QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

Similar documents
QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss LSM 880 Protocol

LSM 800 Confocal Microscope Standard Operation Protocol

Zeiss Deconvolution Microscope: A Quick Guide

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Zeiss LSM 780 Protocol

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Widefield 1. Switching on

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

Guide to Confocal 5. Starting session

LSM 780 Confocal Microscope Standard Operation Protocol

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Leica SP8 TCS Users Manual

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Simplified Instructions: Olympus Widefield Microscope S1230

Olympus Digital Microscope Camera (DP70) checklist

Using Olympus dotslide for polarising microscopy

Olympus xcellence Software - basic user guide

Leica SP8 Resonant Confocal. Quick-Start Guide

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Zeiss Axio Imager.A1 manual

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

LSM 710 Confocal Microscope Standard Operation Protocol

AxioVision 4.5 Brightfield Image Capture Procedure

Microscopy from Carl Zeiss

Motorized Axio Observer Start-up instructions

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Leica SPEII confocal microscope. Short Manual

Title: Leica SP5 Confocal User Manual

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Zeiss 880 Training Notes Zen 2.3

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Zeiss AxioObserver with ApoTome

Title: Nikon A1R Confocal User Manual

Leica SP8 TCS Users Manual

Everest System / Slidebook Operating Procedures

Axioscan - Startup. 1. Turn on the Axioscan on (button to the left) and turn on the computer

Zeiss 780 Training Notes

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Focus Strategies with ZEN2. Sven Terclavers ZEISS Embedded 3D Imaging Specialist Harvard University,

Contents. Introduction

Quick Start Guide. Leica SP5 X

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Zeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005

Photography Protocol. Software: ZEN Blue 2012

Practical work no. 3: Confocal Live Cell Microscopy

Olympus IX71 Microscope and DP71 Camera Instructions

Things to check before start-up.

Nikon SIM-E & A1-R System

Nikon E800 Microscope. Operating Instructions

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Using Autofocus in NIS-Elements

Olympus Time-lapse Microscope Basic operation

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Nasmyth Ultraview Vox User Protocol

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

personal DELTAVISION (pdv)

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Simplified Instructions: Zeiss Brightfield Microscope S1000

Characterization Microscope Nikon LV150

Image-Pro Plus 7.0 Product Note

LSM 510 Training Notes

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Nikon E800 Operating Instructions.

Dante (Microscope) & Beatrice (Guide) Orth Lab

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Nikon TE300 Eclipse Wide-Field Microscope

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

LSM 510 Meta Training Notes

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

SOP: Polyvar Met Light Microscope

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

Leica TCS SP8 Quick Start Guide

Diskovery Spinning Disk Guide

OPERATING INSTRUCTIONS

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Center for Microscopy and Image Analysis Axio Scan.Z1 Operating Manual

CELL-IQ IMAGEN USER MANUAL

Cell Biology and Bioimaging Core

Transcription:

Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Observing Life As It Happens Startup procedure... 2 Viewing your sample... 2 Adjusting brightfield Köhler illumination... 3 Setting user-defined pre-sets... 3 Image Acquisition... 4 Saving... 5 Additional Methods... 5 Software Autofocus... 5 Z stack... 5 Tile and Positions... 6 Time series... 6 Heating & CO 2 units... 7 Shutdown procedure... 7 29 Nov 2017,<qsg-hwf1> p. 1/7

Startup procedure All switches on wall Control units on right-hand side from top to bottom Turn on Camera (right hand side below chamber) Start PC and login Run Zen2012 Acquisition blue and click on Zen Pro If you are the first user after switching on the stage will want to calibrate. Select the 5x objective on the TFT controller and calibrate Colibri controls Screen displays wavelengths and power To switch on/off LEDs: press down on respective channel wheel. These also control the power of each LED. TFT touchscreen control Can be used to: Select objective Switch on/off Transmitted light Switch on/off Reflected light Select reflector cube Control heating and CO2 Additional focussing control Other microscope controls Viewing your sample In the software select locate tab Select BF (brightfield) Turn on Transmitted light either in software or using the TFT touchscreen!!! IMPORTANT!!! Check that the lightpath to the eyepieces is open (slider under eyepieces on binocular head) 29 Nov 2017,<qsg-hwf1> p. 2/7

Adjusting brightfield Köhler illumination Select 10x objective either with the TFT touch screen or software you can move your specimen in xy-direction with the joystick, holding the button on top of the joystick makes the stage moving faster focus on your sample adjust brightness with large black wheel on front / bottom of microscope) make sure condenser iris is not completely closed (two buttons on left top of condenser turret) completely close field iris (top black wheel) (if the image turns completely black, reopen until you see some light, proceed to next step and close again after you have focussed the iris) focus iris (black condenser focus wheel) centre iris (2 silver Köhler screws) reopen field iris until the edge is just not visible anymore Setting user-defined pre-sets (for looking down the microscope) open Light Path tab select Settings Editor button Setup required light path Select save as and give it a name Go to configure on main window Add new button, rename it, and drag your new setting across to the Hardware setting Ref. Customise its appearance New button should be visible and ready to use Close configure 29 Nov 2017,<qsg-hwf1> p. 3/7

Image Acquisition Select the Acquisition tab (make sure show all tools checkbox is ticked) Select Smart Setup or load saved configuration (or select the reuse button from a previous image load image, reuse button appears below image window) Select camera Select required channels Choose appropriate method and click ok Set Exposure will give approximate recording settings Fine adjust using the Channels tab Select each Channel in turn adjusting LED power and exposure time (range indicator button is below image window) Additional camera setting are in Camera tab To capture image click Snap Settings can be saved for future use 29 Nov 2017,<qsg-hwf1> p. 4/7

Saving To select file location and auto naming goto Tools on the menu Next selct options Under the saving tab Set path to D:\users\yourname You can also select the auto-naming format Additional Methods Select check box for Z stack, Tile and Time Series. This will activate the respective tabs To capture images using these methods use the Start Experiment button Software Autofocus Select Focus Strategy tab Options will depend on what other functions are selected eg Z stack and Tile Absolute fixed Z position Software autofocus Definite focus Software autofocus as a reference for Definite focus Definite focus as a reference for Software autofocus Local focus Local focus for software autofocus Z stack Set First and Last positions Set interval Start experiment 29 Nov 2017,<qsg-hwf1> p. 5/7

Tile and Positions Setup tile or positions Start experiment (ask for more instruction) Time series Setup duration and cycles Start experiment For very fast acquisition single channel Method 1: (using the movie recorder under the locate tab) Under the locate tab: Set light path or select fluorochrome from button Set the camera settings as required Go Live and set exposure and focus Start movie to record, stop to end recording Method 2: (using acquisition tab) In smart setup o Select fluorochrome o Select Speed Set the camera settings as required Go Live and set exposure and focus Setup duration under Time Series and tick use Camera streaming if possible Run Start experiment 29 Nov 2017,<qsg-hwf1> p. 6/7

Heating & CO 2 units The heating & CO 2 units can be controlled either by the TFT touchpad or through the software. If it is required to pre-heat the chamber prior to a booking then: Switch on wall plugs Turn on Power supply unit 232 On the TFT Touchpad - Select Incubation - Select unit to control (H insert P small environmental stage insert, H Unit XL Chamber, CO 2 Small V lid for insert.) 1. Set required environmental conditions 2. Set to on Turn off Power supply unit 232 Leave left hand set of wall plugs on If using software, start pc, login, run software, select Incubator tab and set conditions (channel 1 is the stage insert, Channel 3 is the chamber). Logout and shutdown as above Shutdown procedure check if anyone is booked after you within 2 hours If nobody is booked within two hours: remove specimen clean oil objective lenses with lens tissue Important: close incubation chamber completely ( protects from dust) update time in Sharepoint secure your data (e.g. copy them to the server) shut down hardware in reverse order from startup If someone s booked within two hours: update usage in Sharepoint remove your samples clean objective lenses with fresh lens tissue and close incubation chamber clear up the desk save files onto the server or on a mobile harddrive log off 29 Nov 2017,<qsg-hwf1> p. 7/7