The influence of bowtie filtration on cone-beam CT image quality

Similar documents
CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

QC Testing for Computed Tomography (CT) Scanner

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

Test Equipment for Radiology and CT Quality Control Contents

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Wide beam CT dosimetry. Elly Castellano

CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

PHYSICS QUESTIONNAIRE FORM

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

T h e P h a n t o m L a b o r a t o r y

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Nuclear Associates , , CT Head and Body Dose Phantom

PD233: Design of Biomedical Devices and Systems

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Data. microcat +SPECT

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Y11-DR Digital Radiography (DR) Image Quality

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction

Distributed source x-ray tube technology for tomosynthesis imaging

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Wide-Detector CT for TAVR Planning:

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Aim. Images for this section: Page 2 of 13

Quantitation of clinical feedback on image quality differences between two CT scanner models

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion

Tomophan TSP004 Manual

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Diagnostic X-Ray Shielding

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

Nuclear Associates and

Nuclear Associates

Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method

COMPUTED TOMOGRAPHY 1

Teaching Digital Radiography and Fluoroscopic Radiation Protection

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Amorphous Selenium Direct Radiography for Industrial Imaging

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

Fluence Field Modulated X-ray CT using Multiple Aperture Devices. Acknowledgements

I. Introduction.

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

diagnostic examination

A ghost story: Spatio-temporal response characteristics of an indirect-detection flat-panel imager

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Development of new dosimeter for measuring dose distribution in CT

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

MTF and NPS of single-shot dual-energy sandwich detectors

Nathan Childress, Ph.D., DABR

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution

Computerized Medical Imaging and Graphics

Testing a wavelet based noise reduction method using computersimulated

ISO Cube Daily QA Package

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

CyberKnife Iris Beam QA using Fluence Divergence

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

QC by the MPE in Belgium

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements

Electronic Brachytherapy Sources. Thomas W. Rusch

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

CT Basics: Image Quality Module 6

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Improvement of CT image quality with iterative reconstruction idose4

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

Overview of Safety Code 35

Transcription:

The influence of bowtie filtration on cone-beam CT image quality N. Mail Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario M5G 2M9, Canada and Ontario Cancer Institute, University Health Network, Toronto, Ontario M5G 2M9, Canada D. J. Moseley Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario M5G 2M9, Canada; Ontario Cancer Institute, University Health Network, Toronto, Ontario M5G 2M9, Canada; and Department of Radiation Oncology, University of Toronto, Toronto, Ontario M5G 2M9, Canada J. H. Siewerdsen and D. A. Jaffray a Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario M5G 2M9, Canada; Ontario Cancer Institute, University Health Network, Toronto, Ontario M5G 2M9, Canada; and Department of Radiation Oncology and Department of Medical Biophysics, University of Toronto, Toronto, Ontario M5G 2M9, Canada Received 20 February 2008; revised 10 October 2008; accepted for publication 11 October 2008; published 4 December 2008 The large variation of x-ray fluence at the detector in cone-beam CT CBCT poses a significant challenge to detectors limited dynamic range, resulting in the loss of skinline as well as reduction of CT number accuracy, contrast-to-noise ratio, and image uniformity. The authors investigate the performance of a bowtie filter implemented in a system for image-guided radiation therapy Elekta oncology system, XVI as a compensator for improved image quality through fluence modulation, reduction in x-ray scatter, and reduction in patient dose. Dose measurements with and without the bowtie filter were performed on a CTDI Dose and an empirical fit was made to calculate dose for any radial distance from the central axis of the. Regardless of patient size, shape, anatomical site, and field of view, the bowtie filter results in an overall improvement in CT number accuracy, image uniformity, low-contrast detectability, and imaging dose. The implemented bowtie filter offers a significant improvement in imaging performance and is compatible with the current clinical system for image-guided radiation therapy. 2009 American Association of Physicists in Medicine. DOI: 10.1118/1.3017470 Key words: bowtie filter, Elekta synergy CBCT system, image quality, dose I. INTRODUCTION Cone-beam computed tomography CBCT systems are employed in radiation therapy to provide information regarding daily target and normal tissue localization during the course of fractionated radiation treatment. CBCT allows radiation to be directed at tumors with greater accuracy and precision than was previously possible. However, improving CBCT image quality further will allow more precise dose delivery to the tumor. Precision in tumor localization and reduction in setup error is based on image quality of CBCT system. However, flat-panel detectors used in CBCT have limited dynamic range compared to CT. These limitations are exposed by the large variation of x-ray fluence at the detector across the imaged field-of-view. It is often the case that the techniques used overwhelm the signal range of the detector at the periphery of the patient, leading to a loss of information in projections and artifacts in reconstruction due to the truncation of anatomy. As a result, effects include the loss of skinline and reduction in CT number accuracy, contrast, and image uniformity. The purpose of this study is to investigate the relative advantages and limitations of a bowtie filter for improving image quality of a commercially used CBCT Elekta Synergy System. The implementation of a bowtie filter in CBCT offers the method of addressing the issue of x-ray flux variation across the detector. Previously, beam shaping x-ray filters or compensators have been used in computed tomography CT to modify the distribution of x-ray flux across the field of view. Compensator filters have been used in CT for beam hardening 1 5 and reduction 6 8 in dose to the patient. These compensators have been used in CBCT for scatter 9 11 and heel 12 compensation. Initially, compensators have been used to improve the detectability in film radiographs with the delivery of more uniform fluence at the film. 13 Bowtie filters potentially play a larger role in scatter, beam hardening, uniform fluence across the detector, and dose reduction. Quantitative investigations are performed on the Elekta Synergy radiotherapy image-guidance system. These investigations provides quantitative evidence of bowtie filter for improving skinline, image uniformity, CT number accuracy, contrast, and reduction in streak artifacts and patient dose. II. MATERIALS AND METHODS II.A. Cone-beam CT system Investigations of bowtie filter use in CBCT were performed on Elekta oncology system shown in Fig. 1 a. The x-ray source uses a rotating anode x-ray tube DunleeD604, IL with maximum potential of 150 kvp, 14 deg tungsten 22 Med. Phys. 36 1, January 2009 0094-2405/2009/36 1 /22/11/$25.00 2009 Am. Assoc. Phys. Med. 22

23 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 23 8>9 ;<: (!==$7>#!- 4: (7 8&9 8(9!#$%../ 1 23456 (7 ;<: (!==$7>#!- )*!+(%,#%- 3456 (7!&'%(#!&'%(# 3456 (7 )*!+(%,#%- ;<: (!==$7>#!-../ 1 23456 (7!#$% 4: (7 FIG. 1. a A photograph of the cone-beam CT Elekta Synergy CBCT system. b A schematic of the CBCT offset geometry at angle 0 deg and medium FOV. c Schematic of the CBCT offset geometry and medium FOV after 180 deg rotation. target, and 0.8 mm focal spot. The detector used was the RID1640-Al1 Perkin Elmer, Wiesbaden, Germany indirectdetection flat-panel imager with a 1024 1024 array of 0.4 mm 0.4 mm pixels and a 0.55 mm thick CsI:Tl x-ray converter. The geometry of the system was configured to have a source-to-isocenter distance of 100 cm and source-todetector distance of 153.6 cm. The x-ray tube and flat-panel imager were placed orthogonally to the treatment head and its EPID imaging device. The kv system shares a common axis of rotation with the MV treatment source. The kv beam is 425 425 mm 2 incident on the flat-panel detector. Images can be acquired with three different fields-of-view FOV ; small, medium, and large. The difference between the three is the edge to the kv central beam, which is 138 mm for the small FOV, 213 mm for the medium FOV, and 262 mm for the large field of view. During 360 deg rotations the system acquires approximately 650 projection images. The size of the flat-panel detector is 410 410 mm 2 and is located at a fixed source to detector distance, resulting in limited FOVs along the x and z direction. An offset scanning geometry was used, in which the imager was shifted laterally 10 cm and a corresponding asymmetric beam, defined by an M20 collimator, was utilized to scan a larger FOV. This offset geometry Fig. 1 b was used to provide the optimum image sets, in terms of quality and reconstructed FOV. The Feldkamp algorithm for 3D filtered (a) 35 (b) 85mm 117mm Thickness (mm) 25 15 5-5 -10 15 40 65 90 115 140 Length (mm) FIG. 2. a Photograph of the Elekta designed bowtie filter. b Bowtie thickness measured with needle gauge as a function of length.

24 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 24 1622!!# %#&'()*!+,*)-.!,)*# 1621 1673 1670 1604 * $ +$,''-.$,$$' /+01,$0.$,23 4+00,#5.$,30 6+#,-3.$,#1 $ = $ $ '# + 2π! + ' ) 1602 ( /0 1 0 2 3 4 51 50 52 53 89&*9: &*)9+(# ;%!< (#+)#% =(<> '$$ && ##$ && FIG. 3. a A schematic diagram of the CTDI body. b Percentage of dose reduction with bowtie filter is plotted as a function of radial distance from the center. Symbol represents measured data and solid line theoretical fit. c Circular Catphan with an irregular annulus human torso. back-projection algorithm was used for CBCT reconstruction, with a Hann reconstruction filter. The total reconstructed volume used in this work was 410 410 264 mm 3 with voxel size of 1 1 1 mm 3. II.B. Bowtie filter An aluminum bowtie filter Elekta, Filter Cassette Assembly, F1 was employed for all the tests described here. This F1 filter was inserted in the filter tray 30 cm from the x-ray source on an Elekta Synergy XVI unit. For the medium FOV, a cassette of M20 collimator was inserted between the x-ray tube and F1 filter shown in Fig. 2 a. There are various possible designs for bowtie filters implemented on Elekta Synergy XVI systems. Bowtie filters could potentially be optimized based on a variety of imaging tasks while accounting for tradeoffs between image quality and patient dose. The design of the bowtie filter used here is based on patient size, dose, nominal x-ray energy, and material. The bowtie filter used in this study Fig. 2 a has no modulation in the z direction. The filter used for this work was built based on the objective of achieving uniform fluence through a cylindrical water of diameter 300 mm measured at 120 kvp. The manufacturing material was aluminum of size 85 135 mm 2. The bowtie filter thickness measured with a needle gauge as a function of length is shown in Fig. 2 b. The filter was rigidly mounted in the cassette with a 1 mm thick acrylic covering protecting it from physical damage or scratching. II.C. Dosimetry measurements Dose measurements were made with and without a bowtie filter, such that the scanning geometry, size, beam quality, and mas/projections were the same for both cases. The shown in Fig. 3 a was a PMMA CTDI dosimetry with diameter of 320 mm and thickness of 150 mm RTI Electronics, Mölndal, Sweden. Two lowdensity polyethylene LDPE blocks 32 32 15 cm 3 were placed superior and inferior to the acrylic to simulate scattered dose. The center of the was placed at the machine isocenter position. This specially designed allows for accurate positioning of radiation detectors. A 0.6 cc farmer-type ion chamber NE 2571, Nuclear Associates and Fluke 3504 electrometer at atmospheric pressure 756.2 mmhg and room temperature 22 C was used for dose measurements. The dose was measured at five different radial positions of the, starting from the center to the skin. A total of five CBCT scans were obtained, with and without a bowtie filter at 120 kvp and 1.6 mas per projection 660 projections. The ratio of doses with and without bowtie was modeled for all the radial positions of the s center to periphery. This model contained four fitting coefficients and is very useful to compute the dose reduction attributed to the bowtie at any radial position of the CTDI. II.D. Scatter-to-primary ratio measurements Scatter-to-primary ratio was measured for imaging conditions with and without the bowtie filter using a beam-blocker method. 9 The SPR can be defined as SPR = S P, where S is the energy integrated signal of the scattered radiation measured as the average of a 10 10 pixel 2 area on 1

25 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 25 the FPI, and P is the signal from the primary radiation. The same CBCT system/geometry was used as discussed in Sec. II A. The SPR measurements were performed at the center and edge of the projection image for the circular and irregular posterior-anterior PA view s. A 10 10 4 mm 3 lead block was placed on the central axis before the on the side facing the source. The scatter accounting for shadowing a part of the by the blocker was not included for this SPR data. Twenty 20 images of the circular and irregular s were acquired with and without the bowtie filter at the tube voltage of 120 kvp and exposures 1.0 and 1.6 mas, respectively. The SPR measurements were performed on an irregular at equivalent patient dose techniques 0.1 mgy/exposure for with and without the bowtie filter. This required an exposure of 1.6 mas/projection bowtie filter as compared to 1.0 mas/projection without bowtie filter at 120 kvp. The same procedure was used for the SPR measurements at the s edge except the block was moved from the center to the edge of the s projection. II.E. CBCT image quality: Phantom study The Catphan 500 The Phantom Laboratory, Salem, NY was employed in these investigations. Experiments were performed with the Catphan 500 alone referred to as circular and with the Catphan 500 inserted in an irregular annulus referred to as irregular ; Fig. 3 c shaped to reflect a humanoid torso. Two types of configurations were used to evaluate overall CBCT image quality with and without a bowtie filter. The was positioned at the center of the imaging field of view with the help of room lasers. A photograph of the cone-beam CT system is shown in Fig. 1 a. Image acquisition proceeded with gantry rotation over 360 deg for all the cases during which approximately 660 planar images were acquired at the medium FOV. Images of the circular, with and without bowtie filter, were acquired under the same imaging conditions at 120 kvp and tube current 40 ma and exposure time 20 ms per projection. Similarly, images of the irregular with and without bowtie were acquired under the same imaging conditions at 120 kvp and tube current 40 ma and exposure time 40 ms, respectively. The CBCT reconstruction algorithm includes a scatter correction technique as described by Boellaard et al. 14 The irregular was scanned using a conventional helical CT scanner Discovery ST 16 slice, GE Healthcare, Milwaukee, WI at 120 kvp and 300 mas. The total number of CT reconstructed slices were 264 with each slice thickness of 1 mm and voxel size of 1 1 1 mm 3. The CT number for both CBCT and CT images was defined as CT # = object water / water 1000 HU, where object and water are the linear attenuation coefficients for the object and water, respectively. Quantitative comparison of image quality based on skinline recovery, CT number accuracy, CT number uniformity, spatial resolution, and contrast-to-noise ratio were evaluated using the circular tests performed in acquisitions with and without the bowtie filter. Imaging metrics were also performed on images generated on a conventional CT scanner. II.E.1. Uniformity The uniformity measurements were performed on the circular and irregular s. The circular contains a uniform, water-equivalent portion CTP486 with CT number HU within 10 and +10 HU at standard scanning protocols. The CBCT images were analyzed in Matlab. Five ROIs of size 10 10 mm 2 were selected in CBCT images of the one at the center and four at peripheral positions symmetrically around the center, each within the central axial plane. The spatial nonuniformity SNU was defined as SNU = VV max VV min 100, 2 1000 where VV max and VV min are the maximum and minimum mean voxel values, respectively, within each ROI. II.E.2. Skinline artifacts/reduction in CT number at skin zone In CBCT images, the reduction in CT number RCTN at the skin zone is attributed to the detector saturation near the object periphery. Measurements were performed on circular and irregular s. The reduction in CT number denoted CT # was measured using 3 ROIs at skin depths 5, 10, and 20 mm. The size of each ROI was 4 4 mm 2. The average reduction in CT RCTN was calculated as RCTN = HU CT HU CBCT 100, 3 HU CT where HU CT and HU CBCT are the mean voxel values for a given ROI measured in helical CT and CBCT images, respectively. II.E.3. CT number accuracy and linearity The CT number accuracy and linearity measurements were performed on circular and irregular s. The contains seven different cylindrical contrast inserts 12 mm diameter, CTP 404 with known electron densities and CT numbers. The mean HU value was measured for each contrast insert. The CBCT images were converted into polar coordinates for convenient analysis. A 2D ROI size was approximately 4 4 mm 2 covering approximately 16 voxels and was confined to within the contrast inserts. For conventional CT images, the 2D ROI size was 4 4 mm 2. The voxel values in helical CT images of the defined the true CT number. II.E.4. Contrast-to-noise ratio CNR measurements were performed using the circular and irregular s CTP 404 module, including polystyrene PS, 100 HU and low-density polyethylene LDPE, 35 HU inserts of 12 mm diameter. Contrast-tonoise ratio CNR measurements were performed on circular

26 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 26 TABLE I. The scatter, primary, and scatter-to-primary ratios behind the center and edge of the circular Cat- Phan500 using lead strip of size 10 10 3.5 mm 3. Scanning techniques Scatter at center Primary at center Scatter at periphery Primary at periphery Scatter-toprimary ratio at center Scatter-toprimary ratio at edge Bowtie 120 kvp/0.8 mas Without bowtie 120 kvp/1.6 mas 302 801 602 6594 0.377 0.02 0.092 0.02 630 1212 1301 12 730 0.519 0.03 0.102 0.02 and irregular s. For each individual, the imaging conditions including kvp, exposure time, and tube current were the same, with and without the bowtie filter. ROIs of size 4 4 mm 2 4 4 voxels within each insert were used to measure the mean and standard deviation noise in HU value. The mean HU value was measured for each contrast insert. The CNR signal difference divided by average noise values was calculated as HU LDPE HU PS CNR = 2 Noise LDPE + Noise PS, 4 where HU LDPE and HU PS are the mean voxel values in LDPE and PS, respectively, and Noise LDPE and Noise PS are the standard deviation in voxel values in LDPE and PS, respectively. II.E.5. Spatial resolution and modulation transfer function The MTF measurements were performed on circular and irregular s using the CTP528 module, which contains a radial high contrast spatial resolution bar pattern ranging from 1 through 21 line pair per cm. A bar pattern may be considered resolved if the bars can be perceived with some discernible spacing or lowering density among them. The radial design eliminates the possibility of streaking artifacts from other test objects. A circular pattern of pixels passing through the bar patterns was taken to obtain square-wave response function SWRF. The amplitude response at various spatial frequencies was analyzed between 1 and 21 lp/cm. The MTF of the system was determined by deconvolving out the SWRF. 15 II.F. Patient study CBCT image quality measurements including uniformity and skinline reconstruction were performed on ten gynecological patients. But only one patient is reported here because there were no significant differences from patient to patient. Images were acquired with and without a bowtie filter at 120 kvp and 1.6 mas 650 projections acquired across 360 deg. The same CBCT system geometry and procedure was used as explained in the study section. CBCT reconstructions with and without bowtie filter were analyzed quantitatively difference images and qualitatively examination by a radiation oncologist. Of specific interest in these studies were the gynecological patient contrast and the reduction in CT number cupping artifact from the skinline to the center of the image. III. RESULTS AND DISCUSSIONS III.A. Dose measurements for kv CBCT The uncertainty of the ion chamber measurements was determined standard deviation/mean value to be with in 0.55%. The dose reduction define in methods with bowtie filter as a function of radial distance from the center is shown in Fig. 3 b. It shows the achievable decrease in dose at the center and periphery of the body with the bowtie used in this study. Dose reductions by 25% and 43% were found at center and skin depth of 1 cm, respectively. It may be noted that the increase in dose reduction was consistent and followed a Lorentzian peak function LPF. A theoretical fit represented by a solid line is shown in Fig. 3 b. Dose reduction with a bowtie as a TABLE II. The scatter, primary, and scatter to primary ratios behind the center and edge of the irregular Cat-Irreg using lead strip of size 10 10 3.5 mm 3. Scanning techniques Scatter at center Primary at center Scatter at periphery Primary at periphery Scatter to primary ratio at center Scatter to primary ratio at edge Bowtie 120 kvp/1.6 mas Without bowtie 120 kvp/1.6 mas Without bowtie 120 kvp/1.0 mas 1011 2280 2539 20 427 0.443 0.03 0.124 0.01 2125 3221 4240 31 921 0.661 0.03 0.133 0.01 1328 2059 2625 1937 0.645 0.025 0.132 0.01

27 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 27! ) the unwanted signal at the object s periphery and hence improves the SPR behind the s center and prevents detector saturation at the skin zone. III.C. CBCT image quality: Phantom study 45 6789: & & % $ #! function of the radial distance, X, from the center of the dosimetry was calculated using an LPF function: D X = D 0 + (! '!' &' (' %' )*+,- /012,3 FIG. 4. Trans-axial images of uniform water portion of a circular Catphan acquired a without window level 130 to 150 HU and b with bowtie filter window level: 130 to 150 HU. c Profiles through the uniform water portion of the circular with and without bowtie; the profile position is indicated by a dotted line in image. 2AB 4 X C + B 2, where D 0 is the intercept and A, B, and C are the fitting coefficients. The values of the intercept and fitting coefficients are listed in Fig. 3 b. With the help of Eq. 5, the dose reduction with the bowtie filter can be calculated for any radial distance within the body. Also this empirical fit will assist in the design for an optimum bowtie filter. III.B. Scatter-to-primary ratio The SPR at the center of each is higher than the SPR measured at the edges. But the absolute scatter signal is higher at the edges of both s including circular and irregular. The SPR values for the circular and irregular s for with and without bowtie filter are listed in Tables I and II, respectively. It shows that the scatter signal at the s edge is higher than the primary signal at the center for the case without the bowtie filter. The absolute scatter signal at the edges of both s is almost two times less than without the bowtie filter. This high absolute scatter signal strongly affects the SPR behind the center of the and hence image quality. Also, the SPR values at the center and edge are higher for the irregular because of the size and excessive fluence at the edge that creates high scatter and affects the SPR at the center throughout the image. The bowtie filter shapes and modulates the beam to reduce 5 III.C.1. Uniformity and skinline Trans-axial CBCT images of a uniform water-equivalent portion of the circular acquired without and with bowtie filter are shown in Figs. 4 a and 4 b, respectively. Profiles through the images are shown in Fig. 4 c ; the profile position is indicated by a dotted line in the image. Image Fig. 4 a and its profile both show severe reduction in CT number near the skin zone, which reduces as a function of skin depth. This is due to the signal range of the detector saturation at the periphery of the, leading to a loss of information in projections that creates a skinline artifact in the reconstruction. Approximately 15% of average reduction in CT number was measured at 1 cm skin depth. The CBCT image with the bowtie filter Fig. 4 b and its profile Fig. 4 c demonstrate significant improvement in the recovery of CT number at the skin zone compared to the nominal case. Almost 100% of the CT number is restored at 1 cm skin depth. The bowtie filter improves image uniformity by i flattening of the fluence profile at the detector thereby reducing the dynamic range requirement and preventing detector saturation at the skinline and ii attenuates more fluence at the edge of the field which reduces the primary beam on the periphery of the, which, in turn reduces scatter throughout the image. The reduction in scatter throughout the image has the greatest advantage in the center of the image where the bowtie has minimally affected the primary fluence. The mean spatial non-uniformity, as defined in Eq. 2, was 2.1% and 9.8% for image acquired with and without a bowtie filter, respectively. The higher non-uniformity number as defined in Eq. 2 means poorer uniformity. Trans-axial CBCT images of irregular acquired without and with bowtie filter are shown in Figs. 5 a and 5 b, respectively, at the same window/level. The difference image image without bowtie subtracted from the image acquired with bowtie is shown in Fig. 5 c. Profiles through the reconstructed slices of CBCT images are compared with helical CT in Fig. 5 d, with the bowtie filter case providing an improvement in edge definition. The images and profiles show a severe reduction of CT number near the skin zone in the nominal CBCT image. Since the irregular is larger in size than the circular, a higher exposure is required for sufficient signal behind the center of the object, resulting in increased pixel saturation near the periphery, which is the main source of CT number reduction at the skinline. Even with the bowtie filter case, there is still some missing skin on the right shoulder of the image profile in Fig. 5 d. This is due to the detector lag and ghosting effects. As the object is irregularly shaped, the projected location of skinline relative to previous projections causes an accumulated lag signal that degrades the reconstruction and results in reduction in CT number in the skinline zone. The gantry rotates clockwise around an irregular, which

28 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 28 456789: ( ( $ '!& ),* )-* 45 )(*! '!' &' (' )>+*<*>/ 711: 45 6789: F45G 7H: &!! & ( )%* )+* 45 && &% &$ &# ( (& )>+*<*>/ 7112: % ).* 2 @1 +I*/ E,C<= (&!@1 +I*/ E,C<= &% & @1 +I*/ E,C<=!$ &!!!! # 5,@=/*A0,+ BCC-*,E FIG. 5. Trans-axial CBCT images of irregular Cat-Irreg a without window level: 350 to 500 HU and b with bowtie filter window level: 350 to 500 HU. c The difference of image b and a. d Profiles through the uniform water portion of the irregular. e The right shoulder of d is magnified to see the missing skinline more clearly. f Reduction in CT number RCTN at several skin depths for with and without bowtie filter. ()* (+* 45 6 789:! (-* 2>;<*, 45!!! & & ( ( )*+,- /012,3 (,* A33>3B 45 4?45 789:!& # % % (%* D*<=>0< 2>;<*, #!! $! % ' >E,B- 45 /012,3 FIG. 6. Trans-axial images of circular with several contrast inserts acquired a without window level 800 to 500 HU and b with bowtie filter window level 800 to 500 HU. c The same image shown in a, but transformed into polar coordinate window level 800 to 500 HU. d Profiles through the image c with several inserts are compared with CT and without bowtie filter. e The difference between the measured CT and CBCT number plotted as a function of Ideal CT number for several inserts. The empty circle and filled square symbols represent CT number error with and without bowtie. A gray solid line represents a linear fit to bowtie data.

29 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 29 *! ' &%$ &$$ #%$ #$$!! ( ) 4,B+03,E 45 6789:2 &!% # & %! 45 +@B/ B*/,B3 745 +@B/: B*/,B3 7: B*/,B3 7:! % & #!% & >E,B- 45 6789: FIG. 7. Trans-axial CBCT images of irregular Cat Irreg with several inserts, a without bowtie window level 350 to 500 HU and b with bowtie filter window level 350 to 500 HU. c The difference of image b and a. d The same image in a but converted into polar coordinates window level 700 to 300 HU. e The same image in b but converted into polar coordinates to make it simple for analysis such as CT number accuracy and streaking artifacts window level 700 to 300 HU. f The measured CT number plotted as a function of ideal CT number for several inserts. + causes lag on the right side of the image. Mail et al. 16 quantitatively explained this lag effect in CBCT images of irregular acquired with and without bowtie filter. Even without a bowtie filter, the reduction in CT number at the skinline is worse on the right shoulder than the left in Fig. 5 d, due to the lag and ghosting. 16 This clearly explains the bowtie case where the reduction in CT number at the skinline on the right shoulder in Fig. 5 d is due to lag and ghosting 16 not due to the compatibility of the bowtie filter for an irregular shaped. The shoulder on the right hand side of Fig. 5 d is magnified in Fig. 5 e to illustrate the reduced signal at the periphery. The mean reduction in the CT number at the skin region, as defined in Eq. 3, is quantitatively shown in Fig. 5 f for three different skin depths for both cases. The mean signal losses for skin depths of 5, 10, and 20 mm were found 35.1%, 25.15%, and 20.12%, respectively, in cases without bowtie filter images. These numbers in bowtie cases were measured 13.8%, 12.5%, and 11%. The spatial non-uniformities of the irregular without and with bowtie filter Figs. 5 a and 5 b were calculated to be 4.1% and 5.7%, respectively. Profiles in Fig. 5 d show that the CT and CBCT images with the bowtie filter have similar uniformity within the circular region. III.C.2. CT number accuracy and linearity The central slice images of a circular without and with bowtie filter are shown in Figs. 6 a and 6 b, respectively. For convenient analysis, the image in Fig. 6 b was transformed into polar coordinate in Fig. 6 c. A profile through several inserts, indicated by a dotted line in image, is shown in Fig. 6 d. The dotted, solid black, and gray solid lines represent CT and CBCT with and without bowtie filter image profiles, respectively. This signal profile shows that CBCT image acquired with a bowtie filter is comparable to helical CT. The comparison of CT number accuracy and linearity for circular, without and with the bowtie filter, is shown in Fig. 6 e, where the difference of measured CT number to CBCT is plotted versus the ideal CT number. The CBCT with bowtie filter demonstrated a clear improvement in CT number accuracy and an increase in image contrast compared to the nominal cases. For the nominal case, inaccuracy in CT number is related to the assumption that the detected signal is all primary fluence. Scattered fluence clearly violates this assumption. Reducing scatter will lead to more accurate CT numbers by allowing accurate inversion of TABLE III. Comparison of the CT number accuracy and linearity using three different techniques in terms of slope, intercept at ideal CT# water, and R 2 values. Techniques Slope R 2 Intercept CT 0.98 0.995 4 Bowtie 0.85 0.99 43 Without bowtie 0.71 0.98 3

30 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 30 Beer s law. It should be noted that other invalid assumptions e.g., beam hardening can lead to inaccuracies in CT number as well. The improvement in CT number accuracy with the bowtie filter is primarily attributed to the scatter reduction with improved beam-hardening and non-uniformity also contributing to improved CT number accuracy throughout the image and particularly near the s periphery. This reduction in scatter reduces the scatter to primary ratio and hence improves CT number accuracy. The CT number accuracies with and without a bowtie filter were found to be 21 and 68 HU, respectively. Central slice images of the irregular with and without a bowtie filter are shown in Figs. 7 a and 7 b, respectively. These images were then transformed into polar coordinates to make the image analysis simple. The transformed polar coordinate images are shown in Figs. 7 d and 7 e. In Fig. 7 f, the measured CT number for Catphan within irregular annulus is plotted versus the ideal CT number. The measurements performed with a bowtie filter showed greater linear CT number accuracy than nominal. The mean CT number obtained from a conventional CT image showed reasonably good agreement with the expected CT numbers. Detailed comparisons of the CT number accuracy and linearity in terms of slope, R 2 values, and intercepts are shown in Table III. The slope of the fitting line and intercept values are reasonably improved. The CT number accuracy for the helical CT and CBCT acquired with a bowtie is higher than the nominal CBCT images. This is because of the influence of scatter radiation in nominal CBCT, which is more severe as compared to that of a fan beam geometry. In general, x-ray scatter reduces image contrast, increases image noise, and may introduce reconstruction error into CBCT. III.C.3. Contrast-to-noise ratio and streak artifacts CNR measurements were performed on Catphan and irregular s contrast inserts, LDPE and Polystyrene. For Catphan, the CNR, as defined in Eq. 4, was measured as 4.51 0.43 and 3.58 0.35, with and without a bowtie filter, respectively an improvement by a factor of 1.3 at fixed mas. The improvement is likely due primarily to reduced scatter with improved beam-hardening and nonuniformity also contributing to improved CNR. The CNR values for the irregular were found to be 4.22 0.41 and 2.24 0.32, with and without bowtie cases, respectively at equivalent mas. The improvement in CNR is found to be much greater for the irregular than circular. A primarily possible approach is that for the without bowtie case, the amount of scatter in the irregular is higher than in the circular, particularly at the s peripheries. The CNR in the nominal case is significantly reduced for the irregular than for the circular, while this reduction is very small in the case of the bowtie filter. This is because of the scatter compensation. Comparing images acquired at equivalent patient dose MTF 1 0.8 0.6 0.4 0.2 0 0 1 2 3 4 5 6 7 8 9 10 11 Line pairs (cm -1 ) Bowtie without bowtie FIG. 8. Measured MTF performed on irregular s using the CTP528 module Catphan500 plotted as a function of line pairs with and without bowtie filter. 0.1 mgy/exposure, the comparison of CNR improves even further, giving a CNR of 4.22 0.41 for the bowtie, compared to 1.61 0.3 without a bowtie. The central slices of irregular converted into polar coordinates with and without the bowtie filter are shown in Figs. 7 e and 7 d, respectively. The streaking artifacts are more visible around the high-density inserts in the polar coordinate image acquired without the bowtie filter. These streaking artifacts appear due to scatter and beam hardening. 17 Beam hardening and scatter effects are worse because of the large size and irregular shape of the and the presence of high contrast objects. The image in Fig. 7 e acquired with a bowtie filter is almost free of streaking artifacts. The scatter compensation reduces streaking artifacts and improves CNR. The bowtie filter reduces beam hardening effects by minimizing the range of variation in x-ray energies presented to the detector. For example, considering the PA projection of the irregular, the mean energy of the x-ray beam is 75 kev at the center of the detector path length 320 mm acrylic and 64 kev near the periphery path length 100 mm acrylic. With a bowtie filter, however, the range in energy is reduced to 76 kev at the center and 76 kev at the periphery. Beam-hardening artifacts are therefore expected to be mitigated significantly by the bowtie filter. III.C.4. Spatial resolution and modulation transfer function The spatial resolutions for 50% MTF of the circular s images acquired with and without a bowtie filter were measured as 2.3 0.11 and 2.2 0.1 lp/cm, respectively. The MTF is essentially equivalent with or without a bowtie filter, with the curves of Fig. 8 showing consistent shape between the two cases. For the irregular, the spatial resolutions for 50% MTF were measured as 2.1 0.08 and 2.0 0.06 lp/cm with and without a bowtie, respectively. The measured MTFs for with and without the bowtie filter have the same values within the error bar.

31 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 31!! ) ) $ ( 45 6789: & & $ &% % <%! %!!% & &% ( (% )>)*<*>/ 711:! &<$ &'$ (!$ (($ )>)*<*>/ 711: FIG. 9. CBCT images of gynecological patient window level 500 to 500 HU a without bowtie and b with bowtie window level 500 to 500 HU. Profiles through the region of interest, indicated by a solid line. d A portion of c, indicated by a rectangular box, is magnified to illustrate signal loss more clearly at skinline zone. III.D. Patient study CBCT images of a gynecological patient acquired without and with a bowtie filter are shown with the same window level in Figs. 9 a and 9 b, respectively. Profiles through the images are shown in Fig. 9 c. Moderate improvements in uniformity and skinline reconstruction are seen with the bowtie filter. Improvement in skinline reconstruction is clearly shown in Fig. 9 d, which is the magnified portion indicated on the right of Fig. 9 c. The black spots in the images are a brachytherapy applicator used as a fiducial marker. The image in Fig. 9 a demonstrates less clarity due to high scatter-to-primary ratio while the image in Fig. 9 b demonstrates clarity due to reduction in scatter-to-primary with the bowtie filter. IV. DISCUSSIONS A fixed, shaped filter has been used to selectively attenuate x rays at the periphery of the exposed field-of-view to compensate for the patient profile. The filter reduces the dynamic range demands on the detector during a CBCT scan. The resulting flux on the detector is relatively uniform across the detector and does not result in saturation in regions beyond the skinline. Further, the reduction of the excessive fluence near the skin zone with a bowtie filter reduces the absolute scatter signal at the object s edge. This reduction in the absolute scatter signal improves the SPR behind the center of the and hence image quality. The influence of the bowtie filter on both dose and image quality was evaluated. The application of the bowtie filter for circular and irregular s resulted in an improvement in image quality and reduction/alteration in the dose distribution within the patient. Figure 3 b demonstrates that the application of the bowtie filter results in a reduction in dose by 43% at the skinline and 26% at the center. Assessment of contrast-to-noise performance demonstrated an improvement when the bowtie filter is employed. This was specifically observed in the large sized irregular where scatter is increased. Overall, the bowtie improves CT number accuracy and image uniformity. This improvement is attributed to the scatter reduction, beam hardening reduction, and flatness of fluence at the detector. The selection/design of a bowtie filter is challenging and will always be less than ideal. It is not likely to be optimum given that a subject being imaged is significantly less than uniform, not exactly cylindrical in shape, and not necessarily centrally located in the x-ray beam. In such cases, it is possible for one or more disjointed regions of saturation to occur or conversely to over-filter the x-ray flux and unnecessarily create regions of very low signal. Regardless of compromises associated with patient size, shape, anatomical site, and field of view, the use of a simple bowtie filter results in an overall improvement in the quality of CBCT images. V. CONCLUSION The implemented bowtie filter shows reduction in scatterto-primary ratio for the circular and irregular s. The bowtie filter shows an improvement in image quality including uniformity, CT number accuracy, contrast-to-noise ratio, and skinline reconstruction of the circular and irregular s. The implementation of a bowtie filer also reduced streaking artifacts in CBCT images of an irregular compared to nominal. The implemented bowtie filter demonstrated an improvement in skinline reconstruction and uniformity in images of gynecological patients. This compensator

32 Mail et al.: The influence of bowtie filtration on cone-beam CT image quality 32 is static and makes many compromises for anatomical imaging site, patient size, and imaged field of view. The ideal compensator would optimize the fluence profile to account for numerous properties of the patient and imaging system. The investigated performance of the bowtie filter on image quality of CBCT will create confidence in clinical implementation of bowtie filters. a Author to whom correspondence should be addressed. Address for correspondence: Radiation Physics, Radiation Medicine Program, Princess Margaret Hospital, 610 University Avenue, Toronto, Ontario M5G 2M9, Canada. Telephone: 416-946-4501 X5384; Fax: 416-946-6566; Electronic mail: david.jaffray@rmp.uhn.on.ca 1 R. Ning, X. Tang, D. Conover, and R. Yu, Flat panel detector-based cone orbit: Preliminary study, Med. Phys. 30, 1694 1705 2003. 2 E. Seeram, Computed Tomography: Physical Principles, Clinical Applications & Quality Control Saunders, Philadelphia, 1994. 3 R. Ning, University of Rochester, Apparatus and method for x-ray scatter reduction and correction for fan beam CT and cone beam volume CT, U.S. Patent No. 6,618,466 2003. 4 H. H. Barrett and W. Swindell, Radiological Imaging: The Theory of Image Formation, detection, and Processing Academic, New York, 1981. 5 G. X. Ding, D. M. Duggan, and C. W. Coffey, Characteristics of kilovoltage x-ray beams used for cone-beam computed tomography in radiation therapy, Phys. Med. Biol. 52, 1595 1615 2007. 6 R. G. Walters and R. W. Carlson, Technicare Corporation, Computerized tomographic scanner with shaped radiation filter, 1981. U.S. Patent No. 4288695. 7 J. E. Tkaczyk, Y. Du, D. Walter, X. Wu, J. Li, and T. Toth, Simulation of CT dose and contrast-to-noise as a function of bowtie shape, Proc. SPIE 5368, 403 410 2004. 8 S. Itoh, S. Koyama, M. Ikeda, M. Ozaki, A. Sawaki, S. Iwano, and T. Ishigaki, Further reduction of radiation dose in helical CT for lung cancer screening using small tube current and a newly designed filter, J. Thorac. Imaging 16, 81 88 2001. 9 S. A. Graham, D. J. Moseley, J. H. Siewerdsen, and D. A. Jaffray, Compensator for dose and scatter management in cone-beam computed tomography, Med. Phys. 34 7, 2691 2703 2007. 10 G. H. Glover, Compton scatter effects in CT reconstructions, Med. Phys. 9, 860 867 1982. 11 J. Hsieh, GE Medical Systems Global Technology Co., LLC, Method and apparatus for optimizing dosage to scan subject, 2003, U.S. Patent No. 6,647,095. 12 C. D. Smith, X-ray filter, 1940, U.S. Patent No. 2,216,326. 13 S. Mori, M. Endo, K. Nishizawa, M. Ohno, H. Miyazaki, K. Tsujita, and Y. Saito, Prototype heel effect compensation filter cone-beam CT, Phys. Med. Biol. 50, N359 N370 2005. 14 R. Boellaard, M. Van Herk, H. Uiterwaal, and B. Mijnheer, Twodimensional exit dosimetry using a liquid-filled electronic portal imaging device and a convolution model, Radiother. Oncol. 44, 149 157 1997. 15 H. E. Johns and J. R. Cunningham, The Physics of Radiology, 4th ed. Thomas, Springfield, IL, 1983. 16 N. Mail, D. J. Moseley, J. H. Siewerdsen, and D. A. Jaffray, An empirical method for lag correction in cone-beam ct, Med. Phys. 35 11, 5187 5196 2008. 17 J. H. Siewerdsen and D. A. Jaffray, Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter, Med. Phys. 28, 2, 220 231 2001.