Keysight Technologies 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM. Data Sheet

Similar documents
DATA SHEET A Multimeter 5.5 Digit Dual Display, Benchtop DMM

Keysight Technologies 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM. Data Sheet

Agilent Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price. Data Sheet

Agilent Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price. Data Sheet

Agilent 34405A Multimeter

Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price

Keysight Technologies Simultaneous Measurements with a Digital Multimeter

Keysight Technologies 87405C 100 MHz to 18 GHz Preamplifier. Technical Overview

Keysight Technologies U1210 Series Handheld Clamp Meters

Keysight Technologies 8490G Coaxial Attenuators. Technical Overview

Keysight Technologies U3400 Series 4½ and 5½ Digit Digital Multimeters

Keysight Technologies

Keysight Technologies Measuring Low Current Consumption with a Digital Multimeter

Keysight E5063A ENA Vector Network Analyzer

Keysight Technologies N2792A/N2818A 200 MHz and N2793A/N2819A 800 MHz Differential Probes. Data Sheet

Keysight Technologies Migrating Balanced Measurements from the

Keysight Technologies P9400A/C Solid State PIN Diode Transfer Switches

Keysight Technologies Making Simpler DC Power Measurements with a Digital Multimeter

Keysight Technologies

Keysight E5063A ENA Series Network Analyzer

Keysight Technologies Migrating from the 4268A/4288A Capacitance Meter to the E4981A Capacitance Meter. Technical Overview

Keysight N9310A RF Signal Generator

Keysight Technologies Improving the Test Efficiency of MEMS Capacitive Sensors Using the E4980A Precision LCR Meter.

Keysight N9311X RF and Microwave Accessory Kit for Low-cost Handheld and Benchtop Solutions. Technical Overview

Keysight Technologies N1918A Power Analysis Manager and U2000 Series USB Power Sensors. Demo Guide

Keysight Technologies Improving Test Efficiency of MEMS Electrostatic Actuators Using the E4980A Precision LCR Meter.

Keysight Technologies Accurate NBTI Characterization Using Timing-on-the-fly Sampling Mode. Application Note

Keysight Technologies N4985A System Amplifiers

Keysight Technologies Making Field Effect Transistor Characterization Using SMU

Introduction. Part 1. Introduction...2

Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters

Keysight Technologies Differences in Application Between Power Dividers and Power Splitters. Application Note

Keysight Technologies 423B, 8470B, 8472B, 8473B/C Low Barrier Schottky Diode Detectors

Keysight 8474B/C/E Planar-Doped Barrier Diode Detectors 0.01 to 50 GHz. Data Sheet

Keysight Technologies E4980A Precision LCR Meter

Keysight Technologies Waveguide Power Sensors. Data Sheet

Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters

Keysight Technologies USB Preamplifiers

Keysight 34401A Multimeter

Keysight Technologies N432A Thermistor Power Meter. Data Sheet

Keysight Technologies N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes. Data Sheet

Keysight Technologies

Keysight Redefines 50 GHz Portability. Get a $30k Credit When You Move Up to FieldFox

Keysight Technologies 87405C 100 MHz to 18 GHz Preamplifier. Technical Overview

Keysight Technologies N9310A RF Signal Generator

Agilent U2741A USB Modular 5.5 Digits Digital Multimeter. Data Sheet

Keysight U3606B Multimeter I DC Power Supply

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview

Keysight Technologies Accurate Capacitance Characterization at the Wafer Level

Keysight Technologies N6850A Broadband Omnidirectional Antenna. Data Sheet

Keysight Technologies Using a Network and Impedance Analyzer to Evaluate MHz RFID Tags and Readers/Writers

Keysight M9485A PXIe Multiport Vector Network Analyzer

Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size

Keysight Technologies U1230 Series Handheld Digital Multimeters (DMMs) Data Sheet

Keysight Technologies How to Easily Create an Arbitrary Waveform Without Programming. Application Note

Keysight Technologies Power of Impedance Analyzer

Keysight Technologies MEMS On-wafer Evaluation in Mass Production

Keysight Technologies Direct Power MOSFET Capacitance Measurement at 3000 V

Keysight Technologies Active Differential Probes U1818A 100 khz to 7 GHz U1818B 100 khz to 12 GHz. Technical Overview

Keysight Technologies Precise Low Resistance Measurements Using the B2961A and 34420A

Agilent 34410A and 34411A Multimeters Setting the Standard for Next Generation Benchtop and System Testing

Keysight Technologies Make Better AC RMS Measurements with Your Digital Multimeter. Application Note

Keysight HMMC-1002 DC 50 GHz Variable Attenuator

Keysight Technologies N2750A/51A/52A InfiniiMode Differential Active Probes. Data Sheet

Keysight Technologies Precise Current Profile Measurements of Bluetooth Low Energy Devices using the CX3300. Application Brief

Keysight Technologies N4983A Multiplexer and Demultiplexer. Data Sheet

Keysight N2806A Calibration Pulse Generator The world s fastest differential pulse generator. Data Sheet

Keysight N9320B RF Spectrum Analyzer

Keysight Technologies Using an External Trigger to Generate Pulses with the B2960A

Keysight M940xA PXIe Optical Extenders for Instrumentation. Data Sheet

Keysight Technologies Using a Scope s Segmented Memory to Capture Signals More Efficiently. Application Note

Keysight Quickly Generate Power Transients for Testing Automotive Electronics. Application Note

Keysight Technologies Automated Receiver Sensitivity Measurements Using U8903B. Application Note

Keysight Technologies How to Read Your Power Supply s Data Sheet. Application Note

Keysight Technologies MATLAB Data Analysis Software Packages

Keysight DSOXT3FRA/DSOX4FRA/DSOX6FRA Frequency Response Analyzer (FRA) Option

The Keysight Technologies, Inc. U1730C Series handheld LCR meters allow you to measure at frequencies as high as 100 khz a capability typically found

Keysight Technologies N8700 Series System DC Power Supplies

HP 34401A Specifications 8

True RMS Bench Multimeter 5492B

Keysight Technologies U1730C Series Handheld LCR Meters Take your expectations higher with the latest LCR meters. Data Sheet

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview

Keysight Technologies Photodiode Test Using the Keysight B2980A Series

Keysight Technologies Network Analyzer Measurements: Filter and Amplifier Examples. Application Note

Keysight U1882B Measurement Application for Infiniium Oscilloscopes. Data Sheet

Keysight Technologies

Keysight Technologies Understanding the Importance of Maximum Power Point Tracking Efficiency for Solar Inverters.

Keysight N8836A PAM-4 Measurement Application For Infiniium S-Series, 90000A, V-Series, X-Series, Q-Series, and Z-Series Oscilloscopes

Keysight 6800C Series AC Power Source/Analyzers, VA

SDM3045X Digital Multimeter. DataSheet

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Keysight Technologies FFT and Pulsed RF Measurements with 3000T X-Series Oscilloscopes. Application Note

Keysight Technologies x1149 Boundary Scan Analyzer. Data Sheet

Keysight Technologies 89601B-SSA/89601BN-SSA Spectrum Analysis VSA Software

Keysight Technologies HMMC GHz High-Gain Amplifier

Keysight Technologies U9400A/C Solid State FET Transfer Switches

Keysight U9397A/C FET Solid State Switches (SPDT)

Keysight Technologies PXI Digital Multimeters 6½ Digit, High Performance (M9182A, M9183A)

Keysight Technologies Measuring Group Delay of Frequency Converters with Embedded Local Oscillators. Application Note

Keysight Technologies 34410A and 34411A Multimeters

Transcription:

Keysight Technologies 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet

02 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Features Fast reading speed of up to 190 readings/sec 0.015% DCV accuracy Multiple connectivity options USB 2.0, Serial Interface (RS-232) and GPIB 11 measurement functions; DC voltage & current, True RMS AC voltage & current, 2- and 4-wire resistance, frequency, continuity, diode test, capacitance and temperature Ultra-bright OLED with dual display capability Up to 50,000 memory points for data logging Built-in Histogram function With Fluke 45 and Fluke 8808A code compatibility Achieve throughput breakthrough Turbo charge your production line with the 34450A digital multimeter. With its fast speed of up to 190 readings per second you can now increase your manufacturing throughput tremendously. Get highly accurate, repeatable and trustworthy measurements with the 0.015% DCV accuracy designed to meet general industrial and educational needs. Figure 1. Bright OLED display for excellent viewing experience when making measurements.

03 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Intuitive and multipurpose device The 34450A is an intuitive and easy to use tool that provides a wide range of measurement functions such as DC voltage, DC current, True RMS AC voltage and AC current, 2-wire and 4-wire resistance, frequency, diode test, continuity, capacitance and temperature. Simplify your dayto-day task with up to 50,000 memory points, allowing you to capture and log up to 14 hours of data. Built-in with histogram and basic statistical functions, you can now do simple data analysis within the unit itself. With its ultra-bright OLED dual display, different measurements can be performed ensuring you get the right readings at first glance. Fast 190 readings/sec and 0.015% DCV accuracy to help you achieve throughput breakthrough 5.5 digit OLED dual display allows clear and quick view on measurements Input terminals More flexibility with multiple connectivity Multiple connectivity options such as USB 2.0, Serial Interface (RS-232) and GPIB port provides greater flexibility to connect DMM to a PC for data access and retrieval for analysis. With the option to connect to a PC, this enables the 34450A to work with Keysight Connectivity software and can be controlled remotely via SCPI commands or Command Expert. The IVI-COM driver is also included to ensure an easy integration with different programming environments. 11 measurement functions; DC voltage & current, True RMS AC voltage & current, frequency, continuity, diode test, capacitance and temperature Multiple choices of connectivity and trigger in/out capability for production data analysis Built-in math functions Up to 50,000 memory points for data logging and histogram function Kensington lock to better secure the unit Easy migration Upgrading your obsoleted DMM is made easy with minimal changes to test program. To ensure both forward and backward compatibility, the 34450A includes SCPI, Fluke 45 or Fluke 8808A commands. Experience a quick, easy and painless migration as it allows you to quickly transfer your existing test programs on the Keysight 34450A.

04 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Specifications DC specification Table 1. DC accuracy ± (% of reading + % of range) Function Range 1 Test current or burden voltage Input impedance Specifications are for 90 minutes warm-up time, slow mode, and calibration temperature within 18 C - 28 C 1 year 23 C ± 5 C Temperature coefficient 0 C - 18 C 28 C - 55 C DC voltage 100.000 mv 10 MΩ or > 10 GΩ 0.018 + 0.008 0.0020 + 0.0008 Resistance (2-wire) 2 Resistance (4-wire) 2 1.00000 V 10 MΩ or > 10 GΩ 0.015 + 0.005 0.0015 + 0.0008 10.0000 V 10 MΩ 0.015 + 0.005 0.0020 + 0.0008 100.000 V 10 MΩ 0.015 + 0.005 0.0020 + 0.0008 1000.00 V 10 MΩ 0.015 + 0.005 0.0020 + 0.0008 100.000 Ω 1 ma 0.050 + 0.008 0.0060 + 0.0008 1.00000 kω 500 μa 0.050 + 0.008 0.0060 + 0.0005 10.0000 kω 100 μa 0.050 + 0.005 0.0060 + 0.0005 100.000 kω 10 μa 0.050 + 0.005 0.0060 + 0.0005 1.00000 MΩ 1 μa 0.060 + 0.005 0.0060 + 0.0005 10.0000 MΩ 100 na 0.250 + 0.005 0.0250 + 0.0005 100.000 MΩ 100 na/10 MΩ 2.000 + 0.005 0.3000 + 0.0005 DC current 100.000 μa < 0.02 V 0.05 + 0.015 0.007 + 0.0015 1.00000 ma < 0.2 V 0.05 + 0.007 0.007 + 0.0010 10.0000 ma < 0.02 V 0.05 + 0.015 0.008 + 0.0015 100.000 ma < 0.2 V 0.05 + 0.007 0.008 + 0.0010 1.00000 A < 0.2 V 0.10 + 0.015 0.012 + 0.0015 10.0000 A < 0.6 V 0.25 + 0.007 0.015 + 0.0010 Continuity 3 1000 Ω 0.5 ma 0.05 + 0.03 0.005 + 0.005 Diode test 4 1.0000 V 0.5 ma 0.05 + 0.03 0.005 + 0.005 Notes on specifications: 1. 20% over range on all ranges except 1000 VDC and 10 A range. 2. Specifications are for 4-wire Ω or 2-wire Ω using NULL function. If without NULL function, add 0.2 Ω additional error. 3. Continuity thresholds is fixed at less than 10 Ω 4. Specifications are for the voltage measured at the input terminals only.

05 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet AC specifications Table 2. AC Accuracy ± (% of reading + % of range) Function Range 1 Frequency 1 year 23 C ± 5 C Specifications are for 90 minutes warm-up time, slow mode, and calibration temperature within 18 C - 28 C True RMS AC voltage 2 100.000 mv 20 Hz 45 Hz 1.0 + 0.1 0.02 + 0.02 45 Hz 10 khz 0.2 + 0.1 0.02 + 0.02 10 khz 30 khz 1.5 + 0.3 0.05 + 0.02 30 khz 100 khz 3 3.0 + 0.3 0.10 + 0.02 1.00000 V to 750.00 V 20 Hz 45 Hz 1.0 + 0.1 4 0.02 + 0.02 45 Hz 10 khz 0.2 + 0.1 0.02 + 0.02 10 khz 30 khz 1.5 + 0.3 0.05 + 0.02 30 khz 100 khz 3 3.0 + 0.3 5 0.10 + 0.02 True RMS AC current 2 10.0000 ma to 10.0000 A 20 Hz 45 Hz 1.5 + 0.1 0.02 + 0.02 45 Hz 1 khz 0.5 + 0.1 0.02 + 0.02 1 khz 10 khz 6 2.0 + 0.2 0.02 + 0.02 Temperature coefficient 0 C - 18 C 28 C - 55 C Notes on specifications: 1. 20% over range on all ranges except ACV 750 V and ACI 10 A 2. Specifications are for sine-wave inputs more than the 5 % of range except 750 V range. Input signal must be more than 50 Vrms for 750 V range. Maximum crest factor of 3 at full scale. Input impedance is 1 MΩ in parallel with capacitance less than 120 pf, AC couple with up to 400 DCV 3. Additional error to be added as frequency > 30 khz and signal input < 10% of range. 30 khz to 100 khz: 0.003% of full scale per khz 4. For input < 200 V rms 5. For input < 300 V rms 6. Frequencies > 5 khz are typical for all ranges Table 3. Frequency Accuracy ± (% of reading + 3 counts) Function Range 1 Frequency 1 year 23 C ± 5 C Frequency 100.000 mv to 750.00 V 1 20 Hz 300 khz 2 0.02 + 3 0.005 10.0000 ma to 10.0000 A 20 Hz 10 khz 3 0.02 + 3 0.005 Notes on specifications: 1. The frequency can be measured up to 1 MHz as 0.5 V signal to 100 mv/1 V ranges. 2. 10% of range to full scale input on all ranges except where noted. 100 mv range specifications are for full scale or greater inputs. For inputs from 10 mv to 100 mv, multiply total % of reading error by 10. 3. 10% of range to full scale input on all ranges except where noted. 10 ma range specifications are for full scale or greater inputs. For inputs from 1 ma to 10 ma, multiply total % of reading error by 10. Temperature coefficient 0 C - 18 C 28 C - 55 C Table 4. Frequency resolution Function Range Frequency Resolution Frequency 100.000 mv to 750.00 V 1 119.999 Hz 0.001 Hz 1.19999 khz 0.00001 khz 11.9999 khz 0.0001 khz 119.999 khz 0.001 khz 1.19999 MHz 0.00001 MHz 1. The frequency can be measured up to 1 MHz as 0.5 V signal to 100 mv/1 V ranges.

06 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Temperature and capacitance specifications Table 5. Temperature and capacitance accuracy ± (% of reading + % of range) Function Range 1 Probe type or test current 1 year 23 C ± 5 C Temperature 80.0 C to 150 C 5 kω thermistor probe Probe accuracy + 0.2 C 0.002 C 110.0 F to 300.0 F 5 kω thermistor probe Probe accuracy + 0.4 F 0.0036 F Capacitance 1.000 nf 100 na Notes on specifications: 1. 20% over range on all ranges. Specifications are for 90 minutes warm-up time, slow mode, and calibration temperature within 18 C - 28 C Temperature coefficient 0 C - 18 C 28 C - 55 C 10.00 nf 100 na 1 + 0.5 0.02 + 0.001 100.0 nf 1.0 μa 1 + 0.5 0.02 + 0.001 1.0000 μf 1.0 μa 1 + 0.5 0.02 + 0.001 10.000 μf 10 μa 1 + 0.5 0.02 + 0.001 100.00 μf 100 μa 1 + 0.5 0.02 + 0.001 1000.0 μf 0.5 ma 1 + 0.5 0.02 + 0.001 10,000 μf 1.0 ma 2 + 0.5 0.02 + 0.001

07 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Operating specifications Table 6. Operating specifications on single display (approximate) Function Resolution Function change (sec) 1 Range change (sec) 2 Auto range Reading rate/sec 4 (sec) 3 USB GPIB Serial ACV Slow (5.5) 2.6 2.5 4.6 1.9 1.9 1.9 Med (4.5) 1.2 1.2 1.5 19 19 19 Fast (4.5) 1.1 1.1 1.2 160 99 33 DCV Slow (5.5) 1.3 1.3 1.6 1.7 1.7 1.7 Med (4.5) 0.6 0.7 0.8 49 49 24 Fast (4.5) 0.6 0.7 0.7 190 117 34 2-wire Ω Slow (5.5) 1.2 1.3 1.6 1.4 1.4 1.4 Med (4.5) 0.4 0.5 0.6 49 49 24 Fast (4.5) 0.4 0.5 0.5 165 110 32 4-wire Ω Slow (5.5) 1.2 1.4 1.9 1 1 1 Med (4.5) 0.6 0.6 1.1 5.2 5.3 4.7 Fast (4.5) 0.6 0.6 1 5.9 5.9 5.3 Frequency 5 Slow (5.5) 2.1 2.1 2.6 0.9 0.9 0.9 Med (4.5) 1.2 1.2 1.7 0.9 0.9 0.9 Fast (4.5) ACI Slow (5.5) 2.6 2.6 6.2 1.9 1.9 1.9 Med (4.5) 1.2 1.2 1.7 19 19 33 Fast (4.5) 1.1 1.2 1.3 160 99 33 DCI Slow (5.5) 1.3 1.3 1.9 1.7 1.7 1.7 Med (4.5) 0.6 0.7 0.9 49 49 24 Fast (4.5) 0.6 0.7 0.7 190 116 36 Diode 4.5 0.1 190 117 38 Continuity 4.5 0.1 165 111 33 Temperature 4.5 0.5 4.2 4.2 3 Notes on specifications: 1. Time to change from 2-wire resistance to this specified function and to take at least one reading using SCPI FUNC and READ? commands. 2. Time to change from one range to the next higher range and to take at least one reading using SCPI FUNC and READ? commands. 3. Time to automatically change one range and to take at least one reading using SCPI CONF AUTO and READ? commands. 4. Number of measurements using SCPI READ? command when front panel display is off using DISP OFF command. 5. Reading rate depends on signal frequency 20 Hz.

08 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Supplemental measurement specifications Table 7. Supplemental measurement specifications Specification measurement Specification DC voltage Measuring method: Sigma delta A-to-D converter Input resistance: > 10 GΩ ± 2% range (Selectable 100 mv, 1 V ranges) 10 MΩ ± 2% range (typical) 1000 V on all ranges (HI terminal) Resistance Measurement method: 2-wire or 4-wire Ohms Open circuit voltage: Limited to < 2.8 V 1000 V on all ranges (HI terminal) DC current Shunt resistance: 1 Ω for 10 ma, 100 ma 90 Ω for 100 μa, 1 ma 0.01 Ω for 1 A, 10 A Externally accessible at rear panel 0.4 A, 500 V fuse for I terminal Internal 11 A, 1000 V fuse for 10 A terminal Continuity/diode test Measurement method: Uses 0.5 ma ± 0.2% constant current source Response time: Continuity: 165 samples/second with audible tone Diode: 190 samples/second with audible tone Continuity threshold: 10 Ω fixed 1000 V (HI terminal) Temperature Measurement method: 2-wire Ohms measurement of 5 kω thermistor sensor (YSI 4407) with computer conversion Auto-ranging measurement, no manual range selection Measurement noise rejection Input Protection: CMR (Common mode rejection) For 1 kω unbalance LO lead NMR (Normal mode rejection) For 60 Hz (50 Hz) ± 0.1% 1000 V (HI terminal) DC 140 db AC 70 db Slow mode 5½ digits 90 db Medium mode 4½ digits 55 db Fast mode 4½ digits 0 db AC voltage Measurement method: AC coupled true-rms measures the AC component with up to 400 VDC bias any range Crest factor: Input impedance: Maximum 3:1 at full scale 1 MΩ ± 2% in parallel with < 100 pf of all ranges 750 V rms on all ranges (HI terminal) AC current Measurement method: DC coupled to the fuse and current shunt, AC coupled true rms measurement (measures the AC component only) Shunt resistance: 1 Ω for 10 ma, 100 ma 0.01 Ω for 1 A, 10 A Externally accessible at rear panel 0.4 A, 500 V fuse for I terminal Internal 11 A, 1000 V fuse for 10 A terminal

09 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Table 7. Supplemental measurement specifications (continued) Specification measurement Specification Frequency Measurement method: Reciprocal counting technique. AC coupled input using AC voltage function. Math functions Data log Triggering and memory Trigger out Non-volatile memory Signal level: Gate time: Null, dbm, db, Min/Max/Avg, hold, limit test Info, list, histrogram Samples per trigger: Trigger delay: 3.3 V logic output Polarity: Pulse width: 50,000 readings 10% of range to full scale input on all ranges except where noted. Auto or manual range selection 0.1 second or 1 period of the input signal. 750 V rms on all ranges (HI terminal) 1 to 5,000 (typical), 1 to 50,000 (optional) 0 to 3600 sec (100 us step size) Negative pulse Approximately 3 μs Sample timer Range: Up to 3600 sec in 100 μs steps Remote interface Programming language USB 2.0 Standard, GPIB IEEE-488 (Optional) SCPI-1994.0, IEEE-488.2

10 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet General characteristics Table 8. General Characteristics Specification measurement Specification Power supply 100 V/120 V(127 V)/220 V(230 V)/240 V ± 10% AC line frequency 45 Hz - 66 Hz and (360 Hz - 440 Hz, 100/120 V operation) Automatically sensed at power -on Power consumption Operating environment 45 VA maximum, < 11 W average Full accuracy at 0 C to 55 C Full accuracy to 80% RH at 30 C (non-condensing) Altitude up to 3000 meters Storage compliance 40 C to 70 C Safety compliance IEC 61010-1:2001/EN 61010-1:2001 (2nd Edition) Canada: CAN/CSA-C22.2 No. 61010-1 -04 USA: ANSI/UL std No. 61010-1:2004 Measurement category CAT II, 300 V: CAT I 1000 Vdc, 750 Vac rms, 2500 Vpk transient over voltages, Pollution degree 2 EMC compliance Certified to IEC61326-1: 2005/EN61326-1:2006 CISPR 11:2003 / EN 55011:2007 Group 1 Class A Canada: ICES/NMB-001:Issue 4, June 2006 Australia/New Zealand: AS/NZS CISPR 11:2004 Shock and vibration Tested to IEC/EN 60086-2 Dimension (HxWxD) Weight Warm up time Rack: 88.5 mm x 212.6 mm x 272.3 mm Bench: 103.8 mm x 261.1 mm x 303.2 mm 3.75 kg (8.27 lb.) 90 minutes Dimensions Standard shipped accessories Test lead set Power cord USB interface cable Quick Start Guide Product Reference CD Keysight IO Library Suite CD-ROM 261.1 mm Upgradable options 3445GPBU - GPIB connectivity upgrade 3445MEMU - 50,000 memory points upgrade for data logging 103.8 mm 303.2 mm

11 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Keysight optional accessories: 34138A test lead set 34191A 2U dual flange kit: secures the instrument to the front of the rack. This kit can be used with the 34194A dual lock link kit to mount two half-width, 2U height instruments side-by side. E2308A thermistor temperature probe 82350B PCI high-performance GPIB interface card 10833A/B/C/D/F/G GPIB cable in various lengths (0.5 m, 1 m, 2 m, 4 m, 0.5 m, 6 m and 8 m) 34194A dual lock link kit: recommended for side-by-side combinations and includes links for instruments of different depths. This kit can be used with the 34191A 2U dual flange kit to mount two half-width, 2U height instruments side-by-side. 82357B USB/GPIB interface high-speed USB 2.0 34190A rackmount kit: designed for use with only one instrument, mounted on either the left or the right side of the rack.

12 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Evolving Since 1939 Our unique combination of hardware, software, services, and people can help you reach your next breakthrough. We are unlocking the future of technology. From Hewlett-Packard to Agilent to Keysight. For more information on Keysight Technologies products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 mykeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.keysight.com/find/emt_product_registration Register your products to get up-to-date product information and find warranty information. Keysight Services www.keysight.com/find/service Keysight Services can help from acquisition to renewal across your instrument s lifecycle. Our comprehensive service offerings onestop calibration, repair, asset management, technology refresh, consulting, training and more helps you improve product quality and lower costs. Keysight Assurance Plans www.keysight.com/find/assuranceplans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight s measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/34450a Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-9-7-17) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System This information is subject to change without notice. Keysight Technologies, 2014-2017 Published in the USA November 18, 2017 5991-1133EN www.keysight.com