HIGH ACCURACY RESISTANCE DECADE BOXES TYPE RBB 4, 5 & 6 DECADES

Similar documents
1433 SERIES. High-Accuracy Decade Resistor User and Service Manual

R-X SERIES. Decade Resistor

Available at Surplus Sales Contents INTRODUCTION... 1 SPECIFICATIONS... 3 OPERATION... 5 MAINTENANCE... 8 Figures

Testing & Calibration Lab, 204, Diamond Industrial Estate No. 2, Ketki Pada Road, (Near Dahisar Toll Naka), Dahisar (East), Mumbai, Maharashtra

ULTRA-PRECISE AIR RESISTANCE STANDARDS

AC/DC Resistance Standard DRR-112

Standard Resistors Precision DC Shunts Digital Thermometers Non-inductive AC Shunts Intelligent Teraohmmeter Precision Decade Standards Precision Air

(RMG100) TO247 Power Pulse-loading Resistors. Token Electronics Industry Co., Ltd. Version: January 12, Web:

Precision in Practice Achieving the best results with precision Digital Multimeter measurements

PRECISION TEMPERATURE SYSTEMS. Compact Modular and Upgradeable Thermometry Measurement Systems

MODEL INFORMATION. AccuBridge Model 6010D. Automated Primary Resistance/ Thermometry Bridge

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT HARS-LX. Laboratory Standard Decade Resistance Substituter User and Service Manual

742A Series Resistance Standards

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

MODEL INFORMATION MODEL 6010D. Automated Primary Resistance/ Thermometry Bridge

5075 Precision Digital Multimeter Time Electronics Calibration, Test & Measurement

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT. HRRS 5kV AND 10kV SERIES. High Resistance, 5kV and 10kV Decade Substituter User and Service Manual

High-precision process calibrator Model CED7000

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Solutions In Calibration TRANSMILLE PRODUCT GUIDE 2015/16 PROVIDING SOLUTIONS IN CALIBRATION WORLDWIDE. www. transmillecalibration.

Temperature References for Highest Accuracy Industrial Thermocouple Measurements

Types CDG 11 and CDG 16 Inverse Time Overcurrent and Earth Fault Relay

Isolated Linearized 4-Wire RTD Input 5B35 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Primary-standard resistance thermometry bridge Model CTR9000

Dimensional Up to 25 mm/ mm 2.0 µm 0 mm to 100 mm/ 0.01 mm 7.4 μm 100 mm to 300 mm/ 0.01 mm 10 μm

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

AC/DC Current Probe GCP-100 QUICK START GUIDE ISO-9001 CERTIFIED MANUFACTURER

Cost Effective Techniques used to Validate the Performance of the microk Resistance Thermometry Instrument with sub mk Uncertainty

6622A Series. Direct Current Comparator Resistance Bridges 6622A SERIES FEATURES

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:

High-Precision Resistance Decade and Calibrator. High-Precision Resistance Decade and Calibrator. Model 1405, 1406, Decades

1210N 1250N MEGOHMMETER. User Manual 15 V 500V kV. MEGOHMMETER Model 1250N 600V CAT II / 1000V CAT I! 30. x 10.

Model AccuBridge 6242D Resistance Bridge

Time Electronics 1050 PT100 Element Simulator

Maximizing your reference multimeter, minimizing measurement uncertainties

THE ELECTROM itig II MOTOR TESTER AND WINDING ANALYZER

Calsytech, # 38 North Mada Street Nandambakkam, Chennai, Tamil Nadu. Discipline Electro-Technical Calibration Issue Date

5080A Multi-Product Calibrator

LCT100. Load Cell Tester User s Guide (v1711) Anyload Transducer Co. Ltd Website:

CALYS 50. Field multifunction calibrator for basic use

2768 Precision Wheatstone Bridge

DM-45 Digital Multimeter

Measurement of Resistance and Potentiometers

SOLICITUD XXVI CONGRESO NACIONAL DE METROLOGIA

CALYS 150. Advanced documenting multifunction calibrator thermometer

Analogue Demo Multimeter ADM

in the genrad tradition

MD-5060 DIGITAL MEGOHMMETER

CALYS 75. Field documenting multifunction calibrator

Calibration Department Address Yard No 304 Road 4306 Block 343 Mina Salman Industrial Area Manama, Bahrain. 0 mm to 600 mm / 0.01 mm 7.

Model 6622T Series. DCC Thermometry Bridges 6622T SERIES FEATURES

Gertsch Products RatioTrans are high precision AC Voltage Dividers. They have the following useful features. 1. EXTREME ACCURACY

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

PHASE ROTATION METER. Operating and Instruction Manual. a n d A C C E S S O R I E S

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

The NEW itig II Winding and Motor Analyzer A revolution in automated, safe, and comprehensive high and low voltage tests all in a lightweight

MEGOHMMETER. User Manual

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C

(HI80) Ultra-Precision High-Power High-Voltage Resistors. Token Electronics Industry Co., Ltd. Version: November 22, Web:

Automated Thermocouple & RTD Calibrator

Datasheet Platinum Resistance Pt100 In-head (Push Button) Temperature Transmitter TX203P mA output, default range supplied C

Model 6625T Series TEMPERATURE MEASUREMENT SYSTEMS

DIGITAL PANEL METER N24 AND N25 METER SERIES

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10

Technical Services Department, BHEL, Piplani, Bhopal, Madhya Pradesh. Discipline Electro Technical Calibration Issue Date

Discover the. Blue Box. Difference. Electrical and Temperature Metrology Products Guide

M.I. Cable High Resistance Fault Locator

CERTIFICATE OF ACCREDITATION

RCTrms Technical Notes

CALYS 150. Advanced documenting multifunction calibrator thermometer

DIGITAL MULTIMETER OPERATING INSTRUCTIONS MODEL CDM-35. Part No

CALYS 100. Field precision documenting multifunction calibrator

Bulk Metal Foil Technology Hermetically Sealed Power and Current Sensing Resistor with TCR of 2 ppm/ C and Power up to 10 Watts

OPERATION MANUAL OF MODEL TH2681/A HIGH RESISTANCE TESTER

How accurate is a measurement? Why should you care? Dr. Andrew Roscoe

Additel 875 Series Dry Well Calibrators

50W TO220 High Power Resistors

Electromagnetic shakers operating guide Models F3, F4, F5B, F10

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Hand-held temperature calibrator Model CEP3000

HI-Physix Laboratory, L-83, Sector-1, DSIDC Industrial Area, Bawana, Delhi. Discipline Electronics Testing Issue Date

AL ANSARI TRADING ENTERPRISE LLC P.O. BOX: 1832 PC: 112 RUWI, MUSCAT OMAN

Operator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique

Discover. Blue Box. the. Difference. High Resistance Metrology Products Guide

CALYS 150. Advanced documenting multifunction calibrator thermometer

Integrated calibration system for electronic and pneumatic instrumentation model CBS-SC2000

Product Information Measuring resistor with the Pt100-sensor referring to DIN EN Pt100, Pt500, Pt1000

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C

APPLICATION NOTE. Wide Range of Resistance Measurement Solutions from μω to PΩ

UNIT II MEASUREMENT OF POWER & ENERGY

Voltage Transformers type UMZ User s Manual, Operation and Maintenance

MODEL INFORMATION. MODEL AccuBridge. Self Calibration Ratio Bridge

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT Series. Primary Standard Inductor User and Service Manual

Cryogenic Temperature Controllers Model 32/32B

USER MANUAL 600A AC Clamp Meter + NCV Model MA610

Model 32/32B. Major highlights of the Model 32:

Transcription:

HIGH ACCURACY RESISTANCE DECADE BOXES TYPE RBB, & DECADES

CONTENTS HIGH ACCURACY RESISTANCE DECADE BOXES..., & DECADES... Residual Resistance... Low Value Decades... Operation... Connection... Operation & Maintenance... 1

HIGH ACCURACY RESISTANCE DECADE BOXES RBB, & DECADES The RBB series of resistance boxes are available in, & decades. High Accuracy and wide range 0.001 ohm to 11 Megohms are combined in a compact case. The switches have gold plated contacts to ensure a low contact resistance and negligible thermal emf. Some models are particularly suited to Pt100 simulation with resolution as low as 0.001 ohm (±0.00 C). The Switches Cropico Type SP1 Contact Resistance <1mΩ/Decade Insulation Resistance (all paths >10GΩ) Resistance Coils Temperature Co-efficient ±ppm maximum over - C to +1 C 0.1, 0.01 and 0.001 dials 10ppm/ C Full Load Stability ±ppm/10,000 hours ±0ppm/,000 Hours No Load Stability ±ppm/10,000 hours ±ppm/,000 hours over full temperature range; - C to 1 C Power Rating 0. watt (+8 C) 0. watt (+110 C) Maximum Continuous Working Voltage Up to 0V DC Noise Essentially non-measurable Thermal EMF <0.µV/ C typical <1.µV/ C max Encapsulation Moulded Epoxy Leads SWG tinned copper Windings Exclusive air cushion technique provides virtually stressless elements for improved performance. Non inductively wound. Direction of winding reversed at half turns point

Models to choose from Model Number of Decades Total Resistance Resolution Residual Resistance Size Weight RBB-A 1,111 Ω 0.1 Ω 0.01Ω 10x100x10. kg RBB-B 11,110 Ω 1 Ω 0.01Ω 10x100x10. kg RBB-C 111,100 Ω 10 Ω 0.01Ω 10x100x10. kg RBB-B 1,11.1 Ω 0.01 Ω 1Ω 8x100x10. kg RBB-C 11,111 Ω 0.1 Ω 0.01Ω 8x100x10. kg RBB-D 111,110 Ω 1 Ω 0.01Ω 8x100x10. kg RBB-E 1.1111MΩ 10 Ω 0.01Ω 8x100x10. kg RBB-F 11.111MΩ 100 Ω 0.01Ω 8x100x10. kg RBB-B 1,11.11 Ω 0.001 Ω 1Ω 0x100x10.8 kg RBB-C 11,11.1 Ω 0.01 Ω 1Ω 0x100x10.8 kg RBB-D 111,111 Ω 0.1 Ω 0.01Ω 0X100X10.8 kg RBB-E 1.11111MΩ 1 Ω 0.01Ω 0x100x10.8 kg RBB-F 11.1111MΩ 10 Ω 0.01Ω 0X100X10.8 kg Accuracy of Decades Decade (Ohms) Accuracy of adjustment Incremental Steps Power Reading Temperature Coefficient 10 x 1 M Ω ± 0.1 % 0. W 10ppm/ C 10 x 100 k Ω ± 0.1 % 0. W 10ppm/ C 10 x 10 k Ω ± 0.0% 0. W ppm/ C 10 x 1 k Ω ± 0.0 % 0. W ppm/ C 10 x 100 Ω ± 0.0% 0. W ppm/ C 10 x 10 Ω ± 0.0% 0. W ppm/ C 10 x 1 Ω ± 0. % 0. W ppm/ C 10 x 0.1 Ω ± 0. % 0. W 10ppm/ C 10 x 0.01 Ω ± 1 % 0. W 10ppm/ C 10 x 0.001 Ω ± % 0. W 10ppm/ C 10 x 0.0001 Ω ± 10 % 0. W 10ppm/ C The RBB series of decade resistance boxes has been designed to be robust, reliable, and require the minimum maintenance. The switches should serviced and cleaned annually to ensure best performance. The resistance elements are factory adjusted to be within specification and it is not possible to alter their value. In the event of a resistor becoming damaged it is necessary to replace the whole resistor. Residual Resistance With all the decade dials set at 0 there will be a resistance measured at the terminals, this resistance is made up form components. The resistance of the wires connecting the decade

switches to the terminals; this is a constant value. The contact resistance of the decade switches, this will vary each time the switch is turned or disturbed. This residual resistance must be added to the decade values set on the decades, it is therefore important that this residual should be as low and constant as possible. The CROPICO decade boxes have a low and stable residual resistance per decade. This value is controlled by the use of special low resistance switches. It is not practical to manufacture decade resistance boxes where the lowest decade is less than 0.1Ω. To switch in steps of 0.01Ω and lower, special techniques need to be employed. Low Value Decades When switching decade resistance of 0.01Ω and lower the switch contact resistance variations become significant, this variation can be typically 10mΩ for an decade unit and 1mΩ for a decade unit. In addition the total residual resistance which would also include the internal wiring of the decade box could be significantly higher than the lowest decade values. CROPICO overcome this problem by using the Wagner Wolf decade, this technique enables higher resistance values to be used for the decade which are then shunted to give the correct incremental steps. The residual value of the unit is deliberately increased to 1Ω, and the 1Ω decade will start at 1 and not zero. This is not normally a problem, because the applications that require values to be selected with high resolution would have a much greater full scale resistance, typical applications include the simulation of Pt100 temperatures. The Pt100 resistance thermometers have a resistance of 100Ω at 0 C and 0.01 C is approximately equal to 0.001Ω. Operation To select a resistance value turn the dial of the appropriate decade until the desired value is indicated by the switch pointers. The multiplier for each decade is shown below the dial i.e. x 0.1 or x 10 etc. Example to select 1,. Ω x 10kΩ x 1kΩ x 100Ω x 10Ω x 1Ω x 0.1Ω

Connection The connections to the resistance box are made via the two mm terminals marked with the resistance symbol either banana plugs or spade terminals may be used. The third terminal marked with an earth symbol is connected directly to the top panel. RX Connection Terminals Earth Terminal *********************************************************

OPERATING AND MAINTENANCE INSTRUCTIONS These Resistance Boxes are supplied calibrated to specification and ready for use. The specified accuracy refers to the dial setting. To maintain the calibrated accuracy, avoid subjecting the box to thermal shocks or rough handling Prior to having the box re-certified by e.g. a UKAS Laboratory, it is good practice to return the box to us first for service of the switch contacts. 1. Before using, at all times rotate each dial twice in both directions to ensure a constant contact resistance.. When using in a voltage carrying circuit, to determine the correct resistance value to be used, always start by setting the dials at a high value. This avoids the possibility of overloading the resistance box and possible permanent damage to the resistance coils.. Remember that the box has a residual resistance which must be added to the final reading. This value is defined in the specification and is usually insignificant. The low value boxes with decades of 10 x 0.01, 0.00 1 and 0.0001 are constructed using the WAIDNER-WOLFF system, in which case the residual resistance is clearly defined, as the 1 ohm dial starts at I and not 0.. The models with low value decades are often used for resistance thermometer simulation, in which case the following data may be useful. Platinum resistance thermometers ballasted to give a total resistance of 100 ohms at U C. Fundamental interval 8.0 ohms Degrees C Resistance Ohms BS190/I.E.C. 1-100 0. -0 80.1-90.19 0 100.00 ± 109. ±0 119.0 ±100 18.0 ±10 1.1 +00 1.8 +0 19.0 +00 1.0 +0 9. +00.0 +0.11 +00 80.90 +0 9.9 +00 1.9 Calibration Temperature 0 0 C Proofing voltage to the earth connection kv

Prior to despatch, each box is subjected to 100% inspection and issued with a Certificate of Conformity. Information regarding traceability to the National Standard is also supplied. The only maintenance that these Resistance Boxes should require is the cleaning of the switch contacts. They should be inspected annually and if the contacts are dirty, or tarnished due to lack of use, proceed as follows: a) Remove the instrument from its box by unscrewing the screws around the outer edge of the top panel. The instrument can than be removed from its case. b) The switches can now be seen and cleaned as follows: Clean all contacts with a lint-free cloth and lightly lubricate all contacts with SGB Contact Treatment Grease. Do not use aerosol cleaning and degreasing agents unless absolutely necessary. If they are used, make sure that the contact grease is applied to all the stud and centre return contacts before re-assembly. c) After greasing the switch contacts, rotate each switch rapidly backwards and forwards 10 times. The resistors will remain within specification for many years and with normal usage should never need attention. However, accidents do happen, and should they be burnt out due to overloading, the connections to the faulty resistors can be unsoldered and the resistor replaced. Order replacement resistors are as follows: One Note ohm resistor for decade box type RBB or RB a dial marked 10 x 100 would have 10 x 100 ohm resistors wired in series.

Additional Products From Cropico DIGITAL OHMMETERS RESISTANCE BRIDGES RESISTANCE DECADE BOXES RESISTANCE STANDARDS DIGITAL THERMOMETERS THERMOCOUPLE SIMULATORS THERMOCOUPLE REFERENCE JUNCTIONS THERMOCOUPLE SELECTOR SWITCHES PT100 SIMULATORS PRECISION WATTMETERS INSULATION TEST SETS UNIVERSAL CALIBRATORS POWER ANALYSERS HIPOT TESTERS ELECTRICAL SAFETY TESTERS Phone, fax or e-mail for further information on the above or for a copy of our general catalogue:- Phone:- 00 88 0 Fax:- 00 88 09 E-mail:- enquiries@cropico.com Website:- www.cropico.com 8