An Evaluation of Artifact Calibration in the 5700A Multifunction Calibrator

Similar documents
5790A Automated AC Measurement Standard

Using Reference Multimeters for Precision Measurements

Migrating from dc voltage dividers to modern reference multimeters

5080A Multi-Product Calibrator

Oscilloscope Calibration Options for 55XX Series Multi-Product Calibrators

734A and 7001 DC Reference Standards

Fluke 170 Series True-rms Digital Multimeters

5520A. Multi-Product Calibrator. Extended Specifications

Making sense of electrical signals

1586A Super-DAQ Precision Temperature Scanner. The most accurate and flexible temperature data acquisition system on the market. Fluke-Direct.

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications

Fluke 5820A Oscilloscope Calibrator Specifications

5500A. Multi-Product Calibrator. Extended Specifications 2005

5520A. Multi-Product Calibrator. Extended Specifications 2005

Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL

5320A. Multifunction Electrical Tester Calibrator. Extended Specifications

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10

9190A Ultra-Cool Field Metrology Well. Ultra-cool dry-block calibrator with best-in-class stability

742A Series Resistance Standards

A New Method for the Calibration of the mv Ranges of an AC Measurement Standard

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators

Fluke 8845A/8846A Digital Multimeters Extended Specifications

6135A/PMUCAL Phasor Measurement Unit Calibration System

5320A Multifunction Electrical Tester Calibrator. Verify and calibrate electrical test tools with a single instrument

5790B/5/AF AC Measurement Standard

Model 6600A Dual Source High Resistance Bridge

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

Fluke MDA-510 and MDA-550 Motor Drive Analyzer

Fluke 3540 FC Three-Phase Power Monitor

Ultra-cool dry-block calibrator with best-in-class stability Technical Data. Improved process control

5502A Multi-Product Calibrator. Extended specifications

The Fluke 8508A. Reference Multimeter. Reference standard accuracy and stability, in one functionally versatile, easy to use solution

Fluke 83V and 87V Digital Multimeters

PM 6685 PM 6685R Universal Frequency Counter Rubidium Frequency Counter Calibrator

Calibration Technique for SFP10X family of measurement ICs

PRECISION TEMPERATURE SYSTEMS. Compact Modular and Upgradeable Thermometry Measurement Systems

American Association for Laboratory Accreditation

1594A/1595A Super-Thermometers. Recognized worldwide for ease of use and reliable accuracy

5790A/AF AC Measurement Standard

Fluke 712B and 714B Temperature Calibrators

Fluke 1740 Series. Three-Phase Power Quality Loggers Memobox. Assess power quality and conduct long-term studies with ease

Troubleshooting. process loops. All. Application Note. Loop calibrators. ProcessMeter

The Metrology Behind Wideband/RF Improvements to the Fluke Calibration 5790B AC Measurement Standard

Maximizing your reference multimeter, minimizing measurement uncertainties

Technical Services Department, BHEL, Piplani, Bhopal, Madhya Pradesh. Discipline Electro Technical Calibration Issue Date

Fluke 370 FC Series True-rms Wireless AC/DC Clamp Meters

5522A Multi-Product Calibrator Extended specifications

Model AccuBridge 6242D Resistance Bridge

Standard versus Custom calibration of a Fluke 5720A

1507/1503. Insulation Testers. Technical Data. Features and benefits:

ScopeMeter Test Tool CSV files available for FlukeView software and Microsoft Excel Application Note

Reliable Power Meters Multipoint Power Recorder

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK

A Guide to Calibrating Your Spectrum Analyzer

Complimentary Reference Material

Measuring variable-speed motor drive output voltage with a Fluke ScopeMeter 190 Series

1587/1577 Insulation Multimeters

+/- (3 PERCENT +2) AC VOLTAGE, +/- (2 PERCENT +2) DC VOLTAGE, +/- (5 PERCENT +3) RESISTANCE. User Manual, Test Lead Wrap, TP2 Test Probes

Fluke 1660 Series Multifunction Installation Testers

MODEL INFORMATION MODEL 6010D. Automated Primary Resistance/ Thermometry Bridge

Agilent 970-Series Handheld Multimeters Data Sheet

Fluke 1587 FC/1577 Insulation Multimeters

Agilent 8902A Measuring Receiver

ACCREDITED LABORATORY

Using the 55XX Series Oscilloscope Calibration Options

CALYS Transportable documenting multifunction calibrator with high accuracy 0.01%

6100A/6101A - Alternative verification methods

PXIe Contents. Required Software CALIBRATION PROCEDURE

Use of the BVD for traceability of bipolar DC voltage scale from 1 mv up to 1200 V

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER 34401A (HEWLETT-PACKARD)

ULTRA-PRECISE AIR RESISTANCE STANDARDS

Application Note. Calibration of 7000 Series Precision LCR Meters

The future of installation testing is here. The Fluke 1650 Series Multifunction Testers

Standard Resistors Precision DC Shunts Digital Thermometers Non-inductive AC Shunts Intelligent Teraohmmeter Precision Decade Standards Precision Air

Discover. Blue Box. the. Difference. High Resistance Metrology Products Guide

MODEL 5002 PHASE VERIFICATION BRIDGE SET

So advanced, we had to create a new category

Use of infrared thermography in electronics

Voltage Sensors URV5-Z

Agilent 86030A 50 GHz Lightwave Component Analyzer Product Overview

Installation Tester and DMM Kits

Keysight Technologies 1 mw 50 MHz Power Reference Measurement with the N432A Thermistor Power Meter. Application Note

5075 Precision Digital Multimeter Time Electronics Calibration, Test & Measurement

The Fluke 6100A. Electrical Power Standard. The most accurate, comprehensive and flexible source of electrical power signals

Four-Channel Sample-and-Hold Amplifier AD684

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

Model 1140A Thermocouple Simulator-Calibrator

PERFORMANCE ASSESSMENTS OF THERMOMETER RESISTANCE BRIDGES

Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals

Isolated High Level Voltage Output 7B22 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Calibration Techniques for the Home Lab

Manual Supplement Fluke Corporation. All rights reserved. Printed in U.S.A. x

Model 6625T Series TEMPERATURE MEASUREMENT SYSTEMS

Testing & Calibration Lab, 204, Diamond Industrial Estate No. 2, Ketki Pada Road, (Near Dahisar Toll Naka), Dahisar (East), Mumbai, Maharashtra

I STATISTICAL TOOLS IN SIX SIGMA DMAIC PROCESS WITH MINITAB APPLICATIONS

University of Pittsburgh

Statistical Software for Process Validation. Featuring Minitab

Thermocouple Conditioner and Setpoint Controller AD596*/AD597*

Study of the HD target spin rotations during G14

MODEL INFORMATION. MODEL AccuBridge. Self Calibration Ratio Bridge

Transcription:

An Evaluation of Artifact Calibration in the 57A Multifunction Calibrator Application Note Artifact Calibration, as implemented in the Fluke Calibration 57A Multifunction Calibrator, was a revolutionary concept when it was introduced in 1988. Today, with many thousands of instruments deployed throughout the world, questions often arise about how well Artifact Calibration stands up over time in real world applications: Does Artifact Calibration really work? Is the 57A stable enough to support its specifications for periods up to one year between Artifact Calibrations? Is the two year full verification really needed? Can I really expect a 57A calibrated in the factory to meet its specifications? This application note answers these questions. The data base As Found and As Left Data The Fluke Northwest Technical Center is equipped to repair and calibrate the 57A. Upon arrival, As Found data is taken for operational instruments. For this study, data on approximately 26 instruments was available. The number is approximate because some instruments were returned for repair and may not have operated properly in all functions. Factory test data from new instruments was also used. As Found data was processed in Microsoft Excel spreadsheets, using the statistical and graphing functions. As noted above, not all data was useable, and in some cases, outlying data points more than four standard deviations from the mean were discarded. As a result, the As Found data set analyzed consists of those points within four standard deviations of the mean. Before they are returned to their owners, instruments are calibrated using Artifact Calibration, then externally verified where As Left data is collected. Production test data Data obtained in the final production test of 57As is uploaded to a computer and is used to control the production process. Production test stations are maintained under rigorous control through the use of check standards (for process control) and through Process Metrology (to assure processes are maintained properly centered). A key requirement of the quality system is the C pk > 1.33 (four sigma process) which must be maintained for each parameter. From the Fluke Calibration Digital Library @ www.flukecal.com/library

This data is useful for resolving some questions about the error contributions of the service center measuring system. Interpreting the data This is complicated by the need to separate instrument performance from measuring system performance. Complete analysis of instrument performance requires complete knowledge of the contributions of the measuring system. Therefore, it was necessary to accurately assign offsets and variability to the measuring system or to the 57A. In most cases, the combination of As Found and As Left data provides enough information to separate those error parameters with good confidence. Does Artifact Calibration really work? How it works the idea behind it all Artifact Calibration incorporates all the metrology functions usually performed by a standards or calibration laboratory into the instrument. Thus, the 57A contains internal reference standards, an internal ratio device and a sensitive null detectorunder the control of its internal software, the 57A performs the metrology procedures necessary to determine offsets from nominal of its outputs, and to make internal corrections to drive the offsets essentially to zero. Artifact Calibration provides superior accuracy because the calibration environment and procedures are fixed by the design of the instrument, and thus are not subject to the range of conditions and capabilities found in the community of calibration laboratories. Performing Artifact Calibration returns the direct voltage (dv), direct current (dc) and resistance functions to essentially the same accuracy the -1 mv -1 V -1 V Correlation Analysis - As Found 1 mv 2-2 2-2 +1 mv Correlation Analysis - As Found 1 V 2-2 2-2 +1 V Correlation Analysis - As Found 1 V 1-1 1-1 +1 V -1 mv -1 V -1 V Correlation Analysis - As Left 1 mv 2-2 2-2 +1 mv Correlation Analysis - As Left 1 V 2-2 2-2 +1 V Correlation Analysis - As Left 1 V 1-1 1-1 +1 V Figure 1. Scale values are ppm. Shaded boxes show one year specification, As Found, and 24 hour specification, As Left. 2 Fluke Calibration Artifact Calibration in the 57A

instrument had when it was first manufactured. This is accomplished by adjusting internal references by comparison to external standards, verifying ratio accuracy, performing measurements and making the adjustments needed to bring the outputs to nominal. Alternating current (ac) and alternating voltage (av) outputs are returned to nominal referenced to an internal ac/ dc difference standard, which is not adjusted during the process. Artifact calibration of av and ac differ from the other functions only in that the internal reference is not measured or adjusted. Long term accuracy of those parameters depends upon long term stability of the internal ac/dc difference standard. As noted later (Figures 4 and 5) the long term stability of the external ac/dc difference standard would support verification intervals of several years in length. Artifact Calibration is performed at intervals determined by the specifications to be maintained. For example, Artifact Calibration must be performed each 9 days to maintain the 57A s 9 day specifications. The 57A also provides a Cal Check function that may be used to monitor outputs relative to the internal standards. This feature provides users with a means for implementing a measurement quality program as required by paragraph 9.6 of ANSI/NCSL Z54-1 which requires in-service checks between calibrations and verifications. How well it works the results How well Artifact Calibration works is best shown by comparing As Found and As Left data for the direct voltage ranges. There we have equal voltages of opposite sign that are adjusted essentially identically by the Artifact Calibration process. Figures 1 and 2 are scatter plots of As Found and As Left measurement results on 72 instruments in our sample judged not to have been repaired. The measured differences from nominal for dv are such that errors that affect both positive and negative output the same way (that is, internal and external standard error and ratio error) are positively correlated. The first are gain errors and will result in a straight line running at an angle from left to -1 V -1 V Correlation Analysis - As Found 1 V 12-12 12-12 +1 V Correlation Analysis - As Found 1 V 14-14 14-14 +1 V right at about 45. The second are zero errors and results in a line running at an angle from left to right at about -45. Random noise shows up as a more or less circular distribution of points. The boxes indicate specification limits one year for As Found, and 24 hour for As Left. Keep in mind these plots include not only the errors in the 57A, but also errors contributed by the measuring system. All plots show measured difference from nominal in parts per million (ppm). -1 V Correlation Analysis - As Left 1 V 12-12 12-12 +1 V Figure 2. Scale values are ppm. Shaded boxes show one year specification, As Found, and 24 hour specification, As Left. -1 V Correlation Analysis - As Left 1 V 14-14 14-14 +1 V 572A 24 hour specification (99 %) 3 Fluke Calibration Artifact Calibration in the 57A

These plots show three things: Only a few points are outside specification As Found. Do these result from errors in the measuring system? All outputs are adjusted to well within 24 hour specifications. There is significant positive correlation in all but the 1 mv plots (Figure 1). In the As Found plots, these result from drift in internal standards, and to some extent, from less than ideal adjustment in the Artifact Calibration process. Positive correlation in the As Left plots result from less than ideal adjustment by Artifact Calibration. 57A Stability Many of the instruments in the data sample were measured several times during an approximately five year span, and presumably were Artifact Calibrated at one year or shorter intervals. Plotting As Found Data versus time since manufacture will reveal any long term drifts either in the instrument or the measuring system. That being true, it becomes possible to assess worst case limits for time stability of the 57A. The approach is shown in Figure 3. The upper portion of the figure shows the raw data for 2 mv, 1 khz. (This parameter was chosen because it is one of the few having significant drift.) The graph in the 1-1 -2 57A Alternating Voltage - As Found 3 mv, 1 khz -3 5 1 15 2 25 6 3-3 -6 Days Since Manufacture Computation of Time to 1 % Outside 1 Year Spec. 2 mv, 1 khz -9 5 1 15 2 25 3 Time Since Manufacture (Years) OOS @ ~19 Years Figure 3. How time to 1 % Out of Specification is calculated. lower portion shows how time to the 1 %-Out-Of-Specification point is determined. This is the point at which the regression line ± 2.58 σ intercepts the upper or lower specification limit, σ being the standard deviation of points about the regression line through the data. Thus, the time to 1 %-Out-Of-Specification is influenced by the drift and noise in the instrument, and in the measuring system. Offsets from nominal are ignored in this stability analysis, but of course will always impact actual performance. Regression UPL LPL USL LSL 4 Fluke Calibration Artifact Calibration in the 57A

Figures 4 though 8 are bar charts presenting the results of this analysis. They represent points measured in the full verification procedure. There are too many av points to present in a single graph, so only the important specification breakpoints are presented. The charts indicate that with a few exceptions the stability of the 57As, maintained as recommended, is sufficient to support much tighter specifications. Of course, stability is only one contributor to instrument accuracy uncertainty of standards and variability and errors in the adjustment process also contribute. In Figure 7, excessive variability makes it impossible to calculate stability for ± 1 mv and ± 1 V. 1 1 1 Alternating Current Stability 1 Hz 2 Hz 4 Hz 1 khz 5 khz 1 khz Frequency.22 ma 2 ma 2 ma 2 ma 2 A 35 3 25 2 15 1 5 1 mv 57A Time Stability - Direct Voltage -1 mv 14 21 19 1 V -1 V 1 V -1 V 1 V Current Level 24-1 V 27 25 1 V -1 V Figure 4. Alternating current stability in years to 1 % out of tolerance. Figure 7. Direct voltage stability in years to 1 % out of tolerance. 1 1 1 Alternating Voltage Stability 1 Hz 4 Hz 2 khz 1 khz 1 MHz Frequency 2 mv 2 mv 2 mv 2 V 2 V 2 V 7 6 5 4 3 2 1 Ω 1.9 Ω Resistance Time Stability 57 47 41 37 3 32 39 27 32 27 25 24 26 16 13 11 9 3 19 Ω 19 1.9 19 Ω kω kω Resistance Value 19 kω 1.9 MΩ 19 MΩ Figure 5. Alternating voltage stability in years to 1 % out of tolerance. Figure 8. Resistance stability in years to 1 % out of tolerance. 12 57A Time Stability - Direct Current 1 8 6 4 2 84 39 63 39 44 64 7 6 23 27 2 µa -2 µa 2 ma -2 2-2 2 ma ma ma ma Current Level -2 ma 2 A -2 A Figure 6. Direct current stability in years to 1 % out of tolerance. 5 Fluke Calibration Artifact Calibration in the 57A

Is the two year verification really needed? Figure 9 shows why the answer to the question must be yes. The plot shows significant drift and increased variability with increasing time since manufacture. The increase in variability results from different drift rates in different instruments. That this should occur is not surprising, since Artifact Calibration does not adjust the internal ac/ dc difference standards, the reference for all alternating current and voltages. However, stability is more than sufficient to support the instrument's specifications. Will a 57A calibrated by the factory meet its specifications? This is a two part question involving the performance of the 57A and the ability of the factory to properly adjust instruments in the production process. The factory s ability to adjust instruments is closely monitored by the Fluke Standards Laboratory. 57A process metrology was implemented to assure the accuracy of production measurements. That process is described in Baldock's Artifact Calibration, Theory and Application (see bibliography) and will not be addressed further here. The best way to answer the first part of the question is to look at As Found data for a large number of instruments, all of which were initially 25 15 5-15 57A Alternating Voltage vs. Days since Mfg. 2 mv, 3 khz - 9 Day spec = 195 ppm -25 5 1 15 2 25 Time (Days) 8 4 5-4 As Found Alternating Voltage 2 mv, 1 MHz - 1 Year spec = 51 ppm -8 5 1 15 2 25 Days Since Manufacture 45 3 15-15 -3 As Found Alternating Voltage 2 V, 1 MHz - 1 Year spec = 345 ppm -45 5 1 15 2 25 Days Since Manufacture Figure 9. Long term drift and dispersion in alternating voltage outputs. 6 Fluke Calibration Artifact Calibration in the 57A

calibrated by the Fluke production facility. Figures 1 through 14 present As Found data for 26 57As serviced by Fluke. Presented are the averages and 99 % confidence limits for each parameter measured in the recommended full verification. Keeping in mind that this data includes any errors contributed by the measuring system, it is apparent that there is very high probability that an instrument will meet all its accuracy specifications, thus meeting one of the objectives of its designers there should be a 99 % probability that each parameter meets its specifications throughout the interval between Artifact Calibrations. Figure 14 indicates smaller than 99 % probability at 1 mv and 1 Vdc. A further analysis was performed to eliminate Percent of 9 Day Spec 15 12 9 6 3-3 -6-9 -12-15 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45 Voltage Frequency Percent of 1 Year Spec 18 15 12 9 6 3-3 -6-9 -12-15 -18 1 2 3 4 5 6 7 8 9 1 11 12 13 14 15 16 17 18 Resistance Value Figure 1. Deviations from nominal and 99 % limits, all verified voltages and frequencies. Figure 13. Deviations from nominal and 99 % limits, all resistance values. Percent of 9 Day Spec 12 9 6 3-3 -6-9 -12 1 3 5 7 9 11 13 15 17 19 21 23 25 27 Alternating Current/Value Frequency Percent of 1 Year Spec 15 12 9 6 3-3 -6-9 -12-15 1 2 3 4 5 6 7 8 9 1 Voltage Value Figure 11. Deviations from nominal and 99 % limits, all verified currents and frequencies. Figure 14. Deviations from nominal and 99 % limits, all verified direct voltages. Percent of 9 Day Spec 12 9 6 3-3 -6-9 -12 1 2 3 4 5 6 7 8 9 1 Current Level Figure 12. Deviations from nominal and 99 % limits, all verified direct current levels. 7 Fluke Calibration Artifact Calibration in the 57A

the measuring system contributions. The results of the analysis are presented in Table 1. The results indicate > 99 % probability that a 57A will meet its one year direct voltage accuracy specifications, except at 1V, where the probability is 98.4 %. Refer to An Assessment of Artifact Calibration Effectiveness for a Multifunction Calibrator for a more complete discussion of this analysis and its conclusions. Conclusion Clearly, the answer to each of the four questions posed at the beginning of this paper is yes. Artifact Calibration indeed works as designed. Most if not all points were found well within specification at calibration. All outputs were adjusted to well within the 24 hour specification afterward. Two year verification is recommended. Instrument variability increases with time from manufacture. Verification identifies those few instruments that do drift while meeting specifications. With the exception of 1 Vdc, there is a greater than 99 % probability that all 57As will meet their accuracy specifications. At 1 Vdc, the probability is a slightly smaller 98.4 %. 1 year 24 Hour As Found CSS As Left Time Production 57A 57A Sigma Minimum Spec ± Spec ± Mean Std Dev Mean Std Dev C Trend? Offset Std Dev Multiple % Found Voltage ppm ppm pk (24 Hrs) Mean Std Dev C pk (24 Hrs) in Spec.1 17 14.5 1.8 7.2 3. 5.4.7 No -.6 2.1 2.2 -.7 5.1 3.2 99.8 -.1 17 14.5-2.1 6.8.4 4.6 1. No -1.2 2.9 1.6-1.3 5.7 2.8 99.4 1 9.2 7.2-2.1 3.6 -.6 1.2 1.8 No -1.4.6 3.1 -.2 3.5 2.6 99.1-1 9.2 7.2-2.3 3.5 -.5 1.1 2. No..9 2.7-1.8 3.5 2.1 98.4 1 8.4 5.4-1.2 2.1 -.3.6 2.7 No -.9.5 3.2. 2. 4.1 1. -1 8.4 5.4-1.5 2.5 -.2.6 2.7 No -.6.4 3.7 -.6 2.4 3.2 99.9 1 1 7-2.1 2.8 -.9 1.3 1.6 No 1..6 3.4-2.2 2.6 3. 99.9-1 1 7-1.9 2.8 -.6 1.2 1.7 No -.6.6 3.8 -.7 2.5 3.6 1. -1 11.6 8.6-1.7 2.8 -.3 1.6 2.2 No -2.1 1. 3.7.8 2.5 3.9 1. 1 11.6 8.6-1.9 2.8 -.5 1.6 2.2 No -3.1 1. 3.3 1.7 2.5 3.8 1. Note: CSS results have opposite sign from others. Table 1. Summary of direct voltage analysis. Artifact Calibration and the development of the 572A Multifunction Calibrator The design goal of the 572A Calibrator was to specify the lowest possible output uncertainties while preserving the ease-of-use and simplified support qualities that made the 57A so popular. The first task was to determine what changes would be necessary to make the instrument and manufacturing processes to support the improved uncertainties. The study outlined here, along with other factory test data, supported those requirements. The first improvement was in Artifact Calibration. Lower uncertainties would require outputs to be adjusted even more closely to the ideal levels. The reason for this can be seen in the scatter plots in Figures 1 and 2. As the As Left plots show, Artifact Calibration adjusts the 57A back to a small percentage of its 24 hour specification. However, as you can see here, the 24 hour specification of the 572A, described by the smaller box in Figure 2, As Left Data, 1 V, requires even more precise adjustment to nominal. A large part of the 572A development effort concentrated on improving the mean values returned by Artifact Calibration for all parameters. The second challenge was to lower the uncertainty of the factory test system, particularly for those output values affected by thermal emfs. More rigorous testing procedures and enhancements to the Process Metrology system used to maintain the 572A factory test equipment permitted a dramatic improvement in both the mean values of measurements as well as the noise level. In addition to dramatically reduced uncertainty specifications, the 572A permits the user to select either 99 % or 95 % confidence levels for the uncertainties reported by the calibrator when the Spec key is pressed. Also added were remote emulation of the 52A AC Calibrator, a screen saver mode, lower dcv noise and a zero reminder. 8 Fluke Calibration Artifact Calibration in the 57A

References Baldock, Paul, Artifact Calibration, Theory and Application, Application Note B218, Fluke Corporation, 1991. Eccleston, Larry, Advancements in Accurate AC Voltage Sources, Fluke Corporation. Huntley, Les, A Preliminary Assessment of the Effectiveness of 57A Artifact Calibration, Proceedings of the 1996 Measurement Science Conference, Anaheim, CA, January, 1996. Huntley, Les, An Assessment of Artifact Calibration Effectiveness for a Multifunction Calibrator, Proceedings of the 1995 National Conference of Standards Laboratories, Dallas, TX, July, 1995. Huntley, Les, Murray, Jill Ellen, An Application of Process Metrology to the Automatic Calibration of Instruments, Proceedings of the Measurement Science Conference, Irvine, CA, January 1987. Huntley, Les, Process Metrology and the Automated Factory, Test and Measurement World Expo, April, 1986. Fluke Calibration. Precision, performance, confidence. Fluke Calibration PO Box 99, Everett, WA 9826 U.S.A. Fluke Europe B.V. PO Box 1186, 562 BD Eindhoven, The Netherlands Web access: http://www.flukecal.eu For more information call: In the U.S.A. (877) 355-3225 or Fax (425) 446-5116 In Europe/M-East/Africa +31 () 4 2675 2 or Fax +31 () 4 2675 222 In Canada (8)-36-FLUKE or Fax (95) 89-6866 From other countries +1 (425) 446-55 or Fax +1 (425) 446-5116 Web access: http://www.flukecal.com 28-213 Fluke Calibration. Specifications subject to change without notice. Printed in U.S.A. 11/213 1263377C_EN Modification of this document is not permitted without written permission from Fluke Calibration. 9 Fluke Calibration Artifact Calibration in the 57A