Optical Spectrum Analyzers

Similar documents
771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com

It s Our Business to be EXACT

671 Series LASER WAVELENGTH METER. The Power of Precision in Wavelength Measurement. It's Our Business to be Exact! bristol-inst.

Swept Wavelength Testing:

R. J. Jones Optical Sciences OPTI 511L Fall 2017

Optical Power & Energy Meters

OCT. Optical Coherence Tomography

Spectrum Analyzer Compact and robust spectrometers with fully customizable range and resolution parameters, able to measure pulsed and continous

Agilent 86146B Optical Spectrum Analyzer Technical Specifications

Agilent 81600B All-band Tunable Laser Source Technical Specifications December 2002

PB T/R Two-Channel Portable Frequency Domain Terahertz Spectrometer

Wavelength Meter Sensitive and compact wavemeter with a large spectral range for high speed measurements of pulsed and continuous lasers.

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

Agilent 8614xB Optical Spectrum Analyzer Family Technical Specifications

Agilent 81980/ 81940A, Agilent 81989/ 81949A, Agilent 81944A Compact Tunable Laser Sources

Keysight Technologies Using a Wide-band Tunable Laser for Optical Filter Measurements

SA210-Series Scanning Fabry Perot Interferometer

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

A Multiwavelength Interferometer for Geodetic Lengths

Kit for building your own THz Time-Domain Spectrometer

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

PB T/R Two-Channel Portable Frequency Domain Terahertz Spectrometer

Q8384 Q8384. Optical Spectrum Analyzer

INGAAS FAST PIN (RF) AMPLIFIED PHOTODETECTORS

How-to guide. Working with a pre-assembled THz system

3 General Principles of Operation of the S7500 Laser

Miniature collimator for POF fiber: large aperture Model 011-TU2

FIBER EVO. Miniaturized laser module complete with controller and USB power supply all within an incredibly small package

Data sheet for TDS 10XX system THz Time Domain Spectrometer TDS 10XX

TURNKEY, ULTRA STABLE, OEM LASER DIODE SOURCE OZ-1000 & OZ-2000 SERIES

Agilent 81662A DFB Laser Agilent 81663A DFB Laser Agilent Fabry-Perot Lasers

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Agilent 8703B Lightwave Component Analyzer Technical Specifications. 50 MHz to GHz modulation bandwidth

POLARIZATION EXTINCTION RATIO METER

Designing for Femtosecond Pulses

LTE. Tester of laser range finders. Integrator Target slider. Transmitter channel. Receiver channel. Target slider Attenuator 2

POLARIZATION EXTINCTION RATIO MEASUREMENT TEST SETFOR V-GROOVE ASSEMBLIES AND PM PATCHCORDS

Tunable Laser Kits. Features

Agilent 8614xB Optical Spectrum Analyzer Family Technical Specifications

OP710. Multichannel Optical Power Meter Instruction Manual. (Also supports the OP710-ANX)

Agilent 71400C Lightwave Signal Analyzer Product Overview. Calibrated measurements of high-speed modulation, RIN, and laser linewidth

NASTER System Definition Proposal

Ultra high resolutions : 140 MHz/1.12 pm ; 20 MHz/0.16 pm ; 5 MHz/0.04 pm... Features:

Optical Isolator Tutorial (Page 1 of 2) νlh, where ν, L, and H are as defined below. ν: the Verdet Constant, a property of the

OP735. Benchtop Optical Power Meter Instruction Manual

Advanced Test Equipment Rentals ATEC (2832)

Agilent 86142B and 86146B Optical Spectrum Analyzers

Supercontinuum Sources

Fiber-optic Michelson Interferometer Sensor Fabricated by Femtosecond Lasers

THE TUNABLE LASER LIGHT SOURCE C-WAVE. HÜBNER Photonics Coherence Matters.

High-frequency tuning of high-powered DFB MOPA system with diffraction limited power up to 1.5W

THz Components and Systems

pulsecheck The Modular Autocorrelator

TIMING DISTRIBUTION AND SYNCHRONIZATION COMPLETE SOLUTIONS FROM ONE SINGLE SOURCE

Advanced Test Equipment Rentals ATEC (2832) EDFA Testing with the Interpolation Technique Product Note

Spectral phase shaping for high resolution CARS spectroscopy around 3000 cm 1

Modular Fourier Transform Infra-Red Spectrometer (FT-IR) Solution

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

FFP-C Fiber Fabry-Perot Controller OPERATING INSTRUCTIONS. Version 1.0 MICRON OPTICS, INC.

WSM-160 Manual Optical Tunable Filter Quick Reference Guide

Testing with Femtosecond Pulses

TracQ. Basic Data Acquisition and Spectroscopy Software

PERFORMANCE OF PHOTODIGM S DBR SEMICONDUCTOR LASERS FOR PICOSECOND AND NANOSECOND PULSING APPLICATIONS

Instruction manual and data sheet ipca h

Continuous-Wave (CW) Single-Frequency IR Laser. NPRO 125/126 Series

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

APE Autocorrelator Product Family

Continuous Wave (CW) Single-Frequency IR Laser NPRO 125/126 Series

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

Single Frequency Laser/Single Longitudinal Mode Laser

UNMATCHED OUTPUT POWER AND TUNING RANGE

PUV3402 LED multiwave photometer A new approach to online process photometry

OPTICAL MEASURING INSTRUMENTS. MS9710B 0.6 to 1.75 µm GPIB OPTICAL SPECTRUM ANALYZER

Spider Pulse Characterization

BENCHTOP POLARIZATION EXTINCTION RATIO METER

Lecture 6 Fiber Optical Communication Lecture 6, Slide 1

Universal and compact laser stabilization electronics

Add CLUE to your SEM. High-efficiency CL signal-collection. Designed for your SEM and application. Maintains original SEM functionality

THE VERSATILE TERAHERTZ-SPECTROMETERS T-SPECTRALYZER. HÜBNER Photonics Coherence Matters.

MEASUREMENT OF POSITIONING ACCURACY IN 1 AXIS

56:/)'2 :+9: 3+'9;8+3+4:

4GHz / 6GHz Radiation Measurement System

A novel tunable diode laser using volume holographic gratings

NATIONAL INSTITUTE OF RESEARCH AND DEVELOPMENT FOR OPTOELECTRONICS

FFP-TF2 Fiber Fabry-Perot Tunable Filter Technical Reference

레이저의주파수안정화방법및그응용 박상언 ( 한국표준과학연구원, 길이시간센터 )

Terahertz Spectroscopic/ Imaging Analysis Systems

High Peak Power Fiber Seeds & Efficient Stabilized Pumps

CALIBRATION OF TERAHERTZ SPECTROMETERS

Ultrafast instrumentation (No Alignment!)

Quick Start MMRC Nicolet 6700 Revised 10/24/2018

Frequency Scanned Interferometer for LC Tracker Alignment

Multiply Resonant EOM for the LIGO 40-meter Interferometer

FPPO 1000 Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual

3550 Aberdeen Ave SE, Kirtland AFB, NM 87117, USA ABSTRACT 1. INTRODUCTION

Multi-Wavelength Meter AQ6141. Perfect for Wavelength Analysis in DWDM Systems

NOW WITH BROADER TUNING RANGE

Agilent 81980A, 81960A, 81940A, 81989A, 81949A, and 81950A Compact Tunable Laser Sources

OPTICAL BACKSCATTER REFLECTOMETER TM (Model OBR 5T-50)

Intracavity laser absorption spectroscopy using mid-ir quantum cascade laser

Transcription:

Optical Spectrum Analyzers Broadband Spectrometer and Wavelength Meter in One Thorlabs Optical Spectrum Analyzers obtain highly accurate measurements of the spectra of unknown light sources. They are continuously self-calibrated using a built-in stabilized reference HeNe laser and internal temperature and pressure sensors, ensuring repeatable, reliable results across time and differing lab environments. Features u Five Models Spanning Wavelengths from 350 nm to 12.0 µm u 7.5 GHz (0.25 cm -1 ) Resolution in Spectrometer Mode u 0.1 ppm Resolution in Wavelength Meter Mode u Includes Laptop with Full Data Acquisition and Analysis Suite

Visible, NIR, and MIR Spectral Analysis The OSA product family consists of five models, each of which is designed to measure a different spectral range between 350 nm and 12.0 μm. These optical spectrum analyzers are suitable for a wide range of applications, including analyzing the spectrum of a telecom signal, resolving the Fabry-Perot modes of a gain chip, and identifying gas absorption lines in a spectral measurement. The spectral ranges supported by the OSA product family include the majority of Thorlabs light sources, including quantum cascade lasers, blackbody sources, and tunable external cavity lasers. These broad ranges make it possible to measure absorption signatures in the visible and NIR; the C-band, L-band, and other important telecom transmission windows; and molecular absorption lines for carbon monoxide and other atmospheric compounds. Available Models Item # OSA201C OSA202C OSA203C OSA205C OSA207C Wavelength Range 350 1100 nm 600 1700 nm 1.0 2.6 µm (10 000 3846 cm -1 ) 1.0 5.6 µm (10 000 1786 cm -1 ) 1.0 12.0 µm (10 000 833 cm -1 ) Demo Units Available To help ensure that our OSAs meet your application needs, we have designated several units for trial use. If you would like to try out an OSA in your lab, please contact us at techsupport@thorlabs.com with your experimental requirements.

Highly Accurate Spectral Measurements The OSA product family is fully optimized for the characterization of broadband and narrowband light sources. For broadband sources, our OSAs provide ±2 parts-per-million (ppm) spectral accuracy and 7.5 GHz (0.25 cm -1 ) spectral resolution. For sources with linewidth < 10 GHz, the Wavelength Meter mode provides center wavelength measurements with 0.1 ppm resolution and ±1 ppm accuracy. Resolution (pm) 30000 10000 1000 100 10 Resolution in Spectrometer Mode Low Resolution Mode High Resolution Mode 1 0 1 2 3 4 5 6 7 8 9 10 11 12 Wavelength (µm) Noise Floor (dbm) Noise Floor (dbm/nm) -45-55 -65-75 -85-95 OSA Noise Floor in Absolute Power Mode OSA201C OSA202C OSA203C OSA205C OSA207C -105 0.3 1 10 20 Wavelength (µm) -30-40 -50-60 -70-80 OSA Noise Floor in Power Density Mode OSA201C OSA202C OSA203C OSA205C OSA207C -90 0.3 1 10 20 Wavelength (µm) Phase-Locked Loop with Stabilized HeNe Laser Our high spectral accuracy and precision is retained across a wide range of environments by incorporating a stabilized 632.991 nm HeNe laser. The interferogram of this HeNe is measured simultaneously with the unknown source under test, and the result is used to continuously calibrate the instrument. Small deviations in the HeNe wavelength can occur as a result of changing temperature and air pressure. Thorlabs OSAs compensate for these in real time by using internal sensors to measure environmental variations. The sensor output is combined with the Edlén formula to calculate the refractive index and determine the HeNe wavelength on a shot-to-shot basis. The HeNe interferogram is used to clock the 16-bit analog-to-digital converter, such that signals from the unknown source are measured at a fixed, equidistant optical path length interval. The HeNe reference fringe period is digitized, and its frequency is multiplied by a phase-locked loop. This phase-locked loop enables extremely high accuracy and precision over the entire operating range.

Full Software Suite for Data Acquisition and Analysis Coherence Length Module (Digits of Uncertainty are Displayed in Gray) Peak Track Mode with 7.9 µm Quantum Cascade Laser Each Optical Spectrum Analyzer includes a Windows laptop with Thorlabs OSA software suite. This software features an intuitive, responsive, straightforward interface that exposes all functions in one or two clicks. We regularly update this software to add significant new features and make improvements suggested by our users. These software updates are available free of charge at www.thorlabs.com. Built-In Tools for Narrowband and Broadband Signals The OSA software displays either the raw interferogram or the Fourier-transformed spectrum obtained by the instrument. In the main window, it is possible to average multiple spectra; display the X axis in units of nm, cm -1, THz, or ev; compare the live spectrum to previously saved traces; perform algebraic manipulations on data; and calculate common quantities such as transmittance and absorbance. Robust graph manipulation tools include automatic and manual scaling of the displayed portion of the trace and markers for determining exact data values and visualizing data boundaries. Automated peak and valley tracking modules identify up to 2048 peaks or valleys within a userdefined wavelength range and follow them over a long period of time. Acquired data can be saved as a spectrum file that can be loaded quickly into the main window. Data can also be exported into Matlab, Galactic SPC, CSV, and text formats. Adjustable Sensitivity and Resolution Settings The scan sensitivity and resolution can be adjusted by the user to balance the needs of the experiment against the data acquisition rate. These settings vary the number of data points per interferogram from 0.5 million to 16 million. The sensitivity setting modifies the range of detector gain levels, while the resolution setting controls the optical path difference (OPD). The table below shows how the data acquisition rate depends upon the chosen settings. Update Frequency Low Resolution (1 cm -1 ) High Resolution (0.25 cm -1 ) Low Sensitivity 1.9 Hz (0.5 s) 0.6 Hz (1.8 s) Medium Low Sensitivity 1.2 Hz (0.8 s) 0.3 Hz (2.9 s) Medium High Sensitivity 0.7 Hz (1.5 s) 0.2 Hz (5.2 s) High Sensitivity 0.4 Hz (2.7 s) 0.1 Hz (9.5 s) The scan sensitivity and resolution are independent settings controlled from the software.

Output of Wavelength Meter Module over Time. One Division is Equal to 50.0 fm. Carbon Dioxide (CO 2 ) Absorption Before and After Baseline Correction Wavelength Meter Module for Narrowband Sources For sources with <10 GHz linewidth, the Wavelength Meter module enables extremely accurate determinations of the center wavelength (±1 ppm accuracy, 0.2 ppm precision, and 0.1 ppm resolution). This mode allows the system to resolve a fraction of a fringe in the interferogram, using the phase-locked loop that is generated by the internal stabilized reference HeNe laser. The uncertainty in the measurement is continuously determined and displayed as gray numbers. A built-in module plots the output of the wavelength meter measurement as a function of time. If the software determines that the wavelength meter will give inaccurate results (as it would for broadband sources), it is automatically disabled. Coherence Length Module for Broadband Sources Since Thorlabs OSAs obtain the raw interferogram of the unknown source (as opposed to grating-based spectrum analyzers, which cannot offer this capability), the software is able to calculate the coherence length of the input signal. The Coherence Length module considers the envelope of the interferogram and reports the optical path length over which the envelope s amplitude decays to 1/e of its maximum value on both sides. Spectroscopic Analysis from HITRAN Reference Database In environmental sensing and telecom applications, it is often useful to identify atmospheric compounds (such as water vapor, carbon dioxide, and acetylene) whose absorption lines overlap with that of the unknown source being measured. The OSA software includes built-in support for HITRAN line-by-line references (www.cfa.harvard.edu/ hitran), which can be used to calculate absorption cross sections as a function of vapor pressure and temperature. The predictions can be fit to the measured trace for comparison, and fits using mixtures of gases are supported. Apodization and Interferogram Truncation Since the resolution of any Fourier-transformed spectrum is intrinsically constrained by the finite path length over which the interferogram is measured, the software implements several functions to account for the effect of the finite path length on the spectrum that is obtained. The user may select from a number of apodization methods (dampening functions), including cosine, triangular, Blackman-Harris, Gaussian, Hamming, Hann, and Norton-Beer functions. The effective optical path length can also be shortened to eliminate contributions from high-frequency spectral components. The ability to view the raw interferogram in real time allows the user to confirm the coherence length reported by the software and adjust the signal amplitude to avoid saturation. The maximum coherence length measurable by the OSA is limited by the maximum optical path difference of ±4 cm in high-resolution mode, making this module best suited for broadband sources.

Fiber-Coupled and Free-Space Input Ports All OSA models directly accept fiber-coupled input sources. The fiber-coupled input is compatible with single mode and step-index FC/PC multimode patch cables. Single mode patch cables provide the highest contrast. For measurements in the visible and NIR, we recommend core sizes up to Ø50 µm. For measurements from 2 µm to 5.5 µm, we recommend our Indium Fluoride (InF 3 ) patch cables with core sizes up to Ø100 µm, which provide extremely low attenuation in the MIR region of the spectrum. Custom designs with other fiber input receptacles are available upon request. In addition, all OSA models also accept free-space optical inputs up to Ø6 mm. For alignment purposes, a red Class 1 beam is emitted from the aperture. The input beam will need to be collinear with the alignment beam for the best possible measurment accuracy. Four 4-40 taps around the free-space input provide compatibility with Thorlabs 30 mm cage system, which reduces the mechanical degrees of freedom in a setup in order to simplify alignment. OSAs are compatible with single mode and step-index multimode patch cables with cores up to Ø50 µm, as well as fluoride multimode patch cables with cores up to Ø100 µm. The fiber-coupled input on the OSA205C is shown here. All OSAs accept free-space optical inputs up to Ø6 mm. The free-space input on the OSA205C is shown here. A Ø1/2 off-axis parabolic mirror is mounted in a 30 mm cage system on the OSA205C. A rotational mount provides a rotational degree of freedom for alignment.

Compact Interferometer with Precision Design Thorlabs Optical Spectrum Analyzers use a dualretroreflector design, as shown in the figure to the right. These retroreflectors are mounted on a moving carriage that simultaneously adjusts the length of each arm of the interferometer in opposite directions. Reference HeNe Unknown Input Mirrors The advantage of this layout is that it changes the optical path difference by four times the mechanical movement of the platform, reducing the physical package and decreasing the acquisition time. Beamsplitter Detector Assembly The detector assembly is clocked by a phase-locked loop that is generated by the interferogram of the internal stabilized HeNe laser. The interferogram Retroreflectors fringes trigger a 16-bit analog-to-digital converter, and the frequency of the reference fringes is multiplied by 16X, 32X, 64X, or 128X. At 128X, data points are acquired approximately every 1 nm of Mirror Moving Carriage Mirror platform travel. In high-resolution mode, the optical path difference (OPD) is ±4 cm, for a maximum spectral resolution of 0.25 cm -1 (7.5 GHz). Low-resolution mode, which improves the update rate by roughly a factor of three, reduces the OPD to ±1 cm for a spectral resolution of 1 cm -1 (30 GHz). For narrowband sources (linewidth < 10 GHz), the Wavelength Meter module can be separately enabled to determine the center wavelength with 1 ppm accuracy. The Wavelength Meter takes advantage of the phase-locked loop to resolve data points to within a fraction of a fringe in the interferogram. Custom OSAs We invite customers whose needs are not addressed by our standard OSA models to tailor an OSA to a specific application by working with our engineering and manufacturing team. In the past, we have built OSAs with user-specified optical inputs, such as FC/APC and SMA905 fiber receptacles, and we have incorporated optical bandpass and notch filters directly into the optical path to reduce light source noise. In addition, our software team has implemented userdesigned data analysis modules within the standard OSA software suite. We have also worked with our customers to choose detector elements targeted at specific light sources and analytes. Our engineers are well-versed in the tradeoffs between detection bandwidth, sensitivity, and linearity, and can make recommendations based upon the needs of the application and prior customers' experiences. Power Density (dbm/nm) -60-65 -70-75 -80-85 MIR Photoluminescence Detection Wafer #1 Wafer #2 Wafer #3 Wafer #4 Wafer #5 Wafer #6-90 2.0 2.5 3.0 3.5 4.0 4.5 5.0 Wavelength (µm) A user requested an OSA capable of detecting photoluminescence from wafers that emit in the 2-4 µm spectral range. We provided a custom-built OSA with a greatly reduced noise floor as compared to the OSA205C, which easily detected the predicted signal.

Common Specifications Notes Value Spectral Resolution Spectral Accuracy a Spectral Precision b Wavelength Meter Resolution Wavelength Meter Accuracy Wavelength Meter Precision c Spectrometer Mode Wavelength Meter Mode (Linewidth < 10 GHz) 7.5 GHz (0.25 cm -1 ) ±2 ppm 1 ppm 0.1 ppm ±1 ppm 0.2 ppm Input Power (Max) CW Source 10 mw (10 dbm) Input Damage Threshold d 20 mw (13 dbm) Power Level Accuracy e ±1 db Optical Rejection Ratio 30 db Input Fiber Compatibility Free-Space Input FC/PC Connectors f All SM Fiber Patch Cables Step-Index MM Fiber Patch Cables with Ø50 µm Core and NA 0.22 Step-Index Fluoride MM Fiber Patch Cables with Ø100 µm Core and NA 0.26 Accepts Collimated Beams up to Ø6 mm Red Alignment Laser Beam (Class 1) Dimensions 320 mm x 149 mm x 475 mm (12.6" x 5.9" x 18.7") Input Voltage f 100-240 VAC, 47-63 Hz 250 W (Max) Storage Temperature -10 C to 60 C Relative Humidity <80%, Non-Condensing a After a 45 minute warm-up, for a single mode FC/PC-terminated patch cable at an operating temperature of 20-30 C. b Spectral Precision is the repeatability with which a spectral feature can be measured using the peak search tool. c Using the same input single mode fiber for all measurements. d Limited by the damage threshold of the internal components. e Specified using Absolute Power Mode, Zero Fill = 2, and Hann apodization, after a 45-minute warm-up, for an operating temperature of 20-30 C. (The different apodization modes available in the OSA software are described in section 16.2 of the manual.) The specified wavelength range is 400-1000 nm for OSA201C, 600-1600 nm for OSA202C, 1.0-2.4 µm for OSA203C, 1.3-5.0 µm for OSA205C, and 2.0-11.0 µm for OSA207C. Each specification is valid for a single mode FC/PCterminated patch cable, as well as for a collimated free-space beam with diameter < 3 mm and divergence < 3 mrad, assuming the included protective window is installed in the free-space aperture. f Each OSA and Windows laptop comes with a region-specific power cord. Model-Specific Specifications OSA201C OSA202C OSA203C OSA205C OSA207C Wavelength Range a 350 1100 nm 600 1700 nm Level Sensitivity b Operating Temperature -50 dbm/nm (350-500 nm) -60 dbm/nm (500-1100 nm) a Limited by bandwidth of detectors and optics. -65 dbm/nm (600-700 nm) -70 dbm/nm (700-1700 nm) 1.0 2.6 µm (10 000-3846 cm -1 ) 1.0 5.6 µm (10 000-1786 cm -1 ) -70 db/nm c -40 db/nm 1.0 12.0 µm (10 000-833 cm -1 ) -30 db/nm for 1.0 2.0 µm -40 db/nm for 2.0 12.0 µm 10 C to 40 C 10 C to 35 C b Minimum detectable power per nanometer using Zero Fill = 0 and the highest resolution and sensitivity settings. c Specified in low-temperature mode over 1.0 2.5 µm. In high-temperature mode, the level sensitivity is -65 dbm/nm over 1.0-2.6 µm. 56 Sparta Avenue Newton, New Jersey 07860 Sales: 973.300.3000 Fax: 973.300.3600 www.thorlabs.com USA www.thorlabs.com Thorlabs Imaging Systems Phone: 1-703-651-1700 Thorlabs Scientific Imaging (TSI) Thorlabs Vytran Division CANADA www.thorlabs.com Thorlabs Canada (ULC) France www.thorlabs.de Phone : +33 (0) 970 444 844 Germany www.thorlabs.de Thorlabs GmbH Phone: +49 (0) 8131 5956-0 China www.thorlabschina.cn Phone: +86 (0)21-60561122 Sweden www.thorlabs.com Phone: +46 31 733 30 00 Polish Direct Line: +48 22 219 52 30 Thorlabs Quantum Electronics (TQE) UK www.thorlabs.de Phone: +44 (0)1353 654440 Thorlabs GmbH (Lübeck) Phone: +49 (0) 8131-5956-0 Japan www.thorlabs.jp Phone: +81-3-5979-8889 Thorlabs Ultrafast Optoelectronics (UFO) Thorlabs Vytran Europe Phone: +44 (0) 1392-445777 Thorlabs Elliptec GmbH Phone: +498131-595640-880 Brazil www.thorlabs.com Phone: +55 (16) 3413 7062