

RF and Electronic Design Perspective on Ultra-High Field MRI systems

Improving high-field MRI using parallel excitation

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

Image Quality/Artifacts Frequency (MHz)

MRI Metal Artifact Reduction

3D-Printed Microstrip Resonators for 4.7T MRI. Saeed Javidmehr. A thesis submitted in partial fulfillment of the requirements for the degree of

HETERONUCLEAR IMAGING. Topics to be Discussed:

A Conceptual Tour of Pulsed NMR*

MR Basics: Module 8 Image Quality

Chapter 2. The Physics of Magnetic Resonance Imaging

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

High Field MRI: Technology, Applications, Safety, and Limitations

Weber State University Radiologic Technology 4603

Mapping the Flip Angle in Magnetic Resonance Imaging Using the Accelerated 3D Look-Locker Sequence

Steady-state sequences: Spoiled and balanced methods

2 Hardware for Magnetic Resonance Imaging

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

Magnetic Resonance Imaging (MRI)

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

Encoding of inductively measured k-space trajectories in MR raw data

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.

The Pulsed Resistive Low-Field MR Scanner

SIEMENS MAGNETOM Skyra syngo MR D13

Magnetic Resonance Imaging Principles, Methods, and Techniques

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations,

A. SPECIFIC AIMS: phase graph (EPG) algorithms to cover a wide range of MRI

Magnetic Resonance Imaging

INSTRUMENTATION FOR PARALLEL MAGNETIC RESONANCE IMAGING. A Dissertation DAVID GERALD BROWN

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

The promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb.

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

MRI Systems and Coil Technology

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Localization of microscale devices in vivo using addressable transmitters operated as magnetic spins

MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Your first NMR measurement

EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 3. x 2 + y 2 0

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION

Introduction to MR Hardware. RF Coils C M L C T. = g * B 0. Rotating magnetization produces alternating magnetic field

1.5T HIGH FIELD SMALL ANIMAL MRI

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode.

Radio Frequency Detector Arrays for High-Field Magnetic Resonance Imaging

Waveform Encoding - PCM. BY: Dr.AHMED ALKHAYYAT. Chapter Two

Nuclear Magnetic Resonance (NMR)

Challenges of Field Inhomogeneities and a Method for Compensation. Angela Lynn Styczynski Snyder. Michael Garwood, Ph.D., Adviser

Design and Implementation of Calculated Readout by Spectral Parallelism (CRISP) in Magnetic Resonance Imaging (MRI)

ELT Receiver Architectures and Signal Processing Fall Mandatory homework exercises

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Pulse Code Modulation

ISMRM weekend educational course, MR Systems Engineering, Console Electronics

The development of the RF-pulse for the low level SAR used by the MRI.

PULSED/CW NUCLEAR MAGNETIC RESONANCE

BOLD fmri: signal source, data acquisition, and interpretation

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

A Complete Digital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla)

Signals and Systems Lecture 9 Communication Systems Frequency-Division Multiplexing and Frequency Modulation (FM)

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

Transcription:

RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229

Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences and naming: gradient-echo spiral T2* BOLD many, many confusing acronyms Expand EE369B, Complement EE369C, EE469B 2

General Course Logistics website: web.stanford.edu/class/rad229 3 Units, Letter or Cr/No Cr (EE300 Equivalent) Mon/Wed 1:30am-2:50pm CCSR 4107 (see calendar for changes) Texts (NOT required, but useful) Bernstein M. Nishimura D. 3

Prerequisites / Grading Prerequisite: EE369B /equivalent (complements EE369C / EE469B) Paper / Matlab assignments / no MRI scanning Grading: 10% Attendance / Participation 10% Midterm 50% Homework + Project 30% Final Auditing: Please participate, but allow for-credit students to do so first 4

Homework / Project Options Replace a HW Question: Spend <10min explaining how you d do a question Replace it with a problem and solution that you choose, related to recent lectures Project: (details to follow) Approximately 1-2 Homeworks Simulate and present a sequence / signals / recon A sequence we didn t cover or simulate A novel sequence that you devise A sequence/recon with EE369C 5

Lectures 75 min lectures -- Notes online at website PDF, whole slide (print 4-6 per page) Try to keep numbered. Read ahead, but try not to ruin suspense(!) Please no email, texting etc in class I try to stay on time - please help by being on time Come early, I will try to entertain with questions etc! Class participation: questions, exercises 6

Homework Due Wednesday 11:59pm, (minus 10% per day late) Paper: Lucas Center Rm P260 (under door) Frank Chavez (nearest cubicle) Electronically as PDF (encouraged): Email w/ subject RAD229: HW1 or similar, <10MB please! bah@stanford.edu Purpose is to learn the material. Note honor code Please do not share solutions without permission 7

Other Information Instructor: Brian Hargreaves Office Hours - See Calendar Other Lecturers: Jennifer McNab, Others? No Teaching Assistant Web Site: web.stanford.edu/class/rad229 Lecture notes, homework assignments, code Schedule / Room info, Announcements 8

Working Together - Rules Follow Honor Code Work together on homeworks, Discuss freely, but write your own matlab code Use resources, but not solutions No discussion of exams with others In general your responsibility is to learn! You should be able to explain anything you submit 9

Participation! FB: 1/10 (willing to answer, but not likely to be correct)? (unlikely to answer but likely to be correct) Balance??!! 10

Introductions Your name? Who do you work with? Your Research? Comments - What you Hope to Learn? 11

Course Overview / Topics Review of Basic MRI (EE369B) Signal Calculation Tools, System Imperfections Pulse Sequences Advanced Acquisition Methods The RAD229 class will continue to evolve! Things might change, and your input will shape the course! You may know more than me about some topics 12

Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

MRI: Basic Concepts N Static Magnetic Field (B0) 1H S B0 B0 B1 Excitation Precession (Reception) Relaxation (Recovery) Gradients (Relative Precession) 14

Precession and Relaxation Relaxation and precession are independent. Magnetization returns exponentially to equilibrium: Longitudinal recovery time constant is T 1 Transverse decay time constant is T 2 Precession Decay Recovery 15

Magnetic Resonance Imaging (MRI) Polarization Excitation Signal Reception Relaxation 16

MRI Hardware Strong Static Field (B0) ~ 0.5-7.0T Radio-frequency (RF) field (B1) ~ 0.1uT Transmit, often built-in Receive, often many coils Gradients (Gx, Gy, Gz) ~ 50-80 mt/m 17

B0: Static Magnetic Field Goal: Strong AND Homogeneous magnetic field Typically 0.3 to 7.0 T Resonance prοportional to B0 : γ/2π = 42.58 MHz/T Superconducting magnetic fields - always on ~1000 turns, 700 A of current Passively shimmed by adjusting coil locations The following increase with with B0: Polarization, Larmor Frequency, Spectral separation, T1 RF power for given B1 B0 variations due to susceptibility, chemical shift 18

B0: The Rotating Coordinate Frame Usually demodulate by Larmor frequency to baseband Also called the rotating frame 19

B1 + : RF Transmit Field Goal: Homogeneous rotating magnetic field Typically up to about 25 ut (Amplifier, SAR limits) Requires varying power based on subject size Dielectric effects cause B1 + variations at higher B0 Amplifier power: kw to tens of kw Specific Absorption Rate (SAR) Limits: Power proportional to B0 2 and B1 2 Goal is to limit heating to <1 C 20

B1 - : RF Receive Goal: High sensitivity, spatially limited, low noise Birdcage coils Uniform B1 - but single channel Surface coils Varying B1 - but high sensitivity Coil arrays Multiple channels with Varying B1 - Allows some spatial localization: Parallel Imaging 21

RF Coils 22

Receiver System 500 to 1000 k samples/s Complex sampling Low-pass filter capability Typically 32-128 channels Time-varying frequency and phase modulation (Typically single-channel) 23

Gradients Goal: Strong, switchable, linear Bz variation with x,y,z Peak amplitude ~ 50-80 mt/m (~ 200A) Switching 200 mt/m/ms (~1500 V) Limits: Amplifier power, heating, coil heating db/dt limitation due to peripheral nerve stimulation Switching induces Eddy Currents Concomitant terms (Bx and By variations) Non-linearities (often correctable) 24

Gradient Waveforms Mapping of position to frequency, slope = γg Typically waveforms are trapezoidal Constant amplitude and slew-rate limits Frequency G read Amplitude Time Position 25

Shims Goal usually to make B0 more uniform with subject Center frequency Linear shims (Up to ~1% offset to gradients) Higher-order (HO) shims (Spherical Harmonics) Shim arrays, Shim+RF (Current Research) Usually HO shims not dynamically switchable 26

Review Questions Which field is the receive field? B1 - Which field is always on? B0 What receive bandwidth corresponds to 500,000 samples/second? ±250 khz Why might small surface coils (or arrays) be useful? High sensitivity, Low noise, Spatially limiting 27

Image Formation and k-space Gradients and phase Signal equation Sampling / Aliasing Parallel Imaging Many reconstruction methods in EE369C 28

Gradient Strength and Sign Positive Gradient x Negative Gradient Double Strength Half-Duration x x Can control both amplitude and duration 29

Gradient Along Both x and y y Gx Gy x Can also vary along z 30

Ribbon Analogy x Gradients induce phase twist Twist has a number of cycles and a sign Twist can be along any direction 31

Gradients and Phase Control gradient amplitude and duration Can control frequency: Frequency = γ(gx x + Gy y) Can encode phase over duration t Angle = γt (Gx x + Gy y + Gz z) Z Z Generally: = (x G x dt + y G y dt) = (x Z Z G x dt + y G y dt) What are the units of Frequency and Angle (φ) here? 32

Signal Equations For a single spin: = (x Z G x dt + y Z G y dt) Represent as exponential: s = e i (x R G x dt+y R G y dt) Sum over many spins: Signal equation: s = Z 1 Z 1 1 1 s = Z 1 Z 1 1 1 (x, y)e 2 i(k xx+k y y) dxdy (x, y)e i (x R G x dt+y R G y dt) dxdy k x,y (t) = 2 Z t s(t) =FT[ (x, y)] kx (t),k y (t) 0 G x,y ( )d 33

Fourier Transform in MRI s(t) =FT[ (x, y)] kx (t),k y (t) M(k) Fourier Transform ρ(r) Given M(k) at enough k locations, we can find ρ(r) It does not matter how we got to k! What are the units of kx(t) and ky(t)? 34

Fourier Encoding and Reconstruction Encoding k y x Sum over image k x Reconstruction k y Gradient-induced Phase k-space k x x Sum over k-space k-space Spatial Harmonic 35

k-space: Spatial Frequency Map k y k-space k x In terms of pixel-width, what is the width of k-space? 36

Image Formation and Sampling Readout Gradient time Phase-Encode Gradient time k read k phase 37 k-space Readout Direction

k space Extent and Image Resolution Data Acquisition k space Image Space Fourier Transform x =1/(2k max ) 38

Sampling and Field of View Sampling density determines FOV Sparse sampling results in aliasing Phase-Encode FOV =1/ k y Readout FOV FOV k read k read k phase k phase 39

Phase-Encoding with Two Coils k y k-space k x 40

Readout Parameters Bandwidth linked to readout half-bandwidth (GE) = 0.5 x sample rate Same as Filter bandwidth (baseband) Pixel-bandwidth often useful Frequency G read Bandwidth per Pixel Full Bandwidth BW pix = G read x BW half = G read FOV/2 FOV Pixel Position 41

Imaging Example Desired Image Parameters: 256 x 256, over 25cm FOV (±)125 khz bandwidth What are the... Sampling period? Readout duration? Gradient strength? Bandwidth per pixel? k-space extent? 1/(2*125kHz) = 4µs 4µs * 256 = 1ms 250kHz / 0.25m / 42.58kHz/mT 23 mt/m (2.3 G/cm) 250kHz/256 pix = ~ 1kHz/pixel 0.5 / 1mm = 0.5 mm -1 = 5cm -1 42

2D Multislice vs 3D Slab Imaging 2D 3D Shorter scan times, reduced motion artifact Continuous coverage Thinner slices, reformats 43

Imaging Summary Gradients impose time-varying linear phase k-space is time-integral of gradients k-space samples Fourier Transform to/from image Density of k-space <> FOV (image extent) Extent of k-space <> Resolution (image density) 3D k-space is possible Parallel imaging uses coils to extend FOV 44

Excitation General principles of excitation Selective Excitation with gradients Relationships for slice excitation Excitation k-space Much more covered in EE469B 45

Excitation: B1 Field Direction of B1 is perpendicular to B0 Magnetization precesses about B1 Turn on and off B1 to tip magnetization Problem: We can t turn off B0! Precession still around B0 46

Excitation Magnetization precesses about net field (B 0 +B 1 ) B 1 << B 0 Must tune B 1 frequency to Larmor frequency B 1 B 0 Magnetization Static B 1 Field Rotating B 1 Field 47

Excitation: Rotating Frame Excite spins out of their equilibrium state. B 1 << B 0 Transverse RF field (B 1 ) rotates at γb 0 about z-axis. B 1 B 0 Magnetization Static Frame Rotating Frame, On resonance 48

Position Selective Excitation Slope = 1 γ G Frequency 49

Selective Excitation Position Slope = 1 γ G Larmor Frequency Slice width = BWRF / γgz Slice center = Frequency / γgz Magnitude RF Amplitude = + + B 1 Frequency Time 50

Excitation Example Given a 2 khz RF pulse bandwidth, and desired 5mm thick slice Slices at -2cm, 0, 2cm What are the... Gradient strength? (γ/2π)gz Excitation frequencies? 2kHz/5mm = 400kHz/m ~ 9.4 mt/m (0.94 G/cm) BW/slice = 2kHz/5mm, so -8, 0, 8 khz (9.4/50)*5mm ~1mm slice Thinnest slice possible with 50mT/m max gradients? 51

Excitation k-space Excitation k-space goes backwards from end of RF/ gradient pair: k e (t) = 2 Z T t Excited profile = Fourier Transform of excitation k-space Central flip angle = area under pulse (may be zero!): = G( )d Z B 1 ( )d k r (t) = 2 Z t 0 G( )d 52

Excitation Example For a 1ms, constant RF pulse of amplitude 10µT What is the flip angle? (42.58 khz/mt)(0.01mt)(1ms) = 0.4258 cycles = 153º How does RF energy change if the duration is halved and amplitude doubled? Doubles - (2A) 2 (T/2) = 2(A 2 T) 53

Signals and Contrast Simple Bloch Equation Solutions Basic contrast mechanisms: T1, T2, IR, Steady-State T2-Weighted T2-w FLAIR T1-w FLAIR Gradient Echo Diffusion-Weighted Apparent Diffusion Coefficient (ADC) 54

Signals and Contrast Bloch Equation Solutions (Relaxation): M xy (t) =M xy (0)e t/t 2 M z (t) =M 0 +[M z (0) M 0 ]e t/t 1 M z M0 0 M0 time Rotations due to excitation: Mxy 0 time M 0 xy = M xy cos + M z sin M 0 z = M z cos M xy sin 55

Echo Time (TE): T2 weighting 90º 90º RF 1 TE = Time from RF to echo Signal M z 0 Short TE Long TE 56

T2 Contrast TE = 20 Signal TE = 40 TE = 60 TE = 80 Echo Time (ms) Dardzinski BJ, et al. Radiology, 205: 546-550, 1997. 57

Repetition Time (TR) : T 1 Weighting 90º 90º RF 90º 1 M z Signal M z 0 M xy Each excitation starts with reduced M z 58

T1-Weighted Spin Echo Short Repetition Long Repetition Signal Signal Bone Time Joint Fluid Time 59

Basic Contrast Question (TE, TR) Short TE Incomplete Recovery Minimal Decay T1 Weighting Short TR Full Recovery Minimal Decay Proton Density Weighting Long TR Long TE Incomplete Recovery Signal Decay Mixed Contrast (Not used much) Full Recovery Signal Decay T2 Weighting 60 Images Courtesy of Anne Sawyer

RF Inversion-Recovery TI 180º 180º 1 Signal 0-1 Fat suppression based on T 1 Short TI Inversion Recovery (STIR) 61

Long Inversion Time (TI) - FLAIR Long TI suppresses fluid signal 62

RF Signal Question TR TI 180º 90º 180º TE Inversion Recovery Sequence: TR = 1s, TI = 0.5s, TE=50ms What is the signal for T1=0.5s, T2=100ms? Signal (Mxy) decays to 0 At TI, Mz-M0 = 1.63 exp(-1) Mz does not fully recover Mz ~ 0.4 M0 exp(-1) = 037 T2 decay ~ exp(-0.5) ~ 0.6 Mz is 0.63 M0 before 180º Signal = 0.4 x 0.6 = 0.24 M0 63

Steady-State Sequences Repeated sequences always lead to a steady state Sometimes includes equilibrium (easier) Otherwise trace magnetization and solve equations Example: Small-tip, TE=0 TR TE M z (TE)=M z (TR) cos M z (TR)=M 0 +[M z (TE) M 0 ]e TR/T 1... Combining... M z (TR)=M 0 1 e TR/T 1 1 e TR/T 1 cos 64

Summary ~ Background I Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast 65