SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

Similar documents
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z & ANSI/NCSL Z

President For the Accreditation Council Certificate Number Valid to January 31, 2010

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PYLON ELECTRONICS INC. 147 Colonnade Road Ottawa, Ontario Canada K2E 7L9 Jim Mullins Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. AMERICAN GAGE 1131 S. Richfield Rd Placentia, CA Roger Arnold Phone: CALIBRATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. J. A. KING & COMPANY, LLC 7103 Juniper Road Fairview, TN Connie Foster Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z540.3

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PYLON ELECTRONICS INC. 147 Colonnade Road Ottawa, Ontario Canada K2E 7L9 Jim Mullins Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CERTIFICATE OF ACCREDITATION. ANSI/NCSL Z (R2002) and ANSI/NCSL Z (R2013)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSLI Z

CERTIFICATE OF ACCREDITATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z & ANSI/NCSLI Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z & ANSI/NCSLI Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CERTIFICATE OF ACCREDITATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. J. A. KING & COMPANY, LLC South Lakes Drive Charlotte, NC Connie Foster Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. EL AL STANDARDS LABORATORY EL AL Head Office Ben-Gurion Airport Motty Shai Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

ACCREDITED LABORATORY

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Accredited Laboratory

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSLI Z

CERTIFICATE OF ACCREDITATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CERTIFICATE OF ACCREDITATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z540.3

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017. J. A. KING 354 Riverchase Way Lexington, SC Connie Foster Phone: CALIBRATION

Transcription:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 SIMCO ELECTRONICS 8203-J Piedmont Triad Parkway Greensboro, NC 27409 Gene Denmark Phone: 336 605 1665 Fax: 336 605 1934 CALIBRATION Valid To: September 30, 2017 Certificate Number: 1395.01 In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this laboratory to perform the following calibrations 1 : I. Dimensional Parameter/Equipment Range CMC 2, 4 ( ) Comments Micrometers 3 Up to 6 in 64 µin + 0.6R Gage blocks Depth Gages 3 Up to 12 in 68 µin + 0.6R Gage blocks Calipers 3 Up to 12 in (12 to 24) in 580 µin + 0.6R 590 µin + 0.6R Gage blocks Dial Indicators 3 Up to 1 in 96 µin Starrett 716 Rulers Up to 24 in 570 µin Denmark Designs Linear scale calibrator (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 1 of 16

II. Electrical DC/Low Frequency Parameter/Equipment Range CMC 2 ( ) Comments DC Voltage Generate 3 (0 to 330) mv (0.33 to 3.3) V (3.3 to 33) V (33 to 330) V (330 to 1000) V 19 µv/v + 1 V 9.1 µv/v + 2 V 9.8 µv/v + 20 V 21 µv/v + 150 V 14 µv/v + 1.5 mv Fluke 5520A/ SC1100 DC Voltage Measure 3 (0 to 100) mv (0.1 to 1) V (1 to 10) V (10 to 100) V (100 to 1000) V* 6.3 µv/v + 0.3 V 4.5 µv/v + 0.3 V 4.5 µv/v + 0.5 V 6.6 µv/v + 30 V 6.6 µv/v + 100 V* HP 3458A opt-2 *add 12 µv/v x (V in /1000) 2 to all V in > 100 (1 to 90) kv 0.12 % rdg Ross Eng. HV divider, HP 34401A DC Current Generate 3 (0 to 330) A (0.33 to 3.3) ma (3.3 to 33) ma (33 to 330) ma (0.33 to 1.1) A (1.1 to 11) A (11 to 20.5) A 0.013 % rdg + 0.02 A 0.008 % rdg + 0.05 A 0.009 % rdg + 0.25 A 0.008 % rdg + 2.5 A 0.016 % rdg + 40 A 0.04 % rdg + 500 A 0.078 % rdg + 750 A Fluke 5520A/ SC1100 DC Current Measure 3 (0 to 1) A (1 to 10) A (10 to 100) A (0.1 to 1) ma (1 to 10) ma (10 to 100) ma (0.1 to 1) A 0.062 % rdg + 40 pa 0.0067 % rdg + 110 pa 0.0027 % rdg + 800 pa 0.0028 % rdg + 5 na 0.0031 % rdg + 50 na 0.0047 % rdg + 500 na 0.013 % rdg + 10 A HP 3458A opt-2 (1 to 100) A 0.058 % rdg Empro C-1500-50 HP 3458A (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 2 of 16

Parameter/Equipment Range CMC 2 ( ) Comments Resistance Generate 3 (0 to 11) (11 to 33) (33 to 110) (110 to 330) (0.33 to 1.1) k (1.1 to 3.3) k (3.3 to 11) k (11 to 33) k (33 to 110) k (110 to 330) k (0.33 to 1.1) M (1.1 to 3.3) M (3.3 to 11) M (11 to 33) M (33 to 110) M (110 to 330) M (0.11 to 1.1) G 34 / + 1 m 34 / + 1.5 m 24 / + 1.4 m 22 / + 2 m 22 / + 2 m 22 / + 0.02 22 / + 0.02 22 / + 0.2 22 / + 0.2 30 / + 2 29 / + 2 49 / + 30 0.010 % rdg + 50 0.021 % rdg + 2.5 k 0.056 % rdg + 3 k 0.23 % rdg + 100 k 1.2 % rdg + 500 k Fluke 5520A/ SC1100 Resistance Measure 3 10 m to 10 (10 to 100) (0.1 to 1) k (1 to 10) k (10 to 100) k (0.1 to 1) M (1 to 10) M (10 to 100) M (100 to 1000) M 25 / + 50 16 / + 0.5 m 12 / + 0.5 m 12 / + 5 m 12 / + 0.05 20 / + 2 83 / + 100 0.1 % rdg + 1 k 0.71 % rdg + 10 k HP 3458A opt-2 Oscilloscopes 3 Level Sine Amp 50 khz Ref. 5.0 mv to 5.5 V p-p 2.6 % rdg + 300 V Fluke 5520A/SC1100 Level Sine Flatness 5 mv to 5.5 V Relative to 50 khz Reference 50 khz to 100 MHz (100 to 300) MHz (300 to 600) MHz (600 to 1000) MHz 2 % rdg + 100 V 2.5 % rdg + 100 V 4.7 % rdg + 100 V 6 % rdg + 100 V Square Wave Amp 1 mv to 130 V p-p 1 mv to 6.6 V p-p 0.14 % rdg + 40 µv 0.30 % rdg + 40 µv Into 1 M load Into 50 load (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 3 of 16

Parameter/Equipment Range CMC 2 ( ) Comments Oscilloscopes 3 (cont) Fluke 5520A/SC1100 DC Signal Level Up to 130 V Up to 6.6 V 0.06 % rdg + 40 µv 0.3 % rdg + 40 µv Into 1 M load Into 50 load Time Marker Output into 50 1 ns to 20 ms 50 ms to 5 s 2.9 s/s (1.2 + 1000*t) s/s t = time in seconds Edge Transition Time 1 khz to 10 MHz 120 ps Electrical Simulation of Thermocouples Generate 3 Type E -250 C to -100 C -100 C to -25 C -25 C to 350 C 350 C to 650 C 650 C to 1000 C 0.58 C 0.19 C 0.17 C 0.19 C 0.25 C Fluke 5520A/SC1100 Type J -210 C to -100 C -100 C to -30 C -30 C to 150 C 150 C to 760 C 760 C to 1200 C 0.32 C 0.32 C 0.17 C 0.21 C 0.27 C Type K -200 C to -100 C -100 C to -25 C -25 C to 120 C 120 C to 1000 C 1000 C to 1372 C 0.39 C 0.21 C 0.19 C 0.31 C 0.47 C Type T -250 C to -150 C -150 C to 0 C 0 C to 120 C 120 C to 400 C 0.73 C 0.28 C 0.19 C 0.17 C (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 4 of 16

AC Voltage Generate 3 (1 to 33) mv (10 to 45) Hz 45 Hz to 10 khz (10 to 20) khz (20 to 50) khz (50 to 100) khz (100 to 500) khz 0.065 % rdg + 6 V 0.019 % rdg + 6 V 0.021 % rdg + 6 V 0.081 % rdg + 6 V 0.28 % rdg + 12 V 0.29 % rdg + 50 V Fluke 5520A/SC1100 (33 to 330) mv (10 to 45) Hz (45 Hz to 10) khz (10 to 20) khz (20 to 50) khz (50 to 100) khz (100 to 500) khz 0.024 % rdg + 8 V 0.012 % rdg + 8 V 0.013 % rdg + 8 V 0.029 % rdg + 8 V 0.063 % rdg + 32 V 0.29 % rdg + 70 V (0.33 to 3.3) V (10 to 45) Hz 45 Hz to 10 khz (10 to 20) khz (20 to 50) khz (50 to 100) khz (100 to 500) khz 0.024 % rdg + 50 V 0.013 % rdg + 60 V 0.016 % rdg + 60 V 0.024 % rdg + 50 V 0.056 % rdg + 125 V 0.29 % rdg + 600 V (3.3 to 33) V (10 to 45) Hz 45 Hz to 10 khz (10 to 20) khz (20 to 50) khz (50 to 100) khz 0.024 % rdg + 650 V 0.013 % rdg + 600 V 0.019 % rdg + 600 V 0.028 % rdg + 600 V 0.071 % rdg + 1.6 mv (33 to 330) V 45 Hz to 1 khz (1 to 10) khz (10 to 20) khz (20 to 50) khz (50 to 100) khz 0.016 % rdg + 2.0mV 0.017 % rdg + 6.0 mv 0.020 % rdg + 6.0 mv 0.025 % rdg + 6.0 mv 0.16 % rdg + 50 mv (330 to 1020) V 45 Hz to 1 khz (1 to 5) khz (5 to 10) khz 0.024 % rdg + 10 mv 0.02 % rdg + 10 mv 0.024 % rdg + 10 mv (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 5 of 16

AC Voltage Measure 3 (10 to 100) mv, Locked on 100 mv Range (1 to 40) Hz 40 Hz to 1 khz (1 to 20) khz (20 to 100) khz 100 khz to 1 MHz 0.012 % rdg + 4 V 0.013 % rdg + 2 V 0.018 % rdg + 2 V 0.095 % rdg + 2 V 1.2 % rdg + 10 V HP 3458A opt-2 100 mv to 1 V (1 to 40) Hz 40 Hz to 1 khz (1 to 20) khz (20 to 100) khz 100 khz to 1 MHz 0.01 % rdg + 40 V 0.009 % rdg + 20 V 0.017 % rdg + 20 V 0.14 % rdg + 20 V 1.2 % rdg + 100 V (1 to 10) V (1 to 40) Hz 40 Hz to 1 khz (1 to 20) khz (20 to 100) khz (100 khz to 1) MHz 0.0092 % rdg + 0.4 mv 0.023 % rdg + 0.2 mv 0.017 % rdg + 0.2 mv 0.093 % rdg + 0.2 mv 1.2 % rdg + 1.0 mv (10 to 100) V (1 to 40) Hz 40 Hz to 1 khz (1 to 20) khz (20 to 100) khz 0.024 % rdg + 4 mv 0.024 % rdg + 2 mv 0.024 % rdg + 2 mv 0.14 % rdg + 2 mv (100 to 700) V (1 to 45) Hz 40 Hz to 1 khz (1 to 20) khz (20 to 50) khz 0.05 % rdg + 40 mv 0.047 % rdg + 20 mv 0.1 % rdg + 20 mv 0.16 % rdg + 20 mv (1 to 64) kv 60 Hz 0.58 % rdg Ross Eng. HV divider, HP 34001A (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 6 of 16

AC Current Generate 3 (29 to 329.99) A (10 Hz to 1) khz (1 to 5) khz 0.12 % rdg + 0.1 µa 0.24 % rdg + 0.15 A Fluke 5520A/SC1100 (0.33 to 3.2999) ma (10 Hz to 1) khz (1 to 5) khz 0.08 % rdg + 0.15 µa 0.16 % rdg + 0.2 A (3.3 to 32.999) ma (10 Hz to 1) khz (1 to 5) khz 0.03 % rdg + 2 µa 0.06 % rdg + 2 A (33 to 329.99) ma (10 Hz to 1) khz (1 to 5) khz 0.03 % rdg + 20 µa 0.08 % rdg + 50 A (0.33 to 1.09999) A (10 Hz to 1) khz (1 to 5) khz 0.04% % rdg + 100 µa 0.47 % rdg + 1.0 ma (1.1 to 10.9999) A (10 Hz to 1) khz (1 to 5) khz 0.04 % rdg + 2.0 ma 2.4 % rdg + 2.0 ma (11 to 20.5) A (45 Hz to 1) khz (1 to 5) khz 0.15 % rdg + 5.0 ma 2.4 % rdg + 5.0 ma Parameter/Equipment Range CMC 2 ( ) Comments AC Current Measure 3 (45 Hz to 5) khz (5 to 100) A (0.1 to 1) ma (1 to 10) ma (10 to 100) ma (0.1 to 1) A 0.07 % rdg + 0.03 A 0.04 % rdg + 0.2 A 0.04 % rdg + 2 A 0.04 % rdg + 20 A 0.12 % rdg + 0.2 ma HP 3458A opt-2 @ 60 Hz (1 to 100) A 2.3 % Fluke i400s Capacitance 3 Generate (0.19 to 0.3999) nf (0.4 to 1.0999) nf (1.1 to 3.2999) nf (3.3 to 10.9999) nf (11 to 32.9999) nf (33 to 109.999) nf (110 to 329.999) nf (0.33 to 1.09999) uf (1.1 to 3.29999) µf (3.3 to 10.9999) µf (11 to 32.9999) µf 0.62 % rdg + 0.01 nf 0.58 % rdg + 0.01 nf 0.58 % rdg + 0.01 nf 0.29 % rdg + 0.01 nf 0.29 % rdg + 0.1 nf 0.29 % rdg + 0.1 nf 0.29 % rdg + 0.3 nf 0.3 % rdg + 1 nf 0.29 % rdg + 3 nf 0.29 % rdg + 10 nf 0.47 % rdg + 30 nf Fluke 5520A/ SC1100 (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 7 of 16

Parameter/Equipment Range CMC 2 ( ) Comments Capacitance 3 Generate (cont) (33 to 109.999) µf (110 to 329.999) µf (0.33 to 1.09999) mf (1.1 to 3.29999) mf (3.3 to 10.9999) mf (11 to 32.9999) mf (33 to 110) mf 0.52 % rdg + 100 nf 0.52 % rdg + 300 nf 0.52 % rdg + 1 uf 0.52 % rdg + 3 uf 0.52 % rdg + 10 uf 0.87 % rdg + 30 uf 1.3 % rdg + 100 uf Fluke 5520A/ SC1100 III. Electrical RF/Microwave Parameter/Range Frequency CMC 2 ( ) Comments Amplitude Modulation Measure 3 (0.15 to 10) MHz 50 Hz to 10 khz rate Depth: (5 to 99) % 3.5 % rdg + 1 digit HP 8902A measuring receiver 20 Hz to 10 khz rate Depth: Up to 99 % 4.3 % rdg + 1 digit (10 to 1300) MHz 50 Hz to 50 khz rate Depth (5 to 99) % 2.8 % rdg + 1 digit 20 Hz to 100 khz rate Depth: Up to 99 % 4.3 % rdg + 1 digit (1.3 to 18) GHz 50 Hz to 50 khz rate Depth: (5 to 99) % 2.9 % rdg + 1 digit 20 Hz to 100 khz rate Depth: Up to 99 % 4.3 % rdg + 1 digit w/ HP 11793A, HP 8673B (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 8 of 16

Frequency Modulation Measure 3 (0.25 to 10) MHz 20 Hz to 10 khz rate 40 khz peak 2.7 % rdg + 1 digit HP 8902A measuring receiver (10 to 1300) MHz 50 Hz to 100 khz rate 400 khz peak 1.8 % rdg + 1 digit 20 Hz to 200 khz rate 40 khz peak 5.9 % rdg + 1 digit (1.3 to 18) GHz 50 Hz to 100 khz rate 400 khz peak 1.8 % rdg + 1 digit w/ HP 11793A, HP 8673B 20 Hz to 200 khz rate 400 khz peak 5.9 % rdg + 1 digit Phase Modulation Measure 3 150 khz to 10 MHz 200 Hz to 10 khz rate 40 radians 5.9 % rdg + 1 digit HP 8902A measuring receiver (10 to 1300) MHz (1.3 to 18) GHz 200 Hz to 20 khz rate 400 radians 200 Hz to 20 khz rate 400 radians 5.1 % rdg + 1 digit 5.1 % rdg + 1 digit w/ HP 11793A, HP 8673B RF Absolute Power Measure 3 Power Reference 1 mw, Type-N (f), 50 Ω Coaxial Power Sensors 0.02 db (0.53 % rdg) HP 432A w/ HP 478A- H75 (+20 to -30) dbm, 75 Ω 100 khz to 2 GHz 0.08 db (1.8 % rdg) HP 8483A w/ HP 438A (+20 to -30) dbm, 50 Ω 100 khz to 4.2 GHz 0.08 db (1.7 % rdg) HP 8482A w/ HP 438A (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 9 of 16

RF Absolute Power Measure 3 (cont) Coaxial Power Sensors (+20 to -30) dbm, 50 Ω 10 MHz to 18 GHz 0.08 db (1.8 % rdg) HP 8481A w/ HP 438A (-20 to -70) dbm, 50 Ω (-20 to -70) dbm, 50 Ω 10 MHz to 18 GHz to 26.5 GHz 0.14 db (3.3 % rdg) 0.14 db (3.3 % rdg) HP 8484A w/ HP 438A HP 8481D w/ HP 438A CoaxialSensor Modules (+30 to -20) dbm 100 khz to 2.6 GHz 0.12 db (2.9 % rdg) HP 11722A w/ HP 8902A (+30 to -20) dbm Distortion Measure to 26.5 GHz 0.13 db (2.9 % rdg) HP 11792A w/hp 8902A (-99 to 0) db CW Input -13 db 20 Hz to 20 khz (20 to 100) khz 1.2 db 2.3 db HP 8903B distortion analyzer (-100 to 0) db CW Input -13 db 100 khz to 13.25 GHz 3.9 db HP 8563E spectrum analyzer Attenuation, Relative Measure 3 0 db, reference (0 to -10) db (-10 to -20) db (-20 to -30) db (-30 to -40) db (-40 to -50) db (-50 to -60) db (-60 to -70) db (-70 to -80) db (-80 to -90) db (-90 to -100) db (-100 to -110) db (-110 to -120) db (2.5 to 1300) MHz 0.01 db 0.14 db 0.14 db 0.15 db HP 8902A measuring Receiver w/ HP 83630B (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 10 of 16

Attenuation, Relative Measure 3 0 db, reference (0 to -10) db (-10 to -20) db (-20 to -30) db (-30 to -40) db (-40 to -50) db (-50 to -60) db (-60 to -70) db (-70 to -80) db (-80 to -90) db (-90 to -100) db (-100 to -110) db (-110 to -120) db 1300 MHz to 26.5 GHz 0.001 db 0.14 db 0.14 db 0.15 db 0.2 db HP 8902A measuring receiver w/ HP 11793A, HP 83630B RF Absolute Power Generate 3 Sine Wave into 50 Ω (10 to 3) V p-p Sine Wave into 50 Ω 0.001 Hz to 100 khz 0.007 db HP 3325B w/ BNC (f) function generator 2.99 V p-p to 1 mv p-p 0.001 Hz to 100 khz 0.004 db (10 to 3) V p-p 100 khz to 20 MHz 0.12 db 2.999 Vp-p to 1 mv p-p 100 khz to 10 MHz 0.012 db (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 11 of 16

RF Absolute Power Generate 3 (cont) Sine Wave into 50 Ω (2.999 to 0.1) V p-p (99.99 to 1) mv p-p Sine Wave into 50 Ω (10 to 20) MHz (10 to 20) MHz 0.12 db 0.12 db HP 3325B (to 20 MHz) function generator (+13 to -10) dbm (-10 to -30) dbm 100 khz to 2 GHz 100 khz to 2 GHz 1.8 dbm 1.2 dbm HP 83630 synthesized signal generator (-30 to -100) dbm 100 khz to 2 GHz 1.3 dbm (-100 to -143) dbm 100 khz to 2 GHz 1.3 dbm Sine Wave into 50 Ω (+13 to -10) dbm 10 MHz to 1 GHz (1 to 10) GHz (10 to 20) GHz (20 to 26.5) GHz 0.71 dbm 0.82 dbm 0.83 dbm 1.1 dbm Agilent 83630B synthesized sweptsignal generator (-10 to -50) dbm 10 MHz to 1 GHz (1 to 10) GHz (10 to 20) GHz (20 to 26.5) GHz 1.1 dbm 1.2 dbm 1.2 dbm 1.4 dbm (-50 to -143) dbm 10 MHz to 1 GHz (1 to 10) GHz (10 to 20) GHz (20 to 26.5) GHz 1.7 dbm 1.8 dbm 1.8 dbm 2 dbm (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 12 of 16

RF Attenuation Generate 3 Coaxial, 1 db Step (0 to 12) db Reference Reference 0.12 db 0.12 db 0.12 db HP 355C w/bnc (f) attenuation Coaxial, 10 db Step 10 db Reference Reference HP 355D 2/BNC (f) attenuation 20 db 30 db 40 db 50 db 60 db 70 db 80 db (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 13 of 16

RF Attenuation Generate (cont) Coaxial, 10 db Step 90 db HP 355D 2/ BNC (f) attenuator 100 db 110 db 0.25 db 0.25 db 0.25 db 120 db IV. Mechanical Parameter/Equipment Range CMC 2, 4 ( ) Comments Force Gage (0 to 200) lbf 0.057 % rdg + 0.58R Dead weights Vacuum (0 to 30) inhg 0.033 % rdg DPI 620 Pressure (0 to 3) psi (0 to 300) psi (0 to 1500) psi 0.075 % rdg 0.03 % rdg 0.035 % rdg DPI 620 with PM620 transducers (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 14 of 16

V. Thermodynamics Parameter/Equipment Range 5 CMC 2 ( ) Comments Temperature (-80 to 420) o C 0.12 o C Burns Eng. PRT Measure 3 12001, HP 3458A Relative Humidity Measuring Equipment 3 25.01% RH 32.8% RH 49.9% RH 53 % RH 75.6 % RH 0.74 % RH 0.89 % RH 0.84 % RH 0.89 % RH 0.82 % RH Vaisala HMP-155 and RH chamber VI. Time & Frequency Parameter/Equipment Range CMC 2 ( ) Comments Frequency Measuring Equipment 3 1 mhz to 20 MHz (1.9 x 10-10 ) Hz/Hz Fluke 910 w: HP 3325B 20 MHz to 3 GHz 3 GHz to 26.5 GHz (1.9 x 10-10 ) Hz/Hz (1.2 x 10-6 ) Hz/Hz HP 83630B Frequency Measure 3 1 mhz to 3 GHz (1.9 x 10-10 ) Hz/Hz Fluke 910 w: HP 53131A 3 GHz to 26.5 GHz (1.2 x 10-6 ) Hz/Hz HP 5351B 1 This laboratory offers commercial calibration service and field calibration service. 2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer s device and to influences from the circumstances of the specific calibration. (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 15 of 16

3 Field calibration service is available for this calibration and this laboratory meets A2LA R104 General Requirements: Accreditation of Field Testing and Field Calibration Laboratories for these calibrations. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer s site being larger than the CMC. 4 In the statement of CMC, R is the numerical value of the resolution of the device in microinches. (A2LA Cert. No. 1395.01) Revised 08/15/2016 Page 16 of 16

Accredited Laboratory A2LA has accredited SIMCO ELECTRONICS Greensboro, NC for technical competence in the field of Calibration This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005 General requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of ANSI/NCSLI Z540-1-1994, the requirements of ANSI/NCSLI Z540.3-2006, and any additional program requirements in the field of calibration. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated 8 January 2009). Presented this 5 th day of November, 2015. Senior Director, Quality and Communications For the Accreditation Council Certificate Number 1395.01 Valid to September 30, 2017 Revised August 15, 2016 For the calibrations to which this accreditation applies, please refer to the laboratory s Calibration Scope of Accreditation.