Zeiss AxioImager.Z2 Brightfield Protocol

Similar documents
Zeiss LSM 880 Protocol

Zeiss LSM 780 Protocol

Zeiss Axio Imager.A1 manual

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Photography Protocol. Software: ZEN Blue 2012

Zeiss Deconvolution Microscope: A Quick Guide

Nikon E800 Operating Instructions.

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Leica SP8 TCS Users Manual

Simplified Instructions: Olympus Widefield Microscope S1230

Leica SP8 TCS Users Manual

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Everest System / Slidebook Operating Procedures

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

LSM 780 Confocal Microscope Standard Operation Protocol

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

JEOL 6500 User Manual

Nasmyth Ultraview Vox User Protocol

Guide to Confocal 5. Starting session

LSM 710 Confocal Microscope Standard Operation Protocol

Simplified Instructions: Zeiss Brightfield Microscope S1000

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Zeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005

Using the Nikon TE2000 Inverted Microscope

Using the AmScope Microscope Cameras

Nikon E800 Operating Instructions.

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Figure 1. Oil-immersion objectives available for use with the Lionheart FX.

User Operation of JEOL 1200 EX II

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Things to check before start-up.

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Zeiss AxioObserver with ApoTome

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

Title: Nikon A1R Confocal User Manual

AxioVision 4.5 Brightfield Image Capture Procedure

Widefield 1. Switching on

Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope

Nikon E800 Microscope. Operating Instructions

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur

Nikon SIM-E & A1-R System

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

JEOL 6700 User Manual 05/18/2009

Zeiss 880 Training Notes Zen 2.3

Motorized Axio Observer Start-up instructions

Dante (Microscope) & Beatrice (Guide) Orth Lab

A BRIEF INTRODUCTION TO MICROSCOPY The two key properties of a microscope that allow you to see microbes are resolution and magnification.

LSM 800 Confocal Microscope Standard Operation Protocol

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Characterization Microscope Nikon LV150

SOP: Polyvar Met Light Microscope

Diskovery Spinning Disk Guide

Olympus IX71 Microscope and DP71 Camera Instructions

LSM 510 Training Notes

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Renishaw InVia Raman microscope

Olympus xcellence Software - basic user guide

RENISHAW INVIA RAMAN SPECTROMETER

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Olympus Digital Microscope Camera (DP70) checklist

Using Autofocus in NIS-Elements

Zoom Set Too Tight Zoom Set Correctly Zoom Set Too Wide

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Zeiss 780 Training Notes

Confocal Raman Microscopy (WITec Alpha 300R)

Olympus Time-lapse Microscope Basic operation

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

Nikon TE300 Eclipse Wide-Field Microscope

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4

OPERATING INSTRUCTIONS

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Microscopy from Carl Zeiss

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Using the Microscope for a NANSLO Remote Web-based Science Lab Activity

Marine Invertebrate Zoology Microscope Introduction

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

Instructions for Making On-Line Reservations for Microscopes in NB11-204

WITec Alpha 300R Quick Operation Summary October 2018

Leica DB LB Research microscope and Studo Lite Imaging software

GXCapture 8.1 Instruction Manual

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

LSM 510 Meta Training Notes

User manual for Nikon Elements software

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

Transcription:

Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time. The facility will allow for extenuating circumstances (cells dying, sick day, etc.) if you inform us in a timely fashion. Follow each step of the startup poster and wait for the microscope software to fully load. 2) Lens Cleaning Please clean all of the lenses (used and unused) before and after your session. Refer to the lens cleaning poster if you need any help recalling the rules and steps. 3) Microscope Control Take note of the microscope light path: A. Light source B. Direction through microscope C. Condenser assembly F D. Slide E. Lens F. Camera/ Eyepiece a. Note the locking positions of the sliders E D and remember to be delicate! The slider pulls glass. C B A Touch pad control Within the Home menu, select Microscope and then Turret (Stay within this window for the entire session and ignore other tabs) Objective Menu: always begin focusing using the 10x lens. (a) Take care to properly set your focal position with the lower power lens before going to a higher power lens. Proper focusing from low power to high power will help prevent the microscope from breaking your slide.

Reflector Menu: select your first filter channel To open the shutter, turn the TL illumination on the far right of the screen On (a) Remember to close the shutter as often as you open it to prevent rapid bleaching of your sample. (b) Note: The active option is always white. Always begin imaging with the 10x objective and take care for the following issues: Select and inspect each slide If it is dirty, gently clean with a Kim Wipes and/or cotton swab. You should do this with all your slides before you come. Load the slide securely into the stage clip. Focusing will become difficult when the slide is uneven. Joystick Control Use the joystick to find your region of interest Use the F1 button in the upper right to toggle between course and fine x and y control

Focus Control Course and fine focus control are located on both sides of the microscope. If you are focusing on the right side of the microscope, the fine focus is located INSIDE the coarse focus. You can adjust by using the finger positions within the coarse. Focus your sample and check all the channels before imaging. FINE 4) Software Control Open the Zen Software Select the Zen Pro imaging option Click Skip calibration Upon Zen s completed startup, please note the 3 divisions of the interface: Left: acquisition control Middle: Top: image preview and acquired image

Bottom: basic measurements and manipulations Right: image gallery Preview Window Control Panel Image Gallery Display Window 5) Your first image Select the H condenser filter and readjust light intensity using the LED dial Gently push both sliders all the way in to send the light to the eyepiece. Use the touchpad to: Select the 10x lens in the objective menu Select the DIC/TL filter channel and open the TL illumination Use the stage movement wand and focus knobs to focus and center your specimen on a region of interest. Setup Kohler illumination; you can refer to the poster for help. This step is critical and necessary for uniform field illumination in imaging and must be done every time you switch lenses. With one hand on focus and the other on the condenser iris for your desired level of brightness, contrast, and morphological detail. Higher Lens NA larger condenser aperture Lower Lens NA smaller condenser aperture

Use the touchpad RL illumination option to close the shutter. Pull the BOTTOM camera/eyepiece slider out to send light to the camera. Part 1 Part 2 1 3 4 2 Within the Zen software: Setup your experiment in the Control Panel: AVOID THE LOCATE TAB! Click the Acquisition tab. Select the Channels window.

Using the Live mode, focus on your sample using one of the following options: Focus using the computer keyboard Put your cursor over the image and hit Ctrl on the keyboard; use the mouse roller to fine tune the focus (to signify focus activation, a dialog box will appear). Focus using the microscope fine focus Click Set Exposure within the channels window. Click the Acquisition Mode window to set the white balance. Select either the auto option or picker option; find a white region of your sample and click using the picker option. Before White Balance After White Balance Click Stop (the same icon as Live ) When all exposure times are set, Click Snap to acquire the image. Before saving your image set your gamma = 1 using the display tab at the bottom of the display window.

Image Saving Options: Click the Save As button to save the image in your directory within the specified User Data folder. Note: All data is stored in the D: drive under the User Data folder. Be sure to save to the User Data folder and NOT the Users folder. Optional: Auto-save function Note: There is also an auto save function that must be enabled AND disabled before and after your session. You CANNOT export overlay images using this tool. Save raw.czi file Save tiff files Use the same parameters from the batch protocol to export your images correctly.

Exporting options: be sure that you select the following image options to generate the same image you acquired. (a) DO NOT select Original Data export (b) Select Apply Display Curve and Channel Color 6) Higher magnification and switching slides If you switch to the higher magnification oil immersion lenses from a dry immersion lens, go back to the microscope and enter the objective menu on the touchpad. Press the high power lens you want to use. The lens will then move to the correct objective and the stage will drop. The touchpad will prompt you to put immersion on the specimen. At this point, apply one drop of oil and click Done to raise the stage back up. If you switch from an oil lenses to a dry lens, after the touchpad prompts you to remove the immersion, take the slide out, wipe off the oil with a cotton swab. Replace the slide into the stage clip and click up arrow on the left to raise the stage back up. When you need to switch slides, Go back to the 10x and refocus the new sample If you are imaging a similar sample using the same objective lens, click the Load Position button located in the upper right hand corner of the touchpad. This will allow you to lower the stage, replace your slide with a new one, and restore the same focal plane.

7) Data Export Go to the Processing tab (located to the right of the Acquisition tab) Click Batch Under Batch Method select Image Export In the middle of the screen under Batch Processing, click Add and then select all the images you want to export. Highlight one image Under parameters, make the following selections File type: Tagged Image File.tiff Compression: None Unclick convert to 8 bit Check Use Full Set of Dimensions Unclick Create project folder unless you would like all images in separate folders Exporting options: be sure that you select the following image options to generate the same image you acquired. DO NOT select Original Data export Select Apply Display Curve and Channel Color 1 2 3 Underneath Batch Processing Click Copy Parameters Select all images. (Ctrl +Shift) Click Paste Parameters Click Run Selected

8) System Shutdown Back up all your data. Clean all the lenses. Check the microscope calendar to see when the next user has an appointment If the user comes within 2 hours, log off your account Otherwise, follow shutdown poster steps.