Optimizing System Throughput with the NI PXI ½-Digit FlexDMM

Similar documents
6 1 2-Digit FlexDMM and 1.8 MS/s Isolated Digitizer

6 1 2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter

Precision in Practice Achieving the best results with precision Digital Multimeter measurements

Fallstricke präziser DC- Messungen

PXIe Contents. Required Software CALIBRATION PROCEDURE

National Instruments Switches

Contents. Software Requirements

EKT 314/4 LABORATORIES SHEET

IVI STEP TYPES. Contents

High Speed Milliohm Resistance Meter

Contents CALIBRATION PROCEDURE NI PXI-5422

Selecting Telecommunication Test Equipment To Maximize Throughput and Accuracy. By Robert Green Keithley Instruments, Inc.

Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size

2302 Battery Simulator 2306, 2306-PJ Battery/Charger Simulators

National Instruments Switches

Agilent 34410A and 34411A Multimeters Setting the Standard for Next Generation Benchtop and System Testing

Racal Instruments 4152A. 6.5 Digit Digital Multimeter. Product Information

Contents. CALIBRATION PROCEDURE NI PXIe-6555/6556. ni.com/manuals

Chapter 6 Specifications

HP 34401A Specifications 8

CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

Keysight 34401A Multimeter

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

PXI Modules 3066 PXI Multi-Way Active RF Combiner Data Sheet

Low Pass Filtering for Checker with Input Noise from Mechanical Relays

Keysight Technologies How to Take Fast, Simultaneous Measurements of Two or More Signals Using BenchVue Software. Application Note

Agilent DC Power Analyzer

EX FEATURES. Stand-alone 48-channel unit with built-in Ethernet controller. Built-in bridge completion and Excitation

Battery Simulator Battery/Charger Simulators

Keysight Technologies 8 Hints for Making Better Digital Multimeter Measurements. Application Note

Fast and Accurate RF component characterization enabled by FPGA technology

USB-TEMP and TC Series USB-Based Temperature Measurement Devices

NI PXI ½ Digit FlexDMM Calibration Procedure

Advances in Antenna Measurement Instrumentation and Systems

90 Day TCAL ±5 C. = channel 2 reading channel 2 accuracy channel 2 reading

How to Optimize Measurement Speed of Your DAQ

National Instruments Flex II ADC Technology The Flexible Resolution Technology inside the NI PXI-5922 Digitizer

Application Note Series. Solutions for Production Testing of Connectors

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

APPLICATION NOTE. Wide Range of Resistance Measurement Solutions from μω to PΩ

NI PXIe-5601 Specifications

Selecting and Using High-Precision Digital-to-Analog Converters

Features EX1629. High-performance Remote Strain Gage Measurement Unit. Overview. 48 programmable channels in a single unit

5075 Precision Digital Multimeter Time Electronics Calibration, Test & Measurement

DC Resistance Meters 2840 Series

NI PXI-2530 Specifications

SCXI 8-Channel Isolated Analog Input Modules

8-Bit A/D Converter AD673 REV. A FUNCTIONAL BLOCK DIAGRAM

Model ½-Digit Multimeter Datasheet

Agilent DC Power Analyzer

Tel: Fax:

NI 4070/4072 6½-Digit FlexDMM

Agilent N3300 Series DC Electronic Loads

Application Note. 8 Hints. for Making Better Digital Multimeter. Measurements

Contents CALIBRATION PROCEDURE NI 5412

Contents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator

Keysight Technologies 6½ Digit PXI Digital Multimeters (M9181A, M9182A, M9183A)

Precision Flash Lamp Current Measurement Thermal Sensitivity and Analytic Compensation Techniques

Keysight Technologies PXI Digital Multimeters 6½ Digit, High Performance (M9182A, M9183A)

NI 272x Help. Related Documentation. NI 272x Hardware Fundamentals

PRECISION TEMPERATURE SYSTEMS. Compact Modular and Upgradeable Thermometry Measurement Systems

Integrated Diode Laser System DioPower

Correlation of Voltage and Temperature Measurement

Signal Conditioning Fundamentals for PC-Based Data Acquisition Systems

Model Multimeter Specifications ABSOLUTE ACCURACY TYPICAL ACCURACIES 2002 SPECIFIED CALIBRATION INTERVALS

EKT 314/4 LABORATORIES SHEET

2601 System SourceMeter 2602 Multi-Channel I-V Test Solutions

Maximizing your reference multimeter, minimizing measurement uncertainties

2.1 Make Accurate Low-Level Measurements with High-Resolution Instrumentation. Incorporating Best Practices for Ensuring Product Quality

ni.com The NI PXIe-5644R Vector Signal Transceiver World s First Software-Designed Instrument

NI PXI/PXIe-2532 Specifications

Considerations for Analog Input and Output

NI PXI-2521 Specifications 40-Channel DPST Relay Module

Agilent dc Electronic Loads Models N3300A-N3307A

Keysight Technologies 34410A and 34411A Multimeters

Racal Instruments. Product Information

ni.com Sensor Measurement Fundamentals Series

2520 Pulsed Laser Diode Test System

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

NJM4151 V-F / F-V CONVERTOR

CHAPTER 6 DIGITAL INSTRUMENTS

Model 32/32B. Major highlights of the Model 32:

CALIBRATION PROCEDURE NI PXIe-4330/4331. Contents

User s Manual for Integrator Short Pulse ISP16 10JUN2016

IXIAN TM class V 1.5. EC Transmitter This is an evolving document. EC Transmitter IXIAN TM. Features

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

6.5-Digit Multimeter, High- Accuracy, C-Size HP E1410A Technical Specifications

True RMS Bench Multimeters 2831E and 5491B

Keysight Technologies Making Simpler DC Power Measurements with a Digital Multimeter

6530 TeraOhm Bridge-Meters

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24

Oscilloscope Measurement Fundamentals: Vertical-Axis Measurements (Part 1 of 3)

SIGNAL RECOVERY. Model 7265 DSP Lock-in Amplifier

User s Manual for Integrator Long Pulse ILP8 22AUG2016

Laboratory 3 (drawn from lab text by Alciatore)

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

Low Cost 10-Bit Monolithic D/A Converter AD561

INPUT RESISTANCE OR CLAMP VOLTAGE. 2 Years 23 C ±5 Voltage mv 3 10 nv > 10 G

Agilent L4411A Low profile 6½ Digit Multimeter

Data acquisition and instrumentation. Data acquisition

Transcription:

Optimizing System Throughput with the NI PXI-4070 6 ½-Digit FlexDMM

Introduction How do I maximize my system throughput? is a common question posed by many engineers and scientists. For years, engineers and scientists have employed numerous strategies to extract more speed from DMM-based systems in both R&D laboratories and on the manufacturing floor. These optimization techniques have often included brute force procedures, such as cutting down the number of tests, purchasing several DMMs, or purchasing very accurate DMMs and running them at much lower resolution. This paper describes how to achieve superior throughput with the National Instruments PXI-4070 6 ½ -digit FlexDMM, NI switch modules, and NI LabVIEW software, without having to make such sacrifices. DMM Optimization Optimizing both measurement and system speed is the easiest way to improve a DMM-based system throughput. Measurement speed is how fast a DMM can take a measurement. System speed includes how fast a DMM can scan with an external switch, and change ranges (1 V to 10 V) and functions (VDC to VAC), in addition to taking measurements. Addressing both of these speeds can dramatically increase automated test system throughput. Measurement Speed Every application has different requirements. For example, some may require the highest level of accuracy, while others require you to maximize throughput. Consequently, DMMs have selectable features or parameters so you can tweak the DMM to your specific requirements. But before you can make the appropriate tradeoffs, you must first understand how these features and parameters affect measurement speed and accuracy. Resolution The highest resolution of a DMM (i. e. 6 ½-digits) will always correspond to its slowest measurement rate. As you reduce the resolution of your DMM, you achieve faster and faster measurement rates. A very common scenario is for a DMM user to purchase a 6 ½ digit DMM, but primarily use it at 5 ½ digits to achieve higher measurement speeds with respectable accuracy. The unique architecture of the NI FlexDMM delivers superior measurement performance by taking full advantage of both LabVIEW and the high-speed PXI bus. The FlexDMM offers a continuously variable DC reading rate from 5 S/s at 7 digits to 5 ks/s at 4 ½-digits, as shown in Figure 1. With the NI-DMM driver software, you can fine-tune the FlexDMM to meet your exact requirements.

Figure 1. FlexDMM DC Measurement Speed Autoranging Autoranging is used for measuring a signal when you do not know which range to select. The DMM takes a series of measurements and adjusts the range until the measurement falls within the smallest range appropriate for that measurement. If you know the approximate value of the input signal, it is advantageous to use a fixed range because the FlexDMM will only take one measurement. With the FlexDMM, you can also select Autorange Once. Autorange Once performs an autorange and stores the range value for use in all subsequent measurements, as long as Autorange Once remains selected. Autozero Autozero is a technique that removes the effect of several kinds of drift, such as temperature-related drift, long-term drift, and low-frequency drift. The process of autozeroing involves disconnecting the external signal from the input of the DMM, internally connecting zero volts to the DMM instead, and taking a measurement. Ideally, this internal value should be zero volts, but real components can create an offset. While measuring this offset value, autozeroing automatically reduces this error. Autozero requires the FlexDMM to take additional measurements, reducing system speed. Similar to autoranging, you can enable Autozero Once. Autozero Once performs an autozero and stores the value for use in all subsequent measurements, as long as Autozero Once remains selected. Offset-Compensated Ohms Offset-Compensated Ohms (OCO) is a technique that eliminates the thermal offsets in low-level resistance measurements. OCO requires the DMM to take two resistance measurements. The first measurement includes the voltage drop across the resistor created by the current source and any thermal EMFs that may be present caused by dissimilar metal connections in the measurement path.

During the second measurement, the current source is turned off so the voltage drop across the resistor becomes zero volts. The DMM subtracts the second measurement from the first, giving you the correct resistance value. Similar to Autozero, this feature requires the DMM to take additional measurements. ADC Calibration ADC Calibration is a feature exclusive to the FlexDMM, with which you can appropriately trade off measurement speed for long-term accuracy. The NI PXI- 4070 ADC is designed for precision, linearity, and stability. By doing routine calibration of the ADC back to a single well-controlled component, you can ensure absolute accuracy of the conversion. With ADC Calibration disabled, the measurement speed increases by a factor of up to two, but you need to add a temperature coefficient error of 3 ppm/ºc to the appropriate range specification. Although this error is small, you should still consider the effects. If you need to turn off ADC Calibration, as you might when you want optimum speed in 6½ digit resolution, you can recover the specified accuracy by running self-calibration periodically. DC Noise Rejection DC Noise Rejection (DCNR) is another feature exclusive to the FlexDMM that suppresses the noise coupled onto a DC measurement. There are three types of noise rejection normal, second-order, and high-order. Normal emulates the behavior of most traditional DMMs. Second-order provides improved protection against high-frequency interference. High-order enables unprecedented immunity from interfering signal, >46 Hz. How DCNR affects your measurement speed depends on the level of noise you have on your measurement. For example, if you have a very noisy power line frequency, you can run faster using the High-Order DCNR because it only requires you to take a single measurement due to the high level of noise rejection. In Normal DCNR, you would have to average several points to get the same level of noise rejection, thus reducing your overall throughput. With LabVIEW, you can quickly adjust each parameter by simply adding a subvi or a property node, a shown in Figure 2.

Figure 2. FlexDMM Optimization Program To simplify the optimization process, the NI-DMM driver software provides an auto mode, which implements features only when appropriate, as shown in Table 2. Function Aperture Autozero ADC Calibration DC Noise Rejection DC 6 1/2 digits 100 ms ON ON High-Order 5 1/2 digits 500 µs ONCE OFF Second-Order 4 1/2 digits 20 µs ONCE OFF Second-Order Autorange 500 µs OFF OFF Second-Order AC 6 1/2 digits Max (100 ms or 4/minFreq) OFF OFF N/A Autorange Max (500 µs or 4/minFreq) OFF OFF N/A Table 1. Auto Mode Setting System Speed Optimizing the order of your measurements to minimize the number of reconfigurations is the simplest way to increase system speed because every function change (DCV to 2-wire Ω) or range change within a function (1 V to 10 V) requires a DMM reconfiguration. The reconfiguration time can be on the order of a few milliseconds, which can substantially slow down many automated test applications. The reconfiguration overhead consists of both hardware and software latencies. Hardware Latencies The types of relays used in traditional DMMs primarily cause hardware latencies. Each function change and sometimes range change requires switching an

electromechanical relay inside a traditional DMM. Electromechanical relays have switching times of several milliseconds, which can hinder your overall throughput. The FlexDMM reduces this hardware latency by using solid-state relays, which have switching times less than one millisecond. Even the random measurement sequence, shown in Table 2, will execute much faster due the solid-state relays on the FlexDMM. The recommended method is to take similar measurements contiguously, which limits the number of function and range changes. A well-constructed measurement sequence is depicted in Table 2. A Random Measurement Sequence: Switch Channel Function Range 1 DC volts 10 V 2 2-wire ohms 1 kω 3 DC current 1 A 4 DC volts 300 V 5 AC current 100 ma 6 2-wire ohms 100 kω 7 2-wire ohms 1 kω 8 DC volts 300 V 9 AC current 100 ma 10 AC current 1 A 11 2-wire ohms 100 kω 12 AC current 100 ma Function changes: 9 Range changes within a function: 2 Total DMM reconfigurations: 11 An Optimized Measurement Sequence: Switch Channel Function Range 1 DC volts 10 V 2 DC volts 300 V 3 DC volts 300 V 4 2-wire ohms 1 kω 5 2-wire ohms 1 kω 6 2-wire ohms 100 kω 7 2-wire ohms 100 kω 8 DC current 1 A 9 AC current 100 ma 10 AC current 100 ma 11 AC current 100 ma 12 AC current 1 A Function changes: 3 Range changes within a function: 3 Total DMM reconfigurations: 6 Table 2. Correct Sequencing of Measurements Software Latencies The reconfiguration cycle requires you to an open a session with the DMM, configure the DMM to the correct state, and close the session with DMM. However, that doesn t mean you have open and close a session with the DMM each time a range or function change is needed. A more efficient approach is to implement all the range and function changes inside a LabVIEW loop structure, as shown in Figure 3. You can make all of your measurements in a loop, and then close the single session after the measurement sequences are complete. This technique will greatly reduce the software latencies.

Figure 3. A LabVIEW Optimization Technique DMM and Switch Scanning Scanning is the key to maximizing your DMM and switch system throughput. The most efficient scanning method is hardware handshaking. With hardware handshaking, the switch and DMM communicate with each other when the device is ready, which ensures that the switch has settled completely. Note that you may need to add additional trigger delays if your unit under test hasn t completely settled (for reasons such as charging a capacitor). Hardware Handshaking requires both the DMM and switches to have hardware triggers. Table 3 lists the common hardware triggers and their functionality. DMM Triggers Input Trigger Measurement Complete Switch Triggers Input Trigger Output Trigger/Scanner Advanced Function The DMM acquires measurement The DMM emits pulse upon completion of measurement Switch executes next connections in the scan list Switch emits pulse when the connections have settled Table 3. The Common Triggers Found on DMMs and Switches

To implement handshaking, you should connect the hardware triggers in the following manner: Input trigger of the DMM with the output trigger/scanner advanced of the switch. Measurement complete of the DMM with the input trigger of the switch Hardware scanning also requires the switch modules to have a scan list. A scan list is a list of channels supplied to the switch that indicates the order in which channels will be scanned. This scan list is downloaded onto the switch module at run time. With National Instruments switches, you can execute the scan list once or continuously. In your program, configure both devices and initiate the DMM. The DMM will wait for a trigger before taking first measurement. Then initiate the switch, which cause the switch to connect its first scan list entry and emit an output pulse. At that point, the DMM and switch start the handshaking process, as shown in Figure 4. This scanning method delivers improved system speed because all communication between the DMM and Switch is hardware timed and runs without host PC intervention. Start NI PXI-2503 Switch Execute First in Scan List entry Output Trigger Input Trigger NI PXI-4070 FlexDMM Execute Next Scan List Entry Measurement Input Trigger Measurement Complete Figure 4. Hardware Handshaking Conclusion Understanding these techniques doesn t necessarily mean you need to use them in every application. Oftentimes, the FlexDMM provides higher throughput rates and accuracy, without disabling any functions. The FlexDMM achieves this performance because it s optimized for both measurement speed and system speed.