Raman Spectroscopy of TiO2

Similar documents
Renishaw InVia Raman microscope

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

RENISHAW INVIA RAMAN SPECTROMETER

WITec Alpha 300R Quick Operation Summary October 2018

Parameter Selection and Spectral Optimization Using the RamanStation 400

Improving the Collection Efficiency of Raman Scattering

STEM Spectrum Imaging Tutorial

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter.

Instructions for the Experiment

Fast Laser Raman Microscope RAMAN

EDUCATION EXPERIMENT. Determining Color Difference (Delta E) Introduction

Spectroscopy Application: Using Raman Spectroscopy to Detect Art Forgeries Detecting art forgeries using Raman spectroscopy

Fast Laser Raman Microscope RAMAN

Ocean Optics R-2000 Raman Spectrometer Setup and Operating Instructions Arlen Viste and Deanna Donohoue April 2000 Update 2003, DEW

AP Chemistry Cell Phone Spectroscopy Lab Adopted from Alexander Scheeline Department of Chemistry University of Illinois at Urbana-Champaign

Ground Truth for Calibrating Optical Imagery to Reflectance

Swept-Field User Guide

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE

Leica SPEII confocal microscope. Short Manual

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

Construction Manual of Merlin spectrometer

VS7550 VUV/UV Mini Spectrograph Operating Manual

Integrated into Nanowire Waveguides

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

Practical work no. 3: Confocal Live Cell Microscopy

Shimadzu RF-5301 Fluorimeter operation guide for students

Things to check before start-up.

Experiment P-40 Colors of Light

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

The New Techpap NIR spectroscopy for Recycled Paper Bales Inspection

An Activity in Computed Tomography

IBIL setup operation manual for SynerJY software version

An Activity in Computed Tomography

Single Photon Interference Katelynn Sharma and Garrett West University of Rochester, Institute of Optics, 275 Hutchison Rd. Rochester, NY 14627

TriVista. Universal Raman Solution

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Renishaw InVia Quick Operation Summary October 2018

ScanArray Overview. Principle of Operation. Instrument Components

Cell Biology and Bioimaging Core

NAME SECTION PERFORMANCE TASK # 3. Part I. Qualitative Relationships

Zeiss 780 Training Notes

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Composite Thermal Damage Measurement with Handheld FTIR. April 9, 2013 Brian D. Flinn, Ashley Tracey, and Tucker Howie University of Washington

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

Supplemental Method Information Zeiss LSM710

Approachable Raman Solutions The Shortest Path from Problem to Answer

University of Wisconsin Chemistry 524 Spectroscopic Components *

Material analysis by infrared mapping: A case study using a multilayer

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

A Narrow-Band Tunable Diode Laser System with Grating Feedback

Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,

PHYS 1112L - Introductory Physics Laboratory II

It s Our Business to be EXACT

Leica TCS SP8 Quick Start Guide

Mira M-3 Handheld Raman Spectrometer. On-site verification of materials in seconds

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club

Zeiss 880 Training Notes Zen 2.3

Standard Operating Procedure

Terahertz spectroscopy measurements

X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P )

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

Physics 262. Lab #1: Lock-In Amplifier. John Yamrick

ANALYSIS OF ELECTRON CURRENT INSTABILITY IN E-BEAM WRITER. Jan BOK, Miroslav HORÁČEK, Stanislav KRÁL, Vladimír KOLAŘÍK, František MATĚJKA

BV NNET User manual. V0.2 (Draft) Rémi Lecerf, Marie Weiss

Multifluorescence The Crosstalk Problem and Its Solution

LSM 510 Training Notes

Title: Leica SP5 Confocal User Manual

Supplementary Materials for

Pre-Lab 10. Which plan or plans would work? Explain. Which plan is most efficient in regard to light power with the correct polarization? Explain.

PHYS General Physics II Lab Diffraction Grating

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Lab 3: Low-Speed Delta Wing

Technische Universität Chemnitz Physikalische Chemie. Laser Safety

AN INITIAL investigation into the effects of proton irradiation

INSTRUMENTATION BREADBOARDING (VERSION 1.3)

Program for UV Intercomparison 2014 in Davos:

Supplementary Materials for

Ultrafast Surface-Enhanced Raman Probing of the Role of Hot Electrons in Plasmon-Driven Chemistry. Supporting Information

Filter & Spectrometer Electron Optics

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Math Labs. Activity 1: Rectangles and Rectangular Prisms Using Coordinates. Procedure

Camera Requirements For Precision Agriculture

Confocal Raman Microscopy (WITec Alpha 300R)

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Imaging Fourier transform spectrometer

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

PhysicsAndMathsTutor.com 1

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

MUSKY: Multispectral UV Sky camera. Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM

Transcription:

Raman Spectroscopy of TiO2 Experiment #4 Characterization of Materials (96.445/545) Meg Noah Meg Noah 1 of 8 10/21/2010

Objective The purpose of this laboratory is to identify two different phases (rutile and anatase) of TiO2 by raman spectroscopy. A secondary objective is to learn to use the Raman spectroscopy equipment and how to optimize the geometry of the data acquisition and the parameters such as integration time and laser power. Experimental Equipment The experimental equipment used for this lab included: Powdered TiO2 samples Number 8 and Number Raman Systems, Inc. (RSI) RSL Plus R-3000 series and RSI Scan software Raman Systems, Inc. (RSI) RSI-Indent software. The XPowder software on the laboratory computer (to use for finding Raman Line position and ). The RSL we used had a 785 nm laser light, which is just beyond the red visible region. A red LED on the probe turns on when the laser is operating so that the beam can be focused on a sample. Safety goggles should be worn when this system is operated, but we didn t. Dark samples and samples optically opaque in the near infrared will heat up quickly with this laser focus. So, the samples are being turned. The actual spot size of the laser beam is about 100 microns. The typical low power setting for the RSL Plus is 125 mw, and the typical high power setting is 250 mw; for the R-3000 series the low setting is 90 mw and the high setting is 290 mw. The spectrometer is supposed to be calibrated, and during typical operation this calibration is verified by measuring the peaks of a Teflon cap using the verification option of the software. We didn t have the Teflon (polytetrafluoroethylene) cap, and we didn t verify the calibration. Procedure 1. Sample Preparation The samples were already prepared by the instructor, and labeled with a number that didn t identify their phase. We know they are TiO2. 2. Loading the Raman specimen holder The powder samples were loaded into the turning specimen holder. Scotch tape was used to keep the specimen holder from walking around the table during acquisition. We measured the rotation rate to be about 1 rotation in 5 seconds. 3. Acquire exposures. First we tried Number 8, but didn t get quality data. So then we did Number 9. After acquiring data for Number 9, we tried Number 8 again. In both cases, the XRF data identified Ti, so we Meg Noah 2 of 8 10/21/2010

concluded that the compounds were both TiO 2. The system is activated by first turning it on and then turning a key on the laser power control box. The software on an accompanying laptop provides the user interface for operating the system and managing the data. We defined the low power as being 12 O Clock setting on the power knob, medium as being 5 O Clock, and high as being 10 O Clock. To display the control parameters, press the arrow key. We did this and tried exposures of 3, 5, and 10 seconds for sample 8, and 10 and 30 seconds for sample 9. To take a measurement, the take measurement button is clicked: Then, the status bar on the software tells what stage of the acquisition is current: acquiring background, or collecting data, or saving data. The plot is displayed afterword, and automatically saved with sequential filenames. 4. Save Exposures and Data We saved the data to disk, converted the csv (comma separated value text format) for use in XPowder to find the Raman lines and values. Using this hack in XPowder, the offset of the plots is manually set from 3.01 (default) to 0.01. The line positions which XPowder interprets as a 2 value, are calibrated by multiplying by 10 and adding 200 to get to wavenumber ( ) values. The positions of the center of the line, the right and left side half peak positions are found interactively and calibrated to. Meg Noah 3 of 8 10/21/2010

Results and Discussion Determination of Optimal Settings Trial and error seems to be the only method for finding the optimal settings. To get good signal to noise statistics, typically longer integration times are used. But with this system, there is no way to set the gain values. Figure 1 shows the saturated values that are acquired when long integration times are used; here sample 8 was measured with a 20 second integration time at the lowest power setting. To find the optimal integration time, we simply tried different time settings and visually assessed the quality of the scan. For sample 9, Figure 2 shows the difference of a quality spectrum (blue) when integration time was 30 seconds, and one with too low of a signal to noise (red) when 10 seconds was used both for the lowest power laser setting. The optimal integration time was significantly different for sample 9 and sample 8. But it did not seem to depend on the power setting of the laser, although the highest power setting showed the most variation in the data sets from set to set (Figure 3). Good statistics can be acquired by averaging multiple runs, and using low or medium power settings. The next section shows that the did not depend very much on the settings. Figure 4 shows the bizarre behavior of some of the CCD pixels. Occasionally the data are fine, but more often than not, they are either missing or spiked. This is probably telegraph noise (Hung, 1990). Figure 1: Saturated Values (Sample 8, 20 second integration time, lowest power setting) Figure 2: Sample 9 with 30s (good) and 10s (too low) integration times Meg Noah 4 of 8 10/21/2010

Figure 3: Stacked Acquisitions Show Signal Variations (Sample 9, 30 second, highest power) Figure 4: Telegraph Noise Determination of Optimal Settings Using a hacked procedure for XPowder, the Raman lines for Sample 8 and Sample 9 were determined and are in Table 1 and Table 2. The co-added spectra are provided in Figure 5 and Figure 6. Table 1: Sample 8 Raman Line Measurements Number of Runs Power Setting Exposure time Line Center Line Center 3 Medium 5 s 443.8 42 604.9 40 3 Low 5 s 443.5 40 605.7 40 3 High 3 s 443.0 40 605.6 40 Table 2: Sample 9 Raman Line Measurements Number of Runs Power Setting Exposure time Line Center Line Center Line Center 4 Low 30 s 397.5 45 514.6 35 635 42 3 Medium 30 s 395.9 38 514.6 35 633.3 41 3 High 30 s 397.5 43 514.6 35 633.3 40 Meg Noah 5 of 8 10/21/2010

Figure 5: Sample 8 High Power (Bottom), Medium Power (Middle), Low Power (Top) Meg Noah 6 of 8 10/21/2010

Figure 6: Sample 9 High Power (Bottom), Medium Power (Middle), Low Power (Top) Meg Noah 7 of 8 10/21/2010

Conclusions The optimization of RSI Raman equipment settings has already been discussed. The values found for the Raman lines of Sample 8 and Sample 9 were compared to values from the literature. Table 3 shows the values reported by Ma (2007). We can easily conclude that Sample 8 is the rutile phase and sample 9 is the anatase phase. The anatase values are well within experimental error. All three measureable Raman vibration modes (B1g, A1g, and Eg) were active. The other Eg modes are below the measurement capabilities of our system. The multiphonon 235.5 cm-1 mode of the rutile sample was evident in the spectra, but the value and width were not extrapolated. The 825.5 cm-1 was not evident, not even in the saturated exposure (Figure 1). Ma used a 514 nm laser (20 mw) to do the measurements in Table 3. It is likely that our system is too far into the infrared to see this mode. The Eg and A1g modes were measured. Table 3: Raman Shifts Measured by Ma (2007).. References [1] Operation Manual For RSLplus R-3000 series RSIScan software, www.ramansystems.com [2] Hung, K. K., P. K. Ko, Chenming Hu, and Yiu Chung Cheng, Random Telegraph Noise of Deep- Submicrometer MOSFET's, IEEE Electron Device Letters, Vol 11, No. 2, February 1990. [3] Ma H.L., J.Y. Yang, Y. Dai, Y.B. Zhang, B. Lu, G.H. Ma, "Raman study of phase transformation of TiO2 rutile single crystal irradiated by infrared femtosecond laser," Applied Surface Science 253, 2007 Meg Noah 8 of 8 10/21/2010