INSTRUCTION MANUAL DL49

Similar documents
INSTRUCTION MANUAL UTL260

INSTRUCTION MANUAL Fax: (503)

INSTRUCTION MANUAL Fax: (503)

INSTRUCTION MANUAL DL Fax: (503)

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL DLM2

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com # # AAC Clamp Meter. Instruction Manual

INSTRUCTION MANUAL INF Fax: (503)

USER'S MANUAL DMR-6700

User Guide. Digital AC/DC Clamp Meter Model 38394

User s Guide. Digital AC/DC Clamp Meter Model 38394

INSTRUCTION MANUAL. Model Autoranging DMM ProbeMeter TM. Measures voltage, resistance, frequency, capacitance, temperature, and duty cycle.

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter

User s Guide. 400A AC/DC Clamp Meter. Model MA220

600A Clamp Meters w/tightsight Display

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH

INSTRUCTION MANUAL ADM Fax: (503)

USER MANUAL 600A AC Clamp Meter + NCV Model MA610

Autoranging Multimeter Extech EX503

INSTRUCTION MANUAL DL Fax: (503)

NOTE: Fully read and understand this manual before using this Digital Multimeter.

User's Guide Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX330

312, 316, 318. Clamp Meter. Users Manual

User Manual Digital Clamp Multimeter. model no.: MSR-C600

200Amp AC Clamp Meter + NCV Model MA250

TRMS LEAKAGE CURRENT CLAMP-ON METER 565

DM-46 Instruction Manual

R/C SMD Tweezers Multimeter Model RC200

CONTENTS MS2033A. 1.Introduction Safety Information...01

DCM730. Digital Clamp Meter User s Manual

Thank you again for choosing AstroAI, if you have any questions or concerns regarding your product, please contact us at

INSTRUCTION MANUAL DMEG3/IRT Fax: (503)

Instruction Manual ICM 3091N Digital AC Clampmeter EN FR IT DE ES

netzerotools.com 374/375/376 Users Manual Clamp Meter

UT207A/208A/209A Operating Manual. Table of Contents

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español

USER S MANUAL DMR-1500

OPERATING INSTRUCTIONS

User Manual Digital Multimeter. model no.: MSR-U1000

DIGITAL MULTIMETER CONTENTS DIGITAL MULTIMETER CONTENTS

User s Manual. Ground Resistance Clamp On Tester MODEL

Table of Contents Title Page

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual

DVM645BI BENCH MULTIMETER TAFELMULTIMETER MULTIMETRE DE TABLE BANCO MULTÍMETRO TISCHMULTIMETER. User Manual. Gebruikershandleiding

OPERATING INSTRUCTIONS

AC/DC True RMS Power Clamp Meter ELECTRONIC TEST INSTRUMENTS MODEL 325

DIGITAL MULTIMETER OPERATOR'S INSTRUCTION MANUAL HOLD 10A COM LIGHT MS8265 ON/OFF. 200M KHz 2K 20K μ μ μ n.

HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL

USER'S MANUAL DMR-4350

DM-46 Instruction Manual

Digital Clamp Meter Model: &

User s Guide. Model MA A AC Mini Clamp-on Meter

Model: Pro95 TRUE RMS MILLIAMP CLAMP METER

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL- KM 2782 / 2782-T

36 AC/DC True RMS. Clamp Meter. P Read First: Safety Information

80i-110s AC/DC CURRENT PROBE

AC/DC CLAMP METER USER S MANUAL

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

374/375/376 Clamp Meter

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX330

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

DC/AC CLAMP METER ISO-TECH 2000 USERS MANUAL

Sales: Technical: Fax:

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL

Mini Clamp Meter Model:

User's Guide. 800 Amp AC/DC True RMS Clamp Meter. Model EX Washington Street Melrose, MA Phone Toll Free

374/375/376 Clamp Meter

MM V 10A ENGLISH. INSTRUCTION MANUAL Auto-Ranging DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE

MM700. True RMS ENGLISH. INSTRUCTION MANUAL Auto-Ranging. Measurement Technology

CL900. True RMS 1000V 2000A 60MΩ ENGLISH. INSTRUCTION MANUAL 2000A Digital Clamp Meter. Measurement Technology

902 FC. HVAC True-rms Clamp Meter

Model : OPERATING MANUAL Table of Contents (1)

ET-3888 True RMS Clamp Meter. User Manual

12B/18 MultiMeter Instruction Sheet

Pen Multimeter. Model

User's Manual. Clamp-on Power Datalogger. Model

OPERATOR S INSTRUCTION MANUAL

Not for Reproduction

USER'S MANUAL DMR-2400

OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER

MM700. INSTRUCTION MANUAL Auto-Ranging Digital Multimeter True RMS

Analog Technologies. Multimeter 115C and 117C

EM420A/420B DIGITAL MULTIMETER OWNERS MANUAL Read this owners manual thoroughly before use

DVM1190 DIGITAL MULTIMETER

User s Guide. 400A AC/DC Clamp Meter. Model MA Washington Street Melrose, MA Phone Toll Free

Model UT201/202: OPERATING MANUAL. Table of Contents

15B & 17B. Multimeters. Users Manual

MS8250A/B OPERATION MANUAL MS8250A. Hz% FUNC REL RANGE REL HOLD OFF 10A. Hz% A NCV. Hz% COM. A ma 10A FUSED 600V CAT IV.

15B & 17B. Users Manual. Multimeters

User Manual. All rights reserved. Specifications are subject to change without notice.

AC/DC DIGITAL CLAMP METER OPERATION MANUAL

Pen Multimeter. Model

374 FC/375 FC/376 FC Clamp Meter

User's Guide. Extech 800 Series 1000 Amp Clamp Meters with IR Thermometer

User Manual. All rights reserved. Specifications are subject to change without notice.

Safety. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.

Compact Autoranging Clamp Meters. Models (400 A AC), (400 A AC/DC)

Dawson DDM190. Digital Multimeter User s Manual

PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D

LH240 LH1040 LH2040 LH2015

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

Transcription:

INSTRUCTION MANUAL DL49 1-800-547-5740 Fax: (503) 643-6322 www.ueitest.com email: info@ueitest.com

Introduction The DL49 is a hand-held, battery powered, digital meter with clamp-on current measuring ca p a b i l i ty. This instrument is ideal for anyone that needs to make quick, accurate measurements of voltage, resistance, and AC or DC inductive ampera g e. Features include UL listed 3111-1 Six measurement functions 1. AC inductive amps (to 400 amps) 2. DC inductive amps (to 400 amps) 3. AC Volts 4. DC Volts 5. Ohms 6. Continuity (with audible alert) Autoranging AC amperage peak-hold Data hold for volt, ohm, and amperage functions Low battery indicator Auto continuity Auto power off Rugged construction Surge protection Compact size for easy access to tight areas If any of the following indications occur during testing, turn off the power source to the circuit under test: Arcing Flame Smoke Extreme Heat Smell of Burning Materials Discoloration or Melting of Components Read the safety precautions associated with the equipment being tested and seek assistance or advice when performing unfamiliar tasks. Keep your fingers away from the test lead metal probe contacts and bus-bars when making measurements. Always grip the instrument and test-leads behind the hand guards (molded into the probes). In the event of electrical shock, ALWAYS bring the victim to the emergency room for evaluation, regardless of the victim s apparent recovery. Electrical shock can cause an unstable heart rhythm that may need medical attention. International Symbols Safety Notes Before using this meter, read all safety information carefully. In this manual the word "WARNING" is used to indicate conditions or actions that may pose physical hazards to the user. The word "CAUTION" is used to indicate conditions or actions that may damage this instrument. Exceeding the specified limits of this meter is dangerous and can expose the user to serious or possibly fatal injury. DO NOT attempt to measure any voltage that exceeds 600 volts with this meter - UEi offers numerous alternatives for measuring high voltage and current Voltages above 60 volts DC or 25 volts AC may constitute a serious shock hazard DO NOT attempt to use this meter if either the meter or the test leads have been damaged. Send unit in for repair by a qualified repair facility Test leads must be fully inserted prior to taking measurements Always turn off power to a circuit (or assembly) under test before cutting, unsoldering or breaking the current path. Even small amounts of current can be dangerous Always disconnect the live test lead before disconnecting the common test lead from a circuit When measuring high voltage, disconnect the power source before making test lead connections. Connect the test leads to the meter first then to the circuit under test. Reapply power C o n t rols and Indicators 1. C l a m p : Used to measure inductive AC and DC current. Opens to 1 1/2 (40 mm). CAUTION! The clamp uses a high-tension spring to close the jaw. DO NOT allow fingers or objects to become pinched in the base as jaw closes. 2. Conductor Alignment Marks: Used to aid in the visual alignment of a conductor when measuring inductive ampera g e. Greatest accura cy is achieved when the conductor inside the clamp is centered at the intersection of these marks. 3. Hand Guard: Used as a point of reference for the operator s safety. Always keep your hands and fingers behind the hand guards when measuring current on exposed conductors. Contact may result in serious safety. DL49-MAN P. 1

4. Data Hold Push-button: Freezes the value displayed on the digital read-out. 5. Peak Hold Push-button: Used to capture the highest AC inductive amp reading and to automatically zero (approximately) the inductive DC amp readings. 6. Clamp Lever: Opens and closes current clamp jaw. 7. Rotary Function Switch: Used to power the meter on and off, or to select one of the available measurement functions. NOTE: Measure inductive AC and DC current using the clamp. Measure volts AC and DC, resistance and continuity at the test lead inputs. 8. Off Position: Turns the meter off. Always store your meter in the off position. If the meter will not be used for a month or more, remove the batteries. 9. Display: C o m m u n i cates function, range, and value information to the user. 10. Common Terminal: The black test lead is plugged into this terminal to supply the ground or low reference for all measurements. 11. Volt/Ohm (Ω) Terminal: The red lead is plugged into this terminal. It is used for AC/DC volts, ohms, and continuity measurements. 12. Maximum Input Statements: M AX 600V indicates that a m aximum of 600 Volts can be applied between the two terminals of b e tween earth ground and any terminal. LCD Display Functional Description 1. Peak Hold: I n d i cates the meter is displaying the maximum inductive AC current value recorded. 2. Continuity: I n d i cates the meter is in the continuity measurement mode and will sound a tone when measuring resistance below approximately 50 ohms. 3. Data Hold: I n d i cates the value displayed is held on screen (the data hold button is pressed). 4. Low Battery Indicator: This symbol appears when the battery needs replacement. NOTE: A low battery will adversely affect accuracy. 5. AC: I n d i cates that alternating current/voltage is being measured. 6. Minus: I n d i cates the value measured has a negative polarity. 7. DC: I n d i cates that direct current/voltage is being measured. 8. mva: I n d i cates that millivolts (mv), volts (V), milliamps (ma) or amps (A) is being displayed. 9. MKΩ: Indicates that Megohms, Kilohms, or ohms are being displayed. 10. Numerical Value: Displays the total value of the units displayed. 4 1 2 3 2 1 3 5 6 7 8 9 6 4 5 7 8 Operating Instructions Auto-Power Off This instrument automatically shuts off after 30 minutes of inactivity. The meter is considered active when there is a change of at least 10 digits during this period (for example: the meter senses a change from 24.04 volts to 24.14 volts). 10 10 11 9 Autorange Resolution This instrument automatically selects the range that gives you the best resolution for the function and value measured. As the meter seeks the appropriate range, the display may briefly display an overload (OFL) indication or show quickly changing values until it stabilizes. 12 DL49-MAN P. 2

Peak Hold When measuring AC amps you can use the PEAK HOLD button on the right side of the instrument to capture the highest measured value. Press and hold until the word PEAK appears along the top of the display and only the highest measured value will be displayed on the LCD. Press and hold a second time or change functions to return to real-time measurement. Data Hold The DATA HOLD button freezes the reading displayed on the LCD at the moment it is passed. To engage data hold, press the HOLD button, located on the face of the instrument. When this function is active, the symbol D.H appears on the digital display. To cancel data hold, press the data hold button again, or select any other measurement function using the rotary function select switch. Rotary Function Select Switch The rotary function select switch is used to select the measurement mode and to turn the meter on and off. CAUTION! Set the rotary function select switch to the appropriate setting before connecting the test leads to circuits under test. Observe those safety precautions outline in the beginning of this manual. Measuring Inductive Current The inductive current measurement mode relies on the induced electromagnetic field that occurs when electricity flows through a conductor. Prepare for measurement be separating a single live conductor from any other phase, neutral or ground conductor. Squeeze the clamp lever, and place the conductor in the open jaws. To attain the most accurate reading, ensure the conductor is centered in the jaws of the clamp, and the jaws are closed tight. The conductor must be able to fit inside the 1 1/4 (32 mm) fully closed jaws. To measure inductive AC current: 1. Place the function select switch in the AC amp position. 2. Place the clamp jaw around a live conductor (as described above). 3. Allow meter to stabilize - Observe reading. The maximum limit for this function is 400 amps AC. Too much current will saturate the ferrous material in the clamp, and adversely affect accuracy. DO NOT attempt to take any unknown voltage or current measurements that may be in excess of this meter s maximum limits. To avoid the risk of electrical shock and instrument discharge, open circuit voltage for the circuit under test must not exceed 600 volts (RMS). Consider using an optional high voltage probe for high power situations. Measuring Voltage Input voltages must not exceed 600 volts. High voltages and currents require greater awareness of physical safety hazards. Before connecting the test leads; turn off the power to the circuit under test; set the meter to the desired function and range; connect the test leads to the meter first, then to the circuit under test. Reapply pow e r. If an erroneous reading is observed, disconnect power immediately and recheck all settings and connections. When taking voltage measurements your meter must be connected in p a rallel to the circuit, or circuit element, under test. To measure DC or AC volts: 1. Set the rotary function select switch to the desired AC or DC voltage position. 2. Disconnect the power from the circuit to be tested. 3. Connect the test leads to the circuit to be tested. 4. Reapply power to the circuit. The measured voltage will be displayed. 5. If the input to the V/Ω port is lower (more negative) than the black COM port, a minus sign (negative polarity) will appear on the left of the display. 6. Disconnect power to the circuit before removing attached test leads from the circuit. NOTE: During continuity or resistance measurements, polarity does not matter. Measuring Resistance CAUTION! Turn off power and discharge all capacitors on the circuit to be tested before attempting in circuit resistance measurements. Failure to do so may result in equipment or instrument damage. It is critical to both the welfare of the meter, and the accuracy of the measurement, that you remove all power to the circuit under test when making resistance measurements. If any voltage is present in the test circuit, whether from a conventional power supply, or energy stored in a capacitor, an erroneous reading will result. This meter may be damaged if more than 600 volts are present. NOTE: When measuring critically low ohm values, touch tips of test leads together and record the test lead resistance value. Subtract this value from the total circuit resistance to obtain the most accurate reading. NOTE: For all measurements requiring the use of the meter leads, insert the red lead into the V/Ω port and the black lead into the COM port. DL49-MAN P. 3

For resistance measurements above one megohm the display might take a few seconds to stabilize. This is normal for high resistance readings. To measure resistance, follow these steps: 1. Set the rotary switch to the resistance function. 2. Turn off power to the circuit under test and ensure there is no residual voltage present from any source. 3. Touch the probes to the test points and read the display. The meter beeps as it seeks the correct range to measure the circuit s resistance. Be sure you have good contact between the test leads and the circuit. Dirt, oil, solder-flux or other foreign matter alters the reading value. Measuring Continuity Use the continuity mode to make quick checks for connection of electrical circuits, such as electrical wiring, switch contacts, relays and audio cables. In the continuity mode, an audible tone sounds when the value measured is approximately 50Ω or less. To test for continuity, follow these steps: 1. Set the rotary switch to the continuity function. 2. Place one probe to each side of the circuit to be tested. If the circuit measures approximately 50Ω or less, the meter will sound a continuos tone. NOTE: When servicing the meter, use only the replacement parts specified. Battery: 9V, NEDA 1604 or IEC 6LR 61 Test lead set: ATL55 Cleaning and Decontamination Periodically clean your meter s case using a damp cloth. DO NOT use abrasives, cleaning solvents or strong detergents, as they may damage the finish or affect the reliability of the structural components. Battery Replacement Always use a fresh replacement battery of the specified size and type. Immediately remove the old or weak battery from the meter and dispose of it in accordance with your local disposal regulations. Old or defective batteries can leak chemicals that corrode electronic circuits. To avoid electric shock, be sure to turn off the meter s power and disconnect both test leads from any equipment before you remove or install batteries. To install a new battery, follow these procedures: 1. Remove the screw from the battery compartment cover on the back (lower half) of the meter and lift the cover (Fig 1). 2. Remove and discard the old battery. Always dispose of old batteries promptly in a manner consistent with local disposal regulations. Periodic service M a i n t e n a n c e Under NO circumstance should you expose batteries to extreme heat or fire as they may explode and cause injury. 3. Place a fresh 9V battery in the compartment. Repair and service of this instrument is to be performed by qualified personnel only. Improper repair or service could result in physical degradation of the meter. This could alter the protection from electrical shock and personal injury this meter provides to the operator. Perform only those maintenance tasks that you are qualified to do. NOTE: If you do not plan to use the meter for a month or more, remove the battery and store it in an area that won t be damaged by a leaking battery. 4. Reattach the battery compartment cover to the meter and reinstall the screw. These guidelines will help you attain long and reliable service from your meter: 1. Calibrate your meter annually to ensure it meets original performance specifications. 2. Keep your meter dry. If it gets wet, wipe it dry immediately. Liquids damage electronic circuits. 3. Whenever pra c t i cal, keep the meter away from dust and dirt, which can cause premature wear. 4. Although your meter is built to withstand the rigors of daily use, it can be damaged by severe impacts. Use reasonable caution when using and storing the meter. Remove cover DL49-MAN P. 4 (Fig 1)

S p e c i f i c a t i o n s Standard & Optional Accessories Measurement limits Frequency AC Amperage DC Amperage AC Voltage DC Voltage Ohms Continuity General specifications Size (H x W x L) Operating temperature Storage temperature Relative humidity (storage and use) Weight Calibration frequency Ul listing IEC Listing Battery Battery life 50-400 Hz 400 Amps 400 Amps 600 Volts 600 Volts 40 Megohms <50 ohms 7.68 x 3.06 x 1.79 (195 x 77 x 46 mm) 32 to 113 F (0 to 45 C) -4 to 140 F (-20 to 60 C) 0% to 80% RH 13 oz. (370g) Annual 3111-1 1010-1 (CAT III 600 Volts) 9V Alkaline (NEDA 1604 or 6F22 or 006P) 80 hours typical (Alkaline) Standard Soft carrying case.....................................ac259 Test leads............................................atl55 Optional Plug-in line splitter....................................asl1 Temperature measurement adapter......................ta2k Hard carrying case (service kits).........................ac506 High voltage probe....................................khv66 Flame safeguard test adapter...........................dlmat Electrical specifications Function Ohms (Ω) Volts AC (50-400 Hz) Volts DC Amps AC (50-400 Hz) Amps DC Range 400 Ω 4 KΩ 40 KΩ 400 KΩ 4 MΩ 40 MΩ 4 V 40 V 400 V 600 V 4V 40 V 400 V 600 V 40 A 400 A 40 A 400 A Resolution 0.1 Ω 1.0 Ω 10 Ω 100 Ω 1 KΩ 10 KΩ 1 mv 10 mv 100 mv 1 V 1 mv 10 mv 100 mv 1 V 0.01 A 0.1 A 0.01 A 0.1 A Accuracy ±(0.9% + 3 digits) ±(1.2% + 5 digits) ±(1.5% + 3 digits) ±(0.9% + 2 digits) ±(2% + 5 digits) ±(3% + 10 digits) ±(2.5% + 10 digits) ±(1.5% + 5 digits) DL49-MAN P. 5

DL49 Digital Clamp-On Meter Limited Warranty The DL49 is warranted to be free from defects in materials and workmanship for a period of three years from the date of purchase. If within the warra n ty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi s option. This warra n ty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or improper maintenance. Batteries and consequential damage resulting from failed batteries are not covered by warra n ty. Any implied warranties, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the express warranty. UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expenses or economic loss. A purchase receipt or other proof of original purchase date will be required before warra n ty repairs will be rendered. Instruments out of warra n ty will be repaired (when r e p a i rable) for a service charge. Return the unit postage paid and insured to: 1-800-547-5740 FAX: (503) 643-6322 www.ueitest.com Email: info@ueitest.com This warranty gives you specific legal rights. You may also have other rights which vary from state to state. PLEASE RECYCLE Copyright 2007 UEi DL49-MAN 1/07