TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE OPTICAL COMPARATORS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 20 DATED 1 FEBRUARY 1988

Similar documents
DIAL AND VERNIER CALIPERS

TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE THREAD RING GAGES THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 143L DATED 1 OCTOBER 1992

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR THREADED PLUG GAGES. This publication replaces T.O. 33K dated 30 January 2013.

CLAMP-ON METERS THIS PUBLICATION SUPERSEDES NAVAIR 17-20AQ-32 DATED 1 OCTOBER 2009

TABLE OF CONTENTS SECTION I INTRODUCTION AND DESCRIPTION

INSTRUMENT CALIBRATION PROCEDURE GAGE BLOCKS

TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE PEAK POWER ANALYZER HEWLETT-PACKARD 8991AOPT003

Metrology Prof. Dr Kanakuppi Sadashivappa Bapuji Institute of Engineering and Technology Davangere. Lecture 24 Measurement of Screw Thread Element

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR GAGE BLOCKS MECHANICAL COMPARISON. This publication replaces T.O. 33K dated 30 December 2013.

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames TW18 3HR

OPTICAL BENCH - simple type

Dozuki. Written By: Dozuki System. Guide to calibrating the Haas wireless intuitive probing system. How to Calibrate WIPS

Fluid Sealing Association

Components of the Microscope

Metrology and instrumentation Indian Institute of Technology

Setup & Operating INSTRUCTIONS. for (FOR PIN FITTING AND ROD RECONDITIONING)

SWIFT SERIES M2252DGL MICROSCOPE

Each Height Master is supplied with a gauge block for reference-height setting. Fitted wooden case supplied. SPECIFICATIONS

How to Calibrate a CNC Machine's Positioning System

OMM300. Inverted Metallurgical Microscope

DETAIL SPECIFICATION SHEET TERMINAL, WIRE ROPE, SWAGING, FORK END

The old adage seeing is believing is appropriate when

Termination Procedure

نﺎﻤﻟ آ ﻲﻧﺎﭙﻤﻛ يرﺎﺼﺤﻧا هﺪﻨﻳﺎﻤﻧ

Field Inspection Gage. Instruction Manual

OxFAC-08 Angled Fiber Cleaver. User Manual. Issue 1.5

DETAIL SPECIFICATION SHEET

DETAIL SPECIFICATION SHEET

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

LINEAR MEASUREMENT. Prof.H M Prajapati & Prof. A R Sankhla

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

INSTALLATION INSTRUCTIONS

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

BHJ Products, Inc. Parts List & Instructions

FT-KV1SERIES FD-WL48

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER 34401A (HEWLETT-PACKARD)

Ring crush of paperboard (flexible beam method) (REAFFIRMATION of Classical Method T 818 cm-07) (No changes from previous drafts)

DETAIL SPECIFICATION SHEET

CIRRUS AIRPLANE MAINTENANCE MANUAL

ILFORD SPORTSVIEW PROJECTOR INSTRUCTION BOOK

DETAIL SPECIFICATION SHEET

Inside Micrometers and End Measuring Rods

3 Emergency Breakaway Coupling

These thickness gauges are especially handy for measuring thickness of small

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR MULTIMETERS

Understanding Optical Specifications

Additional Information

ROTARY TABLE OPERATION AND SERVICE MANUAL HORIZONTAL AND VERTICAL. Horizontal & Vertical. Rotary Table (HVRT) Tilting Rotary Table

ISH-R150 MANUAL ROCKWELL HARDNESS TESTER OPERATION MANUAL

MEASURING MACHINES. Pratt & Whitney METROLOGY LABORATORY. Measurement Systems, Inc.

DETAIL SPECIFICATION SHEET FITTINGS, PIPE, CAST BRONZE, SILVER-BRAZING; REDUCING COUPLING, BUSHING, AND ADAPTER, PIPE

SQ2 User Instructions SQ2 Overview:

Zoom Stereo Microscope NYMCS-360 Instruction Manual

Metrology Prof.Dr Kanakuppi Sadashivappa Bapuji Institute of Engineering and Technology Davangere

SUPERSEDED REVISION. Reasons for reissue of this instruction sheet are provided in Section 7, REVISION SUMMARY.

Evaluation Instructions Machining Level II Milling: Precision Locations

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

ASSIGNMENT 2. Textbook Assignment: 2-1. Levels are designed for which of the following purposes?

Arc Radius Measuring Gauge

Replacement of Pitch Link Retainer and Service Improvement of the Pitch Control System. Effectivity: Helicopters manufactured prior to January, 1981

Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual

MantelMount. TM1A Installation Instructions IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS

MM540 Installation Instructions IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR HIGH VOLTAGE INSTRUMENTS

Quality Assurance Supplier Reference Guide

TRUE TECHNICAL SERVICE MANUAL - ALL MODELS. DOORS/DRAWERS/LIDS

Optical Engineering 421/521 Sample Questions for Midterm 1

MM340 Installation Instructions IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS

Trade of Toolmaking Module 1: Induction & Bench Fitting Unit 6: Filing Internal & External Radii Phase 2

MaraMeter. Indicating Snap Gages

Procedure for Testing Direct Tension Indicators (DTI) Assemblies

STEINDORFF NYMC Polarizing Microscope

Digital scales description S_System LIN Precision measuring set for new development and retro-fitting Easy to integrate in your projects Small overall

Helpful Alignment Tips for Machine Shops

Frameless Inline Door With Return QCI5263

SURFACE VEHICLE STANDARD

Nikon E800 Operating Instructions.

PERFORMANCE SPECIFICATION SHEET ELECTRON TUBE, NEGATIVE GRID (MICROWAVE) TYPE 8727

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NSCL Z

AUC Cell Assembly Torque Stand. User Manual

Instruction Sheet MCHE 365 & I-Tech 344 Lathe & Mill Machining Operations Pencil Organizer Project, FALL 2015

MADE TO MEASURE. MARGAGE.

Table of Contents. Table of Contents. Preface 11 Prerequisites... 12

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Milling Machine Operations

Calibration of reflectance standards for hemispherical geometry

Duties and Standards. for. Screw Machining--Level III

INSTRUCTION MANUAL SM-330/ 430/ 936/ (Semi-Automatic Divider Rounder)

Certificate of Accreditation

DETAIL SPECIFICATION SHEET CONNECTORS, RECEPTACLE, ELECTRICAL, FLANGE MOUNTING, MINIATURE, CLASSES E, F, G AND R

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

Cylindrical Set Masters Universal Setting Masters. Application Series Page. Setting Discs in Steel and Chrome 801 F.2

CLEANING FOR BETTER OBSERVATION AND PHOTOMICROGRAPHY

Instruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope

DETAIL SPECIFICATION SHEET TERMINAL, WIRE ROPE, SWAGING, STUD

MILITARY SPECIFICATION LIGHTING, INSTRUMENT, INTEGRAL, WHITE GENERAL SPECIFICATION FOR

DETAIL SPECIFICATION SHEET

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR LCR METER 875A, 875B (B & K PRECISION) This publication replaces T.O. 33K dated 30 August 2011.

DETAIL SPECIFICATION SHEET

Transcription:

TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE OPTICAL COMPARATORS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 20 DATED 1 FEBRUARY 1988 DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND THEIR CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMINIS- TRATIVE OR OPERATIONAL PURPOSES ONLY. OTHER REQUESTS FOR THIS DOCUMENT SHALL BE RE- FERRED TO NAVY MEASUREMENT SCIENCE DIRECTORATE, NAVAL WARFARE ASSESSMENT DIVISION, P.O. BOX 5000 CORONA, CA 91718 5000 DESTRUCTION NOTICE FOR UNCLASSIFIED, LIMITED DOCUMENTS, DESTROY BY ANY METHOD THAT WILL PREVENT DISCLOSURE OF CONTENTS OR RECONSTRUCTION OF THE DOCUMENT. PUBLISHED BY DIRECTION OF COMMANDER NAVAL AIR SYSTEMS COMMAND 1 OCTOBER 1993

LIST OF EFFECTIVE PAGES Insert latest changed pages; dispose of superseded pages in accordance with applicable regulations. NOTE: On a changed page, the portion of the text affected by the latest change is indicated by a vertical line, or other change symbol, in the outer margin of the page. Page No. Change No. * Change Date Page No. Change No. * Change Date Title 0 A 0 i 0 1 to 7 0 8 (blank) 0 9, 10 0 A 1 0 * Zero in this column indicates an original page. A

TABLE OF CONTENTS Section Page 1 Introduction and Description...................................................... 1 2 Equipment Requirements......................................................... 2 3 Preliminary Operations.......................................................... 3 3.1 Preliminary Conditions......................................................... 3 3.2 Preliminary Inspection......................................................... 3 4 Calibration Process.............................................................. 4 4.1 Light Beam Adjustment........................................................ 4 4.2 Table Angular Alignment Test................................................... 5 4.3 Magnification Tests............................................................ 5 4.4 Horizontal Micrometer Tests.................................................... 5 4.5 Vertical Micrometer Tests....................................................... 6 4.6 Screen Adjustment............................................................ 7 Navy Calibration Checklist....................................................... 9 Appendix A: Representative Test Instruments........................................ A 1 ILLUSTRATIONS Figure Page 1 Typical TI Component Location and Indentification.................................... 4 TABLES Table Page 1 Calibration Description.......................................................... 1 i

SECTION 1 INTRODUCTION AND DESCRIPTION 1.1 This procedure describes the calibration of Optical Comparators. Instruments that can be calibrated by this procedure include, but are not limited to the models listed in Appendix A. The instrument being calibrated is referred to herein as the TI (Test Instrument). 1.2 All comments concerning this procedure should be directed to Navy Measurement Science Directorate, Naval Warfare Assessment Division, P.O. Box 5000, Corona, CA 91718 5000. 1.3 This procedure includes tests of essential performance parameters only. Any malfunction noticed during calibration, whether specifically tested for or not, should be corrected. TI Characteristics Light beam adjustment Magnification Horizontal micrometer Vertical micrometer Table angular alignment Screen adjustment Table 1. Calibration Description Performance Specifications Collimated and centered on screen 5X, 10X, 20X, 31.25X, 50X, 62.5X, 100X, and 125X Range: 0 to 1 inch Tolerance: +0.0001 inch Range: 0 to 2.5 inches Tolerance: +0.0001 inch Within +0.000291 inch/inch of lens mount diameter Alignment fixture shadow to remain in contact with horizontal screen line Test Method Lamp filament projected on screen and adjusted as required Comparison with ball and pin chart magnification checker Comparison with gage blocks Comparison with gage blocks Measured with a dial indicator Fixture moved horizontally before the projection lens 1

SECTION 2 EQUIPMENT REQUIREMENTS NOTE Minimum use specifications are the principal parameters required for performance of the calibration, and are included to assist in the selection of alternate equipment, which may be used at the discretion of the using laboratory. Satisfactory performance of alternate items shall be verified prior to use. All applicable equipment must bear evidence of current calibration. The instruments utilized in this procedure were selected from those known to be available at Navy calibration facilities, and the listing by make or model number carries no implication of preference, recommendation, or approval for use by other agencies. It is recognized that equivalent equipment produced by other manufacturers may be capable of equally satisfactory performance in this procedure. Item 2.1 Gage block set 2.2 Magnification checker 2.3 Dial indicator 2.4 Universal surface gage 2.5 Steel parallels 2.6 Steel toolmakers flat Minimum Use Specifications Style 2, Grade 2, Class I Range: 0.2 to 3 inches Uncertainty: +10 in. Four prevision ground balls and one pin. Pin 0.0625 inch diameter Balls 0.1875, 0.250, 0.625, and 1.000 inch diameters Range: 0 to 0.008 inch Uncertainty: 0.0001 inch To hold dial indicator Riser for support of V block 1 to 1 1/4 inch Size: 6 inch diameter X 1 inch thick Flatness: 0.000010 inch Calibration Equipment Pratt & Whitney Co. Set #88; or Webber RS88A1 Optical Gaging Products, Inc. Magnification checker catalog #20290 or 120290 Brown & Sharpe Mfg Co. 599 7032 or 599 7032 3 or 599 7035 3 Starrett 257C; or Brown & Sharpe 599 620 42; or Federal 2400 Taft Peirce Mfg Co. Set 9150C Acme Scientific Co. 6 inch diameter X 1 inch thick; or Van Keuren 23 006IG 2

SECTION 3 PRELIMINARY OPERATIONS 3.1 PRELIMINARY CONDITIONS 3.1.1 Ensure that the work area is clean, free from drafts, free from excessive humidity, and that the rate of temperature change does not exceed 2F per hour. 3.1.2 Ensure that the calibration equipment is clean, and that the TI and the applicable calibration equipment have been allowed to stabilize at the ambient temperature. 3.2 PRELIMINARY INSPECTION 3.2.1 Verify that all of the TI surface components are clean and free from damage such as burrs, score marks, etc., that would impair their operation, and take any corrective action, if necessary, before proceeding. 3.2.2 Ensure that the TI glass viewing screen and mirror are clean and free from excessive scratches, chips, fractures, or other damage. If necessary, clean the glass surfaces as follows: 3.2.2.1 Remove all loose foreign material by brushing lightly and carefully with a clean camel s hair brush or by using air forced from a small rubber syringe. CAUTION USE LENS CLEANING FLUID VERY SPARINGLY. IT MAY DETERIORATE COMPONENTS IN CONTACT WITH THE GLASS SURFACES. 3.2.2.2 If additional cleaning is required, lightly and gently wipe the glass surfaces with a clean lens tissue moistened with an approved lens cleaning fluid, then dry by lightly patting with a clean, dry lens tissue. NOTE When wiping a window, prism, or mirror surface, always wipe from the center to the outer edge when the surface is horizontal, or from the top to the bottom when the surface is vertical. To minimize the possibility of scratching the surface with any foreign material that may be picked up on the lens tissue, constantly change the working portion of the tissue. 3.2.2.3 If necessary, rinse the lens cleaning fluid from the glass surfaces by repeating step 3.2.2.2, substituting distilled water for the cleaning fluid, until all of the cleaning fluid is removed. 3

SECTION 4 CALIBRATION PROCESS NOTE Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met, before proceeding. 4.1 LIGHT BEAM ADJUSTMENT 4.1.1 Remove the TI magnification lens and install the TI light aligning diaphragm in the lens bushing. 4.1.2 Turn the TI line switch on. 4.1.3 Turn the TI projection lamp switch on. An image of the lamp filament or arc will appear on the screen. 4.1.4 Adjust the condensing lens assembly to obtain an image that is focused as sharply as possible. Figure 1. Typical TI Component Location and Identification 4

4.1.5 Verify that the lamp filament or arc image is centered on the screen both vertically and horizontally. If necessary, adjust the TI lamp adjusting screws on the lamp house assembly. 4.1.6 Open the TI iris diaphragm fully. (This is located in the condensing lens assembly). 4.1.7 Verify that the TI light beam is collimated from the condensing lens to the light aligning diaphragm. To test the beam, place a white card directly in front of the condensing lens. Measure the diameter of the lighted area of the card. Move the card to place it against the lens bushing. Verify that the lighted area diameter is still the same. If not, move the condensing lens assembly as necessary so that the lighted area in both positions is equal. 4.1.8 Remove the TI light aligning diaphragm from the lens bushing. NAVAIR 17 20MD 20 4.2 TABLE ANGULAR ALIGNMENT TEST 4.2.1 Remove the TI projection lens from the lens mount. 4.2.2 Attach the dial indicator to the universal surface gage. 4.2.3 Adjust the TI table angle vernier to indicate zero. 4.2.4 Place the surface gage on the TI table and indicate from one side of the lens mount to the other side across the flat surface by moving the table from side to side. Verify that the dial indicator indication is within the tolerance limits. 4.3 MAGNIFICATION TESTS 4.3.1 Position the TI 10X magnification lens into the lens housing and rotate the TI stop pin spool as required to position the 10X stop pin. 4.3.2 Place the MAGNIFICATION CHECKER on the TI work table so that the applicable ball or pin may be observed on the screen. 4.3.3 Adjust the TI vertical, horizontal, and focusing controls so that a sharp outline (shadow) of the 1.000 inch steel ball is centered on the screen. 4.3.4 Position the MAGNIFICATION CHECK SCALE over the outline on the TI screen and verify that the outline is within the SCALE markings. 4.3.5 Repeat steps 4.3.1 through 4.3.4 except using the remaining magnification lens systems and stop pin spools and an appropriate size ball or pin to verify that the outline on the TI screen is within the scale markings for each lens. 4.4 HORIZONTAL MICROMETER TESTS 4.4.1 Set the TI horizontal micrometer to 0.0000. 4.4.2 Place a 1.000 inch gage block on the TI table with the gage block length parallel to the table length. 4.4.3 Adjust the TI vertical handwheel so that the gage block is centered vertically on the screen. 4.4.4 Adjust the TI focus control to obtain a sharply defined shadow outline of the gage block. 5

4.4.5 Adjust the TI lateral handwheel so that the right edge of the gage block is tangent to the vertical center line of the screen. 4.4.6 Move the TI adjustable anvil into contact with the micrometer and clamp the anvil to the table. 4.4.7 Push the TI table away from the micrometer spindle and insert a combination of gage blocks in the V trough between the anvil and micrometer to obtain the values in the following table. Allow the table to return so that the TI anvil clamps the gage blocks to the micrometer. NOTE The gage blocks used as a combination must be wrung together. Do not use more than three gage blocks in any combination. 4.4.8 Adjust the TI micrometer so that the right edge of the gage block is tangent to the vertical center line of the screen. Verify that each TI micrometer indication is within the tolerance limits. TI Micrometer Setting and Gage Block Combination (inch) Tolerance Limits (inch) 0.2050 0.2049 to 0.2051 0.4100 0.4099 to 0.4101 0.6150 0.6149 to 0.6151 0.8200 0.8199 to 0.8201 1.0000 0.9999 to 1.0001 4.5 VERTICAL MICROMETER TESTS 4.5.1 Place the toolmakers flat on the TI table. 4.5.2 Wring the 0.500 inch gage block to the toolmakers flat, and position the gage block so that its shadow will occupy one half of the screen. 4.5.3 Adjust the TI lateral handwheel to obtain a sharply defined image on the screen. 4.5.4 Adjust the TI vertical handwheel to place the top of the 0.5 inch gage block image tangent to the horizontal line of the screen. 4.5.5 Loosen the screw on the TI vertical graduated ring, adjust the ring to align the zero mark with the fixed pointer, and tighten the screw. 4.5.6 Position each of the gage blocks, in turn, listed in the following table beside the 0.500 inch gage block and wring the gage block to the toolmakers flat. Ensure that the gage block is positioned so that a sharply defined image is displayed on the screen. 6

4.5.7 Adjust the TI vertical handwheel to place the top of the gage block image tangent to the same horizontal line of the screen referenced in step 4.5.4. Verify that the vertical handwheel graduations are within the tolerance limits. NAVAIR 17 20MD 20 TI Vertical Handwheel Graduations Gage Blocks (inch) Tolerance Limits (inch) 0.5000 1.000 0.4999 to 0.5001 1.5000 2.000 1.4999 to 1.5001 2.5000 3.000 2.4999 to 2.5001 4.6 SCREEN ADJUSTMENT 4.6.1 Place alignment fixture on table top with its tongue in the table groove. 4.6.2 Move table vertically until the top of the vertical pin is approximately in the axis of the lens system. 4.6.3 Use focus control to project sharpest image of the fixture on screen. 4.6.4 Zero the graduated chart ring on the vernier. 4.6.5 Rotate the chart within the ring until the projected radius pin image remains tangent to the horizontal centerline when the test pin is moved laterally in front of lens. 4.6.6 When the alignment is complete, secure screen with clamp if equipped. ( 8 blank) 7

CALIBRATION CHECKLIST TEST INST (S) Jones & Lamson PC 14, PC 14A, or TC 10D Optical Comparators PROC. NO. PROCEDURE STEP NO. (1) NA 17 20MD 20 FUNCTION TESTED (2) MFG. MODEL SER. NO. NOMINAL 3.2 Preliminary Inspection 3.2.1 Clean and undamaged ck ( ) 3.2.2 Screen and mirror ck ( ) (3) MEASURED VALUES FIRST RUN SECOND RUN (4) (5) OUT OF TOL (6) CALIBRATION TOLERANCES (7) 4.1 Light Beam Adjustment 4.1.5 Vert & horiz center ck ( ) 4.1.8 Parallel test ck ( ) 4.2 Table Alignment Test 4.2.4 Indicator 0.000 in. +0.000291 inch 4.3 Magnification Tests 4.3.5 5 X (PC 14A & TC10) ck ( ) 10 X ck ( ) 20 X ck ( ) 31.25 X ck ( ) 50 X ck ( ) 62.5 X ck ( ) 100 X ck ( ) 125 X (PC 14A & TC10) ck ( ) 4.4 Horizontal Micrometer Tests 4.4.8 Micrometer reading0.2050 in. 0.2049 to 0.2051 inch 0.4100 in. 0.4099 to 0.4101 inch 0.6150 in. 0.6149 to 0.6151 inch 0.8200 in. 0.8199 to 0.8201 inch 1.0000 in. 0.9999 to 1.0001 inch 4.5 Vertical Micrometer Tests 4.5.7 Displacement Ring 0.5000 0.4999 to 0.5001 inch 4.5.8 1.5000 1.4999 to 1.5001 inch 2.5000 2.4999 to 2.5001 inch Page 1 of 2 9

CALIBRATION CHECKLIST TEST INST (S) Jones & Lamson PC 14, PC 14A, or TC 10D Optical Comparators PROC. NO. PROCEDURE STEP NO. (1) NA 17 20MD 20 FUNCTION TESTED (2) MFG. MODEL SER. NO. NOMINAL (3) MEASURED VALUES FIRST RUN SECOND RUN (4) (5) OUT OF TOL (6) CALIBRATION TOLERANCES (7) 4.6 Screen adjustment 4.6.5 Horizontal contact ck ( ) 10 Page 2 of 2

APPENDIX A REPRESENTATIVE TEST INSTRUMENTS A.1 TI s that can be calibrated by this procedure include, but are not limited to, the following: Manufacturer Model Nomenclature Jones & Lamson FC 14 Optical Comparator FC 30 PC 14 PC 14A TC 10A TC 10D TC 14 TOPIC 14 E 114 A 1