Experiment 1 Topic: Sensors/Measurement Systems/Calibration Week A Procedure

Similar documents
Experiment A6 Solar Panels I Procedure

Experiment 3 Topic: Dynamic System Response Week A Procedure

Experiment 4 Topic: Solar Panels Week A Procedure

EXPERIMENT 2: STRAIN GAGE DYNAMIC TESTING

DSC Lab 2: Force and Displacement Measurement Page 1

Experiment 3 Topic: Dynamic System Response Week A Procedure

Class #9: Experiment Diodes Part II: LEDs

Experiment P01: Understanding Motion I Distance and Time (Motion Sensor)

Section 2 Lab Experiments

Voltage Current and Resistance II

Experiment A2 Galileo s Inclined Plane Procedure

MAKE SURE TA & TI STAMPS EVERY PAGE BEFORE YOU START

ME 461 Laboratory #5 Characterization and Control of PMDC Motors

Experiment P31: Waves on a String (Power Amplifier)

LabVIEW Based Instrumentation and Experimental Methods Course

LAB #5: Measurement of Strain

Exercise 2: Temperature Measurement

EET 150 Lab Activity 13 On-Line Students Temperature Logging and Display Analog Discovery 2

ENGN/PHYS 207 Fall 2018 Assignment #5 Final Report Due Date: 5pm Wed Oct 31, 2018

EE 210: CIRCUITS AND DEVICES

ENGR 1121 Lab 3 Strain Gauge

Motomatic Servo Control

Ph 3455 The Photoelectric Effect

Onwards and Upwards, Your near space guide

Sensor Calibration Lab

Lab 2b: Dynamic Response of a Rotor with Shaft Imbalance

Experiment P10: Acceleration of a Dynamics Cart II (Motion Sensor)

Experiment A8 Electronics III Procedure

NAME SECTION PERFORMANCE TASK # 3. Part I. Qualitative Relationships

Experiment A3 Electronics I Procedure

Laboratory 14. Lab 14. Vibration Measurement With an Accelerometer

Experiment A8 Electronics III Procedure

Lab 12 Microwave Optics.

Lab 4 Ohm s Law and Resistors

Resistance Apparatus EM-8812

Experiment P55: Light Intensity vs. Position (Light Sensor, Motion Sensor)

Experiment P02: Understanding Motion II Velocity and Time (Motion Sensor)

MICROTC LINEAR THERMOCOUPLE AMPLIFIER OPERATOR S MANUAL

18600 Angular Momentum

MicroLab 500-series Getting Started

Experiment P49: Transistor Lab 2 Current Gain: The NPN Emitter-Follower Amplifier (Power Amplifier, Voltage Sensor)

Experiment P20: Driven Harmonic Motion - Mass on a Spring (Force Sensor, Motion Sensor, Power Amplifier)

Computer Tools for Data Acquisition

Free vibration of cantilever beam FREE VIBRATION OF CANTILEVER BEAM PROCEDURE

Activity P40: Driven Harmonic Motion - Mass on a Spring (Force Sensor, Motion Sensor, Power Amplifier)

Acoustic Resonance Lab

Exercise 2: Temperature Measurement

Vibration Fundamentals Training System

PHOTOGATE 0662I WITH PULLEY ATTACHMENT User s Guide

MAE334 - Introduction to Instrumentation and Computers. Final Exam. December 11, 2006

Lab assignment: Strain gauge

Load Cells, LVDTs and Thermocouples

Name Date: Course number: MAKE SURE TA & TI STAMPS EVERY PAGE BEFORE YOU START EXPERIMENT 10. Electronic Circuits

Introduction to the Op-Amp

Experiment 15: Diode Lab Part 1

Experiment P11: Newton's Second Law Constant Force (Force Sensor, Motion Sensor)

Chapter 1: DC circuit basics

Sensor Calibration Lab

FYSP1110/K1 (FYSP110/K1) USE OF AN OSCILLOSCOPE

Single-Slit Diffraction. = m, (Eq. 1)

EK307 Introduction to the Lab

Experiment P58: Light Intensity in Double-Slit and Single-Slit Diffraction Patterns (Light Sensor, Rotary Motion Sensor)

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

Experiment 13: LR Circuit

sin(wt) y(t) Exciter Vibrating armature ENME599 1

THE AMPLIFIER. A-B = C subtractor. INPUTS Figure 1

Standing Waves. Equipment

LABORATORY 2: Bridge circuits, Superposition, Thevenin Circuits, and Amplifier Circuits

EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES

EE 210 Lab Exercise #5: OP-AMPS I

26 Sep. 10 PHYS102 2

2 Oscilloscope Familiarization

MEC751 Measurement Lab 2 Instrumented Cantilever Beam

Physics 4C Chabot College Scott Hildreth

2 AC and RMS. To pass this lab you must solve tasks 1-2. Tasks 3 and 4 are included in the grading of the course.

Response spectrum Time history Power Spectral Density, PSD

Introduction. ELCT903, Sensor Technology Electronics and Electrical Engineering Department 1. Dr.-Eng. Hisham El-Sherif

Physics 476LW. Advanced Physics Laboratory - Microwave Optics

This manual describes the Motion Sensor hardware and the locally written software that interfaces to it.

Experiment: P34 Resonance Modes 1 Resonance Modes of a Stretched String (Power Amplifier, Voltage Sensor)

Projectile Motion. Equipment

Physics 2020 Lab 9 Wave Interference

How to do the Thermal Noise Lab. And also your DNA melting lab report

Chapter 7: Instrumentation systems

The Photoelectric Effect

ELECTRICAL ENGINEERING TECHNOLOGY PROGRAM EET 433 CONTROL SYSTEMS ANALYSIS AND DESIGN LABORATORY EXPERIENCES

INSTALLATION AND OPERATING INSTRUCTIONS FOR FARRAND ABSOLUTE WITH INCREMENTAL CONVERSION SYSTEM BOARD

EE 241 Experiment #7: NETWORK THEOREMS, LINEARITY, AND THE RESPONSE OF 1 ST ORDER RC CIRCUITS 1

Electronic Simulation Software for Teaching and Learning

Introduction to Measurement Systems

Modern Physics Laboratory MP4 Photoelectric Effect

ANALOG TO DIGITAL CONVERTER ANALOG INPUT

Dumpster Optics BENDING LIGHT REFLECTION

Physics 2310 Lab #2 Speed of Sound & Resonance in Air

electrical noise and interference, environmental changes, instrument resolution, or uncertainties in the measurement process itself.

MFJ-208 VHF SWR Analyzer

Physics 303 Fall Module 4: The Operational Amplifier

Single Slit Diffraction

MEASUREMENT AND INSTRUMENTATION QUESTION BANK UNIT I INTRODUCTION. Part A

Notes on Experiment #1

Transcription:

Experiment 1 Topic: Sensors/Measurement Systems/Calibration Week A Procedure Laboratory Assistant: Michael Wicks and Erik Ross Email: mwicks@nd.edu and eross2@nd.edu Office/hours: 2/7 2/10, 5 pm - 6 pm, B14 E-1 Website: http://www.nd.edu/~jott/measurements_lab/e1/ Prior to performing the experiments, review relevant sensor data sheets and videos. Also, you or your partner should bring a flash drive to lab. Overview: In this laboratory exercise, the students will conduct a series of two experiments, Beerless Pong and Water Conservation, each involving the use and calibration of several sensors. These sensors include a photocell and 3-axis accelerometer. Data sheets for each of these sensors can be found on the course website. Additionally, the students will use strain gages to measure displacement and strain. Various signal processors for the sensors will be used, including a simple sensor interface box(sib). The course website has links to videos demonstrating the sensor and SIB. Part 1: Beerless Pong: Single-handedly responsible for the loss of countless engineers to another unnamed college, beerless pong is one of the most widely played games among college students today. However, despite its mass appeal the game has not changed much over the past decade, until now. Using a cantilever mass measurement system based on strain gages and a Wheatstone bridge, you will perform a proof-of-concept experiment for a Beer Pong 2.0 system, specifically demonstrating a new key design feature, automatic scoring. Setup and Data Acquisition 1. The cantilever apparatus is shown in Figure 1. It consists of a cantilever beam affixed with strain gages, an amplifier, and a integrated voltmeter that reads the output voltage from the bridge. The bridge has already been configured as a full bridge in deflection mode. To set the input voltage to the Wheatstone bridge, set the Signal Select button to Bridge Excite. The red LED display should indicate now the bridge excitation voltage. Set it to 5 V using the turning knob in the upper left corner. To read the bridge output voltage, set the Signal Select button to Amplifier Out. The LED display now shows the output bridge voltage, which has been amplified by an op-amp that is inside the voltmeter box. File: E1a_procedure.docx.docx 1 Last Revision 9/5/2016

Figure 1. Cantilever beam and Wheatstone bridge apparatus 2. Balance the Wheatstone bridge so that the output voltage is ~0 V. Make sure that there is no weight attached to the beam. The range of measured masses should be up to 50 g. Take your calibration weights and place 50 g in the Red Dixie Cup. Then change the amplification factor so that the output voltage is ~2 V. 3. Remove the weights and balance the Wheatstone bridge again. At this point, the settings on the voltmeter should not be modified (the amplification, the bridge balance resistance, or the input voltage) because the calibration will only be valid for these settings. 4. Calibrate the force balance by sequentially adding known weights and recording the resultant voltage. Because ping-pong balls are light, it is suggested to start at no weight and increase in small increments (5-10 g). At least 5 data points are required to generate the calibration curve. 5. Sequentially add ping-pong balls to the cup while recording the resultant voltage after adding each ball. Using this data and the actual weight (published value) of one pingpong ball, one can estimate the accuracy of the measurement system. Part 2: Signal Amplification Study Beerless Pong 3.0: Recall that a measurement device that measures mass served as inspiration for our Beer Pong 2.0 system. However, while this approach would be sufficient for automatic scoring, it does not provide a mechanism for team recognition. So for Beer Pong 3.0, we will take an optical approach. Rather than attempting to measure the mass, we will measure whether the balls actually make it into the cup using the photocell sensor. Further, we will try to distinguish between teams using red balls and those using blue balls. File: E1a_procedure.docx.docx 2 Last Revision 9/5/2016

Background Photocells use semi-conductors whose resistance changes upon exposure to light. Importantly, though, they respond differently to different parts of the electromagnetic spectrum the photocells used in this lab are more sensitive in the visible spectrum as opposed to the infrared and ultraviolet spectrums. Referring to the photocell data sheet, it is clear that a CdS photocell is more sensitive for certain wavelengths than others. Since red and blue are at different wavelengths, a CdS photocell should be able to distinguish between them when they are in the cup. Setup and Data Acquisition 1. Connect the photocell to the Sensor Interface Box. Recall from the YouTube tutorials that all the sensors in this lab require three things: power, signal and ground. The photocell sensor uses the red wire for power(5v), blue wire for signal and black for ground. 2. Take and record a dark and light reading by covering the sensor with your finger and holding the sensor to the overhead light, respectively. This is essentially a two point calibration. 3. Tape the photocell inside the cup (about midway) facing the bottom of the cup and record a baseline reading. 4. Insert the red ball and record a baseline reading. Repeat for the blue ball. Note that these measurements are not too different and that our signal needs to be conditioned to make these two readings more distinguishable. 5. Set one channel of the red dual op amp box(dob) gain to switch setting 2. This will set the gain to 10. For the next three instructions, refer to Figure 6 in Appendix A. 6. Connect the black wire on the DOB(channel A) to GND on the sensor interface box. 7. Connect the red wire on the DOB(channel A) to +5V on the sensor interface box. 8. Connect the output of the sensor interface box to the input of channel A(the remaining solid color wire). 9. Connect the output of the DOB(stripe color wire) to the input of the second channel of the sensor interface box. 10. Record the base photocell readings and amplified readings for the following cases: no ball in cup, red ball in cup, white ball in cup and blue ball in cup. (Note: Experience has taught us that a gain on the order of 10 is necessary to see a difference in the red and blue signals.) File: E1a_procedure.docx.docx 3 Last Revision 9/5/2016

Switch Gain Setting 0 100. 1 20 2 10 3 7 4 5 5 4.2 6 3.4 7 3 8 2.0 9 1.9 Figure 6: Wiring Diagram of photocell and red dual operational amplifier box Part 3: Water Conservation: Water is scarce in many places, especially in parts of Asia and Africa, and only becoming scarcer. One possible solution: Use less water. The average human requires 0.5 gal of water a day to survive. How many people could survive on the amount of water it takes to wash your hands? Using the accelerometer, design and calibrate a flow rate measurement system to determine the average amount of water consumed to wash one s hands. Setup and Data Acquisition 1. Connect the accelerometer to the Sensor Interface Box as shown in Fig. 2. Recall from the YouTube tutorials that all of the sensors in this lab require 3 things: power, ground, and signal. In this station, the red wire red is power, black is ground and white is the y-axis signal. Caution: Do NOT connect the red wire to the 5 V supply. It requires the 3.3 V. Failure to follow this step will result in destroying the sensor. File: E1a_procedure.docx.docx 4 Last Revision 9/5/2016

Figure 2. Wire connections for the accelerometer. 2. Using the angle calibration fixture, record the output voltage as a function of pitch angle θ. The accelerometer should be oriented such that the wires are hanging down when it is in the vertical position. When the y-axis sensor is vertical, it corresponds to 9.81 m/s 2 or 1 g. Conversely when the y-axis sensor is horizontal, it corresponds to 0 g. Be sure to include the horizontal and vertical positions in your data. 3. The output voltage is a linear function of acceleration (i.e. Vout = Aay + B). Draw the free body diagram for the accelerometer pitched at an angle θ, and determine the theoretical formula for output voltage vs. pitch angle θ. Define your coordinates such that horizontal corresponds to θ = 0. 4. As shown in Fig. 3, wrap the sensor in plastic; fix it to the shaft so that the y-axis points down and is parallel to the shaft; clamp to the cold water faucet handle. 5. Calibrate the accelerometer measurement system by relating the angle of the faucet handle to flow rate where flow rate is calculated by dividing the volume of water collected in the beaker by the filling time. Record the flow rate and accelerometer voltage reading for at least 5 different angles. (In your analysis, you will use your formula from step 3 to convert the measured voltages to angles.) File: E1a_procedure.docx.docx 5 Last Revision 9/5/2016

(c) Figure 3. Procedure to (a) place sensor in plastic, (b) affix to extension arm, and (c) clamp to cold water knob. File: E1a_procedure.docx.docx 6 Last Revision 9/5/2016

Week A Deliverables You are required to include the following items in your lab report. (See the E1 score sheet for points.) 1. The equation of best fit for the linear calibration of mass vs. voltage for the Wheatstone bridge on the cantilever beam. A plot of the calibration data for the cantilever beam is not necessary. 2. MakeA plot of measured mass vs. the discrete number of ping pong balls(discrete). with the theoretical, continuous line plotted on top. (Use your calibration equation from deliverable #1 to convert the voltages to mass.) Apply a linear curve fit to the data and use the slope to estimate the mass of a single ping pong ball. Plot the curve fit on top of the data as a continuous line. 3. A table containing the recorded photocell voltages and amplified voltages for the different colored balls and the empty cup. 4. A plot of accelerometer voltage vs. pitch angle θ with a derived theoretical trigonometric curve that you came up with. 5. A plot of flow rate vs. the angle of the faucet handle θ. Use the theoretical equation from deliverable 4 to calculate the faucet angle θ from the measured accelerometer voltage. Include a non-linear curve fit that captures the behavior of the system. You may use any type of function you like for the curve fit, except for a polynomial. (Do not use a polynomial!) Suggested Talking Points Estimate the relative error of your measurement in the mass vs. number of ping pong balls. What does this suggest if this configuration were used in the automatic scoring feature on the beer pong 2.0 system? (Would this feature work? Does this signal need amplification? Is the system sufficiently sensitive? Is the calibration sufficient?) Download the data sheet for the photocell on Prof. Ott s website. Use it to explain why the different colored balls produce yield different output voltages. Estimate the average amount of water consumed to wash your hands. The average human requires 0.5 gal of water a day to survive. How many people could survive on the amount of water it takes to wash your hands? File: E1a_procedure.docx.docx 7 Last Revision 9/5/2016

Equipment Appendix A Cantilever Apparatus: (includes mounted strain gages with wire leads, an amplifier, and an integrated voltmeter, configured to read output voltage from full Wheatstone Bridge in deflection mode Ping Pong Balls: (50 both white and colored) 1 g 50g slotted weight set 3 axis Accelerometer with wire leads(female connectors) Accelerometer extension cord: (3 wire - 18 in length; male to female connectors) Digital Protractor Sensor Interface Box (SIB) w/ 9V battery and 9V power supply Water, Sink, and Faucet Faucet Handle Clamp w/12 extension rod Small Plastic Bag Medium Alligator Clamp Stop watch 2 L beaker Red Op Amp Box Set of 3 - Leads (12 - red, yellow, black)(3 prong plastic-snap connector to 3-wire leads(female end connector for Red Op Amp Box) Set of 3 - Leads (12 - red and white striped, yellow and white striped, black and white striped )(3 prong plastic-snap connector to 3-wire leads(female end connector for Red Op Amp Box ) File: E1a_procedure.docx.docx 8 Last Revision 9/5/2016