MANUAL. Electrostatic Field Meter EFM 235 / 255. System Prof. Dr.-Ing. Hans Kleinwächter

Similar documents
MANUAL. Electrostatic Field Meter EFM 120. System Prof. Dr.-Ing. Hans Kleinwächter

MS6231 DIGITAL ENGINE ANALYZER OPERATOR S MANUAL CONTENTS CONTENTS

MS8211 DIGITAL MULTIMETER INSTRUCTION MANUAL

PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D

OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER

LF Conductivity Measuring Probe

DIGITAL MULTIMETER INSTRUCTION MANUAL

CD770 DIGITAL MULTIMETER INSTRUCTION MANUAL

RAGU 81D DIGITAL MULTIMETER OPERATION MANUAL

MS2030 CAT III 600 V A V AUTO RS232

Operating Instructions / Voltage Tester per IEC with Phase and Polarity Tester and Phase Sequence Indicator

MS8216 DIGITAL MULTIMETER OPERATOR S MANUAL

3B SCIENTIFIC PHYSICS

Operator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique

SIMEAS-T. Operating Instructions Transducer without auxiliary power. 7KG6111 and 7KG6101. Operating Instructions

Schlöder GmbH - EMC Test and Measurement Systems Model #

Operating Manual * * Differential pressure transmitter. Table of Contents. 1 Safety guidelines. 1.1 General Information

DM-46 Instruction Manual

USER'S MANUAL DMR-6700

Item ref: UK MTTR01 TRUE RMS DIGITAL MULTITESTER WITH USB INTERFACE. User Manual

400Amp True RMS AC Clamp Meter + NCV

200Amp AC Clamp Meter + NCV Model MA250

BATTERY GRIP INSTRUCTION MANUAL

GFL-1000 User Manual Ground Fault Locator

AMM-1022 Digital Multimeter USER`S MANUAL

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output

MD-45 AC MD-45 LV/HV INSTALLATION MANUAL

The Meter Structure The Figure 2-1 shows the Meter structure.

TOS5300 SERIES Hipot Tester/Hipot Tester with Insulation Resistance Test

AC/DC DIGITAL CLAMP METER OPERATION MANUAL

MS8250A/B OPERATION MANUAL MS8250A. Hz% FUNC REL RANGE REL HOLD OFF 10A. Hz% A NCV. Hz% COM. A ma 10A FUSED 600V CAT IV.

Rudder or propeller angular position transmitter KINAX WT720

CL900. True RMS 1000V 2000A 60MΩ ENGLISH. INSTRUCTION MANUAL 2000A Digital Clamp Meter. Measurement Technology

MS8250D DUAL DISPLAY DIGITAL MULTIMETER User s Manual

MM V 10A ENGLISH. INSTRUCTION MANUAL Auto-Ranging DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE

CAP ANALOG 410 INSTALLATION & OPERATION MANUAL

Thyristor power controller B DI1002 V1.0 Manual/Operating instructions

AC/DC CLAMP METER USER S MANUAL

MM700. INSTRUCTION MANUAL Auto-Ranging Digital Multimeter True RMS

DVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use

Thermocouple - Transducer

Model DIGITAL MULTIMETER

RS Pro ENGLISH. Datasheet

MM700. True RMS ENGLISH. INSTRUCTION MANUAL Auto-Ranging. Measurement Technology

Analog Technologies VC99. Multimeter FEATURES

audionet amp II G2 Owner's Manual Mono - Amplifier

R-F Skewed Hybrids. Type H1SB and H1SB-R. & R-F Balanced Hybrids Type H1R, H3X and Type H1RB, H3XB and Type H1RB-40. System Manual CH44 VER03

DL6508 User's Guide. 1500Amp True RMS AC/DC Clamp Meter

Starship Lighting Kit

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751

3B SCIENTIFIC PHYSICS

Advanced Test Equipment Rentals ATEC (2832)

Dawson DDM181. Pocket-Size Autorange Digital Meter User s Manual

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español

CONTENTS MS2033A. 1.Introduction Safety Information...01

5XP-A 15XP-A 35XP-A Compact Digital Multimeters

DIGITAL CLAMP METER Use s Manual

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL

Dawson DDM190. Digital Multimeter User s Manual

DIGIT & POINTER MULTIMETER

User's Guide. 800 Amp AC/DC True RMS Clamp Meter. Model EX Washington Street Melrose, MA Phone Toll Free

True RMS AC / DC Power Clamp Meter Model

MDW-45 Converter RS RS-422/485

ProfiScale MULTI Multimeter

Visit us at

Bedienungsanleitung Instruction Manual. testo 815 Sound Level Meter

MIT400/2 CAT IV Insulation testers

MAGNETIC FIELD METER Operator s Manual

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

MAGNETIC FIELD METER Operator s Manual

DDM350 Pen-Type Digital Multimeter User s Manual

MIT400/2. CAT IV Insulation testers MIT430/2: CAT IV Insulation testers

SOLDERING STATION. Thermo-Control Anti-Static. Operating Instructions. 405 S. Pioneer Blvd. Springboro, OH MCMElectronics.

WE-525T Antenna Analyzer Manual and Specification

SL300 Snow Depth Sensor USL300 SNOW DEPTH SENSOR. Revision User Manual

DMM8900 SERIES USERS MANUAL

MI-10kV. Technical specifications Operating instructions. Supplier:

EM420A/420B DIGITAL MULTIMETER OWNERS MANUAL Read this owners manual thoroughly before use

Pressure transmitters MBS 4201, MBS 4251, MBS 4701 and MBS 4751

Model Number Structure. Ordering Information. Solid-state Power OFF-delay Timer H3DE-H. Model Number Legend. List of Models

2015 RIGOL TECHNOLOGIES, INC.

Efcon Water. Installation & User Manual Electro Magnetic Flowmeter (EMF) Efcon Water compact Electro Magnetic Flowmeter (EMF) Manual Version UK

OPERATING INSTRUCTION

OPERATOR S INSTRUCTION MANUAL

Model ST Instruction Manual. True RMS Autoranging Digital Multimeter. reedinstruments. www. com

1. SAFETY INFORMATION.1 2. DESCRIPTION SPECIFICATIONS.6 4. OPERATING INSTRUCTION Voltage measurement Current measurement 10

User s Guide. 400A AC/DC Clamp Meter. Model MA220

Model: Pro95 TRUE RMS MILLIAMP CLAMP METER

PRODUCT SPECIFICATIONS

Instruction manual TAG 2550-C Tester + Analyser and software: PT9 Report Creator Manual for initiation and maintenance No.

Inductive sensor slot-type SI2-K08-AP7

Model True RMS AC/DC 30A Mini Clamp-on Meter. User Guide

Installation of vibration sensors

User Guide. AC/DC Mini Clamp Meter. Model Introduction. Large 3 3/4 digit LCD display with 40 segment bargraph

Agilent U1253B True RMS OLED Multimeter. Quick Start Guide

Instruction Manual for Digital Grounding Resistance Meter. Table of Contents

Computer-14d - xx - 144a

IF30. User's manual. Description. Table of contents IF30

Telemetrie-Messtechnik Schnorrenberg

Thank you again for choosing AstroAI, if you have any questions or concerns regarding your product, please contact us at

Transcription:

MANUAL Electrostatic Field Meter EFM 235 / 255 System Prof. Dr.-Ing. Hans Kleinwächter

2 Manual EFM 120 Contents General information about electrostatics... 3 Origin of electrostatic charge... 3 Product description... 3 Measuring principle... 3 Application Areas... 3 Specifications... 4 Legend... 4 Start Condition... 4 Manual... 5 Startup Menu... 5 Main Menu... 5 Sub Menu => MODE... 6 Sub Menu => ZERO... 6 Sub Menu => START... 7 Sub Menu => OFF... 8 Measure Range... 9 Tripod Connection... 9 Battery Supervision... 9 Grounding... 10 Electric Field Strength... 10 Voltage Metering with MK1 / HMK40... 10 Scope of Delivery... 11 Maintenance... 11 NiMH Storage Battery... 12 Warranty... 12 Warning Notices... 12 Calibration... 12

General information about electrostatics 3 General information about electrostatics Nowadays ESD 1 is a problem at many workstations, because modern microelectronics 2 are easily destroyed by the sequels of ESD. Other branches of industry like e.g. telecommunications-, plastics-, and explosive material industry are also heavily affected by ESD. ESD causes losses of time as well as high financial losses and can endanger the human health. Charges of over 10000 Volts can have effects on people, clothes, materials, and equipment. Devices that are sensitive to electrostatics can be damaged by electrostatic discharges of less than 100 Volts. Charges of 3000 Volts and more can cause sparks. In endangered areas that can cause explosions. Origin of electrostatic charge Triboelectricity 3 is caused by attrition of different materials. Electrons are transferred from one material to the other. As electrons are charged negative the material that releases electrons is charged positive. The material emitting electrons is charged negative. There are different ways to avoid or to discharge electrostatic charges. But to find an effective and reasonable solution first the emitted amount and polarity of the charge must be found. Our devices are suitable for that purpose and for the supervision of desired charge. Product description The unit is housed into an EMV aluminum enclosure. The influence electrode is star shaped. In front of it in a small distance apart a rotating grounded modulation propeller with the same shape as the electrode is located. The influence electrode is enclosed by a ring electrode system that is used as a mechanical shield for the propeller wheel and the sensor plate. Measurement values are displayed on a graphic LC display. The device has an integrated micro processor with USB 2.0 Interface. Measuring principle The electrostatic field meter is a parametric amplifier. The electric field influences a current proportional to the electrostatic field. The current is amplified and measured with a selective amplifier. No energy is taken from the field over time means. No radioactive matters are used! Application Areas Detection and Control of electrostatic fields resp. charges, measuring of electric charges, electrostatic charges, very high-resistance voltage sources, very low currents, or high resistances. 1 electrostatic discharge 2 integrated circuits 3 Greek: tribeia = friction

4 Manual EFM 120 Specifications Dimension (L x W x H): 170mm x 77mm x 73,5mm Weight: 550gr Calibration in a plate capacitor: 400mm x 400mm, Distance 100mm Calibration accuracy: < 2% Interface: Mini USB 2.0 Buchse Power supply: Power supply 230V 50Hz sek. 9V 300mA Battery: 4 xnimh rechargeable Battery AA Operating time: Min. 4h Legend 1 Graphic LC-Display 6 Charge Control-LED 2 Zero - Potentiometer 7 Power Interface 3 ENTER Button 8 Modulator System with cover (protective cap) 4 On Button 9 Connection for MK1 5 Mini USB Interface 10 Grounding Jack Start Condition The protective cap on the modulator system in front must be removed before measuring!

Manual 5 Manual By pressing the «ON» Button (4) on the back of the unit it switches on. In the display it appears for 3 sec. : Startup Menu It shows the Type = EFM 235/EFM 255 and the Version Number V X.X After that, the unit switches to: Main Menu In the dark field it shows the selected Mode. Press the Button (4) to select the Sub Menu. Selected Menu is black => ( START ) Press the ENTER button (3) to go to the selected Sub Menu.

6 Manual EFM 120 Sub Menu => MODE First the upper line is black. By pressing the Button(4) you can change the Mode. Selection => E-Feld => V-Meter MK1 => HV-Met.MK40 => E-Feld => etc. Press the ENTER button (3) to choose the selected Mode and the 2nd line becomes black By pressing the Button(4) you can switch between AUTO and MANUAL. Press the ENTER button (3) to choose the selection and the 3 rd line becomes black. Press the ENTER button (3) again to go back to the Main Menu. By pressing the Button(4) you come back to the 1st. line and you can start a new selection. Sub Menu => ZERO In the dark field it shows an advice to set the cover (protection cap) onto the modulator system. Press the Button (4) to select START or BACK. Press the ENTER button (3) by Start the Zero adjustment starts, by Back you switch back to the Main Menu. The buttons are out of function during the Zero procedure. Do not remove the cover during this procedure! Displayed Range 0-3 is in process.

Manual 7 After successful Zero adjustment it displays Press the ENTER button (3) to go back to the Main Menu. By pressing the Button(4) and select Start then pressing ENTER button (3) you can restart the Zero adjustment. If no Zero adjustment is possible it appears in the display: In this case adjust by the Zero Potentiometer on the back the Voltage to V < 0,10V After that, press the ENTER button (3) to Start the calibration again. Sub Menu => START In the 1st. line it shows the selected Measure Mode => AUTO and the actual Range => 2kV/m. In the 2nd line it shows by E-Feld Measurement E = (actual value) kv/m Press the ENTER button (3) to go back to the Main Menu.

8 Manual EFM 120 In case the MANUAL Mode is chosen it appears in the display: In the 1st. line it shows the selected Measure Mode => MANUAL and the Selected Range => 2kV/m. In the 2nd line it shows by E-Feld Measurement E = (actual value) kv/m By pressing the Button(4) you can change the Measure Range. Press the ENTER button (3) to select BACK and again to go back to the Main Menu. By selected Measure Mode => Voltmeter MK1 or HMK40 it appears in the display: The service is like the E-Field Mode. Only in the 2nd line it shows the voltage V = not E = and V not kv/m. Sub Menu => OFF Press the Button (4) in Main Menu to select OFF Then Press the ENTER button (3) to switch off. After a few seconds the instrument will switch off!

Measure Range 9 Measure Range By E-Field Measurement you can chose the following Ranges: Range EFM255 EFM235 Gain Factor 3 200V/m 2kV/m x 100 2 1kV/m 10kV/m x 20 1 4kV/m 40kV/m x 5 0 20kV/m 200kV/m x 1 By Voltage Measurement with the HV Measuring Heads you get following ranges: Range MK1 255 MK1 235 MK40 235* 3 2 V 20 V 400 V 2 10 V 100 V 2 kv 1 40 V 400 V 8 kv 0 200 V 2 kv 40 kv * The Measuring Head HMK 40 is only applicable with the EFM 235.. In Mode AUTO the ranges switches automatically! Tripod Connection To use the Instrument on a table (with MK1) you can change the handle to a tripod. To release the hand grip, loosen the attaching finger screw (wheel). Then screw the tripod into the recessed thread. Please notice, that the screw thread in the recessed thread is a M6 thread and the thread on the tripod is an imperial thread. Screw it until finger tight. Battery Supervision The device has an integrated battery supervision function. If the battery voltage decreases below 4.8V, in the display appears Low Battery. By decreases below 4.5V it appears in the display Switch OFF To prevent total discharge of the battery the device is automatically powered off. In this case the battery has to be loaded on the power interface (7) with the shipped charger for about 12-14h.

10 Manual EFM 120 Grounding The device must be grounded properly to measure the amount and the polarity of an electrostatic field. For that purpose, the device must be connected to ground by using the grounding jack (10) on the bottom. Grounding the device by a grounded person (e.g. over conductive shoes or wrist strap) touching the handle suffices under most circumstances. Electric Field Strength For measuring the electric field strength at first the protective cap must be removed. Afterwards the electro static field meter must be adjusted with the modulator system parallel to the object to be measured. The distance between the EFM 120 and the object to be measured must be kept constant, as the distance is for the voltage calculation. The measured field strength is displayed in kv/m. The charge in Volt is obtained by multiplying the displayed value with the distance in meters. For example: The distance between the object and the device is d=5cm and the measured field strength E =1,6kV/m, the surface potential U is calculated as follows: Surface charge = field strength x distance d (m) U = E x d (in m) => U = 1600V/m x 0,05m = 80 V Please pay attention to use correct units. To avoid these problems, we recommend using SI units. Voltage Metering with MK1 / HMK40 With the optional voltage gauge head MK1/HMK40 voltages up to ±2kV/40kV can be measured with an internal resistance R i >10 16 Ω. For high voltage measurements the guidelines of the European standard EN 10100 need to be considered. The voltage gauge head must be mounted on the modulator system of the static meter for measurements The generated linear voltmeters have the following outstanding properties: Input capacitance: Input resistance: approx. 5pF >10 16 Ω with appropriate gauge head

Scope of Delivery 11 Scope of Delivery The basic equipment of the electrostatic field meter includes the following components: Carrying case with ESD foam Electro static field meter EFM 235/255 Measuring diaphragm Power supply USB cable with mini plug tripod Grounding spiral cable CD with EFM_ReadOut Software Manual Declaration of factory calibration Optional available: Voltage gauge head MK1 High Voltage gauge head HMK40 Thread measuring head By using our BGT 200 Set including MK1, Hand Electrode and 3m PTFE cable you can make a walking test conform to the standard DIN EN 61340-4-5. Maintenance The modulator system or parts of it must not be touched. Those parts must be protected from dust, isolating impurity layers, color- or paint haze and condensate. If required, the modulator system can be cleaned with ethyl alcohol and a lint-free cotton cloth.

12 Manual EFM 120 NiMH Storage Battery Even if the device is not used for longer terms, the battery must be charged after 6 months to prevent total discharge. If the battery cannot be charged anymore, the battery must be changed. For that purpose, the device must be sent to the manufacturer. Warranty We provide a 24-month warranty in case of proper application according to the manual. Excluded of the guarantee are: The rechargeable battery, damage by electric shock, wrong grounding, and mechanical damage of the device. The guarantee expires if the device was opened. Warning Notices The electrostatic field meter must not be opened. By opening the guarantee expires. The electrostatic field meter must not be used in explosive areas. No admission for explosive areas. If high charges are possible the electrostatic field meter must be grounded. An adequate distance must be kept. Flashovers on the modulator system must be avoided! The use of the device in power plants or comparable areas is prohibited! The device cannot measure alternating fields > 1Hz! Calibration A check of the measurement values is recommended annually. 2017 - Kleinwächter GmbH