Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications

Similar documents
Oscilloscope Calibration Options for 55XX Series Multi-Product Calibrators

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators

Fluke 5820A Oscilloscope Calibrator Specifications

5500A. Multi-Product Calibrator. Extended Specifications 2005

5520A. Multi-Product Calibrator. Extended Specifications 2005

5520A. Multi-Product Calibrator. Extended Specifications

5080A Multi-Product Calibrator

5820A. Service Manual. Oscilloscope Calibrator

5820A. Service Manual. Oscilloscope Calibrator

5522A Multi-Product Calibrator Extended specifications

5502A Multi-Product Calibrator. Extended specifications

Oscilloscope Calibrator

5790A Automated AC Measurement Standard

Fluke 8845A/8846A Digital Multimeters Extended Specifications

Fluke 83V and 87V Digital Multimeters

5520A. Getting Started. Multi-Product Calibrator

Model MHz Arbitrary Waveform Generator Specifications

280 Series Waveform Generators A selection of universal waveform generators offering superior performance and excellent value

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

5320A. Multifunction Electrical Tester Calibrator. Extended Specifications

1587/1577 Insulation Multimeters

G5100A: 50 MHz Arbitrary Function Generator

PXIe Contents. Required Software CALIBRATION PROCEDURE

Synthesized Function Generators with Versatile Arbitrary Waveform Capabilities. Synthesized Function Generator. Technical Data.

Tektronix: Products > AWG 2040 Arbitrary Waveform Generator

Arbitrary/Function Waveform Generators 4075B Series

Model 310H Fast 800V Pulse Generator

8845A/8846A 6.5 Digit Precision Multimeters

The Fluke 6100A. Electrical Power Standard. The most accurate, comprehensive and flexible source of electrical power signals

5500A. Multi-Product Calibrator. Operator Reference Guide. PN December, 1994

Racal Instruments. Product Information

Dual Channel Function/Arbitrary Waveform Generators 4050B Series

GFT Channel Digital Delay Generator

PM 6685 PM 6685R Universal Frequency Counter Rubidium Frequency Counter Calibrator

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK

Arbitrary/Function Generators AFG3000C Series Datasheet

Agilent 8902A Measuring Receiver

5500A. Service Manual. Multi-Product Calibrator

Oscilloscope calibration Application Note

DG5000 Series Specifications

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

PicoSource PG900 Series

AWG-GS bit 2.5GS/s Arbitrary Waveform Generator

This DDS function generator is designed for use in such applications as communication equipment and electronic components production.

PicoSource PG900 Series USB differential pulse generators

FREQUENCY SYNTHESIZERS, SIGNAL GENERATORS

5522A. Getting Started. Multi-Product Calibrator

Fluke Norma 4000/5000

Arbitrary/Function Generator AFG1000 Series Datasheet

Fluke 170 Series True-rms Digital Multimeters

MODELS WW5061/2. 50MS/s Single/Dual Channel Arbitrary Waveform Generators

MG3740A Analog Signal Generator. 100 khz to 2.7 GHz 100 khz to 4.0 GHz 100 khz to 6.0 GHz

Keysight Technologies N9320B RF Spectrum Analyzer

Agilent 81133A/81134A

Agilent 83711B and 83712B Synthesized CW Generators

Agilent N9320B RF Spectrum Analyzer

Installation Tester and DMM Kits

Function/Arbitrary Waveform Generator

1507/1503. Insulation Testers. Technical Data. Features and benefits:

Models 296 and 295 combine sophisticated

Manual Supplement Fluke Corporation. All rights reserved. Printed in U.S.A. x

Chapter 5 Specifications

RIGOL Data Sheet. DG3000 Series Function/Arbitrary Waveform Generator DG3121A, DG3101A, DG3061A. Product Overview. Easy to Use Design.

The Fluke 6100A. Electrical Power Standard. The most accurate, comprehensive and flexible source of electrical power signals

15. Specifications WF1947/WF

TRANSMILLE 3010A PRECISION MULTIPRODUCT CALIBRATOR EXTENDED SPECIFICATIONS

2026Q CDMA/GSM Interferer MultiSource Generator

DG4000 Series Waveform Generators

7051Plus / 5051Plus. Extended Specification

Analog Arts SF900 SF650 SF610 Product Specifications

Traceability for Oscilloscopes and Oscilloscope Calibrators

Model 865 RF / Ultra Low Noise Microwave Signal Generator

DECLARATION OF CONFORMITY

Fluke 1587 FC/1577 Insulation Multimeters

Interpreting Specifications

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

Analog Arts SG985 SG884 SG834 SG814 Product Specifications [1]

Model Hz to 10MHz Precision Phasemeter. Operating Manual

9640A RF Reference Source Calibration made simpler

Green LED (right of 'D' type connector)

Contents. ZT530PCI & PXI Specifications. Arbitrary Waveform Generator. 16-bit, 400 MS/s, 2 Ch

HP 8901B Modulation Analyzer. HP 11722A Sensor Module. 150 khz MHz. 100 khz MHz. Technical Specifications. Four Instruments In One

Analog Arts SF990 SF880 SF830 Product Specifications

TG1010A AIM & THURLBY THANDAR INSTRUMENTS. 10MHz programmable DDS function generator. Direct Digital Synthesis

GT 9000 GT 9000S MICROWAVE

Fluke Norma 4000/5000

HP 8560 E-Series Spectrum Analyzers Technical Specifications

DSA700 Series Spectrum Analyzer

U1571A Ni-MH Battery Pack for U1600A Handheld Oscilloscopes

Racal Instruments. Product Information

DG5000 series Waveform Generators

Analog Arts SL987 SL957 SL937 SL917 Product Specifications [1]

Agilent 8560 E-Series Spectrum Analyzers

RIGOL Data Sheet. DG2000 Series Function/Arbitrary Waveform Generator DG2041A, DG2021A. Product Overview. Main Features.

Dual Channel Function/Arbitrary Waveform Generators 4050 Series

5320A Multifunction Electrical Tester Calibrator. Verify and calibrate electrical test tools with a single instrument

MODEL AND MODEL PULSE/PATTERN GENERATORS

MODEL 9041 EXTENDED SPECIFICATIONS. 25ppm TRANSPORTABLE CALIBRATOR

Specification RIGOL. 6 Specification

Transcription:

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications These specifications apply to the 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Oscilloscope Calibration Options. For general specifications, see the 5500A or 5520A Extended Specifications (downloadable from the Fluke web site or available from your Fluke representative). The specifications are valid when the calibrator is operated under the conditions specified in the operator manual, and has completed a warm-up period of at least twice the length of time the calibrator was powered off, up to a maximum of 30 minutes. The 5520A-SC1100 can only be installed in a 5520A calibrator. The 5500A-SC600 and SC300 can be installed in a 5520A or a 5500A calibrator.

Voltage Function Specifications Volt Function DC Signal Square Wave Signal 1 Load Into 50 W Into 1 MW Into 50 W Into 1 MW Amplitude Characteristics 0 V to ± 6.6 V 0 V to ± 130 V Adjustment range 1 mv to 24.999 mv 25 mv to 109.99 mv 110 mv to 2.1999 V 2.2 V to 10.999 V 11 V to 130 V ± (0.25 % of output + 40 µv) 1-2-5 (e.g., 10 mv, 20 mv, 50 mv) ± (0.05 % of output + 40 µv) Square Wave Frequency Characteristics 1 µv 10 µv 100 µv 1 mv 10 mv Continuous 10 Hz to 10 khz ± (2.5 ppm of setting) ± 1 mv to ± 6.6 V p-p ± (0.25 % of output + 40 µv) Typical abberation (from 50 % of leading/trailing edge) 25 mv to 130 V: within 4 µs < (0.5 % of output + 100 µv) 10 mv to 25 mv: within 8 µs 1 mv to 10 mv: within 14 µs 1 Positive or negative, zero referenced square wave. 2 Above 1 khz, ± (0.25 % of output + 40 µv). Assumes connectors and cables are in good condition. ± 1 mv to ± 130 V p-p ± (0.1 % of output + 40 µv) 2 5500A-SC300 Volt Function DC Signal Square Wave Signal 1 Load Into 50 W Into 1 MW Into 50 W Into 1 MW Amplitude range 0 V to ± 2.2 V 0 V to ± 33 V ± 1.8 mv to ± 2.2 V p-p Frequency range 1 Positive or negative, zero referenced square wave. ± (0.25 % of output + 100 µv) 1-2-5 (e.g., 10 mv, 20 mv, 50 mv) 10 Hz to 10 khz ± 1.8 mv to ± 105 V p-p 2 Fluke Corporation 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Extended Specifications

Edge Function Specifications Edge Characteristics into 50 Ω 1-Year Absolute Uncertainty, tcal ± 5 C Amplitude Rise time < 300 ps + 0/-100 ps (p-p) 5.0 mv to 2.5 V ± (2 % of output + 200 µv) Adjustment range values 4 digits ± 10 % around each sequence value (indicated below) 5 mv, 10 mv, 25 mv, 50 mv, 60 mv, 80 mv, 100 mv, 200 mv, 250 mv, 300 mv, 500 mv, 600 mv, 1 V, 2.5 V Other Edge Characteristics Frequency range 1 khz to 10 MHz 1 ± (2.5 ppm of setting) Frequency range 300 ps 1 (+ 0 ps/-100 ps) Typical jitter, edge to trigger < 5 ps [p-p] Leading edge abberations 2 within 2 ns from 50 % of rising edge] < (3 % of output + 2 mv) 2 ns to 5 ns < (2 % of output + 2 mv) 5 ns to 15 ns < (1 % of output + 2 mv) after 15 ns Typical duty cycle 45 % to 55 % Tunnel diode pulse drive 1 Frequency range above 2 MHz has rise time specification 350 ps. 2 Below 250 mv aberrations are typical. < (0.5 % of output + 2 mv) Square wave at 100 Hz to 100 khz, with variable amplitude of 60 V to 100V p-p 5500A-SC300 Edge Characteristics into 50 Ω 1-Year Absolute Uncertainty, tcal ± 5 C Amplitude range (p-p) 4.5 mv to 2.75 V ± (2 % of output + 200 µv) Frequency range 1 khz to 1 MHz ± (25 ppm of setting + 15 mhz) Rise time 1 ns Typical jitter, edge to trigger < 5 ps (p-p) Leading edge aberrations Within 10 ns < (2 % of output + 2 mv) 10 to 30 ns < (1 % of output + 2 mv) After 30 ns < (0.5 % of output + 2 mv) Typical duty cycle 45 % to 55 % 3 Fluke Corporation 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Extended Specifications

Leveled Sinewave Function Specifications 5520A-SC1100 (> 600 MHz) Characteristics into 50 Ω Adjustment Frequency 50 khz (reference) 600 MHz to 1.1 GHz Amplitude Characteristics 5 mv to 3.5 V < 100 mv: 3 digits; 100 mv: 4 digits Continuously Adjustable 1-year absolute uncertainty, tcal ± 5 C ± (2 % of output + 300 µv) ± (7 % of output + 300 µv) Flatness (relative to 50 MHz) 1 not applicable ± (5 % of output + 100 µv) Short-term amplitude stability 1 % 2 1-year absolute uncertainty, tcal ± 5 C Frequency Characteristics Distortion Characteristics 100 khz ± 2.5 ppm 2 nd harmonic -33 dbc 3 rd and higher harmonic -38 dbc 1 As measured near oscilloscope bandwidth frequency. 2 Within one hour after reference amplitude setting, provided temperature varies no more than ± 5 C. Leveled Sine Wave Characteristics into 50 (p-p) Frequency 50 khz (Reference) 50 khz to 100 MHz 100 MHz to 300 MHz ± (2 % of output + 300 µv) Amplitude ± (3.5 % of output + 300 µv) Flatness (relative to 50 khz) 1 Not applicable ± (1.5 % of output + 100 µv) 5 mv to 5.5 V Short-term amplitude stability 1 % 2 Frequency Distortion 10 khz ± 2.5 ppm 2 nd harmonic -33 dbc 3 rd and higher harmonics -38 dbc 1 As measured near oscilloscope bandwidth frequency. 2 Within one hour after reference amplitude setting, provided temperature varies no more than ± 5 C. ± (4 % of output + 300 µv) ± (2 % of output + 100 µv) 300 MHz to 600 MHz ± (6 % of output + 300 µv) ± (4 % of output + 100 µv) 4 Fluke Corporation 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Extended Specifications

5500A-SC300 Leveled Sine Wave Characteristics Frequency into 50 W 50 khz (Reference) 50 khz to 100 MHz 100 MHz to 300 MHz 1 Amplitude (p-p) 5 mv to 5.5 V 1 ± (2 % of output + 200 µv) ± (3.5 % of output + 300 µv) ± (4 % of output + 300 µv) Flatness (relative to 50 khz) 1 Not applicable ± (1.5 % of output + 100 µv) ± (2 % of output + 100 µv) Short-term amplitude stability 1 % 2 Frequency 10 khz ± 2.5 ppm Distortion 2 nd harmonic -33 dbc 3 rd and higher harmonics -38 dbc 1 Extended frequency range to 350 MHz is provided, but flatness is not specified. Amplitude is limited to 3 V for frequencies above 250 MHz. 2 Within one hour after reference amplitude setting, provided temperature varies no more than ± 5 C. Time Marker Function Specifications Time Marker into 50 W 1 5 s to 50 ms 20 ms to 100 ns 50 ns to 20 ns 10 ns 5 ns to 2 ns 1-year absolute uncertainty, ± (25 + t* X 1000) tcal ± 5 C 2 ppm ± 2.5 ppm ± 2.5 ppm ± 2.5 ppm ± 2.5 ppm Wave shape Spike or square Spike, square, 20 %-pulse Spike or square Square or sine Sine Typical jitter (p-p) < 10 ppm < 1 ppm < 1 ppm < 1 ppm < 1 ppm 5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms) 1 Output amplitude > 1 V pk. 2 Time marker uncertainty is ± 50 ppm when measured off of cardinal points. *t = time in seconds. 5500A-SC300 Time Marker into 50 W 1 5 s to 100 µs 50 µs to 2 µs 1 µs to 20 ns 10 ns to 2 ns 1-year absolute uncertainty, tcal ± 5 C ± (25 + t* X 1000) pmm ± (25 + t* X 15000) pmm ± 25 ppm ± 25 ppm Wave shape Pulsed sawtooth Pulsed sawtooth Pulsed sawtooth Sine 5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms) 1 Typical amplitude > 1 V *t = time in seconds. Trigger Functions 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Available for edge and time marker functions (volt, pulse and video are available for only). 5 Fluke Corporation 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Extended Specifications

Wave Generator Function Specifications 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Wave Generator Characteristics 10 Hz to 10 khz Amplitude Square Wave, Sine Wave, and Triangle Wave into 50 W or 1 MW Into 1 MΩ: 1.8 mv to 55 V p-p Into 50 Ω: 1.8 mv to 2.5 V p-p () Into 50 Ω: 1.8 mv to 2.2 V p-p (5500A-SC300) ± (3 % of p-p output + 100 µv) 1-2-5 (eg., 10 mv, 20 mv, 50 mv,...) Typical dc offset range 0 to ± ( 40 % of p-p amplitude) 1 1 The DC offset plus the wave signal must not exceed 30 V rms. Frequency 10 Hz to 100 khz 4 or 5 digits depending on frequency Pulse Generator Function Specifications The pulse generator is designed for oscilloscope capture function tests and trigger verification applications. Pulse Generator Characteristics Typical rise/fall times Amplitude available Positive Pulse into 50 W 2 ns Pulse Width 1 4 ns to 500 ns 1 Uncertainty 2 Discrete steps: 2.5 V, 1 V, 250 mv, 100 mv, 25 mv, 10 mv 5 % ± 2 ns Pulse Period 1 Pulse width not to exceed 40 % of period. 2 Pulse width uncertainty for periods less than 2 µs are not specified. 20 ms to 200 ns (50 Hz to 6.6 MHz) ± 2.5 ppm 6 Fluke Corporation 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Extended Specifications

TV Trigger Specifications. TV Trigger is provided at the Scope Output Terminal Trigger Signal Type Parameters Frame formats Selectable NTSC, SECAM, PAL, PAL-M Polarity Selectable Inverted or Uninverted Video Amplitude into 50 W (p-p) Adjustable 0 to 1.5 V p-p into 50 W load, (± 7 % accuracy) Line marker Selectable Line Video Marker Input Impedance Measurement Specifications Uncertainty Resistance 40 W to 60 W 0.1 % 500 kw to 1.5 MW 0.1 % Capacitance 5 pf to 50 pf ± (5 % of input + 0.5 pf) 1 1 Measurements made within 30 minutes of capacitance zero reference. Scope option must be selected for at least five minutes prior to capacitance measurement or zero. Overload Measurement Specifications. The overload test function applies dc or ac (1 khz square wave) power into the 50 W oscilloscope input and monitors the current. A time measurement counter indicates the time duration of the applied overload signal. When the oscillosope's input protection circuit reacts and opens up the 50 W load, the calibrator indication is set to off on the right hand display. In order to prevent oscilloscope front end damage, a limited amount of energy is applied by a user settable time limit. Source Voltage Time Limit dc or 1 khz ac 5 V to 9 V Settable from 1 Sec to 60 Sec Typical On Current Indication Typical Off Current Indication 5 V to 9 V Settable from 1 Sec to 60 Sec External Frequency Reference Input (5520A only) The External Reference Input selection allows the user to provide their own high stability 10 MHz reference clock for the 5500A-SC300, 5500A-SC600 and 5520A-SC1100 when fitted in a 5520A mainframe. All functions except Wave Generator and Marker greater than 50ms are then referenced to the external 10 MHz signal. The external reference input must be between 1 V to 5 V p-p. Uncertainty of output = uncertainty of reference + 5 µhz. 7 Fluke Corporation 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Extended Specifications

1000V RMS V,, RTD 20V PK 1V PK A, -SENSE, AUX V 20V RMS 20V PK 200V PK µ n p SHIFT dbm W sec F SETUP NEW REF MEAS TC MULT RESET CE TRIG OUT DIV EDIT Ordering Information Models 5500A Multi-Product Calibrator 5500A/3 Multi-Product Calibrator with 300 MHz Oscilloscope Calibration Option 5500A/6 Multi-Product Calibrator with 600 MHz Oscilloscope Calibration Option 5520A High-Performance Multi-Product Calibrator 5520A/3 High Performance Multi-Product Calibrator with 300 MHz Oscilloscope Calibration Option 5520A/6 High Performance Multi-Product Calibrator with 600 MHz Oscilloscope Calibration Option 5520A/1GHZ High Performance Multi-Product Calibrator with 1.1 GHz Oscilloscope Calibration Option Upgrades 5500A ->/3 UG UPGRADE 5500A TO 5500A/3 5500A ->/6 UG UPGRADE 5500A TO 5500A/6 5500A/3 ->/6 UG UPGRADE 5500A/3 TO 5500A/6 5520A ->/3 UG UPGRADE 5520A TO 5520A/3 5520A ->/6 UG UPGRADE 5520A TO 5520A/6 5520A ->1GHZ UG UPGRADE 5520A TO 5520A/1.1 GHZ 5520A ->PQ UG UPGRADE 5520A TO 5520A-PQ 5520A-PQ ->/3 UG UPGRADE 5520A-PQ TO 5520A-PQ/3 5520A-PQ ->/6 UG UPGRADE 5520A-PQ TO 5520A-PQ/6 5520A-PQ ->1GHZ UG UPGRADE 5520A-PQ TO 5520A-PQ/1GHZ 5520A-PQ/3 ->/6 UG UPGRADE 5520A-PQ/3 TO 5520A-PQ/6 5520A-PQ/3 ->1GHZ UG UPGRADE 5520A-PQ/3 TO 5520A-PQ/1GHZ 5520A-PQ/6 ->1GHZ UG UPGRADE 5520A-PQ/6 TO 5520A-PQ/1GHZ 5520A/1GHZ ->PQ UG UPGRADE 5520A/1GHZ TO 5520A-PQ/1G 5520A/3 ->/6 UG UPGRADE 5520A/3 TO 5520A/6 5520A/3 ->1GHZ UG UPGRADE 5520A/3 TO 5520A/1GHZ 5520A/3 ->PQ UG UPGRADE 5520A/3 TO 5520A-PQ/3 5520A/6 ->1GHZ UG UPGRADE 5520A/6 TO 5520A/1GHZ 5520A/6 ->PQ UG UPGRADE 5520A/6 TO 5520A-PQ/6 5520A->PQ/6 UG UPGRADE 5520A TO 5520A-PQ/6 5520A->PQ/3 UG UPGRADE 5520A TO 5520A-PQ/3 5520A->PQ/1GHZ UG UPGRADE 5520A TO 5520A-PQ/1 GHZ 5520A/3->PQ/6 UG UPGRADE 5520A/3 TO 5520A-PQ/6 5520A/3->PQ/1GHZ UG UPGRADE 5520A/3 TO 5520A-PQ/ 1 GHZ 5520A/6->PQ/1GHZ UG UPGRADE 5520A/6 TO 5520A-PQ/ 1 GHZ Upgrades must be installed and calibrated at a Fluke Service Center and include report of calibration. Accessories 5500A/COIL 50-Turn Current Coil 5500A/CASE Roll-Aboard Transit Case 5500A/LEADS Comprehensive Test Lead Kit for 5500A 5520A-525A/LEADS Comprehensive Test Lead Set for 5520A 5500A/HNDL Side Carry Handle Y5537 Rack Mount Kit 5725A Amplifier (5500A Only) MET/CAL Calibration Software (IEEE and RS-232) 5500/CAL Calibration Software (RS-232) 5800A-7002K Two Piece Replacement Output Cable Set 5800A-7004K Oscilloscope Calibrator BNC Connector Kit Product Compatability Chart Model 1.1 GHz Option 600 MHz Option 300 MHz Option 5520A 5500A 5500A CALIBRATOR NORMAL AUX SCOPE TC HI LO TRIG OUT STBY +/ 7 8 9 4 5 6 1 2 3 0 QUALITY MANAGEMENT SYSTEM ISO 9001 C E RT I FI E D T O M E E T IS O 9 0 01 43.2 cm (17 in) PREV OPR EARTH SCOPE BOOST MENU m V Hz FIELD k M A ENTER Front view and dimensions. C F x POWER I 47.0 cm (18.5 in) 6.4 cm (2.5 in) Side view and dimensions. Note: 5500A and 5520A dimensions are the same. O 17.8 cm (7 in) Fluke. Keeping your world up and running. For Cable Access Fluke Corporation PO Box 9090, Everett, WA 98206 U.S.A. Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com 1999-2009 Fluke Corporation. Specifications subject to change without notice. Printed in U.S.A. 2/2009 1265775 D-EN-N Rev E Pub_ID: 10806-eng, Rev 02 8 Fluke Corporation 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Extended Specifications Modification of this document is not permitted without written permission from Fluke Corporation.