ARU-01CF Portable X-Ray NSN:

Similar documents
Stationary Anode X-Ray Tube. General Data

Rotating Anode X-Ray Tube Housing Assembly. General Data

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I

ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Electrical :

Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL. Version 1.0

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Type B

Product Information PI

Rotating Anode X-ray Tube

Rotating Anode X-ray Tube Assembly

OPERATOR'S INSTRUCTIONS

Printed on: 09 Apr 2013, 07:01:56 pm; Printed by: TOBIAS BAUER

CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

GE AMX 4+ Portable X-Ray

Product Information PI

OPERATOR S INSTRUCTIONS (for USA & Canada)

P R E S E N T E D B Y. K A M A R U L A M I N A B D U L L A H Dip. MED. IMG., BSc. MED. IMG. (UiTM)

OPERATORS MANUAL AND INSTALLATION AND SERVICE MANUAL UNIVERSAL UNI-MATIC 325D GENERATOR MODEL NO. 3488/3550

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

Product Information PI

PROVECTA V. Veterinary X-Ray System. Operator s Manual

Quality Design Because you pay more and more attention to the look and effectiveness of your practice

Minnesota Rules, Chapter 4732 X-ray Revision

Digital Panoramic X- Ray unit. Product Data

Product Information PI

MODEL 3810/2 Line Impedance Stabilization Network

Industrial X-ray Tube KL

ADAE portable X ray machine Operation manual

MaxRay Handheld X-ray Systems Operator Training Exam

PET1606J2F. Pilani Electron Tubes & Devices Pvt. Ltd. Water Cooled Triode. For Industrial RF Heating. Drop in equivalent of BW1606J2F

DENTAL X-RAY OPERATOR'S INSTRUCTIONS. (for USA) Wall Mount Type...WK

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

Minnesota Rules, Chapter 4732 X-ray Revision

PORTABLE HALOGEN WORK LIGHT OWNER S MANUAL

WRM-10 TM TRANSFORMER WINDING RESISTANCE METER

Neo Ultrasound Module Manual

TABLE OF CONTENTS PAGE 1.0 INTRODUCTION

KODAK 2200 Intraoral X-ray System. User s Manual

DCM20 Series. Three-Function DC Power Meters. DCM20 DISPLAY PRODUCT OVERVIEW FEATURES

X-band Magnetron. Cooling (note 5) Water Output coupling (note 6) UG51/U Magnet (note 7) Integral, Permanent

Current Probes. User Manual

MG5223F S-Band Magnetron

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS

Minnesota Rules, Chapter 4732 X-ray Revision

Rotating Anode X-Ray Tube Housing Assembly. General Data

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer

(a) This subchapter applies to installations using analytical x-ray equipment and establishes requirements for their use.

V50D ELECTRIC GUITAR AMPLIFIER WITH DIGITAL DELAY AND CHORUS. User s Manual

TABLE OF CONTENTS PAGE

Elderfield & Hall, Inc., Kama Bandsaw AD 105S. Instruction Manual: Introduction to the Manual. General Precautions. Equipment. Machine.

OPERATOR'S INSTRUCTIONS DENTAL X-RAY MODEL 096

LPX SERIES Constant Potential Portable X-Ray

Nuclear Associates

E2V Technologies MG5223F S-Band Magnetron

Product Information PI

BCA 9SK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

Dynamo Brushless DC Motor and GreenDriveTM Manual

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority

TABLE OF CONTENTS. References

OPERATOR'S MANUAL JUNE 2017

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

Model #2 UVB Midband Ultraviolet Lamp Assembly with Timer

GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier

Angle Grinder. Model Visit our website at:

VACUUM INTERRUPTER APPLICATION NOTES Filename: VIAN X-Rays and Vacuum Interrupters Revision: 0 PAGE 1

MOXTEK. 50kV 10 Watt MAGNUM X-ray Source. X-ray Sources. Contents

Minnesota Rules, Chapter 4732 X-ray Revision

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

CAUTION: WEIGHT ON THIS PRODUCT SHOULD NOT EXCEED 136KG / 300LBS.

POWER SUPPLY MODEL XP-720. Instruction Manual ELENCO

X band Magnetron. Water: Anode cavity Forced-air: Input ceramics and terminals Output coupling (note 6) UG51/U Magnet (note 7) Integral, Permanent

V160R ELECTRIC GUITAR AMPLIFIER WITH A TUBE PREAMP, REVERB AND BUILT-IN CHROMATIC TUNER. User s Manual

1.0 INTRODUCTION WARNINGS 2 DUTY CYCLE 2 STORAGE PHYSICAL DESCRIPTION 3 HIGH VOLTAGE PULSER/TUBEHEAD 3 BASE 3 BATTERY PACK 3

E2V Technologies M5187F X-Band Magnetron

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

Nuclear Associates

CD-Pxx-00-0 Series Duct Mount CO 2 Transmitter

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

ATV CULTIVATOR OWNER S MANUAL

OPERATOR'S MANUAL MARCH Golden Engineering...-- Portable X-ray Technology

BCA 9RK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

Tube Facing Tool.

OPERATION, PARTS & MAINTENANCE MANUAL MODELS HB73-16 HB97-18 HB97-16 HB97-12 HB HB HB HB145-18

4. CONNECTIONS. Make sure all connections to and in your Universal Rectifier are tight.

English. Where quality becomes beauty

1. Oklahoma Administrative Code (OAC) 252:410, titled Radiation Management

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

PET1610F. Pilani Electron Tubes & Devices Pvt. Ltd. Forced-Air Cooled Triode. For Industrial RF Heating. Drop in equivalent of BW1610F

OPERATOR'S INSTRUCTIONS DX-073 DENTAL X-RAY

ELECTRONIC SHEAR PIN MODEL ESP 1 & 2 INSTRUCTION MANUAL

TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS

Service Instructions. The Conductor Controls. Conductor DC15-A, Enclosed Unit CH15-A, Open Chasis Unit

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

TECHNIlCAL DATA. Amperes Cathode Heating Time..300

MILADY. Product Data. Page 1 of 8

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

1.0 INTRODUCTION WARNINGS PHYSICAL DESCRIPTION CONTROL MODULE DESCRIPTION OF OPERATION OPERATING INSTRUCTIONS 7

8121 Power Tube. Linear Beam Power Tube

OPERATING AND MAINTENANCE INSTRUCTIONS FOR CONSTANT CURRENT RECTIFIERS USING SATURABLE CORE REACTORS

Wireless Room Thermostat Installation and Operation Y6630D1007

Transcription:

ARU-01CF Portable X-Ray NSN: 6525-01-468-1672 COMMAND l AIR OPERATION & MAINTENANCE MANUAL

Table of Contents Introduction................................................................3 Purpose.............................................................3 Performance Characteristics............................................3 Items Furnished......................................................3 Safety Precautions..........................................................4 Assembly..................................................................5 Performance Verification.....................................................6 Line Voltage Indicator.................................................6 Timer...............................................................6 Peak Tube Potential...................................................8 Tube Current.........................................................8 Testing Checklist......................................................9 Operation Instructions......................................................10 Operation..........................................................10 Technique Chart.....................................................11 Maintenance..............................................................12 Component Data........................................................13-16 Table of Illustrations Figure A - Component Identification...........................................5 Figure B - Potentiometer Location.............................................6 Figure C - Electrical Pulse Activity During X-Ray Tube Warm-up and Exposure......7 Figure D - X-Ray Tube Electrical Performance...................................8 Chart A - Testing Checklist...................................................9 Figure E - Control Panel....................................................10 Chart B - Technique Chart..................................................11 Figure F - Absolute Maximum Rating Charts...................................14 Figure G - Emission and Filament Characteristics...............................14 Figure H - Anode Thermal Characteristics.....................................14 Figure I - Dimensional Outline...............................................15 Figure J - Schematic Diagram................................................16 Upon Arrival When the equipment is received, each shipping container should be carefully examined for any evidence of mishandling during shipment. Note its condition, if abnormal, carefully unpack all parts and examine for damage. If any damage is noted, immediately report it to the carrier in the proper manner. All printed matter supplied with the unit should be saved for installation, operation, and future reference. Page 2

Introduction Purpose The ARU-01CF Command Air X-ray System provides a portable and practical way to obtain dental radiographs in the field. To receive the best service and longest life from your Aseptico product, follow the instructions detailed in this manual. Performance characteristics Weight:.............................42 lbs. Size (external):.......................17.5 x 16 x 17, 2.75 cubic feet Maximum peak tube potential 1 :.........63kVp ± 15% Leakage technique factors:............63 kvp 12 ma 2 sec. Range of line voltage regulator:........110-130 V AC (220-260 V AC), 50/60 Hz. Current at 12 ma, 63 kvp output:......6 Amps Line voltage regulation 2 :...............2 ~ 5% Generator rating 3 :...12 ma at 63 kvp ± 15% Duty cycle:..........................30 : 1 Maximum deviation from fixed factors:..± 15% X-ray tube:..........................type D-082B Manufactured by Toshiba Items furnished X-ray tubehead and control Mounting handle Dental cone Exposure cord Power cord Mil-spec case with shock resistant foam insert ARU-01S tripod Vinyl carrying case Manual 1 Measurements taken with a 50mm diameter sphere gap. 2 The stated ma and line current are under conditions of input line voltage of 110-130 V (220-260 V) with 2 ~ 5% regulation. If regulation is less, the amps drawn and ma can be higher than stated. 3 Using a precise DC ma meter across points N and NE of the high tension transformer (see schematic diagram) to measure ma. Input is 117 V AC (230 V AC), 50/60 Hz. Tolerance is ± 15%. Page 3

Safety Precautions WARNING - Careless or improper use of x-ray equipment can be extremely hazardous. WARNING - It is imperative that this equipment be operated and serviced only by trained personnel familiar with the safety precautions required to prevent excessive exposure to primary x-ray radiation, the dangers of exposure to x-ray radiation, and the proper use of the equipment discussed in this manual. 1. During exposure, the operator must stand as far as possible from the patient being x-rayed and should wear a lead apron or stand behind a lead shield. 2. The operator must not stand in the primary x-ray beam. 3. The operator must wear a monitoring badge while operating this unit. It should be on the collar, not on an area covered by the lead apron. 4. X-ray exposure should be as short as possible. All personnel authorized to operate or service this equipment should be fully acquainted with the established maximum permissible doses, safety recommendations, and procedures derived from the following sources: A. National Council on Radiation Protection Report No. 33 (Medical X-ray and Gamma Ray Protection for Energies up to 10 MEV - Equipment Design and Use); from NCRP Publications; P.O. Box 30175, Washington, D.C. 20014. B. National Bureau of Standards Handbook No. 76 (Medical X-ray Protection up to Three Million Volts); from the Superintendent of Documents, Government Printing Office, Washington, D.C. 20401. C. All documents relating to the Performance Standard for Diagnostic X-ray Systems, 21 CFR Subchapter J, Part 1020; obtainable from FDA Center for Devices and Radiological Health, Department of HHS, 2098 Gaither Road, Rockville, MD 20850. D. State and local regulations governing radiation protection and the use of diagnostic x-ray equipment. E. Requirements of the user's in-house radiation protection program. F. Instructions and precautionary notices of this manual. Although this equipment incorporates protective design features for limiting both the direct (primary) x-ray beam and the secondary radiation produced by this beam, design factors alone cannot prevent human carelessness, negligence, or lack of knowledge. This apparatus is sold with the understanding that the user assumes sole responsibility for radiation safety and that Aseptico, Inc., its agent and representatives, do not accept any responsibility for: 1. Injury or danger to patient or other personnel from x-ray exposure. 2. Overexposure due to poor operating techniques or procedures. 3. Equipment not properly serviced or maintained in accordance with this manual. 4. Equipment which has been modified or tampered with in any way. Page 4

Assembly 1. Remove the x-ray tube head and control, cone, exposure cord, and power cord from the carrying case. 2. Screw the cone on to the front of the tubehead. 3. Plug the exposure cord into the rear of the tubehead. 4. Plug the power cord into the rear of the tubehead 5. Set up tripod 6. Attach the assembled x-ray unit to the stand. 7. Check the line voltage and the voltage on the rating label; 115 V type requires line voltage between 110 and 130 V. 230 V type requires line voltage between 220 and 260 V. These voltages must be with 15 ampere #12 three-wire service, if not, the films may be light or show lack of sufficient penetration. 8. Plug the power cord into the wall outlet. The unit is now assembled and ready for testing. 8 7 1 X-ray tubehead and control 6 2 3 4 5 Mounting handle Dental cone Exposure cord Power cord 1 6 7 ARU-01S tripod Includes: #510311 5/16-18 wing-nut #510322 5/16 fender washer Vinyl carrying case 2 3 4 5 8 Mil-spec case with shock resistant foam insert Figure A - Component Identification Page 5

Performance Verification The following tests must be conducted without fail before the ARU-01CF Command Air X-Ray System can be used for radiology. Fill out the checklist for these tests when installation of x-ray unit is complete. Line Voltage (LV) Indicator 1-1 LV Indicator Test Method: Measure voltage of wall outlet, and confirm that the following conditions are fulfilled. Wall outlet voltage: 130 ~ 125 V (260 ~ 250 V) Green LED goes on when LV knob turns 1 click clockwise. 125 ~ 120 V (250 ~ 240 V) Green LED goes on when LV knob turns 2 clicks clockwise. 120 ~ 115 V (240 ~ 230 V) Green LED goes on when LV knob turns 3 clicks clockwise. 115 ~ 110 V (230 ~ 220 V) Green LED goes on when LV knob turns 4 clicks clockwise. 110 ~ 105 V (220 ~ 210 V) Green LED goes on when LV knob turns 5 clicks clockwise. If the LV indicator needs adjustment: 1. Remove the 4 screws on the cover. 2. Adjust the potentiometer "A" (See Figure B) on the printed circuit board so that the green LED goes on at the above settings. 3. Replace the cover and the 4 screws. A B - + Figure B - Potentiometer Location Timer Since the filament of the x-ray tube is off when the unit is not emitting radiation, it takes approximately 0.15 seconds (9 electrical pulses) for the x-ray tube filament to warm up be fore an x-ray is emitted (see Figure C). Because of this warm up time, the electronic timer has a built in lag of approximately 0.15 seconds, therefore, when checking the timer with a cycle counter, 0.15 seconds must be added to the time set on the timer. Page 6

Visible Pulses Exposure end Warm-up Time Exposure Time Total Electrical Pulses Exposure start Figure C - Electrical Pulse Activity During X-Ray Tube Warm-up and Exposure TW = P - Dx TW = Filament warm-up time P = Total number of electrical pulses Dx = Number of readable dots on film as determined by a spin top test 2-1 Timer Test Method: 1. Set timer to 0.08 sec. 2. Set NERO to the following settings. SID: 18 inches Wheel range: 50 ~ 85 kv Sensitivity: Low Phase Select: 1Æ Measurement Mode: STAT X-ray values: 63 kv, 0.08 sec. 3. Make exposure and read value of exposure time. 4. Measure value of exposure time at following settings with same procedure: 0.08 sec., 0.10 sec., 0.20 sec., 0.50 sec., 1.00 sec., 2.00 sec. 2-2 Instruments: NERO Model 6000M X-ray Beam Analyzer, Manufactured by Victoreen, Inc. or equivalent 2-3 Rejection Limit: ± 2 pulse (0.08 sec. ~ 2.0 sec.), ± 10% (0.25 sec. ~ 2.0 sec.) If the timer needs adjustment: 1. Remove the 4 screws on the cover. 2. Adjust the potentiometer "B" (See Figure B) on the printed circuit board so that the radiation exposure time equals the time indicated on the timer dial, taking into consideration the accuracy tolerances. 3. Replace the cover and the 4 screws. Page 7

Peak Tube Potential The ARU-01CF has a filament warm-up time of 0.15 sec. (9 electrical pulses) built into its settings. The kv is as shown on Figure D. For that reason, measure kv with 0.15 sec. delay time. 3-1 Peak Tube Potential Test Method: 1. Set timer to 0.2 sec. 2. Set NERO to the following settings: SID: 18 inches Wheel range: 50 ~ 85 kv Sensitivity: High Phase Select: 1Æ Measurement Mode: SGL Time Delay: 0.1 sec. X-ray values: 63 kv, 0.2 sec. 3. Make exposure and measure value of average kv 4. Measure value of average kv 3-2 Instruments: NERO Model 6000M X-ray Beam Analyzer, Manufactured by Victoreen, Inc. or equivalent 3-3 Rejection Limit: 63 kv ± 15% kvp 80 63 Peak Tube Potential Tube Current 0 0.15 sec ma 12 Figure D - X-Ray Tube Electrical Performance Tube Current The ARU-01CF has a filament warm-up time of 0.15 sec. (9 electrical pulses) built into its settings. The kv is as shown on Figure D. For that reason, measure kv with 0.15 sec. delay time. 4-1 Tube Current Test Method: 1. Set timer to 2.0 sec. 2. Connect the DC-mA meter to the pin jack terminal on rear panel. 3. Make exposure and measure the ma value. 4. Measure High value of ch2 at the following kv settings, and timer to 2.0 sec. with same procedure. 4-2 Instruments: Portable standard DC ammeter Model 2011 Yokogawa Electric Works, Ltd. or equivalent 4-3 Rejection Limit: ± 15% Page 8

Chart A - Testing Checklist Complete this check list when assembly and testing are complete. Serial Number: Date Manufactured: Equipment Location: Test Acceptance Results Check Description Limits Line Voltage 110 ~ 130 V V GO NO GO (220-260 V) Line Voltage 2 ~ 5% % GO NO GO Regulation Peak Tube 63 kvp ± 15% kvp GO NO GO Potential Tube Current 12 ma ± 15% ma GO NO GO 9.5 ma ± 15% ma Exposure Time 0.08 sec ± 2 pulse sec GO NO GO 0.1 sec ± 2 pulse sec 0.2 sec ± 2 pulse sec 0.5 sec ± 10% sec 1.0 sec ± 10% sec 2.0 sec ± 10% sec All Mechanical Smooth Movement GO NO GO Movement All Operation See Operation GO NO GO Manual Instruments Used Manufacturer Model Accuracy Last Calibrated Page 9

Operation Instructions It is assumed by the distributors and manufacturer of the equipment that the person responsible for its operation has a general knowledge of the use of x-rays, including the precautions which must be taken. Note: Confirm the wall outlet voltage and the voltage on the rating label of the ARU-01CF before connecting the power cord. Operation: 1. Position the tubehead for the radiograph desired. 2. Line voltage compensation. Turn the LV knob clockwise until the center green LED lights. The green LED must be lit before each exposure. 3. Radiology A. After positioning the patient and adjusting the line voltage compensator, set the electronic timer for the desired time. The following technique chart can be used as a guide. B. Stand as far as possible from the x-ray unit, press the exposure switch to initiate the exposure and keep it pressed for the duration of the exposure. During the exposure, there will be an audible signal and the "X-ray" light will illuminate on the unit. C. The timer will automatically recycle; e.g. successive exposures can be made at the previously set time by merely pressing the exposure button. D. An exposure can be interrupted at any time by releasing the exposure button. E. When an exposure is terminated, the "X-ray" light goes out and the audible signal ends. Line Voltage L.E.D. (Green) X-Ray L.E.D. (Yellow) Line Voltage Selector Timer Control Power Cord Receptacle Exposure Cord Receptacle Figure E - Control Panel Operation Precautions: 1. During exposure, the operator must stand as far as possible from the patient being x-rayed and should wear a lead apron or stand behind a lead shield. 2. The operator must not stand in the primary x-ray beam. 3. The operator must wear a monitoring badge while operating this unit. It should be on the collar, not on an area covered by the lead apron. 4. X-ray exposure should be as short as possible. Page 10

Chart B - Technique Chart Exposure Time and Angulation of Central Ray Note: Exposure times are intended as guides only. They are based on averages and may be modified to suit the individual patient. When modifying exposure times, the increase or decrease should be made in steps of 20-25%. Aseptico ARU-01CF X-Ray System Technique Chart Exposure Time (seconds) Adult Child Angle (degrees) Molars 0.8 0.5 + 20 Maxillary Bicuspids 0.8 -- + 30 Cuspids 0.6 0.4 + 45 Incisors 0.6 0.3 + 40 Molars 0.6 0.4-5 Mandibular Bicuspids 0.6 -- - 10 Cuspids 0.4 0.3-20 Incisors 0.4 0.3-15 Maxillary Occlusals Mandibular Occlusals Incisor Region 1.0 -- +65 Cuspid-Molar Region 1.0 -- + 60 Incisor Region 1.0 -- - 55 Cuspid-Molar Region 1.0 -- - 90 Page 11

Maintenance: The following must be checked at least every six (6) months. 1. Tubehead A. Make sure certification and identification labels are in place. B. Check for any loose or missing screws. C. Check for oil leaks. D. Check for any physical damage. E. Check the LV knob for proper operation. F. Check the timer for proper operation. G. Check all the cord connections. H. Check the exposure button; the exposure must terminate if the button is released during an exposure. 2. Beam Limiting Device (Dental Cone) A. Make sure the certification label is on the cone. B. Check for any physical damage. 3. Indicators A. Make sure the green light goes on when the LV knob is turned to the right from the off position. B. Check the x-ray "ON" light for proper operation. C. Make sure the audible signal operates during an exposure. Page 12

Component Data Toshiba D-082B Stationary Anode X-ray Tube Electrical: Circuit:....................................Self-rectified (Center-Grounded) Operating Tube Voltage:.....................50 to 70 kv Max Focal Spot:................................0.8 mm Input energy (at 1.0 sec):.....................600 W Mechanical: Dimensions:...............................See dimensional outline Overall Length:.............................102mm Max. Diameter:.............................31 mm Target Angle:...............................20 degrees Inherent Filtration:.........................At least 0.8 mm Al equivalent at 50 kv X-ray coverage:.............................354 x 354 mm at SID 490 mm Weight (approx.):...........................95 g Cooling method:............................oil immersed (60 Max.) and convection oil Maximum and Minimum Ratings: These values must not be exceeded at any time. Max. Tube Voltage:.........................70 kv Max. Inverse Tube Voltage:...................80 kv Min. Tube Voltage:..........................50 Kv Max. Tube Current: See rating charts.........................19 ma Max Filament Current:......................2.0 A Filament Voltage: At max. filament current...................2.9 to 4.0 V Thermal Characteristics: Anode Heat Storage Capacity..............7 kj (10 khu) Max. Anode Heat Dissipation Rate.........210 W (300 HU/s) Cautions for the Toshiba D-082B Stationary Anode X-ray Tube Since the X-ray tube will emit x-rays when it is energized with high voltage, special knowledge is required to handle it. The items below show general cautions for the tube. 1. The tube shall be handled or operated only by qualified personnel. Only a specialist who has good knowledge of the x-ray tube should assemble, maintain, and remove the tube. 2. The tube envelope is made of glass. In transporting and handling the tube, sufficient care should be taken to avoid strong impacts or vibrations. 3. Adequate radiation protection must be used in the unit assembled with the tube. The leakage technique factor of the tube unit must not exceed the maximum anode cooling rate of the tube. 4. Fulfill the regulations and standards for the minimum source-skin distance (SSD) and the minimum filtration of the useful beam before using tube. 5. The tube might be broken by only one overload operation. Provide a proper overload protection circuit. Operate the tube by selecting a proper input condition according to the conditions for operation and tube characteristics charts. 6. If any abnormality is found when using this tube, immediately cut off the power supply and contact the TOSHIBA service department. 7. The following data sheet charts indicate standard values. Page 13

Figures F, G, & H - Standard Value Charts Toshiba D-082B Stationary Anode X-Ray Tube 20 18 SELF RECTIFIED 50 kv FOCAL SPOT: 0.8 mm TUBE CURRENT (ma) 16 14 12 10 8 60 kv 70 kv 6 0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10 EXPOSURE TIME (s) Figure F - Absolute Maximum Rating Charts 20 SELF RECTIFIED 3.5 TUBE CURRENT (ma) 15 10 5 60 kv E 3.0 2.5 2.0 FILAMENT VOLTAGE (V) 0 1.5 1.6 1.7 1.8 1.9 2.0 7 6 210 W 170 W FILAMENT CURRENT (A) Figure G - Emission and Filament Characteristics HEAT STORAGE (kj) 5 4 3 2 130 W 1 HEATING COOLING 0 0 1 2 3 4 TIME (min) Figure H - Anode Thermal Characteristics Page 14

Figure I - Dimensional Outline Toshiba D-082B Stationary Anode X-Ray Tube M5, 8 DEEP 9 SCALE= 1:1 5 UNITS= mm 51 51 ± 2 24 FOCAL SPOT 13 MAX 19 50 MIN 5 MAX 102 ± 5 31 MAX F C 60 F C CENTRAL X-RAY C C SHORT PIN 17.25 ± 0.1 F: FILAMENT C: CATHODE Terminals C are connected mutually inside the tube. Figure I - Dimensional Outline Page 15

Figure J - Schematic Diagram Aseptico ARU-01CF X-Ray System (MinXray P200D Mark III) CN3 ma output LV RSW F1 10A off CN1 AC120V 3 CN2 4 3 2 1 65902-004 T1 VAR1 VARISTOR SSR1 X-RAY SW Vcc D1 R1 20W/10 1 off Vcc SSR(-) X-RAY LED AC12V AC12V LV RSW 1 2 3 4 5 6 7 4 3 2 1 CN4 ma+ ma- 0V 110V 110V 115V 120V 120V 125V 130V 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 3191-9R 3191-9P PCB109 JP2 JP1 JP3 12 11 10 9 8 7 6 5 4 3 2 1 ma+ ma- 0V 110V 115V 120V 120V 125V 125V 130V 130V YELLOW EXPOSURE LV RSW T1 RED OVER 2 X-RAY TUBE D-082B R1 GREEN CENTER RED UNDER Figure J - Schematic Diagram Page 16

Notes Page 17

Page 18 Notes

Notes Page 19

P.O. Box 1548 Woodinville, WA 98072-1548 1-800-426-5913 * 425-487-3157 * Fax: 360-668-8722 email: info@aseptico.com * Internet: www.aseptico.com P/N 420263 Rev. A 05/00