CyberKnife Iris Beam QA using Fluence Divergence

Similar documents
A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

(Refer Slide Time: 00:10)

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

A novel and fast method for proton range verification using a step wedge and 2D scintillator

Basic Optics System OS-8515C

Be aware that there is no universal notation for the various quantities.

Check the LCLS Project website to verify 2 of 7 that this is the correct version prior to use.

YOUNGS MODULUS BY UNIFORM & NON UNIFORM BENDING OF A BEAM

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%.

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

Y11-DR Digital Radiography (DR) Image Quality

SCINTILLATING FIBER DOSIMETER ARRAY

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Physics 3340 Spring Fourier Optics

Physics 2020 Lab 9 Wave Interference

An Activity in Computed Tomography

Beam-Restricting Devices

PD233: Design of Biomedical Devices and Systems

QC Testing for Computed Tomography (CT) Scanner

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

IMAGE FORMATION. Light source properties. Sensor characteristics Surface. Surface reflectance properties. Optics

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

IQM Detector Characteristics: Signal reproducibility

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

LOS 1 LASER OPTICS SET

Beam Analysis BeamWatch Non-contact, Focus Spot Size and Position monitor for high power YAG, Diode and Fiber lasers. Disruptive Technology

Opto Engineering S.r.l.

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

An Activity in Computed Tomography

Module-4 Lecture-2 Perpendicularity measurement. (Refer Slide Time: 00:13)

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Devices & Services Company

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

PHYS 1112L - Introductory Physics Laboratory II

Laser and LED retina hazard assessment with an eye simulator. Arie Amitzi and Menachem Margaliot Soreq NRC Yavne 81800, Israel

International Journal of Advance Engineering and Research Development

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

Laser Beam Analysis Using Image Processing

Tutorial Zemax 9: Physical optical modelling I

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

EXPERIENCE WITH AND STUDIES OF THE SNS* TARGET IMAGING SYSTEM

3B SCIENTIFIC PHYSICS

PASS Sample Size Software

Ludlum Medical Physics

Telecentric lenses.

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Nuclear Associates

Physics 345 Pre-lab 1

HCS 50W, 60W & 80W. Data Sheet. Housed Collimated High Power Laser Diode Bar

3B SCIENTIFIC PHYSICS

PHY 431 Homework Set #5 Due Nov. 20 at the start of class

LENSES. a. To study the nature of image formed by spherical lenses. b. To study the defects of spherical lenses.

Initial setup and subsequent temporal position monitoring using implanted RF transponders

PHYSICS QUESTIONNAIRE FORM

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

Exercise 8: Interference and diffraction

AgilEye Manual Version 2.0 February 28, 2007

User s Guide Modulator Alignment Procedure

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

GE AMX 4+ Portable X-Ray

Sensor Calibration Lab

Solution of Exercises Lecture Optical design with Zemax Part 6

PASS Sample Size Software. These options specify the characteristics of the lines, labels, and tick marks along the X and Y axes.

Calibration of KAP meters

User s Guide Modulator Alignment Procedure

Physics 4L Spring 2010 Problem set 1 Due Tuesday 26 January in class

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

P202/219 Laboratory IUPUI Physics Department THIN LENSES

The Henryk Niewodniczański INSTITUTE OF NUCLEAR PHYSICS Polish Academy of Sciences ul. Radzikowskiego 152, Kraków, Poland.

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

10/26/2015. Study Harder

A fast F-number 10.6-micron interferometer arm for transmitted wavefront measurement of optical domes

ECEN 4606, UNDERGRADUATE OPTICS LAB

Corundum C Axis Device for Sample Preparation Timothy Thomas, M.E., M.S.E.E. GIA Laboratory June 4, 2009

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope

Workshop IGLEX Andromède & ThomX 23 June 2016, LAL Orsay. The X-line of ThomX.

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

Test Equipment for Radiology and CT Quality Control Contents

This experiment is under development and thus we appreciate any and all comments as we design an interesting and achievable set of goals.

Computer Generated Holograms for Optical Testing

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

Design Description Document

Howie's Laser Collimator Instructions:

ScanArray Overview. Principle of Operation. Instrument Components

Beam Profiling. Introduction. What is Beam Profiling? by Michael Scaggs. Haas Laser Technologies, Inc.

Name: Date: Math in Special Effects: Try Other Challenges. Student Handout

Dimensional Variations in Tire Tread Extrusions Starrett-Bytewise Measurement Systems May 24, 2013 Abstract

Transcription:

CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery system produces a pencil-thin beam of radiation that is aimed at the cancer tumor from many different angles during the course of a treatment. The Iris collimator used on over half of the CyberKnife systems is a mechanism consisting of 12 tungsten leaves that shapes the diameter of this beam as it diverges toward the patient. Understanding how the radiation beam diverges is useful in verifying that the treatment plans given to a patient are adequately painting the 3D volume of the lesion with radiation and missing nearby vital organs and healthy tissue. This report examines the use of a scintillator-based camera system to measure the CyberKnife X-ray fluence at several distances along the beam path, with analysis of the results. Figure 1 The IBAC converts X-ray beam fluence (blue) to visible light (yellow) using a scintillator module The Iris Beam Aperture Caliper (IBAC) made by Logos Systems (Scotts Valley, CA) consists of a darkened optical enclosure that directly attaches to the CyberKnife Iris. The IBAC allows a module holding X-ray scintillator phosphor to be inserted in the beam path directly below the Iris collimator. As ionizing radiation passes through the scintillator, a beam spot of visible light is formed by Compton scattering. This image is reflected by a mirror toward a digital camera and computer which can be used to measure the intensity profile of the beam diameter to an accuracy of.05 mm. The Iris shapes the radiation into a dodecagon that approximates a circle and is capable of 12 different beam diameters: 5, 7.5, 10, 12.5, 15, 20, 25, 30, 35, 40, 50, and 60 mm. These beam diameters are typically measured 80 centimeters from the linear accelerator source or 80 cm SAD (Source to Axis Distance). The lower set of six collimator leaves where the fluence exits the Iris is located 40 cm from the beam source. Consequently, beam diameters measured at the lower collimator surface are presumed to be half of their 80 cm SAD values. The real-time feedback system that controls the Iris aperture has an accuracy of +/- 0.1 mm at the lower collimator and consequently +/- 0.2 mm at 80 cm SAD. CyberKnife Iris Beam QA using Fluence Divergence Page 1 of 8

Materials and Methods The IBAC camera enclosure was attached to the CyberKnife Iris as shown in Figure 2. The controlling laptop was placed near the CyberKnife control console and a USB extender cable was used to connect the laptop to the IBAC. Figure 2 The IBAC is shown attached to the CyberKnife Iris with the scintillator module access door open. The IBAC is supplied with a calibration module that has a series of precisely aligned 1 mm holes that serve to calibrate camera pixel measurements to physical distances. The calibration module shown on the right in Figure 3 is designed to be inserted through an access panel in the column of the IBAC and is held in place next to the lower Iris collimator by a series of magnets. With the Iris fully opened, the X- ray beam excites the scintillator material behind the mask forming a pattern of bright spots. Figure 3 Scintillator and calibration modules are held near the lower collimator using magnets CyberKnife Iris Beam QA using Fluence Divergence Page 2 of 8

Figure 4 shows how the software finds the position of each spot formed by the calibration module mask pattern. The software uses the position and spacing of the spots to generate horizontal and vertical scaling ratios to convert pixel measurements into millimeters. Figure 4 IBAC calibration reference hole measurements After the IBAC software was calibrated, a script was executed that waited for the presence of a beam spot in the camera s field of view. Once a beam was detected, 30 frames were integrated together to form an image having a superior signal-to-noise ratio than each individual frame. The beam spot in this integrated image was then measured over many different diameters and the average diameter saved to the measurement output file. Each integrated beam image was archived as an individual image file after measurement. Simultaneous to this script running, the CyberKnife was instructed to cycle through all 12 aperture sizes, exposing the IBAC for 50 Monitor Units at each aperture. Next the IBAC was detached from the Iris and placed on a cart directly under the Iris as shown in Figure 5. The laser was used to adjust the position of the cart so that the radiation beam was pointed directly at the center of the scintillator module. The distance from the scintillator to the lower collimator surface was measured at 17.7 cm. Since the lower collimator is 40.0 cm from the source, the scintillator at this position is 57.7 cm from the source. A series of 12 beam sizes was captured at this position. CyberKnife Iris Beam QA using Fluence Divergence Page 3 of 8

Figure 5 IBAC detached from the Iris with the scintillator at 57.7 cm SAD Two other sets of beam diameters were captured at 52.7 cm and 47.7 cm using 5 cm thick spacers placed beneath the IBAC. The 47.7 cm position is shown in Figure 6. During the three sets of measurements with the IBAC disconnected from the Iris, the beam passed through an extra 3 mm of rubber which was used to keep the enclosure darkened. Figure 6 IBAC supported by two 5 cm spacers with the scintillator at 47.7 cm SAD CyberKnife Iris Beam QA using Fluence Divergence Page 4 of 8

Data Representation IBAC beam diameters are measured using a full-width half-maximum (FWHM) algorithm. First the maximum and minimum grayscale intensities of the beam spot are used to determine the grayscale value that represents the half-maximum value. Next, the circular group of pixels that most closely match this grayscale midpoint are identified by the software and the center point is determined. Radii are then calculated every 5 degrees around this circle and these seventy-two values are used to calculate the average beam diameter. This value is recorded in the output text file or displayed in the beam viewer as shown in Figure 7. Figure 7 IBAC beam profile for a 20 mm Iris aperture Shown below is the script output text for the first 4 beam spot sizes with the IBAC attached to the Iris (40 cm SAD). WinLVS Script Version:, 9.02,Time-Date:,11:04:43 02/03/2012,Center XY:, 325.177, 241.795 BWS,1, Time,11:05:04 02/03/2012 1,Beam Spot Metrics,Units:,mm - percent, Diameter:, 2.60, Quality:, 93.95, Min. Gray:, 0.00, Max. Gray:, 81.00, Output Factor:, 0.00 Time,11:05:07 02/03/2012,30 Time,11:06:17 02/03/2012 2,Beam Spot Metrics,Units:,mm - percent, Diameter:, 4.19, Quality:, 96.21, Min. Gray:, 0.00, Max. Gray:, 91.00, Output Factor:, 0.00 Time,11:06:21 02/03/2012,30 Time,11:07:04 02/03/2012 3,Beam Spot Metrics,Units:,mm - percent, Diameter:, 5.41, Quality:, 97.89, Min. Gray:, 0.00, Max. Gray:, 95.00, Output Factor:, 0.00 Time,11:07:08 02/03/2012,30 Time,11:07:55 02/03/2012 4,Beam Spot Metrics,Units:,mm - percent, Diameter:, 6.62, Quality:, 97.92, Min. Gray:, 0.00, Max. Gray:, 97.00, Output Factor:, 0.00 Time,11:07:58 02/03/2012,31 CyberKnife Iris Beam QA using Fluence Divergence Page 5 of 8

Test Results and Analysis The complete IBAC data using the four different SAD positions and twelve aperture sizes are shown in Figure 8. It can be seen that beam width measurements at 40.0 cm SAD are close to half the 80 cm SAD aperture sizes. Also note that beam diameters steadily increase along with distance from the source. The IBAC maximum field of view is 34.5 mm, so divergence measurements were limited with the 50 and 60 mm apertures. For QA purposes, the measured values at 40 cm for these two apertures when doubled are within tolerance of the expected 80 cm SAD values. 5 mm 7.5 mm 10 mm 12.5 mm 15 mm 20 mm 25 mm 30 mm 35 mm 40 mm 50 mm 60 mm 40.0 cm SAD 2.6 4.19 5.41 6.62 7.85 10.32 12.75 15.23 17.62 20.04 24.97 29.92 47.7 cm SAD 3.13 5.06 6.5 7.93 9.41 12.3 15.19 18.06 20.86 23.73 29.56 Max. 52.7 cm SAD 3.47 5.56 7.17 8.79 10.37 13.57 16.69 19.88 22.98 26.14 32.59 Max. 57.7 cm SAD 3.87 6.11 7.88 9.6 11.39 14.86 18.29 21.79 25.17 28.67 Max. Max. Figure 8 IBAC beam width measurements at 4 different distances to source Shown in Figure 9 is a chart of the data in Figure 8 demonstrating that the IBAC has measured a good linear relationship between each beam diameter and its distance from the source of the X-ray fluence. IBAC CyberKnife Beam Divergence Study 35 Beam Diameter (mm) 30 25 20 15 10 5 5 mm 7.5 mm 10 mm 12.5 mm 15 mm 20 mm 25 mm 30 mm 35 mm 40 mm 0 0 10 20 30 40 50 60 70 Distance SAD (cm) Figure 9 Chart of IBAC measured beam spot diameters at the 4 SAD positions Figure 10 applies linear regression to the data and displays the formula in slope intercept format for each set of aperture measurements. CyberKnife Iris Beam QA using Fluence Divergence Page 6 of 8

IBAC CyberKnife Beam Divergence Trend Lines 35 Beam Diameter (mm) y = 0.4865x + 0.5488 30 y = 0.426x + 0.5616 25 y = 0.3698x + 0.4247 20 y = 0.312x + 0.277 y = 0.2563x + 0.069 15 y = 0.1994x - 0.121 10 y = 0.1688x - 0.1252 y = 0.1392x - 0.1517 y = 0.108x - 0.1168 5 y = 0.0712x - 0.2599 0 0 10 20 30 40 50 60 70 Distance SAD (cm) 5 mm 7.5 mm 10 mm 12.5 mm 15 mm 20 mm 25 mm 30 mm 35 mm 40 mm Figure 10 Linear regression analysis applied to all beam diameter trends Figure 11 contains the linear regression coefficients calculated in the previous chart with slope being the m value and intercept the b value in the equation y = mx + b. In this equation y is the beam diameter in millimeters at the distance x centimeters from the source. These coefficients are then used to calculate the expected beam fluence diameter at 80 cm SAD. In the dose domain, the beam diameter at 80 cm SAD is expected to be close to the selected aperture size, and so the far right columns calculate the difference between the projected fluence beam widths and the expected 5 mm to 40 mm dose beam width at 80 cm. Aperture (mm) Linear Regression Beam Divergence Expected Width Dose Width Delta Slope (Degrees) Intercept at 80 cm SAD (mm) Delta (mm) Percent 5 0.0712.4079-0.2599 5.4361 0.4361 8.722 7.5 0.108.6188-0.1168 8.5232 1.0232 13.64267 10 0.1392.7975-0.1517 10.9843 0.9843 9.843 12.5 0.1688.9671-0.1252 13.3788 0.8788 7.0304 15 0.1994 1.142-0.121 15.831 0.831 5.54 20 0.2563 1.468 0.069 20.573 0.573 2.865 25 0.312 1.787 0.277 25.237 0.237 0.948 30 0.3698 2.118 0.4247 30.0087 0.0087 0.029 35 0.426 2.439 0.5616 34.6416-0.3584-1.024 40 0.4865 2.785 0.5488 39.4688-0.5312-1.328 Figure 11 Linear regression coefficients used to predict beam fluence diameter at 80 cm SAD It should be noted that the CyberKnife linear accelerator radiation source is formed by a beam of electrons approximately 3 mm in diameter striking a tungsten plate. A rigorous analysis of the intercept values for the beam apertures should take into account the fact that the beam starts not as a point, but as CyberKnife Iris Beam QA using Fluence Divergence Page 7 of 8

a disk 3 mm in diameter. It should also take into account the mechanical design of the upper and lower Iris collimator banks and the distance that separates each bank from the source. For the purpose of this paper, the Iris is treated as a black box and only the characteristics of the emerging X-ray fluence are discussed. The most concerning data in Figure 11 are the projected 80 cm SAD beam fluence diameters for the 7.5 and 10 mm aperture sizes. These values are 1.02 and 0.98 mm larger than their corresponding nominal dose values. Their percentage differences of 13.6% and 9.8% percent from nominal are the largest of all the aperture widths. These deviations should be investigated more thoroughly as part of regular Iris QA procedures. Another alternative would be to take new water tank dose profile measurements and compare those profiles to commissioning values looking for any out-of-tolerance beam diameter increases. Conclusion The beam fluence divergence angles in air for ten of the Iris apertures have been measured. The extrapolated beam diameters at 80 cm SAD serve as a QA benchmark for comparing these fluence beam measurements with those made using traditional ion chamber and diode dose-based methodologies. At least two of these extrapolated beam diameters indicate the need for additional dose beam width verification with commissioning values. Contact Information Ronald Berg, Ph.D. Senior Medical Physicist Jesse McKay, M.S. Medical Physicist Radiation Oncology Dept. Erlanger Medical Center 975 East 3 rd St Chattanooga, TN 37403 Ph: 423-778-7485 Brett Nelson, M.S. Director of Engineering Logos Systems Int l 175 El Pueblo Road Scotts Valley, CA 95066 www.logosvisionsystem.com Phone: 831-600-6101 BKN 9/30/2012r2 CyberKnife Iris Beam QA using Fluence Divergence Page 8 of 8