Technical Information Memosens CLS82D

Similar documents
Technical Information Condumax CLS15D/CLS15

Technical Information Condumax CLS21D/CLS21

Condumax CLS16 and CLS16D

Condumax CLS15 and CLS15D

Condumax CLS15 and CLS15D

Condumax W CLS15 and CLS15D

Condumax CLS21 and CLS21D

Condumax CLS21 and CLS21D

Condumax W CLS15 and CLS15D

Technical Information Orbipac CPF81D/CPF82D and CPF81/CPF82

Indumax CLS54D. Technical Information

Technical Information Indumax CLS50D/CLS50

Technical Information Indumax CLS54

Technical Information Orbipore CPS91D and CPS91

Technical Information Smartec CLD18

Technical Information Smartec CLD18

Conductivity Measuring Cells for High-Temperature Applications ConduMax H CLS 16

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

Technical Information Memosens CPS96D

Ceramax CPS341D. Technical Information

Indumax CLS50/CLS50D. Technical Information

Technical Information Smartec CLD18

Orbipac CPF81/CPF82 and CPF81D/CPF82D

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

Memosens CPS96D. Technical Information

Indumax H CLS54. Technical Information

Memosens New technology for contactless, inductive, digital transmission of ph signals

Technical Information Memosens CCS50D

Smartec S CLD134. Technical Information

Conductivity Sensor for High-Temperature Applications ConduMax W CLS 13

Conductive conductivity measurement mypro CLM 431 / CLD 431 conductive

Technical Information OUSBT66

Tophit CPS441 and CPS441D

Conductivity sensor for hygienic applications

The transducer component converts the measured signal and displays the actual value.

A Higher Level of Performance. Data Sheet. Series Pressure and Level Transmitters. For more information, please visit >

Flow Measurement SITRANS F M

Magnetic-Inductive flow Meter fmi

OPTIFLUX 5000 Technical Datasheet

Assembly 249 (9.80) 259 (10.2) Example: level probe with Pt100 and HUT temperature head transmitter ( ma) ma. Core colors: RD BK YE BU

OPTIFLUX 5000 Technical Datasheet

Inductive conductivity measurement mypro CLM 431 / CLD 431 inductive

P R o D u c T I n F o R m at I o n. TDR Level Sensor LFP. The clean solution

BATCHFLUX 5500 C Technical Datasheet

Technical Information Liquisys M CLM223/253

Pressure transmitter For sanitary applications Model SA-11

FLOWave SAW-Flowmeter

Digital conductivity transmitter

Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm

Miniature resistance thermometer For sanitary applications Model TR21-B, for orbital welding

In-line resistance thermometer Model TR25

Turbimax CUS71D. Technical Information. Ultrasonic interface sensor Immersion sensor for interface measurement

SITRANS F flowmeters SITRANS F M. Flow sensor MAG /48 Siemens FI Overview

omnigrad M TR 46 Technical Information Hygienic in-line RTD thermometer PCP ( ma), HART or PROFIBUS-PA electronic

Conductivity measurement SmarTec S CLD 132

Tophit CPS491 and CPS491D

Tophit CPS491 and CPS491D

Miniature resistance thermometer For sanitary applications, for orbital welding Model TR21-B

Flow Measurement SITRANS F M. Flow sensor MAG /49. Overview

OEM radar transmitter, for aggressive media level measurement

Level sensor Magnetostrictive, high-resolution measuring principle For sanitary applications, model FLM-H

Pharma Temperature Sensor TFP-EX

Inductive Conductivity Meter ILM

Temperature Sensor for Build-In System G3/8" FLEXadapt

Temperature Sensor for Build-In System G3/8" FLEXadapt

Miniature resistance thermometer For sanitary applications Model TR21-C, with integrated sterile connection

Type F-20-3A Sanitary Pressure Transmitter NEMA 4X with Integral Junction Box

Submersible pressure sensor For oils and fuels Model LF-1

Continuous Level Sensor NSL-M

Temperature Sensor. Temperature Transmitter MPU-LCD with Display Product Information TSM and TSB FOOD SENSORS FOR FOOD AND LIFE SCIENCES.

Liquisys M CLM223/253

Submersible pressure sensor For superior applications Model LF-1

Temperature Sensor Tri-Clamp

Capacitive Limit Switch Pharma NCS

Pressure transmitter with flameproof enclosure For applications in explosion-protected areas Models E-10 and E-11

Capacitive Limit Switch Pharma NCS

Continuous Level Sensor NSL-M-00

ED701 General Industry Pressure Transmitter

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

High-quality pressure transmitter For general industrial applications Model S-10

Temperature Sensor G1/2" Standard

Temperature Sensor with M12 hygienic

Ultrasonic level measuring device, non-contact

Ultra high purity transducer, Ex na nl Models WU-20, WU-25 and WU-26

Submersible pressure sensor For water and wastewater Model LW-1

Pharma Temperature Sensor G3/8"

Process pressure transmitter Model IPT-10, standard version Model IPT-11, flush diaphragm

Level Measurement. Continuous level measurement - Radar transmitters SITRANS LR400 5/229. Overview

Pharma-Temperature Sensor M12 hygienic

Flush pressure transmitter For viscous and solids-containing media Model S-11

Resistance thermometer Model TR30, compact design

JUMO Dtrans T100 Screw-in RTD Temperature Probe with/without Transmitter

prosonic T FMU230, FMU231

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. appmeas.co.uk +44 (0)

Electromagnetic Flow Measuring System Two-wire, loop-powered PROline promag 23 H

CoriolisMaster FCB330, FCB350, FCH330, FCH350 Coriolis mass flowmeter

0 2 1 â Ó. itemp HART TMT182. Technical Information

Temperature Sensor Tri-Clamp

Temperature Sensor with G1/2" hygienic

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F

Transcription:

TI01188C/07/EN/03.16 71318348 Products Solutions Services Technical Information Memosens CLS82D Hygienic conductivity sensors, digital with Memosens technology, cell constant k = 0.57 cm -1 Application For measurements where very diverse conductivities must be measured by one measuring system. Typical applications include: Phase separations Chromatography Fermentations CIP monitoring in small pipes Ultrafiltration The sensor is used with the following transmitters: Liquiline CM44x/R Liquiline CM42 The resistivity in MΩ cm can also be measured using these transmitters. Your benefits High measuring accuracy as cell constant is individually measured Quality certificate stating the individual cell constant Hygienic process connections for installation in pipes or flow vessel IP68 protection Easy to clean thanks to electropolished surfaces Can be sterilized up to 140 C (284 F) Stainless steel 1.4435 (AISI 316L) meets the highest demands of the pharmaceutical industry The entire sensor is EHEDG- and 3A-certified FDA conformity Available with certificate of conformity for pharmaceutical requirements Available with inspection certificate EN 10204-3.1 Other advantages of Memosens technology Maximum process safety thanks to non-contact, inductive signal transmission Data security thanks to digital data transmission Very easy to use as sensor data saved in the sensor Recording of sensor load data in the sensor enables predictive maintenance

Function and system design Measuring principle The measuring cell has four electrodes. An alternating current is applied via the outer electrode pair. At the same time, the voltage applied is measured at the two inner electrodes. The electrolytic conductivity between the electrodes can be reliably established based on the measured voltage and the current flow caused by the liquid's resistance. The advantage of this technology compared to traditional two-electrode sensors is that electrochemical effects at the live electrodes are suppressed by the two additional voltage measuring electrodes. I G U + - - + + - + - + - + - + - 1 Conductivity measurement I Current intensity measurement U Voltage measurement G Generator A0024312 Measuring system A complete measuring system consists of the following components at least: Conductive conductivity sensor CLS82D A transmitter, e.g. Liquiline CM44x A Memosens data cable CYK10 1 2 3 2 Example of a measuring system 1 Memosens CLS82D 2 Liquiline CM44x transmitter 3 Measuring cable A0024179 General properties Temperature measurement In addition, a temperature probe is fitted in the sensor element to measure the temperature. 2 Endress+Hauser

Communication and data processing Communication with the transmitter Always connect digital sensors with Memosens technology to a transmitter with Memosens technology. Data transmission to a transmitter for analog sensors is not possible. Digital sensors can store measuring system data in the sensor. These include the following: Manufacturer data Serial number Order code Date of manufacture Calibration data Calibration date Cell constant Delta cell constant Number of calibrations Serial number of the transmitter used to perform the last calibration Operating data Temperature application range Conductivity application range Date of initial commissioning Maximum temperature value Hours of operation at high temperatures Reliability Dependability Serviceability Interference immunity Memosens technology digitizes the measured values in the sensor and transmits the data to the transmitter using a non-contact connection that is free from potential interference. The result: Automatic error message if sensor fails or connection between sensor and transmitter is interrupted Immediate error detection increases measuring point availability Easy handling Sensors with Memosens technology have an integrated electronics unit that stores calibration data and other information (e.g. total operating hours and operating hours under extreme measuring conditions). Once the sensor has been connected, the sensor data are transferred automatically to the transmitter and used to calculate the current measured value. As the calibration data are stored in the sensor, the sensor can be calibrated and adjusted independently of the measuring point. The result: Easy calibration in the measuring lab under optimum external conditions increases the quality of the calibration. Pre-calibrated sensors can be replaced quickly and easily, resulting in a dramatic increase in the availability of the measuring point. Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive maintenance is possible. The sensor history can be documented on external data carriers and in evaluation programs. Thus, the current application of the sensors can be made to depend on their previous history. With inductive transmission of the measured value using a non-contact connection, Memosens guarantees maximum process safety and offers the following benefits: All problems caused by moisture are eliminated. Plug-in connection free from corrosion Measured value distortion from moisture is not possible. The plug-in system can even be connected under water. The transmitter is galvanically decoupled from the medium. EMC safety is guaranteed by screening measures for the digital transmission of measured values. Endress+Hauser 3

GN/YE Memosens CLS82D Input Measured variable Measuring range Conductivity Temperature Conductivity 1 µs/cm to 500 ms/cm Temperature -5 to 120 C (23 to 248 F) Cell constant CLS82D k = 0.57 cm -1 Temperature compensation Pt1000 in accordance with DIN EN 60751 Power supply Electrical connection The sensor is connected to the transmitter via measuring cable CYK10. GY GND YE Com B GN WH BN Com A + 3 Measuring cable CYK10 A0023956 Performance characteristics Uncertainty of measurement Conductivity response time Temperature response time Maximum measured error Repeatability Each individual sensor is factory-measured in a solution of approx. 50 μs/cm using a reference measuring system traceable to NIST or PTB. The exact cell constant is entered into the quality certificate supplied. The uncertainty of measurement in determining the cell constant is 1.0 %. t 90 3 s t 90 25 s 4 % of reading 0.2% of reading 4 Endress+Hauser

Installation Symmetrical installation is recommended in order to guarantee linearity. The distance to the side walls and opposite walls must be at least 15 mm. 3 2 a 15 1 4 Minimum distance between pipe and end of the measuring cell 1 Sensor CLS82D 2 Pipe 3 Direction of flow A0024621 The ionic current in the liquid is affected by the walls in confined installation conditions. This effect is compensated by what is referred to as the installation factor. The installation factor can be entered in the transmitter for the measurement or the cell constant is corrected by multiplying by the installation factor. The value of the installation factor depends on the diameter and the conductivity of the pipe nozzle as well as the sensor's distance to the wall. The installation factor can be disregarded (f = 1.00) if the distance to the wall is sufficient (a > 15 mm). If the distance to the wall is smaller, the installation factor increases for electrically insulating pipes (f > 1) and decreases for electrically conductive pipes (f < 1). The installation factor can be determined using calibration solutions. a b 5 Schematic drawing of CLS82D in confined installation conditions a Wall distance b Gap width A0024626 Endress+Hauser 5

f 1.20 1.10 1 1.00 0.90 2 0.80 3 5 10 15 20 a [mm] 0.12 0.20 0.39 0.59 0.79 a [inch] 6 Relationship between installation factor f and wall distance a 1 Electrically insulating pipe wall 2 Electrically conductive pipe wall A0024615 f 1.06 1.04 1.02 1 1.00 0.98 0.96 2 0.94 3 5 10 15 20 b [mm] 0.12 0.20 0.39 0.59 0.79 b [inch] 7 Relationship between installation factor f and gap width b 1 Electrically insulating pipe wall 2 Electrically conductive pipe wall A0024616 Hygienic properties For a 3-A compliant installation, the following must be noted: After the instrument is installed its hygienic integrity shall be maintained. All process connections must be 3-A compliant. 6 Endress+Hauser

Environment Ambient temperature range Storage temperature -20 to +60 C (-4 to 140 F) -25 to +80 C (-13 to +176 F) Humidity 5 to 95 % Degree of protection IP 68 / NEMA type 6P (1 m water column, 25 C, 168 h) Process Process temperature Normal operation: -5 to 120 C (23 to 248 F) Sterilization (max. 45 min.): Max. 140 C (284 F) at 6 bar (87 psi) The maximum temperature for communication with the transmitter is 130 C (266 F). Process pressure (absolute) Pressure-temperature ratings 17 bar (247 psi) at 20 C (68 F) 9 bar (131 psi) at 120 C (248 F) p [psi] p [bar] absolute 247 180 131 120 87 60 15 17 16 12 9 8 6 4 A 1 5 20 60 100 120 140 23 60 120 180 240 248 284 T [ C] T [ F] 8 Pressure-temperature ratings A Can be sterilized for a short time (45 min.) A0024322 Endress+Hauser 7

Mechanical construction Design, dimensions 97 (3.82) Ø 84 (3.31) 97 (3.82) Ø 66 (2.60) Ø 11,9 (0.47) Ø 11,9 (0.47) 22 (0.87) 10 (0.39) 22 (0.87) 10 (0.39) A0024208 A0024209 9 Varivent N DN 40 - DN 125 10 Varivent F DN 25 157.3 (6.19) Ø 44,5 (1.75) Ø 44,5 (1.75) 137.8 (5.43) Ø 24,9 (0.98) 24.2 (1.0) 48.8 (1.92) Ø 11.9 (0.47) 71 (2.80) Ø 24,9 (0.98) 24.2 (1.0) 29.3 1.15 Ø 11.9 (0.47) 51.5 (2.02) A0028461 A0028462 11 DN 25 brown 12 DN 25 standard 8 Endress+Hauser

155 (6.10) Ø 66 (2.60) 155 (6.10) Ø 66 (2.60) 78 (3.07) 53 (2.09) G1" Ø 12,2 (0.48) 78 (3.07) 53 (2.09) NPT1" Ø 12,2 (0.48) 22 (0.87) 22 (0.87) A0024212 A0024213 13 G1" 14 NPT1" 103 (4.06) Ø 50,5 (1.99) Ø 11,9 (0.47) 25 (0.98) A0024214 A0024567 15 Clamp 1.5" 16 Clamp 2" 50 (1.97) 120 (4.72) 100.1 (3.94) Ø 11.9 (0.47) Ø 64 (2.52) Ø11,9 (0.47) 13 (0.51) 10 (0.39) A0024216 17 Pg 13.5 18 BioControl DN 25 All designs are supplied without a process seal. A0028463 Endress+Hauser 9

Weight Approx. 0.06 to 0.950 kg (0.13 to 2.09 lbs) depending on the version Materials in contact with the medium Sensor element: Platinum and ceramic Process connection: Stainless steel 1.4435 (AISI 316L) Only for CLS82D-**NA* 1) and CLS82D-**NB* 2) : Seal: EDPM 1) 1. Connection: DN25 standard 2) 2. Connection: DN25 brown Surface roughness R a < 0.38 µm Certificates and approvals Ex approval Ex approvals in conjunction with Liquiline CM42 transmitter ATEX II 1G Ex ia IIC T3/T4/T6 Ga CSA IS/NI Cl.1 Div.1&2 Grp.:A-D FM IS/NI Cl.1 Div.1&2 Grp.:A-D NEPSI Ex ia IIC T3/T4/T6 Ga TIIS Ex ib IIC T4 EAC Ex, OEx ia IIC T6/T4/T3 GaX All of the Ex versions listed here are identified by an orange-red ring on the plug-in head. EHEDG Hygienic process connections Clamp 1.5'', Clamp 2'', Varivent F, Varivent N and BioControl DN 25 are certified in accordance with EHEDG, Document 8 Validated as follows: Cleanability in accordance with EHEDG, Document 2 Sterilizability in accordance with EHEDG, Document 5 Bacteria-tightness in accordance with EHEDG, Document 7 The CLS82D with process connection Pg 13.5 in conjunction with Unifit CPA442 is certified in accordance with EHEDG, Document 8 Validated as follows: Cleanability in accordance with EHEDG, Document 2 FDA All materials in contact with the product are FDA-compliant. Quality certificate Stating the individual cell constant Pharma CoC Certificate of conformity for pharmaceutical requirements, confirms conformity with biological reactivity test USP Class VI, FDA material conformity, TSE-/BSE-free, surface roughness (must be ordered separately) Inspection certificate in acc. with EN10204-3.1 (optional) Available for all process connections (must be ordered separately) ASME BPE-2002 Manufactured in accordance with ASME criteria (American Society of Mechanical Engineers) 3-A Meets the requirements of 3-A Sanitary Standards. Regulation (EC) No. 1935/2004 Meets the requirements of Regulation (EC) No. 1935/2004 10 Endress+Hauser

Ordering information Product page Product Configurator www.endress.com/cls82d On the product page there is a "Configuration" button to the right of the product image. 1. Click this button. The Configurator opens in a separate window. 2. Select all the options to configure the device in line with your requirements. In this way, you receive a valid and complete order code for the device. 3. Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right above the selection window. For many products you also have the option of downloading CAD or 2D drawings of the selected product version. To do so, click the "CAD" tab and select the desired file type using dropdown lists. Endress+Hauser 11

Accessories Seals Only for CLS82D-**NA* 1) and CLS82D-**NB* 2) : EPDM seals for CLS82D (x 2; FDA USP Class VI); Order No. 71307106 FKM (Viton ) seals for CLS82D (x 2; FDA USP Class VI); Order No. 71307105 Silicone seals for CLS82D (x 2, FDA USP Class VI); Order No. 71307107 Welding socket G1¼ Ø28(1.1) G1¼ Ø28 (1.1) 50(1.97) / 30(1.18) 44(1.73) / 24.5(0.96) 59 (1.97) / 39 (1.18) 49(1.93) / 30(1.18) 44(1.73) / 24.5(0.96) 15 Ø25(0.98) Ø25(0.98) A0028344-EN A0028343-EN Safety welding socket DN25, straight, stainless steel 1.4435, L=30; Order No. 51508051 Only for CLS82D-**NA* 1) Safety welding socket DN25, angled, stainless steel 1.4435, L=30/40; Order No. 51508052 Only for CLS82D-**NA* 1) Safety welding socket DN25, straight, stainless steel 1.4435, L=50; Order No. 51508049 Only for CLS82D-**NB* 2) Safety welding socket DN25, angled, stainless steel 1.4435, L=50/60; Order No. 51508050 Only for CLS82D-**NB* 2) Existing standard welding sockets (for CPA440 / CPA441 / CPA460), order nos. 50005192 and 50028446, are also suitable for sensor CLS82D. Connection CYK10 Memosens data cable For digital sensors with Memosens technology Product Configurator on the product page: www.endress.com/cyk10 Technical Information TI00118C Memosens data cable CYK11 Extension cable for digital sensors with Memosens protocol Product Configurator on the product page: www.endress.com/cyk11 Technical Information TI00118C Calibration solutions Conductivity calibration solutions CLY11 Precision solutions referenced to SRM (Standard Reference Material) by NIST for qualified calibration of conductivity measuring systems in accordance with ISO 9000 CLY11-A, 74 μs/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081902 CLY11-B, 149.6 μs/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081903 CLY11-C, 1.406 ms/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081904 CLY11-D, 12.64 ms/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081905 CLY11-E, 107.00 ms/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081906 Technical Information TI00162C 1) Process connection: DN25 standard 2) Process connection: DN25 brown 12 Endress+Hauser

Calibration set Conducal CLY421 Conductivity calibration set (case) for ultrapure water applications Complete, factory-calibrated measuring system with certificate, traceable to SRM by NIST and PTB, for comparison measurement in ultrapure water up to max. 20 μs/cm Product Configurator on the product page: www.endress.com/cly421 Technical Information TI00496C/07/EN Recalibration The conductivity calibration set must be calibrated regularly onsite at the manufacturer's depending on the frequency of use and operating conditions. Recommended period: 1 year Endress+Hauser 13