HP 8921A Cell Site Test Set. Product Note AMPS Base Station Testing

Similar documents
HP 8647A and 8648A/B/C/D Signal Generators Technical Specifications

Permittivity Measurements of PC Board and Substrate Materials using the HP 4291B and HP 16453A


6.5-Digit Multimeter, High- Accuracy, C-Size HP E1410A Technical Specifications

HP 8491A/B, 8492A, 8493A/B/C, 11581A, 11582A and 11583A/C Coaxial Attenuators. Product Overview

HP 8509B Lightwave Polarization Analyzer. Product Overview. Optical polarization measurements of signal and components nm to 1600 nm

HP 8560 E-Series Spectrum Analyzers Technical Specifications

Wide Range DC Current Biased Inductance Measurement. Application Note with HP 4284A Precision LCR Meter / HP 42841A Bias Current Source

Hints. for making. Better. Spectrum Analyzer. Measurements. Application Note

product note Using Power Leveling to Control Test Port Output Power Product Note 8510XF XF Network Analyzer

Better. Hints. Microwave Counter. for making. Measurements

Basic Transceiver tests with the 8800S

E-series E9300 Average Power Sensor Specifications

Optimizing Electronic Component and Material Impedance Measurements

product overview HP 8970 Series Noise Figure Measurement Products

HP 4286A RF LCR Meter. Technical Specifications. Specifications. Test Signal

Fundamentals of Microwave Frequency Counters. Application Note Electronic Counters Series

Network, Spectrum, and Impedance Evaluation of Electronic Circuits and Components

DMR Application Note Testing MOTOTRBO Radios On the R8000 Communications System Analyzer

Application Note: DMR Application Note Testing MOTOTRBO Radios On the Freedom Communications System Analyzer

Introduction to BER testing of WDM systems

ID-5100 User Evaluation & Test Report

ProBER 2 2 Mb/s Handheld Test Set. The handheld that takes 2 Mb/s testing beyond convention

CDPD Conceptual Overview. Product Note

Application Note: Testing P25 Conventional Radios Using the Freedom Communications System Analyzers

Testing Motorola P25 Conventional Radios Using the R8000 Communications System Analyzer

Obtaining Flat Test Port Power with the Agilent 8360 s User Flatness Correction Feature. Product Note

Installed Radio Testing with the 3500

Agilent 8920A RF Communications Test Set Product Overview

Advanced Test Equipment Rentals ATEC (2832)

profile New Generation Precision Impedance Analyzer for functionality and efficiency in engineering 40 Hz to 110 MHz 4294A

200GTL ALIGNMENT REVISION: 1.0 BURKE MODEL: 200GTL REVISION: 1.2 DATE: 02/14/06. Total Pages: 6 pages. Page:1 print date: 9/23/09

PSM 900 Personal Monitor System. Personal Monitoring. Redefined.

RF Communications Test Set, Application Handbook

Quick Site Testing with the 8800SX

HP 4155B/4156B Semiconductor Parameter Analyzer Product Note-3 Prober Connection Guide. HP 4155B/4156B Semiconductor Parameter Analyzer

Agilent 8703A Lightwave Component Analyzer Technical Specifications

FCC ID: AXI IC: 10239A Alignment

EMC Accessories Catalog

Agilent 8657A/8657B Signal Generators

HP 8921A Cell Site Test Set HP 11807B Cell Site Test Software HP 83204A TDMA Cellular Adapter HP 83205A CDMA Cellular Adapter HP 83236B PCS Interface

Specifications. HP 8920A RF Communications Test Set. Key Features:

JD746A/JD786A CellAdvisor RF Analyzer

IC-400pro - RADIOAFICION.COM

Receiver Adjustments

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

Cross-Connect Interface

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Phase Noise Measurements

Understanding the Precision Antenna, Cable, and Power Measurements on the 3550 Radio Test System

Cable and Antenna Analyzer

Agilent 8644A-1 Phase noise test with the Agilent 8644A and 8665A Signal Generators Product Note

Mastr III P25 Base Station Transmitter Tune-up Procedure

PC Tune PC Tune Test Procedures for 5100 Series Portable Radios

Agilent AN 1275 Automatic Frequency Settling Time Measurement Speeds Time-to-Market for RF Designs

PAMS. User s Manual. Portable Attenuation Measurement System. The solution for making easy shielding effectiveness measurements.

HP 8920B RF Communications Test Set

Agilent 8902A Measuring Receiver

User s Manual For PAMS. Portable Attenuation Measurement System /

HF Transceiver Codan NGT SR. HF Transceiver Codan NGT SR

JD723A/JD724B/JD726A Cable and Antenna Analyzers

Agilent Technologies 8935 Series E6381A TDMA Base Station Test Set

AUTOTUNE USER GUIDE. R8000 Series Communications Systems Analyzer

The equipment will provide up to 50W RF output power in the MHz band.

Appendix G: IFR 8800 Test Procedures MAINTENANCE GUIDE.

MEASUREMENT PROCEDURE AND TEST EQUIPMENT USED

Technical Equipment Specification

Using High-Directivity Couplers in Isolatorless Cellular Phone PA Control

Frequency Coverage MHz RF Power Output 30W SSB / 9W AM/ 30W FM Dual Finals on Heat Sink Modes AM, FM, USB, LSB Microprocessor

Agilent 71400C Lightwave Signal Analyzer Product Overview. Calibrated measurements of high-speed modulation, RIN, and laser linewidth

VX-2100/VX-2200 (UHF) Alignment

Cell Extender Antenna System Design Guide Lines

LM1866 Low Voltage AM FM Receiver

Agilent E4438C ESG Vector Signal Generator Differential I/Q outputs. Product Note

Application Note. Understanding PassPort Radio Systems

i2820h (USA) ie2820(europe)

Agilent 8644A-2 Air Navigation Receiver Testing with the Agilent 8644A

IC-F7000. Advanced selective call and ALE make HF communication easier than ever!

Agilent 8902A Measuring Receiver Product Note

Wireless Portable Radio Communications Test Set. Advanced Test Equipment Rentals ATEC (2832)

INSTALLATION and OPERATION INSTRUCTIONS. FOR FiberLink BI-DIRECTIONAL AMPLIFIER WITH DIVERSITY MW-FBDA-800AB-50W-DIV

G7104A Multimaster Base Station Tester

8800SX DMR Repeater Test Option 06

Measurement Procedure & Test Equipment Used

Signal Generators for Anritsu RF and Microwave Handheld Instruments

MASTR II BASE STATION MHz RECEIVER IF/AUDIO/SQUELCH & RF ASSEMBLY (25 khz/12.5 khz CHANNEL SPACING) Maintenance Manual LBI-38506A

Rockwell Collins, Inc. VHF Users Manual

Base Station Installation and Maintenance

Yaesu FT-8800R Alignment

evolution wireless G4 ew 100 G4-ME2 ew 100 G4-ME4 Lavalier Set

Agilent PSA Series Spectrum Analyzers Noise Figure Measurements Personality


INSTALLATION AND OPERATING MANUAL

Frequency and Time Domain Representation of Sinusoidal Signals

Agilent PNA Microwave Network Analyzers

TS-590SG HF/ 50MHz All-Mode TRANSCEIVER_

UHF Wireless Microphone System

Agilent AN Balanced Circuit Measurement with an Impedance Analyzer/LCR Meter/Network Analyzer Application Note

FT-897 Alignment. Local Oscillator Adjustment. PLL Adjustment

TABLE OF CONTENTS. Magnum International PO Box 445 Issaquah, WA 98027

Data and Computer Communications

Transcription:

HP 8921A Cell Site Test Set Product Note 8921-1 AMPS Base Station Testing

AMPS Base Station Testing This product note describes manual operation of the HP 8921A Cell Site Test Set (or HP 8920A RF Communications Test Set with options 001, 002, 003, 004, 005, 013, 014) for testing Advanced Mobile Phone System (AMPS) Cellular Base Station equipment. It will lead you through the steps required to set-up the instrument in the proper state to make particular measurements. Storing these instrument states in memory registers allows you to quickly recall them without having to go through the laborious task of creating them each time you want to make a measurement on your base station equipment. For automated base station testing, Hewlett-Packard offers HP 11807B Cell Site Test Software for use with the HP 8921A Cell Site Test Set. The HP 11807B software has options that cover the most widely used AMPS base stations. Each option uses test and adjustment procedures derived from the manufacturers maintenance manuals. For more information on this and other Hewlett-Packard products, contact your local Hewlett-Packard representative. Cellular System Overview Cell Site Overview A cell site contains three types of radios: control, voice, and locate. When a mobile originates a call or is paged, the cell site control radio sets up the call. This control radio communicates with the central switch (Electronic Mobile Exchange (EMX), Mobile Switching Center (MSC), and Mobile Telephone Switching Office (MTSO)) to connect the call to another mobile or landline phone. The mobile is then instructed by the cell site control channel to tune to a particular voice channel. The conversation takes place on the voice radios with the voice signals traveling back to the MSC and routed to the other party. The locate, or scanner, radio is used to monitor the signal levels of the channels and notifies the switch enabling it to make handoff decisions. This radio is receive only and, like the voice and control radios, is controlled by distributed control functions between the MSC and local cell site controller. Communication of these control functions takes place via either a landline or microwave transmission data link. Cell sites are configured as omnidirectional, sectored, or some combination of the two. In general, an omni-directional cell will have two receive antennas and one transmit antenna. 2

Basic Omni Cell Site Setup The two receive antennas are physically separated to get better immunity to multipath fading conditions and are typically space diversity antennas. Other omni sites may have only two receive antennas to save money. In this case, one receive path is coupled to the same antenna as the transmit path. A tuned cavity duplexer (combined receive and transmit filter) is then used to keep the signal paths separated. A directional, or sectored, cell splits the coverage into sectors which are served by different antennas. Typically, a directional cell will have three sectored receive antennas, one omni receive antenna, three sectored transmit antennas, and one omni transmit antenna. Voice radios are configured on one face of the cell for both transmit and receive and the locate radio accepts inputs from all of the receive antennas. The one omni-transmit antenna is used for control channel transmit. When testing base station equipment in directional cells, you will be working with radios 3 - Sector Directional Site configured to different antennas. Gain/loss for each antenna path will have to be taken into consideration for power measurements and switching between receive antennas for testing receiver sensitivity may also be necessary. 3

AMPS Base Station Testing Base Station Tests Standard tests to make on base station equipment include receive, transmit, and system tests. This product note outlines the following tests for cell site transceivers: Transmitter Tests Frequency Error/Offset Output Power SAT Frequency Error and Deviation Voice Deviation Data Deviation Receiver Tests Sensitivity (SINAD) Squelch Threshold The following System Test procedures are NOT INCLUDED in this product note. However, you can find detailed procedures for these tests in the HP 8920A RF Communications Test Set Applications Handbook (Hewlett-Packard PN 08920-90126). System Tests (not included) Antenna Return Loss (VSWR) Cable Fault Location Field Strength Measurements Cell Site Connections The basic connections required for performing the manual tests in this application note include TX, RX, and audio. Because each manufacturers base station equipment may connect to the HP 8921A Cell Site Test Set differently, it will be up to you to determine how best to connect your transceivers to the HP 8921A. Test Set Operation This product note assumes the user has some familiarity with the general use of the HP 8921A Cell Site Test Set (or HP 8920A RF Communications Test Set). If this is not the case, please refer to the HP 8921A Cell Site Test Set User s Guide and the HP 8920A RF Communications Test Set Applications Handbook for an overview of its functions, features, and capabilities. This product note also assumes that the user is familiar with the operation of the Base Station under test. It is not the intent of this note to show the step-by-step setups required for setting up all base stations for particular measurements since many base stations have unique setups and control options. However, when a particular base station setup is required, it will be called out in the setup procedure. 4

HP 8921A Cell Site Test Set HP 8921A Cell Site Test Set Overview The HP 8921A Cell Site Test Set is a single instrument that combines the features of 22 individual radio test instruments. The HP 8921A is designed to meet the communication needs of both service and manufacturing environments, and has the capability to test AMPS and TACS Base Stations, cellular phones, and various other communications systems. Test and troubleshooting time is decreased by simplifying standard measurement tasks and providing the required measurement capability in a single instrument. Transmitters and receivers are characterized with single-key RX, TX, and duplex tests. Each of these tests displays a specialized screen that provides access to the necessary controls and measurement results. Measured results may be displayed as digital readouts and/or bar graphs. All settings and measurements are easily accessed and changed using the front panel knob and keys. If desired, all settings can be saved in nonvolatile save/recall registers for future access. The various ports on the HP 8921A allow the receiver/transmitter being tested to be quickly connected. A receiver with 2 µv sensitivity (typically <1 µv) is available through the ANT IN port, for off-the-air monitoring of low-level signals. Transmitters of up to 60 W (100 W intermittently for 10 seconds/minute) can be connected to the RF IN/OUT port without the use of external attenuators. Global Measurement Considerations Saving and Recalling Instrument Setups The Save and Recall functions allow you to store different instrument setups and retrieve them later eliminating the task of reconfiguring the HP 8921A each time you wish to make a measurement. You can name save registers anything you want provided the name does not exceed 9 characters. If there is a particular setup you would like the instrument to be in whenever you power it up, save the setup with the name POWER ON. To save an instrument setup: 1. Make any changes to the instrument that you want to SAVE in a register. 2. Press, shift, SAVE. 3. Using the front panel knob, create a new name or choose an existing name from the SAVE menu. 4. Scroll to the top of the SAVE menu and choose DONE. The current instrument state is now saved. Configuring the HP 8921A for AMPS Base Station Testing The Configure screen defines a number of general functions, such as date and time, screen intensity, and beeper volume. It is also used to define some RF signal parameters, such as RF Generator / Analyzer offset, channel standards and characteristics, and signal loss/gain compensation. Changes made to the "Configure" screen will be saved when you save instrument setups. 5

AMPS Base Station Testing Screen Configuration Explanations: A The RF Display field selects the format for entering the RF Analyzer frequencies: * When Freq. is selected, you enter the RF Generator and RF Analyzer frequencies directly using the keypad or knob. * When Chan is selected, the RF Gen Freq and Tune Freq fields on all screens are replaced by the RF Channel field, and only the channel number is entered and displayed. B C The RF Chan Std field allows you to select the channel standard for the radio under test. The RF Generator and RF Analyzer frequencies will be automatically set to correspond to the channel number entered in the RF Channel field. This field also sets the correct TX/RX frequency offset (45 MHz for AMPS). The RF Level Offset field enables/disables the effects of the RF In/Out, Duplex Out, and Antenna In fields below it. These fields are used if you wish to take into account losses due to cables and/or splitters in your setup. * When set to Yes, the RF Generator amplitude and RF Analyzer power measurement are offset by the values entered in these fields. * When set to No, the values in these fields are ignored. The values can, however, be left in the three fields. For example, enter "-3.5" into the RF In/Out field for the approximate total loss due to a 1 meter cable and 3 db splitter. From the front panel of the HP 8921A, enter the "Configure" screen. Set the RF Display and RF Chan Std as seen here and the RF Level Offset as desired. 6

Performing Measurements The following measurement procedures assume the HP 8921A Configure screen is setup as described in the previous section. If the instrument has been preset since then, you will need to reconfigure that screen. Transmitter Tests Transmitter Frequency Error/Offset and Power... page 8 Transmitter SAT Frequency and Deviation... page 9 Transmitter Maximum Voice Deviation... page 10 Transmitter Data Deviation... page 11 Connect the Base Station to the HP 8921A according to the following diagram. From Base Station { TX Output RX Input RX Audio Out } From Base Station TX Audio In 7

Transmitter Frequency Error/Offset and Power Base Station Setup Channel: As desired TX Power Level: As desired TX Power: On SAT: Off Data: Off HP 8921A Cell Site Test Set Setup Step 1: Press the TX key on the HP 8921A to enter the transmitter test screen. Step 2: Change the settings in the TX screen as seen in the highlighted areas. Key instrument considerations A With the front panel keys, enter the same RF Channel as set on the base station. B Disable the FM Deviation and AF Freq measurements by scrolling to their units fields and pressing the On/Off key. Step 3: The transmitter frequency error and power should now be displayed in the appropriate result fields. For TX Output Power in units other than Watts, scroll to and highlight the TX Power units field and press the appropriate units key. Step 4: Save this measurement setup as FREQ. 8

Transmitter SAT Frequency and Deviation Base Station Setup Channel: As desired TX Power Level: As desired TX Power: On SAT: On Data: Off HP 8921A Cell Site Test Set Setup Step 1: Press the TX key on the HP 8921A to enter the transmitter test screen. Step 2: Change the settings in the TX screen as seen in the highlighted areas. Key instrument considerations A Enter the same RF Channel as set on the base station. If you are using an HP 8920A RF Communications Test Set, you B must have Option 014 (6 khz BPF) in order to choose this filter. If not, you can use the 15 khz LPF. The 6 khz BPF is standard on the HP 8921A. C D AFGen1 Lvl must be "OFF" or it may interfere with the SAT measurement. Disable the TX Freq Error and TX Power measurements by scrolling to their units fields and pressing the On/Off key. Step 3: The SAT frequency and deviation should now be displayed in the appropriate result fields. For AF Freq in units other than Hz, scroll to and highlight the AF Freq units field and press the appropriate units key. Step 4: Save this measurement setup as SAT. 9

Transmitter Maximum Voice Deviation Base Station Setup Channel: As desired TX Power Level: As desired TX Power: On SAT: Off Data: Off HP 8921A Cell Site Test Set Setup Step 1: Press the TX key on the HP 8921A to enter the transmitter test screen. Step 2: Change the settings in the TX screen as seen in the highlighted areas. Key instrument considerations A Enter the same RF Channel as set on the base station. For AFGEN1 Lvl, enter an initial value of 50 mv and set an increment for this field of 20 B db. C Disable the TX Freq Error and TX Power by scrolling to their units fields and pressing the On/Off key. Step 3: Scroll to the AFGen1 Lvl field and depress the Up-Arrow key once. This will increment the AFgen1 Lvl by 20 db (as set above). Step 4: The maximum voice deviation should now be displayed in the FM Deviation result field. Step 5: Save this setup as MAX_VOICE. 10

Transmitter Data Deviation Base Station Setup Channel: As desired TX Power Level: As desired TX Power: On SAT: Off Data: On HP 8921A Cell Site Test Set Setup Step 1: Press the TX key on the HP 8921A to enter the transmitter test screen. Step 2: Change the settings in the TX screen as seen in the highlighted areas. Key instrument considerations A Enter the same RF Channel as set on the base station. B AFGen1 Lvl must be "OFF" or it may interfere with the DATA measurement. C Disable the TX Freq Error, TX Power, and AF Freq measurements by scrolling to their units fields and pressing the On/Off key. Step 3: The Data Deviation should now be displayed in the FM Deviation result field. Step 4: Save this measurement setup as DATA. 11

Receiver Measurements Receiver Sensitivity (SINAD)... page 13 Receiver Squelch Threshold... page 14 Connect the Base Station to the HP 8921A according to the following diagram. Note the Audio In connection. This must be connected properly for accurate receiver audio measurements Audio IN From Base Station { TX Output RX Input RX Audio Out } From Base Station TX Audio In Using the front-panel knob, scroll to and select the AF ANL screen, and select 600 TO HI in the Audio IN LO field. 600 TO Hi establishes a 600 Ω impedance between the center pins of the Audio IN LO and Audio IN HI connectors. 12

Receiver Sensitivity (SINAD) Base Station Setup Channel: As desired TX Power Level: As desired TX Power: Off SAT: Off Data: Off Squelch: Off HP 8921A Cell Site Test Set Setup Step 1: Press the RX key on the HP 8921A to enter the receiver test screen. Step 2: Change the settings in the RX screen as seen in the highlighted areas. Key instrument considerations A Enter the same RF Channel as set on the base station. B Enter an initial value in the amplitude field. This value will depend on the sensitivity of the receiver you are testing. Step 3: Using the front-panel knob, adjust the Amplitude until the desired SINAD is measured. To adjust the resolution of the amplitude adjustment, use the INCR SET, INCR 10, or INCR œ10 keys. Step 4: Save this measurement setup as SINAD. 13

Receiver Squelch Threshold Base Station Setup Channel: As desired TX Power Level: As desired TX Power: Off SAT: Off Data: Off Squelch: On HP 8921A Cell Site Test Set Setup Step 1: Press the RX key on the HP 8921A to enter the receiver test screen. Step 2: Change the settings in the RX screen as seen in the highlighted areas. Key instrument considerations A Enter the same RF Channel as set on the base station. B Enter a starting amplitude. This initial amplitude value will depend on the squelch sensitivity of your receiver. C It is important to activate the Atten Hold function of the HP 8921A for this test. If this is not done, the RF signal to the receiver may drop out momentarily as you adjust its amplitude and an attenuator change is required. This will activate the squelch on the receiver and make it difficult to measure the squelch sensitivity. 14

Step 3: Adjust the volume on the HP 8921A to verify audio output. If no audio is heard, the base station may be in a squelched condition. In this case, increase the amplitude until the squelch de-activates and an audio signal is heard. With the amplitude field highlighted, slowly decrease the amplitude of the RF signal and note the point at which the squelch activates and the audio signal is muted. To adjust the resolution of the amplitude adjustment, use the INCR SET, INCR 10, or INCR œ10 keys. For squelch hysteresis testing, slowly increase the RF level until the squelch de-activates and an audio signal is present. Note the RF level at which this occurs. The squelch hysteresis is the difference between this level and the level at which the squelch activates. Step 4: Save this measurement setup as SQUELCH. Appendix A Typical AMPS System Specifications Parameter Specification TX Frequency Error ±1.5 ppm TX Sat Frequency ±1 Hz Deviation 2 khz ±10% TX Voice Nominal Deviation 8 khz ±10% Maximum Deviation 12 khz TX Wideband Data Deviation 8 khz ±10% RX Sensitivity (SINAD) < -116 dbm @ 12 db SINAD 15

For more information on Hewlett- Packard Test & Measurement products, applications or services please call your local Hewlett-Packard sales offices. A current listing is available via Web through http://www.hp.com./go/tmdir If you do not have access to the internet please contact one of the HP centers listed below and they will direct you to your nearest HP representative. United States: Hewlett-Packard Company Test and Measurement Organization 5301 Stevens Creek Blvd. Bldg. 51L-SC Santa Clara, CA 95052-8059 1 800 452 4844 Canada: Hewlett-Packard Canada Ltd. 5150 Spectrum Way Mississauga, Ontario L4W 5G1 (905) 206 4725 Europe: Hewlett-Packard European Marketing Center P.O. Box 999 1180 AZ Amstelveen The Netherlands Japan: Hewlett-Packard Japan Ltd. Measurement Assistance Center 9-1, Takakura-Cho, Hachioji-Shi, Tokyo 192, Japan Tel: (81-426) 56-7832 Fax: (81-426) 56-7840 Latin America: Hewlett-Packard Latin American Region Headquarters 5200 Blue Lagoon Drive 9th Floor Miami, Florida 33126 U.S.A. (305) 267 4245/4220 Australia/New Zealand: Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130 Australia 1 800 629 485 Asia Pacific: Hewlett-Packard Asia Pacific Ltd 17-21/F Shell Tower, Times Square, 1 Matheson Street, Causeway Bay, Hong Kong Fax: (852) 2506 9285 Data subject to change Hewlett-Packard Company 1997 Printed in U.S.A. 5/97 5962-9475E 16