Initial setup and subsequent temporal position monitoring using implanted RF transponders

Similar documents
8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

ISO Cube Daily QA Package

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

Electromagnetic Tracking for Medical Imaging

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA

THE STORAGE RING CONTROL NETWORK OF NSLS-II

IQM Detector Characteristics: Signal reproducibility

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA

Laser Systems for Patient Alignment

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

New Long Stroke Vibration Shaker Design using Linear Motor Technology

Distributed source x-ray tube technology for tomosynthesis imaging

CyberKnife Iris Beam QA using Fluence Divergence

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

QC Testing for Computed Tomography (CT) Scanner

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

EnSite Precision Cardiac Mapping System ACCURACY OF THE ENSITE NAVX NAVIGATION AND VISUALIZATION TECHNOLOGY, SENSOR ENABLED

Nathan Childress, Ph.D., DABR

COMPUTED TOMOGRAPHY 1

Mobius3D. Software based IMRT QA

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom

Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion

END-TO-END TESTING OF THE LEKSELL GAMMA KNIFE ICON SYSTEM

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

DIT-5200L. Non-Contact Displacement Differential Measuring System User s Manual

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Beam Commissioning and Annual QA Phantoms, Detectors & Accessories. iba-dosimetry.com

Physics Requirements Document Document Title: SCRF 1.3 GHz Cryomodule Document Number: LCLSII-4.1-PR-0146-R0 Page 1 of 7

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it

Orbit Stability Challenges for Storage Rings. Glenn Decker Advanced Photon Source Beam Diagnostics March 8, 2012

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

HITACHI Proton Therapy System with Spot Scanning

MOVING LASER SYSTEMS IN RT

SOLEIL Libera Performance

Stretched Wire Test Setup 1)

Nuclear Associates , , CT Head and Body Dose Phantom

Cavity BPMs for the NLC

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung

An Activity in Computed Tomography

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program

The Henryk Niewodniczański INSTITUTE OF NUCLEAR PHYSICS Polish Academy of Sciences ul. Radzikowskiego 152, Kraków, Poland.

Measurement and compensation of displacement errors by non-stop synchronized data collection

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy

MRI SYSTEM COMPONENTS Module One

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

Product Range Electronic Units

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

Review of the magnetic measurement technique (experience of the SLC, LEP, CEBAF)

Accuracy of SRS dose delivery using the TomoTherapy Hi-Art System

Proposal of test setup

Sensitivity study of an automated system for daily patient QA using EPID exit dose images

CT Scanner Dose Survey

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines

Measurement of table feed speed in modern CT

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

MR3T LASER SYSTEMS IN RT ESSENTIAL FOR PRECISE PATIENT ALIGNMENT

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

An Activity in Computed Tomography

PHYSICS QUESTIONNAIRE FORM

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Recent studies of the electron cloud-induced beam instability at the Los Alamos PSR

Chapter 3. Material and Methods

OPERATION AND MAINTENANCE MANUAL TRIAXIAL ACCELEROMETER MODEL PA-23 STOCK NO

Introducing the LumiCoil Platinum Fiducial Marker. Features of LumiCoil Platinum Fiducial Markers. Expanding Innovation in EUS

Aim. Images for this section: Page 2 of 13

3D Brachytherapy with Afterloading Machines

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

ArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools

MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed.

Relative Navigation, Timing & Data. Communications for CubeSat Clusters. Nestor Voronka, Tyrel Newton

This is a preview - click here to buy the full publication

Diagnostic X-Ray Shielding

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

Experience with Insertion Device Photon Beam Position Monitors at the APS

Transcription:

Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies

Acknowledgements Edward Vertatschitsch 3, J Nelson Wright 3, Steven Dimmer 3, Barry Friemel, PhD 3, Laurence J. Newell 3, Andrew Silber, PhD 3, Y Cheng 3, Lisa Levine, PhD 3, *Timothy Mate 1, Daniel Low 4, 1 Swedish Medical Center, Seattle WA 3 Calypso Medical Technologies, Inc., Seattle, WA 4 Washington University, St. Louis MO * Has financial interest in Calypso

Objectives Most current volumetric imaging paradigms require significant time to acquire and process information prior to making a decision Radiographic and fluoroscopic methods have shown the feasibility of localization and rapid fiducial position monitoring, although dose may be an issue for fractionated treatments A wireless electromagnetic localization is under development for positioning and tracking during radiotherapy

Implantable Sensor - Beacon TM Transponder Wireless AC electromagnetic resonant circuit No external lead wires No internal power supply Designed for permanent implantation Implanted prior to therapy Positioned in soft tissue in or near treatment target Remains inactive until energized by system console 1.85 mm x 8 mm for initial prostate application

Basic Principle of Operation Magnetic Source Coils Magnetic Sensor Coils Preamplifiers f1 f2 f3 f1 f2 f3 Excitation Response

Step 1: Implant Beacon Transponders

Step 2: Treatment Planning CT Reference coordinates of transponders with respect to isocenter are established from a treatment planning CT scan Transponders X Isocenter Z Y x T,y T,z T Relative Position (xt,yt,zt) = transponder location (x,y,z) isocenter (x,y,z)

Step 3: Localization Localization system components at the linac: 1. Wireless Transponders 2. Array 3. Console 4. Infrared Cameras 5. Tracking Station - Cameras determine array position relative to isocenter - Transponders are continuously localized relative to the array - The relative offset between expected and actual transponder positions is reported for feedback in positioning

Flintstone, Fred

Accuracy and linac compatibility Evaluate the system in the radiation therapy environment for: Accuracy of setup localization Stability over time Effect on localization accuracy during linac operation (IMRT) Effect on localization accuracy with displacement of transponders (to simulate minor deformations/volume changes)

Test Assembly Prior to linac testing, accuracy was established using a calibrated benchtop system Custom designed stand with precision mounts at fixed offsets (- 8, -4, 0, 4, 8 cm) from center and distances from the array selected by the length of precision-machined mounting posts Constructed using Ultem 1000 (GE Thermoplastics) high rigidity, low thermal expansion Machined positions and inserts ( <0.1 mm precision) Calibrated by reference alignment marks and validated by a null position transponder

Test assembly Source Array Transponder

Experimental setup Transponder

Localization experiments Comparison of actual versus predicted position at distances of 80 and 270 mm from the source array (minimum and maximum distances determined from retrospective evaluation of prostate position from clinical sites). Tests performed in air and separately in a tank of 0.9% saline solution to mimic properties of human tissue Tests of tracking rate Reporting position over multiple readings to report stability and precision at 10 Hz sampling at 20s versus 20 minutes Tracking of linear trajectories at varying speed

Stability 20 s (10 Hz) (80 mm from source array) 0.03 e rro r (c m ) 0.02 0.01 0-0.01 x axis y axis z axis 0 2 4 6 8 10 12 14 16 18 20-0.02-0.03 time (s) σ x = 0.006 mm, σ y = 0.01 mm, σ z = 0.006 mm

Stability 20 s (10 Hz) (270 mm from source array) Error (cm ). 0.5 0.4 0.3 0.2 0.1-0.1 0-0.2-0.3-0.4-0.5 0 5 10 15 20 time (s) σ x = 0.27mm, σ y = 0.36 mm, σ z = 0.48 mm x axis y axis z axis

Stability over 20 minutes Stability does not change measurably over 20 minutes 80 mm from source Change in average position from 20 s: 0-.01 mm Change in standard deviation: ~0 (10^-4) mm 270 mm from source Change in average position from 20 s:.08-0.15 mm Change in standard deviation:.001 -.06 mm

Accuracy 80 mm distance to array E rro r ( cm ). dx dy dz 0.03 0.025 0.02 0.015 0.01 0.005 0-10 -8-6 -4-2 -0.005 0 2 4 6 8 10 Offset (cm)

Accuracy 270 mm distance to array 0.3 Error (cm). 0.2 0.1 0-10 -5 0 5 10-0.1-0.2-0.3 ofset (cm) dx dy dz

Accuracy in saline Tested under conditions of maximum likely error (270 mm distance to array) in 0.9% saline (140 mm depth to transponder) Centered position Accuracy: dx 0.11 mm, dy 0.06 mm, dz 0.32 mm Precision: σ x 0.27 mm, σ y 0.36 mm, σ z 0.69 mm Offset 8 cm laterally Accuracy: dx 0.29 mm, dy 0.43 mm, dz 0.27 mm Precision: σ x 0.54 mm, σ y 0.41 mm, σ z 0.43 mm

Accuracy with 3 transponders To demonstrate lack of cross talk, a mount with three transponders was tested (separation of 1 cm between pairs of transponders) Accuracy of localization: 80 mm from source array: dx 0.17 mm, dy 0.03 mm, dz 0.05 mm σ x 0.01 mm, σ y 0.01 mm, σ z 0.01 mm 270 mm from source array: dx 0.16 mm, dy 0.18 mm, dz 0.12 mm σ x 0.26 mm, σ y 0.49 mm, σ z 0.62 mm.

Phantom design A tissue equivalent phantom was developed for accuracy studies containing external marks for laser-based alignment and a precision slot for test inserts Inserts were machined containing transponders in known configurations The configuration of transponders for a given insert generates an average coordinate with a known (nonzero) offset relative to isocenter (when the phantom is aligned to lasers using the external marks)

Accuracy Demonstration - Method Laser accuracy was established using a stereotactic isocenter standard (~0.25 mm) Target position was accuracy was defined by absolute error from baseline position established by laser alignment and phantom design ( truth ) Comparison to existing standard for localization was accomplished by diagnostic radiographic localization at UM using inroom X-Ray system Perform both a kv and Calypso isocenter localization and compare each to the measured ( absolute ) isocenter offset by CT

Static Accuracy Results Absolute Isocenter Offset: -0.70 mm lat, 4.70mm long, 0.01 mm vert Reported Position (mm) Reported & truth difference (mm) Localization Method Lat Long Vert Lat Long Vert 3-D error Radiographic Localization -1.49 5.44 0.92 0.79-0.74-0.91 1.17 Calypso Localization -0.67 4.74-0.01-0.03 0.04 0.02 0.05

System Stability in RT Environment - Beam Off Continuous localization (gantry angle 180 degrees), no radiation beam Initial positioning using manual couch adjustment guided by Calypso interface Average offsets: (-0.13,0.13,0.26) mm Variation (σ): Location (mm) 2 1.8 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0-0.2-0.4-0.6-0.8 Target Position - Beam Off Lat Long Vert (0.04,0.02,0.05) mm -1-1.2-1.4-1.6-1.8-2 0 2 4 6 8 10 Time (Minutes)

System Stability in RT Environment - Beam On Same setup process as beam off experiment Continuous monitoring during delivery of 5 field IMRT (step and shoot) treatment over 15 minutes Slight (<0.2 mm) dependence of gantry angle on average offset measurements Variation (σ): (0.8,0.6,0.6) mm Location (mm) 2.0 1.5 1.0 0.5 0.0-0.5 Target Position during IMRT 225 o gantry 290 o gantry 330 o gantry Lat Vert 20 o gantry Long 65 o gantry -1.0-1.5-2.0 0 2 4 6 8 10 12 14 16 Time (minutes)

Static Accuracy Table Shift Localizations Reported Calypso Position (mm) Calypso & Radiographic Difference (mm) 10 mm Table Shifts Lat Long Vert Lat Long Vert 3D difference (mm) At isocenter align -0.27 0.01 0.01 1.01-0.88-0.87 1.60 Posterior shift 0.01 0.16-10.00 1.05-1.07-0.81 1.70 Anterior shift -0.03 0.01 10.26 0.89-0.76-0.68 1.35 Right shift -10.39 0.00 0.15 0.62-0.82-0.39 1.10 Left shift 9.89 0.00 0.00 0.71-1.08-0.50 1.39 Inferior shift -0.36-10.10 0.14 0.41-1.49-0.29 1.57 Superior shift -0.28 10.32 0.00 1.05-0.95-0.24 1.44 When 10 mm table shifts were made (assuming a precise shift), The system could detect these shifts to within 0.4 mm, well within the uncertainty of radiographic verification (mean error 1.5 mm)

Study - Conclusions The Calypso System was able to localize the isocenter with sub-millimeter (< 1mm) accuracy After initial setup, the system was capable of performing stable continuous localization (drift <0.2 mm) Minimal localization error (<0.2mm) induced by linac operation Accuracy of tracking offsets due to table shifts up to 10 mm was within the accuracy of reference standard (radiographic) localization technique

Wash U Phantom (Low) Validation Digitizer Beacon Phantom 3X 1D Stages

Comparison Along Longitudinal Axis: Patient Motion File Low, Wash. U.

Parikh and Low Accuracy vs speed 0.9 2x4x2 cm ellipse - 20 bpm - 1 transponder 33ms 0.8 0.7 0.6 error (mm) 0.5 0.4 0.3 0.2 0.1 0 0 10 20 30 40 50 60 70 speed (mm/sec)

Initial clinical trials - prostate First stage assess marker position stability over time UC-02: Beacon Transponder Positional Stability Inter-Transponder Distance (mm) 50 40 30 20 10 0 0 25 50 75 100 125 CT images are denoted w ith an "open symbol" # Days Post-Implant A-L L-R A-R Treatment planning CT 2 weeks post implant

Initial trials - prostate Standard Deviations of Inter-Beacon Distances All Beacon Geometry Data Day 14 - Fraction 20 A-L L-R A-R Avg A-L L-R A-R Avg UC-01 1.3 1.3 0.7 1.1 1.4 1.2 0.7 1.1 MD-01 0.5 0.7 0.8 0.7 0.4 0.4 0.6 0.5 MD-02 0.6 0.9 0.5 0.7 0.4 0.4 0.4 0.4 MD-03 0.9 1.9 2.9 1.9 0.5 0.2 0.5 0.4 UM-01 0.8 1.3 1.1 1.1 0.7 1.0 1.1 0.9 MD-04 5.0 0.9 2.1 2.7 1.8 1.0 1.2 1.3 UC-02 1.2 1.7 0.9 1.3 0.7 1.6 0.6 1.0 MD-05 0.9 0.8 0.8 0.8 0.5 0.4 0.8 0.5 MD-06 0.9 1.7 2.0 1.5 1.1 1.6 0.8 1.1 UM-02 2.3 1.6 1.5 1.8 0.9 1.4 1.4 1.2 UM-03 1.5 1.0 1.3 1.3 0.7 1.2 1.2 1.0 UM-04 0.6 1.3 0.8 0.9 0.4 0.5 0.6 0.5 MD-07 0.4 0.7 0.7 0.6 0.4 0.4 0.4 0.4 MD-08 0.7 0.8 1.5 1.0 0.6 0.4 0.5 0.5 UM-05 -- 1.9 -- 1.9 -- 1.2 -- 1.2 UM-06 0.9 1.3 1.1 1.1 0.6 1.0 0.9 0.8 MD-09 1.5 0.8 1.2 1.2 0.7 0.7 1.2 0.9 UC-03 1.3 1.1 0.9 1.1 0.5 0.7 0.5 0.6 UM-07 1.0 1.8 1.4 1.4 0.5 0.9 0.5 0.7 MD-10 1.1 2.3 1.8 1.8 0.6 0.4 0.6 0.5 Mean 1.2 1.3 1.3 N/A 0.7 0.8 0.8 N/A

2 nd phase Beacon-based positioning Initial positioning using Beacons (35 seconds for an untrained operator) Radiographic verification of position Starting 2-5 minutes after initial positioning, 8 minutes of position data is collected 1 P a tie n t A lig n m e n t U s in g th e C a lyp s o S ys te m MD-08 0.5 Target Position (cm) 0-0.5-1 -1.5 Longitudinal, lateral, and vertical table shifts m ade sequentially (~35 seconds) -2 L a te r a l Longitudinal V e r tic a l 0 1 0 2 0 3 0 4 0 5 0 6 0 Time (sec)

8 Minutes of Continuous Calypso Monitoring MD-05 1 0.8 Target Position (cm) 0.6 0.4 0.2 0 Lateral X Y Z -0.2 Longitudinal -0.4 Vertical 0 100 200 300 400 500 600 Time (sec)

0.5 8 Minutes of Continuous Calypso Monitoring MD-06 0.4 Target Position (cm) 0.3 0.2 0.1 0-0.1 Longitudinal Lateral Vertical X Y Z -0.2-0.3 0 100 200 300 400 500 600 Time (sec)

8 Minutes of Continuous Calypso Monitoring 0.2 UM-04 0.1 Lateral Target Position (cm) 0-0.1-0.2-0.3-0.4 Longitudinal Vertical X Y Z -0.5-0.6 0 100 200 300 400 500 600 Time (sec)

Table 2. Maximum motion during 8 minute tracking sessions (meas in mm) Willoughby (MDA Orlando) X (Lateral) Y (Ant/Post) Z (Sup/Inf) md04 1.0 2.9 1.6 md05 1.4 9.9 13.9 md06 0.7 2.5 2.4 md08 0.4 2.0 1.7 md09 0.4 1.2 1.1 md10 0.9 1.4 1.6 um01 0.5 1.5 1.7 um02 1.0 3.2 2.4 um03 1.0 3.4 3.1 um04 1.2 2.2 2.4 um06 1.3 9.6 11.2

Summary RF localization using implanted transponders is feasible This system has shown the potential to provide rapid positioning based on transponder location Intratreatment monitoring is possible, and early studies show the potential value for detecting large transient shifts, as well as slower trends in position variation Developments are underway for other body sites