Introduction to Charge Mode Accelerometers

Similar documents
Introduction to LIVM Accelerometers

Technical Information

Anthony Chu. Basic Accelerometer types There are two classes of accelerometer in general: AC-response DC-response

CHOOSING THE RIGHT TYPE OF ACCELEROMETER

Laboratory 6. Lab 6. Operational Amplifier Circuits. Required Components: op amp 2 1k resistor 4 10k resistors 1 100k resistor 1 0.

A high temperature 100 mv/g triaxial accelerometer. Endevco technical paper 329

Laboratory 9. Required Components: Objectives. Optional Components: Operational Amplifier Circuits (modified from lab text by Alciatore)

Shock sensor PKGS series Application manual of peripheral circuit Feb. 3, 2003 Example circuit for charge sensitivity type shock sensor.

Linear electronic. Lecture No. 1

Lesson number one. Operational Amplifier Basics

OPERATING GUIDE MODEL 3093M32 MINIATURE HIGH SENSITIVITY TRIAXIAL LIVM ACCELEROMETER WITH SINGLE 4-PIN CONNECTOR INTERNALLY CASE GROUND ISOLATED

Operational Amplifier BME 360 Lecture Notes Ying Sun

Department of Mechanical Engineering

Chapter.8: Oscillators

Dytran Instruments, Inc. 1

Chapter 9: Operational Amplifiers

1. Consider the closed loop system shown in the figure below. Select the appropriate option to implement the system shown in dotted lines using

EET 438a Automatic Control Systems Technology Laboratory 1 Analog Sensor Signal Conditioning

UNIVERSITI MALAYSIA PERLIS

Design of Accelerometer Pre-regulation Circuit and Performance Analysis of the Key Components

INTER PLANT STANDARD STEEL INDUSTRY

Chapter 2. Operational Amplifiers

Vibration Isolation for Scanning Tunneling Microscopy

SC Series. SC Series High Voltage Power Supply

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

Basic Electronics Learning by doing Prof. T.S. Natarajan Department of Physics Indian Institute of Technology, Madras

Chapter 9: Operational Amplifiers

Physics 303 Fall Module 4: The Operational Amplifier

C H A P T E R 02. Operational Amplifiers

GUIDE TO DYNAMIC FORCE SENSORS

Group: Names: (1) In this step you will examine the effects of AC coupling of an oscilloscope.

Laboratory 3 (drawn from lab text by Alciatore)

PSD Characteristics. Position Sensing Detectors

Analytical Chemistry II

ASC IMU 7.X.Y. Inertial Measurement Unit (IMU) Description.

PHYS 536 The Golden Rules of Op Amps. Characteristics of an Ideal Op Amp

ICL MHz, Four Quadrant Analog Multiplier. Features. Ordering Information. Pinout. Functional Diagram. September 1998 File Number 2863.

Ground. Input: 0-24VDC

Special-Purpose Operational Amplifier Circuits

Velocity and Acceleration Measurements

26 Endevco Isotron accelerometers

Models Z7, Z11, Z602WA and Z820WA Impedance head operating guide

Group: Names: voltage calculated measured V out (w/o R 3 ) V out (w/ R 3 )

Oscillators. An oscillator may be described as a source of alternating voltage. It is different than amplifier.

Figure 1: Closed Loop System

HAQ Series High Temperature High Voltage Power Supply

Lecture 28 RC Phase Shift Oscillator using Op-amp

Homework Assignment 03

ELT 215 Operational Amplifiers (LECTURE) Chapter 5

Analysis of Electrical Noise in Piezoelectric Sensors

Operational amplifiers

VCSO Mechanical Shock Compensation

2775AM4 Signal conditioner Instruction manual. IM2775AM4, Revision BC

Last Time. P and N type semiconductors Diode internals Transistors NPN PNP

Chapter 1 Introduction to the MAE 221 Laboratory

5. Transducers Definition and General Concept of Transducer Classification of Transducers

PAiA 4780 Twelve Stage Analog Sequencer Design Analysis Originally published 1974

SmartSenseCom Introduces Next Generation Seismic Sensor Systems

Chapter 2. Operational Amplifiers

731A seismic accelerometer and P31 power unit/amplifier Operating guide

GENERAL PURPOSE PIEZOELECTRIC LOAD CELL

ACTUATORS AND SENSORS. Joint actuating system. Servomotors. Sensors

High-Speed, Low-Power Dual Operational Amplifier AD826

ELEC207 LINEAR INTEGRATED CIRCUITS

LOW COST SDI 2210, 2260 & 2266 HIGH PERFORMANCE SDI 2220 & 2276

Introduction To Impulse Hammers

+ power. V out. - power +12 V -12 V +12 V -12 V

UNIT I. Operational Amplifiers

UNIT-I CIRCUIT CONFIGURATION FOR LINEAR

4/30/2012. General Class Element 3 Course Presentation. Practical Circuits. Practical Circuits. Subelement G7. 2 Exam Questions, 2 Groups

Low Cost 100 g Single Axis Accelerometer with Analog Output ADXL190*

Circuit Applications of Multiplying CMOS D to A Converters

Operational Amplifiers

Lab 4: Transmission Line

Analog Circuits Prof. Jayanta Mukherjee Department of Electrical Engineering Indian Institute of Technology-Bombay

Exercise 9: inductor-resistor-capacitor (LRC) circuits

Acceleration Sensor AS - 022

When input, output and feedback voltages are all symmetric bipolar signals with respect to ground, no biasing is required.

Chapter 7. Response of First-Order RL and RC Circuits

Interface Electronic Circuits

OPERATIONAL AMPLIFIERS (OP-AMPS) II

MAHALAKSHMI ENGINEERING COLLEGE TIRUCHIRAPALLI

Table of Contents Lesson One Lesson Two Lesson Three Lesson Four Lesson Five PREVIEW COPY

EE2210 Laboratory Project 1 Fall 2013 Function Generator and Oscilloscope

Voltage Controlled SAW Oscillator Mechanical Shock Compensator

Lecture #4 Basic Op-Amp Circuits

DEPARTMENT OF ELECTRICAL ENGINEERING LAB WORK EE301 ELECTRONIC CIRCUITS

MFJ 259 Operation & Simplified Calibration

EE301 ELECTRONIC CIRCUITS CHAPTER 2 : OSCILLATORS. Lecturer : Engr. Muhammad Muizz Bin Mohd Nawawi

CEC VIBRATION TRANSMITTER

Week 8 AM Modulation and the AM Receiver

For the filter shown (suitable for bandpass audio use) with bandwidth B and center frequency f, and gain A:

Operational Amplifiers

MAE334 - Introduction to Instrumentation and Computers. Final Exam. December 11, 2006

Analog front-end electronics

using dc inputs. You will verify circuit operation with a multimeter.

Techniques for Passive Circuit Analysis for. State Space Differential Equations

Industrial vibration sensor selection: piezovelocity transducers

Lab 7: DELTA AND SIGMA-DELTA A/D CONVERTERS

Tactical grade MEMS accelerometer

Transcription:

Introduction to Charge Mode Accelerometers Dytran charge mode accelerometers are designed to measure shock and vibration phenomena over a broad temperature range. These accelerometers, unlike the Low Impedance Voltage Mode (LIVM) types, contain no built-in amplifiers. Dytran s charge mode accelerometers utilize high sensitivity piezoceramic crystals, of the lead zirconate titinate (PZT) family, to produce a relatively high charge output in response to stress created by input vibration or shock acting upon the seismic system. Because of the high impedance level of the charge mode signal generated by the crystals, a special type of amplifier, called a charge amplifier, is used to extract the very high impedance electrostatic charge signal from the crystals. The charge amplifier has the ability to convert the charge signal to a low impedance voltage mode signal without modifying it. Figure 1 is a cross-section of a typical charge mode compression design accelerometer, model 3100C6. The sensitivity is 100 pc/g (pc = pico coulomb = 1 x 10-12 Coulomb) and the useful frequency range is up to 5 khz. The 3100C6 operates at temperatures up to +500 F. A heavy metal seismic mass is preloaded against the piezoceramic crystals with an elastic preload screw. The mass converts the input acceleration into analogous stress on the crystals producing an output charge signal in direct proportion to instantaneous acceleration.

When to Use Charge Mode Accelerometers The question may be asked, When should I consider using charge mode accelerometers vs. LIVM types with built-in electronics? Charge mode accelerometers should be considered: 1. when making measurements at temperatures above +250 F, the maximum temperature for most LIVM instruments, 2. when the versatility of the laboratory charge amplifier is desired for system standardization, ranging, filtering, integrating for velocity and displacement, etc. and, 3. when adding or replacing accelerometers where existing charge amplifiers must be used for economic or other reasons. Two System Concepts Charge mode accelerometers may be combined with a variety of electronic components to create two basic measurement system classifications: 1. The conventional charge mode system, and 2. The Hybrid system. The conventional charge mode system utilizes a sophisticated laboratory charge amplifier while the hybrid system features simple dedicated range miniature in-line charge and voltage amplifiers operating in conjunction with LIVM current source power units. The Conventional Charge Mode System The versatile laboratory charge amplifier is the main feature of the conventional charge mode system. This section will familiarize you with the theory, operating characteristics and features of the basic laboratory charge amplifier. Figure 2 illustrates a laboratory charge amplifier, model 4165 in use with a model 3100C6 charge mode accelerometer. Series 6019A low-noise coaxial cable is used to minimize

triboelectric noise generated by cable motion. This is a very versatile system whose signal conditioning options include: 1. Standardization of system sensitivity 2. Full Scale range selection 3. Discharge time constant choices 4. Filter options 5. 0-10 VDC out for sinusoidal input 6. Overload indication 7. Instant system zeroing (or reset) 8. External calibrate signal insertion 9. Front panel meter for observation of DC level of output signal Charge Amplifier, Basic Theory A charge amplifier is a special high gain, high input impedance inverting voltage amplifier with capacitive feedback. The amplifier is usually an operational amplifier (op-amp) with near infinite voltage gain. Referring to Figure 3, the input charge qin is applied to the summing junction (inverting input) of the charge amplifier and is distributed to the input capacitance of the amplifier CA and the feedback capacitor Cf. We may write the equation: Using the electrostatic equation q = Cv and substituting in equation 1:

Using equation 2 and making the appropriate substitutions and solving for the output voltage of the amplifier in terms of input charge, amplifier loop gain, and input and feedback capacitance we have: Now, letting gain A approach infinity, we have: This result (Eq 4) shows clearly that the transfer function (gain) of a charge amplifier is a function only of the value of the feedback capacitor Cf. Notice that input capacitance CA has no effect on the sensitivity of the charge amplifier. This means that cable capacitance, for example, has no effect on the sensitivity, a significant find when switching cable lengths and types. Adding Versatility to The Charge Amplifier Standardization of system sensitivity, say to exactly 10.00 or 100.00 mv/g is accomplished by adjusting the amount of feedback with a potentiometer as shown. Standardization is often necessary because accelerometers are rarely ever made to exact sensitivities. The use of a precision multi-turn potentiometer with turns counting dial allows the standardization of the system sensitivity by dialing in the accelerometer sensitivity. Referring to Figure 3b, changing the range of the charge amplifier is accomplished by switching various values of feedback capacitor into the feedback path. This is accomplished with a rotary switch which has maybe 10 values of precision capacitor arrayed around it. Again referring to Figure 3b, feedback resistor Rf gives DC stability to the circuit and establishes the discharge time constant (TC) of the amplifier thereby setting the low frequency response of the amplifier. A momentary reset switch (S1) discharges the residual charge in the feedback capacitor returning the system output to zero. The circuit shown in Figure 3b represents only the first stage of a rudimentary charge amplifier. One or more stages of filtering, integration and other features can be added all in one compact package.

The Hybrid System A Dytran hybrid system combines charge mode accelerometers with miniature in-line fixed sensitivity charge amplifiers. These charge amplifiers are powered by standard 2-wire LIVM power units. The hybrid system, (refer to Figure 4) is ideal for field use because of its small size and rugged construction of the miniature charge amplifiers, for example, Models 4751B and 4505A. These amplifiers are powered by conventional LIVM constant current power units and transmit the output signal over the same two wires as do conventional LIVM systems. The power unit separates the signal information from the DC bias of the amplifier and couples it to the readout instrument. As with conventional charge mode systems, low noise coaxial cable is used to couple the accelerometer to the charge amplifier to minimize triboelectric noise. When to Use The Hybrid System The hybrid system should be considered when: 1. The accelerometer is used to measure events at a temperature which is above that recommended for LIVM instruments, i.e., instruments that have built-in amplifiers and the environment is not favorable to laboratory charge amplifiers. 2. System cost is an important factor. Per channel cost of the hybrid system is a fraction of that for the conventional charge amplifier system. 3. LIVM current source power units are in-hand and must be utilized.

4. Ruggedness and small size of the measurement system is imperative. 5. A dedicated system without the versatility of a laboratory charge is sufficient for the measurement task.