MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

Similar documents
MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

Basic Users Manual for Tecnai-F20 TEM

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Transmission Electron Microscopy 9. The Instrument. Outline

STEM alignment procedures

2. Raise HT to 200kVby following the procedure explained in 1.6.

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Instructions for Tecnai a brief start up manual

2 How to operate the microscope/obtain an image

User Operation of JEOL 1200 EX II

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1. Specimen Holder Removal, Loading, and Insertion

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

FEI Tecnai G 2 F20 Operating Procedures

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

Transmission electron Microscopy

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

SEM Training Notebook

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

NANO 703-Notes. Chapter 9-The Instrument

SEM Training Notebook

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

Tecnai T12 Operating Procedures

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

05/20/14 1. Philips CM200T. Standby Condition

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

Tecnai on-line help manual --

S200 Course LECTURE 1 TEM

STANDARD OPERATING PROCEDURE: JEOL TEM-2100

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7. Chapter 2 CHEM Fall L. Ma

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.

INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY

Section 1: TEM parts and functions... 2

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Physics 3340 Spring 2005

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

1.6 Beam Wander vs. Image Jitter

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

Introduction: Why electrons?

CM20 USER GUIDE. Duncan Alexander, CIME 2010

ECEN 4606, UNDERGRADUATE OPTICS LAB

HOLOGRAPHY EXPERIMENT 25. Equipment List:-

Instruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

Systems Biology. Optical Train, Köhler Illumination

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope

SUBJECT: PHYSICS. Use and Succeed.

Chapter 2 Alignment C. Robert Bagnell, Jr., Ph.D., 2012

BEAM HALO OBSERVATION BY CORONAGRAPH

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

A few concepts in TEM and STEM explained

Physics 1520, Spring 2013 Quiz 2, Form: A

p q p f f f q f p q f NANO 703-Notes Chapter 5-Magnification and Electron Sources

Model SU3500 Scanning Electron Microscope

JEOL 2010 FasTEM & DigitalMicrograph User's Guide

FEI Falcon Direct Electron Detector. Best Practice Document

Introduction to Electron Microscopy

Be aware that there is no universal notation for the various quantities.

Using the Hitachi 3400-N VP-SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

TEM theory Basic optics, image formation and key elements

JEOL 6500 User Manual

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Week IX: INTERFEROMETER EXPERIMENTS

Basics of Light Microscopy and Metallography

The following article is a translation of parts of the original publication of Karl-Ludwig Bath in the german astronomical magazine:

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

OPTICAL SYSTEMS OBJECTIVES

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS. GUI Simulation Diffraction: Focused Beams and Resolution for a lens system

7. Michelson Interferometer

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

ELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu

Marine Invertebrate Zoology Microscope Introduction

Operating F20/F30 with SerialEM

Physics 3340 Spring Fourier Optics

Exam 4. Name: Class: Date: Multiple Choice Identify the choice that best completes the statement or answers the question.

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Laboratory Introduction

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

1. Preliminary sample preparation

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. JY/T

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

JEOL 6700 User Manual 05/18/2009

FE-SEM SU-8020 Operating manual (Preliminary version)

SEM OPERATION IN LOW VACUUM MODE

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com


Transcription:

MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an image Report Requirements No report is required for this lab. You should, however, make extensive notes for your own use. Alignment Procedures I. Pivot Point Adjustment In the next step, we will use the BRIGHT TILT controls to set the beam parallel to the optic axis, but it is first necessary to ensure that the tilting operations are done about a pivot point located in the specimen plane. 1. Load a specimen in the specimen holder and complete all of the alignments that you learned in Lab. 2. The adjustments in this Lab will not work unless these are completed first. 2. With the specimen in focus, and at the correct height, select a magnification of about 40kX. [Alternately, with no specimen in the beam path, adjust the objective focus so that the objective lens current reads 7.04 ma.] 3. Focus the illumination to a small spot (BRIGHTNESS) and center it on the screen (SHIFT). 4. Engage the IMAGE X wobbler. This tilts the beam to its maximum deflection in the X-axis, alternating between positive and negative beam tilts. You may see two illumination spots, corresponding to these deflections. 5. Bring the two spots together, using the IMAGE WOBBLER ADJ X controls. [note always make certain that you are grasping the correct pair of knobs.]

6. Disengage the IMAGE X wobbler. 7. Engage the IMAGE Y wobbler. This tilts the beam to its maximum deflection in the Y-axis, alternating between positive and negative beam tilts. You may see two illumination spots, corresponding to these deflections. 8. Bring the two spots together, using the IMAGE WOBBLER ADJ Y controls. 9. Disengage the IMAGE Y wobbler. 10. Recenter the illumination with the SHIFT controls. II. Current Centering This procedure makes the electron beam parallel to the optic axis of the objective lens. This is accomplished by changing the focal length of the lens. If the beam is parallel to the optic axis, then an object at the center of the screen should remain at the center. If the beam is tilted, the image at the center of the screen will move. 1. Select a magnification below 50kX. 2. Locate a small, distinct image feature and place it at the center of the screen. 3. Center the illumination and spread it to at least fill the screen, thus providing planewave illumination. 4. Select BRIGHT TILT on the DEFLECTOR control. 5. Engage the OBJ WOBBLER. This continuously cycles the objective lens excitation above and below its set point. The image will go in and out of focus, and will rotate as the spiraling action of the lens varies. 6. Use the DEFLECTOR controls to make the image stationary at its center, so it rotates about this point. You may wish to use the small screen and binoculars; this allows you to concentrate on the center point of the screen, allowing less distraction. 7. Disengage the BRIGHT TILT and OBJ WOBBLER controls. III. High Voltage Centering This technique is used to set the illumination parallel to the objective lens axis at higher magnifications. It is more sensitive than current centering. Instead of varying the objective lens excitation, we now vary the accelerating voltage, in a saw-tooth fashion, causing the electron wavelength to be changed. The fixed lens excitation then provides a varying focal length according to the varying electron wavelengths. 1. Select a magnification above 50kX. 2. Locate a distinct image feature, place it at the center of the screen and focus. 3. Center the illumination and spread it to fill the screen, thus providing plane-wave illumination. 4. Select BRIGHT TILT on the DEFLECTOR control. 5. Engage the HT WOBBLER. This continuously cycles the accelerating voltage. The image will go in and out of focus, and will rotate as the spiraling action of the lens varies.

6. Use the DEFLECTOR controls to make the image stationary at its center, so it rotates about this point. Again, you may wish to use the small screen and binoculars. 7. Disengage the BRIGHT TILT and HT WOBBLER controls. This adjustment must be repeated periodically during any high magnification or high resolution work. IV. Image Focus. This is the process of changing the objective lens current, to alter its focal length, and bring the image into focus. The objective lens current value that corresponds to the proper focus position is about 7.04mA. This value, of course, depends upon the specimen height first being set correctly, using the Z control. There are two methods of focusing the image: 1. Rely upon your experience in identifying a focused image. You should select an area which is the thinnest, most-curved region near an edge, if possible. An overfocused image (lens too strong) produces a dark Fresnel fringe around denser objects in the specimen. An underfocused image (lens too weak) produces a bright Fresnel fringe. The ideal focus is obtained when there is no fringe. Contrast is also minimized at the point of focus. Both of these effects will be most noticeable at an edge of a very thin specimen region. 2. Alternatively (though not ideal for high-magnification work) you can use the IMAGE WOBBLER, either X or Y. When the wobbler is engaged, the image appears to oscillate if it is out of focus. Adjust the FOCUS control to eliminate the oscillation. Do not forget to disengage the WOBBLER. V. Obtaining a Diffraction Pattern 1. Select MAG1 or MAG2. 2. Center the electron beam using the SHIFT knobs. 3. Spread the beam to fill the screen (or more) using the BRIGHTNESS control. 4. Locate a suitable part of the specimen, and place it at the center of the field of view. 5. Insert a Selected Area Diffraction (SAD) aperture, and center it carefully. 6. Press DIFF. A diffraction pattern will appear on the screen. 7. Focus the pattern using the DIFF FOCUS control. Note that you can return to the image by strengthening the diffraction lens. Note the effect of the BRIGHTNESS control on your ability to form a wellfocused diffraction pattern. VI. Obtaining Convergent Beam Diffraction Pattern 1. Select MAG1 or MAG2.

2. Center the electron beam using the SHIFT knobs. 3. Spread the beam to fill the screen (or more) using the BRIGHTNESS control. 4. Locate a suitable part of the specimen, and place it at the center of the field of view. 5. Using the BRIGHTNESS control, focus the illumination to the smallest possible spot (crossover). You may wish to turn off the Bright 16X button when doing this. 6. Press DIFF. A diffraction pattern will appear on the screen. Instead of spots, however, the pattern should consist of discs. 7. Focus the pattern using the DIFF FOCUS control. In this case, the edge of the central disk will be sharpest at good focus, fuzzy when out-of-focus. Obtaining a Bright Field Image This is normally done after obtaining a diffraction pattern on the screen. 1. Insert an objective aperture and center it on the transmitted beam. 2. Select MAG1 to return to image mode. 3. Select the desired magnification. 4. Focus the image.

Note: The following sections are provided for your information. Image recording on film is not part of this laboratory exercise, but is something you may wish to do in future. VII. Image Recording on Film While the computer is convenient and suitable for recording many images, the best quality, resolution and permanence is obtained by using sheet film. 1. Select the desired magnification. 2. Bring the features to be recorded within the borders of the framed screen. 3. In order to measure the exposure for a small area, place the small screen in the beam path and place the area of interest on the small screen; leave the small screen in place for the remainder of the process. For an exposure time averaged over the whole image, remove the small screen from the beam path. 4. Adjust the image brightness so the EXPOSURE TIME displayed on Page 1 reads less than 5 seconds, if possible. Longer exposures tend to exhibit the effects of specimen drift. 5. If the camera is in MANUAL advance mode, press the PHOTO button once to advance a piece of film into position for exposure. In AUTO film advance mode, this process is handled automatically. 6. Cover the viewing port. 7. Depress PHOTO a second time to expose the film, being careful to avoid any vibration of the microscope. The PHOTO lamp will turn on while the shutter is in place. VIII. Film Processing 1. Open the camera chamber by turning the handle clockwise by 90 degrees. Open the valve on the nitrogen tank in order to fill the camera to atmospheric pressure. The door will open in about two minutes. 2. Close the valve on the nitrogen tank after the door opens. 3. Wearing gloves, remove the receiving cassette from the camera chamber. Replace it with an empty one from the desiccator. 4. Remove the dispensing cassette and replace it with a full one from the desiccator. 5. Check the O-ring for any lint or deposits. 6. Close the camera door and turn the handle anti-clockwise by 90 degrees. This starts the pumping cycle. 7. Take the camera cassettes to the dark room. Process the used film. Reload the dispensing cassette. Place the filled dispensing cassette and the empty receiving cassette in the desiccator.