LEO RECORD Ei Intrinsically Safe Digital Manometer with Record Function for use in Hazardous Applications. Accuracy: 0,1 %FS

Similar documents
EV-120 PRESSURE INDICATOR OPERATING MANUAL

Pressure transmitters MBS 4201, MBS 4251, MBS 4701 and MBS 4751

Submersible pressure sensor For oils and fuels Model LF-1

Submersible pressure sensor For water and wastewater Model LW-1

ED701 General Industry Pressure Transmitter

Intrinsically Safe Compact Controller CTR 210i. Intrinsically Safe Bargraph Indicator. BGI 210i

... for hydraulic and general applications in hazardous areas

Ultrasonic level measuring device, non-contact

Pressure Transmitters and Pressure Transducers

LF Conductivity Measuring Probe

MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400

Field indicator for current loops with HART communication Models DIH50, DIH52

ICARUS ELECTRONIC GAS VOLUME CONVERTER

SALES SERVICE +358 (0)

Submersible pressure sensor For superior applications Model LF-1

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751

Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm

Transducer transmitter BILT 4 II 2(1) G. Technical Manual

Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters

Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted

MINI MCR-SL-UI-I-LP-NC

OEM radar transmitter, for aggressive media level measurement

Relative and absolute pressure transmitter type 528

IP26 Series Current to Pneumatic Converter

PRetrans Table of contents

Specifications. Specifications. Transfer Characteristics Accuracy at 20 C (68 F) Temperature Drift

Pressure transmitter with flameproof enclosure For applications in explosion-protected areas Models E-10 and E-11

2-WIRE UNIVERSAL TEMPERATURE TRANSMITTER

SEM210 SERIES. Programmable In-head Universal Temperature Transmitter INDEX SECTION CONTENTS PAGE NO.

DS 400 DS 400. Intelligent Electronic Pressure Switch Completely in Stainless Steel. piezoresistive stainless steel sensor

General Specifications

Technical Information Nivector FTC968, FTC968Z

Relative pressure transmitter type 522 for shipbuilding

Loop Powered Indicator

KSR-Level sensor KSR-Schwimmer-Magnetschalter KSR Contact Protection Relays

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure

General Specifications

Intrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments Model IS-20-H

KFS. Flap-type flow meter. Design and range of application

Intrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F, IS-21-F

OPTITEMP TT 10 C/R Technical Datasheet

Rev ECN Description Approval Date AA Release to Approvals RCS 10/9/12

Power supply. Calibration. Surface finishing. Seal. Process connection

Relative and absolute pressure transmitter type 528

TR 104, TR 104-Ex Rail mounted Temperature Transmitters

VTM Local Display Unit with Frequency and Analogue Output

D1061S INSTRUCTION MANUAL. RS422/RS485 Fieldbus Isolating Repeater Din-Rail Model D1061S. D RS422 / RS485 Fieldbus Isolating Repeater

FieldIT Temperature Transmitter, rail mounted TH102/TH102-Ex

Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters

Mobrey Ultrasonic. MSP422, MSP400RH, and MSP900GH Level Transmitters. Product Data Sheet February 2015 IP2045, Rev EA

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. appmeas.co.uk +44 (0)

MODEL: B6U-B B6-UNIT. [2] LCD DISPLAY 0: Without 1: With

BAR-ANT-N-N-EX. Antenna barrier for the hazardous area. Data sheet. 1 Description

Digital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version

MANUAL HYDROBAR. Manufactured by: Fax: H/US/Hydrobar/ /00 Klay Instruments Page 1

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

[ Rosemount 648 Wireless Temperature Transmitter. Rosemount 648 Wireless. Quick Installation Guide , Rev CA August 2011

MINI MCR-SL-UI-f(-SP)

DOC. NO. : C3200-E04A APPROVED. Smart PressureTransmitter for Gauge and Absolute Pressure Measurement MODEL APT3200

General Specifications

SINEAX TI 807 Passive DC signal isolator. SINEAX TI 807 Passive DC signal isolator

JUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter

System Components Liquiphant M (NAMUR)/PROFIBUS-PA interface FXA 164

Design and applications

SINEAX TI 807 Passive DC signal isolator

User manual. Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD

Indicomp 2, Indicomp 2-Ex. Indicator and alarm signalling unit 10/ EN

Isolators A3/1.

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F

User Manual mantracourt.com

Pressure transmitter for general industrial applications Model A-10

Rail mounted Temperature Transmitter TH102/TH102-Ex

Operating Manual * * Differential pressure transmitter. Table of Contents. 1 Safety guidelines. 1.1 General Information

SUNSTAR 传感与控制 TEL: FAX: Relative and absolute pressure transmitter type 528

MANUAL HYDROBAR. Manufactured by: 7991 CZ DWINGELOO 7990 AA DWINGELOO The Netherlands

KDS. All Metal Variable Area Flowmeter. for Liquids and Gases

Field indicator for current loops with HART communication Models DIH50, DIH52

SET Installation and Operating Instructions. Level switch for two sensors

Trimod Besta Level Switch type I 8 / IE9 8

Certification of EXPLOSION PROTECTED ELECTRICAL EQUIPMENT

Pressure transmitter for general industrial applications Model A-10

INSTRUCTION MANUAL. SIL 3 - SIL 2 Digital Output Loop/Bus Powered DIN-Rail Models D1040Q, D1041Q, D1042Q, D1043Q, D1040Q - D1041Q - D1042Q - D1043Q

3 Ex i Isolators. Temperature Transmitter with Output 0/ ma, without Limit Value Contact (Field Circuit Ex i) Type 9182/.0-5.

Flush pressure transmitter For viscous and solids-containing media Model S-11

JUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter

Operating Instructions for Pressure Sensor with Ceramic Sensor Element. Model: SEN-9601

Mid-West. Instrument. Model 140 Electrical Installation and Operating Instructions. Gauge Front ELECTRICAL

Pressure transmitter For general industrial applications Model A-10

Operating Instructions

E instruments. MAXiCAL. ATEX Calibration Lab Multifunction Pressure and Temperature ATEX Calibrators

Nivotester FTR325. Technical Information. 1-channel switch amplifier for the microwave barrier Soliwave FDR50/FQR50

D1035S INSTRUCTION MANUAL. Frequency-Pulse Isolating Repeater DIN Rail Model D1035S. D Frequency-Pulse Isolating Repeater

DPharp Differential Pressure and Pressure Transmitters

Nivotester FTR325. Technical Information. 1-channel switch amplifier for the Soliwave M microwave barrier

SINEAX B 811 Power Pack with additional Functions

Pressure Transmitters for Industrial Applications

RMA42. Technical Information

Positioner with HART Communication Type 3780

Pressure transmitter For general industrial applications Model A-10

D1072. SIL 2 Temperature Signal Converter, Duplicator, Adder/Subtractor DIN-Rail Models D1072S, D1072D. Front Panel and Features:

Transcription:

LEO RECORD Ei Actual Value Status Indication SELECT ENTER G1/4 II 2 G Ex ia IIC T4 PTB 05 ATEX 2012 X Intrinsically Safe Digital Manometer with Record Function for use in Hazardous Applications. Accuracy: 0,1 %FS

Description and Application Intrinsically safe manometer with record function for use in hazardous environments. The technical data of the digital manometer can be taken from the corresponding data sheet or from the agreed specifications. Turn-On and Functions LEO RECORD Ei has two operating keys. The left key (SELECT) serves to select the functions and the pressure units. The right key (ENTER) activates the selected function or pressure unit. Turn-on: Pressing the SELECT key turns the instrument on. It first shows the full-scale pressure range (top display) and the software version (bottom display). The instrument is then ready for use and indicates the actual pressure (top display) and the record status (bottom display). The instrument has the following functions: OFF: Turns off the instrument and ends an active record MANO: Releases the following functions: rec Strt: Starts a record with the actual configuration (set via the software) rec StoP: Ends an active record ZERO SEt: Sets a new Zero reference ZERO res: Sets the Zero to factory setting, followed by the unit selection: bar, mbar/hpa, kpa, MPa, PSI, kp/cm 2, (m)h 2 O Example: Setting a new Zero Reference: Turn on the instrument by shortly pressing the SELECT-key Wait for the instrument s measuring mode (approx. 3 seconds) Press the SELECT-key 2 times: MANO appears Press ENTER: rec Strt appears - 2 -

Press the SELECT-key 2 times: ZERO SEt appears Press ENTER: The new Zero reference is set. The instrument returns to the measuring mode. Notes - The functions and units can also be called up by keeping the SELECT-key depressed. Releasing the key enables the displayed function or unit to be activated with the ENTER-key. - If the selected function or unit is not activated within 5 seconds with the ENTER-key, LEO RECORD Ei returns to the measuring mode without changing any settings. - Turning LEO RECORD Ei on and off does not influence any of the previous settings. - In the measuring mode, the record status is always indicated on the bottom display (run, prep, end). - If a pressure can not be represented on the display, OFL (overflow) or UFL (underflow) appears on the display. - If the actual pressure goes beyond the measuring range, the last valid pressure value starts flashing on the display. - Both the PC-Software and the LEO RECORD Ei indicate the battery condition in only 2 ways: 0% (BatLow display) and 100%. - The internal clock of LEO RECORD Ei is also updated when the instrument is turned off. The clock is automatically upated by the PC-software (Logger 4.x) when writing a record configuration. - After a battery change, LEO RECORD Ei has to be reconfigured with the PC-Software (Logger 4.x). - If a record is active, the display of the pressure is updated according to the set record interval (at least every 5 seconds) - 3 -

S E L E C T KELLER Installation Screw the G1/4 male port of LEO RECORD Ei into the female pressure port and tighten using the lower hexagon of the transducer. Loosen the upper hexagon and rotate the LEO RECORD Ei to the desired position. Retighten. The face of LEO KELLER E N T E R RECORD Ei can be rotated through 355. This feature allows the instrument to be mounted in all possible positions; vertical, horizontal or upside down. Battery Change / Battery Life When the battery starts weakening, a low battery warning will appear (BAT LOW). Battery change (please end current records and turn off the instrument before changing the battery!): Open the instrument by turning the display ring beyond the limit stop. Disconnect the battery. Remove weak battery and insert new one. Close the instrument. The battery life is up to 5 years, depending on measuring cycle. The following battery type must be used in explosive atmospheres: 3,6 V Lithium battery, type SL-760 or Tadiran SL-760. Ranges / Calibration The factory setting of the Zero for sealed gauge instruments (i.e. Range in bar abs. is at 0 bar absolute and for reference pressure instruments (i.e. Range in bar rel at ambient pressure. Sealed gauge pressure instruments can be activated with ZERO SET at ambient pressure for referential pressure measurements. - 4 -

Interface (RS485) The interface is located on the back of the manometer. The actual data can be transferred to the PC via interface converter K103A (RS232) or K104A (USB), both available separately at KELLER. The corresponding software can be found on our website. The connection for the RS 485 interface may only be used outside the zone with a potentially explosive atmosphere. General Safety Instructions When installing and operating the digital manometer, attention should be paid to the corresponding national safety regulations and to the relative country regulations concerning the Ex-application. Only mount the digital manometer onto unpressurized systems. On pressure ranges > 30 bar, the pressure connections could show residual hydraulic oil. Please note that temperatures outside of 0 60 C could impair the readability of the display. Special Conditions for Safe Use The digital manometer LEO RECORD Ei is an intrinsically safe apparatus ; it can be operated in explosive atmospheres. The operating ambient temperature are included between -10 C and +60 C. The connection for the RS 485 interface may only be used outside the zone with a potentially explosive atmosphere. Due to the internal capacitances only a safe maximum voltage of Um = 5,7 V may be applied. The permissible capacitance for Ui = 5,7 V according to EN 50020:2002, Table A.2 is Ci = 50 µf. This guarantees that the capacitance limit for the basic voltage level is not exceeded on reintroducing the equipment to the hazardous zone. - 5 -

Declaration of Conformity Herewith we declare, that the following products Intrinsically Safe Digital Manometer LEO RECORD Ei meet the basic requirements, which are established in the guidelines of the European Community: Directive EMC 89/336/EEC Directive ATEX 94/9/CE As criteria, the following norms for this Intrinsically Safe Manometer LEO RECORD Ei are applied EN 61000-6-1 / -6-2: 2001 EN 61000-6-3 / -6-4: 2001 EN 60079-0: 2004 EN 50020: 2002 and the EC examination certificate is given under: PTB 05 ATEX 2012 X This declaration is given for the manufacturer Keller AG für Druckmesstechnik St. Gallerstrasse 119 CH-8404 Winterthur in full responsibility by Keller GmbH Schwarzwaldstrasse 17 D 79798 Jestetten Jestetten, 02.06.08 H. W. Keller Geschäftsführer / CEO (with legally effective signature) 06/2008 www.keller-druck.com KELLER AG St. Gallerstrasse 119 CH-8404 Winterthur Tel. 052 235 25 25 Fax 052 235 25 00 KELLER GmbH Schwarzwaldstrasse 17 D-79798 Jestetten Tel. 07745 9214 0 Fax 07745 9214 50