Specifications for the NI PXI/PCI-6552/6551

Similar documents
NI PXI/PCI-6561/6562 Specifications

100 MHz, 32-Channel, 5 V PXI Digital Waveform Instrument

NI 6143 Specifications

NI PXI/PCI-5411/5431 Specifications

This document provides the specifications for the PXIe-6544.

NI 5421 Specifications NI PXI/PCI Bit 100 MS/s Arbitrary Waveform Generator

This document provides the specifications for the PXIe-6547.

This document provides the specifications for the PXIe-6548.

NI PXI/PCI-5411/5431 Specifications

200 MHz, 16-Channel PXI Digital Waveform Instrument

NI PXI-4461 Specifications

NI PXI/PCI-5122 Specifications

PXI Timing and Synchronization Control

These specifications apply to the PXIe-5113 with 64 MB and 512 MB of memory.

NI PXI/PCI-5421 Specifications 16-Bit 100 MS/s Arbitrary Waveform Generator

NI PXI/PCI-5412 Specifications 14-Bit 100 MS/s Arbitrary Waveform Generator

NI 6023E/6024E/6025E Family Specifications

NI 6731/6733 Specifications

NI PXI/PCI-5402/5406 Specifications 14/16-Bit, 20/40 MHz Arbitrary Function Generator

NI PXI-5422 Specifications 16-Bit 200 MS/s Arbitrary Waveform Generator

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

NI 6013/6014 Family Specifications

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

NI 6238/6239 Specifications

NI PXI-2532 Specifications

NI PXI-2557 Specifications

NI DAQPad -6020E Family Specifications

Analog Characteristics

NI PXI-2555 Specifications

NI PXIe-5601 Specifications

NI 6040E Family Specifications

NI PXIe-5630 Specifications

PCI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PCI Oscilloscope Device

NI PXI-2521 Specifications 40-Channel DPST Relay Module

100 MHz Bandwidth, 12-Bit PXI Waveform Generator

PXI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PXI Oscilloscope

SCC-FV01 Frequency Input Module

NI PXI/PXIe-2540 Specifications

NI PXI/PXIe-2541 Specifications 300 MHz Ω Matrix

8-Channel, 12-Bit, 60 MHz PXI Express Oscilloscope

NI 6230 Specifications

Note Using the PXIe-5785 in a manner not described in this document might impair the protection the PXIe-5785 provides.

PXI Contents SPECIFICATIONS. 1 GHz Bandwidth, 2 GS/s, 8-Bit PXI Oscilloscope

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

NI PXIe-5663 Specifications

PXI Contents SPECIFICATIONS. 300 MHz Bandwidth, 2 GS/s, 8-Bit PXI Oscilloscope

Specifications. PCI Bus. Analog Input Input Characteristics

Contents. ZT530PCI & PXI Specifications. Arbitrary Waveform Generator. 16-bit, 400 MS/s, 2 Ch

PXIe, 24-Channel, ±24 V, 50 ma Precision PXI Source Measure Unit

NI PXI-2530 Specifications

NI PXI-2797 Specifications 40 GHz 6 1 Terminated 50 Ω Multiplexer

NI PXI-2520 Specifications 80-Channel SPST Relay Module

DEVICE SPECIFICATIONS NI PXIe Ch, 24-bit, 25.6 ks/s Simultaneous AC LVDT Input Module

NI 622x Specifications

Analog Input. DEVICE SPECIFICATIONS NI PXIe-4300

250 MS/s, 125 MHz, 8-Bit Digitizers NI PCI-5114, PXI-5114

NI PXI/PXIe-2532 Specifications

NI PXI/PXIe-2529 Specifications

PXI Contents SPECIFICATIONS. 1.3 GHz RF Analog Signal Generator

NI PXI-5441 Specifications

PXIe Contents SPECIFICATIONS. 250 khz to 20 GHz RF Analog Signal Generator

PXI Contents SPECIFICATIONS. 6.6 GHz RF Analog Signal Generator

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

PXIe Contents SPECIFICATIONS. 500 MHz, 2.5 GS/s, 10-Bit PXI Express Oscilloscope

80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

SCC-DI01 Isolated Digital Input Module

SCXI MHz Quad 8x1 50 Ω Multiplexer

This document lists the specifications for the NI PXI The PXI-5900 is a differential amplifier.

PXIe Contents SPECIFICATIONS. 3.3 GHz RF Analog Signal Generator

Contents. USER GUIDE Differential Digital Flying Lead Cable Accessory for Differential NI Digital Waveform Generator/Analyzers

PXI Contents SPECIFICATIONS. 100 MHz, 100 MS/s, 14-Bit PXI Oscilloscope

PXIe Contents SPECIFICATIONS. 1.5 GHz, 5 GS/s, 10-Bit PXI Express Oscilloscope

NI 6124 Specifications

SPECIFICATIONS FOR THE NI PXI/PCI-4060

PXIe, 40 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5423.

NI USB Contents DEVICE SPECIFICATIONS. 50 MS/s Bus-Powered USB Oscilloscope

4-Channel, ±6 V, 500 ma Precision PXI Source Measure Unit

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

PXIe Contents SPECIFICATIONS. 125 MHz, 2-Channel, 8-Bit, PXI Express Oscilloscope

PCI Contents SPECIFICATIONS. 100 MHz, 100 MS/s, 14-Bit PCI Oscilloscope Device

PXI Contents SPECIFICATIONS. 125 MHz, 2-Channel, 8-Bit PXI Oscilloscope

SCC-ACC01 Accelerometer Input Module

4-Channel, ±24 V, 150 ma Precision PXI Source Measure Unit

PXI Contents. Definitions SPECIFICATIONS. ±100 V, 2 W Precision PXI Source Measure Unit. These specifications apply to the PXI-4132.

Refer to the NI Switches Help at ni.com/manuals for detailed topology information.

PXI Contents SPECIFICATIONS. 24-Bit, Flexible Resolution PXI Oscilloscope

PXI Contents SPECIFICATIONS. 24-Bit, Flexible Resolution PXI Oscilloscope

NI SCXI Specifications

PXIe Contents SPECIFICATIONS. RF Amplifier

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

PXIe, 80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

4-Channel, ±24 V, 150 ma PXI Source Measure Unit

NI PXIe-5442 Specifications 16-Bit 100 MS/s Arbitrary Waveform Generator with Onboard Signal Processing (OSP)

DEVICE SPECIFICATIONS NI PXIe-4302/ Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module

DEVICE SPECIFICATIONS NI PXIe-4304/ Ch, 24-bit, ±42 V, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

NI PXI/PXIe-2527 Specifications

6 1 2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter

NI PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. 3 GHz Ω Multiplexer (SP16T)

PXI Contents SPECIFICATIONS. 150 MHz Bandwidth, 200 MS/s, 12-Bit PXI Oscilloscope

Transcription:

Specifications for the NI PXI/PCI-6552/6551 100/50 MHz Digital Waveform Generator/Analyzer Channel Characteristics These specifications are valid for the operating temperature range, unless otherwise noted. Number of data channels Direction control of data channels Number of Programmable Function Interface (PFI) channels Direction control of PFI channels Number of clock terminals 20 Per channel 4 Refer to the Waveform Characteristics section for more details. Per channel 3 input 2 output Refer to the Timing Characteristics section for more details.

Generation Signal Characteristics (Data, DDC CLK OUT, and PFI <0:3> Channels) Generation voltage range Generation signal type Number of programmable voltage levels Generation voltage range restrictions Generation voltage swing Generation voltage level resolution DC generation voltage level accuracy Output impedance Output impedance temperature coefficient Maximum DC drive strength Data channel driver enable/disable control 2.0 V to 5.5 V Into 1 MΩ Single-ended 1 voltage low level 1 voltage high level Note: While you can only set one voltage low level and one voltage high level for all generation channels, you can set a different voltage low level and voltage high level for all acqusition channels. 0.5 V to 5.5 V (up to 50 MHz clock rate) 2.0 V to 3.7 V (up to 50 MHz clock rate) 0.5 V to 3.7 V (50 to 100 MHz clock rate; NI 6552 only) 400 mv to 6 V (up to 50 MHz clock rate) 400 mv to 4.2 V (50 to 100 MHz clock rate; NI 6552 only) For all data, CLK OUT (Sample clock only), and PFI channels Into 1 MΩ Into 1 MΩ 10 mv Into 1 MΩ ±20 mv Into 1 MΩ; does not include system crosstalk 50 Ω nominal At 25 C 0.2 Ω/ C Typical ±50 ma maximum per channel ±600 ma maximum for all data, clock, and PFI channels Per channel Softwareselectable Specifications for the NI PXI/PCI-6552/6551 2 ni.com

Channel power-up state Drivers disabled, 10 kω input impedance Output protection The device can indefinitely sustain a short to any voltage in the generation voltage range. Acquisition Signal Characteristics (Data, STROBE, and PFI <0..3> Channels) Number of voltage comparators per channel Acquisition voltage range Number of programmable acquisition thresholds Minimum detectable voltage swing Acquisition voltage threshold resolution DC acquisition voltage threshold accuracy 2 2.0 V to 5.5 V 1 voltage low threshold 1 voltage high threshold Note: While you can only set one voltage low level and one voltage high level for all acquisition channels, you can set a different voltage low level and voltage high level for all generation channels. For all data, STROBE, and PFI channels 50 mv 10 kω input impedance, measured with 50% duty cycle input signal 10 mv 10 kω input impedance ±30 mv 10 kω input impedance, does not include system crosstalk National Instruments Corporation 3 Specifications for the NI PXI/PCI-6552/6551

Input impedance Input protection 50 Ω nominal or 10 kω (default) Softwareselectable per channel, when powered on and within valid voltage range 2.3 V to 6.8 V Diode clamps in the design may provide additional protection outside this range. Timing Characteristics Sample Clock Sample clock sources On Board Clock frequency range CLK IN frequency range 1. On Board Clock (internal voltage-controlled crystal oscillator (VCXO) with divider) 2. CLK IN (SMB jack connector) 3. PXI_STAR (PXI backplanepxi only) 4. STROBE (DDC connector; acquisition only) NI 6552: 48 Hz to 100 MHz Configurable to 200 MHz/N, where 2 N 4,194,304 NI 6551: 48 Hz to 50 MHz Configurable to 200 MHz/N, where 4 N 4,194,304 NI 6552: 20 khz to 100 MHz NI 6551: 20 khz to 50 MHz Refer to the CLK IN (SMB Jack Connector) section for restrictions based on waveform type. Specifications for the NI PXI/PCI-6552/6551 4 ni.com

PXI_STAR frequency range (PXI only) STROBE frequency range Sample clock relative delay adjustment Sample clock relative delay adjustment resolution Exported Sample clock destinations Exported Sample clock delay range (δ C ) Exported Sample clock delay resolution (δ C ) Exported Sample clock jitter NI 6552: 48 Hz to 100 MHz NI 6551: 48 Hz to 50 MHz NI 6552: 48 Hz to 100 MHz NI 6551: 48 Hz to 50 MHz Refer to the PXI_STAR (PXI BackplanePXI only) section. Refer to the STROBE (Digital Data & Control (DDC) Connector) section. 0 to 1 Sample clock period You can apply a delay or phase adjustment to the On Board Clock 10 ps to align multiple devices. 1. DDC CLK OUT (DDC connector) 2. CLK OUT (SMB jack connector) Sample clocks with sources other than STROBE can be exported. 0 to 1 Sample clock periods For clock frequencies 25 MHz 1/256 of Sample clock period For clock frequencies 25 MHz Period Jitter Cycle-to-Cycle Jitter Typical; using 20 ps rms 35 ps rms On Board Clock National Instruments Corporation 5 Specifications for the NI PXI/PCI-6552/6551

Generation Signal Characteristics (Data, DDC CLK OUT, and PFI <0..3> Channels) Data channel-tochannel skew ±300 ps Typical skew across all data channels ±900 ps Maximum skew across all data channels Maximum data channel toggle rate NI 6552: 50 MHz NI 6551: 25 MHz Data formats NRZ Data position modes Rising edge, Falling edge, or Delayed Relative to Sample clock, per channel Generation data delay range (δ G ) Generation data delay resolution (δ G ) 0 to 1 Sample clock period For clock frequencies 25 MHz 1/256 of Sample clock period For clock frequencies 25 MHz Specifications for the NI PXI/PCI-6552/6551 6 ni.com

Figure 1. Eye Diagram 1 Rise time (0 V to 3.3 V swing) Fall time (0 V to 3.3 V swing) Exported Sample clock offset (t CO ) Time delay from Sample clock (internal) to DDC connector (t SCDDC ) Into 50 Ω 2.25 ns 20% to 80%, Into 1 MΩ 2.75 ns into 475 pf test typical system capacitance Into 50 Ω 2.25 ns 20% to 80%, Into1MΩ 2.75 ns into 475 pf test typical system capacitance 0ns or 2.5ns (default) Softwareselectable 32.5 ns Typical 1 This eye diagram was captured on DIO 0 (100 MHz clock rate) at 3.3 V at room temperature into 50 Ω termination. National Instruments Corporation 7 Specifications for the NI PXI/PCI-6552/6551

Sample Clock Exported Sample Clock Noninverted Exported Sample Clock Inverted Exported Sample Clock Delayed δ C t SCDDC t P DATA CHANNELS Data Position Rising Edge Data Position Falling Edge t CO Sample n Sample n+1 Sample n+2 t CO Sample n Sample n+1 Sample n+2 Data Position Delayed Sample n Sample n+1 Sample n+2 δ G t SCDDC : Time delay from Sample Clock (internal) to DDC Connector 0 δ C 1 : Exported Sample Clock Delay (fraction of t P ) 0 δ G 1 : Pattern Generation Data Delay (fraction of t P ) t P = 1 ƒ = Period of Sample Clock t CO = Exported Sample Clock Offset; 0 or 2.5 ns, software-selectable Figure 2. Generation Timing Diagram Specifications for the NI PXI/PCI-6552/6551 8 ni.com

Acquisition Signal Characteristics (Data, STROBE, and PFI <0..3> Channels) Channel-tochannel skew Minimum detectable pulse width Set-up time to STROBE (t SUS ) Hold time to STROBE (t HS ) Time delay from DDC connector to internal Sample clock (t DDCSC ) Set-up time to Sample clock (t SUSC ) ±400 ps Typical skew across all data channels ±900 ps Maximum skew across all data channels 4 ns Required at both acquisition voltage thresholds 2.3 ns Maximum; includes maximum data channel-tochannel skew 1.9 ns Maximum; includes maximum data channel-tochannel skew 27.5 ns Typical 0.4 ns Does not include data channel-tochannel skew, t DDCSC, or t SCDDC National Instruments Corporation 9 Specifications for the NI PXI/PCI-6552/6551

Hold time to Sample clock (t HSC ) Data position modes Acquisition data delay range (δ A ) Acquisition data delay resolution (δ A ) 0ns Rising edge, Falling edge, or Delayed Does not include data channel-tochannel skew, t DDCSC, or t SCDDC Relative to Sample clock, per channel 0 to 1 Sample clock periods For clock frequencies 25 MHz 1/256 of Sample clock period For clock frequencies 25 MHz Specifications for the NI PXI/PCI-6552/6551 10 ni.com

Sample Clock t P Virtual Sample Clock Projected to DDC Connector t DDCSC DATA CHANNELS Data Position Rising Edge t SUSC t HSC Data Position Falling Edge t SUSC t HSC Virtual Sample Clock Projected to DDC Connector δ A DATA CHANNELS Data Position Delayed t SUSC t HSC t DDCSC : Time Delay from DDC Connector to Internal Sample Clock 0 δ A 1 : Pattern Acquisition Data Delay (fraction of t P ) t P = 1 = Period of Sample Clock ƒ t SUSC = Set-Up Time to Sample Clock t HSC = Hold Time to Sample Clock Figure 3. Acquisition Timing Diagram National Instruments Corporation 11 Specifications for the NI PXI/PCI-6552/6551

STROBE t P DATA CHANNELS t SUS Data Position Rising Edge t HS t SUS Data Position Falling Edge t HS t SUS Data Position Delayed δ A t HS t SUS = Set-up Time to STROBE t HS = Hold Time from STROBE 0 δ A 1 : Pattern Acquisition Data Delay (percentage of t P ) t P = 1 ƒ = Period of Sample Clock Figure 4. Acquisition Timing Diagram Using STROBE as the Sample Clock CLK IN (SMB Jack Connector) Direction Input into device Destinations 1. Reference clock (for the phase lock loop (PLL)) 2. Sample clock Input coupling AC Input protection ±10 VDC Input impedance 50 Ω (default) or 1 kω Softwareselectable Specifications for the NI PXI/PCI-6552/6551 12 ni.com

Minimum detectable pulse width Clock requirements 4 ns Required at V rms mean Clock must be continuous and free-running As Sample clock External Sample clock requirements As Reference Clock Reference clock frequency range Reference clock voltage range Reference clock duty cycle Square Waves Voltage range 0.65 V pp to 5.0 V pp Frequency range Duty cycle range Voltage range Frequency range NI 6552: 20 khz to 100 MHz NI 6551: 20 khz to 50 MHz f < 50 MHz: 25% to 75% f 50 MHz: 40% to 60% Sine Waves 0.65 V pp to 5.0 V pp NI 6552: 5.5 MHz to 100 MHz NI 6551: 5.5 MHz to 50 MHz 1.0 V pp to 5.0 V pp NI 6552: 3.5 MHz to 100 MHz NI 6551: 3.5 MHz to 50 MHz 2.0 V pp to 5.0 V pp NI 6552: 1.8 MHz to 100 MHz NI 6551: 1.8 MHz to 50 MHz 10 MHz ±50 ppm 0.65 V pp to 5.0 V pp 25% to 75% National Instruments Corporation 13 Specifications for the NI PXI/PCI-6552/6551

STROBE (Digital Data & Control (DDC) Connector) Direction Input into device Destinations Sample clock (acquisition only) STROBE frequency range STROBE duty cycle range Minimum detectable pulse width Voltage thresholds NI 6552: 48 Hz to 100 MHz NI 6551: 48 Hz to 50 MHz NI 6552: f 50 MHz: 25% to 75% f > 50 MHz: 40% to 60% NI 6551: 40% to 60% 4ns Refer to the Acquisition Signal Characteristics (Data, STROBE, and PFI <0..3> Channels) specifications in the Channel Characteristics section. At the programmed thresholds Required at both acquisition voltage thresholds Clock requirements Input impedance Clock must be continuous and free-running 50 Ω or 10 kω (default) Softwareselectable PXI_STAR (PXI BackplanePXI only) Direction Input into device Destinations 1. Sample clock 2. Start trigger 3. Pause trigger (generation sessions only) 4. Script trigger (generation sessions only) 5. Reference trigger (acquisition sessions only) Specifications for the NI PXI/PCI-6552/6551 14 ni.com

PXI_STAR frequency range NI 6552: 48 Hz to 100 MHz NI 6551: 48 Hz to 50 MHz Clock requirements Clock must be continuous and free-running CLK OUT (SMB Jack Connector) Direction Output Sources 1. Sample clock (excluding STROBE) 2. Reference clock (PLL) Output impedance 50 Ω nominal As Sample Clock Electrical characteristics Refer to the Generation Signal Characteristics (Data, DDC CLK OUT, and PFI <0..3> Channels) specifications in the Channel Characteristics section. As Reference Clock Maximum drive current 24 ma Logic type 3.3 V CMOS DDC CLK OUT (Digital Data & Control (DDC) Connector) Direction Output Sources Sample clock STROBE cannot be routed to DDC CLK OUT Electrical characteristics Refer to the Generation Signal Characteristics (Data, DDC CLK OUT, and PFI <0..3> Channels) specifications in the Channel Characteristics section. National Instruments Corporation 15 Specifications for the NI PXI/PCI-6552/6551

Reference Clock (PLL) Reference clock sources 1. PXI_CLK10 (PXI backplanepxi only) 2. RTSI 7 (RTSI buspci only) 3. CLK IN (SMB jack connector) 4. None (internal oscillator not locked to a reference) Provides the reference frequency for the phase lock loop Lock time 400 ms Typical Reference clock frequencies Reference clock duty cycle Reference clock destinations 10 MHz ±50 ppm 25% to 75% CLK OUT (SMB jack connector) Waveform Characteristics Memory and Scripting Memory architecture The NI 655X uses the Synchronization and Memory Core (SMC) technology in which waveforms and instructions share onboard memory. Parameters such as number of script instructions, maximum number of waveforms in memory, and number of samples (S) available for waveform storage are flexible and user-defined. Refer to the NI Digital Waveform Generator/ Analyzer Help for more information. Onboard memory size 1 Mbit/channel (for generation sessions) 1 Mbit/channel (for acquisition sessions) 8 Mbit/channel (for generation sessions) 8 Mbit/channel (for acquisition sessions) 64 Mbit/channel (for generation sessions) 64 Mbit/channel (for acquisition sessions) Maximum limit for generation sessions assumes no scripting instructions. Specifications for the NI PXI/PCI-6552/6551 16 ni.com

Generation modes Generation minimum waveform size Generation finite repeat count Generation waveform quantum Single-waveform mode: Generate a single waveform once, N times, or continuously. Scripted mode: Generate a simple or complex sequence of waveforms. Use scripts to describe the waveforms to be generated, the order in which the waveforms are generated, how many times the waveforms are generated, and how the device responds to Script triggers. Configuration 100 MHz (NI 6552 only) Sample Rate Finite waveform 2 2 Continuous waveform Stepped triggered script Burst triggered script 32 16 128 64 512 256 50 MHz Sample rate dependent. Increasing sample rate increases minimum waveform size requirement. For information on these configurations, refer to the Common Scripting Use Cases topic in the NI Digital Waveform Generator/ Analyzer Help. 1 to 16,777,216 Waveform size must be an integer multiple of two samples. Regardless of waveform size, NI-HSDIO allocates waveforms into block sizes of 32 S of physical memory. National Instruments Corporation 17 Specifications for the NI PXI/PCI-6552/6551

Acquisition minimum record size Acquisition record quantum Acquisition number of pre-reference trigger samples Acquisition number of post- Reference trigger samples 1 sample 1 record 0 up to full record 0 up to full record Triggers (Inputs to the NI 655X) Specification Values Comments Trigger types Sources 1. Start trigger 2. Pause trigger 3. Script trigger (generation sessions only) 4. Reference trigger (acquisition sessions only) 1. PFI 0 (SMB jack connector) 2. PFI <1..3> (DDC connector) 3. PXI_TRIG<0..7> (PXI backplanepxi only)/ RTSI <0..7> (RTSI buspci only) 4. PXI_STAR (PXI backplanepxi only) 5. Pattern match (acquisition sessions only) 6. Software (user function call) 7. Disabled (do not wait for a trigger) Specifications for the NI PXI/PCI-6552/6551 18 ni.com

Specification Values Comments Trigger detection 1. Start trigger (edge detection: rising or falling) 2. Pause trigger (level detection: high or low) 3. Script trigger (edge detection: rising or falling; level detection: high or low) 4. Reference trigger (edge detection: rising or falling) Minimum required trigger pulse width 30 ns Acquisition triggers must meet set-up and hold time requirements. Destinations 1. PFI 0 (SMB jack connectors) 2. PFI <1..3> (DDC connector) 3. PXI_TRIG<0..7> (PXI backplanepxi only)/ RTSI <0..7> (RTSI buspci only) Each trigger can be routed to any destination except the Pause trigger. The Pause trigger cannot be exported. Delay from Pause trigger to Paused state Delay from trigger to digital data output Generation Sessions Acquisition Sessions 32 Sample clock periods + 150 ns Synchronous to the data Use the Data Active event during generation to determine when the NI 655X enters the Paused state. 32 Sample clock periods + 160 ns National Instruments Corporation 19 Specifications for the NI PXI/PCI-6552/6551

Events (Output from the NI 655X) Event type Destinations 1. Marker (generation sessions only) 2. Data Active event (generation sessions only) 3. Ready for Start event 1. PFI 0 (SMB jack connectors) 2. PFI <1..3> (DDC connector) 3. PXI_TRIG<0..7> (PXI backplanepxi only)/ RTSI <0..7> (RTSI buspci only) Each event, except the Data Active event, can be routed to any destination. The Data Active event can only be routed to the PFI channels. Marker time resolution (placement) Markers must be placed at an integer multiple of two samples. Calibration Interval for external calibration 2 years Warm-up time 15 minutes Onboard calibration voltage reference Temperature coefficient Long-term stability ±5 ppm/ C 90 ppm/ khr Typical On Board Clock characteristics (only valid when PLL reference source is set to None) Frequency accuracy ±100 ppm Typical Specifications for the NI PXI/PCI-6552/6551 20 ni.com

Temperature stability ±30 ppm Typical Aging ±5 ppm first year Typical Power Value Specification Typical Maximum Comments +3.3 VDC 2.0 A 2.0 A +5 VDC 1.8 A PXI PCI 2.3 A 2.4 A +12 VDC 0.3 A 0.5 A 12 VDC 0.2 A 0.2 A Total power 21.6 W PXI PCI 26.5 W 27 W Software Specifications Driver software Application software Test panel NI-HSDIO driver software. NI-HSDIO allows you to configure, control, and calibrate the NI 655X. NI-HSDIO provides application interfaces for many development environments. NI-HSDIO follows IVI API guidelines. NI-HSDIO provides programming interfaces for the following application development environments: National Instruments LabVIEW 7.0 or later National Instruments LabWindows /CVI 6.0 or later Microsoft Visual C/C++ 6.0 or later National Instruments Measurement & Automation Explorer (MAX) provides test panels with basic acquisition and generation functionality for the NI 655X. MAX is included on the NI-HSDIO driver CD. National Instruments Corporation 21 Specifications for the NI PXI/PCI-6552/6551

Physical Specifications Dimensions Front Panel Connectors CLK IN PXI: 18.6 cm 13.1 cm (7.32 in. 5.16 in.) Single 3U CompactPCI slot; PXI compatible PCI: 12.6 cm 35.5 cm (4.95 in. 13.9 in.) Label Function(s) Connector Type External Sample clock, external PLL reference input Environment and Compliance SMB jack connector PFI 0 Events, triggers SMB jack connector CLK OUT DIGITAL DATA & CONTROL Exported Sample clock, exported Reference clock Digital data channels, exported Sample clock, STROBE, events, triggers SMB jack connector 68-pin VHDCI connector Operating/ storage environment Operating temperature Storage temperature Operating relative humidity Storage relative humidity Indoor use only PXI: 0 ºC to +55 ºC in all NI PXI chassis except the following: 0 ºC to +45 ºC when installed in an NI PXI-1000/B and NI PXI-101X chassis (Meets IEC-60068-2-1 and IEC-60068-2-2) PCI: 0 ºC to +45 ºC 20 C to 70 C 10% to 90% relative humidly, noncondensing (Meets IEC-60068-2-56) 5% to 95% relative humidity, noncondensing (Meets IEC-60068-2-56) Specifications for the NI PXI/PCI-6552/6551 22 ni.com

Operating shock Storage shock 30 g, half-sine, 11 ms pulse (Meets IEC-60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) 50 g, half-size, 11 ms pulse (Meets IEC-60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) NI PXI-655X only NI PXI-655X only Operating vibration 5 Hz to 500 Hz, 0.31 g rms (Meets IEC-60068-2-64) NI PXI-655X only Storage vibration 5 Hz to 500 Hz, 2.46 g rms (Meets 60068-2-64. Test profile exceeds requirements of MIL-PRF-28800F, Class B) NI PXI-655X only Altitude 0 m to 2000 m above sea level (at 25 ºC ambient temperature) Pollution Degree 2 Safety Emissions The NI 655X meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: IEC 61010-1, EN 61010-1 UL 3111-1, UL 61010B-1 CAN/CSA C22.2 No. 1010.1 EN 55011 Class A at 10 m FCC Part 15A above 1 GHz For UL and other safety certifications, refer to the product label or to ni.com. Immunity EN 61326:1997 + A2:2001, Table 1 EMC/EMI CE, C-Tick, and FCC Part 15 (Class A) Compliant Note: For EMC compliance, you must operate this device with shielded cabling. This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows: Low-Voltage Directive (safety) 73/23/EEC National Instruments Corporation 23 Specifications for the NI PXI/PCI-6552/6551

Electromagnetic Compatibility Directive (EMC) 89/336/EEC For full EMC compliance, you must operate this device with shielded cabling. In addition, all covers and filler panels must be installed. Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory compliance information. To obtain the DoC for this product, visit ni.com/hardref.nsf, and search by model number or product line, and click the appropriate link in the Certification column. CVI, IVI, LabVIEW, National Instruments, NI, ni.com, and RTSI are trademarks of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents. 2003 2004 National Instruments Corp. All rights reserved. *323309B-01* 323309B-01 Jan04